Abstract
The recent availability of novel dyes and alternative light sources to facilitate complex tissue immunofluorescence studies such as multiplex labelling has not been matched by reports critically evaluating the considerations and relative benefits of these new tools, particularly in combination. Product information is often limited to wavelengths used for older fluorophores (FITC, TRITC & corresponding Alexa dyes family). Consequently, novel agents such as Quantum dots are not widely appreciated or used, despite highly favourable properties including extremely bright emission, stability and potentially reduced tissue autofluorescence at the excitation wavelength. Using spectral analysis, we report here a detailed critical appraisal and comparative evaluation of different light sources and fluorophores in multiplex immunofluorescence of clinical biopsy sections. The comparison includes mercury light, metal halide and 3 different LED-based systems, using 7 Qdots (525, 565, 585, 605, 625, 705), Cy3 and Cy5. We discuss the considerations relevant to achieving the best combination of light source and fluorophore for accurate multiplex fluorescence quantitation. We highlight practical limitations and confounders to quantitation with filter-based approaches.
Introduction
From beginnings in the early 20th century, fluorescence microscopy has evolved to a widely used research and diagnostic tool capable of high quality image generation for the study of tissues. Fluorophores, illumination sources and optics continue to improve with the development of new dyes and hardware to meet demands from researchers. Investigators in cancer immunotherapy and tissue immunology now seek to discriminate and quantify multiple (5 or more) fluorophore labels applied to single samples, especially tissue sections, for in situ cell phenotyping. To achieve these goals with clinical formalin-fixed paraffin-embedded (FFPE) tissue samples, researchers must also address the challenge posed by auto-fluorescence. Many studies are based on fluorescence microscopy of tissue sections, but relatively little data is available critically evaluating the capacity and limitations of standard approaches compared with the range of newer methodological options, and in particular the potential benefits and limitations for multiplex tissue immunofluorescence.
Among newer fluorophores, the non-organic Quantum dots (Qdots) have highly favourable characteristics, outperforming traditional dyes such as FITC, TRITC or their more stable recent counterparts such as the Alexa dyes, in key metrics. These include brightness and photostability, a large Stokes shift and a broad choice of narrowband emission spectra across the visible spectrum into infrared [1–5] (S Prost Manuscript PONE-D-16-16178 under revision). The Qdots excitation spectrum also peaks around 400nm, which potentially reduces autofluorescence with FFPE tissue (see results). Nevertheless, while fluorophore performance is critical, the illumination system, labelling protocol and analysis technology also affect the quality of results. Traditionally, mercury light (HBO) with selectable excitation filters has been favoured for fluorescence microscopy as a powerful wideband source of excitation wavelengths across the visible spectrum to ultraviolet (UV). However, newer metal halide and LED systems lack the drawbacks of HBO that include explosive hazard, cumbersome alignment, non-uniformity, temporal instability and fading. Metal halide provides a similar excitation spectrum, but with stronger emission outwith the peaks of mercury light, is brighter overall, has controllable intensity, greater uniformity and stability [6]. New LED systems also offer uniform and dimmable illumination, typically of narrow band width, which is potentially advantageous for multi-colour fluorescence. LED emission is more intense than mercury for some, but not all wavelengths (for example green excitation spectra) [7].
In the present study, we consider the factors contributing to an optimal imaging system for multiplex immunostaining of human FFPE tissue, focusing on light sources, fluorophores and their interaction. We compare in detail the relative merits of 3 different light sources (HBO, metal halide and LED) and different fluorophores, including 7 Qdot labels, Cy3 & Cy5. We then further compare 3 different LED systems with a particular focus on stability of the fluorescence, autofluorescence and specificity of emission. We use spectral unmixing and image analysis to accurately determine the occurrence and magnitude of specific fluorescence, bleed-through and autofluorescence with different setups, which are otherwise not separately measurable in more conventional filter-based imaging systems. The findings are applicable to any system of image capture and analysis. We highlight key areas of consideration and potential weakness for investigators wishing to accurately measure multiplex-stained tissues, and recommend an approach to maximise the combined benefits of the different components.
Materials and Methods
Light Sources
Illumination sources tested were mercury light (HBO100 watts), metal halide (Olympus), Sola (Lumencor), CoolLED PE2 with 425 & 535/615 LAMs, Sola 2 (Lumencor, CoolLED PE4000) [8,9]. The Sola and Sola 2 produce LED-based white light continuous spectrum from 380 to 680 nm; by contrast the PE2 and the PE4000 produce selected wave lengths between 365nm and 770nm [8,9]. The PE2 unit can drive up to 4 LED wavelengths from 2 LAMs (LED Array modules) while the PE4000 has 16 selectable LED wavelengths which can be activated singly or in groups up to 4 [10]. The PE4000 can also be used as a source of white light but this was not tested here. Both companies provide alternative systems that were not tested. All excitation was done with the light source at full power unless specified.
Microscope and Filters
The microscope used is an Olympus BX51, with a Chroma filter cube “Qdot set with long pass emission filter“(set 32013 Exciter E460SPUVv2; emitter E500lp; Dichroic 475dcxru). For the comparison of LED illumination sources in the 2nd part of the study, the set was improved by changing the excitation filter with a Semrock FF01-417/60-25 (laser2000) (average transmission >90% over 60nm compared with 70% for E460SPUVv2) (S1 File). The data presented was captured using a x40 objective (Olympus UplanFl 40x/0.75 ∞/0.17); brighter objectives (40x Olympus UPlanSApo 40x/0.95 ∞/0.11–0.23/FN26.5 and 20x Olympus UPlanSApo 20x/0.75 ∞/0.17/FN26.5) reduced exposure times by 30–50% (S2 Table).
The Cy5 filter was Cy5-4040C (Semrock) and Cy3 filter Cy3-4040C (Semrock) adapted with a reduced band emission filter FF01- 543/22 (Semrock) to prevent bleed through).
Multiplex Immunofluorescence and Fluorophores
We used the Labelled StrepAvidin Biotin (LSAB) method with streptavidin-conjugated Qdots for single colour staining or a combination of LSAB technique and indirect immunofluorescence with secondary Ab conjugated with Qdots (all Life technology). Briefly, 4μm FFPE sections of human liver or kidney were treated with histosolve (Fisher ThermoScientific) and rehydrated. Antigen retrieval was accomplished with 0.1mM EDTA pH8 in a microwave pressure cooker for 10 minutes at pressure before TBST washes. The sections were treated with the Vectorlabs avidin/biotin blocking kit (SP-2001) where appropriate and incubated with antiCD68 (macrophage marker, Dako, M0876) at 1/150 at room temperature for 45 minutes. After TBST washes, sections were incubated with biotinylated antimouse IgG3 (AbCam ab97258) 1/200 for 30 minutes, followed by incubation with streptavidin-conjugated Qdots (Life technology) for 30 minutes or with the appropriate secondary Ab conjugated with Qdot. Qdots were pre-diluted to give a comparable final intensity of positive fluorescence (preferable for quantification) (Table 1 & S1 Fig). Slides were then dehydrated, cleared, and mounted in Qmount (Life technology). For 5 color immunostaining (Qdots 525, 565, 655, Cy3 & Cy5) the first two fluorophore steps (Cy3 & Cy5) were completed on a Bond immunostainer (Leica) using TSA plus fluorescence system (NEL744 & NEL745 Perkin Elmer), according to the manufacturer’s instructions. The antigen retrieval step was performed with the epitope retrieval 2 (Leica, EDTA pH8) and repeated after the Cy3 (Ki67) and the Cy5 (CD31) staining, followed with manual staining using an indirect method for Qdot 585 (CD163) and 655 (CD86) and a final streptavidin-Qdot 525 (CD68) step as previously described (S Prost Manuscript PONE-D-16-16178 under revision). TSA Alexa 488 & 655 (Thermofisher) were also tested, but the fluorescence intensity was too weak compared with Qdots and those fluorophores were not investigated further.
Table 1. Qdots molar extinction coefficient ε (cm-1M-1 determined at 405nm) and dilutions used to achieve a similar level of fluorescence (adapted from [11]).
Product & catalogue name | ε(cm-1M-1) | dilution |
---|---|---|
Qdot 525 Q10141MP | 360,000 | 6/100 |
Qdot 565 Q10131MP | 1,100,000 | 4/100 |
Qdot 585 Q10111MP | 2,200,000 | 2/100 |
Qdot 605 Q10101MP | 2,800,000 | 1/100 |
Qdot 625 A10196 | 9,900,000 | 1/750 |
Qdot 655 Q10121MP | 5,700,000 | 1/200 |
Qdot 705 Q10161MP | 8,300,000 | 2/100 |
Note that the there is a logarithmic correlation between ε and the chosen dilutions apart for Qdot 705 which requires a higher titter to achieved a similar level of fluorescence. This is likely due to the lower performance of the camera to detect emission in the infrared spectrum. (dilution = a-bxlog(ε) r = 097149 & r = 0.90376 without and with Qdot 705 respectively; see S1 Fig).
Imaging and Analysis
Investigators often rely on filters to mitigate relative spectral contamination from auto-fluorescence by truncating the spectrum captured, but this approach has no effect on auto-fluorescence within the captured bandwidth. Using spectral unmixing [12] with a tuneable camera (Nuance FX camera, CRI; now PerkinElmer [13]) we separated the complete auto-fluorescence spectrum for individual wavelength analysis (Fig 1), allowing us to accurately compare the effect of different illumination sources on autofluorescence across the fluorophore emission spectrum, 500-720nm. Measurement data are Scaled Counts/Second which represents the counts after scaling for exposure time, binning, camera gain, and bit depth (Scaled counts/s = (counts/full scale) x (1/exp2) x (1/bin2)x (1/gain); with the full scale being 4096 for 12 bits acquisition. The value is therefore independent of those settings.
Various Worldwide Web based modelling tools are available to help match fluorophores, filters and illumination systems. We used Semrocksearchlight [14] to model the best fluorescence set-up for a 7 Qdots multiplex and obtain a theoretical comparison of illumination with mercury light, metal halide light, Sola LED white light and a series of single wave lengths LED illumination from COOL-LED. An example of theoretical data obtained for Qdot 605 is shown in Table 2. This was repeated for the 7 Qdots (S1 Table). From the theoretical analysis, PE2 LED system with a 405nm LAM was predicted to give the best fluorescence signal with a good signal to noise ratio for all the Qdots (Table 2 and S1 Table). Real comparisons were then performed using archival FFPE human liver and kidney tissues, known for their strong autofluorescence.
Table 2. Comparison of theoretical fluorescence parameters using semrocksearchlight.com; example of Qdot 605.
Fluorophore: Qdot 605 | |||||
Filter Set: Exciter FF01-405/150; Emitter FF01-500/LP; Dichroic FF510-Di02 | |||||
Light Source | 405 LED Single Wavelength CoolLED | 425 LED Single Wavelength CoolLED | Metal Halide 120mW | Lamp Mercury Arc Lamp | LED Broadband Lumencor SOLA |
Fluorescence Signal (mW) | 1 | 0.82 | 0.47 | 0.46 | 0.28 |
Excitation Light Noise (mW) | 1 | 4.10 | * | * | * |
Autofluorescence Noise (mW) | 1 | 1.01 | 0.49 | 0.51 | 0.37 |
Signal-to-Noise Ratio | 1 | 0.81 | 0.96 | 0.90 | 0.77 |
A | |||||
Fluorophore: Qdot 605 | |||||
Filter Set: Exciter FF01-405/150; Emitter FF01-500/LP; Dichroic FF510-Di02 | |||||
Light Source | 405nm LED CoolLED pE-4000 | 435nm LED CoolLED pE-4000 | Metal Halide X-Cite 120mW | HBO 100 Mercury Lamp | LED Broadband Lumencor SOLA light engine |
Fluorescence Signal (mW) | 1 | 0.79 | 0.493 | 0.555 | 0.414 |
Excitation Light Noise (mW) | 1 | 1.16 | 22 | 171000 | 22.4 |
Autofluorescence Noise (mW) | 1 | 1.01 | 0.501 | 0.527 | 0.518 |
Signal-to-Noise Ratio | 1 | 0.785 | 0.984 | 0.217 | 0.8 |
B |
The test was performed for all Qdots, a selection of possible filters, and various available illumination sources. Example for Qdot 605 is shown compared with PE2 LED 405 & PE2 435nm (A) and compared with PE-4000 405 & 435nm (B) (data for the other Qdots in S1 Table)
LED 405nm is taken as reference. The Light Source Power (mW) was set at 100, Numerical Aperture 0.5, Index of Refraction 1, Reflected Excitation Light Factor 0.07, Transmitted Excitation Light Factor, 0.95, Fluorophore Optical Depth (M cm) 1.00E-12, Fluorophore DMAC (M-1cm-1) 80000, Fluorophore Quantum Yield 0.5, Autofluorescence Light Factor 0.1 and Transmitted Emission Factor 0.95
Test A was performed in 2014, test B in 2016.
*No data available
Tissue Sourcing and Ethical Consent
The clinical tissue samples (inflamed human liver and kidney samples) used in this study were surplus archival paraffin block material originally procured for diagnosis. The archival tissue block samples were fully anonymised to the researchers and as such do not require patient consent. Their use for this study was fully approved by the local clinical governance and ethics committee (Lothian SAHSC and Tissue Governance Unit) with Approval Reference: 07/S1102/21.
Results
Initial Illumination Comparison for Qdot Fluorescence: Mercury versus Metal Halide versus LED
Metal Halide causes photobleaching of Qdots and strong autofluorescence of red blood cells
FFPE sections were labelled with Qdots and imaged successively using mercury light, LED 405 and Metal Halide source. Metal halide or mercury bulb illumination (HBO100W) of FFPE sections labelled with Qdots each produced more autofluorescence than did LED 405 (Fig 2). Furthermore, with metal halide illumination, photobleaching of Qdots was clearly appreciable with simple eye-balling. Photobleaching was still observed even with the illumination reduced to 10% intensity, so metal halide was discarded as inappropriate for quantification of Qdot intensity.
Mercury light provides good illumination of Qdots but higher autofluorescence
Mercury light illumination of Qdots also gave greater autofluorescence than LED 405, especially of red blood cells (Fig 2), but as no photobleaching was apparent by eye, further evaluations were made. Slides labelled with each Qdots were imaged using mercury bulb or CoolLED PE2 with either a 405 or a 425nm LAM (LED Array Module). Intensity images were generated by Nuance and the specific and background (autofluorescence) intensities were recorded.
The autofluorescence of the tissue and associated red blood cells (RBC) had similar spectra for a given illumination source in both the liver (Fig 3A) and the kidney samples (Fig 3B). However, the autofluorescence spectrum under UV illumination differed from that with LED 405 & 425, having a peak of intensity around 525nm for UV compared with 550nm for LEDs (Fig 3C).
The overall intensity of autofluorescence between 500 and 720 nm was greatest in UV-illuminated samples (UV: 0.964 counts/s versus 0.135 and 0.146 counts/s in the same region of interest illuminated with LED 405 & 425nm respectively). However, this observation is not necessarily pertinent to researchers using filter blocks to evaluate only fluorescence around the emission peak of their specific fluorophore. We therefore measured the levels of autofluorescence and of Qdot fluorescence on images taken at specific wavelengths: serial sections were either unlabelled (autofluorescence test) or labelled with the same antibody and one of the 7 Qdots tested. The same areas from stained and unstained serial sections were illuminated using the different light sources and intensity images were capture for emission wave length corresponding to the peak of emission of each specific Qdot. The average specific fluorescence intensities (CD68 in macrophages) and autofluorescence intensities (in proximal tubules) were quantified using the Nuance analysis software. Table 3 gives the average intensities of autofluorescence and corresponding specific fluorescence recorded at the peak of emission of the selected Qdot (i.e. autofluorescence intensity at 525nm when stained with Qdot 525; autofluorescence intensity at 565nm with Qdot 565 etc.).
Table 3. Comparison of fluorescence intensities when illuminating Qdots with mercury light, LED 405nm or LED 425nm.
Mercury bulb | LED 405 nm | LED 425 nm | |||||||
---|---|---|---|---|---|---|---|---|---|
Qdot | positive cells | autofluo | Ratio | positive cells | autofluo | Ratio | positive cells | autofluo | Ratio |
525 | 1.557 | 1.183 | 1.317 | 0.566 | 0.201 | 2.819 | 1.289 | 0.158 | 8.158 |
565 | 1.483 | 1.177 | 1.259 | 0.329 | 0.170 | 1.940 | 0.445 | 0.177 | 2.517 |
585 | 1.342 | 1.107 | 1.213 | 0.497 | 0.152 | 3.270 | 0.419 | 0.154 | 2.715 |
605 | 3.466 | 0.888 | 3.904 | 0.545 | 0.118 | 4.624 | 0.615 | 0.117 | 5.273 |
625 | 4.252 | 0.697 | 6.097 | 0.436 | 0.091 | 4.805 | 0.764 | 0.094 | 8.139 |
655 | 2.032 | 0.537 | 3.786 | 0.898 | 0.069 | 13.032 | 0.875 | 0.067 | 13.094 |
705 | 0.852 | 0.266 | 3.205 | 0.305 | 0.011 | 27.869 | 0.250 | 0.010 | 23.396 |
The table gives the levels of fluorescence (average counts/s) in positive cells stained with the indicated Qdot (positive cells) and in the unstained control on a serial section (autofluo), under illumination by 3 systems (Mercury light, LEDs PE2-405nm & PE2-425nm). The levels of positive fluorescence and autofluorescence are recorded using the spectral camera with tunable filter at one unique wave length corresponding to the peak of fluorescence emission of each Qdot (ie 525nm for Qdot 525; 565nm for Qdot 565 etc).
The Ratio of specific fluorescence to autofluorescence (pink column) is an indication of the discrimination between autofluorescence and specific fluorescence: the higher the ratio specific:autofluorescence, the better. Note that LED 405 is generally a poorer exciter than LED 425.
Please note that the conditions used for the present test lead to a weaker specific signal than normally obtained with Qdots and were chosen to test the limits of the system.
Unlike suggested by the theoretical data (Table 2 & S1 Table), mercury light was a much stronger exciter of Qdots when compared with either LED source (compare the intensity of fluorescence levels in positive cells excited by mercury bulb with LED Table 3). However, the intensity of autofluorescence in surrounding tissue was also more intense under mercury bulb illumination (Table 3), leading to a much poorer overall signal-to-noise ratio compared with LED. The ratio for LED 425nm being over 2–7 fold greater than mercury bulb with different Qdots (Table 3 Ratio).
LED-425 gives better Qdot illumination than LED-405
Interestingly, the excitation profile for Qdots (Fig 4) suggests that the lower wavelengths give greater excitation, as did the theoretical assessment using Semrock searchlight (S1 Table). However, we found that LED-425nm was a similar or better exciter than LED-405nm for 5 out of 7 Qdots (Table 3) and moreover that the specific:autofluorescence ratios were consequently better when using 425nm, especially for the green and yellow wave lengths (525nm & 565nm), where autofluorescence from FFPE is the strongest (Fig 3). Simultaneous illumination with 405 and 425nm LAMs was not better than LED-425 alone (data not shown). LED illumination therefore had clear advantages compared with mercury bulb and metal halide for fluorescence studies using Qdots in FFPE samples. In addition to excellent temporal and special stability, there was no significant photobleaching of the Qdots unlike with Metal Halide, and tissue autofluorescence was reduced compared with mercury bulb illumination.
Comparison of white light LED and single wavelength LED for multicolour fluorescence using Qdots with other fluorophores
Having initially settled on a single excitation system with 425nm (PE2-CoolLED) excitation which was found to be optimum for Qdots work, we investigated the possibility of performing multicolour fluorescence combining Qdots with other fluorophores. Commercial LED illumination systems are either white light equivalent or with selectable single colour emission diodes. Qdots share a common excitation spectrum (350-480nm, with emission between 525 and 800nm depending on the Qdot), but other fluorophores impose additional excitation requirements on selectable illumination systems, requiring additional LED Array Modules (LAM). For testing, we evaluated a modern white light LED source (Sola2), having improved power especially in green wavelengths, together with two selectable LED sources: firstly, the PE4000 from CoolLED, with capacity for 16 wavelengths that can be turned on individually or in groups of 4; secondly, the PE2 with 2 or 4 LAM. The Sola2 and PE4000 were tested with illumination of Qdots, Cy3, Cy5, Alexa 488 and 555, and further compared with the PE2 for Qdots, Cy3 & Cy5. Using spectral unmixing we analysed the overall fluorescence intensity between 500 and 720nm, the peak emission intensity and the relative level of specific and autofluorescence at the emission peak.
Intensity of autofluorescence and specific fluorescence with Qdots
Illumination with each LED source, PE4000 (405+460nm), Sola2 and PE2 (425nm), gave similar autofluorescence spectra (Fig 5D). However, autofluorescence increased when illuminated at increasing wavelengths between 540 and 650nm with PE4000 (405<435<460 Fig 5A, 5B and 5C).
We quantified autofluorescence and specific fluorescence from Qdots 525, 565, 585, 605, 625, 655 and 705 on serial FFPE kidney sections labelled with the same specific primary antibody (Ab) against CD68 and a secondary Ab complexed with the indicated Qdot. Matching fields of the stained and unstained serial sections were illuminated with the indicated light source and fluorescence images were taken every 10 nm from 500 to 720nm using the Nuance tuneable camera. The exposure time was set to achieve 70% of the dynamic range (autoexposure function) and the exposure times were recorded (Table 4). The autoexposure time reflects the strength of excitation provided by each illumination system to each fluorophore as well as the level of autofluorescence where this is not negligible. It is dependent on the strength of the immunofluorescence (function of the extinction coefficient and quantum yield of each fluorophore as well as the level of the target labelled in the tissue), the autofluorescence of the tissue and the microscope configuration, including the optical characteristics of the objective.
Table 4. Autoexposure times to achieve 70% of the dynamic range.
Autoexposure time (ms) | PE2 | Sola 2 | PE 4000 | ||||
---|---|---|---|---|---|---|---|
425 nm | White LED light | 405nm | 435nm | 460nm | 405+460 nm | 435+460 nm | |
Qdot 525 | 149.26 | 141.97 | 404.11 | 236.61 | 501.15 | 250.31 | 176.70 |
Qdot 565 | 71.20 | 69.22 | 213.18 | 122.14 | 311.52 | 137.70 | 86.78 |
Qdot 585 | 76.12 | 79.61 | 257.76 | 145.01 | 356.85 | 164.00 | 99.17 |
Qdot 605 | 273.6 | 265.3 | 541.392 | 365.376 | 994.624 | 1018.86 | 409.425 |
Qdot 625 | 55.14 | 55.95 | 178.83 | 88.66 | 224.56 | 100.12 | 68.08 |
Qdot 655 | 193.47 | 151.28 | 311.52 | 251.26 | 546.54 | 306.51 | 220.99 |
Qdot 705 | 252.08 | 335.94 | 1035.70 | 553.14 | 1609.26 | 865.58 | 508.94 |
autofluo | 514.66 | 383.09 | 1609.73 | 737.09 | 2000.00 | 1029.58 | 467.90 |
A field of view of sections stained with the indicated Qdot was successively illuminated by all indicated light sources in the order indicated in the table.
The exposure time is automatically calculated by the Nuance system to prevent saturation of pixels by restricting the exposure time to 70% percentage of the dynamic range of pixel values. A second autoexposure value was recorded for PE2 before moving to another section stained with the next color to confirm the absence of photobleaching between the first and the last exposure (data not shown).
The intensities of autofluorescence and of specific fluorescence were measured at the peak fluorescence for each fluorophore (Fig 6 and Table 5). Excitation with Sola 2 & PE2 gave the strongest autofluorescence (Table 5A) and specific fluorescence (Table 5B) for each fluorophore. However, the ratio of specific to autofluorescence was best for most fluorophores with the PE4000 405 nm (Table 5C & Fig 7).
Table 5. Autofluorescence and specific fluorescence emission.
A | |||||||
Autofluo at | 525 nm | 565 nm | 585 nm | 605 nm | 625 nm | 655 nm | 705 nm |
405 | 0.061 +/- 0.027 | 0.043 +/- 0.016 | 0.035 +/- 0.012 | 0.026 +/- 0.008 | 0.02 +/- 0.005 | 0.019 +/- 0.003 | 0.012 +/- 0.002 |
435 | 0.137 +/- 0.059 | 0.101 +/- 0.038 | 0.083 +/- 0.029 | 0.061 +/- 0.019 | 0.047 +/- 0.012 | 0.041 +/- 0.008 | 0.027 +/- 0.004 |
460 | 0.043 +/- 0.019 | 0.037 +/- 0.015 | 0.032 +/- 0.012 | 0.024 +/- 0.008 | 0.018 +/- 0.006 | 0.014 +/- 0.004 | 0.009 +/- 0.002 |
405+460 | 0.098 +/- 0.043 | 0.075 +/- 0.029 | 0.063 +/- 0.022 | 0.047 +/- 0.015 | 0.036 +/- 0.01 | 0.031 +/- 0.007 | 0.02 +/- 0.003 |
435+460 | 0.175 +/- 0.076 | 0.135 +/- 0.051 | 0.112 +/- 0.039 | 0.083 +/- 0.026 | 0.064 +/- 0.017 | 0.054 +/- 0.011 | 0.035 +/- 0.006 |
PE2 | 0.28 +/- 0.112 | 0.201 +/- 0.07 | 0.164 +/- 0.052 | 0.12 +/- 0.034 | 0.092 +/- 0.022 | 0.079 +/- 0.014 | 0.05 +/- 0.005 |
Sola 2 | 0.207 +/- 0.037 | 0.167 +/- 0.045 | 0.146 +/- 0.054 | 0.114 +/- 0.055 | 0.091 +/- 0.052 | 0.079 +/- 0.051 | 0.047 +/- 0.03 |
B | |||||||
Specific fluo at | 525 nm | 565 nm | 585 nm | 605 nm | 625 nm | 655 nm | 705 nm |
405 | 0.282 +/- 0.093 | 0.917 +/- 0.269 | 0.898 +/- 0.191 | 0.183 +/- 0.034 | 0.988 +/- 0.362 | 0.542 +/- 0.125 | 0.238 +/- 0.044 |
435 | 0.536 +/- 0.171 | 1.686 +/- 0.492 | 1.68 +/- 0.354 | 0.335 +/- 0.068 | 1.926 +/- 0.715 | 0.789 +/- 0.195 | 0.351 +/- 0.067 |
460 | 0.205 +/- 0.069 | 0.584 +/- 0.172 | 0.589 +/- 0.123 | 0.147 +/- 0.033 | 0.757 +/- 0.284 | 0.302 +/- 0.075 | 0.154 +/- 0.03 |
405+460 | 0.49 +/- 0.162 | 1.459 +/- 0.432 | 1.467 +/- 0.312 | 0.282 +/- 0.055 | 1.676 +/- 0.625 | 0.63 +/- 0.167 | 0.298 +/- 0.061 |
435+460 | 0.753 +/- 0.241 | 2.2 +/- 0.65 | 2.221 +/- 0.474 | 0.447 +/- 0.09 | 2.575 +/- 0.96 | 0.905 +/- 0.249 | 0.392 +/- 0.082 |
PE2 | 1.033 +/- 0.299 | 3.137 +/- 0.711 | 2.287 +/- 0.603 | 0.51 +/- 0.096 | 3.402 +/- 1.054 | 1.319 +/- 0.335 | 0.594 +/- 0.094 |
Sola 2 | 0.839 +/- 0.312 | 2.659 +/- 0.846 | 2.184 +/- 0.69 | 0.521 +/- 0.126 | 3.153 +/- 1.282 | 1.168 +/- 0.302 | 0.888 +/- 0.146 |
C | |||||||
ratio | 525 | 565 | 585 | 605 | 625 | 655 | 705 |
405 | 4.60 | 21.51 | 25.79 | 7.10 | 48.24 | 29.22 | 19.62 |
435 | 3.91 | 16.63 | 20.18 | 5.48 | 40.92 | 19.36 | 12.92 |
460 | 4.74 | 15.64 | 18.32 | 6.12 | 42.12 | 21.46 | 18.02 |
405+460 | 5.02 | 19.52 | 23.43 | 6.05 | 46.31 | 20.25 | 14.93 |
435+460 | 4.31 | 16.28 | 19.77 | 5.38 | 40.45 | 16.83 | 11.09 |
PE2 | 3.69 | 15.63 | 13.95 | 4.27 | 37.07 | 16.78 | 11.89 |
Sola 2 | 4.06 | 15.97 | 14.93 | 4.57 | 34.67 | 14.78 | 18.80 |
A: intensity of autofluorescence at indicated wave lengths (average counts/s +/- sdv) as quantified in unstained kidney tubules (Fig 6A). red: worse levels of autofluorescence.
B: specific fluorescence emission at indicated wave lengths (average counts/s +/- sdv) as quantified in CD68 stained macrophages (Fig 6B). green: best levels of specific fluorescence.
C: ratio of specific emission over autofluorescence at indicated wave lengths.
The ratio gives the level of specific fluorescence at the peak of fluorescence of each Qdot.
Highlighted are the best ratios for each specific Qdot at peak of fluorescence. Note that the autoexposure time (with a long pass emission filter Chroma emitter E500lp) for Qdot605 is often high and close to that of the autofluorescence (Table 4) showing that the intensity of Q605 fluorescence in the stain is low, however, as the autofluorescence intensity around this wavelength is lower than in the green emission area (Fig 5) the specific fluorescence ratio is good (D) at this specific wave length.
When using spectral unmixing, increased autofluorescence is not an issue as it can be separated from the specific fluorescence (Fig 1) and a fast capture is beneficial. However, without access to spectral unmixing, the best specific:autofluorescence ratio should be sought–especially when dealing with low level target signal. The example given for Qdot 605 (Fig 8) shows that LED 405nm would be the best illumination option in that circumstance: the unmixed images (left and middle columns) and their intensity plot show the underlying composition of data that would be generated with simple epifluorescence image acquisition (right column): both autofluorescence (left column) and specific fluorescence (middle column) are visible contributors at the emission peak of 605nm, and this autofluorescence is particularly strong with 435nm illumination (magenta line). By contrast, the plot shows that illumination at 405nm gives a high specific fluorescence (red line profile) and the lowest autofluorescence (green line), while 460nm illumination has similarly low autofluorescence (yellow line) but lower specific fluorescence (dark blue line) (Fig 8 & Table 5C). Given that the target used for the study is a high density label (CD68 in macrophages) he data show that illumination quality is a sensitive determinant of signal dynamic range: suboptimal illumination wavelengths can render low density targets unreliable for quantification as signal levels approach closer to autofluorescence, generating a low dynamic range of quantifiable intensity. LED 405 here represents the best combination of bright specific fluorescence with low autofluorescence.
Illumination of Qdots in combination with other fluorophores
The possibility of using other fluorophores together with Qdots was tested with Cy3, Cy5, Alexa 488 and Alexa 555. We found that the intensities of Alexa 488 and Alexa 555 were too weak (even with tyramide amplification) compared with Qdots and the bleed through too contaminating—including bleed through of Alexa 488 into the Alexa 555 detection filter—to use for reliable quantification even with spectral unmixing (data not shown). In contrast, Cy3 and Cy5 gave good signal intensity. We compared the excitation using PE2 (535 & 615nm respectively), Sola white light, or PE4000 (555 and 635nm). Although the LAM 615nm is not optimum for excitation of Cy3, PE2 can host LAMs for only 4 different wavelengths, which restricted the choice. LAM 615nm was already used for imaging the Qnuclear Deep Red (LifeTechnology) counterstain in our Qdot multiplex staining, so this was tested for excitation of Cy5. In these conditions and despite the suboptimal excitation wavelength, PE2 and PE4000 were comparable and gave better excitation of Cy3 and Cy5 than Sola white light, which was less effective than the single intensity LAMS (Table 6, Fig 9).
Table 6. Autofluorescence and specific fluorescence at peak of emission for Cy3 & Cy5 excited by white light LED (SOLA), PE2 (535nm & 615nm) or PE4000 (555 & 635nm).
Cy3 at 570nm | PE2 535nm | PE4000 555nm | Sola | |||
BK | specific | BK | specific | BK | specific | |
average | 1.989757 | 9.160652 | 1.588668 | 6.795557 | 1.327651 | 5.540776 |
sem | 0.253373 | 0.182925 | 0.014931 | 0.123194 | 0.224278 | 0.105282 |
ratio specific/bk | 4.603906 | 4.277519 | 4.173367 | |||
Cy5 at 670nm | PE2 615nm | PE4000 635nm | Sola | |||
BK | specific | BK | specific | BK | specific | |
average | 0.272805 | 0.919322 | 0.26374 | 1.122252 | 0.128934 | 0.52169 |
sem | 0.01203 | 0.022824 | 0.025742 | 0.02747 | 0.010783 | 0.014581 |
ratio specific/BK | 3.369884 | 4.255147 | 4.046169 |
The specific staining was quantified in thresholded positive cells and 3 regions of interest were drawn manually in unstained tissue for quantification of the background fluorescence (BK).
The average fluorescence intensity (counts/s) was quantified at the peak of fluorescence on images taken at the appropriate wave length by the Nuance camera with Tunable filter (ie 570nm for Cy3 & 670nm for Cy5).
Peaks of excitation for Cy3 & Cy5 are 554nm & 649nm respectively.
We next tested for confounding excitation/emission interactions among the fluorophores under illumination relevant to a multiplex imaging setting. Sections stained with one per section of the 7 study Qdots or Cy3 or Cy5 were each illuminated with all the wave lengths together that would be potentially used in a multiplex staining involving Qdots, Cy3 and Cy5. The resulting emitted fluorescence (specific, autofluorescence, or bleed through of specific fluorescence) was visualised directly through the microscope eyepieces and scored semiquantitatively (Table 7). Qdots under single wavelength excitation (PE4000 & PE2) gave a clear signal in the Qdot filter (in Table 7; green cells = good level of specific fluorescence) without any apparent bleed through across the Cy3 or Cy5 filters (in Table 7; black cells represent no fluorescence going through a given filter; all Qdots excited by all single wave length LEDs are “black cells” for Cy3 and Cy5 filters). Illumination of Qdots with the white light LED also gave no apparent bleed through across the Cy5 filter, apart from Qdot 655 (Table 7, row “Qdot655/Sola”; column “Cy5 filter” is colored pink for non expected fluorescence). By contrast, all Qdots under white light LED illumination showed specific fluorescence bleed-through across the Cy3 filter (pink cells for all “Qdots/Sola” in “Cy 3 filter” column), sometimes accompanied by autofluorescence (Table 7, S2 Fig). This bleed-through was particularly strong from Qdot 565, having a similar order of magnitude to fluorescence seen with the specific Qdot filter (Table 7, Fig 10). This effect may be due to the small absorbance peak for Qdot565 at 545nm (S3 Fig). Illumination with the single LEDs 405, 435 or 460nm does not activate Cy3, but the Sola white light LED can still illuminate and excite Qdots due to the passband within the Cy3 excitation filter between 500 & 550nm (S4 Fig). The emission peaks of Cy3 and Qdot 565 are close together, while Qdot 585 emission is fully passed by the emission filter for Cy3 (S3 Fig & S4 Fig). This then causes the observed bleed through of both Qdot 565 and 585nm emissions across the Cy3 filter under white light (Table 7). Conversely, Cy3 illuminated with the appropriate LED single wave lengths (PE2 & PE400 535nm) was excited with no bleed-through across the other filters (green for expected fluorescence in Cy3 filter, and black in other filters). However, illumination of Cy3 by white light LED or any single LED wave length suitable for Qdot excitation led to visible fluorescence bleed-through the Qdot filter (Table 7, pink rows in Qdot filter column). In these conditions, quantification of a multiplex using Qdots and Cy3 would not be possible without spectral unmixing (see further). These findings differ from the software modeller’s prediction that more bleed-through would occur with PE2 or PE4000 illumination than with Sola white light illumination (Table 8). Cy5 excitation by any of the light sources did not bleed through Qdot or Cy3 filters, as predicted (Table 7).
Table 7. Bleed through test.
Dye | Illumination | Cy5 filter | Qdot filter | Cy3 filter |
---|---|---|---|---|
Qdot 525 | PE4000 405nm | |||
PE4000 435nm | ||||
PE4000 460nm | + | |||
PE2 425nm | ||||
Sola | AUTO | |||
Qdot 565 | PE4000 405nm | |||
PE4000 435nm | ||||
PE4000 460nm | ||||
PE2 425nm | ||||
Sola | +++ (**) | |||
Qdot 585 | PE4000 405nm | |||
PE4000 435nm | ||||
PE4000 460nm | ||||
PE2 425nm | ||||
Sola | ++ | |||
Qdot 605 | PE4000 405nm | |||
PE4000 435nm | ||||
PE4000 460nm | ||||
PE2 425nm | ||||
Sola | ++ | |||
Qdot 625 | PE4000 405nm | |||
PE4000 435nm | ||||
PE4000 460nm | ||||
PE2 425nm | ||||
Sola | + & AUTO (***) | |||
Qdot 655 | PE4000 405nm | |||
PE4000 435nm | ||||
PE4000 460nm | ||||
PE2 425nm | ||||
Sola | + | AUTO | ||
Cy5 | PE4000 635nm | |||
Sola | ||||
PE2 615nm | (*) | |||
PE2 425nm | ||||
Cy3 | PE 4000 535nm | |||
PE4000 405nm | ++ | |||
PE4000 435nm | ++ | |||
PE4000 460nm | ++ | |||
Sola | ++ | |||
PE2 535nm | ||||
PE2 425nm | ++ |
Serial sections stained with the indicated dye (column 1) were illuminated with the indicated light (column 2) and of the observation of specific fluorescence (+/++/+++) and autofluorescence (AUTO) going through each filter block (Cy5; Qdot & Cy3 filter blocks) was recorded.
+/++/+++ reflect the intensity of the specific fluorescence; AUTO reflects the presence of noticeable autofluorescence.
grey cells: no visible fluorescence; green cells: expected specific fluorescence, strong unless indicated; pink cells: bleed through (i.e. observation of fluorescence in a set up where fluorescence is not desirable).
Excitation wave lengths that would not normally be used for a dye are indicated by a red cell.
(*) more autofluorescence was observed with PE2 615nm than with Sola & PE4000 635nm
(**) see Fig 10;
(***) see image in S2 Fig
Table 8. Theoretical bleed through for Qdot 565 in Cy3 filter, using semrocksearchlight.com.
Theoretical data | single wave length LED | LED white light | ||||
Name | Qdot565 in Qdot filter | Cy3 in Cy3 filter | Qdot in Cy3 filter | Qdot565 in Qdot filter | Cy3 in Cy3 filter | Qdot in Cy3 filter |
Filter Set | specific | specific | bleed through | specific | specific | bleed through |
Exciter | FF02-435/40 | FF01-543/22 | FF01-543/22 | FF02-435/40 | FF01-543/22 | FF01-543/22 |
Emitter | FF01-496/LP | FF01-593/40 | FF01-593/40 | FF01-496/LP | FF01-593/40 | FF01-593/40 |
Dichroic | FF482-Di01 | FF562-Di03 | FF562-Di03 | FF482-Di01 | FF562-Di03 | FF562-Di03 |
Fluorophore Emission | Qdot 565 | Cy3 | Qdot 565 | Qdot 565 | Cy3 | Qdot 565 |
Light Source | CoolLED pE4000 | Lumencor SOLA light engine | ||||
435 nm | 525 nm | 525 nm | ||||
Fluorescence Signal (mW) | 2.71E-06 | 3.54E-06 | 1.86E-07 | 7.60E-07 | 8.39E-07 | 4.22E-08 |
Excitation Light Noise (mW) | 1.29E-08 | 2.72E-09 | 2.72E-09 | 2.15E-08 | 1.49E-09 | 1.49E-09 |
Autofluorescence Noise (mW) | 0.0000325 | 0.0000034 | 0.0000034 | 0.00000937 | 0.000000747 | 0.000000747 |
Signal-to-Noise Ratio | 0.083442 | 1.039 | 0.054611 | 0.080983 | 1.1212 | 0.056396 |
Ratio single wave length /LED white light | Qdot565 in Qdot filter | Cy3 in Cy3 filter | Qdot in Cy3 filter | |||
specific | specific | bleed through | ||||
Fluorescence Signal (mW) | 3.565789474 | 4.219308701 | 4.407582938 | (1) | ||
Excitation Light Noise (mW) | 0.6 | 1.825503356 | 1.825503356 | (2) | ||
Autofluorescence Noise (mW) | 3.468516542 | 4.551539491 | 4.551539491 | (3) | ||
Signal-to-Noise Ratio | 1.030364397 | 0.926685694 | 0.968348819 | (4) | ||
Single wave length | LED white light | Ratio single wave length/white light LED (7) | ||||
Bleed through signal Qdot/specific signal Qdot (5) | 0.06863 | 0.05553 | 1.23607 | |||
Bleed though into Cy3 / specific Cy3 (6) | 0.05254 | 0.0503 | 1.04462 |
From the theoretical data, ratio between the data given for single wave length and the white light illumination were calculated. They suggest that:
(1) The fluorescence signal would be stronger with single wave length LED but this would lead to more bleed through too
(2) The light noise for Qdot would be better with single wave length LED but worse than the white light illumination for Cy3
(3) The autofluorescence for Qdot would be worse with single wave length LED for Cy3 and Qdot
(4) The signal to noise ratio would slightly better for Qdot with single wave length LED but worse for Cy3
(5) The equivalent of 6.86% of Qdot specific fluorescence would be observed through Cy3 filter with single wave length LED & 5.55% with LED white light
(6) The equivalent of 5.25% of Qdot specific would go through Cy3 filter with LED white light & 5.03% with LED white light
(7) There would be more bleed through with single wave length LED, about 1.2 time more than with white light LED
From direct visual inspection it was therefore clear that white light LED, but apparently not single wavelength LED, produced visible bleed through in many cases. However, measuring with the Nuance multispectral camera, we found that bleed through was also present with single LED illumination (Fig 11) albeit at a much lower level. Spectral unmixing systems like Nuance [13] can account for bleed-through and provide real intensity images (see example of a 5 color multiplex using 3 Qdots together with Cy3 & Cy5 in Fig 12), because the intensity is calculated according to the entire spectrum of each fluorophore. By contrast, the present data shows that quantification of this type of mixed fluorophore multiplex staining would be jeopardised with systems relying on filter blocks only.
Discussion
We aimed here to identify the best system for accurate discrimination and quantification of fluorescence signals from immuno-detection of multiple antigens in FFPE tissue sections when using epifluorescence and a standard microscope. The present data explore in detail the interaction of light source, different fluorophores and tissue autofluorescence; they identify favourable combinations of different fluorophore types (Q dots and tyramide-amplified Cy3 & Cy5), while other combinations were found unsuitable (Qdots with Alexa fluorophores).
The ideal fluorophore would be bright, stable, without bleed through, with its excitation wavelength causing minimum autofluorescence at the emission wave length(s). Compared with conventional organic fluorophores, Qdots fluorophores have highly favourable physical properties for multiplexing studies, including high extinction coefficients and quantum yields, long fluorescence life time and stability [1] (Prost Manuscript PONE-D-16-16178 Under revision), as well as large Stokes shifts with narrow bright symmetrical emission spectra [2,11,17]. The present data validate the utility of those qualities in practice and confirm Qdots as well-suited to multiplex immunofluorescence study of tissue sections.
Bleed-through, for example of green into red [18], is an important confounder of epifluorescence evaluation of multiplex staining. In particular, even low percentage bleed-through of a bright fluorophore to a channel whose target is only weakly positive can obscure the latter with false signal. Mitigation strategies include attempting to equalise signal strengths in multiplex staining by careful combination of fluorophore, antibody concentrations and labelling technique. Varying illumination strength within an experiment is best avoided if possible but can ameliorate strong fluorophore bleed-through, as can reducing excitation filter bandwidth (e.g. from 531/40 for QnuclearDeepRed/Cy3 to 543/22-25, to reduce bleed through of Qdot 585 or 605). Nevertheless, filters including band-pass filters are not 100% efficient and may not be sufficient. The present data show that changing white light sources for monochromatic LED much reduces some instances of bleed through but that even so, measurable contamination can remain, detectable after spectral unmixing. Moreover, some fluorophore combinations such as Cy3 with Qdots were not remediable even with this strategy. The Qdots narrow and symmetrical emission spectra were an optimal solution here to prevent the problems of bleed-through, and although conventional fluorophores such as Alexa dyes were too weak in our setting to use alongside Qdots, tyramide amplification allowed effective combination with Cy5, if needed. Combination with Cy3 is also possible, when analysis is done with spectral unmixing.
Autofluorescence is another major confounder of accurate evaluation of fluorescence signals, especially with FFPE clinical samples [19]. Mammalian tissue autofluorescence derives from flavin coenzymes, reduced pyrimidine nucleotides, reticulin fibres, collagens and elastin as well as lipofuscin, to which is added fluorescence induced by formalin fixation, ubiquitous to standard clinical biopsy samples [20]. The level of endogenous fluorescence varies between tissues, being strong in liver and kidney. We found here that illumination source and wavelength has an impact on autofluorescence, with an increase between 560 and 680nm when using mercury bulb, LED 460 or Sola white light. This may in part reflect increased excitation of tissue flavins (excitation 360–520; emission 500–560) [21,22]. An understanding of the autofluorescence spectrum in the tissue of interest allows the investigator to select fluorophores with excitation and emission properties that minimise its impact, especially for detection of low density targets. Here, we found Qdots highly favourable compared with organic fluorophores such as Alexa 488, due to both reduced autofluorescence with the low wavelength excitation requirement of Qdots and because of the brightness of the Qdots.
We found that the choice of illumination source affected the performance of Qdots and other fluorophores: metal halide caused photobleaching of Qdots (and other fluorophores–data not shown) at such a rate that quantification of multiplex staining was not reliable in our configuration of manual capture of cubes of images for spectral unmixing. However, in a setting without spectral unmixing, others have successfully used rapid scanning detection of multiplexed Qdots under metal halide illumination [23]. When considering excitation, the light source would ideally match the excitation profile of the fluorophore without superfluous “contaminating”wavelengths. Broad spectrum light can be passed through carefully chosen filter sets, but the filter efficiency is never 100% and bleed-through of nonspecific wave lengths can significantly degrade the reliability of epifluorescence data, as observed here. “Monochromatic “LED reduce the likelihood of cross-illumination and autofluorescence compared with broad spectrum illumination (metal halide, mercury or LED white light), allowing a high contrast image with high signal-to-noise ratio. The outstanding temporal stability of the LED illumination compared with gas discharge lamps that need time to reach an equilibrium and decay throughout their life and which intensity fluctuates during use [24,25] is a further advantage for quantification over project lifetimes. However, the bandwidth of “monochromatic” LED varies from 10 to 85nm (example for PE -4000 [8]) so appropriate filters may still be required. Good Web-based tools are available to predict the performance of illumination, fluorophore and filter combinations and to estimate bleed-through and fluorescence intensity [14,26–29]. These provide a good starting point and acquaint the user with parameters that affect fluorescence quality. However, the estimates provided do not replace validation through practical testing: we observed significant discrepancies in fluorescence intensity and bleed-through with different light sources. Discussion with the technical representative for SemRock highlighted that Semrocksearchlight [14] was, at the time, assuming equal intensity for all CoolLED modules (personal communication). It is likely that the better excitation provided by PE2 LED 425 compared with PE2 LED 405 reflected different intensity of the 2 LAMs; such issues are to be expected with constantly evolving hardware iterations. Most vendors are willing to loan lights, cubes and other hardware for users to test before purchase (we had loans from Cool LED, Olympus, Semrock, Leica, Sola/ Lumencor for this study and from CRI and Zeiss previously) and we would recommend on-site testing after theoretical analysis.
Of course, other technical considerations not primarily the focus of this study also contribute to achieving a bright, specific and reliable signal [30,31], including configuration of the filter set (S1 File), microscope objectives (S2 Table), camera and the immunohistochemical detection method. There are several detection methods for successful multiple immunostaining, each with particular merits [18,32–35]. We have had success with different methods individually (indirect method, ABC, and tyramide amplification), and in combination (indirect with Qdots together with ABC with Qdots, and tyramide amplification with Cy3 & Cy5). Factors affecting choice of method include the nature of the target, including density and variability. The methodologic requirements to establish and maintain reproducible reliable staining are exacting [18,36]; they include standardised sample preparation (fixation, processing, section thickness, optimal antigen retrieval), [33] consistent positive and negative controls with each experiment, including for reproducibility, effect of the staining order, saturation and cross reactivity tests. Nevertheless, with a good labelling technique controlling for cross binding, the major limitations to quantification of multiplex fluorescence are the stability and uniformity of the fluorescence, confounding autofluorescence and bleed through, which are key outcome measures considered in this study. We have found Qdots illuminated with LED greatly superior in all these areas and would recommend their use with monochromatic LED. Broad spectrum LED can be more powerful and advantageous for qualitative fluorescence, or possibly where autofluorescence is not an issue (e.g. cultured cells) (high specific fluorescence emission of all Qdots illuminated by Sola in Table 5), but do not match the specificity and flexibility of monochromatic LED in this setting. We tested here lights from Lumencor (Sola broad light) and CoolLED (monochromatic), but both manufacturers make white light (broad spectra) and individually selectable LEDs. We found the material we tested from both companies to be excellent, with excellent technical and customer support. We therefore recommend a multi-monochromatic LED illumination system, but not a specific brand.
The present study shows that workable epifluorescence solutions to quantifying multiplex fluorescence can be achieved with careful planning, but the best solution to accurate quantification involves spectral unmixing, which accounts for bleed-through and autofluorescence. We use the Nuance FX system [12,13], but various techniques have been used both for fluorescence [23,37–39] and chromogenic stains [32,40,41]. Attempting quantification without spectral unmixing depends on a relatively crude “subtraction of background fluorescence”, which precludes reliable quantification of low level targets in particular. Our analyses of fluorescence when using the recommended filter cubes showed that autofluorescence or bleed-through were often so close to specific fluorescence (because of the typically restricted emission filter bandwidth) that they could only be discriminated using a camera with tunable filter and spectral unmixing with a long pass emission filter (tested for Cy 3, Cy5, Alexa 488, Alexa555 through specific single, double and triple band pass conventionally used in qualitative multicolour fluorescence). However, capture of cubes of images for multispectral analysis is time consuming: multiple pictures have to be taken at 10-20nm intervals across the emission bandwidth of each cube, which we like to be wide to obtain well defined spectra of emission. We have found this to benefit from the very bright staining that Qdots deliver, without needing over-amplification steps that easily saturate. It is possible to combine multiple single LAMS to increase the excitation intensity (435+460nm for excitation of Qdots for example); if this improves specific fluorescence without compromise from autofluorescence (Table 5) the reduced capture time may justify the compromise [31], especially if the fluorescence can be spectrally unmixed. Binning (S5 Fig) improves signal-to-noise ratio to the detriment of resolution, although we have not found it necessary with Qdots and LED.
In conclusion, we believe that with good wet lab protocols and controls, Qdots fluorophores (525 565 585 605 625 655 & 705), used alone or together with selected organic fluorophores, in a system using selectable monochromatic light LED and analysed with a spectral unmixing system is the system of choice for quantification of multiplex immunostaining on FFPE tissues. Such an approach is sensitive and specific, including for FFPE tissues considered particularly problematic such as liver and kidney.
Supporting Information
Acknowledgments
The work was supported by a project grant from the CSO (Chief Scientist Office Scotland ETM/189 to David Kluth, Christopher Bellamy & Sandrine Prost, a Moray Endowment Fund award to Sandrine Prost (University of Edinburgh), and an Innovation grant (GR01858) from the University of Edinburgh Development Trust.
We thank Leica, Semrock, CoolLED and Olympus for the loan of equipment as well as discussions and advices. We thank Clare Ainslie and the LifeTechnology technical application Scientists for their help with Qdots.
The TSA Cy3 & Cy5 were performed by Histoplex (Edinburgh)
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
The work was supported by a project grant from the CSO (Chief Scientist Office Scotland) ETM/189 to David Kluth, Christopher Bellamy & Sandrine Prost; a Moray Endowment Fund award to Sandrine Prost (University of Edinburgh), and an Innovation grant (GR01858) from the University of Edinburgh Development Trust. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1.Resch-Genger U, Grabolle M, Cavaliere-Jaricot S, Nitschke R, Nann T. Quantum dots versus organic dyes as fluorescent labels. Nature methods. 2008;5(9):763–75. Epub 2008/08/30. 10.1038/nmeth.1248 . [DOI] [PubMed] [Google Scholar]
- 2.Pinaud F, Michalet X, Bentolila LA, Tsay JM, Doose S, Li JJ, et al. Advances in fluorescence imaging with quantum dot bio-probes. Biomaterials. 2006;27(9):1679–87. Epub 2005/12/02. 10.1016/j.biomaterials.2005.11.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Michalet X, Pinaud FF, Bentolila LA, Tsay JM, Doose S, Li JJ, et al. Quantum dots for live cells, in vivo imaging, and diagnostics. Science. 2005;307(5709):538–44. Epub 2005/02/01. 10.1126/science.1104274 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Goldman ER, Mattoussi H, Anderson GP, Medintz IL, Mauro JM. Fluoroimmunoassays using antibody-conjugated quantum dots. Methods in molecular biology (Clifton, NJ). 2005;303:19–34. Epub 2005/06/01. 10.1385/1-59259-901-X:019 . [DOI] [PubMed] [Google Scholar]
- 5.Medintz IL, Uyeda HT, Goldman ER, Mattoussi H. Quantum dot bioconjugates for imaging, labelling and sensing. Nature materials. 2005;4(6):435–46. Epub 2005/06/02. 10.1038/nmat1390 . [DOI] [PubMed] [Google Scholar]
- 6.Zeiss. Microscope Light Sources 2016. Available: http://zeiss-campus.magnet.fsu.edu/articles/lightsources/.
- 7.Aswani KJ, T.; Brown C.M. Fluorescence Microscopy Light Sources. Microscopy Today. 2012;20(04):18–21. [Google Scholar]
- 8.CoolLED. PE-4000 product detail 2016. Available: http://www.coolled.com/product-detail/pe-4000/.
- 9.Lumencor. Sola Light Engine—Solid state white light excitation subsystems [cited 2016]. Available: http://lumencor.com/products/sola-light-engine/.
- 10.CoolLED. Products 2016. Available: http://www.coolled.com/products.
- 11.InVitrogen. Molecular Probes® Handbook Section 6.6. 2010.
- 12.Mansfield JR, Hoyt C, Levenson RM. Visualization of Microscopy-Based Spectral Imaging Data from Multi-Label Tissue Sections Current Protocols in Molecular Biology: John Wiley & Sons, Inc.; 2001. [DOI] [PubMed] [Google Scholar]
- 13.PerkinElmer. User's Manual for Nuance 3.0.2. 2012.
- 14.Semrock. Searchlight [2016]. Available: http://searchlight.semrock.com/.
- 15.Chroma. E460SPUVv2 2016. Available: https://www.chroma.com/products/parts/e460spuvv2.
- 16.Semrock. Typical Measured Spectrum of Semrock FF01-593/40 Optical Filter.
- 17.Tholouli E, Sweeney E, Barrow E, Clay V, Hoyland JA, Byers RJ. Quantum dots light up pathology. The Journal of pathology. 2008;216(3):275–85. Epub 2008/09/25. 10.1002/path.2421 . [DOI] [PubMed] [Google Scholar]
- 18.Hoffman G. Seeing is believing: use of antibodies in immunocytochemistry and in situ hybridization. Short Course II. 2008:1. [Google Scholar]
- 19.Viegas MS, Martins TC, Seco F, do Carmo A. An improved and cost-effective methodology for the reduction of autofluorescence in direct immunofluorescence studies on formalin-fixed paraffin-embedded tissues. European journal of histochemistry: EJH. 2007;51(1):59–66. Epub 2007/06/06. . [PubMed] [Google Scholar]
- 20.Monici M. Cell and tissue autofluorescence research and diagnostic applications. Biotechnology annual review. 2005;11:227–56. Epub 2005/10/12. 10.1016/S1387-2656(05)11007-2 . [DOI] [PubMed] [Google Scholar]
- 21.Croce AC, Bottiroli G. Autofluorescence spectroscopy and imaging: a tool for biomedical research and diagnosis. European journal of histochemistry: EJH. 2014;58(4):2461 Epub 2015/01/13. 10.4081/ejh.2014.2461 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Evans EW, Dodson CA, Maeda K, Biskup T, Wedge CJ, Timmel CR. Magnetic field effects in flavoproteins and related systems. Interface focus. 2013;3(5):20130037 Epub 2014/02/11. 10.1098/rsfs.2013.0037 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Isse K, Lesniak A, Grama K, Maier J, Specht S, Castillo-Rama M, et al. Preexisting epithelial diversity in normal human livers: a tissue-tethered cytometric analysis in portal/periportal epithelial cells. Hepatology. 2013;57(4):1632–43. Epub 2012/11/15. 10.1002/hep.26131 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Beacher J. LEDs for fluorescence microscopy. Biophotonics International. 2008;15:34–7. [Google Scholar]
- 25.Zeiss. Fundamentals of Light-Emitting Diodes. 2016.
- 26.Biolegend. Fluorescence Spectra Analyzer 2016. Available: http://www.biolegend.com/spectraanalyzer.
- 27.Leica. FluoScout 2016. Available: http://www.leica-microsystems.com/applications/life-science/fluorescence/learn-more-about-the-leica-fluoscout/.
- 28.Zeiss. Filter Assistant 2016. Available: https://www.micro-shop.zeiss.com/index.php?l=en&p=de&f=f.
- 29.ThermoFisher. Fluorescence SpetraViewer 2016. Available: http://www.thermofisher.com/uk/en/home/life-science/cell-analysis/labeling-chemistry/fluorescence-spectraviewer.html.
- 30.Waters JC. Accuracy and precision in quantitative fluorescence microscopy. The Journal of cell biology. 2009;185(7):1135–48. Epub 2009/07/01. 10.1083/jcb.200903097 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Brown CM. Fluorescence microscopy–avoiding the pitfalls. Journal of Cell Science. 2007;120(19):3488–. [DOI] [PubMed] [Google Scholar]
- 32.van der Loos CM. Multiple Immunoenzyme Staining: Methods and Visualizations for the Observation With Spectral Imaging. Journal of Histochemistry and Cytochemistry. 2008;56(4):313–28. 10.1369/jhc.2007.950170 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Taylor CR, Levenson RM. Quantification of immunohistochemistry—issues concerning methods, utility and semiquantitative assessment II. Histopathology. 2006;49(4):411–24. Epub 2006/09/19. 10.1111/j.1365-2559.2006.02513.x . [DOI] [PubMed] [Google Scholar]
- 34.Osman TA, Oijordsbakken G, Costea DE, Johannessen AC. Successful triple immunoenzymatic method employing primary antibodies from same species and same immunoglobulin subclass. European journal of histochemistry: EJH. 2013;57(3):e22 Epub 2013/10/03. 10.4081/ejh.2013.e22 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Ma B, Winkelbach S, Lindenmaier W, Dittmar KE. Six-colour fluorescent imaging of lymphoid tissue based on colour addition theory. Acta histochemica. 2006;108(4):243–57. Epub 2006/05/30. 10.1016/j.acthis.2006.02.002 . [DOI] [PubMed] [Google Scholar]
- 36.Hoffman GE, Le WW, Sita LV. The importance of titrating antibodies for immunocytochemical methods. Current Protocols in Neuroscience. 2008:2.12. 1–2. 26. [DOI] [PubMed] [Google Scholar]
- 37.Schieker M, Pautke C, Haasters F, Schieker J, Docheva D, Bocker W, et al. Human mesenchymal stem cells at the single-cell level: simultaneous seven-colour immunofluorescence. Journal of anatomy. 2007;210(5):592–9. Epub 2007/04/25. 10.1111/j.1469-7580.2007.00716.x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Pautke C, Tischer T, Vogt S, Haczek C, Deppe H, Neff A, et al. New advances in fluorochrome sequential labelling of teeth using seven different fluorochromes and spectral image analysis. Journal of anatomy. 2007;210(1):117–21. Epub 2007/01/19. 10.1111/j.1469-7580.2006.00660.x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Stack EC, Wang C, Roman KA, Hoyt CC. Multiplexed immunohistochemistry, imaging, and quantitation: a review, with an assessment of Tyramide signal amplification, multispectral imaging and multiplex analysis. Methods (San Diego, Calif). 2014;70(1):46–58. Epub 2014/09/23. 10.1016/j.ymeth.2014.08.016 . [DOI] [PubMed] [Google Scholar]
- 40.Levenson RM. Spectral imaging perspective on cytomics. Cytometry Part A: the journal of the International Society for Analytical Cytology. 2006;69(7):592–600. Epub 2006/05/09. 10.1002/cyto.a.20292 . [DOI] [PubMed] [Google Scholar]
- 41.Diebele I, Kuzmina I, Lihachev A, Kapostinsh J, Derjabo A, Valeine L, et al. Clinical evaluation of melanomas and common nevi by spectral imaging. Biomedical optics express. 2012;3(3):467–72. Epub 2012/03/22. 10.1364/BOE.3.000467 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the paper and its Supporting Information files.