Skip to main content
Sensors (Basel, Switzerland) logoLink to Sensors (Basel, Switzerland)
. 2016 Aug 4;16(8):1229. doi: 10.3390/s16081229

Frequency Splitting Analysis and Compensation Method for Inductive Wireless Powering of Implantable Biosensors

Matthew Schormans 1,*, Virgilio Valente 1, Andreas Demosthenous 1
Editor: Alexander Star1
PMCID: PMC5017394  PMID: 27527174

Abstract

Inductive powering for implanted medical devices, such as implantable biosensors, is a safe and effective technique that allows power to be delivered to implants wirelessly, avoiding the use of transcutaneous wires or implanted batteries. Wireless powering is very sensitive to a number of link parameters, including coil distance, alignment, shape, and load conditions. The optimum drive frequency of an inductive link varies depending on the coil spacing and load. This paper presents an optimum frequency tracking (OFT) method, in which an inductive power link is driven at a frequency that is maintained at an optimum value to ensure that the link is working at resonance, and the output voltage is maximised. The method is shown to provide significant improvements in maintained secondary voltage and system efficiency for a range of loads when the link is overcoupled. The OFT method does not require the use of variable capacitors or inductors. When tested at frequencies around a nominal frequency of 5 MHz, the OFT method provides up to a twofold efficiency improvement compared to a fixed frequency drive. The system can be readily interfaced with passive implants or implantable biosensors, and lends itself to interfacing with designs such as distributed implanted sensor networks, where each implant is operating at a different frequency.

Keywords: frequency splitting, frequency tuning, implantable biosensors, inductive link, inductive powering, medical implants, wireless power transfer

1. Introduction

Short-range wireless power transfer (WPT) by means of inductive coupling is a safe and established technique, extensively used to transfer power to implanted medical devices (IMDs). WPT avoids the use of transcutaneous wires in applications where implanted batteries do not represent a viable alternative, for example, in high-density visual and brain implants [1,2,3,4], neuromuscular interfaces [5,6], and implanted biosensors [7,8,9].

Inductive links are based on weak coupling between an external transmitter coil and an implanted receiver coil optimized for operation at a specific resonant frequency and distance between the coils. Deviations in coil separation, alignment, shape, and load conditions greatly affect the link gain, resulting in significant variations in the power delivered to the implant. In particular, a phenomenon known as frequency splitting occurs in WPT systems when the coils are driven in the overcoupled regime [10]. Under these conditions, the link gain shows two peaks at different frequencies, which deviate from the tuned resonant frequency. Fixed-frequency WPT systems therefore suffer from dramatic changes in the output voltage as the relative distance between the coils varies. In order to maintain high system efficiency and constant supply voltage in the implant despite varying conditions, adaptive methods of tracking the splitting frequency and providing coupling-insensitive gain are required [11].

In fixed frequency WPT systems, uniform power delivery is commonly achieved by sensing the DC load voltage, transmitting its value to the transmitter via back-telemetry, and appropriately adjusting the supply voltage at the transmitter [12,13,14]. This method is suitable as long as the transmitter and receiver units are working close to resonance. Under overcoupled conditions, a significant increase in primary supply voltage is required to compensate for the occurrence of frequency splitting. This leads to a consequent drop in system efficiency. Alternative methods rely on careful optimization of the coil design to allow for coupling insensitive power transfer [15], but fail to address deviations from optimal operation arising from drifts in coil electrical parameters and geometric deformations. A number of adaptive systems have been described that accomplish power regulation in the secondary by electronically varying inductance [16,17] or capacitance [18] in either the primary or the secondary tank circuit. However, component variations and drift may further reduce the link efficiency, in particular where flexible and stretchable coils are used, whose parameters may fall outside the available compensation range. In addition, the achievable frequency resolution is determined by the number of taps in the tuning circuit, which limits the improvement in system efficiency that can be obtained. A triple-loop automatic tuning system including transmitter and receiver tuning and power control is reported in [19]. In addition to the limited resolution available, this system is only suitable for operation at a fixed frequency, and relies heavily on back telemetry to inform local control units.

An alternative approach to these methods is to directly control the drive frequency of the WPT system [20]. This has the benefit of requiring no variable link components, only a variable frequency synthesizer. There are several benefits to this approach. Firstly, the design of the power transmitter can be considered largely separately from the design of the coils, as long as the frequency synthesizer has a wide range. Secondly, a variable frequency synthesizer can be integrated on chip, saving valuable space in a compact transmitter compared to variable link components. Additionally, this approach requires no back-telemetry, as the link state is measured and maintained from the primary side. For these reasons, control of the drive frequency is the basis for the method described in this paper.

This paper presents an active compensation method for frequency splitting, based on optimum frequency tracking (OFT) in the overcoupled regime [21,22]. It can operate over a wide frequency range, and does not require the use of variable capacitors or inductors. A simple analytical derivation of the link gain is presented, that accounts for different coupling and load conditions. Theoretical link gain profiles are derived for a series-parallel link operating at fixed and variable frequencies. A compact closed-loop automatic frequency tuning system is presented. The link efficiency is optimized by sensing the phase between the primary voltage and current, and adjusting the frequency of the transmitter driver. This implementation uses a custom integrated, high-power Class-D amplifier with an on-chip frequency synthesizer [23], which allows for very fine frequency resolution to be obtained.

The rest of the paper is organised as follows. Section 2 presents some fundamentals of inductive link design and a simple theoretical analysis of link gain. Section 3 describes the proposed system. Section 4 describes the testing methodology and the measured results from employing the OFT compensation method. Section 5 concludes the paper, and provides a comparison with other work.

2. Inductive Link Theory

2.1. Inductive Link Fundamentals

A series-parallel model of a resonant inductive link is shown in Figure 1. This configuration allows the input to be driven with an AC voltage source Vin, and the output can be considered a voltage source Vout as a result of current in the output load RL.

Figure 1.

Figure 1

Idealised series-parallel inductive link model.

The primary and secondary sides consist of the coil inductances, LP and LS, the coil series resistances, RP and RS, and the tuning capacitors, CP and CS. M is the mutual inductance between the coils, which defines the coupling coefficient, k=M/LPLS. RL represents the equivalent load presented by the implant at the secondary. The tuned resonant frequency of both sides is defined in Equation (1):

ω0=1LPCP=1LSCS (1)

The link gain, A=Vout/Vin, can be determined as follows, in accordance with [24]. Firstly, the secondary circuit can be considered as a reflected impedance Zrefl, presented in series with the primary circuit:

Zrefl=(ωM)2ZS (2)

where ZS is the equivalent impedance of the secondary side of the link, and ω is the drive frequency. Considering the effect of Zrefl, the current in the primary side is therefore:

IP=VinZP+(ωM)2ZS (3)

where ZP is the equivalent impedance of the primary circuit (not including Zrefl). The voltage induced in the secondary by the primary circuit is jωMIP, and from Equation (3):

IS=jωMVinZPZS+(ωM)2 (4)

Equation (4) defines the output voltage Vout:

Vout=ISjωCS+1RL (5)

Therefore, since in the series-parallel configuration Vout appears across CS||RL, the gain A can be written as follows:

A=VoutVin=jωMZPZS+(ωM)2jωCS+1RL=jωkLPLSZPZS+ω2k2LPLSjωCS+1RL (6)

Equation (6) can be greatly simplified by assuming an ideal resonant condition of ω=ω0. However, this assumption does not hold true when the link coils are overcoupled, where the link will resonate at an optimum frequency ωoptω0 due to frequency splitting. It is possible to express the drive frequency in terms of a deviation from ω0 termed the ‘gamma factor’ [24]:

γ=ωω0 (7)

The primary and secondary impedances ZP and ZS can also be defined in terms of γ:

Z1=R1+jωL11ωC1=R1+jωL111γ2 (8a)
Z2=R2+jωL2+1/RLjωC21/RL2+(ωC2)2=R2+RL1+α+jωL2γ2RL2ωL2(1+α) (8b)

where α=ωCSRL2 as defined in [25]. By substituting the expressions in Equation (8) into Equation (6), and using the definition of Q-factor such that QP=ωLP/RP and QS=ωLS/RS, the gain can be defined in terms of Q-factor, α, and γ:

A(γ,k)=jkLS/LP1QP+j11γ2γ21QSα1+j1QS+1α+1+k2jγ2+1α (9)

Equation (9) provides a convenient expression for the link gain for different values of k, RL (contained within α), and γ. Using the definition of gain in Equation (9), the effect of frequency splitting on the link gain can be considered.

2.2. Frequency Splitting in Overcoupled Inductive Links

Frequency splitting can be described as a deviation of the link resonant frequency, referred to from here on as the optimum frequency ωopt, from ω0. Frequency splitting occurs when an inductive link is overcoupled, i.e., when the coupling k is greater than the critical coupling kcrit. For k=kcrit, the link gain is maximised when the link is driven at the tuned frequency, i.e., ωopt=ω0|k=kcrit [26]. While determination of the exact splitting frequencies is unnecessary in this context, kcrit for a given link must be identified, so that the overcoupled and undercoupled regions can be defined.

2.3. Relationship between Coupling and Coil Separation

When designing a pair of link coils to power an IMD, it is important to know the value of critical coupling kcrit, so that the over/undercoupled regions can be defined. kcrit can be defined in terms of the Q-factors [24]:

kcrit=1QPQS (10)

kcrit, however, is not a value that is immediately applicable to a design procedure; the value of k for a given distance d between two coils depends strongly on the coil geometries. Additionally, to measure the coupling k between a pair of coupled coils requires a complex empirical test process [25]. The distance d, however, can be directly related to a design by comparing dcrit with the expected implantation depth. In order to avoid this complex procedure for measuring k, it is convenient to translate from k to d using Equation (11) [27], which translates between M and d for given coil geometries (where M=kLPLS):

Mij=2μarirj1a22K(a)E(a) (11a)
a=2rirjri+rj2+d2 (11b)
M=gi=1NPj=1NSMijri,rj,d (11c)

where ri and rj are turn radii of the primary and secondary coils, respectively, μ is the magnetic permeability of the surrounding material, K(a) and E(a) are complete elliptic integrals of the first and second kind, respectively, NP and NS are the number of primary turns and secondary turns, respectively, and g is a shape factor that must be experimentally determined (for printed square coils g=1.1 [27]).

2.4. Effects of Frequency Splitting

Figure 2 shows the profile of the link gain versus variations in coupling and frequency, as modelled by Equations (9) and (11), for the coils specified in Table 1.

Figure 2.

Figure 2

(a) Normalized link gain against normalized coil spacing and (b) normalized link gain against gamma factor, for the coil parameters in Table 1 with RL=100 kΩ.

Table 1.

Measured coil parameters for operation at 5 MHz.

    Tx Coil Rx Coil
Outer Diameter do 64.7 mm 20.0 mm
Inner Diameter di 37.2 mm 4.7 mm
Number of Turns n 12 13
Inductance L 10.150 μH 2.255 μH
Q-factor Q 47.5 26.0

Figure 2a shows normalized link gain vs. coil spacing distance. When driving at ω0 (γ=1), the gain is defined by a bell shaped curve with a peak at critical coupling (d=dcrit). As the distance between the coils moves above or below dcrit, a significant loss of gain occurs. In the undercoupled case (d>dcrit), the loss is due to a reduction in coupling k. In the overcoupled region (d<dcrit), the loss is due to frequency splitting, and can be compensated for by adjusting the input frequency of Vin to the optimum frequency ωopt (γopt=ωopt/ω0). For instance, in Figure 2a, if the coil spacing reduces to half the critical spacing, a 50% loss in gain occurs if the drive frequency is maintained at ω=ω0. By reducing the driving frequency such that γ=0.96, the link gain increases by 5% compared to the value at dcrit, and increases by more than 100% for the same distance of d/dcrit=0.5. The γopt trend shown in Figure 2a illustrates that by maintaining the driving frequency at ωopt while d<dcrit, the link gain is maximised at a value equal to or greater than the gain when d=dcrit.

Figure 2b highlights how frequency splitting manifests at different coil spacing distances. When the link is critically coupled or undercoupled (ddcrit), the gain function is again represented by a bell-shaped curve, with a peak at ω0 (γ=1). When the link is overcoupled, however, this single peak splits into a pair of peaks at frequencies either side of ω0, one of which represents the optimum frequency, ωopt for Vin. This is generally the lower frequency peak, as there will be less impact from parasitic high frequency effects. The frequencies at which these peaks occur can be determined analytically [10].

2.5. Effects of RL

The definition of kcrit in Equation (10) is only accurate if RL is assumed to be infinite. This assumption holds up for some very low power implants [28]. However, it is more realistic to assume that RL.

The effect of RL is to add more parallel loss at the receiver, reducing QS. The breakdown of Equation (10) is highlighted by the plots in Figure 3.

Figure 3.

Figure 3

Link gain against variable distance and frequency for two distinctly different load resistances. (a) RL = 100 kΩ; (b) RL = 1 kΩ.

For very large resistances RL, frequency splitting is prominent, and the calculated value of dcrit is very close to the correct value. As this resistance approaches zero, splitting becomes less prominent, and the actual critical distance varies considerably from the calculated value: Figure 3b shows an error of approximately 50% compared with the value calculated from Equation (10).

To make Equation (10) more accurate when considering a loaded link, the way QS is determined must include the additional parallel loss from RL. If the parasitic secondary series loss RS is converted to a parallel loss, it can be considered in parallel with RL to find a ‘loaded Q’ value for QS, which accounts for the presence of RL. Equation (12) provides a means of recalculating QS to account for RL:

QS(loaded)=RLQS+1/QSωLSQS+1/QS+RLRLQSωLSQS+RL|QS>10 (12)

Therefore, by replacing QS in Equation (10) with QS(loaded) from Equation (12), it is possible to calculate much more accurate values for kcrit and dcrit when RL.

2.6. Considerations for Tracking ωopt

The results in Figure 2 and Figure 3 show that ωopt for a given link depends strongly on both the coil spacing and link load. Seeing that the spacing and load are both variable factors, it is insufficient to assume nominal fixed values and calculate a nominal fixed value of ω if the goal is to maintain a maximised output voltage. It is necessary, therefore, to devise a system that is capable of tracking ωopt regardless of changes in d and RL. The system proposed in Section 3 achieves this by monitoring the phase difference between the primary current and voltage; allowing it to track ωopt regardless of d and RL.

3. Frequency Tracking System

A closed-loop transmitter system was designed to compensate for frequency splitting that occurs in the overcoupled region of inductive links. The aim of the system is to adjust the drive frequency so that the phase between the voltage and current in the primary side is close to 0, ensuring operation at resonance.

Figure 4 shows a block diagram of the system, which includes a Class-D amplifier chip [23]. This chip includes a high resolution phase-locked loop (PLL) for clock generation, a programmable delay-line for optimizing the drive signal dead time, a phase detector (PD) referenced against the PLL clock, and a Class-D output stage capable of an output power of up to 30 W (at a supply voltage of up to 30 V). The chip was fabricated in a 0.18 μm CMOS technology. The inductive link formed by L1, L2, C1, and C2 is driven by the power amplifier (PA). In theory, the high Q of the resonant tanks means that the link currents can be considered as sinusoidal at the fundamental harmonic of the PLL frequency. The AC signal induced in the secondary coil is converted to DC via a discrete Schottky diode rectifier bridge, smoothed by the capacitor CREC, and delivered to the load RL. The current in L1 is monitored by using a capacitive divider formed from Ca and Cb that give the voltage VDIV. The purpose of the divider is to reduce the load on the primary from the monitoring circuit, and protect the monitoring circuit from the high voltage that can develop in the primary. VDIV lags the primary current by 90, and serves two purposes.

Figure 4.

Figure 4

Architecture of the closed-loop OFT system. Inset: photo of the transmitter prototype. The prototype contains the primary side of the inductive link, ED, LC, EA, and the Class-D amplifier. The microcontroller and ADC are connected externally.

First, the amplitude of VDIV is checked using an envelope detector (ED) and a ‘lock comparator’ (LC) (TLV3502, Texas Instruments, Dallas, TX, United States), which compares the ED output level with a pre-set offset voltage Voff. The output of the LC, Vlock, is then fed to a microcontroller (MSP430, Texas Instruments). Vlock is high when the link is close enough to resonance for the ED output to be higher than Voff; this allows room for calibration. Second, VDIV is fed to the on-chip PD for comparison with the clock signal, VPLL, generated by the on-chip PLL. The output of the PD is applied to an error amplifier (EA), which compares it with a target error voltage, Vtarg, which should represent 90 phase difference between VDIV and VPLL when calibrated correctly. The resulting voltage from the EA, Vϕ, is digitized by a 12-bit analogue to digital converter (ADC) (AD7091, Analog Devices) and fed back to the microcontroller. The EA consists of a differential gain stage and a single ended integrating stage. The integrating stage at the end of the chain damps the transient response to prevent instability. By making use of Vlock, to confirm the link is near resonance, and Vϕ, to monitor the phase angle between the primary current and voltage, the microcontroller can track the link state and adjust the PLL frequency so it stays at ωopt. Section 3.1 and Section 3.2, respectively, describe the operation of the transmitter system and the specifics of the inductive link coils.

3.1. Transmitter System Operation

The operation of the transmitter system can be set to either fixed-frequency mode (at ω0) or OFT mode, where the input frequency should automatically be set to ωopt. When operating in OFT mode, the system can be considered as a control system employing negative feedback, with the feedback signals represented by Vϕ and Vlock. The primary coil current is measured by using a capacitive divider formed of Ca and Cb (Ca = 10 pF, Cb = 33 pF), as part of the primary resonance capacitor C1. The phase difference between the coil current and the PLL clock is measured by the on-chip PD [23]. The photo inset in Figure 4 shows an example of a miniaturised prototype of the transmitter system, used to confirm successful operation of the feedback system even in very close proximity to the link’s magnetic field. The feedback action of the circuit operates under the control of the microcontroller as detailed in the flow diagram in Figure 5.

Figure 5.

Figure 5

Flow diagram of the operation of the microcontroller control unit.

Initially the driving frequency is set to ω0, which represents the resonant frequency of a specific coil pair at critical coupling. The PLL frequency is then varied either side of ω0 in 1 kHz steps to determine the working range of frequencies for the link. The upper and lower frequency bounds are then defined while the coils are brought together (recall that the frequency splitting phenomenon shows two peaks in the link gain curve either side of ω0, so only one is chosen). This process makes use of the feedback signal Vlock, so that checking whether the envelope of VDIV is greater than a specified offset Voff, determines if the link is close to resonance. After locating this working frequency range, the microcontroller enters the control loop routine (Main Loop in Figure 5). The error signal Vϕ is read from the ADC, and translated by the microcontroller into a frequency code that updates the PLL frequency.

3.2. Coil Design and Implementation

The link was tested using square printed spiral coils, fabricated on a standard FR4 substrate, and optimised for 5 MHz operation. To obtain the design parameters for the coils, the iterative process described in [27] was employed through a script in MATLAB. Initial constraints were applied to the following parameters: implanted coil outer diameter do2, typical load RL, working coil spacing d, and minimum PCB track spacing s. The constraints were do2=d= 20 mm, RL=10kΩ, s=150μm. The constraint on spacing was a conservative estimate of the limitations of the available PCB fabrication facility, while the other constraints were common example values in the context of biomedical implant size, implantation depth, and power requirements [29]. The measured coil parameters of the fabricated coils are listed in Table 1.

4. Testing and Results

4.1. Test Procedure

The operation of the closed-loop system was tested with a bench setup, in which the printed coils were mounted on a jig. This provided control over the lateral displacement between the coils from 5 mm to 70 mm. The system efficiency was calculated in terms of the DC power drawn from the supply and the secondary voltage VOUT across the load RL. Therefore, the efficiencies stated are system efficiencies, not link efficiencies.

Initially, the coils were aligned on the jig, and placed far enough apart to ensure power transfer in the undercoupled region (d = 70 mm); the value of ω0 could then be empirically determined through manual frequency adjustment. The distance between the coils was then manually adjusted between 5 mm and 50 mm in steps of 5 mm, and VOUT at the load was measured with OFT disabled and enabled. In a second test, the distance was varied as before, but the power supply was also manually adjusted such that VOUT = 10 V; the efficiency was recorded. In a third test, the link coils were fixed in an overcoupled position (d = 5 mm), and the load RL was varied; again VOUT and system efficiency were recorded, with OFT enabled and disabled.

4.2. Link Measurements

The operation of the OFT mode was tested at frequencies around the nominal 5 MHz, with the link parameters specified in Table 1, and a supply voltage of 5 V.

Figure 6a shows the measured (data points) and calculated (solid lines) relationships between coil spacing and link gain when RL = 100 kΩ. In the overcoupled region, the OFT compensation scheme demonstrates an increase in link gain when compared with the fixed-frequency approach; the experimental data closely match the predicted results. The overcoupled region for the coils used in this study extends to a distance of approximately 28 mm, beyond which the trends converge as ωoptω0. At d = 5 mm, there is a notable deviation in the experimental link gain from the calculated value. This deviation is attributed to capacitive coupling between the coils becoming significant, particularly given the relatively high values of RL used. Over the range d = 5 mm to 27 mm, the results show that OFT can provide a significant improvement in link gain in the overcoupled region, with measured gain improvements greater than twofold at d = 5 mm. This improvement in gain allows for reduction in the supply voltage at the transmitter, while still achieving the target VOUT, allowing for reduced overall power consumption. Figure 6b shows the measured and calculated link gain versus variations in load resistance RL with OFT enabled/disabled. The OFT system can automatically compensate for changes in RL in the same way it can compensate for changes in coil coupling, with a twofold increase in link gain for OFT vs. fixed frequency when RL = 100 kΩ. The small discrepancy in measured results at d = 5 mm and RL = 100 kΩ in Figure 6b is due to mechanical and component tolerances in the link, resulting in small systematic errors between measurement runs. The convergence of the OFT on and off trends in Figure 6b occurs at approximately RL = 1 kΩ. This convergence indicates the point at which the load becomes significantly lower than the parallel loss resistance of the receiver coil, which was calculated, from the values in Table 1, to be approximately 1.85 kΩ. Figure 6b presents a similar deviation between measurement and prediction for RL>50kΩ to that seen in Figure 6a. The fact that this deviation occurs only at the highest load values is further evidence that it is the result of stray capacitive coupling effects.

Figure 6.

Figure 6

Link gain against changes in coil spacing d and link load RL. Solid lines represent calculated behaviour, data points are experimentally determined. (a) Spacing d vs. Gain A, RL = 100 kΩ; (b) Load RL vs. Gain A, d = 5 mm.

The benefit of the OFT compensation scheme is more evident with the aid of Figure 7a, which shows the difference in normalized measured efficiency with and without OFT, while manually varying the supply to the transmitter PA in order to maintain a target VOUT = 10 V. The gain in system efficiency using OFT is significant in the overcoupled region, doubling at the smallest distance of 5 mm. Figure 7b shows the direct effect of OFT compensation on the efficiency of the system against changes in RL. Figure 7b illustrates that by employing OFT, the system efficiency can be increased for almost all load values. Under specific link conditions, where RL=500Ω and d=5mm, ωopt was found to be equal to ω0, and so the overall system efficiency was equal for OFT on and off.

Figure 7.

Figure 7

Plots to show experimental efficiencies against variations in load and coil spacing, with OFT enabled and disabled. (a) Normalized efficiency vs. spacing, RL=100kΩ; and (b) system efficiency vs. load, d=5mm.

Figure 6 and Figure 7 together show that while there are some gains in system efficiency, the primary benefit of OFT is that it maintains a more constant output voltage at the receiver than the fixed frequency mode.

4.3. Real-Time System Operation

Figure 8 shows the transient operation of the system with OFT disabled (Figure 8a) and enabled (Figure 8b). In the former case, the coils are placed at a distance of 20 mm, and the power supply is regulated to obtain 25 V across the load (100 kΩ). If the coils are brought closer to a distance of 5 mm, the voltage across the load will drop to 12.5 V (in accordance with Figure 6a). The inset sections labelled A and B in Figure 8a show the profile of VDIV, which represents the current in the primary coil. The amplitude of VDIV drops from 5Vpk-pk to 2Vpk-pk when the coils are moved closer with OFT disabled. This drop in the amplitude will track the optimum frequency to keep the load voltage maximised. As the coils are brought closer together, the load voltage increases to 30 V after a short settling period. The ringing visible in Figure 8b occurs due to a combination of mechanical vibration of the coil displacement jig and overshoot in the control circuit. This behaviour is in accordance with the results in Figure 6a, where the gain increases as the link spacing decreases when OFT is enabled.

Figure 8.

Figure 8

Scope screenshots to illustrate variations in VOUT in fixed mode and OFT mode as the coils are brought closer together. (a) OFT disabled; (b) OFT enabled.

Figure 9 shows a zoomed version of the inset sections labelled A and B in Figure 8. This highlights the variation in phase difference between VDIV and VPLL as the link is brought out of resonance by reducing the distance between the coils. The traces in Figure 9a show a phase difference ϕ of 95 between VDIV and VPLL, and this indicates resonance in the link; when the phase between VDIV and VPLL is 90, the phase between the primary current and primary voltage is at 0. The traces in Figure 9b show the effect of driving the link in a fixed mode such that ωωopt. This manifests in a phase difference of ϕ=180 and a drop in the amplitude of VDIV as the coils are moved closer together.

Figure 9.

Figure 9

Close-up view of the labelled areas A and B in Figure 8, illustrating the difference between VPLL and VDIV while near resonance and far from resonance. (a) Near resonance (ϕ90); (b) Far from resonance (ϕ180).

Together Figure 8 and Figure 9 show how, by employing OFT, the output voltage is maintained at a high level even as the coils are displaced.

5. Conclusions

This paper has presented the implementation and operation of a closed-loop OFT system capable of compensating for frequency splitting and therefore improving the resilience and system efficiency of inductive WPT systems. The system requires no variable capacitors or inductors, and is suitable for compact systems where space is valuable, such as for IMDs. The system is capable of adapting to changes in coil coupling and load conditions, maintaining a constant output voltage and improved system efficiency compared to fixed frequency equivalents.

Measurements have shown close matching with the predicted behaviour of the system. Discrepancies were only observed at very small distances between the coils, where capacitative coupling becomes a relevant factor. These effects are currently under analytical and experimental investigation. This system is suited to powering implanted medical devices, particularly those in which the receiver moves significantly with respect to the transmitter, and has time variant power requirements, such as in endoscopic capsules [30]. This approach could also be useful in the case of distributed implants [2], where the distances between the transmitter and the receivers can be quite variable. Since OFT requires no back-telemetry to operate, this work is suitable for systems with very little complexity at the receiver, e.g. for direct inductive stimulation [31]. Additionally, in the case of implants employing flexible coils, the proposed OFT system would be capable of adapting to the altered link impedance presented by coil flexion [16] as a result of its ability to drive across a wide range of frequencies. OFT could also be used to interrogate remote units to establish the deviation in their true resonant frequencies from the designed value, or operate multiple receivers tuned at different frequencies [32]. Such an interrogation technique could also be used to confirm each receiver’s post-implantation optimum frequency.

Table 2 provides some recent examples of comparable biomedical WPT systems to contrast with the work presented here, highlighting the ability of this work to compensate for load and spacing changes, regardless of component variations, without requiring any data telemetry.

Table 2.

Comparison of this work to some recent examples of comparable biomedical WPT systems.

[13] [12] [19] [31] This Work
Operating Frequency 1 MHz 13.56 MHz 13.56 MHz 10 MHz Variable (4.5–5.5 MHz)
Effective Range 15 mm 20 mm 20 mm N/A 30 mm
Compensation Parameter Supply Voltage Supply Voltage C1, C2, and Supply Voltage None Drive Frequency
System Efficiency (Best) 65.8% 14% 14% N/A 15%
Component Mismatch Insensitive No No Yes No Yes
Passive Implant Compatible No No No Yes Yes

Acknowledgments

This work was supported by a Ph.D. scholarship awarded to Matthew Schormans from University College London, and the European Community’s Seventh Framework Program [FP7/2007-2013] under Grant 258654 (Project NEUWalk; www.neuwalk.eu).

Abbreviations

The following abbreviations are used in this manuscript:

WPT

Wireless Power Transfer

IMD

Implanted Medical Device

OFT

Optimum Frequency Tracking

PA

Power Amplifier

ED

Envelope Detector

PD

Phase Detector

LC

Lock Comparator

PLL

Phase-Locked Loop

EA

Error Amplifier

ADC

Analogue to Digital Converter

Author Contributions

Matthew Schormans conceived and designed the system, performed the experiments, collected and compared the analytical and experimental data, and wrote the paper; Virgilio Valente designed and provided the integrated power amplifier, supervised the system design, experiments, and data analysis, and wrote the paper; Andreas Demosthenous supervised the system design and experiment design, and revised and edited the paper.

Conflicts of Interest

The authors declare no conflict of interest.

References

  • 1.Rhew H.G., Jeong J., Fredenburg J.A., Dodani S., Patil P.G., Flynn M.P. A fully self-contained logarithmic closed-loop deep brain stimulation SoC with wireless telemetry and wireless power management. IEEE J. Solid-St. Circ. 2014;49:2213–2227. doi: 10.1109/JSSC.2014.2346779. [DOI] [Google Scholar]
  • 2.Lee B., Ahn D., Ghovanloo M. Three-phase time-multiplexed planar power transmission to distributed implants. IEEE J. Emerg. Sel. Top. Power Electron. 2015;4:263–272. doi: 10.1109/JESTPE.2015.2436391. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Stoecklin S., Yousaf A., Volk T., Reindl L. Efficient wireless powering of biomedical sensor systems for multichannel brain implants. IEEE Trans. Instrum. Meas. 2015;65:754–764. doi: 10.1109/TIM.2015.2482278. [DOI] [Google Scholar]
  • 4.Moradi E., Björninen T., Sydänheimo L., Rabaey J.M. Analysis of wireless powering of mm-size neural recording tags in rfid-inspired wireless brain- machine interface systems; Proceedings of the IEEE International Conference on RFID; Penang, Malaysia. 30 April–2 May 2013; pp. 8–15. [Google Scholar]
  • 5.Lin Y.P., Yeh C.Y., Huang P.Y., Wang Z.Y., Cheng H.H., Li Y.T., Chuang C.F., Huang P.C., Tang K.T., Ma H.P., et al. A Battery-Less, Implantable Neuro-Electronic Interface for Studying the Mechanisms of Deep Brain Stimulation in Rat Models. IEEE Trans. Biomed. Circuits Syst. 2015;10:98–112. doi: 10.1109/TBCAS.2015.2403282. [DOI] [PubMed] [Google Scholar]
  • 6.Cleven N.J., Müntjes J.A., Fassbender H., Urban U., Görtz M., Vogt H., Gräfe M., Göttsche T., Penzkofer T., Schmitz-Rode T., et al. A novel fully implantable wireless sensor system for monitoring hypertension patients. IEEE Trans. Biomed. Eng. 2012;59:3124–3130. doi: 10.1109/TBME.2012.2216262. [DOI] [PubMed] [Google Scholar]
  • 7.Chang C.W., Chiou J.C. A wireless and batteryless microsystem with implantable grid electrode/3-dimensional probe array for ECoG and extracellular neural recording in rats. Sensors. 2013;13:4624–4639. doi: 10.3390/s130404624. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Laskovski A.N., Yuce M.R., Dissanayake T. Stacked spirals for biosensor telemetry. IEEE Sens. J. 2011;11:1484–1490. doi: 10.1109/JSEN.2010.2091123. [DOI] [Google Scholar]
  • 9.Kang S.K., Murphy R.K.J., Hwang S.W., Lee S.M., Harburg D.V., Krueger N.A., Shin J., Gamble P., Cheng H., Yu S., et al. Bioresorbable silicon electronic sensors for the brain. Nature. 2016;530:71–76. doi: 10.1038/nature16492. [DOI] [PubMed] [Google Scholar]
  • 10.Niu W.Q., Chu J.X., Gu W., Shen A.D. Exact analysis of frequency splitting phenomena of contactless power transfer systems. IEEE Trans. Circuits Syst. I Regul. Pap. 2013;60:1670–1677. doi: 10.1109/TCSI.2012.2221172. [DOI] [Google Scholar]
  • 11.Bocan K., Sejdić E. Adaptive transcutaneous power transfer to implantable devices: A state of the art review. Sensors. 2016;16:393. doi: 10.3390/s16030393. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Kiani M., Ghovanloo M. An RFID-based closed-loop wireless power transmission system for biomedical applications. IEEE Trans. Circ. Syst. II Exp. Briefs. 2010;57:260–264. doi: 10.1109/TCSII.2010.2043470. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Wang G., Liu W., Sivaprakasam M., Kendir G.A. Design and analysis of an adaptive transcutaneous power telemetry for biomedical implants. IEEE Trans. Circuits Syst. I Regul. Pap. 2005;52:2109–2117. doi: 10.1109/TCSI.2005.852923. [DOI] [Google Scholar]
  • 14.Li X., Tsui C.Y., Ki W.H. A 13.56 MHz wireless power transfer system with reconfigurable resonant regulating rectifier and wireless power control for implantable medical devices. IEEE J. Solid-St. Circ. 2015;50:978–989. doi: 10.1109/JSSC.2014.2387832. [DOI] [Google Scholar]
  • 15.Lyu Y.I., Meng F.Y., Yang G.H., Che B.J., Wu Q., Sun L., Erni D., Li J.L.W. A method of using nonidentical resonant coils for frequency splitting elimination in wireless power transfer. IEEE Trans. Power Electron. 2015;30:6097–6107. doi: 10.1109/TPEL.2014.2387835. [DOI] [Google Scholar]
  • 16.Lenaerts B., Peeters F., Puers R. Closed-loop transductor-compensated Class-E driver for inductive links; Proceedings of the TRANSDUCERS and EUROSENSORS 2007–4th International Conference on Solid-State Sensors, Actuators and Microsystems; Lyon, France. 10–14 June 2007; pp. 65–68. [Google Scholar]
  • 17.Aldhaher S., Chi-kwong Luk P., Whidborne J.F. Tuning class E inverters applied in inductive links using saturable reactors. IEEE Trans. Power Electron. 2014;29:2969–2978. doi: 10.1109/TPEL.2013.2272764. [DOI] [Google Scholar]
  • 18.Si P., Hu A.P., Malpas S., Budgett D. A frequency control method for regulating wireless power to implantable devices. IEEE Trans. Biomed. Circuits Syst. 2008;2:22–29. doi: 10.1109/TBCAS.2008.918284. [DOI] [PubMed] [Google Scholar]
  • 19.Lee B., Kiani M., Ghovanloo M. A triple-loop inductive power transmission system for biomedical applications. IEEE Trans. Biomed. Circuits Syst. 2015;10:138–148. doi: 10.1109/TBCAS.2014.2376965. [DOI] [PubMed] [Google Scholar]
  • 20.Hannan M.A., Hussein H.A., Mutashar S., Samad S.A., Hussain A. Automatic frequency controller for power amplifiers used in bio-implanted applications: Issues and challenges. Sensors. 2014;14:23843–23870. doi: 10.3390/s141223843. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Schormans M., Valente V., Demosthenous A. Efficiency optimization of class-D biomedical inductive wireless power transfer systems by means of frequency adjustment; Proceedings of the 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Milan, Italy. 25–29 August 2015; pp. 5473–5476. [DOI] [PubMed] [Google Scholar]
  • 22.Sample A.P., Meyer D.A., Smith J.R. Analysis, experimental results, and range adaptation of magnetically coupled resonators for wireless power transfer. IEEE Trans. Ind. Electron. 2011;58:544–554. doi: 10.1109/TIE.2010.2046002. [DOI] [Google Scholar]
  • 23.Valente V., Eder C., Donaldson N., Demosthenous A. A high-power CMOS class-D amplifier for inductive-link medical transmitters. IEEE Trans. Power Electron. 2014;30:1–12. doi: 10.1109/TPEL.2014.2353214. [DOI] [Google Scholar]
  • 24.Terman F.E. Radio Engineers’ Handbook. McGraw-Hill Book Company; New York, NY, USA: 1943. [Google Scholar]
  • 25.Van Schuylenbergh K., Puers R., editors. Inductive Powering. Springer Netherlands; Dordrecht, The Netherlands: 2009. [Google Scholar]
  • 26.Donaldson N.D.N., Perkins T.A. Analysis of resonant coupled coils in the design of radio frequency transcutaneous links. Med. Biol. Eng. Comput. 1983;21:612–627. doi: 10.1007/BF02442388. [DOI] [PubMed] [Google Scholar]
  • 27.Jow U.M.M., Ghovanloo M. Design and optimization of printed spiral coils for efficient inductive power transmission; Proceedings of the IEEE International Conference on Electronics, Circuits, and Systems; Marrakech, Morocco. 11–14 December 2007; pp. 70–73. [Google Scholar]
  • 28.Chandrakasan A.P., Verma N., Daly D.C. Ultralow-power electronics for biomedical applications. Ann. Rev. Biomed. Eng. 2008;10:247–274. doi: 10.1146/annurev.bioeng.10.061807.160547. [DOI] [PubMed] [Google Scholar]
  • 29.Xu H., Handwerker J., Ortmanns M. Telemetry for implantable medical devices. IEEE Solid-St. Circ. Mag. 2014;6:60–63. doi: 10.1109/MSSC.2014.2327714. [DOI] [Google Scholar]
  • 30.Lenaerts B., Puers R. An inductive power link for a wireless endoscope. Biosens. Bioelectron. 2007;22:1390–1395. doi: 10.1016/j.bios.2006.06.015. [DOI] [PubMed] [Google Scholar]
  • 31.Ha S., Khraiche M.L., Silva G.A., Cauwenberghs G. Direct inductive stimulation for energy-efficient wireless neural interfaces; Proceedings of the 2012 IEEE-EMBS International Conference on Biomedical and Health Informatics; Hong Kong, China. 5–7 January 2012; pp. 883–886. [DOI] [PubMed] [Google Scholar]
  • 32.Mirbozorgi S.A., Bahrami H., Sawan M., Gosselin B. A smart multi-receiver power transmission system for long-term biological monitoring; Proceedings of the 2014 IEEE Biomedical Circuits and Systems Conference (BioCAS); Lausanne, Switzerland. 22–24 October 2014; pp. 412–415. [Google Scholar]

Articles from Sensors (Basel, Switzerland) are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES