Multicentre evaluation of the Thrombotest International Sensitivity Index used with a steel ball coagulometer.

A M van den Besselaar 1, R H Peters 1
PMCID: PMC500484  PMID: 8707959

Abstract

AIM: To compare the International Sensitivity Index (ISI) of the Thrombotest reagent used with a steel ball coagulometer (KC) to the ISI of the same reagent used with the manual (tilt tube) technique. METHODS: The study was carried out by eight laboratories using their own KC instrument and method of testing. All laboratories used the same batch of Thrombotest to determine the clotting times of fresh blood samples from 20 local healthy volunteers and 60 patients on long term oral anticoagulant therapy. KC clotting times were plotted against manual clotting times on double logarithmic scales. Orthogonal regression lines were calculated to assess the ISI. RESULTS: In two laboratories the ISI of the KC method was lower than that of the manual method; these differences, however, were 2% or less. In the other laboratories no clinically important differences were observed between ISI values obtained. However, the clotting times determined with the KC methods were shorter than the manual values. CONCLUSIONS: The ISI of Thrombotest determined with the KC methods was very similar to the manual value. Therefore, use of the ISI value supplied by the manufacturer without adjustment is justified. The mean normal prothrombin time, however, must be determined locally.

Full text

PDF
414

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Chantarangkul V., Tripodi A., Mannucci P. M., Bonomi A. B. The effect of instrumentation on thromboplastin calibration. Thromb Haemost. 1992 May 4;67(5):588–589. [PubMed] [Google Scholar]
  2. Clarke K., Taberner D. A., Thomson J. M., Morris J. A., Poller L. Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers. J Clin Pathol. 1992 Jan;45(1):58–60. doi: 10.1136/jcp.45.1.58. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Crespo B., Ankri A., Lesty C. Evaluation de l'appareil modulaire semi automatique de coagulation KC10. Nouv Rev Fr Hematol. 1983;25(2):113–121. [PubMed] [Google Scholar]
  4. Hermans J., van den Besselaar A. M., Loeliger E. A., van der Velde E. A. A collaborative calibration study of reference materials for thromboplastins. Thromb Haemost. 1983 Oct 31;50(3):712–717. [PubMed] [Google Scholar]
  5. International Committee for Standardization in Haematology, International Committee on Thrombosis and Haemostasis: ICSH/ICTH recommendations for reporting prothrombin time in oral anticoagulant control. Thromb Haemost. 1985 Feb 18;53(1):155–156. [PubMed] [Google Scholar]
  6. Kitchen S., Walker I. D., Woods T. A., Preston F. E. Thromboplastin related differences in the determination of international normalised ratio: a cause for concern? Steering Committee of the UK National External Quality Assessment Scheme in Blood Coagulation. Thromb Haemost. 1994 Sep;72(3):426–429. [PubMed] [Google Scholar]
  7. Peters R. H., van den Besselaar A. M., Olthuis F. M. A Multi-centre study to evaluate method dependency of the international sensitivity index of bovine thromboplastin. Thromb Haemost. 1989 Apr 25;61(2):166–169. [PubMed] [Google Scholar]
  8. Peters R. H., van den Besselaar A. M., Olthuis F. M. Determination of the mean normal prothrombin time for assessment of international normalized ratios. Usefulness of lyophilized plasma. Thromb Haemost. 1991 Oct 1;66(4):442–445. [PubMed] [Google Scholar]
  9. Poggio M., van den Besselaar A. M., van der Velde E. A., Bertina R. M. The effect of some instruments for prothrombin time testing on the International Sensitivity Index (ISI) of two rabbit tissue thromboplastin reagents. Thromb Haemost. 1989 Nov 24;62(3):868–874. [PubMed] [Google Scholar]
  10. Poller L., Thomson J. M., Taberner D. A., Clarke D. K. The correction of coagulometer effects on international normalized ratios: a multicentre evaluation. Br J Haematol. 1994 Jan;86(1):112–117. doi: 10.1111/j.1365-2141.1994.tb03260.x. [DOI] [PubMed] [Google Scholar]
  11. Poller L., Thomson J. M., Taberner D. A. Effect of automation on prothrombin time test in NEQAS surveys. J Clin Pathol. 1989 Jan;42(1):97–100. doi: 10.1136/jcp.42.1.97. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Preston F. E. Quality control and oral anticoagulation. Thromb Haemost. 1995 Jul;74(1):515–520. [PubMed] [Google Scholar]
  13. Ray M. J., Smith I. R. The dependence of the International Sensitivity Index on the coagulometer used to perform the prothrombin time. Thromb Haemost. 1990 Jun 28;63(3):424–429. [PubMed] [Google Scholar]
  14. van Rijn J. L., Schmidt N. A., Rutten W. P. Correction of instrument- and reagent-based differences in determination of the International Normalized Ratio (INR) for monitoring anticoagulant therapy. Clin Chem. 1989 May;35(5):840–843. [PubMed] [Google Scholar]
  15. van den Besselaar A. M., Bertina R. M. Multi-center calibration of the second reference material for thromboplastin rabbit, plain, coded CRM 149R. Thromb Haemost. 1991 Mar 4;65(3):263–267. [PubMed] [Google Scholar]
  16. van den Besselaar A. M., Bertina R. M. Multi-center study of thromboplastin calibration precision--influence of reagent species, composition, and International Sensitivity Index (ISI). Thromb Haemost. 1993 Jan 11;69(1):35–40. [PubMed] [Google Scholar]
  17. van den Besselaar A. M., Lewis S. M., Mannucci P. M., Poller L. Status of present and candidate international reference preparations (IRP) of thromboplastin for the prothrombin time. A report of the Subcommittee for Control of Anticoagulation. Thromb Haemost. 1993 Jan 11;69(1):85–85. [PubMed] [Google Scholar]
  18. van den Besselaar A. M. Multi-center study of replacement of the international reference preparation for thromboplastin, rabbit, plain. Thromb Haemost. 1993 Nov 15;70(5):794–799. [PubMed] [Google Scholar]
  19. van den Besselaar A. M. Use of lyophilized calibration plasmas and control blood for international normalized ratio calculation in external quality assessment of the prothrombin time. Am J Clin Pathol. 1994 Jul;102(1):123–127. doi: 10.1093/ajcp/102.1.123. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES