Can Transcutaneous CO2 Tension Be Used to Calculate Ventilatory Dead Space? A Pilot Study

Pradeep H Lakshminarayana 1,*, Adiba A Geeti 1, Umer M Darr 2, David A Kaufman 1
PMCID: PMC5027368  PMID: 27688911

Abstract

Dead space fraction (V d/V t) measurement performed using volumetric capnography requires arterial blood gas (ABG) sampling to estimate the partial pressure of carbon dioxide (PaCO2). In recent years, transcutaneous capnography (PtcCO2) has emerged as a noninvasive method of estimating PaCO2. We hypothesized that PtcCO2 can be used as a substitute for PaCO2 in the calculation of V d/V t. In this prospective pilot comparison study, 30 consecutive postcardiac surgery mechanically ventilated patients had V d/V t calculated separately using volumetric capnography by substituting PtcCO2 for PaCO2. The mean V d/V t calculated using PaCO2 and PtcCO2 was 0.48 ± 0.09 and 0.53 ± 0.08, respectively, with a strong positive correlation between the two methods of calculation (Pearson's correlation = 0.87, p < 0.05). Bland-Altman analysis showed a mean difference of −0.05 (95% CI: −0.01 to −0.09) between the two methods. PtcCO2 measurements can provide a noninvasive means to measure V d/V t, thus accessing important physiologic information and prognostic assessment in patients receiving mechanical ventilation.

1. Introduction

The ventilatory dead space fraction (V d/V t) is defined as the portion of tidal volume that does not participate in gas exchange because it does not reach perfused lung units. Clinically, V d/V t may be measured using volumetric capnography and the Enghoff modification of the Bohr equation, which states

VdVt=PaCO2PeCO2PaCO2, (1)

where PaCO2 is the partial pressure of carbon dioxide in the arterial blood and PeCO2 is the partial pressure of carbon dioxide in exhaled air [1]. V d/V t may increase in a variety of pathological situations, especially conditions with obstruction of the pulmonary vasculature [2]. Measuring V d/V t requires arterial blood gas (ABG) sampling to measure PaCO2 [3]. Transcutaneous capnography (PtcCO2) has emerged as a noninvasive method of estimating PaCO2 [4]. The objective of our study was to evaluate the relationship between PtcCO2 and PaCO2 in calculating V d/V t. We hypothesized that PtcCO2 can be used as a substitute for PaCO2 in the calculation of V d/V t.

2. Methods

This was a prospective pilot comparison study conducted at a single 383-bed urban, community-teaching hospital. We recruited thirty consecutive adult patients admitted to the Surgical Intensive Care Unit (SICU) after cardiac surgery from May 1, 2014, until December 1, 2014. We excluded patients who were extubated within 4 hours of arrival to the SICU. The Bridgeport Hospital Institutional Review Board (IRB) approved this study.

Volumetric capnography was performed within 12 hours of patient arrival to the SICU. A CAPNOSTAT® CO2 sensor was attached between the subject's endotracheal tube and the Y-piece of the ventilator circuit to obtain breath-by-breath volumetric capnography using NICO2 volumetric capnography (Philips-Respironics®, Wallingford, CT). We measured PtcCO2 using the Tosca 500 (Radiometer®, Copenhagen, Denmark), according to the manufacturer's instructions. A fully automated calibration of the transcutaneous CO2 sensor (Tosca sensor 92) containing a Stow-Severinghaus type electrode was performed using CAL-Gas (Radiometer), which contained 12% O2, 7% CO2, and 81% N2 prior to each monitoring period and every time the membrane of the transcutaneous sensor was changed (every 14 days). First, the sensor probe and the subject's earlobe were cleaned with alcohol and dried prior to each measurement. Then, the sensor was placed on the earlobe after placing a drop of contact solution provided by the manufacturer and secured by means of an adhesive clip. No PtcCO2 values were collected until the ear lobe reached an appropriate temperature of 42°C and stable signals for both PtcCO2 and PeCO2 were achieved after at least 15 minutes. The PtcCO2 was recorded at the same time an arterial blood sample was collected from an existing radial arterial line, placed on ice, and sent immediately for standard blood gas analysis. PaCO2 was measured using the arterial blood gas sample analyzed with Cobas b221 blood gas analyzer (Roche®). Turnaround time for the result was under 10 minutes for all samples. We calculated V d/V t using both simultaneously measured values for PaCO2 and PtcCO2 using NICO2 volumetric capnography according to the manufacturer's instructions. In this study, we used single measurements of PeCO2, PaCO2, and PtcCO2 to calculate V d/V t, with all values obtained simultaneously. The V d/V t calculated using PaCO2 was considered the reference standard [5].

2.1. Statistical Analysis

Statistical analysis was performed using SPSS version 9.0. Results from descriptive statistics are presented as mean ± standard deviation (SD). Linear regression and Bland-Altman analyses were performed to compare the values of V d/V t using PtcCO2 and PaCO2 measurements.

3. Results

30 consecutive postcardiac surgery patients were included and 30 separate volumetric capnography measurements were obtained. Statistical description of the study population can be found in Table 1. 24 patients were males, mean age was 68.25 ± 6.36 years, and the mean Acute Physiology and Chronic Health Evaluation (APACHE) II score was 14.5 ± 3.92. The mean V d/V t calculated using PaCO2 and PtcCO2 was 0.48 ± 0.09 and 0.53 ± 0.08, respectively, with a strong positive correlation between the two methods of calculation (Pearson's correlation = 0.87, p < 0.05) as shown in Figure 1. Bland-Altman analysis showed a mean difference of −0.05 (95% CI: −0.01 to −0.09) between the two methods of V d/V t measurements as shown in Figure 2.

Table 1.

Patient demographic and clinical characteristics (n = 30).

Mean age (years) ± SD 68.25 ± 6.36
Males 24
Types of cardiac surgery
 Valve replacement 12
 CABG 12
 Valve + CABG 4
 Aneurysm 2
Cardiopulmonary bypass status
 On bypass 26
 Off bypass 4
Mean APACHE II score ± SD 14.5 ± 3.92
Patient comorbidities1
 Cardiovascular 28
 Respiratory 6
 Endocrine 8
 Neurology 3
 Miscellaneous 4
Vasopressor or inotrope support 14
Mean tidal volume on ventilator2 (mL) ± SD 660 ± 95.95
Mean PaCO2 (mm of Hg) ± SD 35.93 ± 5.19
Mean PtcCO2 (mm of Hg) ± SD 39.73 ± 5.82
Mean shunt fraction (Q s/Q t)3 0.19 ± 0.08

1Patients may have more than one comorbidity.

Miscellaneous comorbidities included cirrhosis, chronic kidney disease, HIV and malignancy.

2All patients maintained on volume assist control mode of mechanical ventilation.

3Measured only in 27 patients who had a pulmonary artery catheter.

Abbreviations: CABG: coronary artery bypass graft; APACHE: acute physiology and chronic health evaluation.

Figure 1.

Figure 1

Scatter plot of V d/V t using PaCO2 against V d/V t using PtcCO2.

Figure 2.

Figure 2

Modified Bland-Altman plot of V d/V t(PaCO2) − V d/V t(PtcCO2) against mean of V d/V t(PaCO2) and V d/V t(PtcCO2).

4. Discussion

Measuring V d/V t at the bedside can be easily performed with volumetric capnography, in which both exhaled tidal volume and PeCO2 are measured. The PeCO2 is compared to PaCO2 determined from the arterial blood, and the Enghoff modification of the Bohr equation is used to calculate V d/V t [3]. V d/V t has been shown to be useful in identifying cardiac surgery patients with microthrombosis of the pulmonary circulation [6] and ARDS patients who have an increased risk for death [7]. It has also been proposed to help identify patients with pulmonary embolism [8] and to risk-stratify intubated patients for extubation failure [9]. A drawback of this method is that it requires sampling of the arterial blood, which can be painful and may cause complications. Thus, a noninvasive test to estimate V d/V t might be helpful.

A transcutaneous measurement of CO2 is based on the principle that an increase in skin capillary blood flow facilitates diffusion of CO2, hence allowing its detection by a sensor located at the skin surface. The sensor is also equipped with a thermostatically controlled heater unit, which allows an increase in temperature of the skin surface. The Stove-Severinghaus type CO2 sensor is a potentiometric sensor combining a silver/silver chloride reference electrode and a miniature glass pH electrode. PtcCO2 is determined by a change in pH of the electrolyte solution [4, 10].

PtcCO2 is a noninvasive method of estimating PaCO2 used frequently in clinical practice. Its application has been widely noted during mechanical ventilation, anesthesia, bronchoscopy, and sleep studies [4]. Whether substituting PtcCO2 for PaCO2 yields accurate measurements of V d/V t is unreported in the scientific literature. We performed a prospective pilot comparison to explore whether using PtcCO2 in place of PaCO2 in the Enghoff-Bohr calculation would yield accurate estimates of V d/V t. Our observations show a strong positive correlation between both methods of estimating V d/V t with good agreement between the two techniques.

Our study has some limitations. First, the sample size is small. We consider our findings to be preliminary. They will require confirmation in a larger cohort of patients. Second, we conducted our study exclusively in patients who had undergone cardiac surgery, both emergent and elective. We chose this group because it is relatively homogeneous with regard to complicating respiratory and metabolic disorders, thereby minimizing residual confounders that might bias the results. Whether our findings can be extrapolated to other critically ill populations will require further investigation. For example, it is plausible that transcutaneous CO2 measurements will be less accurate in patients with circulatory failure or shock, as peripheral blood flow may be diminished. Finally, our goal was limited. We sought to test how well V d/V t calculations using PtcCO2 measurements compared to V d/V t calculations using PaCO2. We did not employ a reference standard measurement of V d/V t such as mixed inert gas elimination technique, as such techniques are found only in highly specialized research centers. It is possible that unmeasured confounders that affect volumetric capnography might result in random or systematic biases that make our findings unreliable. Indeed, PtcCO2 could be influenced by advanced age, hypothermia, hypercapnia, and use of vasopressor and inotropic support. A larger observational study would be essential to determine the influence of these potential confounders on V d/V t measurement. Nonetheless, we believe that our methods represent a real-world approach to use bedside V d/V t measurements and that PtcCO2 may be an acceptable substitute for PaCO2 to calculate V d/V t.

Thus, we propose that substituting PtcCO2 for PaCO2 may be clinically useful and provide a noninvasive means for estimating V d/V t. Using a noninvasive CO2 measurement to calculate V d/V t potentially could permit more frequent estimates of V d/V t when multiple samples of arterial blood are unavailable or invasive procedures are not desired. For example, it is plausible that following the V d/V t over the course of a spontaneous breathing trial could be useful at identifying patients at high risk for requiring continued mechanical ventilator support, as has been shown in pediatric patients with respiratory failure [11]. Future work incorporating noninvasive CO2 measurements into V d/V t calculations should focus on whether these estimates are valid in a broader population of critically ill patients and whether V d/V t measurements can be incorporated into treatment strategies to yield improved patient-centered outcomes.

5. Conclusion

PtcCO2 may be a useful substitute for PaCO2 to calculate the dead space fraction. PtcCO2 measurements can provide a noninvasive means to measure the dead space fraction, thus allowing us to obtain important physiologic information and prognostic assessment in critically ill patients on mechanical ventilation.

Competing Interests

The authors declare that they have no competing interests.

References

  • 1.Sinha P., Flower O., Soni N. Deadspace ventilation: a waste of breath! Intensive Care Medicine. 2011;37(5):735–746. doi: 10.1007/s00134-011-2194-4. [DOI] [PubMed] [Google Scholar]
  • 2.Blanch L., Romero P. V., Lucangelo U. Volumetric capnography in the mechanically ventilated patient. Minerva Anestesiologica. 2006;72(6):577–585. [PubMed] [Google Scholar]
  • 3.Tusman G., Sipmann F. S., Bohm S. H. Rationale of dead space measurement by volumetric capnography. Anesthesia & Analgesia. 2012;114(4):866–874. doi: 10.1213/ane.0b013e318247f6cc. [DOI] [PubMed] [Google Scholar]
  • 4.Eberhard P. The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions. Anesthesia & Analgesia. 2007;105(6):S48–S52. doi: 10.1213/01.ane.0000278642.16117.f8. [DOI] [PubMed] [Google Scholar]
  • 5.Suarez-Sipmann F., Santos A., Böhm S. H., Borges J. B., Hedenstierna G., Tusman G. Corrections of Enghoff's dead space formula for shunt effects still overestimate Bohr's dead space. Respiratory Physiology and Neurobiology. 2013;189(1):99–105. doi: 10.1016/j.resp.2013.06.020. [DOI] [PubMed] [Google Scholar]
  • 6.Dixon B., Campbell D. J., Santamaria J. D. Elevated pulmonary dead space and coagulation abnormalities suggest lung microvascular thrombosis in patients undergoing cardiac surgery. Intensive Care Medicine. 2008;34(7):1216–1223. doi: 10.1007/s00134-008-1042-7. [DOI] [PubMed] [Google Scholar]
  • 7.Nuckton T. J., Alonso J. A., Kallet R. H., et al. Pulmonary dead-space fraction as a risk factor for death in the acute respiratory distress syndrome. The New England Journal of Medicine. 2002;346(17):1281–1286. doi: 10.1056/nejmoa012835. [DOI] [PubMed] [Google Scholar]
  • 8.Kline J. A., Meek S., Boudrow D., Warner D., Colucciello S. Use of the alveolar dead space fraction (Vd/Vt) and plasma D-dimers to exclude acute pulmonary embolism in ambulatory patients. Academic Emergency Medicine. 1997;4(9):856–863. doi: 10.1111/j.1553-2712.1997.tb03809.x. [DOI] [PubMed] [Google Scholar]
  • 9.González-Castro A., Suárez-Lopez V., Gómez-Marcos V., et al. Utility of the dead space fraction (Vd/Vt) as a predictor of extubation success. Medicina Intensiva. 2011;35(9):529–538. doi: 10.1016/j.medin.2011.05.016. [DOI] [PubMed] [Google Scholar]
  • 10.Rubortone S. A., De Carolis M. P., Lacerenza S., Bersani I., Occhipinti F., Romagnoli C. Use of a combined SpO2/PtcCO2 sensor in the delivery room. Sensors. 2012;12(8):10980–10989. doi: 10.3390/s120810980. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Hubble C. L., Gentile M. A., Tripp D. S., Craig D. M., Meliones J. N., Cheifetz I. M. Deadspace to tidal volume ratio predicts successful extubation in infants and children. Critical Care Medicine. 2000;28(6):2034–2040. doi: 10.1097/00003246-200006000-00059. [DOI] [PubMed] [Google Scholar]

Articles from Critical Care Research and Practice are provided here courtesy of Wiley