Problems in the disinfection of class 1 microbiology safety cabinets.

P H Everall, C A Morris, P R Oliver, J F Becker
PMCID: PMC497759  PMID: 7047573

Abstract

Microbiology safety cabinet disinfection procedures using formaldehyde have been tested. Tubercle bacilli were killed by concentrations of formaldehyde obtained by heating commercial formalin irrespective of whether the bacilli were in the cabinet free space or above the prefilters. However, Bacillus stearothermophilus spore papers for for the testing of low temperature steam/formaldehyde sterilisers were almost never sterilised and a strain of Staphylococcus epidermidis (NCTC 7944) showed a resistance intermediate between the B stearothermophilus spores and the tubercle bacilli. Tests using a vaccine strain of poliovirus type 3 indicated a considerable degree of resistance of the virus to the action of formaldehyde. No such resistance was demonstrated by vaccinia virus or echovirus 14. Chemical and biological evidence is presented which indicates that filter paper discs are an unsuitable carrier material for a challenge organism in testing the efficiency of any formaldehyde sterilising process. Recommendations are made towards developing a satisfactory test procedure.

Full text

PDF
698

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Everall P. H., Morris C. A. Quantitative recovery of spores from thermophilic spore papers. J Clin Pathol. 1978 May;31(5):423–425. doi: 10.1136/jcp.31.5.423. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. HIATT C. W. KINETICS OF THE INACTIVATION OF VIRUSES. Bacteriol Rev. 1964 Jun;28:150–163. doi: 10.1128/br.28.2.150-163.1964. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Marcos D., Wiseman D. Measurement of formaldehyde concentrations in a subatmospheric steam-formaldehyde autoclave. J Clin Pathol. 1979 Jun;32(6):567–575. doi: 10.1136/jcp.32.6.567. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Newsom S. W., Walsingham B. M. Sterilization of the biological safety cabinet. J Clin Pathol. 1974 Nov;27(11):921–924. doi: 10.1136/jcp.27.11.921. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES