Clinical comparison of the ProTon and Tono-Pen tonometers with the Goldmann applanation tonometer.

A Midelfart 1, A Wigers 1
PMCID: PMC504987  PMID: 7819170

Abstract

A clinical evaluation of a new electron ProTon tonometer was performed comparing the values of intraocular pressure (IOP) measured using this instrument with those determined by a similar instrument, Tono-Pen XL, and by Goldmann applanation tonometry. The mean IOP measured in 106 eyes with the ProTon tonometer was not significantly different from that determined with Goldmann applanation, while the IOP values measured with Tono-Pen XL were significantly lower. The 95% limits of agreement between applanation tonometry and ProTon tonometry were between -4 mm Hg and 5 mm Hg and between applanation tonometry and Tono-Pen XL tonometry between -3 mm Hg and 8 mm Hg. The ProTon tonometer appears to have a higher level of accuracy than the Tono-Pen XL tonometer in clinical practice.

Full text

PDF
895

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Armstrong T. A. Evaluation of the Tono-Pen and the Pulsair tonometers. Am J Ophthalmol. 1990 Jun 15;109(6):716–720. doi: 10.1016/s0002-9394(14)72442-0. [DOI] [PubMed] [Google Scholar]
  2. Bland J. M., Altman D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986 Feb 8;1(8476):307–310. [PubMed] [Google Scholar]
  3. Boothe W. A., Lee D. A., Panek W. C., Pettit T. H. The Tono-Pen. A manometric and clinical study. Arch Ophthalmol. 1988 Sep;106(9):1214–1217. doi: 10.1001/archopht.1988.01060140374035. [DOI] [PubMed] [Google Scholar]
  4. Christoffersen T., Fors T., Ringberg U., Holtedahl K. Tonometry in the general practice setting (I): Tono-Pen compared to Goldman applanation tonometry. Acta Ophthalmol (Copenh) 1993 Feb;71(1):103–108. doi: 10.1111/j.1755-3768.1993.tb04970.x. [DOI] [PubMed] [Google Scholar]
  5. Denis P., Nordmann J. P., Bertin V., Gayraud J. M., Laroche L., Saraux H. Evaluation of the Tono-Pen 2 and the X-Pert noncontact tonometers in cataract surgery. Ophthalmologica. 1993;207(3):155–161. doi: 10.1159/000310423. [DOI] [PubMed] [Google Scholar]
  6. Farrar S. M., Miller K. N., Shields M. B., Stoup C. M. An evaluation of the Tono-Pen for the measurement of diurnal intraocular pressure. Am J Ophthalmol. 1989 Apr 15;107(4):411–416. doi: 10.1016/0002-9394(89)90666-1. [DOI] [PubMed] [Google Scholar]
  7. Geyer O., Mayron Y., Loewenstein A., Neudorfer M., Rothkoff L., Lazar M. Tono-Pen tonometry in normal and in post-keratoplasty eyes. Br J Ophthalmol. 1992 Sep;76(9):538–540. doi: 10.1136/bjo.76.9.538. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Hessemer V., Rössler R., Jacobi K. W. Comparison of intraocular pressure measurements with the Oculab Tono-Pen vs manometry in humans shortly after death. Am J Ophthalmol. 1988 Jun 15;105(6):678–682. doi: 10.1016/0002-9394(88)90064-5. [DOI] [PubMed] [Google Scholar]
  9. Higginbotham E. J. Clinical evaluation of the Oculab Tono-Pen. Am J Ophthalmol. 1988 Jan 15;105(1):101–101. doi: 10.1016/0002-9394(88)90139-0. [DOI] [PubMed] [Google Scholar]
  10. Hines M. W., Jost B. F., Fogelman K. L. Oculab Tono-Pen, Goldmann applanation tonometry, and pneumatic tonometry for intraocular pressure assessment in gas-filled eyes. Am J Ophthalmol. 1988 Aug 15;106(2):174–179. doi: 10.1016/0002-9394(88)90830-6. [DOI] [PubMed] [Google Scholar]
  11. Kao S. F., Lichter P. R., Bergstrom T. J., Rowe S., Musch D. C. Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer. Ophthalmology. 1987 Dec;94(12):1541–1544. doi: 10.1016/s0161-6420(87)33249-x. [DOI] [PubMed] [Google Scholar]
  12. Lim J. I., Blair N. P., Higginbotham E. J., Farber M. D., Shaw W. E., Garretson B. R. Assessment of intraocular pressure in vitrectomized gas-containing eyes. A clinical and manometric comparison of the Tono-Pen to the pneumotonometer. Arch Ophthalmol. 1990 May;108(5):684–688. doi: 10.1001/archopht.1990.01070070070037. [DOI] [PubMed] [Google Scholar]
  13. Minckler D. S., Baerveldt G., Heuer D. K., Quillen-Thomas B., Walonker A. F., Weiner J. Clinical evaluation of the Oculab Tono-Pen. Am J Ophthalmol. 1987 Aug 15;104(2):168–173. doi: 10.1016/0002-9394(87)90010-9. [DOI] [PubMed] [Google Scholar]
  14. Moore C. G., Milne S. T., Morrison J. C. Noninvasive measurement of rat intraocular pressure with the Tono-Pen. Invest Ophthalmol Vis Sci. 1993 Feb;34(2):363–369. [PubMed] [Google Scholar]
  15. Pen Q. F., Ran R. T., Yao X. Y. [Experimental study of the seromuscular pedicle flap graft in repair of large bile duct defects]. Zhonghua Wai Ke Za Zhi. 1988 Feb;26(2):101-2, 126. [PubMed] [Google Scholar]
  16. Whitacre M. M., Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol. 1993 Jul-Aug;38(1):1–30. doi: 10.1016/0039-6257(93)90053-a. [DOI] [PubMed] [Google Scholar]
  17. Wilson M. R., Baker R. S., Mohammadi P., Wheeler N. C., Lee D. A., Scott C. Reproducibility of postural changes in intraocular pressure with the Tono-Pen and Pulsair tonometers. Am J Ophthalmol. 1993 Oct 15;116(4):479–483. doi: 10.1016/s0002-9394(14)71408-4. [DOI] [PubMed] [Google Scholar]

Articles from The British Journal of Ophthalmology are provided here courtesy of BMJ Publishing Group

RESOURCES