AIMS--A study was conducted to determine the capability of the recently developed National Eye Institute (NEI) Scheimpflug cataract imaging system in detecting changes in the nuclear region of the lens over a 1 year period. METHODS--Twenty five eyes with pure nuclear cataracts with mean nuclear densities < or = 0.30 optical density units (ODU) as well as 30 normal control eyes were each examined at baseline and 12 months later. Computerised densitometry using the NEI Scheimpflug cataract imaging system was performed. Clinical grading of the lenses was also done using the Lens Opacities Classification System II (LOCS II). For densitometry, a change of plus or minus 0.023 ODU (the 99% range) in mean density in the nuclear area was considered a progression or regression at 1 year. RESULTS--Using the Scheimpflug densitometry, 14 of the 25 cataractous eyes showed significant progression at 1 year. In the normal control group, only three of the 30 eyes showed significant progression. In contrast, using the LOCS II clinical grading, only two of the 25 cataractous eyes showed a one step increase, two of the 30 controls progressed at 1 year, and none regressed. There was no significant difference in visual acuity. CONCLUSION--This study suggests the value of the NEI Scheimpflug cataract imaging system in detecting nuclear change within 1 year. However, clinically significant changes may require longer follow up periods. These data will be useful in planning future longitudinal studies of nuclear cataracts, such as for clinical trials of anticataract drugs.

Full text

PDF
527

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Brown N. A., Bron A. J., Ayliffe W., Sparrow J., Hill A. R. The objective assessment of cataract. Eye (Lond) 1987;1(Pt 2):234–246. doi: 10.1038/eye.1987.43. [DOI] [PubMed] [Google Scholar]
  2. Brown N. Slit-image photography. Trans Ophthalmol Soc U K. 1970;89:397–408. [PubMed] [Google Scholar]
  3. Chylack L. T., Jr, Leske M. C., McCarthy D., Khu P., Kashiwagi T., Sperduto R. Lens opacities classification system II (LOCS II) Arch Ophthalmol. 1989 Jul;107(7):991–997. doi: 10.1001/archopht.1989.01070020053028. [DOI] [PubMed] [Google Scholar]
  4. Chylack L. T., Jr, Wolfe J. K., Singer D. M., Leske M. C., Bullimore M. A., Bailey I. L., Friend J., McCarthy D., Wu S. Y. The Lens Opacities Classification System III. The Longitudinal Study of Cataract Study Group. Arch Ophthalmol. 1993 Jun;111(6):831–836. doi: 10.1001/archopht.1993.01090060119035. [DOI] [PubMed] [Google Scholar]
  5. Datiles M. B., Edwards P. A., Trus B. L., Green S. B. In vivo studies on cataracts using the Scheimpflug slit lamp camera. Invest Ophthalmol Vis Sci. 1987 Oct;28(10):1707–1710. [PubMed] [Google Scholar]
  6. Drews-Bankiewicz M. A., Caruso R. C., Datiles M. B., Kaiser-Kupfer M. I. Contrast sensitivity in patients with nuclear cataracts. Arch Ophthalmol. 1992 Jul;110(7):953–959. doi: 10.1001/archopht.1992.01080190059029. [DOI] [PubMed] [Google Scholar]
  7. Edwards P. A., Datiles M. B., Green S. B. Reproducibility study on the Scheimpflug Cataract Video Camera. Curr Eye Res. 1988 Oct;7(10):955–960. doi: 10.3109/02713688809015140. [DOI] [PubMed] [Google Scholar]
  8. Edwards P. A., Datiles M. B., Unser M., Trus B. L., Freidlin V., Kashima K. Computerized cataract detection and classification. Curr Eye Res. 1990 Jun;9(6):517–524. doi: 10.3109/02713689008999591. [DOI] [PubMed] [Google Scholar]
  9. Ferris F. L., 3rd, Kassoff A., Bresnick G. H., Bailey I. New visual acuity charts for clinical research. Am J Ophthalmol. 1982 Jul;94(1):91–96. [PubMed] [Google Scholar]
  10. Harris M. L., Smith G. T., Brown N. A. Inter and intra observer reproducibility of the new Oxford CCD Scheimpflug camera. Eye (Lond) 1991;5(Pt 4):487–490. doi: 10.1038/eye.1991.79. [DOI] [PubMed] [Google Scholar]
  11. Hockwin O., Dragomirescu V., Laser H. Measurements of lens transparency or its disturbances by densitometric image analysis of Scheimpflug photographs. Graefes Arch Clin Exp Ophthalmol. 1982;219(6):255–262. doi: 10.1007/BF00231409. [DOI] [PubMed] [Google Scholar]
  12. Karbassi M., Khu P. M., Singer D. M., Chylack L. T., Jr Evaluation of lens opacities classification system III applied at the slitlamp. Optom Vis Sci. 1993 Nov;70(11):923–928. doi: 10.1097/00006324-199311000-00009. [DOI] [PubMed] [Google Scholar]
  13. Kashima K., Trus B. L., Unser M., Edwards P. A., Datiles M. B. Aging studies on normal lens using the Scheimpflug slit-lamp camera. Invest Ophthalmol Vis Sci. 1993 Jan;34(1):263–269. [PubMed] [Google Scholar]
  14. Klein B. E., Klein R., Linton K. L., Magli Y. L., Neider M. W. Assessment of cataracts from photographs in the Beaver Dam Eye Study. Ophthalmology. 1990 Nov;97(11):1428–1433. doi: 10.1016/s0161-6420(90)32391-6. [DOI] [PubMed] [Google Scholar]
  15. Knott F. A., Blaikley J. B. Penicillin Sprays and Upper Respiratory Tract Infection. Br Med J. 1946 Mar 9;1(4444):349–350. [PMC free article] [PubMed] [Google Scholar]
  16. Lasa M. S., Datiles M. B., 3rd, Podgor M. J., Magno B. V. Contrast and glare sensitivity. Association with the type and severity of the cataract. Ophthalmology. 1992 Jul;99(7):1045–1049. [PubMed] [Google Scholar]
  17. Lasa M. S., Podgor M. J., Datiles M. B., 3rd, Caruso R. C., Magno B. V. Glare sensitivity in early cataracts. Br J Ophthalmol. 1993 Aug;77(8):489–491. doi: 10.1136/bjo.77.8.489. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Magno B. V., Datiles M. B., 3rd, Lasa S. M. Senile cataract progression studies using the Lens Opacities Classification System II. Invest Ophthalmol Vis Sci. 1993 May;34(6):2138–2141. [PubMed] [Google Scholar]
  19. Magno B. V., Freidlin V., Datiles M. B., 3rd Reproducibility of the NEI Scheimpflug Cataract Imaging System. Invest Ophthalmol Vis Sci. 1994 Jun;35(7):3078–3084. [PubMed] [Google Scholar]
  20. Niesel P. Spaltlampenphotographie mit der Haag-Streit-Spaltlampe 900. Ophthalmologica. 1966;151(4):489–504. doi: 10.1159/000304907. [DOI] [PubMed] [Google Scholar]
  21. Sparrow J. M., Bron A. J., Brown N. A., Ayliffe W., Hill A. R. The Oxford Clinical Cataract Classification and Grading System. Int Ophthalmol. 1986 Dec;9(4):207–225. doi: 10.1007/BF00137534. [DOI] [PubMed] [Google Scholar]
  22. Sparrow J. M., Brown N. A., Shun-Shin G. A., Bron A. J. The Oxford modular cataract image analysis system. Eye (Lond) 1990;4(Pt 4):638–648. doi: 10.1038/eye.1990.89. [DOI] [PubMed] [Google Scholar]
  23. Taylor H. R., West S. K. The clinical grading of lens opacities. Aust N Z J Ophthalmol. 1989 Feb;17(1):81–86. [PubMed] [Google Scholar]
  24. Vivino M. A., Chintalagiri S., Trus B., Datiles M. Development of a Scheimpflug slit lamp camera system for quantitative densitometric analysis. Eye (Lond) 1993;7(Pt 6):791–798. doi: 10.1038/eye.1993.186. [DOI] [PubMed] [Google Scholar]
  25. West S. K., Rosenthal F., Newland H. S., Taylor H. R. Use of photographic techniques to grade nuclear cataracts. Invest Ophthalmol Vis Sci. 1988 Jan;29(1):73–77. [PubMed] [Google Scholar]

Articles from The British Journal of Ophthalmology are provided here courtesy of BMJ Publishing Group

RESOURCES