. 2016 Apr 15;24(8):9053–9066. doi: 10.1364/OE.24.009053

High-accuracy reference standards for two-photon absorption in the 680–1050 nm wavelength range

Sophie de Reguardati 1, Juri Pahapill 1, Alexander Mikhailov 2, Yuriy Stepanenko 3,4, Aleksander Rebane 1,2,*
PMCID: PMC5025204  PMID: 27137334

Abstract

Degenerate two-photon absorption (2PA) of a series of organic fluorophores is measured using femtosecond fluorescence excitation method in the wavelength range, λ2PA = 680–1050 nm, and ~100 MHz pulse repetition rate. The function of relative 2PA spectral shape is obtained with estimated accuracy 5%, and the absolute 2PA cross section is measured at selected wavelengths with the accuracy 8%. Significant improvement of the accuracy is achieved by means of rigorous evaluation of the quadratic dependence of the fluorescence signal on the incident photon flux in the whole wavelength range, by comparing results obtained from two independent experiments, as well as due to meticulous evaluation of critical experimental parameters, including the excitation spatial- and temporal pulse shape, laser power and sample geometry. Application of the reference standards in nonlinear transmittance measurements is discussed.

OCIS codes: (190.0190) Nonlinear optics; (300.6410) Spectroscopy, multiphoton; (190.4710) Optical nonlinearities in organic materials

1. Introduction

Spectroscopic reference standards facilitate carrying out measurements of molecular absorption- and scattering cross sections, quantum yields, etc. under circumstances, where absolute methods are overly involved or inapplicable [1]. Reference standards are even more vital in nonlinear-optical spectroscopy [2], where accurate absolute determination of the nonlinear molecular parameters requires knowledge on the instantaneous photon flux, i.e. the number of photons incident on the sample per unit time and per unit area. For this, one would need to perform equally accurate measurements of the spatial- and temporal properties of the excitation beam in a broad range of wavelengths, which, given the notorious inconsistency of tunable lasers, may pose a challenging task.

Degenerate 2-photon absorption (2PA) is a process where two photons of the same wavelength (frequency) and polarization are absorbed simultaneously [3]. The concept of reference standards in the 2PA spectroscopy is well established in the literature [47], and consists in calibrating the measurements performed with the sample under study with respect to the measurements performed with a suitable reference under identical conditions. In a generic 2-photon excited fluorescence (2PEF) experiment, one uses a reference standards whose 2PA spectral shape is known to find the so-called 2PA spectral shape correction function, which adjusts for the relative variation of the excitation photon flux in the wavelength range under study. In the second step, the absolute 2PA cross section value, σ2PA, is determined at select wavelengths using a reference standard, whose absolute σ2PA is known, and whose emission spectrum overlaps with that of the system under investigation. The final 2PA spectrum is found by scaling the shape function according to the 2PA cross section. Reference standards are also increasingly used in nonlinear transmittance (NLT) experiments [8], including z-scan [9].

In [7] a set of 2PA fluorophores were characterized using a femtosecond optical parametric amplifier (OPA) operating at 1 kHz pulse repetition rate. However, because the OPA wavelength tuning was inherently discontinuous, and the excitation pulse parameters changed abruptly between the different tuning ranges, especially at the degeneracy point around 800 nm, this previous data may contain sizable uncertainties. Experimental errors may also occur when multi-photon absorption exhibits dependence on the excitation pulse repetition rate. Since many applications use mode-locked femtosecond oscillators operating at, ~100 MHz, it is advisable that applicability of the 2PA values obtained at 1 kHz [7] is independently ascertained at higher pulse rates. Most importantly, using up to 5 orders of magnitude higher pulse repetition rates is usually accompanied by much lower peak photon flux, which, in turn, means that the relative magnitude one-photon excited fluorescence increases relative to 2PEF signal. This issue becomes most critical in case of potential overlap between the 1PA and 2PA spectra [10]. Nevertheless, due to lack of better alternatives, the data presented in [7] continues to be used under many different conditions, even if the consistency and reliability of the standards is not yet fully verified.

In this work, we strive to substantially improve the accuracy of the reference standards, both in terms of the relative 2PA spectral shape as well as regarding the absolute cross section values. Augmented accuracy is most critical for calibration and optimization of fluorophores used in multi-photon microscopy [1113], as well as in the emerging area of quantitative 2PA spectroscopy for measuring the strength of intra- and intermolecular electric fields [14] For this purpose, we have constructed two independent experimental setups, using different 76 – 80 MHz pulse repetition rate femtosecond lasers, where one setup is optimized for the measurement of 2PA spectral shapes and the other is optimized for the absolute cross section measurement, and where we have increased the accuracy of characterization of all critical temporal-, spatial- and spectral parameters of the excitation beam. For the 2PA shape measurement, we use a ~80 MHz pulse repetition rate laser that is continuously tunable without gaps over the wavelength range of commonly-used mode-locked femtosecond sources, 680 – 1050 nm. The quadratic dependence is measured with high fidelity for each wavelength, thus minimizing potential artifacts. The new reference fluorophores set comprises both commercial organic dyes such as Prodan, Coumarin 153 (C153), Fluorescein and Rhodamine 590 (Rh 6G), but also two custom-synthesized compounds 4,4’-Bis-(diphenylamino)-stilbene (BDPAS) and 7,7',7”-(1,3,5-triazine-2,4,6-triyl)tris[9,9-didecyl-N,N-diphenyl 9H-Fluoren-2-amine CAS Registry Number, 517874-02-13 (AF455), were the latter two possess a superior peak 2PA value compared to the commercial counterparts. Perylene, Lucifer yellow and chloroanthracenes show relatively low peak 2PA cross section, σ2PA < 10 GM (1 GM = 10−50 cm4 s photon−1), and were therefore not considered in the present set. Solutions of Rhodamine 610 (Rhodamine B), on the other hand, even though showing a relative large peak σ2PA, exhibit undesired dependence of the absorbance on the concentration and temperature, most likely due to the presence of different equilibrium forms (protonated cation, zwitterion and lactone) [15,16], and were therefore excluded from current measurements. The fluorescence emission spectral range, 375 - 600 nm, is chosen to match the emission wavelengths of common fluorescent microscopy probes such as green fluorescent proteins. Finally, in order to cover a sufficiently broad wavelength range, we take advantage of large solvatochromic and fluorosolvatochromic shifts of the 1PA and 2PA spectra of some of the chromophores.

2. Theoretical considerations

When a monochromatic beam of light propagates through a thin slab of 2-photon absorbing medium, then the difference between the input- and output photon flux (in photon cm−2 s−1) may be expressed as:

ΔI2PA=σ2PANcΔzI2PA2, (1)

where the σ2PA is in cm4 s photon−1, Nc is the 2-photon fluorophore concentration (in cm−3), Δz is the thickness (in cm) and I2PA is the incident photon flux. In most cases, the flux changes only by a relatively small amount, ΔI2PA << I2PA, which makes measurement of ΔI2PA, and, accordingly, accurate determination σ2PA difficult. Alternatively, one can use the relation between the 2PA cross section and the corresponding 2-photon excitation rate,

n˙2PA=12σ2PAI2PA2, (2)

where the latter is determined from experiment by detecting 2PEF emitted by the chromophores, whereas some other detection schemes involving e.g. phosphorescence [17], generation of heat [18] or acoustic waves [19], have also been demonstrated.

Suppose that the 2PEF medium is excited by a periodic train of ultrashort pulses at the rate g (in Hz), and the wavelength, λ2PA. Then the average fluorescence signal may be expressed as:

F2PA(λ2PA)=Δt2PA[gNc12σ2PA(λ2PA)ΔzI2PA2(t,x,y;λ2PA)dxdydt][λminλmaxη(λem)ϕ(λem)dλem], (3)

where Δt2PA is the fluorescence detector integration time, φ(λem) is the differential quantum efficiency of fluorescence emission at the emission wavelength, λem, and η(λem) is the aggregate detection efficiency that accounts for the efficiency of fluorescence collection, spectrometer/diffraction grating through-put, efficiency of the detector, etc. The differential quantum efficiency is defined as,

0φ(λ)dλ=Q, (4)

where Q is the standard quantum yield, equal to the ratio of the radiative decay rate to the total decay rate of the S1 state. The quantity enclosed in the first square brackets in Eq. (3) stands for the average number of molecules excited per second, while the quantity inside the second square brackets accounts for the fluorescence signal that is detected within a finite wavelength interval. If the photon flux, I2PA(t, x, y;λem), and the parameters, η(λem) and φ(λem), were known, then the relation Eq. (3) could be immediately used to evaluate the cross section, σ2PA(λ2PA). As was pointed out above, this information is, however, rarely available. At this point it is convenient to present the photon flux as a product of four factors:

I2PA(t,x,y;λ2PA)=P2PA(λ2PA)taveSaveftime(t)farea(x,y)fcorr(λ2PA), (5)

where P2PA is the total number of photons in the excitation pulse, tave is the average pulse duration, Save is the average beam area, ftime and farea are the normalized functions describing, respectively, the temporal pulse shape and spatial beam profile and fcorr is the so-called wavelength-dependent correction function that quantifies the deviation of the pulse parameters from the average value as a function of λ2PA.

If the Kasha-Vavilov rule is obeyed, then we are allowed to assume that the differential quantum efficiency is independent of the excitation wavelength, and the relative 2-photon absorption spectrum may be obtained then measuring F2PA as function of λ2PA:

σ2PArel(λ2PA)=cnormF2PA(λ2PA)P2PA2(λ2PA)fcorr2(λ2PA). (6)

Since we are dealing here only with the 2PA spectral shape, we can set the empirical normalization factor, cnorm, such that the peak value equals unity, max(σ2PArel) = 1.

To obtain the absolute 2-photon cross section, we need to know the value of the quantity enclosed in the second square bracket in Eq. (3). If we excite the fluorescence by 1-photon absorption in the same sample, using exact same experimental geometry and same fluorescence detection as in the 2-photon measurement, then the corresponding 1-photon excited fluorescence signal may be expressed as:

F1PA=Δt1PA[(110OD)I1PA(x,y)dxdy][λminλmaxη(λem)ϕ(λem)dλem], (7)

where I1PA is the time-average photon flux, Δt1PA is the fluorescence signal integration time and OD = Ncσ1PAΔz is the optical density of the sample and σ1PA is the 1-photon absorption cross section at the 1-photon excitation wavelength. Here the quantity in the first square brackets represents the number of molecules excited per second, whereas the second term is the same as in Eq. (3). Note that the one-photon excitation rate depends neither on the beam spatial profile nor on its temporal structure. By combining Eqs. (3) and (7), we can express the absolute 2PA cross section as:

σ2PA(λ2PA)=F2PA(λex)F1PA(λ1PA)Δt1PAΔt2PA2(110OD)I1PA(x,y)dxdygNcΔz[P2PA(λ2PA)fcorr(λ2PA)ttimeSarea]2ftime2(t)farea2(x,y)dxdydt. (8)

3. Experimental

3.1 Materials, linear spectroscopy and sample preparation

Prodan, C153, Fluorescein and Rh 6G were obtained from Aldrich and were used as received. BDPAS was custom-synthesized by K. Schanze group (U of Florida) as described in [8]. AF455 was provided by Dr. S. Tan from the Air Force Research Laboratory. The synthesis of AF455 is described in [20, 21]. All solvents were purchased from Sigma-Aldrich and were used without further purification. Stock solutions were prepared by mixing the solvent with 1 − 3 mg of dry dye, where the latter was weighed using Mettler-Toledo Model AT2611 analytical balance. Linear absorption spectra were obtained with Shimadzu UV-3600Plus spectrophotometer and corrected fluorescence spectra were measured with Perkin-Elmer Fluorimeter LS55. Extinction coefficients were determined by the dilution method, where a set of daughter solutions with maximum absorbance in the range OD = 0.5 - 1.5 were prepared from the stock solution. The samples were contained 1 cm quartz cuvettes. The chromophore concentration used in the 2PA and 1 PA measurements was in the range 10−6 - 10−3 M.

3.2 Measurement of relative 2PA spectral shape function

Schematic of the 2PA spectral shape measurement setup is shown in Fig. 1(a) . In this part of the experiment, we use 80-MHz repetition rate femtosecond laser (Spectra-Physics InSight DeepSee) continuously tunable in the 680-1300 nm wavelength range with average output power 0.6 – 1.5 W. The laser output was spatially-filtered by focusing the beam through a 50 μm diameter pinhole. To avoid thermal lensing effects in the sample the average laser power was reduced by factor 10 using a 100 Hz chopper wheel. The collimated beam was focused with f = 400 mm achromatic lens (Thorlabs) and passed through a motorized variable transmittance circular neutral density filter (OD = 0.1 – 2.0). Beam reflected from a flat glass plate positioned in front of the sample was detected with a pyroelectric pulse energy probe (Molectron) and was used as reference. The variation of the responsivity of the reference detector at different wavelengths was calibrated relative to a thermoelectric probe (Ophir P1) and did not exceed +/− 2% in the 680 −1100 nm range. The focused beam intensity profile at the sample was measured using a microscope objective and CCD camera (Allied Stingray) in the 680 −990 nm range and InGaAs SWIR camera (Xenics Bobcat 320) in the 970 – 1100 nm range. The measured spot size varied in the range 0.2 – 0.4 mm depending on the wavelength. The pulse duration was measured with custom-adapted auto-correlator using a scanning delay line (ODL-150, Clark MXR) and 0.1 mm thick BBO crystal set up for non-collinear SHG with computer-controlled phase matching angle adjustment. The laser spectrum was measured with a diffraction grating spectrometer (OceanOptics USB4000). The fluorescence from the sample was collected at 90° angle with respect to the propagation direction and vertical linear polarization of the excitation beam. The fluorescence was directed through a stack of short-pass filters to a photomultiplier (Hamamatsu R777) working in current detection mode. The output voltage from the photomultiplier and the reference detector were directed to respective A/D converters (National Instruments USB6009). The sample solutions were contained in 1-2 mm path length spectroscopic cuvettes to minimize effect of absorption by the solvent. Iris diaphragms were used to monitor alignment of the laser beam.

Fig. 1.

Fig. 1

Schematics of experimental set-ups; (a) Measurement of relative 2PA spectra; (b) Measurement of absolute 2PA cross section. SF – spatial filter; L1, L2, L3, L4 – focusing lenses; MO – microscope objective; ND – neutral density filter wheel; SPF – short-pass glass filter; LPF – long-pass glass filter; GP – glass plate; PM – photomultiplier; DAQ – data acquisition, A/D converters; Pol - Glan-Taylor polarizer; PD – photodetector.

The relative 2PA cross section spectra were measured by varying the laser wavelength with 2nm steps in the 680 – 1100 nm range and by measuring at each wavelength the dependence of the fluorescence signal on the incident photon flux by varying the transmittance through the OD filter in 20 discrete steps. The control of the laser wavelength, setting of the OD filter wheel and acquisition of the fluorescence- and reference signals was accomplished using PC via LabView program. The average time for tuning the laser and collecting the data was about 40 - 60 s per one wavelength step.

3.3 Measurement of absolute 2PA cross sections

Schematic of the setup is shown in Fig. 1(b). Here we used a 76-MHz pulse repetition rate mode-locked Ti:Sapphire femtosecond oscillator (Coherent Mira 900) pumped by 10 W cw frequency-doubled Nd:YVO4 laser (Coherent Verdi V-10). The femtosecond laser wavelength was tuned manually in the range 690 - 960 nm with the average output power varying in the range 0.5 – 1.5 W.

The fundamental laser spectrum was measured with diffraction grating spectrometer (OceanOptics USB4000). The pulse temporal shape was measured with a modified optical auto-correlator (INRAD 5-14A), where the variable delay was produced by rotating glass plates (glass thickness 1 mm) and non-collinear second harmonic generation was produced in 0.1 mm BBO crystal. The spatial beam profile at the sample location was measured with the CCD-camera based beam profiler (Thorlabs BC106-VIS) (the sample was removed for these measurements).

To minimize detrimental effect of thermal lensing in the optical elements and in the sample, the average fundamental laser power was reduced by factor 10 using a 100 Hz optical chopper (Thorlabs MC2000). The fundamental power reaching the sample was further varied by manually rotating a λ/2 plate that was positioned in front of a Glan-Taylor polarizer (GL10-B Thorlabs) (Pol). The relative average power of the fundamental beam at the sample was monitored by reflecting a portion of the incident beam to integrating sphere silicon photodetector (Thorlabs S140C) coupled to optical power meter (Thorlabs PM100A). Absolute fundamental power was measured with optical power meter (Coherent FieldMate) with thermoelectric probe (Coherent Powermax PM10) placed directly in front of the sample. Single reflection (~4%) off a glass plate (GP1) was focused on a Type I phase matched BBO crystal that generated second harmonic (blue) light, which was then recombined with the main fundamental wavelength beam using the second glass plate (GP2). At short wavelengths, ~700 nm, the glass plate was replaced by a flipping mirror to compensate for drop in laser output power. A λ/2 plate in front of the SHG crystal rotated the pump beam polarization to the horizontal direction in order to assure that the second harmonic beam had the same (vertical) polarization as the fundamental beam, After GP2 the two beams followed the same path and were incident on the same spot at sample. The blue beam power was adjusted by using continuously variable metallic-coated filter wheel (ND1), and the corresponding power was measured with integrating sphere silicon photodetector (Thorlabs S140C) or with standard silicon photodetector (Thorlabs S120VC) placed directly in front of the sample.

A combination of focusing and collimating lenses (L1 - L3) were used to shape both beams so that they have approximately the same spot size, ~0.3 mm. A color glass long-pass filter (LPF1) was used to cut off residual short-wavelength pump laser light and a glass short-pass filter (SPF1) was used to cut off residual fundamental wavelength after the SHG crystal.

Fluorescence signal was collected in 90° geometry and focused on the entrance slit of a scanning diffraction grating spectrometer (LOMO MDR-12). Scattered laser light was additionally suppressed by a stack of short-pass color glass filters (SPF2). The fluorescence signal was detected with a photon counting module (Hamamatsu H6240-01) coupled to a frequency counter with PC readout.

The absolute cross-section was determined at a fixed wavelength by measuring the fluorescence signal on the incident photon flux in both 1PA and 2PA successively and all the laser characterization, e.g. spectral profile, beam profile and temporal profile. Average time needed to evaluate absolute 2PA cross section at 1 wavelength, including stabilizing the laser mode-locking, setting the laser wavelengths and performing the beam characterization measurements, was about 4 h.

4. Results

The normalized fluorescence spectra are shown in Fig. 2 . The corresponding peak- and min/max fluorescence wavelengths are listed in Table 1 .

Fig. 2.

Fig. 2

Corrected fluorescence spectra of (1) BDPAS in methylene chloride; (2) Prodan in toluene; (3) Prodan in DMSO; (4) C153 in toluene; (5) C153 in DMSO; (6) AF455 in toluene; (7) AF455 in THF; (8) Fluorescein in H2O pH11 buffer; (9) Rh 6G in Methanol.

Table 1. 1-photon and 2-photon photophysical properties of the systems studied. The σ2PA and Δσ2PA values are obtained by averaging over all measurements performed.

Comp. Solvent ε (λ1PA) min λem–max λ em (peakλem) σ2PA(λ2PA) Δσ2PA
M-1 cm-1 (nm) nm GM (nm) ±%
1 BDPAS DCM 52.6x103 (388) 415 – 540 (443) 175 (690)
138 (700)
8
8
2 Prodan toluene 19.8x103 (349) 390 – 480 (414) 19 (700) 6
3 Prodan DMSO 17.5x103 (358) 420 – 550 (459) 20 (723) 8
4 C153 toluene 20.5x103 (408) 440 – 600 (468) 17 (816) 5
5 C153 DMSO 18.5x103 (427) 480 – 650 (540) 17 (851) 7
6 AF455 toluene 117x103 (419) 430 – 550 (453) 404 (784) 7
7 AF455 THF 106x103 (415) 450 – 630 (504) 392 (784) 6
8 Fluorescein H2O pH11 88.6x103 (491)
8.7x103 (322)
490 – 580 (514) 26 (785)
6.5(860)
5
8
9 Rh 6G MetOH 122x103 (528)
12.1x103 (347)
580 – 630 (554) 79 (812)
202 (692)
8
8

The 2PA spectral shapes are presented in Fig. 3 by black symbols, and the corresponding absolute cross section measurements are shown as red squares. The shape functions are scaled to give the best match with average absolute cross sections at the select wavelengths. The linear extinction spectra are shown by blue solid line. The two panels (left and right) present the same data in the linear and logarithmic vertical scales. The lower horizontal axis of the plots is calibrated in the wavelength of the 2-photon excitation (λ2PA), while the upper axis is calibrated in 1-photon absorption wavelength (λ1PA). The same data is presented in Table 2 in the Appendix.

Fig. 3.

Fig. 3

2PA spectra of (1) BDPAS in methylene chloride; (2) Prodan in toluene; (3) Prodan in DMSO; (4) C153 in toluene; (5) C153 in DMSO; (6) AF455 in toluene; (7) AF455 in THF; (8) Fluorescein in H2O pH11 buffer; (9) Rh 6G in methanol.

Table 2. 2-photon cross sections (GM) of the dyes at selected wavelengths (nm). The maximum relative error of the numbers shown here is given in Table 1. Please note that the values in Table 1 are obtained directly from measuring the absolute cross section at few select wavelengths, and as shown in Fig. 3, whereas the values shown here are the best fit by scaling the experimentally measured shape function according to the absolute cross section data.

λ2PA
(nm)
σ2PA (GM)
BDPAS
MethCl
Prodan
toluene
Prodan
DMSO
C153
toluene
C153
DMSO
AF455
toluene
AF455
THF
Fluor.
H2O pH11
Rh6G
MetOH
680 179 17 16 4.9 6.4 122 123 17 162
690 166 19 19 5.0 5.8 119 120 12 248
700 146 19 20 5.0 5.3 120 123 11 237
710 108 17 20 4.8 4.6 120 126 11 145
720 74 15 20 4.9 4.2 137 144 13 76
730 46 14 20 5.4 4.2 177 186 16 51
740 28 13 19 6.4 4.5 231 240 19 43
750 20 12 18 7.6 5.2 284 288 21 44
760 16 10 17 9.1 6.3 331 327 23 50
770 15 8.9 16 11 7.7 374 366 24 60
780 15 6.6 14 13 9.3 403 390 26 65
790 13 4.0 12 14 11 396 388 26 70
800 11 2.1 8.7 16 13 372 374 24 73
810 10 1.1 6.4 17 14 350 361 21 78
820 8.3 0.62 4.3 17 15 330 341 16 76
830 6.3 0.39 2.6 16 16 309 310 11 62
840 4.4 0.31 1.5 16 17 290 285 7.8 44
850 2.6 0.26 0.83 15 17 260 260 6.2 26
860 1.3 0.21 0.42 14 17 210 223 5.8 14
870 0.64 0.15 0.20 12 16 148 180 6.2 7.5
880 0.24 0.08 0.09 8.6 15 83 131 7.0 4.7
890   0.04 0.04 5.7 14 42 89 8.3 4.3
900     0.02 3.3 13 20 59 10 5.1
910       1.7 11 7.6 36 12 6.3
920       0.75 9.0 2.8 20 13 7.5
930       0.31 7.0   11 12 8.9
940       0.10 5.2   4.9 9.8 11
950         3.7   2.3 7.3 14
960         2.5   1.1 6.1 18
970         1.6   0.54 5.4 21
980         0.94     5.2 24
990         0.51     4.6 24
1000               3.3 21
1010               2.1 17
1020               1.2 13
1030                 11
1040                 11
1050 12

Key 2-photon data for all 9 standards, along with the peak molecular extinction coefficient, and estimated maximum fluorophore concentration and the solution stability assessment, is collected in Table 1. The peak cross section value for BDPAS in methylene chloride (1), σ2PA = 175 ± 14 GM, is less than was reported earlier [7], which we attribute to relatively rapid photo-degradation and low dark stability of the solution [8]. In our current measurements all precautions were taken by continuously monitoring the sample for signs of potential degradation. Note that BDPAS is especially useful in case of blue-emitting fluorophores, and in the wavelength range, λ2PA = 680 – 860 nm. At longer wavelengths, the cross section drops below 1 GM. Prodan in toluene (2) and in DMSO (3) have also relatively short wavelength fluorescence emission. These solutions showed no measurable degradation over a period of several months, even though their peak cross section is also lower, σ2PA ~17 GM. Remarkable feature of C153 in DMSO is that its 2PA and 1PA profiles practically coincide in the range λ2PA = 740 - 1000 nm. Physical background of this phenomenon was recently discussed in [14]. The peak cross section of Fluorescein (aqueous, pH11) (8) is σ2PA = 26 ± 1.2 GM at 780 nm. This is again about factor 2 less than the earlier reported value [7], most likely because this wavelength coincides with degeneracy of the OPA, where the beam parameters may deteriorate. In our current measurement this issue did not occur. We should note that relative 2-photon cross sections reported for Prodan and C153 in [14] were measured using the Fluorescein data from [7], which may have led to over-estimation of σ2PA. Rh 6G in methanol (9) absorbs and emits at the longest wavelengths,compared to other fluorophores in the current set. The absolute cross section at 810 nm is, σ2PA = 79 ± 6 GM, and correlates well with the value reported in [5]. However, both the absolute value and the shape function are higher at 690 nm. This discrepancy is most likely caused by the narrow spectral band at, λ2PA = 680 - 710 nm. The spectral FWHM of the excitation pulses in that particular wavelength range was measured to be Δλ2PA = 3 - 4 nm, i.e. is comparable to the half-width of the named band. For comparison, in the earlier measurements the pulses were spectrally about factor of two broader [5], which may explain why the spectral feature appears more pronounced in the current data. We must underline that,contrary to 1-photon spectroscopy, where spectral retrieval via deconvolution can be quite effective, inherent nonlinear nature of the 2PA precludes application of such straightforward techniques [22].

5. Evaluation of experimental uncertainty

Main causes of experimental uncertainty in the relative spectral shape measurement are due to (a) deviation from the quadratic dependence of the 2PEF signal on the photon flux and (b) errors in the determination of the temporal- and spatial shape of the laser beam. The quadratic power dependence was ascertained at each wavelength by varying the average incident power over ~2 orders of magnitude, and by fitting the resulting 2PEF signal vs. power data, presented in double-logarithmic scale, with a linear function. The measurements accepted have a power law coefficient in the 1.96-2.04 range. Upper panels in Fig. 4 shows the power law coefficient in C153 in DMSO (left) and in Fluorescein (right) obtained in the relative 2PA spectral shape experiment (empty symbols). The power dependence measurement was performed also on the absolute 2PA experimental systems (filled rectangles), and showed very similar behavior. Even though different measurement systems behave slightly differently, the maximum deviation from the exact quadratic dependence does not exceed 3%. In order to further verify the measured spectral shape functions, we performed a relative 2PA shape measurement for a few select standards using the absolute 2PA measurement setup. Lower panels in Fig. 4 show the corresponding normalized shape functions for C153 in DMSO (left) and in Fluorescein (right), measured by the two complementary experiments. The discrepancy between the two measurements does not exceed 4%. This allows us to estimate that the overall uncertainty of the 2-photon shape functions presented here is about 5%. Since the two experiments were truly independent, we conclude that the relative accuracy of measuring the photon flux and, accordingly, the relative beam spatial- and temporal profiles, was also about5% or less. In the case of the absolute cross section measurement, there is an additional uncertainty due to the measurement of the 1-photon excited fluorescence. Firstly, for this measurement, the 1-photon excitation beam should be aligned to illuminate the exact same sample volume as the 2PEF excitation beam. If the beams are even slightly misaligned, then the two fluorescence signals may no longer be collected from the exact same sample volume. Secondly, because the maxima of the 2-photon and 1-photon spectra do not always coincide, when we tune the 2-photon excitation wavelength near the peak of the 2PA spectrum, then the corresponding 1-photon excitation wavelength may be located where the linear absorbance is very low or changes abruptly, thus making it difficult to accurately determine how many photons are absorbed in the sample When combined with the above 5% error due to the characterization of the excitation beam, we arrived at the estimated maximum uncertainty of the 2-photon cross section value of about 8%.

Fig. 4.

Fig. 4

. Comparison between independent 2PEF measurement techniques in two reference samples: C153 in DMSO (left) and Fluorescein in aqueous, pH11 (right). Upper panel: Experimentally determined power law coefficient as a function of laser wavelength measured using the scanning laser setup (empty symbols) and manually-tuned laser setup (filled symbols). Lower panel: 2PA shape functions measured by the scanning laser setup (empty symbols) and manually tuned laser setup (filled symbols). The maximum value of the shape functions is normalized to unity. The manually-tuned data is averaged over 9 measurements in C153 and 3 measurements in Fluorescein.

6. Application notes

How to apply reference standards in the 2PEF-based measurements was described previously in [7]. Here, we would like to briefly discuss the utilization of the reference standards for augmenting nonlinear transmittance-type experiments [2, 9, 24, 25]. According to the Eq. (1), the photon flux passing through a thin layer of 2-photon absorbers decreases in proportion to the number density of the absorbers. In case of finite thickness, it is useful to introduce the effective NLT strength,

κ2PA=σ2PANcd, (9)

where d is the sample thickness. If d is expressed in cm, the chromophore concentration Nc is given in mM (10−3 M) and σ2PA is expressed in GM, then the maximum effective NLT strength would be typically in the range, κ2PA = 1 - 102. The chromophores described here have been selected to exhibit good solubility in their respective solvents, with the maximum concentration, ~5 – 10 mM. By using d = 1 cm and selecting the standards that have σ2PA > 10 GM, an sufficiently large value, κ2PA > 10, may be achieved in the whole wavelength range 680 - 1050 nm. Also, because the reference samples are often exposed to intense laser light, the solutions were evaluated for photo-stability as well as for dark storage stability over a period of about 1 month. Out of the samples tested, only BDPAS in DCM showed an increased photo-induced decomposition [8], which was accompanied by relatively short dark stability of less than ~1 week. One might assume that larger κ2PA would allow larger maximum absolute change of the transmittance, |ΔTmax|, which, in turn, would facilitate more accurate determination of σ2PA values. However, depending on the beam parameters, the nonlinear transmittance may exhibit a quite complicated dependence on the incident photon flux [25]. For this reason, special care should be taken not to use excessively large κ2PA values, especially if the sample and the reference have very different NLT signals. Based on practical experience, if the maximum transmittance change of both the sample and the reference is, |ΔTmax| < 10%, then one can obtain the 2PA spectrum of the sample under study by using the relation,

σ2PA(λ2PA)=σ2PAref(λ2PA)BNLT(λ2PA)BNLTref(λ2PA)NcrefdrefNcd, (10)

where σ2PAref is the 2PA spectrum of the reference standard, Ncref and dref are, respectively, the concentration and the thickness of the reference sample and the coefficients BNLT and BrefNLT are obtained from fitting the measured nonlinear transmittance as a function of the number of incident photons with the linear function,

flin=1BNLTI2PA(t,x,y;λ2PA)dxdydt. (11)

Still another potential issue in the NLT measurements stems from near-IR absorption of common solvents. For example, at the wavelengths > 900 nm both toluene and DMSO show peak absorbance, Amax~0.1, in 1 cm cuvette. One way to minimize the uncertainty caused by solvent absorption would be to either use the same solvent for the reference as for the system under study, or if that is not feasible, then to subtract from the 2PA spectrum given by the Eq. (10) the artifacts that may be present when the measurement is performed with the neat solvent.

7. Conclusions

We presented absolute two-photon absorption spectra of a series of organic fluorophores in the excitation wavelength range, 680 - 1050 nm, By using stable femtosecond lasers and by cross-checking independently performed measurements, we have achieved accuracy of at least 5% for the shape of the 2PA spectra and 8% for in the absolute 2PA cross section values. This constitutes at least a factor of 4 - 5 improvement compared to the previously established 2-photon reference standards. The chromophores were selected to provide improved solubility and stability and are therefore well suited not only for the 2PEF-based experiments, but also for calibration of nonlinear transmission measurement, which often require higher sample concentration. The new data further alleviates the need for tedious characterization of the excitation laser parameters and allows for quantitative comparison and optimization of molecular probes used in multiphoton microscopy and imaging, for quantifying the multiphoton absorption efficiency of chromophores in different environments, as well as many other applications of nonlinear optics and -spectroscopy.

Acknowledgments

This work was supported by Ministry of Education and Science, Republic of Estonia grant IUT23-9. Part of this work was supported by European Union ITN project TOPBIO and by NIH Grants R01 GM098083 and 1U01NS094246-01. We thank Jake Lindquist and Caleb Stoltzfus for their help with LabView programming.

Appendix.

References and links

  • 1.Lakowicz J. R., Principles of Fluorescence Spectroscopy, 3rd ed. (Springer, 2006). [Google Scholar]
  • 2.Southerland R. L., Handbook of Nonlinear Optics, 2nd ed. (Marcel Dekker, 2003). [Google Scholar]
  • 3.Göpper-Maier M., “Über Elementarakte mit zwei Quantensprüngen,” Ann. Phys. 9(3), 273–294 (1931). 10.1002/andp.19314010303 [DOI] [Google Scholar]
  • 4.Hermann J. P., Ducuing J., “Absolute measurement of two-photon cross sections,” Phys. Rev. A 5(6), 2557–2568 (1972). 10.1103/PhysRevA.5.2557 [DOI] [Google Scholar]
  • 5.Xu C., Webb W. W., “Measurement of two-photon excitation cross sections of molecular fluorophores with data from 690 to 1050 nm,” J. Opt. Soc. Am. B 13(3), 481–491 (1996). 10.1364/JOSAB.13.000481 [DOI] [Google Scholar]
  • 6.Albota M. A., Xu C., Webb W. W., “Two-photon fluorescence excitation cross sections of biomolecular probes from 690 to 960 nm,” Appl. Opt. 37(31), 7352–7356 (1998). 10.1364/AO.37.007352 [DOI] [PubMed] [Google Scholar]
  • 7.Makarov N. S., Drobizhev M., Rebane A., “Two-photon absorption standards in the 550-1600 nm excitation wavelength range,” Opt. Express 16(6), 4029–4047 (2008). 10.1364/OE.16.004029 [DOI] [PubMed] [Google Scholar]
  • 8.Dubinina G. G., Price R. S., Abboud K. A., Wicks G., Wnuk P., Stepanenko Y., Drobizhev M., Rebane A., Schanze K. S., “Phenylene vinylene platinum(II) acetylides with prodigious two-photon absorption,” J. Am. Chem. Soc. 134(47), 19346–19349 (2012). 10.1021/ja309393c [DOI] [PubMed] [Google Scholar]
  • 9.Bridges R. E., Fischer G. L., Boyd R. W., “Z-scan measurement technique for non-Gaussian beams and arbitrary sample thicknesses,” Opt. Lett. 20(17), 1821–1824 (1995). 10.1364/OL.20.001821 [DOI] [PubMed] [Google Scholar]
  • 10.Drobizhev M., Karotki A., Kruk M., Krivokapic A., Anderson H. L., Rebane A., “Photon energy upconversion fluorescence in porphyrins: One-photon hot-band absorption versus two-photon absorption,” Chem. Phys. Lett. 370(5-6), 690–699 (2003). 10.1016/S0009-2614(03)00162-3 [DOI] [Google Scholar]
  • 11.Mütze J., Iyer V., Macklin J. J., Colonell J., Karsh B., Petrášek Z., Schwille P., Looger L. L., Lavis L. D., Harris T. D., “Excitation spectra and brightness optimization of two-photon excited probes,” Biophys. J. 102(4), 934–944 (2012). 10.1016/j.bpj.2011.12.056 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Velasco M. G., Allgeyer E. S., Yuan P., Grutzendler J., Bewersdorf J., “Absolute two-photon excitation spectra of red and far-red fluorescent probes,” Opt. Lett. 40(21), 4915–4918 (2015). 10.1364/OL.40.004915 [DOI] [PubMed] [Google Scholar]
  • 13.Cheng L.-C., Horton N. G., Wang K., Chen S.-J., Xu C., “Measurements of multiphoton action cross sections for multiphoton microscopy,” Biomed. Opt. Express 5(10), 3427–3433 (2014). 10.1364/BOE.5.003427 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Rebane A., Wicks G., Drobizhev M., Cooper T., Trummal A., Uudsemaa M., “Two-photon voltmeter for measuring a molecular electric field,” Angew. Chem. Int. Ed. Engl. 54(26), 7582–7586 (2015). 10.1002/anie.201502157 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Arbeloa I. L., Ojeda P. R., “Molecular forms of rhodamine B,” Chem. Phys. Lett. 79(2), 347–350 (1981). 10.1016/0009-2614(81)80219-9 [DOI] [Google Scholar]
  • 16.Hinckley D. A., Seybold P. G., Borris D. P., “Solvatochromism and thermochromism of rhodamine solutions,” Spectrochimica Acta 42A, 741–754 (1986). [Google Scholar]
  • 17.McIlroy S. P., Cló E., Nikolajsen L., Frederiksen P. K., Nielsen C. B., Mikkelsen K. V., Gothelf K. V., Ogilby P. R., “Two-photon photosensitized production of singlet oxygen: sensitizers with phenylene-ethynylene-based chromophores,” J. Org. Chem. 70(4), 1134–1146 (2005). 10.1021/jo0482099 [DOI] [PubMed] [Google Scholar]
  • 18.Rodriguez L., Ahn H.-Y., Belfield K. D., “Femtosecond two-photon absorption measurements based on the accumulative photo-thermal effect and the Rayleigh interferometer,” Opt. Express 17(22), 19617–19628 (2009). 10.1364/OE.17.019617 [DOI] [PubMed] [Google Scholar]
  • 19.Bae Y., Song J., Kim Y., “Photoacoustic study of two‐photon absorption in hexagonal ZnS,” J. Appl. Phys. 53(1), 615–619 (1982). 10.1063/1.329967 [DOI] [Google Scholar]
  • 20.R. Kannan, L.-S. Tan, and R. A. Vaia, “Two-photon responsive chromophores containing electron accepting cores,” US Patent 6,555,682 (2003).
  • 21.Kannan R., He G. S., Lin T.-C., Prasad P. N., Vaia R. A., Tan L.-S., “Toward highly active two-photon absorbing liquids. Synthesis and characterization of 1,3,5-triazine-based octupolar molecules,” Chem. Mater. 16(1), 185–194 (2004). 10.1021/cm034358g [DOI] [Google Scholar]
  • 22.Xu B., Coello Y., Lozovoy V. V., Dantus M., “Two-photon fluorescence excitation spectroscopy by pulse shaping ultrabroad-bandwidth femtosecond laser pulses,” Appl. Opt. 49(32), 6348–6353 (2010). 10.1364/AO.49.006348 [DOI] [PubMed] [Google Scholar]
  • 23.Rebane A., Drobizhev M., Makarov N. S., Wicks G., Wnuk P., Stepanenko Y., Haley J. E., Krein D. M., Fore J. L., Burke A. R., Slagle J. E., McLean D. G., Cooper T. M., “Symmetry breaking in platinum acetylide chromophores studied by femtosecond two-photon absorption spectroscopy,” J. Phys. Chem. A 118(21), 3749–3759 (2014). 10.1021/jp5009658 [DOI] [PubMed] [Google Scholar]
  • 24.Sheik-Bahae M., Said A. A., Wei T.-H., Hagan D. J., Van Stryland E. W., “Sensitive measurement of optical nonlinearities using a single beam,” IEEE J. Quantum Electron. 26(4), 760–769 (1990). 10.1109/3.53394 [DOI] [Google Scholar]
  • 25.Rumi M., Perry J. W., “Two-photon absorption: an overview of measurements and principles,” Adv. Opt. Phot. 2, 451–518 (2010). [Google Scholar]

Articles from Optics Express are provided here courtesy of Optica Publishing Group