Abstract
Purpose
Modification of a clinical MRI scanner to enable simultaneous or rapid interleaved acquisition of signals from two different nuclei.
Methods
A device was developed to modify the local oscillator signal fed to the receive channel(s) of an MRI console. This enables external modification of the frequency at which the receiver is sensitive and rapid switching between different frequencies. Use of the device was demonstrated with interleaved and simultaneous 31P and 1H spectroscopic acquisitions, and with interleaved 31P and 1H imaging.
Results
Signal amplitudes and signal‐to‐noise ratios were found to be unchanged for the modified system, compared with data acquired with the MRI system in the standard configuration.
Conclusion
Interleaved and simultaneous 1H and 31P signal acquisition was successfully demonstrated with a clinical MRI scanner, with only minor modification of the RF architecture. While demonstrated with 31P, the modification is applicable to any detectable nucleus without further modification, enabling a wide range of simultaneous and interleaved experiments to be performed within a clinical setting. Magn Reson Med 76:1636–1641, 2016. © 2015 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
Keywords: X‐nucleus, interleaved, simultaneous, multinuclear, MRS, MRI
INTRODUCTION
Magnetic resonance imaging and spectroscopy are powerful tools for investigating human anatomy and physiology in vivo. While the majority of studies and clinical routine MRI focus on 1H, other nuclei are also detectable, and all major hardware manufacturers provide system options to enable detection of these nuclei using clinical scanners. Nonproton, or X‐nuclear, studies give access to information not available from proton detection alone. For example, 31P is used to investigate metabolism via high energy phosphates 1, 2, 13C provides information about glucose and glycogen metabolism 3, 4, 5, and 23Na may be used to assess the viability of healthy and tumorous tissue 6, 7, 8 or cartilage defects 9.
Most in vivo multinuclear MR experiments essentially involve a series of mono‐nuclear acquisitions. However, simultaneous detection of signals from more than one nuclear species is possible 10, 11, 12, offering a reduction in total measurement time 13, 14. In practice, simultaneous acquisition is difficult because field gradients always act on both nuclei so that, for example, images acquired simultaneously from two different nuclei will have different fields of view, scaled by the respective gyromagnetic ratios 15. An alternative approach is to rapidly interleave acquisitions of two nuclear species 16, 17, 18, which offers similar benefit to simultaneous acquisition but without this intrinsic challenge. Interleaving also allows one nucleus to be sampled more frequently than the other 18, 19, 20, which is important for optimal signal‐to‐noise ratio (SNR) per unit time, given the potentially very large differences in intrinsic sensitivity and T 1 between 1H and X.
Simultaneous or interleaved acquisition can be exploited to acquire complementary data from transient states that are difficult or even impossible to repeat precisely, e.g., in patients, or when repetition of the experiment affects physiological parameters. For example, interleaved multinuclear 31P MRS and 1H MRI have been used to investigate the regulation of oxidative and glycolytic metabolism in skeletal muscle during a single exercise 18, 20, 21. Recently developed methods for quantitation of phosphorylated metabolites 22, 23 and pH 24 by 31P MRI and 31P MRS 25 would benefit by combination with interleaved measurements of perfusion and blood oxygenation by 1H MRI 26. Further, coregistration problems between 1H and X‐nuclear data sets could be alleviated and promising applications, such as dynamic shimming or 1H‐based motion correction of X‐nuclear data 12, require interleaved acquisition of 1H imaging data. However, most of these prior works on multinuclear simultaneous and interleaved detection have been accomplished on research systems or with significant modification to the spectrometer.
To run simultaneous or interleaved dual‐nuclear experiments, the MRI system must be capable of exciting the sample, and recording the resulting NMR signals, at the two different frequencies. Transmit pulses must be generated simultaneously or in quick succession. Most multinuclear‐capable MRI systems already have this capability, intended to be used for nuclear Overhauser enhancement (nOe), polarization transfer, and J‐decoupling. Also, the RF probe must be simultaneously tuned to both frequencies of interest. A standard dual‐resonant probe 27, 28, 29 may be used here, including those commercially available. Finally, the MRI console must be able to process and digitize signals at two different frequencies or, for interleaved acquisition, rapidly switch between two different acquisition frequencies. It is this last part that is generally not possible using a standard clinical MRI scanner, and that we address in this work.
METHODS
Functional Principle
MR signals can occur over a wide range of frequencies, typically on the order of tens to hundreds of megahertz. Rather than digitizing NMR signals directly at the Larmor frequency, most MRI spectrometers use the superheterodyne principle to first convert the signal down to a lower frequency of the order of a few megahertz 30. This is done by mixing the received signal with a reference signal, called the local oscillator (LO), to produce an output at the frequency difference between the original signal and the LO, known as the intermediate frequency (IF). By adjusting the LO, which is generated by the system synthesizer, the IF can always be set to a fixed frequency, independent of the frequency of the NMR signal. The signal is then directly digitized at the IF by the analogue‐to‐digital converter (ADC).
By externally modifying the LO, the frequency at which the spectrometer is sensitive can be manipulated. This is the principle behind the operation of the interleaving device, built to our specification by Pure Devices (Würzburg, Germany), shown schematically in Figure 1. It is inserted into the system receive chain between the synthesizer and the receivers. By default, the interleaving device rests in the passive state, passing the system LO from input to output. All system adjustments and mononuclear experiments can be performed in the normal manner when the interleaving device is in this state. The device is activated by a digital control signal provided by the MRI console and controlled via the pulse sequence. Once active, the interleaving device replaces the system LO with a signal generated by an auxiliary synthesizer. This synthesizer has a frequency resolution of 1 kHz and is synchronized to the MRI scanner via a console‐provided 10 MHz reference signal to ensure frequency and phase stability relative to the console. The frequency of the auxiliary synthesizer is set using an external computer via a Universal Serial Bus (USB) connection. When the interleaving device is active, the frequency of the signal digitized by the scanner is determined by the auxiliary synthesizer, rather than the console.
Figure 1.

Schematic diagram showing the RF pathway modification. The interleaving device is inserted between the system synthesizer and LO input of the receiver. In passive mode, the system LO is passed through the interleaving device, while in active mode, a replacement LO signal is generated by the auxiliary synthesizer, which is fed to the receiver in place of the system LO. The interleaving device is switched between passive and active mode via a control signal generated by the scanner and controlled via the pulse sequence.
The frequency synthesizer of the MRI console used in this work has a resolution of 2.5 MHz, limiting how accurately the LO can be offset from the Larmor frequency, resulting in different IFs for different nuclei. The ADC and RF stage following the mixer have a much wider bandwidth than the MR signal. Provided that the IF offset plus the bandwidth of the MR signal fits inside this wider bandwidth, the signal may be recorded at full fidelity. The frequency offset is then removed in digital postprocessing. The offset frequency used in this digital stage is not user‐configurable, so the frequency of the auxiliary synthesizer must be set to reproduce the IF of the nuclei that the console is configured to detect. For example, to digitize a 31P signal at 7 T (f 0 = 120.32 MHz), the console provides a LO at 122.50 MHz to give an IF of 2.18 MHz. To record a 1H signal (f 0 = 297.22 MHz), the interleaving device LO must be set to 299.41 MHz to reproduce the same IF. The receive bandwidth of the final signal is controlled per ADC event via the dwell time as a pulse sequence parameter and can be selected independently for each nucleus.
The receive channels of the MRI system used in this work are grouped into four banks of eight channels, each bank having a separate LO input. For interleaved acquisition, the frequency of any or all banks of receive channels can be switched. For simultaneous acquisition at two frequencies, at least one bank must continue to operate unchanged.
The interleaving device only modifies the receive chain of the scanner. The RF coil and its interface to the scanner require no customization. Any coil capable of X and 1H acquisition is suitable for interleaved acquisition, and any coil capable of heteronuclear decoupling can be used for simultaneous acquisition. The MRI console is already capable of generating transmit pulses at the Larmor frequency and at a second “decoupler” frequency, within a single pulse sequence. This feature is intended for nOe and J‐decoupling, but can be used to generate excitation pulses for a second nuclear species.
The safety‐critical transmit chain is not modified in any way, so the scanner RF safety system, including SAR monitoring, continues to operate in the normal manner. The interleaving device, therefore, has no impact on system safety or safety monitoring.
MR Experiments
To evaluate the function and performance of the interleaving device, spectroscopy and imaging experiments were conducted on a Siemens Magnetom 7 T (Siemens Medical, Erlangen, Germany). All experiments on the 7 T system were performed using an in‐house built RF probe, consisting of two 1H and three 31P transceive elements, shaped to a half cylinder (d = 14 cm, l = 10 cm) 29. The test object was a spherical phantom containing 500 mL aqueous phosphate solution (K2HPO4, 100 mmol/L). Functionality was also tested on a Siemens Tim Trio, a 3 T MRI system equipped with a concentric 31P/1H single loop coil (d = 10 cm, Rapid Biomedical, Würzburg, Germany).
To examine the interleaving device's influence on SNR and signal amplitude in NMR experiments as directly as possible, several series of pulse‐acquire 1H and 31P spectra were compared. 1H‐only and 31P‐only measurements were conducted without the interleaving device present, and with the interleaving device connected but passive (i.e., passing the system LO to the receivers). Interleaved and simultaneous multinuclear acquisitions were then made using the interleaving device with the sequences shown in Figure 2a and b. The interleaving device is only active during 1H reception. For simultaneous multinuclear acquisition, the LO fed to the receivers recording the 1H signal is modified by the interleaving device, while the receivers recording the 31P signal use the system LO unmodified. As the MRI console is not able to generate simultaneous RF pulses at two different frequencies, fast sequential excitation of both nuclei was used 31, as shown in Figure 2b. To eliminate the influence of nOe 18, experiments were repeated with the 1H transmit pulse amplitude set to zero for 31P data quantification.
Figure 2.

Pulse‐acquire sequences, for (a) interleaved and (b) simultaneous acquisition of 1H and 31P spectra; (c) shows a 3D gradient‐echo imaging sequence, which alternately acquires k‐space lines for 1H and 31P.
Because both the system and the auxiliary LO are running phase continuously, a constant phase will be accumulated in every consecutive acquisition of data (i.e., each k‐space line or averaging step, spaced by T R) that are subject to mixing with the interleaving device's LO. This phase accrual can be accounted for by prospectively incrementing the phase of the 1H acquisition by
| (1) |
where and f X are the respective Larmor frequencies and T R is the repetition time.
For reproducible SNR quantification, 50 spectra were acquired in each state (2048 complex points, 5000 Hz spectral bandwidth), using a repetition time of 2 s. SNR was quantified in the frequency domain as peak amplitude divided by the standard deviation of data points from an artefact‐free flat baseline region of the spectra. For 1H free induction decays, small nutations were used (∼0.1° near the coil), resulting in a constant signal amplitude throughout each measurement series and a flat spectral baseline. Phosphorous free induction decays, with inherently lower SNR, were excited with higher flip angles (∼70°). The first 10 spectra were excluded from each dataset to ensure that the system had reached steady‐state.
Images were also acquired to demonstrate use of the interleaving device with a more complicated acquisition sequence involving faster frequency switching and the use of field gradients. A 3D gradient‐echo sequence was modified to alternately acquire k‐space lines for each nucleus (Fig. 2c). The echo time and effective repetition time for each nucleus was T E = 3.8 ms and T R = 16 ms. Equal gradient amplitudes were used for both nuclei, resulting in 2.47 times ( ) larger voxels and field of view for the 31P images than for 1H images, helping to compensate for the lower sensitivity and abundance of 31P. The nominal voxel size was mm3 for the 31P images and mm3 for the 1H images, with the same matrix size of for both nuclei. Four averages were acquired in interleaved and 31P‐only scans, leading to a total measurement time of 3 min 17 s. For 1H MRI, unaveraged data are shown. Images were reconstructed offline using software written in‐house in the Perl Data Language (PDL, http://pdl.perl.org).
RESULTS
NMR data were successfully collected with interleaved and simultaneous multinuclear acquisitions on a Magnetom 7 T and a Tim Trio 3 T system. No adverse effects of the interleaving device on signal amplitude, SNR or phase stability were found. The switching time of the interleaving device, measured using an oscilloscope (TDS 3052, Tektronix, Beaverton, OR), was found to be within 10 μs of the control signal changing state.
The results of the performance tests using pulse‐acquire experiments at 7 T are shown in Figure 3. Measurement precision was better than for the signal amplitude for both nuclei, approximately for 31P SNR, and for 1H SNR. Looping the system LO through the interleaving device in the passive state made no measurable difference to SNR or signal intensity in free induction decays obtained for either nucleus. For simultaneous and interleaved 31P/1H acquisitions, an nOe‐induced increase of 31P signal and SNR on the order of 10% was observed when 1H magnetization was excited 18. SNRs obtained for 31P during interleaved and simultaneous measurements without 1H excitation were identical to those measured in mononuclear experiments without the interleaving device connected, to within measurement precision, while the differences in signal amplitudes were below 1%. Exciting 31P magnetization produced no measurable nOe in 1H spectra. Spectra used to quantify 1H signal intensity and SNR in simultaneous and interleaved acquisitions were measured in the presence of 31P excitation and showed no change in signal intensity or SNR when compared with mononuclear acquisitions made without the interleaving device present. In summary, the interleaving device did not affect the SNR or amplitude of 31P or 1H signals in any of the above measurements.
Figure 3.

Signal‐to‐noise ratio and signal amplitude of (a) 31P and (b) 1H pulse‐acquire experiments using a test object. Performance with the scanner in the standard configuration was not degraded when the interleaving device was added in passive mode, and when it was used to enable interleaved or simultaneous acquisitions. Note that the interleaving device is only ever active during 1H acquisitions. For interleaved acquisitions, the device is always passive during 31P readout, while for simultaneous acquisitions 31P is measured on a receive bank not connected to the device.
Images acquired at 7 T are shown in Figure 4. No difference is seen between 1H images acquired with and without interleaving. An increase in signal intensity is seen in the 31P image with interleaving, relative to the image without, due to nOe from the 1H pulses. It was verified that the larger 31P FOV in interleaved acquisitions did not have an effect on SNR, 31P images were also acquired with the same voxel size but with a reduced matrix size of , corresponding to the same FOV as 1H images. The number of averages was increased to 24, for comparable total measurement time (3 min 20 s).
Figure 4.

1H (a, c) and 31P (b, d) images acquired with (a, b) and without (c, d) interleaving. Images in the top row were acquired in a single experiment. Images in the bottom row were acquired in two separate experiments, using the same sequence, but with transmission at the unused frequency omitted, to demonstrate the effect of nOe. 31P images were cropped to have the same FOV as 1H images. The increased brightness of (b) relative to (d) can be ascribed to nOe from the 1H excitation pulses.
A constant phase accrual was found between consecutive T R steps in 1H data. The phase increment between 30 free induction decays acquired in a single scan was linearly T R‐dependent ( ) and could be corrected by introducing an incremental offset to the receiver phase for 1H acquisitions using Eq. (1). 31P data, acquired using the system LO routed via the interleaving device, showed no phase drift.
DISCUSSION AND CONCLUSION
In this work, we have demonstrated a simple hardware modification to a clinical MRI scanner that enables interleaved and simultaneous acquisition of signals from two different nuclei. Using 31P and 1H, signal amplitudes and signal‐to‐noise ratios quantified in spectra and images acquired using the interleaving device were found to be indistinguishable from data acquired in separate, mononuclear acquisitions with the system in its standard configuration.
While the device was demonstrated using 31P and 1H, the presented method can be used to combine acquisitions of any pair of nuclei the host system is able to excite. The interleaving device has many possible applications, such as 31P or 13C spectroscopy concurrent with 1H spectroscopy or imaging, simultaneous 23Na/1H imaging, combined 31P/1H spectroscopic imaging or the use of 1H navigators for nonproton MRI. Decoupling at the Larmor frequency of one of the interleaved nuclei can be applied during the readout of the other nucleus (e.g., during an interleaved 13C/1H acquisition, 1H decoupling of the 13C readout is possible). However, this effectively reduces the T R at the decoupling frequency, which will impact the SNR efficiency for that nucleus. Developing pulse sequences that integrate excitation and readout of signal from two different nuclei is potentially complicated, but the only modification required to use the interleaving device is the addition of a continuous control signal for the duration of the frequency‐shifted acquisition. Most MRI consoles provide such user‐programmable control signals, generally used for triggering and system debugging.
The modification presented here can be applied to any system using superheterodyne receivers where the LO is accessible, where the console allows transmission of RF pulses at different frequencies during a single pulse sequence (as is usually possible on systems equipped with X‐nuclear capabilities for decoupling and nOe), and that provides a digital control signal that can be used to activate the interleaving device during reception.
The method we present here is very similar to the approach recently presented by Jeong and Kaggie 32, 33. Their method introduces an extra mixer, used to shift the frequency of the interleaved 1H signal down to the X frequency, which the scanner then records as if it was an X signal. This adds an extra component to the RF chain, which may have some impact on system SNR, but allows all system modifications to be made between the RF coil and the receive plug of the scanner. Our approach is simpler, in that only the receiver LO is modified, with all mixing done by the standard system components in the normal manner. As no additional components are inserted into the receive chain, the impact on SNR is negligible. An advantage of the method presented here is its inherent array coil capability, as all channels of a receiver bank (eight in case of the MRI system used here) are controlled with a single device.
In summary, the device presented in this work enables simultaneous and rapidly interleaved acquisition of signals from two different nuclei, without loss of signal intensity or SNR for either nucleus. The interleaving device is the technical prerequisite for the combining dual‐nuclear imaging and spectroscopy techniques into one sequence, which had been impossible on many clinical platforms, thus opening up exciting possibilities for future research and potential clinical application.
ACKNOWLEDGMENT
Martin Meyerspeer and Arthur William Magill contributed equally to this work.
The first authorship is shared between Meyerspeer and Magill.
REFERENCES
- 1. Chance B, Leigh JS, Kent J, McCully K, Nioka S, Clark BJ, Maris JM, Graham T. Multiple controls of oxidative metabolism in living tissues as studied by phosphorus magnetic resonance. Proc Natl Acad Sci USA 1986;83:9458–9462. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Kemp GJ, Ahmad RE, Nicolay K, Prompers JJ. Quantification of skeletal muscle mitochondrial function by 31P magnetic resonance spectroscopy techniques: a quantitative review. Acta Physiol (Oxf) 2015;213:107–144. [DOI] [PubMed] [Google Scholar]
- 3. Chance B, Eleff S, Leigh JS. Noninvasive, nondestructive approaches to cell bioenergetics. Proc Natl Acad Sci USA 1980;77:7430–7434. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Rothman DL, Magnusson I, Katz LD, Shulman RG, Shulman GI. Quantitation of hepatic glycogenolysis and gluconeogenesis in fasting humans with 13C NMR. Science 1991;254:573–576. [DOI] [PubMed] [Google Scholar]
- 5. Gruetter R, Adriany G, Choi IY, Henry PG, Lei H, Oz G. Localized in vivo 13C NMR spectroscopy of the brain. NMR Biomed 2003;16:313–338. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Babsky AM, Topper S, Zhang H, Gao Y, James JR, Hekmatyar SK, Bansal N. Evaluation of extra‐ and intracellular apparent diffusion coefficient of sodium in rat skeletal muscle: effects of prolonged ischemia. Magn Reson Med 2008;59:485–491. [DOI] [PubMed] [Google Scholar]
- 7. Hussain MS, Stobbe RW, Bhagat YA, Emery D, Butcher KS, Manawadu D, Rizvi N, Maheshwari P, Scozzafava J, Shuaib A, et al. Sodium imaging intensity increases with time after human ischemic stroke. Ann Neurol 2009;66:55–62. [DOI] [PubMed] [Google Scholar]
- 8. Madelin G, Regatte RR. Biomedical applications of sodium MRI in vivo. J Magn Reson Imaging 2013;38:511–529. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Trattnig S, Welsch GH, Juras V, Szomolanyi P, Mayerhoefer ME, Stelzeneder D, Mamisch TC, Bieri O, Scheffler K, Zbn S. 23Na MR imaging at 7 T after knee matrix‐associated autologous chondrocyte transplantation preliminary results. Radiology 2010;257:175–184. [DOI] [PubMed] [Google Scholar]
- 10. Thulborn KR, Soffe NF, Radda GK. Simultaneous in vivo measurement of oxygen utilization and high‐energy phosphate metabolism in rabbit skeletal muscle by multinuclear 1H and 31P NMR. J Magn Reson 1981;45:362–366. [Google Scholar]
- 11. Gonen O, Murphy‐Boesch J, Srinivasan R, Hu J, Jiang H, Stoyanova R, Brown TR. Simultaneous and interleaved multinuclear chemical‐shift imaging, a method for concurrent, localized spectroscopy. J Magn Reson B 1994;104:26–33. [DOI] [PubMed] [Google Scholar]
- 12. Keupp J, Rahmer J, Grässlin I, Mazurkewitz PC, Schaeffter T, Lanza GM, Wickline SA, Caruthers SD. Simultaneous dual‐nuclei imaging for motion corrected detection and quantification of 19F imaging agents. Magn Reson Med 2011;66:1116–1122. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Gonen O, Hu J, Murphy‐Boesch J, Stoyanova R, Brown TR. Dual interleaved 1H and proton‐decoupled‐31P in vivo chemical shift imaging of human brain. Magn Reson Med 1994;32:104–109. [DOI] [PubMed] [Google Scholar]
- 14. de Bruin PW, Koken P, Versluis MJ, Aussenhofer SA, Meulenbelt I, Börnert P, Webb AG. Time‐efficient interleaved human 23Na and 1H data acquisition at 7 T. NMR Biomed 2015;28:1228–1235. [DOI] [PubMed] [Google Scholar]
- 15. Stehning C, Keupp J, Rahmer J. Simultaneous 23Na/1H imaging with dual excitation and double tuned birdcage coil In Proceedings of the 19th Annual Meeting of ISMRM, Montréal, Canada, 2011. p 1501. [Google Scholar]
- 16. Richardson RS, Noyszewski EA, Kendrick KF, Leigh JS, Wagner PD. Myoglobin O2 desaturation during exercise. Evidence of limited O2 transport. J Clin Invest 1995;96:1916–1926. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Vanderthommen M, Duteil S, Wary C, Raynaud JS, Leroy‐Willig A, Crielaard JM, Carlier PG. A comparison of voluntary and electrically induced contractions by interleaved 1H‐ and 31P‐NMRS in humans. J Appl Physiol 2003;94:1012–1024. [DOI] [PubMed] [Google Scholar]
- 18. Meyerspeer M, Krššák M, Kemp GJ, Roden M, Moser E. Dynamic interleaved 1H/31P STEAM MRS at 3 Tesla using a pneumatic force‐controlled plantar flexion exercise rig. Magn Reson Mater Phys 2005;18:257–262. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Carlier PG, Brillault‐Salvat C, Giacomini E, Wary C, Bloch G. How to investigate oxygen supply, uptake, and utilization simultaneously by interleaved NMR imaging and spectroscopy of the skeletal muscle. Magn Reson Med 2005;54:1010–1013. [DOI] [PubMed] [Google Scholar]
- 20. Meyerspeer M, Kemp GJ, Mlynárik V, Krššák M, Szendroedi J, Nowotny P, Roden M, Moser E. Direct noninvasive quantification of lactate and high energy phosphates simultaneously in exercising human skeletal muscle by localized magnetic resonance spectroscopy. Magn Reson Med 2007;57:654–660. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Brillault‐Salvat C, Giacomini E, Jouvensal L, Wary C, Bloch G, Carlier PG. Simultaneous determination of muscle perfusion and oxygenation by interleaved NMR plethysmography and deoxymyoglobin spectroscopy. NMR Biomed 1997;10:315–323. [DOI] [PubMed] [Google Scholar]
- 22. Parasoglou P, Xia D, Regatte RR. Feasibility of mapping unidirectional Pi‐to‐ATP fluxes in muscles of the lower leg at 7.0 Tesla. Magn Reson Med 2014;74:225–230. [DOI] [PubMed] [Google Scholar]
- 23. Parasoglou P, Xia D, Chang G, Regatte RR. Dynamic three‐dimensional imaging of phosphocreatine recovery kinetics in the human lower leg muscles at 3T and 7T: a preliminary study. NMR Biomed 2013;26:348–356. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Schmid AI, Meyerspeer M, Robinson SD, Goluch S, Wolzt M, Fiedler GB, Bogner W, Laistler E, Krššák M, Moser E, et al. Dynamic PCr and pH imaging of human calf muscles during exercise and recovery using 31P gradient‐Echo MRI at 7 Tesla. Magn Reson Med, in press. [DOI] [PubMed] [Google Scholar]
- 25. Fiedler GB, Meyerspeer M, Schmid AI, Goluch S, Schewzow K, Laistler E, Mirzahosseini A, Niess F, Unger E, Wolzt M, et al. Localized semi‐LASER dynamic 31P magnetic resonance spectroscopy of the soleus during and following exercise at 7 T. Magn Reson Mater Phy 2015;28:493–501. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Schmid AI, Schewzow K, Fiedler GB, Goluch S, Laistler E, Wolzt M, Moser E, Meyerspeer M. Exercising calf muscle changes correlate with pH, PCr recovery and maximum oxidative phosphorylation. NMR Biomed 2014;27:553–560. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Magill AW, Gruetter R. Nested surface coils for multinuclear NMR In: Grant DM, Harris RK, editors. Encyclopedia of magnetic resonance. Chichester: Wiley; 2011. [Google Scholar]
- 28. Mispelter J, Lupu M, Briguet A. NMR probeheads for biophysical and biomedical experiments. London: Imperial College Press; 2006. [Google Scholar]
- 29. Goluch S, Kuehne A, Meyerspeer M, Kriegl R, Schmid AI, Fiedler GB, Herrmann T, Mallow J, Hong SM, Cho ZH, et al. A form‐fitted three channel 31P, two channel 1H transceiver coil array for calf muscle studies at 7T. Magn Reson Med 2014;73:1190–1195. [DOI] [PubMed] [Google Scholar]
- 30. Hoult DI. Receiver design for MR In: Vaughan JT, Griffiths JR, editors. RF coils for MRI. Wiley; Chichester, UK, 2012. [Google Scholar]
- 31. Klomp DW, Kentgens AP, Heerschap A. Polarization transfer for sensitivity‐enhanced MRS using a single radio frequency transmit channel. NMR Biomed 2008;21:444–452. [DOI] [PubMed] [Google Scholar]
- 32. Jeong EK, Sapkota N, Kaggie J, Shi XF. Simultaneous dual‐nuclear 31P/1H MRS at a clinical MRI System with time‐sharing second RF channel. In Proceedings of the 21st Annual Meeting of ISMRM, Salt Lake City, Utah, USA, 2013. p 2781. [Google Scholar]
- 33. Kaggie JD, Sapkota N, Jeong K, Shi X, Morrell GR, Bangerter NK, Jeong EK. Synchronous 1H and 23Na dual‐nuclear MRI on a clinical MRI system, equipped with a time‐shared second transmit channel. In Proceedings of the 22nd Annual Meeting of ISMRM, Milan, Italy, 2014. p 3789. [Google Scholar]