Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2017 Nov 1.
Published in final edited form as: Pract Radiat Oncol. 2016 Feb 13;6(6):e269–e275. doi: 10.1016/j.prro.2016.02.002

Robustness Quantification Methods Comparison in Volumetric-Modulated Arc Therapy to treat Head and Neck Cancer

Wei Liu 1, Samir H Patel 1, Jiajian (Jason) Shen 1, Yanle Hu 1, Daniel P Harrington 1, Xiaoning Ding 1, Michele Y Halyard 1, Steven E Schild 1, William W Wong 1, Gary A Ezzell 1, Martin Bues 1
PMCID: PMC4983261  NIHMSID: NIHMS760269  PMID: 27025166

Abstract

Background

To compare plan robustness of volumetric-modulated arc therapy (VMAT) with intensity-modulated radiation therapy (IMRT) and to compare the effectiveness of 3 plan robustness quantification methods.

Methods

VMAT and IMRT plans were created for 9 head and neck cancer patients. For each plan, 6 new perturbed dose distributions were computed using ±3mm setup deviations along each of the 3 orientations. Worst-case analysis (WCA), dose-volume histogram (DVH) band (DVHB), and root-mean-square-dose (RMSD) volume histogram (RVH) were used to quantify plan robustness. In WCA a shaded area in the DVH plot bounded by the DVHs from the lowest and highest dose per voxel was displayed. In DVHB we displayed the envelope of all DVHs in band graphs of all the 7 dose distributions. RVH represents the relative volume on the vertical axis and the RMSD on the horizontal axis. The width from the first two methods at different target DVH indices (such as D95% and D5%) and the area under the RVH curves (AUC) for the target were used to indicate plan robustness. Results were compared using Wilcoxon signed-rank test.

Results

DVHB showed that the width at D95% of IMRT was larger than that of VMAT (unit Gy) [1.59 vs 1.18] and the width at D5% of IMRT was comparable to that of VMAT [0.59 vs 0.54]. WCA showed similar results between IMRT and VMAT plans [D95%: 3.28 vs 3.00; D5%: 1.68 vs 1.95]. RVH showed the AUC of IMRT was comparable to that of VMAT [1.13 vs 1.15]. No statistical significance was found in plan robustness between IMRT and VMAT.

Conclusions

VMAT is comparable to IMRT in terms of plan robustness. For the 3 quantification methods, WCA and DVHB are DVH-parameter dependent whereas RVH captures the overall effect of uncertainties.

Keywords: VMAT, Plan Robustness, root-mean-square dose volume histogram, head and neck

INTRODUCTION

Volumetric-modulated arc therapy (VMAT) is an advanced form of intensity-modulated radiation therapy (IMRT) that delivers a precisely sculpted 3-dimensional dose distribution using a single or multi-arc treatment. VMAT has rapidly gained popularity in treating head and neck (H&N) cancer patients due to its lower integral dose and faster delivery compared to conventional static-field IMRT (hereinafter referred to as IMRT).(15) Despite technological advances in radiation treatment of H&N cancer, local relapse still remains a significant problem due to various factors.(611) Among them treatment delivery uncertainties such as patient setup uncertainties and organ motion might lead to target underdose and therefore are considered to be important factors contributing to local relapse.(12, 13) To ensure sufficient dose coverage to the treatment target and meet the clinical requirement for normal tissues, it is critically important to evaluate sensitivity of treatment plans to uncertainties.

Plan robustness quantification is a quantitative way to evaluate treatment plan sensitivity to uncertainties. Representative plan robustness quantification methods previously discussed in the literatures include worst-case analysis (WCA),(14, 15) dose-volume histogram band (DVHB),(16, 17) and root-mean-square-dose volume histogram (RVH).(1821) The need for plan robustness quantification has been articulated for almost 3 decades;(22) but not implemented in routine clinical practice due to high computational cost. In external beam therapy planning, the influence of setup uncertainties and organ motion is addressed by adding predefined fixed margins to the clinical target volume (CTV) to form the planning target volume (PTV) in treatment planning and evaluated by the PTV dose distribution after planning.(23) The PTV concept relies on the assumption that dose cloud is static relative to the room coordinate system.(24) The validity of this assumption in photon therapy remains an active research topic. Recently some groups have proposed to use robust probabilistic planning to replace the concept of PTV in photon therapy.(2531) For H&N cancer treatment, sensitivity of IMRT plans generated using PTV margins to patient setup uncertainties and organ motion has been extensively studied (3242). However, there are few studies investigating sensitivity of VMAT plans to uncertainties (34). It is important to make sure that the superior dose distribution of VMAT can be delivered in the presence of uncertainties.

The goal of this study is to evaluate sensitivity of VMAT to patient setup uncertainties and to compare plan robustness of VMAT with IMRT for H&N cancer patients. A secondary goal of this work is to compare the performance between plan robustness quantification methods for photon therapy.

MATERIALS AND METHODS

Patient Data and Treatment Planning

We retrospectively evaluated treatment plans for 9 H&N cancer patients who were treated at our institution using VMAT. Patient and treatment characteristics for these patients are shown in Table I. Institutional review board approval was obtained for the use of this data (IRB No. XXXX). Patients were treated with radiation alone, or as part of multimodal therapy in combination with surgery with or without chemotherapy (Table I).

Table I.

Patient Characteristics

Patient Age Gender Tumor Site Treatment
With
Prescription
Dose (in Gy)
and Fractions
T
Stage
N
Stage
1 73 M Nasopharynx (involving soft palate) 3-Arc VMAT (70/33) T1 N0
2 63 F Central skull base Chemo-3-Arc VMAT (70/35) T4b N0
3 80 M Supraglottis 2 Arc VMAT (66/33) T2s N2c
4 65 F Oral tongue surgery + combined modality chemoRT (3 Arc VMAT [70/35]) T2 N2b
5 62 M Nasopharynx Concurrent chemo-RT (Arc VMAT [70/33]) T4 N2
6 69 M Base of tongue Definitive chemo-RT (3 Arc VMAT [70/35]) T2/T3 N2b
7 69 M Base of tongue Surgery + chemo-RT (3 Arc VMAT [70/35]) T2 N2b
8 50 F Base of tongue Surgery + chemo-RT (3 Arc VMAT [70/35]) T1 N1
9 78 M Oral Cavity Surgery + adjuvant RT (2 Field VMAT [60/30]) T1 N1

Abbreviations: F, female; M, male; VMAT, volumetric-modulated arc therapy; RT, Radiotherapy; Chemo-RT, Chemo-radiotherapy

As for the radiation therapy, VMAT with 2 or 3 arcs was used (Table I). All 9 patients had been prescribed at 2 dose levels administered using simultaneous integrated boost technique. The target region receiving a high prescribed dose was referred to as CTVhigh and the region receiving a low prescribed dose as CTVlow. CTVs were delineated by a physician, with CTVhigh defined as the high risk microscopic disease volume (gross tumor volume or post-operative tumor bed with non-uniform 5–10 mm margin) including the high-risk nodal volume adjacent to gross disease considered to be at risk of harboring subclinical disease. CTVlow typically encompassed a 10 to 15 mm margin beyond CTVhigh and low-risk nodal volumes. PTVhigh and PTVlow were formed by uniform expansion of the corresponding CTV by 3 mm. Doses to targets and critical normal structures (eg, brainstem, optic chiasm, optic nerves) were constrained to meet acceptable tolerance dose values whenever possible as defined in the departmental H&N cancer treatment protocol (Supplemental Table I). The dose covering a percentage of the structure’s volume (D%) and the volume of the structure receiving a certain dose (VGy) were used for dosimetric evaluation and planning purposes.

All patients were re-planned using IMRT with 7 or 9 non-opposed, equally spaced, co-planar fields. Both VMAT and IMRT plans used the same structure sets, prescription doses, and numbers of fractions as shown in Table I. IMRT plans were normalized to have the same D95% of the PTVhigh as in the VMAT plans. All VMAT and IMRT plans were generated using the Eclipse version 11 (Varian Medical Systems) by experienced dosimetrists or physicists and were approved by physicians.

Robustness Quantification Methods

Inter-fractional patient set-up uncertainties were modeled by applying both positive and negative shifts of the iso-center of the patient in the antero-posterior, superior-inferior, and lateral directions by the same margin as was used for defining the PTV (i.e., 3 mm). The original VMAT and IMRT plans were used for all the recalculations, yielding 7 (nominal plus 6 perturbed) dose distributions per plan per patient. In total 126 dose distributions were calculated. We used 3 in-house developed robustness quantification methods:

  1. WCA (14, 15): In the WCA method, the highest and the lowest dose values in each voxel from the original and the perturbed dose distributions formed a hot dose distribution with the highest values and a cold dose distribution with the lowest values. Each structure was displayed by a shaded area that was bound by the dose-volume histograms (DVHs) from the cold and hot dose distributions. The width of the DVH bands corresponded to the plan robustness for the structure indicated (Figure 1a).

  2. DVHB (16, 17): In the DVHB method, the envelope of all DVHs in band graphs of all dose distributions associated with the corresponding uncertainties was displayed. The width of the DVH bands was also used to indicate the plan robustness for the structure (Figure 1b). In the WCA and DVHB methods, the width varied if they were chosen at different DVH parameters. In this work, DVH parameters were chosen to be at D95% (indication of cold spots induced by uncertainties) and D5% (indication of hot spots induced by uncertainties).

  3. RVH (1820) (analogous to the error-bar volume histograms proposed by Albertini et al.(21)): In the RVH method, for the targets the root-mean-square dose deviation (RMSD) of voxels was calculated as the square root of the sum square of the differences between the dose calculated under the uncertainty scenarios and the nominal scenario and the mean dose of those 7 doses. This was then used to construct the RVH, which represented the relative volume (y) on the vertical axis and the RMSD (x) on the horizontal axis (Figure 1c). Similar to DVH, this meant that y% of the volume of the indicated structure had the RMSD at least × Gy. For example, the point in Fig. 1(c) as indicated by the green star showed that 55% of the volume of the indicated volume had the RMSD of at least 1.0 Gy.

Figure 1.

Figure 1

Comparison of plan robustness of the CTVhigh for patient 1 using 3 robustness quantification methods: (a) WCA, (b) DVHB, and (c) RVH.

For normal tissues, only the perturbed doses larger than the nominal dose per voxel were considered. The positive differences between the perturbed doses and nominal dose were used to calculate the RMSD and then RVH was constructed accordingly. We further simplified the RVH for both targets and normal tissues to compare the 2 plans’ robustness by calculating the area under the curve (AUC) of the RVH, which gave a single numerical value for the plan’s robustness for a given volume of interest: the smaller the AUC; the more robust the plan was for the structure between competing plans. For an ideally robust plan the RVH curve of a volume of interest should coincide with the vertical axis. Therefore the AUC of this structure would be zero.

Statistical Analysis

The width at D95% and D5%, of CTVhigh and CTVlow derived from WCA and DVHB and the AUC of the CTVhigh and CTVlow RVH were used to assess plan robustness of targets. The AUC of the spinal cord, brainstem, left parotid, right parotid, and mandible RVH was used to assess plan robustness of normal tissues.

We performed statistical comparison analyses of the results from both plans. The means of all plan robustness indices i.e., the widths of DVHs from the WCA and DVHB methods and the AUC from the RVH method were calculated. The data were compared with the non-parameter Wilcoxon signed-rank test, using JMP Pro 10 software (SAS Institute Inc.). A P value less than 0.05 was considered statistically significant.

RESULTS

Figure 1 shows the plan robustness comparison between VMAT and IMRT for patient #1 using the 3 plan robustness quantification methods: (a) WCA, (b) DVHB, and (c) RVH. From Figure 1 the width at D95% of the CTVhigh from WCA was 4.2 vs. 3.6 (IMRT vs. VMAT; unit Gy) and the width at D95% of the CTVhigh from DVHB was 1.6 vs. 0.76, which suggested that VMAT was more robust than IMRT (Figure 1a). However, the width at D5% of the CTVhigh from WCA was 1.2 vs. 2.1, and the width at D5% of the CTVhigh from DVHB was 0.34 vs. 0.38, suggesting that IMRT was more robust than VMAT (Figure 1b). From Figure 1c, RVH shows that if RMSD was less than 1.6 Gy (blue star), IMRT had smaller volumes than VMAT given the same RMSD (i.e., IMRT was more robust than VMAT). However, if RMSD was larger than 1.6 Gy, IMRT had a very slightly larger volume than VMAT given the same RMSD (i.e., VMAT was more robust than IMRT).

Figure 2 shows the comparison of robustness results for CTVhigh for 9 patients between IMRT (blue) and VMAT (red) plans: (a) the width at D95% from WCA, (b) the width at D5% from WCA, (c) the width at D95% from DVHB, (d) the width at D5% from DVHB, and (e) the AUC from the RVH curve.

Figure 2.

Figure 2

Comparison of plan robustness of the CTVhigh for 9 patients using 3 robustness quantification methods at different DVH parameters: (a) width at D95% from WCA, (b) width at D5% from WCA, (c) width at D95% from DVHB, (d) width at D5% from DVHB, (e) AUC of the CTVhigh RVH curves.

Within the same plan robustness quantification method, different conclusions would be drawn if the widths from the WCA and DVHB method were chosen at different DVH parameters. For example, the WCA method derived different results for patient 1, 3, and 4 if the results were chosen at different DVH parameters. See Figure 2a with Figure 2b. Likewise the DVHB method derived different results for patient 1, 4, and 5. See Figure 2c with Figure 2d.

Different conclusions would be made from different plan robustness quantification methods. If D95% was chosen, WCA showed consistent results with DVHB for every patient (Figure 2a vs. Figure 2c). However, the AUC of the CTVhigh RVH curve suggested opposite results compared to WCA and DVHB for patient 1, 2, 5, and 7 (comparing Figure 2e with Figure 2a and Figure 2c). If D5% was chosen, WCA indicated different results from DVHB for patient 2, 3, and 5 (Figure 2b vs. Figure 2d). The results from the AUC of the CTVhigh RVH curve compared to the results either from WCA or from DVHB also varied from patient to patient (comparing Figure 2e with Figure 2b and Figure 2d).

Figure 3 shows the comparison of the mean values of the widths from the WCA and DVHB methods and the AUC from the RVH method of the CTVhigh averaged over 9 patients between the IMRT (blue) and VMAT (red) plan: (a) widths from WCA and DVHB at D95% and D5% respectively, and (b) the AUC. P values from non-parametric Wilcoxon signed-rank tests were shown on top of columns. From all plan robustness evaluation metrics, the plan robustness of CTVhigh from VMAT was not significantly different from IMRT, i.e., the plan robustness of VMAT was comparable to IMRT with respect to CTVhigh. Similar results were also found for CTVlow (Supplemental Figure 1 and Supplemental Figure 2) except that the widths at D95% of the DVHB method from IMRT were significantly smaller than those from VMAT (P=0.0039) (Supplemental Figure 2).

Figure 3.

Figure 3

(a) Comparison of plan robustness of the CTVhigh averaged over 9 patients using (a) width at D95% from WCA and DVHB, width at D5% from WCA and DVHB and (b) AUC from RVH curves. Numbers at the top of the columns are P values.

We also compared the plan robustness of normal tissues between the IMRT (blue) and VMAT (red) plan as shown in Figure 4. For simplicity only results from the AUC of RVH of the normal tissues like (a) spinal cord, (b) brainstem, (c) left parotid, (d) right parotid, and (e) mandible were included. Compared to IMRT, VMAT resulted in more robust dose distribution for spinal cord (8 out of 9), brainstem (4 out of 9), mandible (7 out of 9), left parotid (4 out of 9), and right parotid (3 out of 9) (Figure 4).

Figure 4.

Figure 4

Comparison of plan robustness of (a) spinal cord, (b) brainstem, (c) left parotid, (d) right parotid and (e) mandible for 9 patients using the corresponding AUCs.

Patient averaged results of plan robustness comparison of the above normal tissues between the IMRT (blue) and VMAT (red) plan are shown in Figure 5. P values from non-parametric Wilcoxon signed-rank tests are shown on top of the columns. For all normal tissues, the plan robustness from VMAT was not significantly different from IMRT except mandible.

Figure 5.

Figure 5

Comparison of plan robustness of spinal cord, brainstem, left parotid, right parotid, and mandible averaged over 9 patients using the corresponding AUCs. P values are shown on the top of the columns.

DISCUSSION

In this study, we have successfully implemented 3 robustness quantification methods to assess plan robustness in photon therapy. We found that VMAT plan robustness is statistically comparable to IMRT plan robustness for both target and normal tissues. The 3 robustness quantification methods presented here have their advantages and disadvantages.

We noticed that the comparison of WCA and DVHB depended on the selected DVH parameters, while the AUC of the RVH did not rely on DVH parameters. Therefore RVH provides a more comprehensive indicator of the plan robustness for the structure of interest with a simple numerical value (i.e., AUC). In clinical practice, it is preferable to have a single numerical value to indicate plan robustness because it can be easily incorporated into statistical analysis for comparison. In cases where the RVH curves from the competing plans intersect with each other and have similar AUCs, physicians and physicists need to examine the entire RVH curves closely and make a patient-specific judgment: either a larger volume with a small RMSD or a smaller volume with a large RMSD. This might also explain why plan robustness comparison results from WCA and DVHB depend on the DVH parameters. The RVH plot captures the overall effect of uncertainties on the dose to a volume of interest, which is analogous to the DVH for assessing the nominal dose.

More importantly, RVH opens a door to explicitly control plan robustness in optimization. The desired robustness may be specified in terms of a RMSD-volume constraint on the targets. Penalty parameters of RMSD-volume constraints may be adjusted to explicitly control the tradeoff between robustness and plan optimality.

It would be better to have more dose distributions included in all 3 robustness quantification methods. However, it may not be practical to do so in routine clinical practice using a commercial treatment planning system, which makes the computation time formidable. In order to show the influence of more uncertainty scenarios, we tried 14 uncertainty scenarios as defined in Casiraghi et al (43) in a couple of patients and found that the conclusions were still valid using those dose distributions. It is also important to evaluate the effect of rotational perturbation on plan robustness; this will be included in a future study.

Plan robustness represents our confidence level that the nominal plan can be delivered without significant quality degradation in the presence of uncertainties. It has been extensively investigated in proton therapy because proton therapy, especially scanning beam proton therapy, is very sensitive to various uncertainties (29, 24). In photon therapy, integrating plan robustness into plan quality check is essential because it provides an indication of real-life plan quality, instead of computed plan quality. An important clinical application of plan robustness quantification is regarding selection of appropriate treatment modalities (photon vs. proton) for individual patients, which should be based on both plan quality and plan robustness. This remains an important research topic that requires further investigation and will be included in our future studies.

In summary, VMAT is comparable to IMRT in terms of plan robustness. Among the 3 robustness quantification methods, WCA and DVHB are DVH-parameter dependent whereas RVH captures the overall effect of uncertainties. Given lower integral dose and higher delivery efficiency, VMAT seems to be the preferred choice over IMRT at least for the H&N patients.

Supplementary Material

Acknowledgments

This research was supported by the National Cancer Institute (NCI) Career Developmental Award K25CA168984, the Fraternal Order of Eagles Cancer Research Fund Career Development Award, the Lawrence W. and Marilyn W. Matteson Fund for Cancer Research, Mayo Arizona State University Seed Grant, and the Kemper Marley Foundation.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Conflicts of Interest Notification

None.

REFERENCES

  • 1.Vanetti E, Clivio A, Nicolini G, et al. Volumetric Arc modulated Radiotherapy for carcinomas of the Oropharynx, Hypopharynx and Larynx. A Treatment Planning Dosimetric Comparison with Fixed Field IMRT. Clinical Oncology. 2009;21:246–246. [Google Scholar]
  • 2.Verbakel WFAR, Cuijpers JP, Hoffmans D, et al. VOLUMETRIC INTENSITYMODULATED ARC THERAPY VS. CONVENTIONAL IMRT IN HEAD-AND-NECK CANCER: A COMPARATIVE PLANNING AND DOSIMETRIC STUDY. International Journal of Radiation Oncology Biology Physics. 2009;74:252–259. doi: 10.1016/j.ijrobp.2008.12.033. [DOI] [PubMed] [Google Scholar]
  • 3.Rao M, Yang W, Chen F, et al. Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: Plan quality, delivery efficiency and accuracy. Medical Physics. 2010;37:1350–1359. doi: 10.1118/1.3326965. [DOI] [PubMed] [Google Scholar]
  • 4.Bertelsen A, Hansen CR, Johansen J, et al. Single Arc Volumetric Modulated Arc Therapy of head and neck cancer. Radiotherapy and Oncology. 2010;95:142–148. doi: 10.1016/j.radonc.2010.01.011. [DOI] [PubMed] [Google Scholar]
  • 5.Gomez-Millan Barrachina J, Jerez Sainz I, Perez Rozos A, et al. Potential advantages of volumetric arc therapy in head and neck cancer. Head & neck. 2015;37:909–914. doi: 10.1002/hed.23685. [DOI] [PubMed] [Google Scholar]
  • 6.Fried DV, Zanation AM, Huang B, et al. Patterns of local failure for sinonasal malignancies. Practical Radiation Oncology. 2013;3:e113–e120. doi: 10.1016/j.prro.2012.07.001. [DOI] [PubMed] [Google Scholar]
  • 7.Forastiere AA, Goepfert H, Maor M, et al. Concurrent chemotherapy and radiotherapy for organ preservation in advanced laryngeal cancer. New England Journal of Medicine. 2003;349:2091–2098. doi: 10.1056/NEJMoa031317. [DOI] [PubMed] [Google Scholar]
  • 8.Overgaard J, Hansen HS, Specht L, et al. Five compared with six fractions per week of conventional radiotherapy of squamous-cell carcinoma of head and neck: DAHANCA 6&7 randomised controlled trial. Lancet. 2003;362:933–940. doi: 10.1016/s0140-6736(03)14361-9. [DOI] [PubMed] [Google Scholar]
  • 9.Bernier J, Domenge C, Ozsahin M, et al. Postoperative irradiation with or without concomitant chemotherapy for locally advanced head and neck cancer. New England Journal of Medicine. 2004;350:1945–1952. doi: 10.1056/NEJMoa032641. [DOI] [PubMed] [Google Scholar]
  • 10.Lefebvre JL, Rolland F, Tesselaar M, et al. Phase 3 Randomized Trial on Larynx Preservation Comparing Sequential vs Alternating Chemotherapy and Radiotherapy. Journal of the National Cancer Institute. 2009;101:142–152. doi: 10.1093/jnci/djn460. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Spector JG, Sessions DG, Haughey BH, et al. Delayed regional metastases, distant metastases, and second primary malignancies in squamous cell carcinomas of the larynx and hypopharynx. Laryngoscope. 2001;111:1079–1087. doi: 10.1097/00005537-200106000-00028. [DOI] [PubMed] [Google Scholar]
  • 12.Mueller BS, Duma MN, Kampfer S, et al. Impact of interfractional changes in head and neck cancer patients on the delivered dose in intensity modulated radiotherapy with protons and photons. Physica Medica-European Journal of Medical Physics. 2015;31:266–272. doi: 10.1016/j.ejmp.2015.02.007. [DOI] [PubMed] [Google Scholar]
  • 13.Hong TS, Tome WA, Chappell RJ, et al. The impact of daily setup variations on head-and-neck intensity-modulated radiation therapy. International Journal of Radiation Oncology Biology Physics. 2005;61:779–788. doi: 10.1016/j.ijrobp.2004.07.696. [DOI] [PubMed] [Google Scholar]
  • 14.Lomax AJ, Pedroni E, Rutz H, et al. The clinical potential of intensity modulated proton therapy. Z Med Phys. 2004;14:147–152. doi: 10.1078/0939-3889-00217. [DOI] [PubMed] [Google Scholar]
  • 15.Quan M, Liu W, Wu R, et al. Preliminary evaluation of multi-field and single-field optimization for the treatment planning of spot-scanning proton therapy of head and neck cancer. Med. Phys. 2013;40:081709. doi: 10.1118/1.4813900. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Liu W, Zhang X, Li Y, et al. Robust optimization in intensity-modulated proton therapy. Med. Phys. 2012;39:1079–1091. doi: 10.1118/1.3679340. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Trofimov A, Kang J, Unkelbach J, et al. Evaluation of dosimetric gain and uncertainties in proton therapy delivery with scanned pencil beam in treatment of base-of-skull and spinal Tumors. International Journal of Radiation Oncology Biology Physics. 2010;78:S133–S134. [Google Scholar]
  • 18.Liu W, Frank SJ, Li X, et al. Effectiveness of Robust Optimization in Intensity-Modulated Proton Therapy Planning for Head and Neck Cancers. Med. Phys. 2013;40:051711–051718. doi: 10.1118/1.4801899. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Liu W, Liao Z, Schild SE, et al. Impact of respiratory motion on worst-case scenario optimized intensity modulated proton therapy for lung cancers. Practical Radiation Oncology. 2015;5:e77–e86. doi: 10.1016/j.prro.2014.08.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Liu W, Mohan R, Park P, et al. Dosimetric benefits of robust treatment planning for intensity modulated proton therapy for base-of-skull cancers. Practical Radiation Oncology. 2014;4:384–391. doi: 10.1016/j.prro.2013.12.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Albertini F, Hug EB, Lomax AJ. Is it necessary to plan with safety margins for actively scanned proton therapy? PHYSICS IN MEDICINE AND BIOLOGY. 2011;56:4399–4413. doi: 10.1088/0031-9155/56/14/011. [DOI] [PubMed] [Google Scholar]
  • 22.Goitein M. CALCULATION OF THE UNCERTAINTY IN THE DOSE DELIVERED DURING RADIATION-THERAPY. Medical Physics. 1985;12:608–612. doi: 10.1118/1.595762. [DOI] [PubMed] [Google Scholar]
  • 23.van Herk M, Remeijer P, Rasch C, et al. The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. International Journal of Radiation Oncology*Biology*Physics. 2000;47:1121–1135. doi: 10.1016/s0360-3016(00)00518-6. [DOI] [PubMed] [Google Scholar]
  • 24.Schwarz M. Treatment planning in proton therapy. European Physical Journal Plus. 2011;126:67–76. [Google Scholar]
  • 25.Bohoslavsky R, Witte MG, Janssen TM, et al. Probabilistic objective functions for margin-less IMRT planning. Physics in Medicine and Biology. 2013;58:3563–3580. doi: 10.1088/0031-9155/58/11/3563. [DOI] [PubMed] [Google Scholar]
  • 26.Gordon JJ, Sayah N, Weiss E, et al. Coverage optimized planning: Probabilistic treatment planning based on dose coverage histogram criteria. Medical Physics. 2010;37:550–563. doi: 10.1118/1.3273063. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Xu HJ, Gordon JJ, Siebers JV. Coverage-based treatment planning to accommodate delineation uncertainties in prostate cancer treatment. Medical Physics. 2015;42:5435–5443. doi: 10.1118/1.4928490. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Xua HJ, Vile DJ, Sharma M, et al. Coverage-based treatment planning to accommodate deformable organ variations in prostate cancer treatment. Medical Physics. 2014;41 doi: 10.1118/1.4894701. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Fredriksson A, Forsgren A, Hardemark B. Maximizing the probability of satisfying the clinical goals in radiation therapy treatment planning under setup uncertainty. Medical Physics. 2015;42:3992–3999. doi: 10.1118/1.4921998. [DOI] [PubMed] [Google Scholar]
  • 30.Baum C, Alber M, Birkner M, et al. Robust treatment planning for intensity modulated radiotherapy of prostate cancer based on coverage probabilities. Radiotherapy and Oncology. 2006;78:27–35. doi: 10.1016/j.radonc.2005.09.005. [DOI] [PubMed] [Google Scholar]
  • 31.Zhang P, Yorke E, Hu Y-C, et al. Predictive Treatment Management: Incorporating a Predictive Tumor Response Model Into Robust Prospective Treatment Planning for Non-Small Cell Lung Cancer. International Journal of Radiation Oncology Biology Physics. 2014;88:446–452. doi: 10.1016/j.ijrobp.2013.10.038. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Astreinidou E, Bel A, Raaijmakers CPJ, et al. Adequate margins for random setup uncertainties in head-and-neck IMRT. International Journal of Radiation Oncology*Biology*Physics. 2005;61:938–944. doi: 10.1016/j.ijrobp.2004.11.016. [DOI] [PubMed] [Google Scholar]
  • 33.Maleike D, Unkelbach J, Oelfke U. Simulation and visualization of dose uncertainties due to interfractional organ motion. Physics in Medicine and Biology. 2006;51:2237–2252. doi: 10.1088/0031-9155/51/9/009. [DOI] [PubMed] [Google Scholar]
  • 34.Stroom J, Vieira S, Mateus D, et al. On the robustness of VMAT-SABR treatment plans against isocentre positioning uncertainties. Radiation Oncology. 2014;9 doi: 10.1186/1748-717X-9-196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Siebers JV, Keall PJ, Wu Q, et al. Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans. International Journal of Radiation Oncology*Biology*Physics. 2005;63:422–433. doi: 10.1016/j.ijrobp.2005.02.029. [DOI] [PubMed] [Google Scholar]
  • 36.Bortfeld T. IMRT: a review and preview. Physics in Medicine and Biology. 2006;51:R363–R379. doi: 10.1088/0031-9155/51/13/R21. [DOI] [PubMed] [Google Scholar]
  • 37.Manning MA, Wu Q, Cardinale RM, et al. The effect of setup uncertainty on normal tissue sparing with IMRT for head-and-neck cancer. International Journal of Radiation Oncology*Biology*Physics. 2001;51:1400–1409. doi: 10.1016/s0360-3016(01)01740-0. [DOI] [PubMed] [Google Scholar]
  • 38.Samuelsson A, Mercke C, Johansson K-A. Systematic set-up errors for IMRT in the head and neck region: effect on dose distribution. Radiotherapy and Oncology. 2003;66:303–311. doi: 10.1016/s0167-8140(02)00389-4. [DOI] [PubMed] [Google Scholar]
  • 39.Hong TS, Tomé WA, Chappell RJ, et al. The impact of daily setup variations on head-and-neck intensity-modulated radiation therapy. International Journal of Radiation Oncology*Biology*Physics. 2005;61:779–788. doi: 10.1016/j.ijrobp.2004.07.696. [DOI] [PubMed] [Google Scholar]
  • 40.Suzuki M, Nishimura Y, Nakamatsu K, et al. Analysis of interfractional set-up errors and intrafractional organ motions during IMRT for head and neck tumors to define an appropriate planning target volume (PTV)- and planning organs at risk volume (PRV)-margins. Radiotherapy and Oncology. 2006;78:283–290. doi: 10.1016/j.radonc.2006.03.006. [DOI] [PubMed] [Google Scholar]
  • 41.O'Daniel JC, Garden AS, Schwartz DL, et al. Parotid Gland Dose in Intensity-Modulated Radiotherapy for Head and Neck Cancer: Is What You Plan What You Get? International Journal of Radiation Oncology*Biology*Physics. 2007;69:1290–1296. doi: 10.1016/j.ijrobp.2007.07.2345. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Zhang L, Garden AS, Lo J, et al. Multiple regions-of-interest analysis of setup uncertainties for head-and-neck cancer radiotherapy. International Journal of Radiation Oncology*Biology*Physics. 2006;64:1559–1569. doi: 10.1016/j.ijrobp.2005.12.023. [DOI] [PubMed] [Google Scholar]
  • 43.Casiraghi M, Albertini F, Lomax AJ. Advantages and limitations of the ‘worst case scenario’ approach in IMPT treatment planning. Physics in Medicine and Biology. 2013;58:1323. doi: 10.1088/0031-9155/58/5/1323. [DOI] [PubMed] [Google Scholar]
  • 44.Park PC, Cheung JP, Zhu XR, et al. Statistical Assessment of Proton Treatment Plans Under Setup and Range Uncertainties. International Journal of Radiation Oncology*Biology*Physics. 2013;86:1007–1013. doi: 10.1016/j.ijrobp.2013.04.009. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

RESOURCES