Abstract
A Bactometer model 32 was evaluated for use in early detection of bacterial growth. Experiments with simulated cultures showed that 2 ml of broth introduced into the Bactometer module wells could detect 10(2) and 10(6) CFU/ml in 6 h and 2 h respectively. Both Brain Heart Infusion (BHI) and Fastidious Anaerobic broths supported good growth. Detection of nine of 10 organisms inoculated at approximately 10(6) CFU/ml in BHI were detected within 8.5 h. A culture of Bacteroides fragilis failed to grow under these conditions. Of 189 blood cultures, tested by incubation of 2 ml of BHI, 18 were positive by both conventional and Bactometer methods. False-positive or false-negative specimens were not observed using the Bactometer. Use of the Bactometer enables growth detection at least 12 h earlier than culture methods.
Full text
PDF





Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Cady P., Dufour S. W., Lawless P., Nunke B., Kraeger S. J. Impedimetric screening for bacteriuria. J Clin Microbiol. 1978 Mar;7(3):273–278. doi: 10.1128/jcm.7.3.273-278.1978. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cady P., Dufour S. W., Shaw J., Kraeger S. J. Electrical impedance measurements: rapid method for detecting and monitoring microorganisms. J Clin Microbiol. 1978 Mar;7(3):265–272. doi: 10.1128/jcm.7.3.265-272.1978. [DOI] [PMC free article] [PubMed] [Google Scholar]
- DeBlanc H. J., Jr, DeLand F., Wagner H. N., Jr Automated radiometric detection of bacteria in 2,967 blood cultures. Appl Microbiol. 1971 Nov;22(5):846–849. doi: 10.1128/am.22.5.846-849.1971. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Martins S. B., Selby M. J. Evaluation of a rapid method for the quantitative estimation of coliforms in meat by impedimetric procedures. Appl Environ Microbiol. 1980 Mar;39(3):518–524. doi: 10.1128/aem.39.3.518-524.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Silverman M. P., Munoz E. F. Automated electrical impedance technique for rapid enumeration of fecal coliforms in effluents from sewage treatment plants. Appl Environ Microbiol. 1979 Mar;37(3):521–526. doi: 10.1128/aem.37.3.521-526.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Specter S., Throm R., Strauss R., Friedman H. Rapid detection of bacterial growth in blood samples by a continuous-monitoring electrical impedance apparatus. J Clin Microbiol. 1977 Nov;6(5):489–493. doi: 10.1128/jcm.6.5.489-493.1977. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zafari Y., Martin W. J. Comparison of the BACTOMETER microbial monitoring system with conventional methods for detection of microorganisms in urine specimens. J Clin Microbiol. 1977 May;5(5):545–547. doi: 10.1128/jcm.5.5.545-547.1977. [DOI] [PMC free article] [PubMed] [Google Scholar]