Abstract
Artificial ligaments such as the Leeds-Keio ligament are inserted into knee joints to treat cruciate ligament trauma. Three cases of prosthetic ligament failure are reported, showing foreign body granulomatous inflammation around implanted fibres, together with abundant fibre particles within giant cells. It is suggested that intra-operative fibre fragmentation can produce a sufficient quantity of dendritic particles to elicit a granulomatous reaction, which can ultimately lead to ligament instability and failure.
Full text
PDF

Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Fujikawa K., Iseki F., Seedhom B. B. Arthroscopy after anterior cruciate reconstruction with the Leeds-Keio ligament. J Bone Joint Surg Br. 1989 Aug;71(4):566–570. doi: 10.1302/0301-620X.71B4.2768298. [DOI] [PubMed] [Google Scholar]
- Marcacci M., Gubellini P., Buda R., De Pasquale V., Strocchi R., Molgora A. P., Zaffagnini S., Guizzardi S., Ruggeri A. Histologic and ultrastructural findings of tissue ingrowth. The Leeds-Keio prosthetic anterior cruciate ligament. Clin Orthop Relat Res. 1991 Jun;(267):115–121. [PubMed] [Google Scholar]