Corneal biomechanics: Where are we?

Allan Luz 1,2,3,4,5, Fernando Faria-Correia 1,2,3,4,5, Marcela Q Salomão 1,2,3,4,5, Bernardo T Lopes 1,2,3,4,5, Renato Ambrósio Jr 1,2,3,4,5,
PMCID: PMC4992120  PMID: 27579450

Corneal biomechanics emerged as a very hot topic for research in Ophthalmology.1, 2 In this issue, Sharifipour and coworkers3 present an observational cross-sectional study in healthy individuals that evaluated the correlations of corneal viscoelastic properties measured by the Ocular Response Analyzer (ORA, Reichert, Buffalo, NY) with central corneal thickness (CCT) in different age groups. In this study, corneal hysteresis (CH) and corneal resistance factor (CRF) had a significant negative correlation with age (P < 0.001). Other studies had already demonstrated that CH and CRF have a negative correlation with age and a positive correlation with CCT.4 While statistical methodology was done properly in these studies, it is very important to note that CH and CRF do not express stiffness or the stress–strain behavior of corneal tissue. It is well recognized that corneal stiffness increases with age due to enzymatic pathways such as transglutaminase and lysyl oxidase, generating natural collagen crosslinking.5 In agreement with such concept, Elsheikh and coworkers analyzed human corneal specimens ranging in age between 50 and 95 years with an inflation test, finding a significant increase in stiffness associated with age and also with the load rate.6

The ORA was introduced in 2005 as the first instrument for assessing corneal biomechanics in vivo.7 The ORA is a non-contact tonometer (NCT) that monitors corneal deformation using the infrared reflex of the corneal apex. The applanation is detected as a peak on this reflex and is correlated with the pressure of the air puff. The inward phase (P1) and in the outward phase (P2) applanations are registered. The air puff maximal pressure is related to P1 as the system has an integrated loop to control the pump, which generates a collimated air pulse or puff with a symmetrical configuration. P1 and P2 pressure measurements are the basis for the first generation variables as reported by the original ORA software. The difference between the two pressures is called CH. Hysteresis is derived from the Greek, meaning ‘‘lagging behind’’.8 Corneal resistance factor is also derived from the inward and outward pressure values, based on a formula (P1 – kP2), where k is a constant that was empirically developed to augment the correlation with CCT. The concept was to develop a parameter (CRF) that reflects the resistance to deformation, but this is still related to IOP.7 Shah and coworkers first reported that hysteresis was significantly lower in keratoconic eyes compared to normal eyes.9 However, accuracy of CH and CRF is not good enough for using these values as single parameters for ectasia diagnosis.10 Even though pressure-dependent variables are relatively limited for detecting ectatic corneal disease, low CH is very relevant in glaucoma. CH is associated with optic nerve neuropathy in glaucoma. Interestingly, CH is more strongly associated with glaucoma presence, risk of progression, and effectiveness of glaucoma treatments than CCT.11

Beyond pressure-dependent parameters, different waveform-derived parameters were introduced for characterizing corneal deformation.8, 12, 13 Dupps and coworkers developed parameters related to the intensity of deformation in accordance to the pressure applied, along with other aspects of response and combinations of these variables.14 These parameters were found as better representatives of corneal structure in studies related to the diagnosis of ectatic disease.14, 15 Interestingly, integration of biomechanical data with corneal tomography has been demonstrated to augment accuracy for the identification of milder forms of ectasia.16, 17

The Corvis ST (Oculus, Wetzlar, Germany) was introduced as the second instrument for “in vivo” biomechanical assessment of the cornea. This is also a NCT system with a collimated symmetrical air puff pressure profile. However, unlike the ORA, the Corvis ST produces a consistent air puff maximal pressure for every examination. In addition, it has an ultra-fast Scheimpflug camera that takes 140 frames during the 33 ms of the measurement, which allows for a more detailed evaluation of corneal deformation.18 The Corvis ST provides a set of corneal deformation parameters based on the dynamic inspection of the corneal response.19 Deformation amplitude refers to the movement of the corneal apex in the anterior-posterior direction and is determined as the highest displacement of the apex at the highest concavity moment. Studies involving contact lenses with different material properties mounted on an anterior chamber model with adjusted pressures have demonstrated the impact of the chamber pressure on the deformation response.20 Novel corneal deformation parameters (CDP) were developed to improve ectasia detection.21 The Vinciguerra Screening Report provides these data along with correlations of normality values and biomechanically adjusted intraocular pressure. In addition, the horizontal Scheimpflug image of the undisturbed cornea provides data for calculating the profile or the rate of increase of corneal thickness from the apex towards nasal and temporal sides. In addition, the characterization of the thickness data on the horizontal Scheimpflug image enables the calculation of the Ambrósio Relational Thickness through the horizontal meridian (ARTh).22 ARTh has been combined with the CDP to generate the Corvis Biomechanical Index (CBI) that optimizes accuracy for detecting keratoconus. Vinciguerra and coworkers23 demonstrated that with a cut off of 0.5, CBI was able to correctly identify 98.2% of the keratoconic cases with 100% specificity. Further integration of the Corvis ST and Pentacam has already been tested and also provides the most accurate platform for early ectasia diagnosis. (Integration of corneal tomography and biomechanical parameters for diagnosis of ectatic disease. Ambrósio and coworkers, Poster ESCRS 2015).

Other approaches that combine deformation of the cornea with analysis of high-speed imaging, such as swept-source OCT or supersonic shear-wave imaging technology, have been proposed.8 The Brillouin optical microscopy is another technology that has been proposed to measure in vivo corneal biomechanics through the analysis of light scatter. This technology led to new insights regarding corneal biomechanics in ectatic diseases. Brillouin imaging showed differences between healthy and keratoconic corneas. Interestingly, it revealed that the mechanical weakening is primarily concentrated within the area of the corneal protrusion. Brillouin shift measures outside the protruded or conic area were comparable with that of healthy corneas.24 These findings are in agreement with the concept of focal weakening, starting a biomechanical decompensation cycle as proposed by Roberts.2 We predict fast developments and a bright future for corneal biomechanical assessments for ectasia detection, refining refractive surgery, glaucoma, and other applications for Ophthalmology.

Footnotes

Conflicts of interest: Renato Ambrósio Jr., MD, PhD is a consult of OCULUS Optikgeräte GmbH. For the remaining authors, none were declared.

Peer review under responsibility of the Iranian Society of Ophthalmology.

References

  • 1.Torres R., Merayo-Lloves J., Jaramillo M., Galvis V. Biomecánica de la córnea. Archivos de la Sociedad Española de Oftalmología. 2005;80:215–223. doi: 10.4321/s0365-66912005000400004. [DOI] [PubMed] [Google Scholar]
  • 2.Roberts C.J., Dupps W.J., Jr. Biomechanics of corneal ectasia and biomechanical treatments. J Cataract Refract Surg. 2014;40:991–998. doi: 10.1016/j.jcrs.2014.04.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Sharifipour F., Panahi-bazaz M., Bidar R., Idani A., Cheraghian B. Age-related variations in corneal biomechanical properties. J Curr Ophthalmol. 2016;28:117–122. doi: 10.1016/j.joco.2016.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Fontes B.M., Ambrósio_ R., Jr., Alonso R.S., Jardim D., Nosé W., Velarde G.C. Corneal biomechanical metrics in eyes with refraction of −19.00 to +9.00 D in healthy Brazilian patients. J Refract Surg. 2008;24:941–945. doi: 10.3928/1081597X-20081101-14. [DOI] [PubMed] [Google Scholar]
  • 5.O'Brart D.P. Corneal collagen cross-linking: a review. J Optom. 2014;7:113–124. doi: 10.1016/j.optom.2013.12.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Elsheikh A., Wang D., Brown M., Rama P., Campanelli M., Pye D. Assessment of corneal biomechanical properties and their variation with age. Curr Eye Res. 2007;32:11–19. doi: 10.1080/02713680601077145. [DOI] [PubMed] [Google Scholar]
  • 7.Luce D.A. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31:156–162. doi: 10.1016/j.jcrs.2004.10.044. [DOI] [PubMed] [Google Scholar]
  • 8.Dupps W.J., Roberts C.J. Corneal biomechanics: a decade later. J Cataract Refract Surg. 2014;40:857. doi: 10.1016/j.jcrs.2014.04.012. [DOI] [PubMed] [Google Scholar]
  • 9.Shah S., Laiquzzaman M., Bhojwani R., Mantry S., Cunliffe I. Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes. Investig Ophthalmol Vis Sci. 2007;48:3026–3031. doi: 10.1167/iovs.04-0694. [DOI] [PubMed] [Google Scholar]
  • 10.Fontes B.M., Ambrosio R., Jr., Jardim D., Velarde G.C., Nose W. Corneal biomechanical metrics and anterior segment parameters in mild keratoconus. Ophthalmology. 2010;117:673–679. doi: 10.1016/j.ophtha.2009.09.023. [DOI] [PubMed] [Google Scholar]
  • 11.Deol M., Taylor D.A., Radcliffe N.M. Corneal hysteresis and its relevance to glaucoma. Curr Opin Ophthalmol. 2015;26:96–102. doi: 10.1097/ICU.0000000000000130. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Mikielewicz M., Kotliar K., Barraquer R.I., Michael R. Air-pulse corneal applanation signal curve parameters for the characterisation of keratoconus. Br J Ophthalmol. 2011;95(6):793–798. doi: 10.1136/bjo.2010.188300. [DOI] [PubMed] [Google Scholar]
  • 13.Ventura B.V., Machado A.P., Ambrósio R. Analysis of waveform-derived ORA parameters in early forms of keratoconus and normal corneas. J Refract Surg. 2013;29:637–643. doi: 10.3928/1081597X-20130819-05. [DOI] [PubMed] [Google Scholar]
  • 14.Hallahan K.M., Roy A.S., Ambrosio R., Salomao M., Dupps W.J. Discriminant value of custom ocular response analyzer waveform derivatives in keratoconus. Ophthalmology. 2014;121:459–468. doi: 10.1016/j.ophtha.2013.09.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Luz A., Lopes B., Hallahan K.M. Discriminant value of custom ocular response analyzer waveform derivatives in forme fruste keratoconus. Am J Ophthalmol. 2016;164:14–21. doi: 10.1016/j.ajo.2015.12.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Ambrosio R., Jr., Nogueira L.P., Caldas D.L. Evaluation of corneal shape and biomechanics before LASIK. Int Ophthalmol Clin. 2011;51:11–38. doi: 10.1097/IIO.0b013e31820f1d2d. [DOI] [PubMed] [Google Scholar]
  • 17.Luz A., Lopes B., Hallahan K.M. Enhanced Combined Tomography and Biomechanics Data for Distinguishing Forme Fruste Keratoconus. J Refract Surg. 2016;32(7):479–494. doi: 10.3928/1081597X-20160502-02. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Ambrósio R., Jr., Ramos I., Luz A. Dynamic ultra high speed Scheimpflug imaging for assessing corneal biomechanical properties. Revista Brasileira de Oftalmologia. 2013;72:99–102. [Google Scholar]
  • 19.Piñero D.P., Alcón N. In vivo characterization of corneal biomechanics. J Cataract Refract Surg. 2014;40:870–887. doi: 10.1016/j.jcrs.2014.03.021. [DOI] [PubMed] [Google Scholar]
  • 20.Correia F.F., Ramos I., Roberts C.J., Steinmueller A., Krug M., Ambrósio R., Jr. Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed Scheimpflug imaging. Arquivos Brasileiros de Oftalmologia. 2013;76:278–281. doi: 10.1590/s0004-27492013000500005. [DOI] [PubMed] [Google Scholar]
  • 21.Ambrosio R., Jr., Lopes B., Faria-Correia F. Ectasia detection by the assessment of corneal biomechanics. Cornea. 2016;35:e18–20. doi: 10.1097/ICO.0000000000000875. [DOI] [PubMed] [Google Scholar]
  • 22.Lopes B.T., IdC Ramos, Salomão M.Q., Canedo A.L.C., Ambrósio R., Jr. Perfil paquimétrico horizontal para a detecção do ceratocone. Rev Bras Oftalmol. 2015;74:382–385. [Google Scholar]
  • 23.Vinciguerra R, Ambrósio Jr R, Elsheikh A, et al. Detection of keratoconus with a new corvis ST biomechanical index. J Refract Surg. in press. [DOI] [PubMed]
  • 24.Scarcelli G., Besner S., Pineda R., Yun S.H. Biomechanical characterization of keratoconus corneas ex vivo with Brillouin microscopyevaluation of Brillouin microscopy for keratoconus. Investig Ophthalmol Vis Sci. 2014;55:4490–4495. doi: 10.1167/iovs.14-14450. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Current Ophthalmology are provided here courtesy of Wolters Kluwer -- Medknow Publications

RESOURCES