Skip to main content
Journal of Clinical Pathology logoLink to Journal of Clinical Pathology
. 1984 Dec;37(12):1327–1335. doi: 10.1136/jcp.37.12.1327

Evaluation of a commercial radioassay for the simultaneous estimation of vitamin B12 and folate, with subsequent derivation of the normal reference range.

E Raniolo, G Phillipou, G Paltridge, R E Sage
PMCID: PMC499007  PMID: 6392346

Abstract

A commercial assay kit method for the simultaneous estimation of vitamin B12 and folate concentrations has been evaluated. Values derived for folate by a microbiological assay and vitamin B12 by a verified radioassay showed good correlation with the investigated method. The clinical sensitivity of the assay for detecting deficient concentrations of vitamin B12 and folate was comparable to that of the non-commercial methods and other more definitive clinical procedures. Establishment of reference ranges, based on accepted statistical criteria, are discussed and such ranges are contrasted with those proposed by the manufacturer. The kit method is time and labour saving compared with the non-commercial methods.

Full text

PDF
1327

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Cornbleet P. J., Gochman N. Incorrect least-squares regression coefficients in method-comparison analysis. Clin Chem. 1979 Mar;25(3):432–438. [PubMed] [Google Scholar]
  2. Dawson D. W., Delamore I. W., Fish D. I., Flaherty T. A., Gowenlock A. H., Hunt L. P., Hyde K., MacIver J. E., Thornton J. A., Waters H. M. An evaluation of commercial radioisotope methods for the determination of folate and vitamin B12. J Clin Pathol. 1980 Mar;33(3):234–242. doi: 10.1136/jcp.33.3.234. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Fish D. I., Dawson D. W. Comparison of methods used in commercial kits for the assay of serum vitamin B12. Clin Lab Haematol. 1983;5(3):271–277. doi: 10.1111/j.1365-2257.1983.tb01364.x. [DOI] [PubMed] [Google Scholar]
  4. Lindenbaum J. Status of laboratory testing in the diagnosis of megaloblastic anemia. Blood. 1983 Apr;61(4):624–627. [PubMed] [Google Scholar]
  5. Olesen H. Vitamin B12 assays: call for a definitive method. Clin Chem. 1983 Jul;29(7):1442–1442. [PubMed] [Google Scholar]
  6. Paltridge G., Rudzki Z., Ryall R. G. Validity of transcobalamin II-based radioassay for the determination of serum vitamin B12 concentrations. Ann Clin Biochem. 1980 Nov;17(6):287–292. doi: 10.1177/000456328001700602. [DOI] [PubMed] [Google Scholar]
  7. Schilling R. F., Fairbanks V. F., Miller R., Schmitt K., Smith M. J. "Improved" vitamin B12 assays: a report on two commercial kits. Clin Chem. 1983 Mar;29(3):582–583. [PubMed] [Google Scholar]
  8. Tibbling G. A method for determination of vitamin B 12 in serum by radioassay. Clin Chim Acta. 1969 Jan;23(1):209–218. doi: 10.1016/0009-8981(69)90028-x. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group