. 2016 Jun 13;79(17):1113–1130. doi: 10.1007/s10337-016-3116-y

Table 2.

Steps and time for the 8 min automated mini-SPE method

Step Description Time (s)
1 Wash the 1 mL syringe with MeCN (2 pumps of 0.5 mL each) 30
2 Load 300 µL extract from AS vial in Tray1 into 1 mL syringe 10
3 Place mini-cartridge above collection AS vial (with glass insert) in Tray2 10
4 Elute extract through mini-cartridge at 2 µL s−1 150
5 Discard mini-cartridge into waste receptacle 5
6 Wash the 1 mL syringe with 1/1/1 MeCN/MeOH/water (2 pumps of 0.5 mL each) 30
7 Wash the 1 mL syringe with MeCN (4 pumps of 0.5 mL each) 45
8 Switch to 100 µL syringe and wash with MeCN (2 pumps of 50 µL each) 50
9 Add 25 µL MeCN to collection AS vial (with glass insert) in Tray2 10
10 Add 25 µL AP + QC sol’n to collection AS vial (with glass insert) in Tray2 10
11 Wash the 100 µL syringe with 1/1/1 MeCN/MeOH/water (5 pumps of 50 µL each) 50
12 Wash the 100 µL syringe with MeCN (3 pumps of 50 µL each) 40
13 Switch to 1 mL syringe and move to home position 40

The cleanup itself only took ≈3 min, and syringe wash and exchange steps took ≈5 min