High-Resolution Time-Frequency Spectrum-Based Lung Function Test from a Smartphone Microphone

Tharoeun Thap 1, Heewon Chung 1, Changwon Jeong 2, Ki-Eun Hwang 3, Hak-Ryul Kim 3, Kwon-Ha Yoon 4,*, Jinseok Lee 1,*
Editor: Ki H Chon
PMCID: PMC5017470  PMID: 27548164

Abstract

In this paper, a smartphone-based lung function test, developed to estimate lung function parameters using a high-resolution time-frequency spectrum from a smartphone built-in microphone is presented. A method of estimation of the forced expiratory volume in 1 s divided by forced vital capacity (FEV1/FVC) based on the variable frequency complex demodulation method (VFCDM) is first proposed. We evaluated our proposed method on 26 subjects, including 13 healthy subjects and 13 chronic obstructive pulmonary disease (COPD) patients, by comparing with the parameters clinically obtained from pulmonary function tests (PFTs). For the healthy subjects, we found that an absolute error (AE) and a root mean squared error (RMSE) of the FEV1/FVC ratio were 4.49% ± 3.38% and 5.54%, respectively. For the COPD patients, we found that AE and RMSE from COPD patients were 10.30% ± 10.59% and 14.48%, respectively. For both groups, we compared the results using the continuous wavelet transform (CWT) and short-time Fourier transform (STFT), and found that VFCDM was superior to CWT and STFT. Further, to estimate other parameters, including forced vital capacity (FVC), forced expiratory volume in 1 s (FEV1), and peak expiratory flow (PEF), regression analysis was conducted to establish a linear transformation. However, the parameters FVC, FEV1, and PEF had correlation factor r values of 0.323, 0.275, and −0.257, respectively, while FEV1/FVC had an r value of 0.814. The results obtained suggest that only the FEV1/FVC ratio can be accurately estimated from a smartphone built-in microphone. The other parameters, including FVC, FEV1, and PEF, were subjective and dependent on the subject’s familiarization with the test and performance of forced exhalation toward the microphone.

Keywords: smartphone microphone, pulmonary function test, FEV1/FVC, COPD, high-resolution time-frequency

1. Introduction

Lung diseases are some of the most common medical conditions [1]. Many people suffer from lung diseases stemming from smoking, infections, and genetics. The lungs are the primary organs expanding and relaxing thousands of times each day to bring in oxygen and push out carbon dioxide. As one inhales, the breath travels through the mouth or nose to the windpipe and lungs, and further moves through smaller tubes to deliver oxygen throughout the organ. A number of lung diseases adversely affect lung function by affecting the airways. These include asthma, chronic obstructive pulmonary disease (COPD), chronic bronchitis, emphysema, acute bronchitis, and cystic fibrosis [2].

For lung function evaluation, pulmonary function tests (PFTs) have found widespread clinical use. PFTs can help diagnose asthma, allergies, chronic bronchitis, respiratory infections, lung fibrosis, bronchiectasis, and COPD by evaluating how well the lungs take in and release air. For basic PFTs, a spirometer is the main piece of equipment, and an important clinical diagnostic device that should be used by all primary care and most specialist physicians [3]. Indeed, it was reported that 66% of primary care offices owned and used their own spirometer at the primary contact point. This low amount should reflect, in part, the need for alternatives to the spirometer [4,5]. A standard spirometer measures forced expiratory flow and displays a volume-time curve and a flow–volume loop (or curve), as shown in Figure 1. From the measurements, the following parameters are quantized: the forced vital capacity (FVC); the forced expiratory volume in one second (FEV1); the FEV1/FVC ratio; the peak expiratory flow (PEF); the forced expiratory flow at 25%, 50%, or 75% of FVC (FEF25, FEF50, and FEF75); and the forced expiratory flow between 25% and 75% (FEF25%–75%) [6,7,8]. All of these parameters are clinically important in assessing lung function.

Figure 1.

Figure 1

Examples of forced expiratory flow measure, (a) volume-time plot; and (b) volume–flow plot [7].

Among these parameters, the most common clinical measures are FVC, FEV1, FEV1/FVC, and PEF, as they are used to quantify the degree of airflow limitation in chronic lung diseases, such as asthma, COPD, and cystic fibrosis. The FEV1/FVC ratio has been widely accepted, especially for assessing COPD, which obstructs the main airway. In 1983, the European Community for Coal and Steel (ECCS) initially defined airway obstruction with an FEV1/FVC ratio below the lower fifth percentile of a large healthy reference group, which is the statistically-defined lower limit of normal (LLN) [9]. The same definition was followed by the American Thoracic Society (ATS) in 1991 [10], the European Respiratory Society (ERS) in 1993 [11], and the National Lung Health Education Program (NLHEP) in 2000 [12]. Depending on the organization and year, the definition has occasionally been modified by using a fixed value instead of the LLN. In 1987, ATS defined airway obstruction with an FEV1/FVC ratio lower than the fixed value of 0.75 [13], followed by the British Thoracic Society (BTS) with 0.70 in 1997 [14], the National Institute for Health and Clinical Excellence (NICE) with 0.75 in 2004 [15], the ATS and ERS with 0.70 in 2004 [16], and the Global Initiative for Chronic Obstructive Lung Disease (GOLD) with 0.75 in 2007 [15]. In this way, the FEV1/FVC ratio has been a widely accepted standard for assessing COPD, as summarized in Table 1 [15].

Table 1.

Criteria for assessing COPD according to various organizations [15].

Organization Year Criterion/Reference
ECCS 1983 FEV1/FVC < LLN [9]
ATS 1987 FEV1/FVC < 0.75 [13]
ATS 1991 FEV1/FVC < LLN [10]
ECCS/ERS 1993 FEV1/FVC < LLN [11]
BTS 1997 FEV1/FVC < 0.70 and FEV1 < 80% predicted [14]
NLHEP 2000 FEV1/FVC < LLN and FEV1 < LLN [12]
NICE 2004 FEV1/FVC < 0.75 and FEV1 < 80% predicted [15]
ATS/ERS 2004 FEV1/FVC < 0.70 post-bronchodilator [16]
GOLD 2007 FEV1/FVC < 0.75 post-bronchodilator [15]

Home spirometry may provide the convenient way for patients who live great distances from their clinics and research facilities. It also allows patients to monitor more frequently for detecting changes in lung function, which results in in earlier treatment of exacerbations, more rapid recovery, reduced healthcare cost, and improved outcomes [17,18]. Even though there are portable spirometers available, they are expensive, which limits the popularity to many people. Since the smartphone is popular and capable for users to measure forced exhalation with its built-in microphone, it can be a cheap home spirometer available to smartphone users. In addition, since smartphone technology has the capacity for users to communicate with physicians (e.g., using cloud technology), smartphone-based home spirometry has the potential in providing early diagnosis of lung disease for those who may have breathing problems on a day-to-day basis [19,20,21,22,23]. Recently, a smartphone-based approach that measures lung function using the built-in microphone has been introduced [24]. In their pilot study, the researchers demonstrated the extraction of the FVC, FEV1, PEF, and FEV1/FVC parameters. However, the approach used the spectrogram of the signal based on the short-time Fourier transform (STFT) with 30 ms frames. The STFT has a limitation in its time-frequency resolution capability, which is due to the uncertainty principle. The Heisenberg-Gabor inequality states that the time–bandwidth product of a signal is the lower bound by some constant. This means that a signal cannot be simultaneously narrow in time and frequency domains. These limitations in resolution were one of the reasons for the invention of wavelet theory. Recently, the variable frequency complex demodulation method (VFCDM) has shown a higher resolution than any other time-frequency spectrum methods, such as the smoothed pseudo Wigner-Ville (SPWV), short-time Fourier transform (STFT), and the wavelet transform (WT) methods [25,26,27,28].

In this paper, VFCDM-based lung function parameter estimation is presented by using an audio signal recorded from a smartphone built-in microphone. The estimation method of the FEV1/FVC ratio is first presented, and the accuracy of the VFCDM is then compared with that of CWT and STFT. The main contribution is to introduce the smartphone built-in microphone-based FEV1/FVC ratio estimation by using the high-resolution time-frequency spectrum, VFCDM, which accurately detects the main time-varying resonance frequency. To evaluate our proposed method, we compared the estimated FEV1/FVC ratio with the parameters clinically obtained from PFTs. In addition, we evaluated the other parameter estimation of FVC, FEV1, and PEF.

2. Materials and Methods

2.1. Subjects

Twenty-six subjects took part in this study, including 13 healthy subjects and 13 COPD patients. The healthy subjects (10 males, three females), with ages ranging from 24–47 years (30.85 ± 7.74; mean ± STD), had an average weight of 68.15 ± 16.01 kg and an average height of 170.23 ± 9.18 cm. The COPD patients (10 males, three females), with ages ranging from 51–89 years (71.92 ± 10.49), had an average weight of 61.00 ± 10.54 kg and an average height of 163.00 ± 6.30 cm. They were all severe COPD. Table 2 lists the subject details including their average lung function parameters. The study protocol and data analysis were approved by the Institutional Review Board (IRB) at Wonkwang University Hospital (WKUH), and all subjects consented to participate in the experiment. Since the forced expiration could aggravate certain medical conditions, individuals with the following medical conditions were excluded: hemoptysis of unknown origin; pneumothorax; angina; recent myocardial infarction; thoracic, abdominal, and cerebral aneurysms; cataracts or recent eye surgery; recent thoracic or abdominal surgery; nausea; vomiting; acute illness; and recent viral infection.

Table 2.

Subjects’ information.

Subjects Healthy Patient Total
(n = 13) (n = 13) (n = 26)
Gender (n) Male 10 Male 10 Male 20
Female 3 Female 3 Female 6
Age (years) 30.85 ± 7.74 71.92 ± 10.49 51.35 ± 22.85
Height (cm) 170.23 ± 9.18 163.00 ± 6.30 166.58 ± 8.51
Weight (kg) 68.15 ± 16.01 61.00 ± 10.54 64.58 ± 13.77
FVC (Liters) 4.89 ± 1.09 2.64 ± 0.82 3.77 ± 1.49
FEV1 (Liters) 4.02 ± 0.84 1.25 ± 0.52 2.63 ± 1.57
PEF (Liter/s) 10.62 ± 3.09 2.97 ± 1.06 6.80 ± 4.51
FEV1/FVC (%) 87.77 ± 6.61 48.31 ± 15.32 65.54 ± 21.03

Values are means ± standard deviations.

2.2. Data Collection and Preprocessing

The participants were instructed to follow the rules before the spirometry test [29]:

  • Do not smoke during the hour before the test;

  • Do not drink alcohol during the four hours before the test;

  • Do not eat a large meal during the two hours before the test;

  • Please wear loose clothing;

  • Do not perform vigorous exercise within 30 min of the test; and

  • If on puffer (inhaler) medication, you may be asked to refrain from taking it for a few hours before the test.

The audio data was measured and collected using an iPhone 5S (manufactured by Apple Inc., Cupertino, CA, USA). The subjects’ forced exhalation sounds were recorded using the built-in microphone and converted into digital audio signals at a sampling rate of 44,100 Hz. The subjects were required to breathe in their full lung capacity and then forcefully exhale as much air from the lungs as possible through the mouthpiece (normally taking over 6 s). The mouthpiece was attached to the iPhone microphone using a custom-made 3D-printed adapter (see Figure 2a,b). To achieve the spirometry measurement, much effort has been carried out on the vortex whistle [30,31,32,33]. We modified the vortex whistle available for the smartphone built-in microphone. The mouthpiece used in this experiment was VBMax PFT Filter 33 mm from A-M Systems (Sequim, WA, USA). The adapter was printed with polylactic acid (PLA) material. It is an end-opened device which allows the air to pass through to the microphone and exit freely at the end. Rubber material was inserted between the mouthpiece and an adapter to prevent air leakage. The mouthpiece plays an important role because it essentially helps to maintain lip posture, fixes the distance between the lips and the microphone, and reduces surrounding background noise by guiding the airflow directly towards the microphone. Furthermore, it attaches and detaches easily to an iPhone 5S for an older person, and enables the test to be performed in the same manner as that performed with clinical equipment at a hospital. The clinical test was performed at Wonkwang University Hospital and the smartphone-based test was performed in an open-space environment. Participants performed the clinical tests followed by smartphone-based test under two trained pulmonologists. Performing clinical test at first allows participants to familiarize with the smartphone-based test. Participants performed three trial tests with the smartphone, as shown in Figure 2c. Each test was completely effort-dependent, and each participant was instructed by a trained specialist using gestures. The original audio signals with a sampling rate of 44,100 Hz were down-sampled to 2450 Hz, and a seventh-order elliptic low-pass filter with a cutoff frequency of 800 Hz was applied to remove unwanted signal components. These parameters were found based on the prior-evaluation tests under a very quiet room without any noise interference. In addition, to reduce the computational complexity, we down-sampled the signal to 2450 Hz, which still preserved the recorded data. Among the three trials from the smartphone-based test, the parameter set was selected based on the Morris/Polgar standards [34,35,36,37] under the two trained pulmonologists. The participants also performed three trial tests with clinical equipment as a reference, with some times between tests to allow recovery to a rested condition. The pulmonary function test (PFT) data was captured using Vmax software version IVS-0101-21-2B (manufactured by CareFusion Corporation, San Diego, CA, USA), which automatically provided the PFT parameters. Each participant performing PFT was seated in the cabin of a Vmax Autobox (Body Plethysmography System, V62J) manufactured by CareFusion, and performed the test.

Figure 2.

Figure 2

Audio data acquisition setup: (a) 3D-printing mouthpiece-to-iPhone adapter; (b) mouthpiece equipped with adapter; and (c) complete set of experimental and data acquisition equipment.

2.3. Resonance Frequency Estimation Using High Resolution Time-Frequency Spectrum

Using the audio signal, we analyzed the dominant frequency of the forced exhalation sound by using the variable frequency complex demodulation (VFCDM) method [26]. The VFCDM method has shown higher resolution than any other time-frequency spectrum methods, such as smoothed pseudo Wigner-Ville (SPWV), short-time Fourier transform (STFT) and wavelet transform (WT) methods [25,26,27,28]. Considering a sinusoidal signal x(t) to be a narrow band oscillation with an instantaneous amplitude A(t), a center frequency f0, a phase ϕ(t), and a direct current component dc(t):

x(t)=dc(t)+A(t)cos(2πf0t+φ(t)) (1)

For a given center frequency, information on the instantaneous amplitude A(t) and phase ϕ(t) are extracted by multiplying Equation (1) by ej2πf0t, providing the following formula:

z(t)=x(t)ej2πf0t=dc(t)ej2πf0t+A(t)2ejφ(t)+A(t)2ej(4πf0t+φ(t)) (2)

The shift by ej2πf0t results in moving the center frequency, f0, to zero frequency in the spectrum of z(t). If z(t) in Equation (2) is subjected to an ideal low-pass filter (LPF) with a cutoff frequency fc < f0, the filtered signal zlp(t) contains only the component of interest, and the following formulae are obtained:

zlp(t)=A(t)2ejφ(t) (3)
A(t)=2|zlp(t)| (4)
φ(t)= dc(t)+A(t)cos(0t2πf(τ)dτ+φ(t)) (5)

The acoustic signal x(t) contains the modulating frequency, which changes according to time. Therefore, x(t) can be formulated as:

x(t)=dc(t)+A(t)cos(0t2πf(τ)dτ+φ(t)) (6)

Similarly, multiplying Equation (6) by ej0t2πf(τ)dτ provides the instantaneous amplitude, A(t), and instantaneous phase, ϕ(t):

z(t)=x(t)ej0t2πf(τ)dτ=dc(t)ej0t2πf(τ)dτ+A(t)2ejφ(t)+A(t)2ej(0t4πf(τ)dτ+φ(t)) (7)

From Equation (7), when z(t) is filtered with an ideal LPF with a cutoff frequency fc < f0, the filtered signal zlp(t) contains the same instantaneous amplitude A(t) and phase ϕ(t) as in Equations (4) and (5). Details regarding the VFCDM algorithm are described in [26].

Figure 3 shows an example of the main procedure to obtain the volume-flow curve and time-volume curve from the audio signal, as well as the comparison among the resultant time-frequency spectra of VFCDM, CWT, and STFT. Figure 3a shows the original audio signal with a sampling rate of 44,100 Hz from the built-in microphone. The original audio signal was then down-sampled to 2450 Hz, and the seventh-order elliptic LPF with a cutoff frequency of 800 Hz was applied. Figure 3bd show the resultant time-frequency spectra using VFCDM, CWT, and STFT, respectively. The filter parameters of the VFCDM were with the bandwidth Fw = 0.03 Hz (normalized frequency) and the length of the filter Nω = 64. For CWT, the Morlet wavelet scalogram was chosen with the lowest and highest normalized frequencies set to 0.01 Hz and 0.5 Hz. Morlet wavelet transform allows multi-resolution analysis in the time-frequency domain of a non-stationary, transient signal. For STFT, the data was buffered into 30 ms frames and shifted by every sample. Each frame is then windowed using a Hamming window and then computed the magnitude squared of the FFT to produce a spectrogram. In Figure 3b,d the frequencies providing the maximum power for each sample, with an interval of 1/2450 s, are also plotted. These show that VFCDM provides a higher resolution in both time and frequency than CWT and STFT. With the time-frequency spectrum from VFCDM, we plotted maximum power versus accumulated maximum power, as shown in Figure 3e, which shows a similar pattern to the flow-volume curve depicted in Figure 1b. In addition, we plotted the accumulated maximum power versus the sample (time), as shown in Figure 3f, which shows a similar pattern to the volume-time curve depicted in Figure 1a.

Figure 3.

Figure 3

Figure 3

Main procedure to obtain flow-volume curve and volume-time curve from healthy subjects and comparison of VFCDM, CWT, and STFT: (a) original audio signal; (b) resultant time-frequency spectrum from VFCDM; (c) resultant time-frequency spectrum from CWT; (d) resultant time-frequency spectrum from STFT; (e) maximum power vs. accumulated maximum power from VFCDM; and (f) accumulated maximum power vs. time from VFCDM.

2.4. Instantaneous Power and Performance Evaluation

VFCDM, CWT, and STFT were applied for the estimation of the FEV1/FVC ratio for healthy subjects (N = 13) and COPD patients (N = 13). To investigate the effect of instantaneous power, the summation of power values equal to or greater than a varying percentage THP of the maximum power at each sample was considered. THP was altered to 50%, 70%, 80%, 90%, and 100%, and the estimated FEV1/FVC ratio was evaluated. Using the power summation for each sample, power summation vs. accumulated power summation was plotted representing the flow–volume curve, and time vs. accumulated power summation was plotted representing the volume-time curve.

To further investigate the VFCDM based estimation of the other lung function parameters including FVC, FEV1, and PEF, the values of total power, total power in 1 s, and global maximum power were extracted, each of which corresponds to FVC, FEV1, and PEF, respectively. The extracted parameters are denoted by SFVC, SFEV1, and SPEF. It was assumed that SFVC, SFEV1, and SPEF were related by the following linear transformation:

TFVCSFVC=KFVC·EFVC+CFVC (8)
TFEV1SFEV1=KFEV1·EFEV1+CFEV1 (9)
TPEFSPEF=KPEF·EPEF+CPEF (10)

where TFVC, TFEV1, and TPEF are the reference values, and SFVC, SFEV1, and SPEF are the estimated values of FVC, FEV1, and PEF, respectively. All the reference values were obtained from the clinical results. EFVC is the summation of maximum power (THP = 100%) for the entire recorded samples, EFEV1 is the summation of maximum power for the first one second, and EPEF is the global maximum power in the entire recorded samples; KFVC, KFEV1, and KPEF are the scaling constants; and CFVC, CFEV1, and CPEF are offset constants. For each equation, we have a dependent variable corresponding to a reference value and an independent variable corresponding to a predictor. Then, the relationship with the constant values between a dependent variable (TFVC, TFEV1, or TPEF) and an independent variable (EFVC, EFEV1, or EPEF) can be estimated by linear least squares regression. Statistical analyses were performed using IBM SPSS Statistics for Windows, version 22.0 (manufactured by IBM Corp., Armonk, NY, USA, 2012). Pearson’s correlation coefficients were used for performing linear regression analysis. For comparison with the correlation coefficient from the FEV1/FVC ratio, we similarly formulated the linear transformation as:

TFEV1/FVCSFEV1/FVC=KFEV1/FVC·EFEV1/FVC+CFEV1/FVC (11)

where EFEV1/FVC=EFEV1/EFVC.

For the evaluation, we used absolute error (AE) and root mean squared error (RMSE) defined by:

Absolue error (AE)= |EFEV1/FVC TFEV1/FVC| (12)
Root Mean Squared Error (RMSE)= (EFEV1/FVC TFEV1/FVC)2N (13)

where N is the number of subjects; and EFEV1/FVC and TFEV1/FVC are the estimated FEV1/FVC ratio and its reference value, respectively.

3. Results

3.1. Estimation of FEV1/FVC Ratio

Figure 4a–c show the absolute error distributions of the estimated FEV1/FVC ratios from healthy subjects according to THP based on VFCDM, CWT and STFT, respectively. The central red circle mark is the median, the lower and upper whiskers are the 25th and 75th percentiles, the blue square marks are the 10th and 90th percentiles, and the blue diamond marks are the 5th and 95th percentiles. A significant difference between VFCDM and CWT at p < 0.01 was found based on the t-test. Similarly, a significant difference between VFCDM and STFT was found. In addition, it was found that the absolute error decreased as THP increased up to 100% for all VFCDM, CWT, and STFT. Thus, VFCDM with THP = 100% provided the best accuracy for the FEV1/FVC ratio estimation. Table 3 summarizes the absolute error (AE) and RMSE values from healthy subjects according to THP for VFCDM, CWT, and STFT, respectively. It shows that RMSE from VFCDM with THP = 100% is 2.07 and 2.19 times lower than the CWT and STFT with THP = 100%, respectively. For mean AE, VFCDM with THP = 100% provided 2.39 and 2.57 times lower than CWT and STFT with THP = 100%, respectively.

Figure 4.

Figure 4

Absolute error distribution of FEV1/FVC from healthy subjects according to THP based on (a) VFCDM; (b) CWT; and (c) STFT.

Table 3.

Healthy subjects’ absolute errors (AE) and RMSEs of FEV1/FVC ratio according to THP for VFCDM, CWT, and STFT methods.

THP 50% 70% 80% 90% 100%
AE RMSE AE RMSE AE RMSE AE RMSE AE RMSE
VFCDM (N = 13) 4.29 ± 4.44 6.05 4.44 ± 3.94 5.84 4.54 ± 3.72 5.78 4.65 ± 3.39 5.68 4.49 ± 3.38 5.54
CWT (N = 13) 11.76 ± 4.09 12.40 11.65 ± 4.10 12.30 11.60 ± 4.10 12.25 11.55 ± 4.10 12.20 10.77 ± 4.14 11.48
STFT (N = 13) 11.86 ± 4.18 12.52 11.75 ± 4.18 12.42 11.71 ± 4.19 12.38 11.67 ± 4.2 12.35 11.43 ± 4.31 12.16

Figure 5a–c show the absolute error distributions of the estimated FEV1/FVC ratios from COPD patients according to THP based on VFCDM, CWT and STFT, respectively. A significant difference between VFCDM and CWT at p < 0.01 was found based on the t-test. Similarly, a significant difference between VFCDM and STFT was found. Table 4 summarizes the absolute error (AE) and RMSE values from COPD subjects according to THP for VFCDM, CWT, and STFT, respectively. It shows that RMSE from VFCDM with THP = 100% is 1.67 and 1.58 times lower than the CWT and STFT with THP = 100%, respectively. For mean AE, VFCDM with THP = 100% provided 2.03 and 1.91 times lower than CWT and STFT with THP = 100%, respectively.

Figure 5.

Figure 5

Absolute error distribution of FEV1/FVC from COPD patients according to THP based on (a) VFCDM; (b) CWT; and (c) STFT.

Table 4.

COPD patients’ absolute errors (AE) and RMSEs of FEV1/FVC ratio according to THP for VFCDM, CWT, and STFT methods.

THP 50% 70% 80% 90% 100%
AE RMSE AE RMSE AE RMSE AE RMSE AE RMSE
VFCDM (N = 13) 10.01 ± 10.56 14.26 10.29 ± 10.57 14.46 10.25 ± 10.63 14.47 10.17 ± 10.23 14.15 10.30 ± 10.59 14.48
CWT (N = 13) 18.85 ± 13.14 22.69 18.87 ± 12.79 22.52 18.85 ± 12.59 22.39 18.88 ± 12.39 22.32 20.93 ± 12.55 24.15
STFT (N = 13) 19.43 ± 11.95 22.56 19.37 ± 11.86 22.48 19.33 ± 11.90 22.47 19.34 ± 11.98 22.51 19.63 ± 12.14 22.83

Table 5 summarizes the whole estimated FEV1/FVC ratios from each subject at THP = 100% for VFCDM, CWT, and STFT. For healthy subjects and COPD patients, the VFCDM provided higher accuracy. In addition, for healthy subjects, VFCDM showed reasonably accurate estimation of FEV1/FVC. However, for COPD patients, the estimation results were not satisfied with the clinical needs. Nevertheless, all of the COPD results with VFCDM were less than 0.70, which met the criteria for assessing COPD summarized in Table 1.

Table 5.

Individual estimation values of FEV1/FVC at THP = 100% using VFCDM, CWT, and STFT methods.

Healthy #1 #2 1 #3 #4 #5 #6 #7 #8 #9 1 #10 #11 #12 1 #13
Reference 88 84 80 77 76 91 94 90 91 80 70 84 86
VFCDM 87.50 90.59 75.11 79.54 70.45 91.06 85.63 89.67 82.97 80.71 61.88 75.08 89.86
CWT 98.35 99.62 87.99 95.81 85.92 99.04 98.67 98.12 96.57 89.87 83.53 98.26 99.63
STFT 98.85 99.74 86.21 95.51 89.24 99.20 98.70 97.44 98.24 94.05 84.83 98.34 99.67
COPD #1 1 #2 #3 #4 #5 #6 #7 #8 #9 #10 1 #11 #12 #13 1
Reference 37 59 26 41 51 67 61 69 29 56 61 60 26
VFCDM 17.61 30.56 17.90 42.43 57.06 66.02 66.22 69.66 25.31 54.97 42.30 28.51 18.97
CWT 27.16 55.96 52.27 81.10 77.59 89.92 89.94 93.26 70.91 58.03 78.10 40.64 35.72
STFT 29.15 40.88 36.31 83.44 79.11 89.51 89.91 93.79 62.07 58.87 75.86 39.68 24.94

1 Female subject.

Figure 6a–c show screen snapshots of the iPhone 5S application for FEV1/FVC ratio estimation, developed using the Objective-C programming language. Once recording is completed, it takes five seconds for the app to visualize the FEV1/FVC ratio estimation result in real-time. Figure 6a shows the screen at the commencement of recording the forced exhalation through the built-in microphone. Figure 6b shows the forced exhalation in the process of being recorded. Recording can be stopped by clicking the circle button at the bottom of the screen. After the recording is complete, the flow-volume curve and the estimated FEV1/FVC ratio are displayed, as shown in Figure 6c.

Figure 6.

Figure 6

Snapshots of iPhone application for FEV1/FVC ratio estimation, developed using the Objective-C programming language: (a) before recording; (b) forced exhalation being recorded; and (c) flow-volume curve and estimated FEV1/FVC ratio displayed.

3.2. Regression and Estimation of FVC, FEV1, and PEF

Figure 7a–d show the regression plots based on VFCDM for FEV1/FVC, FVC, FEV1, and PEF, respectively, from healthy subjects. As expected, in Figure 7a, EFEV1/FVC, and TFEV1/FVC were linearly correlated with r = 0.814. The constants KFEV1/FVC and CFEV1/FVC were found to be 0.64 and 31.59, respectively. On the other hand, as shown in Figure 7b–d, FVC, FEV1, and PEF did not show strong correlation, respectively.

Figure 7.

Figure 7

Regression plots of: (a) EFEV1/FVC vs. TFEV1/FVC, (b) EFVC vs. TFVC, (c) EFEV1 vs. TFEV1; and (d) EPEF vs. TPEF.

4. Discussion

We presented a smartphone-based lung function test using a high-resolution time-frequency spectrum from a smartphone built-in microphone. Even though the time-frequency spectrum is not a sole component for the result accuracy, it is the key algorithm to increase the accuracy. In [24], the audio from a phone was buffered into 30 ms frames and each frame was windowed to quantify the magnitude spectrogram of the signal. On the other hand, we down-sampled to 2450 Hz and visualized the time-frequency spectrum with the time interval of 1/2450 s, which is equivalent to 0.41 ms, approximately. Hence, our method has approximately 73 times higher time resolution than the previous work. More specifically, to estimate the FEV1/FVC ratio, VFCDM provided lower absolute errors than CWT and STFT by 6.52 and 6.67, respectively. In addition, VFCDM provided lower RMSEs than CWT and STFT by 5.94 and 6.62, respectively. The results suggest that the VFCDM approach provided higher accuracy of the FEV1/FVC ratio than CWT and STFT due to high resolution of the time-frequency spectrum. However, we also found that only the FEV1/FVC ratio can be accurately estimated using the smartphone built-in microphone since the ratio as a relative value can be obtained directly without the estimates of FVC and FEV1. These two factors, as well as PEF, are subjective and dependent on the subject’s familiarization with the test and performance of the forced exhalation. More specifically, each individual may exhale with slightly different angles toward a built-in microphone even with the add-on mouthpiece. In addition, each individual may bite the mouthpiece deeply or shallowly. Each individual may also move or shake the smartphone during the forced exhalation test. Those factors affect the time-frequency spectrum, and may result in inaccurate estimation of FVC, FEV1, and PEF. Healthy subjects #2, #9, and #11 embodied such cases. Regarding FEV1/FVC, if the three subjects #2, #9, and #11 were excluded, the estimation results from the ten other healthy subjects provided very low absolute errors and RMSEs of 3.573 and 4.760, respectively. Regarding FVC, FEV1, and PEF, the correlation r value could increase to 0.763, 0.928, and 0.823, respectively. Thus, the limited and cautious force of exhalation toward the built-in microphone is necessary to increase the estimation accuracy. Furthermore, the future research needs to be toward operation without the add-on adapter. Furthermore, the additional condition needs to be considered: each individual should keep the distance constant between a lip and a microphone. Then, a user may have a trouble keeping the fixed distance between the lip and microphone before the test every time.

We also found that AE and RMSE of FEV1/FVC ratios from COPD patients were relatively high with 10.30% ± 10.59% and 14.48%, respectively, even with VFCDM. These high errors were mainly from the artifact sound caused by the narrow windpipe, especially for COPD patients. Most COPD patients have a narrow windpipe, which causes artifact sound even in routine life. In the case of forced exhalation, the artifact sound is more dominating and the real forced exhalation sound is severely affected by the artifacts. Then, the time-varying main frequency cannot be accurately detected, and eventually the instantaneous frequency with maximum power at each sample cannot appropriately represent the volume of the PFT test. More specifically, in the portion of the sound without the artifacts, our proposed method with VFCDM provides higher accuracy than CWT and STFT. In the portion of the sound with the artifacts, there is no difference among VFCDM, CWT, and STFT. This reflects that the results from healthy subjects have higher accuracy than those from COPD patients. In addition, this reflects that the results with VFCDM have higher accuracy than CWT and STFT for both healthy subjects and COPD patients. Thus, the dispersion trend according to THp is observed only in healthy subjects as shown in Figure 4 and Figure 5. More research regarding artifact sound detection and filtering should be directed to be clinically available in the future.

In our study, among the healthy subjects, seven subjects were aged between 20 and 29 (group A), and the other six subjects were aged between 30 and 39 (group B). For groups A and B, RMSE values were 6.13 and 4.76, and AE values were 4.96 ± 3.89 and 3.93 ± 2.93. However, due to a small number of subjects, the effect of age is not clear from the results. Similarly, for male (N = 10) and female (N = 3) groups from healthy subjects, RMSE values were 4.61 and 7.90, and AE values were 3.48 ± 3.18 and 7.84 ± 1.17. From COPD patients, RMSE values were 15.17 and 11.92, and AE values were 10.65 ± 11.38 and 9.14 ± 99.36, respectively. Similarly with the gender effect, due to a small number of subjects and patients, the effect of gender is not clear from the results. Thus, further studies need to rigorously validate the FEV1/FVC ratio and revise the regression model for FVC, FEV1, and PEF by considering a larger number of subjects with age and gender matching.

5. Conclusions

In this paper, we developed to estimate lung function parameters using a high-resolution time-frequency spectrum from a smartphone built-in microphone. For both healthy and COPD groups, we evaluated the estimation performance of FEV1/FVC, and found that VFCDM was superior to CWT and STFT. Further, regression analysis was conducted to establish a linear transformation on FVC, FEV1 and PEF. However, correlation factor values were not high comparing to FEV1/FVC since they were subjective and dependent on the subject’s familiarization with the test and performance of forced exhalation toward the built-in microphone.

Acknowledgments

This study was partially supported by a grant from the Korean Health Technology R&D Project, Ministry of Health and Welfare, Republic of Korea (HI12C0110) and Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT & Future Planning (NRF-2013R1A1A1005775 and NRF-2015M3A9D7067215).

Abbreviations

PFT Pulmonary function test
FVC Forced vital capacity
FEV1 Forced expiratory volume in 1 s
FEV1/FVC Forced expiratory volume in 1 s/forced vital capacity
PEF peak expiratory flow
ATS American Thoracic Society
BTS British Thoracic Society
ECCS European Community for Coal and Steel
ERS European Respiratory Society
GOLD Global Initiative for Chronic Obstructive Lung Disease
LLN Lower limit of normal
NICE National Institute for Health and Clinical Excellence
VFCDM Variable frequency complex demodulation method
CWT Continuous wavelet transform
RMSE Root mean squared error
STFT short time Fourier transform
SPWV Smoothed pseudo Wigner-Ville
COPD Chronic obstructive pulmonary disease

Author Contributions

J.L. conceived and designed the experiments; T.T., K.H. and H.K. collected dataset with IRB approval, T.T., H.C. and C.J. developed smartphone application and analyzed the data; T.T., K.-H.Y. and J.L. wrote the paper.

Conflicts of Interest

The authors declare no conflict of interest.

References

  • 1.Delzell J.E., Jr. Common lung conditions: Environmental pollutants and lung disease. FP Essent. 2013;409:32–42. [PubMed] [Google Scholar]
  • 2.Gualano R.C., Vlahos R., Anderson G.P. What is the contribution of respiratory viruses and lung proteases to airway remodelling in asthma and chronic obstructive pulmonary disease? Pulm. Pharmacol. Therap. 2006;19:18–23. doi: 10.1016/j.pupt.2005.02.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Petty T.L. Benefits of and barriers to the widespread use of spirometry. Curr. Opin. Pulm. Med. 2005;11:115–120. doi: 10.1097/01.mcp.0000150948.13034.78. [DOI] [PubMed] [Google Scholar]
  • 4.Dombkowski K.J., Hassan F., Wasilevich E.A., Clark S.J. Spirometry use among pediatric primary care physicians. Pediatrics. 2010;126:682–687. doi: 10.1542/peds.2010-0362. [DOI] [PubMed] [Google Scholar]
  • 5.Kaminsky D.A., Marcy T.W., Bachand M., Irvin C.G. Knowledge and use of office spirometry for the detection of chronic obstructive pulmonary disease by primary care physicians. Respir. Care. 2005;50:1639–1648. doi: 10.1016/j.rmedu.2006.03.016. [DOI] [PubMed] [Google Scholar]
  • 6.Pierce R., Johns D. Spirometry: The Measurement and Interpretation of Ventilatory Function in Clinical Practice. 1995. [(accessed on 15 August 2016)]. Originally commissioned by The Thoracic Society of Australia and New Zealand, First published. Available online: https://goo.gl/Sv5MQr. [Google Scholar]
  • 7.Pierce R. Spirometry: An essential clinical measurement. Aust. Fam. Phys. 2005;34:535–539. [PubMed] [Google Scholar]
  • 8.Miller M.R., Hankinson J., Brusasco V., Burgos F., Casaburi R., Coates A., Crapo R., Enright P., Van Der Grinten C., Gustafsson P. Standardisation of spirometry. Eur. Respir. J. 2005;26:319–338. doi: 10.1183/09031936.05.00034805. [DOI] [PubMed] [Google Scholar]
  • 9.Koch G. Standardization of lung function testing in infants, children and adolescents. Bull. Eur. Physiopathol. Respir. 1983;19:15–24. [PubMed] [Google Scholar]
  • 10.American Thoracic Society Lung function testing: Selection of reference values and interpretative strategies. Am. Rev. Respir. Dis. 1991;144:1202–1218. doi: 10.1164/ajrccm/144.5.1202. [DOI] [PubMed] [Google Scholar]
  • 11.Quanjer P.H., Tammeling G.J., Cotes J.E., Pedersen O.F., Peslin R., Yernault J.C. Lung volumes and forced ventilatory flows. Eur. Respir. J. 1993;6(Suppl. S16):5–40. doi: 10.1183/09041950.005s1693. [DOI] [PubMed] [Google Scholar]
  • 12.Ferguson G.T., Enright P.L., Buist A.S., Higgins M.W. Office spirometry for lung health assessment in adults: A consensus statement from the national lung health education program. Chest. 2000;117:1146–1161. doi: 10.1378/chest.117.4.1146. [DOI] [PubMed] [Google Scholar]
  • 13.American Thoracic Society Standards for the diagnosis and care of patients with chronic obstructive pulmonary disease (COPD) and asthma. Am. Rev. Respir. Dis. 1987;136:225–244. doi: 10.1164/ajrccm/136.1.225. [DOI] [PubMed] [Google Scholar]
  • 14.The COPD Guidelines Group of the Standards of Care Committee of the BTS BTS guidelines for the management of chronic obstructive pulmonary disease. Thorax. 1997;52(Suppl. S5):S1–S28. [PMC free article] [PubMed] [Google Scholar]
  • 15.Swanney M.P., Ruppel G., Enright P.L., Pedersen O.F., Crapo R.O., Miller M.R., Jensen R.L., Falaschetti E., Schouten J.P., Hankinson J.L., et al. Using the lower limit of normal for the fev1/fvc ratio reduces the misclassification of airway obstruction. Thorax. 2008;63:1046–1051. doi: 10.1136/thx.2008.098483. [DOI] [PubMed] [Google Scholar]
  • 16.Celli B.R., MacNee W., Force A.E.T. Standards for the diagnosis and treatment of patients with copd: A summary of the ats/ers position paper. Eur. Respir. J. 2004;23:932–946. doi: 10.1183/09031936.04.00014304. [DOI] [PubMed] [Google Scholar]
  • 17.Sevick M.A., Trauth J.M., Ling B.S., Anderson R.T., Piatt G.A., Kilbourne A.M., Goodman R.M. Patients with complex chronic diseases: Perspectives on supporting self-management. J. Gen. Internal Med. 2007;22:438–444. doi: 10.1007/s11606-007-0316-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Seemungal T.A., Donaldson G.C., Bhowmik A., Jeffries D.J., Wedzicha J.A. Time course and recovery of exacerbations in patients with chronic obstructive pulmonary disease. Am. J. Respir. Crit. Care Med. 2000;161:1608–1613. doi: 10.1164/ajrccm.161.5.9908022. [DOI] [PubMed] [Google Scholar]
  • 19.Lindgren B.R., Finkelstein S.M., Prasad B., Dutta P., Killoren T., Scherber J., Edin Stibbe C.L., Snyder M., Hertz M.I. Determination of reliability and validity in home monitoring data of pulmonary function tests following lung transplantation. Res. Nurs. Health. 1997;20:539–550. doi: 10.1002/(SICI)1098-240X(199712)20:6&#x0003c;539::AID-NUR8&#x0003e;3.0.CO;2-N. [DOI] [PubMed] [Google Scholar]
  • 20.Otulana B.A., Higenbottam T., Ferrari L., Scott J., Igboaka G., Wallwork J. The use of home spirometry in detecting acute lung rejection and infection following heart-lung transplantation. Chest J. 1990;97:353–357. doi: 10.1378/chest.97.2.353. [DOI] [PubMed] [Google Scholar]
  • 21.Bjortuft O., Johansen B., Boe J., Foerster A., Holter E., Geiran O. Daily home spirometry facilitates early detection of rejection in single lung transplant recipients with emphysema. Eur. Respir. J. 1993;6:705–708. [PubMed] [Google Scholar]
  • 22.Pelkonen A.S., Nikander K., Turpeinen M. Reproducibility of home spirometry in children with newly diagnosed asthma. Pediatr. Pulmonol. 2000;29:34–38. doi: 10.1002/(SICI)1099-0496(200001)29:1&#x0003c;34::AID-PPUL6&#x0003e;3.0.CO;2-O. [DOI] [PubMed] [Google Scholar]
  • 23.Morlion B., Knoop C., Paiva M., Estenne M. Internet-based home monitoring of pulmonary function after lung transplantation. Am. J. Respir. Crit. Care Med. 2002;165:694–697. doi: 10.1164/ajrccm.165.5.2107059. [DOI] [PubMed] [Google Scholar]
  • 24.Larson E.C., Goel M., Boriello G., Heltshe S., Rosenfeld M., Patel S.N. Spirosmart: Using a microphone to measure lung function on a mobile phone; Proceedings of the UbiComp; Pittsburgh, PA, USA. 5–8 September 2012; pp. 280–289. [Google Scholar]
  • 25.Lee J., McManus D.D., Bourrell P., Sornmo L., Chon K.H. Atrial flutter and atrial tachycardia detection using bayesian approach with high resolution time-frequency spectrum from ecg recordings. Biomed. Signal Process. 2013;8:992–999. doi: 10.1016/j.bspc.2013.04.002. [DOI] [Google Scholar]
  • 26.Wang H., Siu K., Ju K., Chon K.H. A high resolution approach to estimating time-frequency spectra and their amplitudes. Ann. Biomed. Eng. 2006;34:326–338. doi: 10.1007/s10439-005-9035-y. [DOI] [PubMed] [Google Scholar]
  • 27.Chon K.H., Dash S., Ju K. Estimation of respiratory rate from photoplethysmogram data using time-frequency spectral estimation. IEEE Trans. Bio-Med. Eng. 2009;56:2054–2063. doi: 10.1109/TBME.2009.2019766. [DOI] [PubMed] [Google Scholar]
  • 28.Selvaraj N., Shelley K.H., Silverman D.G., Stachenfeld N., Galante N., Florian J.P., Mendelson Y., Chon K. A novel approach using time-frequency analysis of pulse-oximeter data to detect progressive hypovolemia in spontaneously breathing healthy subjects. IEEE Trans. Bio-Med. Eng. 2011;58:2272–2279. doi: 10.1109/TBME.2011.2144981. [DOI] [PubMed] [Google Scholar]
  • 29.Miller M.R., Crapo R., Hankinson J., Brusasco V., Burgos F., Casaburi R., Coates A., Enright P., van der Grinten C.M., Gustafsson P. General considerations for lung function testing. Eur. Respir. J. 2005;26:153–161. doi: 10.1183/09031936.05.00034505. [DOI] [PubMed] [Google Scholar]
  • 30.Larson E.C., Saba E., Kaiser S., Goel M., Patel S.N. Pulmonary Monitoring Using Smartphones. [(accessed on 29 January 2016)]. Available online: http://lyle.smu.edu/~eclarson/teaching/ubicomp/papers/PulmonarySmartphones.pdf.
  • 31.Goel M., Saba E., Stiber M., Whitmire E., Fromm J., Larson E.C., Borriello G., Patel S.N. Spirocall: Measuring lung function over a phone call; Proceedings of the 2016 CHI Conference on Human Factors in Computing Systems; San Jose, CA, USA. 7–12 May 2016; pp. 5675–5685. [Google Scholar]
  • 32.Sato H., Ohara M., Watanabe K., Sato H. Application of the vortex whistle to the spirometer. Trans. Soc. Instrum. Control Eng. 1999;35:840–845. doi: 10.9746/sicetr1965.35.840. [DOI] [Google Scholar]
  • 33.Kristiansen U., Amielh M. Measurements on a little known sound source-the vortex whistle. 2016 [Google Scholar]
  • 34.Crapo R.O., Morris A.H., Gardner R.M. Reference spirometric values using techniques and equipment that meet ats recommendations 1–3. Am. Rev. Respir. Dis. 1981;123:659–664. doi: 10.1164/arrd.1981.123.6.659. [DOI] [PubMed] [Google Scholar]
  • 35.Morris J.F., Koski A., Johnson L.C. Spirometric standards for healthy nonsmoking adults 1–3. Am. Rev. Respir. Dis. 1971;103:57–67. doi: 10.1097/00043764-197107000-00021. [DOI] [PubMed] [Google Scholar]
  • 36.Polgar G., Promadhat V. Standard values in pulmonary function testing in children. Phila.: WB Saunders Go. 1971;179:87–212. [Google Scholar]
  • 37.Polgar G., Varuni P. Pulmonary Function Testing in Children: Techniques and Standards. Saunders Limited.; Condell Park, Australia: 1971. [Google Scholar]

Articles from Sensors (Basel, Switzerland) are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)