Sample dilution to resolve mistaken identification of haemoglobin D as haemoglobin E using the variant automated system.

S Thomas 1
PMCID: PMC500653  PMID: 9659274

Abstract

High performance liquid chromatography (HPLC) is increasingly being used to estimate variant haemoglobins. A case of haemoglobin S/D (HbS/D) is presented, which was misdiagnosed as haemoglobin S/E (HbS/E) by HPLC. The patient was a 22 year old woman with sickle cell anaemia. Subsequent haemodilution by blood transfusion clearly elucidated the haemoglobin D peak on HPLC. Sample dilution experiments, using the pretransfusion sample, were done resulting in correct elution of the peak in the D window. Troubleshooting in similar problematic haemoglobin variant peaks seen on HPLC can be done by sample dilution.

Full text

PDF
253

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Samperi P., Mancuso G. R., Dibenedetto S. P., Di Cataldo A., Ragusa R., Schilirò G. High performance liquid chromatography (HPLC): a simple method to quantify Hb C, O-Arab, Agenogi and F. Clin Lab Haematol. 1991;13(2):169–175. doi: 10.1111/j.1365-2257.1991.tb00266.x. [DOI] [PubMed] [Google Scholar]
  2. Tan G. B., Aw T. C., Dunstan R. A., Lee S. H. Evaluation of high performance liquid chromatography for routine estimation of haemoglobins A2 and F. J Clin Pathol. 1993 Sep;46(9):852–856. doi: 10.1136/jcp.46.9.852. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Wilson J. B., Headlee M. E., Huisman T. H. A new high-performance liquid chromatographic procedure for the separation and quantitation of various hemoglobin variants in adults and newborn babies. J Lab Clin Med. 1983 Aug;102(2):174–186. [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES