Abstract
In an in vitro study, 10/0 nylon was found to require a significantly lower laser energy density to produce suture lysis following a single shot than either 10/0 Dacron or 10/0 prolene. Nylon and Dacron monofilament sutures ruptured at reproducible energy levels without significant observable changes at subthreshold irradiation. Prolene, however, was observed to stretch at energy levels below the lysis threshold, under the standard tension produced by a weight of 0.36 (SD 0.02) g. This feature may be of value when performing laser treatment to trabeculectomy flap sutures in the early postoperative period.
Full text
PDF

Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Hoskins H. D., Jr, Migliazzo C. Management of failing filtering blebs with the Argon laser. Ophthalmic Surg. 1984 Sep;15(9):731–733. [PubMed] [Google Scholar]
- Johnstone M. A., Wellington D. P., Ziel C. J. A releasable scleral-flap tamponade suture for guarded filtration surgery. Arch Ophthalmol. 1993 Mar;111(3):398–403. doi: 10.1001/archopht.1993.01090030118052. [DOI] [PubMed] [Google Scholar]
- Melamed S., Ashkenazi I., Glovinski J., Blumenthal M. Tight scleral flap trabeculectomy with postoperative laser suture lysis. Am J Ophthalmol. 1990 Mar 15;109(3):303–309. doi: 10.1016/s0002-9394(14)74555-6. [DOI] [PubMed] [Google Scholar]
- Savage J. A., Condon G. P., Lytle R. A., Simmons R. J. Laser suture lysis after trabeculectomy. Ophthalmology. 1988 Dec;95(12):1631–1638. doi: 10.1016/s0161-6420(88)32964-7. [DOI] [PubMed] [Google Scholar]