Abstract
Objective:
Evaluation of potential dose savings by implementing adaptive statistical iterative reconstruction (ASiR) on a gemstone-based scintillator in a clinical 64-row whole-body CT (WBCT) protocol after multiple trauma.
Methods:
Dose reports of 152 WBCT scans were analysed for two 64-row multidetector CT scanners (Scanners A and B); the main scanning parameters were kept constant. ASiR and a gemstone-based scintillator were used in Scanner B, and the noise index was adjusted (head: 5.2 vs 6.0; thorax/abdomen: 29.0 vs 46.0). The scan length, CT dose index (CTDI) and dose–length product (DLP) were analysed. The estimated mean effective dose was calculated using normalized conversion factors. Student's t-test was used for statistics.
Results:
Both the mean CTDI (mGy) (Scanner A: 53.8 ± 2.0, 10.3 ± 2.5, 14.4 ± 3.7; Scanner B: 48.7 ± 2.2, 7.1 ± 2.3, 9.1 ± 3.6; p < 0.001, respectively) and the mean DLP (mGy cm) (Scanner A: 1318.9 ± 167.8, 509.3 ± 134.7, 848.8 ± 254.0; Scanner B: 1190.6 ± 172.6, 354.6 ± 128.3, 561.0 ± 246.7; p < 0.001, respectively) for the head, thorax and abdomen were significantly reduced with Scanner B. There was no relevant difference in scan length. The total mean effective dose (mSv) was significantly decreased with Scanner B (24.4 ± 6.0, 17.2 ± 5.8; p < 0.001).
Conclusion:
The implementation of ASiR and a gemstone-based scintillator allows for significant dose savings in a clinical WBCT protocol.
Advances in knowledge:
Recent technical developments can significantly reduce radiation dose of WBCT in multiple trauma. Dose reductions of 10–34% can be achieved.
INTRODUCTION
Since its introduction, multidetector CT (MDCT) use has dramatically risen as the first-line imaging tool in emergency radiology, mostly because of its wide availability, short scan time, duration, robust technique, and high spatial and temporal resolution due to isotropic voxels. The number of CT scans has exceeded 70 million per year in the USA in recent years.1 Although the emergency department patient volumes grew in the USA at about 13% between 2000 and 2005, the number of CT examinations had increased from 51% to 463%, depending on the institution and the relevant body region.2 MDCT is widely accepted as the whole-body imaging procedure in multiple trauma and is indisputably the imaging modality of choice.3,4 A detailed analysis in a clinical outcome study has proven that standardized whole-body CT (WBCT) clearly improves the survival rate after severe trauma.5 Despite this strong benefit, the trauma patient is still exposed to a considerable radiation dose of about 10–20 mSv, with some published reports of >30 mSv for a single WBCT scan.5
Concerns about radiation dose and its potentially harmful and carcinogenic effects are being widely discussed.6–8 There is evidence that cumulative doses to an organ of 30–90 mSv or more can increase the risk of developing cancer.9 Consequently, several technical approaches to dose reduction have already been established and implemented successfully in clinical practice, such as tube current modulation and automatic exposure control.10–12
High computational power now also offers improved techniques of image reconstruction. Filtered back projection (FBP) has been the standard mathematical reconstruction algorithm since the beginning of CT. Its main advantage is the fast calculation, even of CT images, which is essential for routine CT scanning; the main disadvantage is a limited potential of dose reduction. There is a well-known trade-off between radiation dose and image noise: the lower the dose, the higher the image noise, which is considered as deterioration of image quality. This was an indisputable fact in clinical practice for a long time.
Recently, iterative image reconstructions have become available, such as adaptive statistical iterative reconstruction (ASiR). The main advantage of these algorithms is that they assume a far more realistic data model than conventional FBP to reduce radiation dose by decreasing image noise.13 Despite its higher mathematical complexity, the use of ASiR does not result in longer reconstruction times than FBP.14 Therefore, it is also suited for emergency conditions when time is of the essence.
The purpose of this retrospective controlled and blinded clinical study was to evaluate the potential dose savings by combining ASiR and a new gemstone-based scintillator, in a clinical 64-row WBCT protocol of patients sustaining major and multiple trauma.
METHODS AND MATERIALS
Patients
The institutional review board has waived the need for informed consent due to the retrospective nature of the study. All patients who fulfilled the clinical criteria of multiple trauma and underwent a WBCT examination according to an advanced routine CT protocol between December 2008 and October 2010 were included. The indication of WBCT is based on widely accepted modified Advanced Trauma Life Support® criteria referring to vital signs, injury pattern and mechanism of trauma (Table 1).3 Patients who underwent a different CT protocol based on specific clinical indication or had an incomplete dose report were excluded.
Table 1.
Established clinical indication rule for whole-body CT
| Impaired vital signs |
|---|
| Pattern of injury |
| Open thoracic injury |
| Instable thorax |
| Instable pelvic fracture |
| Fractures of at least two major long bones |
| Macroamputation |
| Mechanism of trauma |
| Explosion |
| Entrapment or entombment |
| Fall >3 m height |
| Death of the codriver |
| Rollover of the vehicle |
| Ejection out of the vehicle |
| Accident with (motor-)cyclist, pedestrian |
| High-velocity accident |
CT technique
The WBCT examinations were performed on two different 64-row MDCT scanners from one manufacturer. Scanner A (LightSpeed® VCT® XT; GE Healthcare, Waukesha, WI) was replaced by Scanner B (Discovery® CT750 HD; GE Healthcare) in November 2009. Scanner A was only capable of FBP, whereas Scanner B was capable of ASiR for image reconstruction also using a new CT detector. Both scanners were otherwise technically comparable, and the use of Scanner B was an upgrade on the platform of Scanner A using the same technique of tube current modulation in both xy-plane (Auto mA) and the z-direction (Smart mA). However, there was also a different detector in Scanner B; the gemstone-based scintillator has a low afterglow and a primary decay of about 0.03 µs, which is reported to be 100 times faster than conventional scintillator material.15 The newly introduced dedicated high-definition mode on Scanner B with 2.5 more projections per rotation was not used for this study to keep further parameters constant.
Whole-body CT protocol
Both protocols used have been the routine examination standard for each particular CT device at our emergency radiology department and have not been altered for study purposes.
The main scanning parameters were kept constant for all CT examinations (Table 2). The head, which also includes the mid-face, was scanned in helical mode without gantry tilt (Phase 1, non-contrast). The helical MDCT scans of the thorax and abdomen were conducted separately from each other. The cervical spine was included in the chest volume during the arterial contrast medium phase (Phase 2, arterial); the abdomen, including the pelvis, was examined during the portovenous phase (Phase 3, portovenous). The arms were placed crosswise on the trunk during the examination in order to reduce artefacts, following a long-established institutional protocol. Prior to this study, in order to make the results of both scanners comparable, the protocol for the CT scanner replacing the first one was modified with the help of the manufacturer to preserve the same constant clinical image quality and allow for a objective comparison: based on prior experience, noise index was adapted and increased from 5.2 to 6.0 (head) and from 29.0 to 46.0 (thorax and abdomen). The ASiR level was set at 30% (head) and 50% (thorax and abdomen).
Table 2.
CT scan parameters for two whole-body CT scan protocols each of 64-row multidetector CT (VCT 64 and HD 750; GE Healthcare, Waukesha, WI)
| CT parameters | |
|---|---|
| Two plane scout | |
| Tube voltage | 120 kV |
| Tube current | 40 mA (anteroposterior), 80 mA (lateral) |
| Head | |
| Time of rotation | 0.5 s |
| Collimation | 64 × 0.625 mm |
| Slice thickness | 2.5 mm |
| Tilt | 0° |
| Tube voltage | 120 kV |
| Tube current, mAs product | 100–320 mA, modulated |
| Table feed/rotation | 19.4 mm |
| Contrast medium injection | |
| Volume | 140 ml |
| Saline chaser | 40 ml |
| Flow rate | 3.5 ml s−1 |
| Bolus tracking (monitoring) | Thoracic aorta |
| Cervical spine/thorax | |
| Rotation time | 0.5 s |
| Collimation | 64 × 0.625 mm |
| Slice thickness | 1.25 mm |
| Tube voltage | 120 kV |
| Tube current | 100–700 mA, modulated |
| Table feed | 55 mm |
| Delay | 3 s, after monitoring |
| Abdomen/pelvis | |
| Rotation time | 0.5 s |
| Collimation | 64 × 0.625 mm |
| Slice thickness | 1.25 mm |
| Tube voltage | 120 kV |
| Tube current | 100–700 mA, modulated |
| Table feed | 55 mm |
| Delay | ca. 50 s after thorax |
Image reconstruction
FBP has been used as a standard reconstruction method for almost 30 years. Owing to the limited processing capabilities, CT images were reconstructed from measured attenuation profiles by using the analytic Radon transformation. By contrast, modern CT systems provide an impressive improved processing power for statistical reconstruction algorithms. The ASiR generates CT images by taking photon statistics into account in addition to measured attenuation data. Consecutively, the mean image noise decreases correlating to the applied ASiR level from 0% to 100%. Two strategies could now be used: improving image quality or, as desired by our study concept, reducing the radiation dose by maintaining the accustomed image quality at the same time. We refer to previous publications for a detailed description of the ASiR algorithm.16,17
Radiation dose
The scan length (mm), CT dose index (CTDI, mGy) and dose–length product (DLP, mGy cm) of the head, thorax and abdomen were retrieved and calculated on the basis of the dose reports. The calculation of the mean effective dose (mSv) was carried out by conversion of the DLP with the normalization coefficient corresponding to the European guidelines on quality criteria for CT (Table 3).19 As radiation dose is dependent on the patient's constitution among other things, the coronal and sagittal diameters were measured at the level of the first vertebral body of the lumbar spine.
Table 3.
Effective dose per dose–length product (DLP); normalized values of over various body regions19
| Body region | Normalized effective dose, EDLP (mSv mGy−1 cm−1) |
|---|---|
| Head | 0.0023 |
| Thorax | 0.017 |
| Abdomen | 0.015 |
Image quality
As various instruments, for example, rib spreaders, chest tubes, electrocardiography and monitoring cables, are present depending on the clinical situation, the occurrence of artefacts is very variable during an emergency CT examination. Therefore, we did not perform a detailed assessment of image quality. However, we distinguished between diagnostic and non-diagnostic overall image quality, the latter of which was defined indication for a repeat CT scan without specific medical indication.
Statistical analysis
Statistical analysis was performed using SPSS® v. 18 (IBM Corp., New York, NY; formerly SPSS Inc., Chicago, IL). An a priori estimation of the sample size was conducted resulting in a minimum need of 38 patients (one-sided, alpha error 0.025, power 0.95). Significant differences in age distribution, patients' diameters, dose parameters and mean effective dose were tested by the student's t-test. The gender distribution was analyzed by the χ2test. The significance level was p < 0.05.
RESULTS
The data from 76 patients (54 males, 22 females; mean age 44 ± 19 years) who were examined on Scanner A were compared with the data of 76 patients (49 males, 27 females; mean age 46 ± 19 years) who were examined on Scanner B. There was no significant difference between both the groups in terms of gender (p = 0.386), age distribution (p = 0.510), sagittal diameter (23.2 ± 4.6 vs 23.6 ± 3.8 cm; p = 0.580) and coronal diameter (32.4 ± 4.3 vs 32.7 ± 3.4 cm; p = 0.580).
There was no significant difference in scan length in all examined body regions, comparing Scanners A and B, which would have had influence on the following results.
Table 4 gives an overview of the dose characteristics. The mean CTDI and mean DLP for the head, thorax and abdomen were statistically significantly lower for all regions using Scanner B including ASiR than using Scanner A without ASiR. The largest differences in CTDI (mGy) were for the thorax (reduction of 31.1%; 10.3 ± 2.5 vs 7.1 ± 2.3) and abdomen (reduction of 36.8%; 14.4 ± 3.7 vs 9.1 ± 3.6), whereas in the head section (reduction of 9.5%; 53.8 ± 2.0 vs 48.7 ± 2.2), dose reduction effects were only minor. Considering the CTDI as an indicator for the radiation dose applied per slice, without taking into account the scan length, there were dose savings of 9.5–36.8%, depending on the anatomical region (Table 4).
Table 4.
Results concerning radiation dose parameters: scan length, CT dose index (CTDI) and dose–length product (DLP); mean values ± standard deviation
| CT parameters | Head | Thorax | Abdomen |
|---|---|---|---|
| Scan length (mm) | |||
| Scanner A | 225.3 ± 30.8 | 430.1 ± 40.7 | 520.5 ± 51.2 |
| Scanner B | 221.6 ± 33.3 | 434.1 ± 52.0 | 540.6 ± 87.0 |
| CTDI (mGy) | |||
| Scanner A | 53.8 ± 2.0 | 10.3 ± 2.5 | 14.4 ± 3.7 |
| Scanner B | 48.7 ± 2.2a | 7.1 ± 2.3a | 9.1 ± 3.6a |
| DLP (mGy cm) | |||
| Scanner A | 1318.9 ± 167.8 | 509.3 ± 134.7 | 848.8 ± 254.0 |
| Scanner B | 1190.6 ± 172.6a | 354.6 ± 128.3a | 561.0 ± 246.7a |
p < 0.001.
Again, the largest differences with DLP (mGy cm) were found in the thorax (reduction of 30.0%; 509.3 ± 134.7 vs 354.6 ± 128.3) and abdomen (reduction of 33.8%; 848.8 ± 254.0 vs 561.0 ± 246.7); however, the differences in the head section were only small (reduction of 9.7%; 1318.9 ± 167.8 vs 1190.6 ± 172.6).
The resulting estimated effective dose was reduced by about 10% in the head, 30% in the thorax and 34% in the abdomen. The mean effective dose calculated for the entire WBCT examination was reduced by about 30% from 24.4 mSv with Scanner A down to 17.2 mSv with Scanner B (Table 5).
Table 5.
Mean estimated effective dose ± standard deviation (mSv)
| Radiation dose | Head | Thorax | Abdomen | Total |
|---|---|---|---|---|
| Scanner A | 3.0 ± 0.4 | 8.7 ± 2.3 | 12.7 ± 3.8 | 24.4 ± 6.0 |
| Scanner B | 2.7 ± 0.4a | 6.0 ± 2.2a | 8.4 ± 3.7a | 17.2 ± 5.8a |
p < 0.001.
DISCUSSION
MDCT is considered as the modality of choice in the imaging of patients with polytrauma.20 Despite its proven effect on survival rate,5 the radiation dose applied is not only a major concern only for radiologists but also widely discussed with many alert clinical partners. The popular media in many countries is also focusing on radiation dose issues and the possible medical risks for the patients.
As the use of MDCT has also risen in other radiological subspecialties such as cardiac imaging, recent developments have focused on strategies of dose reduction, including tube current modulation and automatic exposure control.10,11,18 The main limitation of dose reduction so far has been that by lowering the dose, the resulting higher noise level has impaired the image quality. Modern highly capable CT systems offer dose savings using iterative image reconstructions, such as ASiR™ (GE Healthcare), iDose4 (Philips Healthcare) and IRIS (Siemens Healthcare). According to Silva et al,17 ASiR allows for a significant reduction in radiation dose without losing image quality compared with FBP. The study by Hara et al16 showed a significant reduction in the CTDI of 32–65% by using ASiR in 12 abdominal CT scans. Two studies by Prakash et al21,22 showed a significant dose reduction of 27.6% for thorax scan and about 25.1% for the abdomen scan. In our study, the dose reduction has been evaluated for a dedicated WBCT protocol in patients with multiple injuries.
Using a trauma protocol with ASiR, the radiation dose decreased significantly from 24.4 to 17.2 mSv, which means a dose reduction of 30% for the entire CT examination. The mean effective dose of WBCT in the literature is estimated between 10 and 30 mSv.5,23 The different scan parameters and different patient positions applied make an important contribution. Ptak et al24 and Fanucci et al25 have shown that a monophasic examination provides a significant dose reduction and shorter duration of the scan than a segmented scan protocol owing to missing overlap. Neither of these studies were conducted with 64-row MDCT scanners. Moreover, 64-row MDCT has markedly reduced the duration of the scan compared with former MDCT scanners.26 The scan protocols published have been heterogeneous in terms of scan parameters, which have been adjusted to the institutional needs of image quality and patients' characteristics. However, comparable scan protocols resulted in a drop of the mean effective dose of 21.2–24.69 mSv,27,28 which is similar to our result of 24.4 mSv with FBP.
Our results could show in detail that there was a decrease for each anatomical body region. Although the dose reduction of the head CT was only 10%, there was a marked decrease of about 30–34% for the chest and abdominal CT, respectively. The different ASiR levels of 30% for head CT and 50% for thorax and abdomen CT, respectively, might explain these different capabilities of dose reduction. Moreover, the potential dose savings are limited by the higher noise caused by the skull, which has to be penetrated by photons in head CT. Although the scan length influences the DLP with the number of rotations, the CTDI is considered a more accurate parameter correlating with the applied radiation dose per standardized slice thickness. A detailed analysis showed that the dose reduction per slice was at a maximum of 36.8% for the abdomen and 31.1% for the thorax, and lower dose savings of 9.5% in the head CT section.
In addition to the patients' dose exposure, there are medicolegal aspects that have to be taken into account. The European Guidelines, as well as national agencies such as the Federal Office of Radiation Protection in Germany, regulate the application of radiation by publishing dose reference levels.19 In view of this background, all radiological institutions aim to stay within these reference values, based on the analysis of an average European patient of about 70 kg weight. In particular in this context, the dose indicators (CTDI and DLP) that are displayed on the CT workstation are regarded as reference values. As the European Guidelines also state a good correspondence between the displayed and measured dose values,29 we did not perform further analysis of radiation exposure by CT dosimetry software. However, the authors would like to communicate that the CTDI and DLP in this study (as in others) do not represent the exact radiation dose of an individual patient but serve as a valid and reliable reference point of a clinical CT protocol for estimation.
The retrospective study design has to be put into context. However, even without randomization, the biometric characteristics (mean age and gender distribution) were comparable in both study groups and match epidemiological data. Owing to normal distribution, it can be assumed that the patients' size and height did not affect the considerations about radiation exposure in both cohorts. Moreover, a prospective study design should be well considered in terms of ethical aspects, in particular with respect to possible non-diagnosed injuries using a low-dose CT protocol without adjusting the scanner's capabilities.
As both scan protocols were established as routine WBCT protocols in a large Level 1 trauma centre, there were no non-diagnostic CT examinations, and none of the CT examinations had to be repeated because of insufficient image quality. In general, the evaluation of image quality using a multiple trauma CT protocol differs from dose studies, which focus on a single body region. Besides artefacts that can be caused by technical applications, such as electrocardiography cables, the arm positioning influences the image quality.28 With respect to the factor time, patient's safety and standardization of procedures, the arms were positioned in front of the trunk according to our institutional WBCT protocol for trauma patients.
We cannot clearly separate the single effects of ASiR from other technical components, such as the scintillator, as we used two different CT scanners in this study; both scanners were otherwise widely comparable as Scanner B was an upgrade on the platform of Scanner A.
The potential dose savings by the use of ASiR is supported by recent studies, which have shown similar results: dose reduction of about 30% in chest CT22 and about 33% in abdominal CT.30 On the other hand, Jiang et al15 also state that the gemstone-based detector allows for dose savings. However, a recent study by Yanagawa et al31 did not confirm a significant influence of the gemstone-based detector using a standard dose and a very low dose protocol (10 mA, 120 kVp), particularly if the high-definition mode was not enabled. They hypothesize that the detector characteristics could contribute more effectively in a moderate to low dose protocol. Further analysis is needed to clarify this topic. Apart from that, the aim of this study was not to evaluate a single parameter but the capability of the complete CT system for a clinical purpose. Namely, to introduce the new technique safely into clinical practice and make it available for patients with complex WBCT scans that benefit best from this new dose-reduced scan technique.
CONCLUSION
In conclusion, it is essential to keep the dose of the examination as low as reasonably achievable. In routine WBCT of patients with multiple injuries, it is possible to reduce the radiation dose by a combination of recent technical developments, such as ASiR, tube current modulation and a gemstone-based scintillator.
CONFLICTS OF INTEREST
LLG was on the speaker's bureau for GE Healthcare from 2010 to 2012.
FUNDING
ZD received a research grant from GE Healthcare.
Contributor Information
Lucas L Geyer, Email: Lucas.Geyer@med.lmu.de.
Markus Körner, Email: markus.koerner@med.uni-muenchen.de.
Andreas Harrieder, Email: andreas@harrieder.com.
Fabian G Mueck, Email: fabian.mueck@med.uni-muenchen.de.
Zsuzsanna Deak, Email: ulrich.linsenmaier@helios-kliniken.de.
Stefan Wirth, Email: stefan.wirth@med.uni-muenchen.de.
Ulrich Linsenmaier, Email: ulrich.linsenmaier@helios-kliniken.de.
REFERENCES
- 1.Larson DB, Johnson LW, Schnell BM, Salisbury SR, Forman HP. National trends in CT use in the emergency department: 1995–2007. Radiology 2011; 258: 164–73. doi: 10.1148/radiol.10100640 [DOI] [PubMed] [Google Scholar]
- 2.Broder J, Warshauer DM. Increasing utilization of computed tomography in the adult emergency department, 2000–2005. Emerg Radiol 2006; 13: 25–30. doi: 10.1007/s10140-006-0493-9 [DOI] [PubMed] [Google Scholar]
- 3.Linsenmaier U, Krötz M, Häuser H, Rock C, Rieger J, Bohndorf K, et al. Whole-body computed tomography in polytrauma: techniques and management. Eur Radiol 2002; 12: 1728–40. doi: 10.1007/s00330-001-1225-x [DOI] [PubMed] [Google Scholar]
- 4.Körner M, Reiser M, Linsenmaier U. Imaging of trauma with multi-detector computed tomography [in German.] Radiologe 2009; 49: 510–15. doi: 10.1007/s00117-008-1807-6 [DOI] [PubMed] [Google Scholar]
- 5.Huber-Wagner S, Lefering R, Qvick LM, Körner M, Kay MV, Pfeifer KJ, et al. Effect of whole-body CT during trauma resuscitation on survival: a retrospective, multicentre study. Lancet 2009; 373: 1455–61. doi: 10.1016/S0140-6736(09)60232-4 [DOI] [PubMed] [Google Scholar]
- 6.Brenner DJ, Elliston CD. Estimated radiation risks potentially associated with full-body CT screening. Radiology 2004; 232: 735–8. doi: 10.1148/radiol.2323031095 [DOI] [PubMed] [Google Scholar]
- 7.Hall EJ, Brenner DJ. Cancer risks from diagnostic radiology. Br J Radiol 2008; 81: 362–78. doi: 10.1259/bjr/01948454 [DOI] [PubMed] [Google Scholar]
- 8.McCollough CH, Guimaraes L, Fletcher JG. In defense of body CT. AJR Am J Roentgenol 2009; 193: 28–39. doi: 10.2214/AJR.09.2754 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Brenner DJ, Hall EJ. Computed tomography—an increasing source of radiation exposure. N Engl J Med 2007; 357: 2277–84. doi: 10.1056/NEJMra072149 [DOI] [PubMed] [Google Scholar]
- 10.Gunn ML, Kohr JR. State of the art: technologies for computed tomography dose reduction. Emerg Radiol 2010; 17: 209–18. doi: 10.1007/s10140-009-0850-6 [DOI] [PubMed] [Google Scholar]
- 11.Hamberg LM, Rhea JT, Hunter GJ, Thrall JH. Multi-detector row CT: radiation dose characteristics. Radiology 2003; 226: 762–72. doi: 10.1148/radiol.2263020205 [DOI] [PubMed] [Google Scholar]
- 12.Kalra MK, Maher MM, Toth TL, Hamberg LM, Blake MA, Shepard JA, et al. Strategies for CT radiation dose optimization. Radiology 2004; 230: 619–28. doi: 10.1148/radiol.2303021726 [DOI] [PubMed] [Google Scholar]
- 13.Geyer LL, Schoepf UJ, Meinel FG, Nance JW, Jr, Bastarrika G, Leipsic JA, et al. State of the art: iterative CT reconstruction techniques. Radiology 2015; 276: 339–57. doi: 10.1148/radiol.2015132766 [DOI] [PubMed] [Google Scholar]
- 14.Mueck FG, Korner M, Scherr MK, Geyer LL, Deak Z, Linsenmaier U, et al. Upgrade to iterative image reconstruction (IR) in abdominal MDCT imaging: a clinical study for detailed parameter optimization beyond vendor recommendations using the adaptive statistical iterative reconstruction environment (ASIR). Rofo 2012; 184: 229–38. doi: 10.1055/s-0031-1282032 [DOI] [PubMed] [Google Scholar]
- 15.Jiang H, Vartuli J, Vess C. Gemstone–the ultimate scintillator for computed tomography: GE white paper. Waukesha, WI: GE Healthcare; 2008. [Google Scholar]
- 16.Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 2009; 193: 764–71. doi: 10.2214/AJR.09.2397 [DOI] [PubMed] [Google Scholar]
- 17.Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W. Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol 2010; 194: 191–9. doi: 10.2214/AJR.09.2953 [DOI] [PubMed] [Google Scholar]
- 18.McCollough CH, Bruesewitz MR, Kofler JM, Jr. CT dose reduction and dose management tools: overview of available options. Radiographics 2006; 26: 503–12. doi: 10.1148/rg.262055138 [DOI] [PubMed] [Google Scholar]
- 19.European Commission. European guidelines on quality criteria for computed tomography. Report EUR 16262. Brussels, Belgium; 1999. [Cited 17 February 2016.] Available from: http://www.drs.dk/guidelines/ct/quality/mainindex.htm [Google Scholar]
- 20.Kanz KG, Korner M, Linsenmaier U, Kay MV, Huber-Wagner SM, Kreimeier U, et al. Priority-oriented shock trauma room management with the integration of multiple-view spiral computed tomography [in German.] Unfallchirurg 2004; 107: 937–44. doi: 10.1007/s00113-004-0845-4 [DOI] [PubMed] [Google Scholar]
- 21.Prakash P, Kalra MK, Kambadakone AK, Pien H, Hsieh J, Blake MA, et al. Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Invest Radiol 2010; 45: 202–10. doi: 10.1097/RLI.ob013e3181dzfeec [DOI] [PubMed] [Google Scholar]
- 22.Prakash P, Kalra MK, Digumarthy SR, Hsieh J, Pien H, Singh S, et al. Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience. J Comput Assist Tomogr 2010; 34: 40–5. doi: 10.1097/RCT.0b013e3181b26c67 [DOI] [PubMed] [Google Scholar]
- 23.Ruchholtz S, Waydhas C, Schroeder T, Piepenbrink K, Kuhl H, Nast-Kolb D. The value of computed tomography in the early treatment of seriously injured patients [in German.] Chirurg 2002; 73: 1005–12. doi: 10.1007/s00104-002-0429-1 [DOI] [PubMed] [Google Scholar]
- 24.Ptak T, Rhea JT, Novelline RA. Radiation dose is reduced with a single-pass whole-body multi-detector row CT trauma protocol compared with a conventional segmented method: initial experience. Radiology 2003; 229: 902–5. doi: 10.1148/radiol.2293021651 [DOI] [PubMed] [Google Scholar]
- 25.Fanucci E, Fiaschetti V, Rotili A, Floris R, Simonetti G. Whole body 16-row multislice CT in emergency room: effects of different protocols on scanning time, image quality and radiation exposure. Emerg Radiol 2007; 13: 251–7. doi: 10.1007/s10140-006-0554-0 [DOI] [PubMed] [Google Scholar]
- 26.Körner M, Geyer LL, Wirth S, Reiser MF, Linsenmaier U. 64-MDCT in mass casualty incidents: volume image reading boosts radiological workflow. AJR Am J Roentgenol 2011; 197: W399–404. doi: 10.2214/AJR.10.5716 [DOI] [PubMed] [Google Scholar]
- 27.Bayer J, Pache G, Strohm PC, Zwingmann J, Blanke P, Baumann T, et al. Influence of arm positioning on radiation dose for whole body computed tomography in trauma patients. J Trauma 2011; 70: 900–5. doi: 10.1097/TA.0b013e3181edc80e [DOI] [PubMed] [Google Scholar]
- 28.Karlo C, Gnannt R, Frauenfelder T, Leschka S, Brüesch M, Wanner GA, et al. Whole-body CT in polytrauma patients: effect of arm positioning on thoracic and abdominal image quality. Emerg Radiol 2011; 18: 285–93. doi: 10.1007/s10140-011-0948-5 [DOI] [PubMed] [Google Scholar]
- 29.Bongartz G, Golding SJ, Jurik AG, et al. European guidelines for multislice computed tomography, funded by the European Commission, contract number FIGM-CT2000-20078-CT-TIP March 2004. [Cited 17 February 2016.] Available from: http://www.msct.eu/CT_Quality_Criteria.htm
- 30.Sagara Y, Hara AK, Pavlicek W, Silva AC, Paden RG, Wu Q. Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol 2010; 195: 713–19. doi: 10.2214/AJR.09.2989 [DOI] [PubMed] [Google Scholar]
- 31.Yanagawa M, Tomiyama N, Honda O, Kikuyama A, Sumikawa H, Inoue A, et al. Multidetector CT of the lung: image quality with garnet-based detectors. Radiology 2010; 255: 944–54. doi: 10.1148/radiol.10091010 [DOI] [PubMed] [Google Scholar]
