Abstract
Two quantitative, automated methods for the determination of C reactive protein (CRP) were compared: turbidimetry (Cobas Fara II, Roche, Welwyn Garden City, UK) and fluorescence polarisation TDx, Abbott, Wokingham, UK). One hundred and twenty routine serum samples submitted for measurement of CRP were tested using both procedures. The results were compared using regression line analysis and showed a high degree of correlation (r2 = 0.99, X coefficient = 1.01, constant = 0.11). C reactive protein can be accurately measured using the automated turbidimetric method which can be recommended as an alternative to fluorescence polarisation.
Full text
PDF

Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Ballou S. P., Kushner I. C-reactive protein and the acute phase response. Adv Intern Med. 1992;37:313–336. [PubMed] [Google Scholar]
- Benattar C., Francoual J., Magny J. F., Lindenbaum A. Evaluation d'une méthode rapide d'immunodosage de la protéine C réactive par polarisation de fluorescence. Ann Biol Clin (Paris) 1989;47(4):181–185. [PubMed] [Google Scholar]
- Johnson A. M. A new international reference preparation for proteins in human serum. Arch Pathol Lab Med. 1993 Jan;117(1):29–31. [PubMed] [Google Scholar]
- Young B., Gleeson M., Cripps A. W. C-reactive protein: a critical review. Pathology. 1991 Apr;23(2):118–124. doi: 10.3109/00313029109060809. [DOI] [PubMed] [Google Scholar]