Table 2.

Interexaminer Reproducibility and Intereye Variability of Optic Nerve Parameters Measured with Handheld Optical Coherence Tomography

Optic Nerve Parameter Interexaminer Intraclass Correlation Coefficient Intereye Intraclass Correlation Coefficient
Disc diameter 0.915 (0.947) 0.869 (0.892)
Cup diameter 0.892 (0.925) 0.793 (0.806)
Cup-to-disc ratio C/DOCT 0.830 (0.896) 0.741 (0.755)
BMO-MRW
 Nasal 0.980 (0.982) 0.872 (0.834)
 Temporal 0.960 (0.979) 0.717 (0.712)
Retinal thickness
 Nasal 0.923 (0.957) 0.655 (0.636)
 Temporal 0.907 (0.921) 0.793 (0.794)
RNFL
 Nasal 0.773 (0.657) 0.736 (0.505)
 Temporal 0.681 (0.430) 0.713 (0.607)

BMO-MRW = Bruch's membrane opening–minimum rim width; RNFL = retinal nerve fiber layer.

Values in brackets are for participants younger than 4 years (n = 20). Of the 30 patients, 8 were younger than 1 year, 6 were 1–2 years of age, 6 were 2–4 years of age, and 10 were older than 4 years of age.