Abstract
High performance liquid chromatography (HPLC) is increasingly being used to estimate variant haemoglobins. A case of haemoglobin S/D (HbS/D) is presented, which was misdiagnosed as haemoglobin S/E (HbS/E) by HPLC. The patient was a 22 year old woman with sickle cell anaemia. Subsequent haemodilution by blood transfusion clearly elucidated the haemoglobin D peak on HPLC. Sample dilution experiments, using the pretransfusion sample, were done resulting in correct elution of the peak in the D window. Troubleshooting in similar problematic haemoglobin variant peaks seen on HPLC can be done by sample dilution.
Full text
PDF

Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Samperi P., Mancuso G. R., Dibenedetto S. P., Di Cataldo A., Ragusa R., Schilirò G. High performance liquid chromatography (HPLC): a simple method to quantify Hb C, O-Arab, Agenogi and F. Clin Lab Haematol. 1991;13(2):169–175. doi: 10.1111/j.1365-2257.1991.tb00266.x. [DOI] [PubMed] [Google Scholar]