Skip to main content
HHS Author Manuscripts logoLink to HHS Author Manuscripts
. Author manuscript; available in PMC: 2016 Sep 22.
Published in final edited form as: Clin Biochem. 2015 Feb 7;48(4-5):204–212. doi: 10.1016/j.clinbiochem.2015.01.014

Effectiveness of Practices for Improving the Diagnostic Accuracy of Non-ST Elevation Myocardial Infarction in the Emergency Department: A Laboratory Medicine Best Practices Systematic Review

Christopher Layfield 1, John Rose 1, Aaron Alford 1, Susan R Snyder 2, Fred S Apple 3,4, Farah M Chowdhury 5, Michael C Kontos 6, L Kristin Newby 7, Alan B Storrow 8, Milenko Tanasijevic 9,10, Elizabeth Leibach 5, Edward B Liebow 11, Robert H Christenson 12
PMCID: PMC5033652  NIHMSID: NIHMS804219  PMID: 25661303

Abstract

Objectives

This article presents evidence from a systematic review of the effectiveness of four practices (assay selection, decision point cardiac troponin (cTn) threshold selection, serial testing, and point of care testing) for improving the diagnostic accuracy for Non-ST-Segment Elevation Myocardial Infarction (NSTEMI) in the Emergency Department.

Design and Methods

The CDC-funded Laboratory Medicine Best Practices (LMBP™) Initiative systematic review A6 Method for Laboratory Best Practices was used.

Results

The current guidelines (e.g., ACC/AHA) recommend using cardiac troponin assays with a 99th percentile upper reference limit (URL) diagnostic threshold to diagnose NSTEMI. The evidence in this systematic review indicates that contemporary sensitive cTn assays meet the assay profile requirements (sensitivity, specificity, PPV, and NPV) to more accurately diagnose NSTEMI than alternate tests. Additional biomarkers did not increase diagnostic effectiveness of cTn assays. Sensitivity, specificity, and negative predictive value (NPV) were consistently high and low positive predictive value (PPV) improved with serial sampling. Evidence for use of cTn point of care testing (POCT) was insufficient to make recommendations, though some evidence suggests cTn POCT may result in reduction to patient length of stay and costs.

Conclusions

Two best practice recommendations emerged from the systematic review and meta-analysis of literature conducted using the LMBP™ A6 Method criteria: Testing with cardiac troponin assays, using the 99th percentile URL as the clinical diagnostic threshold for the diagnosis of NSTEMI and without additional biomarkers, is recommended. Also recommended is serial cardiac troponin sampling with one sample at presentation and at least one additional sample taken a minimum of 6 hours later to identify a rise or fall in the troponin level. Testing with high-sensitivity cardiac troponin assays, at presentation and again within 6 hours, is the recommended evidence-based best practice testing algorithm for optimized NSTEMI diagnosis.

Keywords: Acute Coronary Syndrome, Cardiac Troponin, Non-ST-Segment Elevation, Myocardial Infarction

1.0 Introduction

Type 1 myocardial infarction (MI) is one component of the Acute Coronary Syndromes (ACS) [1], a continuum of disease spanning from unstable angina to non-ST-segment elevation myocardial infarction (NSTEMI) and cell death, to ST-segment elevation myocardial infarction (STEMI). The primary cause of most type 1 MIs is the unstable coronary plaque. Type 2 MI is caused by a mismatch of oxygen supply and demand from a variety of causes including coronary spasm, coronary embolism, arrhythmia, anemia, and hypotension [1][2]. Here we consider type 1 and type 2 MI as the single entity of acute myocardial infarction (AMI). The American Heart Association (AHA) conservatively estimated 1,190,000 unique hospitalizations for new or recurrent MI (approximately 70% NSTEMI) that occurred in the United States in 2009 [3]. Classifying NSTEMI separately from STEMI is important because biomarkers play a central role in the diagnosis and management of NSTEMI, whereas STEMI classification is based on the electrocardiogram and biomarkers serve only a confirmatory role in diagnosis.

Biomarkers have evolved to be the cornerstone for the diagnosis of MI [1]; the preferred screening biomarker is cardiac troponin (cTn) [1][4][5]. However, cTn reporting is complicated by the multiple generations of cTn assays that have evolved with varying analytic characteristics and differences among cTn assays which have led, in part, to different diagnostic algorithms that can vary widely. Additionally cTn assays may use subtype I or T as the diagnostic component. Though sensitive assays exist for each subtype, we did not assess these differences, but did include the subtype information where necessary to inform assay selections. The relative diagnostic value of these subtype-specific algorithms in clinical decision making has not yet been assessed.

Our aim was to conduct an evidence-based review of four areas for cTn use for diagnosis of NSTEMI. These areas were selected by an expert panel following a preliminary literature review using the CDC Laboratory Medicine Best Practices A-6 Method [6]. These areas represent potential opportunities for quality improvement in NSTEMI diagnostic practices and include: (1) cTn assay selection, (2) cTn assay diagnostic threshold, (3) use of serial cTn samples to confirm NSTEMI diagnosis, and (4) POCT testing and timely receipt of cTn results.

1.1 Quality Gap 1: Selection of an appropriate biomarker to diagnose NSTEMI

Cardiac troponin assays have evolved with improvements in analytical performance since their introduction in the early 1990s. Early cTn assays were intended to replace for measurements of the MB isoenzyme of creatine kinase (CK-MB, muscle and brain subunits produced by heat muscle). However, by the year 2000, cTn was recognized in the first task force on redefinition of MI document as contributing additional information to CK-MB measurement [7], and is currently the preferred cardiac biomarker [1][4][5]. Based on this report, a consensus-based quality specification of a 10% total coefficient of variation (CV) at the decision point, defined as the 99th percentile of a reference control population became a target performance specification [7]. The International Federation for Clinical Chemistry and Laboratory Medicine (IFCC) has established a table of current, commercially available and research assays, available at www.ifcc.org [8].

First generation assays allowed for cTn detection within 4 to 12 hours of onset of myocardial injury with peak values between 12 and 48 h post-onset, as they were designed to be similar to CK-MB using ROC curve decision cutoffs. Current, more sensitive cTn assays have receiver operator characteristics curve (ROC) areas exceeding 90% sensitivity for AMI diagnosis within 2 h of symptom onset [9]. Cardiac troponin levels in early presenters still may not exceed the diagnostic threshold, though many assays are sufficiently sensitive to detect very small amounts accurately, i.e. ~10 ng/L or ~0.01 μg/L, of cTn within hours of MI symptom onset [9]. The imprecision and lack of diagnostic sensitivity of early, first generation cTn measurements led to combining this marker with other biomarkers such as myoglobin and CK-MB in an effort to improve medical decision making. The relative, marginal increase in diagnostic value in clinical decision support of adding these additional biomarkers to algorithms using currently available high sensitivity cTn tests has not been thoroughly assessed [10].

1.2 Quality Gap 2: Appropriate cTn assay threshold to diagnose NSTEMI

Although the initial Joint European Society of Cardiology (ESC)/ACC redefinition report indicated by consensus that a 10% total CV was necessary at the 99th percentile cutoff [7], a 2003 report [11] found that none of the then-current cTn assays could meet the 10% CV specification. Since that time cTn assays have improved, and the most recent global task force redefinition of MI report reiterated the recommended, consensus-based decision limit for myocardial injury as the concentration corresponding to the 99th percentile of the reference distribution in healthy people and that assays with ≤20% CV at the 99th percentile URL are clinically acceptable [1]. However, assay values between the 99th percentile and a cTn concentration with a 10% total CV can result in unnecessary, missed, or delayed treatment pending subsequent evaluation [12]. Kaplan-Meier curves and Cox proportional hazard analyses indicated significant risk for death or AMI recurrence as peak concentration of cTnI increased from low (<0.04 μg/L) to intermediate (0.04–0.10 μg/L) to high (>0.10 μg/L) in a study using the AccuTnI assay (Beckman Coulter Inc.) [13]. In contrast, changing the theoretical CV (up to 20% CV) at the 99th percentile resulted in minimal to no overall effect on diagnosis or outcomes in epidemiological studies using simulated evidence [14][15]. Limited evidence using population studies suggests no difference in diagnostic accuracy using the 99th percentile regardless of CV, although some STEMI patients were included [9]. Comparisons of the World Health Organization (WHO) Monitoring of Trends and Determinants in Cardiovascular Disease (MONICA) and ESC/ACC definitions of MI found significant differences in the prevalence of AMI based on the biomarker used (CK-MB versus cTn) and change in biomarker concentration, but found no difference between the 99th percentile and 10% CV cutoff for cTn [16]. A 2012 study in European hospitals indicated that diagnosis was based on the 99th percentile (38%), the 10% CV (41%), or another cut-off (62%) (lower limit of detection (LLOD), manufacturer’s recommendation, or literature citation) [17].

1.3 Quality Gap 3: Timing of serial samples to confirm ACS diagnosis

Any cTn result above a defined AMI cutoff has been shown to be a significant predictor of 30-day mortality [18]. Serial blood samples may improve timely and accurate diagnosis while reducing adverse outcomes among patients who initially have negative cTn results [18][19][20]. The 2012 task force consensus recommendations specify a criterion for diagnosis of AMI as a rise and/or fall of cTn with at least one value above the 99th percentile URL. However this guidance [10] does not specify the exact temporal sequence of sample collection or magnitude of the rise and/or fall of cTn indicating AMI. Current ACC/AHA consensus-based guidelines recommend initial samples collected at presentation with an additional sample between 6 and 8h post-presentation to identify a rise/fall in cTn level [4]. Other organizations [21][22][23][24][42] also recommend serial sampling but at different time points. Several studies of accelerated diagnostic protocols using contemporary sensitive cTn assays suggest that early serial sampling times (e.g., 0 and 1, 2, or 3 h) combined with use of either absolute or relative cTn concentration change for differential diagnosis may still allow safe clinical decision making [25][26][27][28][29]. Studies using high sensitivity cTn assays have shown that absolute concentrations are preferable in improving clinical specificity compared with percentage changes that were used with contemporary, current generation assays.

1.4 Quality Gap 4: Reduced turnaround time within 60 minutes

The recommended turnaround time (TAT) for cTn results is ≤1 h, with 30 minutes as the ideal [30]. Point of care testing (POCT) may substantially reduce TAT [31][32][33], to achieve the ACC/AHA recommendation of 30 minutes (87.3% of cases) [31]. One study was able to achieve 100% delivery of cTnI results to a nurse or clinician within 20 to 33 minutes after blood draw with a median time of 35 minutes from registration [37]. Reduced TAT has been associated with shorter length of stay (potentially between 8% [34] to 36% [35] reduction) in the emergency department (ED), concomitant cost savings [34], and more efficient coronary care use [35]. However it is unclear whether use of POCT improves clinical outcomes.

2.0 Methods

This evidence review followed the CDC’s LMBP™ A6 Method for conducting systematic reviews for evaluating laboratory quality improvement practices [6]. This approach is tailored to the evaluation of laboratory medicine practice effectiveness studies to support evidence-based best practice recommendations and has been reported in detail elsewhere [6]. A review team, including co-coordinators and staff specifically trained in the A6 Method application, conducted the systematic review. An expert panel, contributing medical and diagnostic perspectives provided additional guidance on the conduct of the systematic review and drafted recommendations based on the review results. The expert panel consisted of individuals selected for their diverse perspectives and subject matter expertise in the review topic, laboratory management, and evidence review methods. (See Appendix A for a listing of Expert Panel members.) In accordance with the A6 Method, the results of the evidence review and best practice recommendations were reviewed by the LMBP™ Workgroup, an independent, multi-disciplinary group comprised of 15 members with broad expertise in laboratory medicine, clinical practice, health services research or health policy (See Appendix B for LMBP™ Workgroup members).

The question addressed in this evidence review is: What cTn testing practices are effective at increasing timely and accurate AMI diagnosis for patients presenting in the ED with signs and symptoms suggestive of NSTEMI? An analytic framework for this quality issue of ACS diagnosis is displayed in Figure 1. The elements of the analytic framework were defined using a PICO (population, intervention, comparator, and outcomes) approach that defined the pertinent study parameters [36]. For this review specific study setting and population inclusion criteria were: studies published between 1996 and 2013, performed in the ED with adult patients suspected of NSTEMI, respectively. STEMI patients were specifically excluded.

Figure 1.

Figure 1

LMBP™ QI Analytical Framework: Cardiac Biomarkers

The review of practice effectiveness studies was based on a literature search strategy using terms developed with the assistance of a research librarian. An initial systematic search was completed in July 2011 which was subsequently updated in 2013 using four electronic databases (PubMed, Cochrane, Embase, and CINAHL), and additional sources including public and private-source professional guideline electronic databases (AHRQ, CLSI, ISO, NACB) for English language, human subjects, practice-relevant articles, and AMI guidelines from 1996 to 2013. We conducted hand searches of bibliographies from relevant secondary literature, consultation with and references from experts in the field including expert panelists, and solicitation of unpublished quality improvement studies resulting in direct data submissions to the LMBP™ website (http://wwwn.cdc.gov/futurelabmedicine/). A separate search for each of the interventions was performed and the results for all searches were then combined. Search terms and strings for each practice are available in Appendix C.

To reduce subjectivity and the potential for bias, all screenings, abstractions, and evaluations were conducted independently by at least two reviewers, and all differences were resolved through consensus. Each study was assigned one of three quality ratings (Good, Fair, Poor). Study quality ratings were based on the LMBP™ scoring methods described previously [6]. A study was included if it were considered to provide valid and useful information, met the PICO study inclusion criteria above, and evaluated a specific intervention/practice with at least one finding for a relevant outcome measure reported in a form which was useful for statistical analysis. Search result records that did not meet the inclusion criteria (i.e., not considered studies, not including a practice/outcome of interest), were excluded from further review. Studies not meeting the LMBP™ study quality criteria (Fair or Good quality rating) were also excluded. The Expert Panel then assigned one of three effect size ratings (Substantial, Moderate, or Minimal/None) for each study. Full tables and evaluation information from the systematic review are available in Appendix E.

3.0 Evidence review synthesis and results

Complete information related to the search results is provided in Figure 2. Of the total of 1,376 separate bibliographic records screened for assay selection, serial testing, assay threshold, and POCT, a full text review of 68 potentially eligible studies was conducted. A total of 24 published and 2 unpublished studies (Storrow 2015 [42] was unpublished at time of evaluation) were deemed acceptable for inclusion in this evidence review which totalled 13 published studies on assay selection, 8 published studies and 1 unpublished study on assay threshold, 5 published studies and 2 unpublished studies on serial testing, and 11 published studies on POCT. Six published studies and 1 unpublished study contained data evaluating two practices and two contained data evaluating three practices. While the data presented in Figure 2 is inclusive of all studies included for evaluation, several studies were removed for failing to meet a specific evaluation criterion or due to cTn assay issues. These instances are explained later in this manuscript. Analyses and characterizations of included and excluded studies are provided in Appendix D.

Figure 2.

Figure 2

Systematic Review Flow Diagram

“Body of Evidence” summary tables for each practice are provided for each practice below, including abstracted and standardized information, study quality ratings, and bibliographic reference information. Detailed evidence tables comprise Appendix E.

3.1 Assay selection practice effectiveness evidence

Table 1 summarizes information on 13 published studies that comprise the practice effectiveness body of evidence for comparing use of a cTn assay alone versus use of a multi-biomarker approach. Five studies were eliminated because they used early generation cTn assays [43][44][45][46][47] as evaluated by the Expert Panel. Four studies [48][49][50][65] were disqualified for failing to meet quality criteria. The remaining four studies were rated as “Fair” with “Minimal/None” effect [39] or “Good” with “Moderate” effect [26][38][40]. Two studies [26][40] included a marginal number (7% of the total population) of STEMI patients, which were included in the analysis as supporting material.

Table 1.

Assay selection practice effectiveness evidence

Practice: Assay Selection Study Quality Rating Effect Size Rating Overall Strength of Body of Evidence
Study Practice Measures Results Total Rating
Published 3 Studies = Good/Moderate
1 Study = Fair/Moderate
5 Studies = Fair/Minimal
4 Studies = Poor Excluded
Chang 2010a 2 2 1 2 7 Fair Minimal/None
Collinson 2011 2 2 2 3 9 Good Moderate
Eggers 2004 2 2 1 1 6 Fair Minimal/None
Engel 2007a 2 2 1 2 7 Fair Minimal/None
Hsu 2000 0 2 1 0 3 Poor
Jernberg 2000a 2 2 1 2 7 Fair Minimal/None
Jurlander 2000 0 0 2 0 2 Poor
Keller 2011 2 2 2 2 8 Good Moderate
Keller 2009 2 2 2 2 8 Good Moderate
McCord 2001a 1 2 1 1 5 Fair Minimal/None
Meune 2011 0 2 2 2 6 Poor
Quilici 2004 0 1 1 0 2 Poor
Straface 2008a 1 2 2 2 7 Fair Moderate
a

Early generation cTn assay; excluded from analysis

The RATPAC study [38], a randomized controlled trial, indicates that myoglobin and CK-MB did not add to the diagnostic value provided by a contemporary, sensitive cTnI assay. Eggers et al. [39] showed a similar result, that CK-MB and in particular, myoglobin, did not offer additional diagnostic value when added to a sensitive cTn assay. This evidence suggests minimal to no benefit from measuring any biomarkers in lieu of or addition to use of cTn to diagnose NSTEMI because the increase in diagnostic sensitivity comes at a similar or greater decrease in specificity. As displayed in Figure 3 (the cumulative assay specificity at 95% sensitivity by sampling time), the sensitivity and specificity from use of cTn far exceeds that of CK-MB or myoglobin at all-time points. Two other studies [26][40] showed a similar result but generalizability is limited by the inclusion of STEMI patients in their population. The findings are in agreement with the consensus recommendation that cTn is the preferred biomarker for MI diagnosis from the 2012 Global Task Force [1].

Figure 3.

Figure 3

Cumulative Assay Specificity at 95% Sensitivity by Sampling Time

3.2 Assay threshold practice effectiveness evidence

Information on the eight published studies and one unpublished study of practice effectiveness comparing use of the 99th percentile diagnostic threshold with other possible thresholds (% CV concentrations, LLOD, or ROC area) is summarized in Table 2. A subset of data from an earlier publication [51], from which STEMI patients have been removed, leaving only NSTEMI subjects was provided for analysis.

Table 2.

Assay threshold practice effectiveness evidence

Practice: Assay Threshold Study Quality Rating Effect Size Rating Overall Strength of Body of Evidence
Study Practice Measures Results Total Rating
Published
Apple 2006 1 2 2 1 6 Fair Moderate
Body 2011 2 2 2 3 9 Good Moderate
Collinson 2012 2 2 2 2 8 Good Moderate 6 Studies = Good/Moderate
Kontos 2004 1 2 2 3 8 Good Moderate
Mills 2011 2 2 2 3 8 Good Moderate 1 Study = Good/Minimal
Morrow 2001 3 2 1 3 9 Good Minimal/None
Mozina 2010 1 2 2 1 6 Fair Minimal/None 1 Study = Fair/Moderate
Straface 2008a 2 2 2 2 8 Good Moderate
Unpublished 1 Study = Fair/Minimal
Storrow 2014b 2 2 2 2 8 Good Moderate
a

Early generation cTn assay; excluded from analysis

b

Storrow study data were analyzed as unpublished prior to publication in 2015 [42].

Of the eight published studies, one study [43] was eliminated because it used an early generation cTn assay. Three studies [37][52][53] were eliminated, after a thorough review, because they did not directly address the diagnostic threshold question. One study [41] was rated as “Fair” quality with “Minimal/None” effect and three published studies [38] [51][54] and one unpublished study [42] were rated “Good” quality with a “Moderate” effect size, for a total of five studies included in the practice effectiveness body of evidence.

Mills et al. [51] provided a re-analysis of previously reported data that shows significantly (p<0.001) improved clinical outcomes from the use of a lower diagnostic threshold of cTn very near the 99th percentile. Here the cTnI assay implemented was reformulated to achieve higher sensitivity, which allowed a lowering of the cutoff at the 10% CV threshold from 0.2 ng/mL (old assay) to 0.05 ng/mL (new assay).

Data from the RATPAC study [38] were re-analyzed to examine various diagnostic cutoffs for the Stratus CS (Siemens Healthcare Diagnostics, Camberley, UK) including 0.07 ng/mL (99th percentile), 0.15 ng/mL (reported ROC cutoff [55]), and use of an arbitrary high cut-off, 0.30 ng/mL. (Re-analysis is not shown.) The 99th percentile cutoff for the cTnI diagnostic performance at presentation for the RATPAC cohort was superior to use of either the ROC decision point (65.2% sensitivity) or the higher cutoff (43.9% sensitivity). Use of the 99th percentile allowed for diagnosis of AMI in 82% (81.8% sensitivity) of RATPAC subjects at presentation, which was significantly more effective than either the ROC cutoff of 0.15 ng/mL (p<0.05) or the 0.30 ng/mL (p<0.0001) cutoff.

The effect of threshold selection was also examined with data from an unpublished multicenter study [42] conducted in urgent care settings. Patients (n=1, 929) presenting with signs and symptoms of acute cardiac ischemia were evaluated using the AccuTnI+3 assay (Beckman Coulter, Inc.) at threshold cut-offs corresponding to the 10% CV (0.04 ng/mL), 15% CV (0.03 ng/mL), 20% CV (0.02 ng/mL) and the 99th percentile URL threshold (0.02 ng/mL). The study findings demonstrated a progressively increasing sensitivity through the range of threshold values with slight loss in specificity, which was recovered through serial sampling. The PPV decreased from 55.1% to 47.0% at baseline for the 10% CV and 99th percentile concentrations, respectively, while the NPV increased from 97.9% to 98.6%.

Our findings are consistent with recommendations from the Global Task Force that the 99th percentile of a reference control population should be utilized for diagnosis of AMI [1].

3.3 Serial testing practice effectiveness evidence

Five published studies (Storrow [42] was unpublished at time of review) and two unpublished studies, as shown in Table 3, had outcomes of interest and were reviewed. Two of the published studies were rated “Fair” and three were rated “Good”. Four of these studies [26][38][40][42][56] had a “Moderate” effect size and one[39], had a “Minimal/None” effect size rating. The two unpublished studies, [42] and (Christenson 2014), were rated “Good” with a “Moderate” effect size.

Table 3.

Serial testing practice effectiveness evidence

Practice: Serial Sampling Study Quality Rating Effect Size Rating Overall Strength of Body of Evidence
Study Practice Measures Results Total Rating
Published
Eggers 2004 2 2 1 1 6 Fair Minimal/None
Collinson 2011 2 2 2 3 9 Good Moderate
Kelly 2011 2 1 2 1 6 Fair Moderate 5 Studies = Good/Moderate
Keller 2011 2 2 2 2 8 Good Moderate
Keller 2009 2 2 2 2 8 Good Moderate 1 Study = Fair/Moderate
Unpublished
Christenson 2014 2 2 2 2 8 Good Moderate 1 Study = Fair/Minimal/None
Storrow 2014 2 2 2 2 8 Good Moderate
b

Storrow study data were analyzed as unpublished prior to publication in 2015 [42].

Serial sampling increases diagnostic accuracy as evidenced by the studies analyzed. A subset of these assay improvements are illustrated with Figure 4, showing increases in sensitivity and specificity with serial sampling.

Figure 4.

Figure 4

Serial Testing Evaluation of cTn Assays

Comparing the diagnostic sensitivity at presentation with that of the peak cTnI level in the RATPAC data set demonstrated significant diagnostic improvement from serial sampling (p<0.001) [38].

Kelly et al. [56] used a sensitive cTn assay for measurement at enrollment and in later serial samples. Of 129 patients with a final diagnosis of AMI (13.6% of the population), 99 had cTnI values above the 99th percentile value at enrollment for a diagnostic sensitivity of 76.7% (95% CI 68.5–83.7%). For the 823 patients who ruled-out for AMI, 53 subjects had cTnI values exceeding the 99th percentile at enrollment; specificity 93.6% (95% CI 91.7–95.1%). A subgroup of 109 MI patients from this cohort had serial cTnI measurements. The cTnI sensitivity for these patients at presentation was similar to the overall cohort at 72.5% (95% CI 63.1–80.6). However, in later serial samples, all of these patients were diagnosed with MI at 100% sensitivity, while the specificity of this subgroup remained high at 92.4% (95% CI 89.9–94.4%). These data suggest that a single troponin measurement at ED presentation, cTnI in this trial, has insufficient diagnostic sensitivity for AMI diagnosis, and that serial biomarker measurement is required to achieve appropriate high sensitivity [56].

A pattern of increasing NPV and decreasing PPV was observed in a multi-center study of 1818 patients presenting with possible AMI [40]. Measurements of a sensitive cTnI were obtained at presentation, 3h and 6h time points. The NPV increased from 94%, to 95.3% to 96.0 over the hours of serial sampling, while the PPV decreased from 82.0%, to 79.3%, to 78.7% over this period; neither the changes in NPV nor PPV were significant. Although this cohort included 127 STEMI patients, which is a limitation of this study, the STEMI patients represented a small proportion (7%) of the total cohort. These patterns were supported by a study conducted by Eggers et al. [39] which included 197 patients, 27 (14%) of which were diagnosed with MI. Cumulative NPV increased from admission (97%) through 2 hour (99%) and 3 hour (100%) sampling. Collinson et al. [38]) showed increasing sensitivity (84.5% to 94.1%), specificity (97.6% to 98.4%), and NPV (98.7% to 99.9%) with a decreasing PPV (74.7% to 55.2%) with serial sampling taken at 90 minutes after presentation.

The contribution of serial sampling was also examined in a multicenter study [42] of 1,929 patients presenting acutely to urgent care settings with signs and symptoms consistent with acute cardiac ischemia. Serial samples were collected at enrollment, 1–3h, 3–6h, 6–9h; cTnI was quantified in these samples using a contemporary sensitive cTnI method. Of the 1,929 subjects, 253 (13%) were diagnosed as having acute MI. Of these, 1,504 (78%) were elevated above the 99th percentile at enrollment and demonstrated an increasing or decreasing temporal cTnI pattern at 3–6h. However 13% of the AMI patients did not show a cTnI elevation in the enrollment sample and the diagnosis may have been missed were it not for serial sampling. On the other hand, for the 1,676 cases for which AMI was ruled out, 20% did not demonstrate a rising or falling cTnI pattern at 3–6h. Of these non-AMI subjects, 11% had values above the cTnI cutoff at enrollment and may have received the incorrect diagnosis were it not for serial sampling. Importantly, because of the large number of non-AMI patients relative to AMI patients, the proportion of patients diagnosed with AMI when enrollment cTnI values were elevated was only 55% (PPV); i.e. 45% of patients with elevated cTnI enrollment values were non-AMI and would possibly have undergone unnecessary work-up or treatment if not for serial sampling.

An unpublished multicenter study (Christenson 2014) included 619 patients presenting to the ED with signs and symptoms of AMI. Serial samples were collected at presentation, 6h (± 3h) and 12h (± 3h). Cardiac troponin I was quantified with a contemporary sensitive assay. Of the 619 patients, 109 (17.7%) were diagnosed as having NSTEMI. Fifty-three (49%) of these AMI patients had cTnI values below the 99th percentile normal value of 40 ng/L on presentation. Some or all of these 53 patients may not have received therapy known to benefit NSTEMI patients had serial sampling not been performed. Seventy-one (65.1%) of the AMI patients showed a rise and/or fall of ⩾20 ng/L. Of the remaining 510 patients (82.3%) who ruled-out for AMI, 28 (5.5%) had baseline values above the 99th percentile of normal in the enrollment sample. Only two of the rule-out patients (0.4%) had a change of ⩾20 ng/L; the remaining 508 non-AMI patients had a change <20 ng/L. The PPV of cTnI for diagnosing AMI in the presentation sample was 65.4%. Thus, if the cTnI value at enrollment presentation were used to rule-in or rule-out AMI, a large proportion of patients may have been diagnosed as having AMI and could have undergone unnecessary work-up or treatment if not for serial sampling.

Serial sampling allows more accurate diagnostic use of cTn. These findings are supported by the consensus based guidance from the Global Task force that recommends serial sampling for diagnosis of MI [1].

3.4 Point of care testing practice effectiveness evidence

Data on the eleven published studies that comprised the practice effectiveness body of evidence comparing use of a cTn POCT with the use of a central laboratory test are summarized in Table 4. Of the eleven studies, three [32][57][58] were excluded for failing (i.e. receiving a 0 score) on at least one of the study quality rating criteria. Of the included studies, three were rated “Good” quality and five were rated “Fair” quality. The effect size of two studies that measured TAT [41][59] were rated “Substantial”, and one study that measured length of stay (LOS) [59] was rated “Minimal/None”. The results are suggestive of decreased TAT with POCT practice implementation.

Table 4.

Point of care testing practice effectiveness evidence

Practice: Point of Care Testing Study Quality Rating Effect Size Rating Overall Strength of Body of Evidence
Study Practice Measures Results Total Rating
Published
Apple 2006 1 2 1 2 6 Fair Moderate
Bock 2008 1 2 1 2 6 Fair Substantial (Neg)
Caragher 2002 2 2 2 1 7 Fair Substantial 1 Study = Good/Moderate
Collinson 2012 2 2 2 3 9 Good Moderate
Di Serio 2009 1 2 2 0 5 Fair 2 Studies = Good/Minimal
Hallani 2005 0 2 1 2 5 Fair
Lee-Lewandrowski 2011 2 2 2 1 7 Fair Minimal/None 3 Studies = Fair/Substantial
Loten 2010 2 2 1 3 8 Good Minimal/None
Mozina 2010 1 2 2 1 6 Fair Substantial 1 Study = Fair/Moderate
Singer 2005 0 2 2 1 5 Fair 1 Study = Fair/Minimal
Straface 2008 2 2 2 2 8 Good Minimal/None 3 Studies = Excluded

For the diagnostic accuracy outcome effect size, one study [61] documented an adverse effect rated “Substantial”, one study [59]) had a favorable effect rated “Substantial”, one study [38] had a favorable effect rated “Moderate”, and three [43][60][62] had effect sizes rated “Minimal/None”. One study [37], rated “Fair” with “Moderate” effect size, found improvement in TAT, but there was not sufficient evidence to evaluate further. The evidence on the effect of the POCT practice on diagnostic accuracy was inconsistent, and no recommendation can be made for or against this practice at this time.

4.0 Additional considerations

4.1 Feasibility of implementation

The data from this evidence review are consistent with the consensus based recommendations for use of cTn alone for the diagnosis of AMI, use of the 99th percentile as the AMI decision point for MI and the use of serial samples for evaluating AMI [1][4]. Use of cardiac markers to diagnose NSTEMI is standard practice [5][10]. Recommendations for the use of cardiac marker assays to diagnose NSTEMI have been part of the NACB [5], ACC/AHA guidelines [4] and ACCF/AHA guidelines [10] for many years; evidence evaluated as part of this review supports those guideline recommendations. Here we examined three questions finding that diagnosis of NSTEMI should be performed using a sensitive cTn assay with a diagnostic threshold at the 99th percentile of normal with serial samples to detect a rise/fall in the cTn level with time. No evidence of diagnostic improvement when using additional biomarkers was discovered, and therefore, the use of any additional biomarkers when a sensitive or contemporary sensitive cTn assay was used is not recommended. The evidence indicates that cTn measurement with a sensitive assay provides important diagnostic information for assessing patient risk of adverse clinical outcomes. Only minimal decrease in diagnostic specificity is incurred. The studies included in this evidence review focused on assessment of NSTEMI and were performed in EDs and cardiac short stay units, all settings in which the diagnosis is determined and where treatment is normally initiated.

4.2 Future research needs

Cost of care is also likely reduced but this was not evaluated in this review. While POCT cTn has the potential to reduce TAT [59][41], and LOS [60], the body of evidence regarding this practice was insufficient and inconsistent for conclusions. However, there is sufficient published evidence to determine the equivalency of POCT assays to laboratory assays and we recommend additional studies be conducted to measure the effect of POCT cTn implementation. Future studies should also establish an evidence base for the equivalency of cTI and cTn assays.

4.3 Study limitations

There were several limitations to the approach taken for evaluating the evidence within this review. The studies used in this review did not report all outcome measures consistently and utilized an array of assays which made quantitative data pooling in meta-analytic techniques impractical. Though the specific assays and outcomes were different, the assays used were appropriate at the time of study and comparisons were made which demonstrated the superiority of sensitive assays that are capable of good performance at the 99th percentile of a reference control population. We did not have sufficient evidence to differentiate between sensitive and high-sensitivity cTn assays, which could potentially improve risk stratification. The outcome measures varied according to the specific interest of study, but in all cases the use of sensitive assays resulted in improved outcomes.

The LMBP™ A6 Method is consistent with practice standards for systematic reviews, but all such methods are imperfect and include subjective assessments at multiple points that may produce bias. Rating study quality depends on consensus assessments that may be affected by such things as rater experience and the criteria used. As is the case with most systematic reviews, publication bias must be considered. The restriction to English language studies, while satisfying the requirement for multiple reviewers for each study, may also introduce bias.

5.0 Conclusions and best practices recommendations

The overall evidence included in this review indicates that a stand-alone, contemporary sensitive cTn assay performs significantly better than alternative biomarkers [26][38][39][40] and that the addition of those biomarkers to the cTn assay does not improve diagnosis [38][39]. In addition, recent economic evaluation studies found the new high sensitivity cTn assay without additional biomarkers is likely to be cost effective relative to the earlier conventional cTn assay or a combination of high sensitivity cTn assays with additional biomarker(s) [63][64]

Serial cTn sampling provides a means to reduce risk of misdiagnosis due to limitations in assays, variations in time of presentation after onset of symptoms and other external factors that can effect optimal assay performance. The evidence included in this review indicates that using a strategy for repeat testing results increases the diagnostic sensitivity [26][56] and NPV with a modest decrease in the PPV [39][56]. Our evaluation of the published [38][42] and unpublished (Christenson 2014) evidence also shows that this approach would more effectively diagnose NSTEMI patients for transition to appropriate care and would increase the efficiency of rule-out testing.

The overall strength of the body of evidence included in this review regarding POCT cTn was inconclusive. On one outcome (diagnostic accuracy effect size), one study [61] with “Substantial” effect rating, documented a negative effect direction while another study [59]), also “Substantial effect rating, documented a favorable effect. An additional three studies [38][43][62] documented effect sizes rated “Minimal/None”. Only one study of the eight included [37] found improvement in TAT. Evidence was insufficient and inconsistent for recommendation either for or against the use of point of care cTn testing.

Testing with high-sensitivity cardiac troponin assays, using the 99th percentile URL as the clinical diagnostic threshold for the diagnosis of NSTEMI and without additional biomarkers, is recommended. Also recommended is serial cardiac troponin sampling with one sample at presentation and at least one additional sample collected a maximum of 6 hours later to identify a rise or fall in the troponin level. Testing with high-sensitivity cardiac troponin assays, without other biomarkers, at presentation and again within 6 hours is the evidence-based best practice testing algorithm for most accurate and timely NSTEMI diagnosis.

Acknowledgments

Melissa Gustafson, Jim Derzon, Diana Mass, Paul Epner, Judy Berkowitz, Barbara Zehnbauer, Colleen Shaw, the LMBP™ Workgroup

Funding Source: CDC funding for the Laboratory Medicine Best Practices Initiative to Battelle Centers for Public Health Research and Evaluation under contract No. SP0700-00-D-3180, Delivery Order 0723.

Abbreviations

ACC

American College of Cardiology

ACS

Acute coronary syndrome

AHA

American Heart Association

AMI

Acute myocardial infarction

AUC

Area under the receiver-operating-characteristic curve

CDC

U.S. Centers for Disease Control and Prevention

CI

Confidence Interval

CK-MB

Creatine kinase – myoglobin

cTn

Cardiac troponin

cTnI

Cardiac troponin subtype I

cTnT

Cardiac troponin subtype T

CV

Coefficient of variation

ECG

Electrocardiogram

ED

Emergency department

ESC

European Society of Cardiology

h

Hour

ICD

International Classification of Diseases

LMBP

Laboratory Medicine Best Practices Initiative

MI

Myocardial Infarction

NSTEMI

Non-ST-segment elevation myocardial infarction

PICO

Population, Intervention, Comparison, Outcome

POCT

Point of care test

QI

Quality improvement

ROC

Receiver operator curve

STEMI

Myocardial infarction with ST elevation

ST Segment

In an ECG, the isoelectric line after the QRS complex (Q, R, and S waves representing ventricular polarization) which represents phase 2 cardiac action potential

URL

Upper reference limit

Appendix A: LMBP™ Cardiac Markers Systematic Review Expert Panel Members

Fred S. Apple, PhD (Hennepin County Medical Center and University of Minnesota Medical School)

Farah M. Chowdhury, MD, MPH (Centers for Disease Control and Prevention, Division for Heart Disease and Stroke Prevention)

Robert Christenson, PhD, DABCC, FACB* (University of Maryland School of Medicine)

Michael C. Kontos, MD (Virginia Commonwealth University Medical Center)

Qaiser Mukhtar, PhD (Centers for Disease Control and Prevention, Division for Heart Disease and Stroke Prevention)

L. Kristin Newby, MD, MHS (Duke University Medical Center)

Alan B. Storrow, MD (Vanderbilt University Medical Center)

Milenko Tanasijevic, MD, MBA* (Brigham and Women’s Hospital & Harvard Medical School)

*Also Laboratory Medicine Best Practices Workgroup member

Appendix B: LMBP™ Cardiac Markers Systematic Review Workgroup Members

Robert H. Christenson, PhD, DABCC, FACB, Professor of Medical and Research Technology Director, Rapid Response Laboratories, University of Maryland School of Medicine, Baltimore, MD

John Fontanesi, PhD, Director, Center for Management Science in Health, University of California, San Diego, School of Medicine, La Jolla, California

Julie Gayken, MT (ASCP), Director of Laboratory Services (Retired), Anatomic & Clinical Pathology, Regions Hospital, St. Paul, MN

Sharon Geaghan, MD, Associate Professor, Department of Pathology, Pediatrics Division, Stanford University School of Medicine, Stanford, CA

Christine M. Litwin, M.D. Professor, Pathology and Laboratory Medicine, Medical Director Clinical Immunology and Referral Testing, Medical University of South Carolina, Charleston, SC

Thomas Lorey, MD, Medical Director, TPMG Regional Reference Laboratory, Kaiser Permanente, Northern California Region, Berkeley CA

Bernadette Mazurek Melnyk, PhD, RN, CPNP, Dean of the College of Nursing, Associate Vice President for Health Promotion, Chief Wellness Officer, The Ohio State University, Columbus, OH

James Nichols, PhD, Professor of Pathology, Microbiology and Immunology, Medical Director of Clinical Chemistry, Pathology Labs/Blood Bank, Vanderbilt University School of Medicine, Nashville, TN

Mary Nix, MS, MT(ASCP)SBB, Project Officer, National Guideline Clearinghouse, Agency for Healthcare Research and Quality, Rockville, MD

Anton Piskac (Tony), MD, Vice President for Administration and Performance Improvement, Methodist Health System, Corporate Offices, Omaha, NE

Jennifer Rhamy, MBA, MA, MT(ASCP)SBB, HP, Director, Regional Blood Donor Center, St. Mary’s Hospital, Grand Junction, CO

Christopher Lee Roy, MD, Assistant Professor of Medicine at Harvard Medical School, Director Hospitalist Service, Brigham and Women’s Hospital, Boston, MA

Melissa Singer, MT(ASCP), Centers for Medicare and Medicaid Services, Center for Medicaid and State Operations, Survey and Certification Group, Division of Laboratory Services, Baltimore, MD

Milenko Tanasijevic, MD, MBA, Director, Clinical Laboratories Division and Clinical Program Development, Pathology Department, Brigham and Women’s Hospital, Boston, MA

Appendix C: LMBP™ Cardiac Markers Systematic Review Structured Search Databases and Terms

Serial Testing Search String

(“angina”[title/abstract] OR “acute coronary syndrome”[mesh] OR “myocardial ischemia”[mesh] OR “Angina Pectoris”[mesh] OR “angina, unstable”[mesh] OR “myocardial infarction”[mesh] OR “chest pain”[mesh] OR “heart attack”[all text] OR “heart attack”[title/abstract] OR “myocardial injury”[all text] OR “myocardial injury”[title/abstract]) AND (“troponin”[MeSH Terms] OR “troponin”[All Fields] OR “cTn”[All Fields]) AND (“serial” [ALL TEXT] OR “serial testing” [TITLE/ABSTRACT] OR “repeat” [TEXT WORD] OR “repeated”[TEXT WORD] OR “repeated testing”[TITLE/ABSTRACT] OR “time”[TEXT WORD] OR “temporal”[TEXT WORD]) AND (“diagnosis”[Subheading] OR “diagnosis”[All Fields] OR “diagnosis”[MeSH Terms] OR “risk factors”[mesh] OR “prognosis”[mesh])

Serial Testing Search String

“Acute Coronary Syndrome”[Mesh] AND (“angina”[title/abstract] OR “myocardial ischemia”[mesh] OR “Angina Pectoris”[mesh] OR “angina, unstable”[mesh] OR “myocardial infarction”[mesh] OR “chest pain”[mesh] OR “heart attack”[all text] OR “heart attack”[title/abstract] OR “myocardial injury”[all text] OR “myocardial injury”[title/abstract]) AND (“troponin”[MeSH Terms] OR “troponin”[All Fields] OR “cTn”[All Fields]) AND (“serial” [ALL TEXT] OR “serial testing” [TITLE/ABSTRACT] OR “repeat” [TEXT WORD] OR “repeated”[TEXT WORD] OR “repeated testing”[TITLE/ABSTRACT] OR “time”[TEXT WORD] OR “temporal”[TEXT WORD]) AND (“diagnosis”[Subheading] OR “diagnosis”[All Fields] OR “diagnosis”[MeSH Terms] OR “prognosis”[mesh] OR “prognosis”[title/abstract] OR “risk factors”[mesh] OR “risk factors”[mesh] OR “risk factors”[title/abstract])

Assay Selection Search String

(“angina”[title/abstract] OR “acute coronary syndrome”[mesh] OR “myocardial ischemia”[mesh] OR “Angina Pectoris”[mesh] OR “angina, unstable”[mesh] OR “myocardial infarction”[mesh] OR “chest pain”[mesh] OR “heart attack”[all] OR “heart attack”[title/abstract] OR “myocardial injury”[all] OR “myocardial injury”[title/abstract]) AND (“troponin”[Major] OR “troponin”[All Fields] OR “cTn”[All Fields]) AND (“myoglobin”[Major] OR “myoglobin”[title/abstract]) OR (“Creatine Kinase, MB Form”[Major] OR “Creatine Kinase”[title/abstract]) OR “CK-MB”[All Fields] OR “Biological Markers”[Major] AND (“diagnosis”[Subheading] OR “diagnosis”[All Fields] OR “diagnosis”[MeSH Terms] OR “prognosis”[mesh] OR “prognosis”[title/abstract] OR “risk factors”[mesh] OR “risk factors”[title/abstract]) AND “heart”[major]

Assay Threshold Search String

(“angina”[title/abstract] OR “acute coronary syndrome”[mesh] OR “myocardial ischemia”[mesh] OR “Angina Pectoris”[mesh] OR “angina, unstable”[mesh] OR “myocardial infarction”[mesh] OR “chest pain”[mesh] OR “heart attack”[all text] OR “heart attack”[title/abstract] OR “myocardial injury”[all text] OR “myocardial injury”[title/abstract]) AND (“troponin”[Mesh] OR “troponin”[All Fields] OR “trans-crotonin”[Supplementary Concept] OR cTn[Text Word]OR “analysis”[Subheading] OR “analysis”[title/abstract] OR “assay”[title/abstract] OR “biological assay”[major] OR “biological assay”[All Fields]) AND (“cut-off”[title/abstract] OR “roc curve”[Major] OR “roc curve”[title/abstract] OR “decision point”[title/abstract] OR “reference values”[Major] OR “reference value”[title/abstract] OR “upper limit normal”[title/abstract]) AND (“prognosis”[mesh] OR “prognosis”[title/abstract] OR “risk factors”[mesh] OR “risk factors”[title/abstract])

Point of Care Testing Search String

(“angina”[title/abstract] OR “acute coronary syndrome”[mesh] OR “myocardial ischemia”[mesh] OR “Angina Pectoris”[mesh] OR “angina, unstable”[mesh] OR “myocardial infarction”[mesh] OR “chest pain”[mesh] OR “heart attack”[all] OR “heart attack”[title/abstract] OR “myocardial injury”[all] OR “myocardial injury”[title/abstract]) AND (“patient care”[Mesh] OR “patients”[MeSH Terms] OR “patients”[All Fields] OR “patient”[All Fields]) AND (“Biological Markers”[mesh] OR “troponin”[MESH] OR “troponin”[all fields] OR “cTn”[all fields]) AND (“Point-of-Care Systems”[MESH] OR “point of care”[all] OR “point-of-care”[all] OR “bedside”[All Fields] OR “at the bedside”[ALL FIELDS] OR “bedside testing”[all] OR “bed side testing”[all] OR “testing”[all]) AND (“laboratory techniques and procedures”[Mesh] OR “laboratory techniques and procedures”[All Fields] OR “laboratory testing”[All Fields] OR “testing”[all]) AND (“diagnosis”[mesh] OR “diagnosis”[all fields] OR “diagnosis”[subheading] OR “prognosis”[mesh] OR “prognosis”[title/abstract] OR “risk factors”[mesh] OR “risk factors”[title/abstract])

Appendix D: LMBP™ Cardiac Markers Systematic Review Study Characterizations and Analyses

Full Text Review Studies1,2 Included Studies2,3
Assay Selection 13 published studies Five studies were disqualified that reported data from early generation cTn assays [43][44][45][46][47] as evaluated by the Expert Panel. Four studies [48][49][50][65] were disqualified for failing to meet quality criteria.
 Published studies included:
  13 total − 5 − 4 = 4
Assay Threshold 8 published studies One study [43] was disqualified that reported data from an early generation cTn assay. Three studies [37][52][53] were disqualified for failing to directly address the diagnostic threshold question.
 Published studies included:
  8 total − 1 − 3 = 4
1 unpublished study One study [42] was rated “Good” quality with a “Moderate” effect size. (Study [42] also reported a serial sampling analysis.)
 Unpublished studies included:
  1 total − 0 = 1
Serial Sampling 5 published studies Two studies were rated “Fair” [39][56] and three were rated “Good” [26][38][40][42]. Four studies [26][38][40][42][56] had a “Moderate” effect size and one [39], had a “Minimal/None” effect size rating.
 Published studies included:
  5 total − 0 = 5
2 unpublished studies Two studies [42][Christenson, 2014] were rated “Good” with a “Moderate” effect size. (One study [42] also reported an assay threshold analysis.)
 Unpublished studies included:
  2 total − 0 = 2
Point of Care Testing 11 published studies Three studies [32][57][58] were excluded for failing to meet quality criteria. Other studies are reported as follows:
1 Study = Good/Moderate [38]
2 Studies = Good/Minimal [43][60]
3 Studies = 2, Fair/Substantial Pos [ 41][59];
  1, Fair/Substantial Neg [61]
1 Study = Fair/Moderate [37]
1 Study = Fair/Minimal [62]
 Graphically represented as follows:
Inline graphic
Eight studies reported inconsistent effects and direction. No recommendation was made.
  Published studies included:
  11 total − 3 = 8
Total Studies Included = 151
1 In this particular instance, authors use “studies” to denote “reports” or “articles.”
2 While Figure 2 is inclusive of all studies included for evaluation, several studies were removed for failing to meet a specific evaluation criterion or due to cTn assay issues.

Published Studies Included = 4 + 4 + 5 = 13
Unpublished Studies Included = 3 − 1 (2 analyses from same reference study) = 2

Total Studies Included = 13 + 2 = 151
3 Later, authors use “studies” to denote practice “analyses”: For example, “A total of 24 published and 2 unpublished studies (Storrow 2014 [42] was unpublished at time of evaluation) were deemed acceptable for inclusion.” See section “3.0 Evidence review synthesis and results”
(1) Six published studies contained data evaluating two practices.
  Math: 151 + 6 extra analyses = 21 analyses
(2) One unpublished study contained data evaluating two practices.
  Math: 21 + 1 extra analysis = 22 analyses
(3) Two published studies contained data evaluating three practices.
  Math: 22 + 4 extra analyses =
    26 Total Analyses

Appendix E: LMBP™ Cardiac Markers Systematic Review Evidence Tables

Laboratory Medicine Best Practices Cardiac Marker Body of Evidence Tables

Practice: Assay Selection Study Quality Rating Effect Size Rating Overall Consistency Overall Strength of Body of Evidence
Study Practice Measures Results Total Rating
Published 2 Studies = Good/Moderate
1 Study = Fair/Moderate
6 Studies = Fair/Minimal
4 Studies = Poor – Excluded
Chang 2010* 2 2 1 2 7 Fair Minimal/None
Collinson 2011 2 2 2 3 9 Good Moderate
Eggers 2004 2 2 1 1 6 Fair Minimal/None
Engel 2007* 2 2 1 2 7 Fair Minimal/None
Hsu 2000 0 2 1 0 3 Poor
Jernberg 2000* 2 2 1 2 7 Fair Minimal/None
Jurlander 2000 0 0 2 0 2 Poor
Keller 2011 2 2 2 2 8 Good Moderate
Keller 2009 2 2 2 2 8 Good Moderate
McCord 2001* 1 2 1 1 5 Fair Minimal/None
Meune 2011 0 2 2 2 6 Poor
Quilici 2004 0 1 1 0 2 Poor
Straface 2008* 1 2 2 2 7 Fair Moderate
*

Early generation cTn assay; excluded from analysis

Footnotes

Disclaimer: The findings and conclusions in this article are those of the authors and do not necessarily represent the official position of the Centers for Disease Control and Prevention/the Agency for Toxic Substances and Disease Registry (CDC/ATSDR).

References

  • 1.Thygesen K, Alpert JS, Jaffe AS, Simons ML, Chaitman BR, White HD, the writing group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction Third universal definition of myocardial infarction. J Am Coll Cardiol. 2012;60:1581–98. doi: 10.1016/j.jacc.2012.08.001. [DOI] [PubMed] [Google Scholar]
  • 2.Sandoval Y, Smith SW, Apple FS. Supply/demand type 2 myocardial infarction: should we be paying more attention? J Amer Coll Card. 2014;63:2079–87. doi: 10.1016/j.jacc.2014.02.541. [DOI] [PubMed] [Google Scholar]
  • 3.Roger VL, Go AS, Lloyd-Jones DM, Benjamin EJ, Berry JD, et al. Heart disease and stroke statistics – 2012 update: a report from the American Heart Association. Circulation. 2012;125(1):e2–220. doi: 10.1161/CIR.0b013e31823ac046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Amsterdam EA, Wenger NK, Brindis RG, Casey DE, Jr, Ganiats TG, et al. 2014 AHA/ACC guideline for the management of patients with non-ST-elevation acute coronary syndromes: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines. J Am Coll Cardiol. 2014 doi: 10.1016/.jacc.2014.09.017. [DOI] [PubMed] [Google Scholar]
  • 5.Morrow DA, Cannon CP, Jesse RL, Newby LK, Ravkilde J, et al. National Academy of Clinical Biochemistry Laboratory Medicine Practice Guidelines: Clinical Characteristics and Utilization of Biochemical Markers in Acute Coronary Syndromes. Circulation. 2007;115:e356–e375. doi: 10.1161/CIRCULATIONAHA.107.182882. [DOI] [PubMed] [Google Scholar]
  • 6.Christenson RH, Snyder SR, Shaw CS, Derzon JH, Black RS, et al. Laboratory medicine best practices: systematic evidence review and evaluation methods for quality improvement. Clin Chem. 2011;57(6):816–825. doi: 10.1373/clinchem.2010.157131. [DOI] [PubMed] [Google Scholar]
  • 7.Alpert JS, Thygesen K, Antman E, Bassand JP. Myocardial infarction redefined – a consensus document of The Joint European Society of Cardiology/American College of Cardiology Committee for the redefinition of myocardial infarction. J Am Coll Cardiol. 2000;36(3):959–69. doi: 10.1016/s0735-1097(00)00804-4. [DOI] [PubMed] [Google Scholar]
  • 8.Apple FS, Collinson PO, IFCC Task Force on Clinical Applications of Cardiac Biomarkers Analytical characteristics of high-sensitivity cardiac troponin assays. Clin Chem. 2012;57(1):54–61. doi: 10.1373/clinchem.2011.165795. [DOI] [PubMed] [Google Scholar]
  • 9.Reichlin T, Hochholzer W, Bassetti S, Steuer S, Stelzig C, et al. Early Diagnosis of Myocardial Infarction with Sensitive Cardiac Troponin Assays. N Engl J Med. 2009;361:858–67. doi: 10.1056/NEJMoa0900428. [DOI] [PubMed] [Google Scholar]
  • 10.Anderson JL, Adams CD, Antman EM, Bridges CR, Calif RM, et al. 2012 ACCF/AHA focused update incorporated into the ACCF/AHA 2007 guidelines for the management of patients with unstable angina/non-ST –elevation myocardial infarction. A report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2013;61(23):e179–347. doi: 10.1016/j.jacc.2013.01.014. [DOI] [PubMed] [Google Scholar]
  • 11.Panteghini M, Pagani F, Yeo KT, Apple FS, Christenson RH, et al. Evaluation of imprecision for cardiac troponin assays at low-range concentrations. Clin Chem. 2004;5D(2):327–32. doi: 10.1373/clinchem.2003.026815. [DOI] [PubMed] [Google Scholar]
  • 12.Bonham M, Miller S. Clinical comparison of 99th percentile and 10% coefficient of variation cutoff values for four commercially available troponin I assays. LABMEDICINE. 2009;40(8):470–3. doi: 10.1309/LMFBC68TIZGVHOHV. [DOI] [Google Scholar]
  • 13.Kavsak PA, MacRae AR, Palomaki GF, Newman AM, Ko DT, et al. Health outcomes categorized by current and previous definitions of acute myocardial infarction in an unselected cohort of troponin-naive emergency department patients. Clin Chem. 2006;52(11):2028–35. doi: 10.1373/clinchem.2006.073403. [DOI] [PubMed] [Google Scholar]
  • 14.Apple FS, Parvin CA, Buechler KF, Christenson RH, Wi AHB, et al. Validation of the 99th Percentile Cutoff Independent of Assay Imprecision (CV) for Cardiac Troponin Monitoring for Ruling Out Myocardial Infarction. Clin Chem. 2005;51(11):2198–2200. doi: 10.1373/clinchem.2005.052886. [DOI] [PubMed] [Google Scholar]
  • 15.Kupchak P, Wu AWB, Ghani F, Newby KL, Ohman EM, et al. Influence of Imprecision on ROC Curve Analysis for Cardiac Markers. Clin Chem. 2006;52(4):752–3. doi: 10.1373/clinchem.2005.064477. [DOI] [PubMed] [Google Scholar]
  • 16.Kavsak PA, MacRae AR, Lustig V, Bhargava, Vandersluis, et al. The impact of the ESC/ACC redefinition of myocardial infarction and new sensitive troponin assays on the frequency of acute myocardial infarction. Am Hear J. 2006;152(1):118–25. doi: 10.1016/j.ahj.2005.09.022. [DOI] [PubMed] [Google Scholar]
  • 17.Collinson PO, van Dieijen-Visser MP, Pulkki K, Hammerer-Lercher A, Suvisaari J, et al. Evidence-Based Laboratory Medicine: How Well Do Laboratories Follow Recommendations and Guidelines? The Cardiac Marker Guideline Uptake in Europe (CARMAGUE) Study. Clin Chem. 2012;58(1):305–6. doi: 10.1373/clinchem.2011.171439. [DOI] [PubMed] [Google Scholar]
  • 18.Newby LK, Christenson RH, Ohman EM, Armstrong PW, Thompson TD, et al. Value of Serial Troponin T Measures for Early and Late Risk Stratification in Patients with Acute Coronary Syndromes. Circulation. 1998;98:1853–1859. doi: 10.1161/01.cir.98.18.1853. [DOI] [PubMed] [Google Scholar]
  • 19.Loughrey BV, Spedding RL, Herity NA, Murphy JC, Walker E, Graham WR, McDonnell M. Prospective Survey of Serial Troponin T Requesting in an Acute Teaching Hospital. Ulster Med J. 2007;76(3):150–153. [PMC free article] [PubMed] [Google Scholar]
  • 20.MacRae AR, Kavask PA, Lustig V, Bharqava R, Vandersluis R, et al. Assessing the Requirement for the 6-Hour Interval between Specimens in the American Heart Association Classification of Myocardial Infarction in Epidemiology and Clinical Research Studies. Clin Chem. 2006;52(5):812–818. doi: 10.1373/clinchem.2005.059550. [DOI] [PubMed] [Google Scholar]
  • 21.Archar SA, Kundu S, Norcross WA. Diagnosis of Acute Coronary Syndrome. Am Fam Physician. 2005;72(1):119–126. http://www.aafp.org/afp/2005/0701/p119.html Accessed 6/27/2015. [PubMed] [Google Scholar]
  • 22.Institute for Clinical Systems Improvement (ICSI) Diagnosis and Treatment of Chest Pain and Acute Coronary Syndrome (ACS) 2011 http://www.ICSI.org. Accessed 6/27/2015.
  • 23.Kumar A, Cannon CP. Acute Coronary Syndromes: Diagnosis and Management, Part I. Mayo Clin Proc. 2009;84(10):917–38. doi: 10.4065/84.10.917. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Kavsak PA, Allen LC, Apple FS, Booth RA, Chan PC, et al. Cardiac troponin testing in the acute care setting: ordering, reporting, and high sensitivity assays – an update from the Canadian Society of Clinical Chemists (CSCC) Clin Biochem. 2011;44(16):1273–7. doi: 10.1016/j.clinbiochem.2011.08.001. [DOI] [PubMed] [Google Scholar]
  • 25.Haaf P, Drexler B, Reichlin T, Twerenbld R, Reiter M, et al. High-Sensitivity Cardiac Troponin in the Distinction of Acute Myocardial Infarction From Acute Cardiac Noncoronary Artery Disease. Circulation. 2012;126(1):31–40. doi: 10.1161/CIRCULATIONAHA.112.100867. 126. [DOI] [PubMed] [Google Scholar]
  • 26.Keller T, Zeller T, Ojeda F, et al. Serial changes in highly sensitive troponin I assay and early diagnosis of myocardial infarction. JAMA. 2011;306(24):2684–93. doi: 10.1001/jama.2011.1896. [DOI] [PubMed] [Google Scholar]
  • 27.Than M, Cullen L, Reid CM, Lim SH, Aldous S, et al. A 2-h diagnostic protocol to assess patients with chest pain symptoms in the Asia-Pacific region (ASPECT): a prospective observational validation study. Lancet. 2011;377(9771):1077–84. doi: 10.1016/S0140-6736(11)60310-3. [DOI] [PubMed] [Google Scholar]
  • 28.Than M, Cullen L, Aldous S, Parsonage WA, Reid CM, et al. 2-Hour accelerated diagnostic protocol to assess patients with chest pain symptoms using contemporary troponins as the only biomarker: the ADAPT trial. JACC. 2012;59:2091–8. doi: 10.1016/j.jacc.2012.02.035. [DOI] [PubMed] [Google Scholar]
  • 29.Reichlin T, Irfan A, Twerenbold R, Reiter M, Hockholzer W, et al. Utility of absolute and relative changes in cardiac troponin concentrations in the early diagnosis of acute myocardial infarction. Circulation. 2011;124:136–45. doi: 10.1161/CIRCULATIONHA.111.023937. [DOI] [PubMed] [Google Scholar]
  • 30.Nichols JH, Christenson RH, Clarke W, Gronowski A, Hammett-Stabler CA, et al. Executive Summary. The National Academy of Clinical Biochemistry Laboratory Medicine Practice Guideline: Evidence-based Practice for Point-of-care Testing. Clinical Chimica Acta. 2007;379:14–28. doi: 10.1016/j.cca.2006.12.025. [DOI] [PubMed] [Google Scholar]
  • 31.Ryan RJ, Lindsell CJ, Hollander JE, O’Neill B, Jackson R, et al. A Multicenter Randomized Controlled Trial Comparing Central Laboratory and Point-of-Care Cardiac Marker Testing Strategies: The Disposition Impacted by Serial Point of Care Markers in Acute Coronary Syndromes (DISPO-ACS) Trial. Ann Emer Med. 2009;53(3):321–328. doi: 10.1016/j.annemergmed.2008.06.464. [DOI] [PubMed] [Google Scholar]
  • 32.Singer AJ, Ardise J, Gulla J, Cangro J. Point-of-Care Testing Reduces Length of Stay in Emergency Department Chest Pain Patients. Ann Emer Med. 2005;45(6):587–591. doi: 10.1016/j.annemergmed.2004.11.020. [DOI] [PubMed] [Google Scholar]
  • 33.Storrow AB, Lyon JA, Porter MW, Zhou C, et al. A systematic review of emergency department point-of-care cardiac markers and efficiency measures. Point of Care. 2009;8:121–5. [Google Scholar]
  • 34.Apple FS, Chung AY, Kogut ME, Bubany S, Murakami MM. Decreased Patient Charges Following Implementation of Point-of-Care Cardiac Troponin Monitoring in Acute Coronary Syndrome Patients in a Community Hospital Cardiology Unit. Clinica Chimica Acta. 2006;370(1–2):191–195. doi: 10.1016/j.cca.2006.02.011. [DOI] [PubMed] [Google Scholar]
  • 35.Goodacre SW, Bradburn M, Cross E, Collinson P, Gray A, Hall AS. The Randomised Assessment of Treatment Using Panel Assay of Cardiac Markers (RATPAC) Trial: A Randomised Controlled Trial of Point-of-Care Cardiac Markers in the Emergency Department. Heart. 2010;97(3):190–6. doi: 10.1136/hrt.2010.203166. [DOI] [PubMed] [Google Scholar]
  • 36.Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gøtzsche PC, et al. The PRISMA Statement for Reporting Systematic Reviews and Meta-Analyses of Studies That Evaluate Health Care Interventions: Explanation and Elaboration. PLoS Med. 2009;6(7):e1000100. doi: 10.1371/journal.pmed.1000100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Apple FS. Point-of-Care Cardiac Troponin Testing: Process Improvements for Detection of Acute Myocardial Infarction. Point of Care. 2006;5(1):25–7. [Google Scholar]
  • 38.Collinson P, Goodacre S, Gaze D, Gray A. Very early diagnosis of chest pain by point-of-care testing: comparison of the diagnostic efficiency of a panel of cardiac biomarkers compared with troponin measurement along in the RATPAC trial. Heart. 2012;98(4):312–8. doi: 10.1136/heartjnl-2011-300723. [DOI] [PubMed] [Google Scholar]
  • 39.Eggers KM, Oldgren J, Nordenskjold A, Lindahl B. Diagnostic value of serial measurement of cardiac markers in patients with chest pain: limited value of adding myoglobin to troponin I for exclusion of myocardial infarction. Am Hear J. 2004;148(4):574–81. doi: 10.1016/j.ahj.2004.04.030. [DOI] [PubMed] [Google Scholar]
  • 40.Keller T, Zeller T, Peetz D, et al. Sensitive troponin I assay in early diagnosis of acute myocardial infarction. NEJM. 2009;361:868–77. doi: 10.1056/NEJMoa0903515. [DOI] [PubMed] [Google Scholar]
  • 41.Mozina H, Vukan V, Lenart K, Skitek M, Osredkar J. Quantitative point-of-care troponin I in emergency department in comparison with troponin I in central laboratory. Point of Care. 2010;9(1):8–11. [Google Scholar]
  • 42.Storrow A, Nowak R, Diercks DB, et al. Absolute and relative changes (delta) in troponin I for early diagnosis of myocardial infarction: Results of a prospective multicenter trial. Clin Biochem. 2015;48:260–7. doi: 10.1016/j.clinbiochem.2014.09.012. [DOI] [PubMed] [Google Scholar]
  • 43.Straface AL, Myers JH, Kirchick HJ, Blick KE. A rapid point-of-care cardiac marker testing strategy facilitates the rapid diagnosis and management of chest pain patients in the emergency department. Am J Clin Pathol. 2008;129(5):788–95. doi: 10.1309/9GGNMURLJWJD88W3. [DOI] [PubMed] [Google Scholar]
  • 44.Chang SS, Lee SH, Wu JY, Ning HC, Chiu TF, et al. Evaluation of the value of rapid D-dimer test in conjunction with cardiac troponin I test for early risk stratification of myocardial infarction. J Thromb Thrombolysis. 2010;40(4):427–8. doi: 10.1007/s11239-010-0469-1. [DOI] [PubMed] [Google Scholar]
  • 45.Engel G, Rockson SG. Rapid diagnosis of myocardial injury with troponin T and CK-MB relative index. Mol Diagn Ther. 2007;11(2):109–16. doi: 10.1007/BF03256230. [DOI] [PubMed] [Google Scholar]
  • 46.Jernberg T, Lindahl B, James S, Ronquist G, Wallentin L. Comparison between strategies using creatine kinase-MB (mass), myoglobin, and troponin T in the early detection or exclusion of acute myocardial infarction in patients with chest pain and a nondiagnostic electrocardiogram. Am J Cardiol. 2000;86(12):1367–71. doi: 10.1016/s0002-9149(00)01245-5. [DOI] [PubMed] [Google Scholar]
  • 47.McCord J, Nowak RM, McCullough PA, Foreback C, Borzack S, et al. Ninety-Minute Exclusion of Acute Myocardial Infarction By Use of Quantitative Point-of-Care Testing of Myoglobin and Troponin. Circulation. 2001;104:1483–8. doi: 10.1161/hc3801.096336. [DOI] [PubMed] [Google Scholar]
  • 48.Hsu LF, Koh TH, Lim YL. Cardiac marker point-of-care testing: evaluation of rapid on-site biochemical marker analysis for diagnosis of acute myocardial infarction. Ann Acad Med Singapore. 2000;29:421–7. [PubMed] [Google Scholar]
  • 49.Jurlander B, Clemmensen P, Wagner GS, Grande P. Very early diagnosis and risk stratification of patients admitted with suspected acute myocardial infarction by the combined evaluation of a single serum value of cardiac troponin-T, myoglobin, and creatine kinase MBmass. European Heart Journal. 2001;21:382–9. doi: 10.1053/euhj.1999.1760. [DOI] [PubMed] [Google Scholar]
  • 50.Meune C, Zuily S, Wahbi K, Claessens Y, Weber S, Cheneview-Gobeaux C. Combination of copeptin and high-sensitivity cardiac troponin T assay in unstable angina and non-ST-segment elevation myocardial infarction: A pilot study. Arch Cardiovasc Dis. 2011;104:4–10. doi: 10.1016/j.acvd.2010.11.002. [DOI] [PubMed] [Google Scholar]
  • 51.Mills N, Churchhouse A, Lee K, et al. Implementation of a sensitive troponin I assay and risk of recurrent myocardial infarction and death in patients with suspected acute coronary syndrome. JAMA. 2011;305(12):1210–6. doi: 10.1001/jama.2011.338. [DOI] [PubMed] [Google Scholar]
  • 52.Kontos MC, Shah R, Fritz LM, Anderson FL, Tatum JL, et al. Implication of different cardiac troponin I levels for clinical outcomes and prognosis of acute chest pain patients. J Am Coll Cardiol. 2004;42(6):958–65. doi: 10.1016/j.jacc.2003.10.036. [DOI] [PubMed] [Google Scholar]
  • 53.Morrow DA, Cannon CP, Rifai N, Frey MK, Vicari R, et al. Ability of minor elevations of troponin I and T to predict benefit from an early invasive strategy in patients with unstable angina and non-ST elevation myocardial infarction: results from a randomized trial. JAMA. 2001;286(19):2405–12. doi: 10.1001/jama.286.19.2405. [DOI] [PubMed] [Google Scholar]
  • 54.Body R, Carley S, McDowell, et al. Rapid exclusion of acute myocardial infarction in patients with undetectable troponin using a high-sensitivity assay. JACC. 2011;58(13):1332–9. doi: 10.1016/j.jacc.2011.06.026. [DOI] [PubMed] [Google Scholar]
  • 55.Heeschen C, Goldmann BU, Langenbrink L, Matschuck G, Hamm CW. Evaluation of a rapid whole blood ELISA for quantification of troponin I in patients with acute chest pain. Clin Chem. 1999;45:1789–96. [PubMed] [Google Scholar]
  • 56.Kelly AM. Performance of a sensitive troponin assay in the early diagnosis of acute myocardial infarction in the emergency department. Emerg Med Australias. 2011;23(2):181–5. doi: 10.1111/j.1742-6723.2011.01388.x. [DOI] [PubMed] [Google Scholar]
  • 57.Di Serio F, Caputo M, Zaninotto M, Ottomano C, Plebani M. Evaluation of Analytical Performance of the Pathfast Cardiac Troponin I. Clinical Chemistry and Laboratory Medicine. 2009;47(7):829–833. doi: 10.1515/CCLM.2009.182. [DOI] [PubMed] [Google Scholar]
  • 58.Hallani H, Leung DY, Newland E, Jeurgens CP. Use of quantitative point-of-care test for the detection of serum cardiac troponin T in patients with suspected acute coronary syndromes. Int Med J. 2005;35:560–2. doi: 10.1111/j.1445-5994.2005.00897.x. [DOI] [PubMed] [Google Scholar]
  • 59.Caragher T, Fernandez BB, Jacobs F, Barr LA. Evaluation of quantitative cardiac biomarker point-of-case testing in the emergency department. J Emerg Med. 2002;22(1):1–7. doi: 10.1016/s0736-4679(01)00429-2. [DOI] [PubMed] [Google Scholar]
  • 60.Loten C, Attia J, Hullick C, Marley J, McElduff P. Point of care troponin decreases time in the emergency department for patients with possible acute coronary syndrome: a randomized controlled trial. Emerg Med J. 2010;27:194–8. doi: 10.1136/emj.2008.069427. [DOI] [PubMed] [Google Scholar]
  • 61.Bock JL, Singer AJ, Thode HC. Comparison of Emergency Department Patient Classification by Point-of-Care and Central Laboratory Methods for Cardiac Troponin I. Clin Chem. 2008;130(1):132–5. doi: 10.1309/NVXH8DL5HWFDNB74. [DOI] [PubMed] [Google Scholar]
  • 62.Lee-Lewandrowski E, Januzzi JL, Grisson R, Mohammed AA, Lewandrowski G, Lewandrowski K. Evaluation of first-draw whole blood, point-of-care cardiac markers in the context of the universal definition of myocardial infarction: a comparison of a multimarker panel to troponin alone and to testing in the central laboratory. Arch Pathol Lab Med. 2011;135(4):459–63. doi: 10.5858/2010-0112-OA.1. [DOI] [PubMed] [Google Scholar]
  • 63.Vaidya A, Severens JL, Bongaerts BW, Cleutiens KB, Nelemans PJ, et al. High-sensitive troponin T assay for the diagnosis of acute myocardial infarction: an economic evaluation. BMC Cardiovasc-Disord. 2014:14–77. doi: 10.1186/1471-2261-14-77. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Goodacre S, Thokala P, Carroll C, Stevens JW, Leaviss J, et al. Systematic review, meta-analysis and economic modelling of diagnostic strategies for suspected acute coronary syndrome. Health Technol Assess. 2013;17(1):v–vi. 1–188. doi: 10.3310/hta17010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Quilici J, Banzet N, Paule P, Meynard J-B, Mutin M, et al. Circulating Endothelial Cell Count as a Diagnostic Marker for Non–ST-Elevation Acute Coronary Syndromes. Circulation. 2004;110:1586–1591. doi: 10.1161/01.CIR.0000142295.85740.98. [DOI] [PubMed] [Google Scholar]

RESOURCES