Skip to main content
Journal of Intensive Care logoLink to Journal of Intensive Care
. 2016 Oct 6;4:63. doi: 10.1186/s40560-016-0188-6

Fluid responsiveness prediction using Vigileo FloTrac measured cardiac output changes during passive leg raise test

Anton Krige 1,, Martin Bland 1, Thomas Fanshawe 2
PMCID: PMC5052799  PMID: 27721980

Abstract

Background

Passive leg raising (PLR) is a so called self-volume challenge used to test for fluid responsiveness. Changes in cardiac output (CO) or stroke volume (SV) measured during PLR are used to predict the need for subsequent fluid loading. This requires a device that can measure CO changes rapidly. The Vigileo™ monitor, using third-generation software, allows continuous CO monitoring. The aim of this study was to compare changes in CO (measured with the Vigileo device) during a PLR manoeuvre to calculate the accuracy for predicting fluid responsiveness.

Methods

This is a prospective study in a 20-bedded mixed general critical care unit in a large non-university regional referral hospital. Fluid responders were defined as having an increase in CO of greater than 15 % following a fluid challenge. Patients meeting the criteria for circulatory shock with a Vigileo™ monitor (Vigileo™; FloTrac; Edwards™; Lifesciences, Irvine, CA, USA) already in situ, and assessed as requiring volume expansion by the clinical team based on clinical criteria, were included. All patients underwent a PLR manoeuvre followed by a fluid challenge.

Results

Data was collected and analysed on stroke volume variation (SVV) at baseline and CO and SVV changes during the PLR manoeuvre and following a subsequent fluid challenge in 33 patients. The majority had septic shock. Patient characteristics, baseline haemodynamic variables and baseline vasoactive infusion requirements were similar between fluid responders (10 patients) and non-responders (23 patients). Peak increase in CO occurred within 120 s during the PLR in all cases. Using an optimal cut point of 9 % increase in CO during the PLR produced an area under the receiver operating characteristic curve of 0.85 (95 % CI 0.63 to 1.00) with a sensitivity of 80 % (95 % CI 44 to 96 %) and a specificity of 91 % (95 % CI 70 to 98 %).

Conclusions

CO changes measured by the Vigileo™ monitor using third-generation software during a PLR test predict fluid responsiveness in mixed medical and surgical patients with vasopressor-dependent circulatory shock.

Electronic supplementary material

The online version of this article (doi:10.1186/s40560-016-0188-6) contains supplementary material, which is available to authorized users.

Keywords: Passive leg raising, Edwards Vigileo FloTrac monitoring, Fluid responsiveness, Cardiac output monitoring, Vasoplegic shock, Septic shock

Background

Circulatory insufficiency is common in critically ill patients and may lead to organ dysfunction. Increasing cardiac preload may increase stroke volume (SV) and consequently cardiac output (CO) and thus tissue perfusion [1]. However, it has been shown that only half of critically ill patients thought to be preload responsive, based on static predictors of preload, actually show an increase in SV, and therefore CO, following volume expansion (VE) [2, 3]. The administration of fluid in the group that is not preload responsive not only delays the appropriate management of their circulatory insufficiency but is also an independent predictor of delayed respiratory weaning and survival [411].

Functional haemodynamic monitoring has been shown to accurately predict fluid responsiveness but unfortunately several caveats, which include constant tidal volumes of adequate size and sinus rhythm, must be present [1217]. Unfortunately, this precludes a large proportion of critically ill patients.

Passive leg raising (PLR), whereby the patients legs are transiently raised by 45°, and the torso flattened, results in a reversible flow of 150–300 ml of blood from the venous capacitance vessels of the lower body, to the thoracic compartment, a so called self-volume challenge [1820]. Boulain et al. [21] first demonstrated the potential of this phenomenon to predict fluid responsiveness by measuring blood pressure changes during a PLR.

Subsequent studies have shown even greater accuracy, without the limitations of functional haemodynamic monitoring, using a variety of minimally invasive devices capable of rapid measurement of changes in flow or stroke volume during PLR. Examples include esophageal Doppler [22, 23], transthoracic Doppler ultrasound [24], transthoracic echocardiography (TTE) [2529], pulse contour techniques [28, 3033], bioreactance [3436], end tidal carbon dioxide change [37] and bio-impedance cardiography [38]. These studies have been summarised in three recent systematic reviews [3941], and all reported a high sensitivity regarding prediction of fluid responsiveness (all three reported the same pooled area under the receiver operating characteristic curve (AUC) of 0.95, with similar confidence intervals).

Only one of these [28] studies used the Vigileo™ monitor but with second-generation software.

The aim of this study was to assess the accuracy of the Vigileo™ monitor (Vigileo™; FloTrac; Edwards™; Lifesciences, Irvine, CA, USA) using third-generation software (version 3.02), in predicting preload responsiveness by measuring changes in CO during a PLR manoeuvre, in a mixture of medical and surgical critically ill patients with circulatory shock, with or without spontaneous breathing efforts or arrhythmias. In addition, we included an assessment of the accuracy of stroke volume variation (SVV) measured by the same device to predict responsiveness. We evaluated a cut point of 9.6 % at baseline (derived by Li et al. [42]), and the change in SVV during PLR were both evaluated as predictors of fluid responsiveness.

Methods

Patients

This prospective study was conducted on a mixed medical and surgical general critical care unit in a large non-university regional referral hospital. The study received research ethics committee approval and patients were enrolled following written informed consent.

Inclusion criteria

Critical care patients, over 18 years of age, requiring a fluid challenge as decided by the attending critical care physician, were included. The research team had no influence over this decision, and patients were only approached for enrolment following the decision to administer a fluid challenge.

This decision was based on the presence of at least one clinical sign of inadequate tissue perfusion, i.e. (a) systolic blood pressure <90 mmHg (or a decrease of >50 mmHg in previously hypertensive patients) or the need for vasoconstrictor drugs (vasopressin or norepinephrine); (b) urine output <0.5 ml/kg/h for ≥2 h; (c) tachycardia (heart rate >100/min); or (d) presence of skin mottling.

The Vigileo system (Edwards Lifescience, Irvine, CA) with arterial pressure waveform analysis device via special blood flow sensor (FloTrac Sensor, Edwards Lifesciences, Irvine, CA) using third-generation software must already be in situ.

Exclusion criteria

Any patients that were unable to perform PLR and who had any contraindications to fluid challenge, defined as life-threatening hypoxaemia, and evidence of blood volume overload and/or hydrostatic pulmonary oedema, were excluded.

Measurements

All haemodynamic data was continuously recorded on a Draeger monitoring system and a Vigileo system (Edwards Lifesciences, Irvine, CA) using third-generation software (version 3.02).

VigileoTM monitor measurements

The FloTrac transducer (FloTracTM, Edwards Lifesciences, Irvine, CA, USA) connected the indwelling arterial line to the VigileoTM System (Edwards Lifesciences, Irvine, CA, USA). This non-calibrated continuous CO monitor software analyses the arterial waveform with a frequency of a 100 Hz over 20 s. SV is calculated as k × pulsatility, where pulsatility is the standard deviation of arterial pressure over the preceding 20 s, and k is a factor which describes vascular compliance and resistance over the preceding 1 to 5 min, depending on the setting. This factor (so called proprietary Dynamic Tone Technology) is derived from a multivariate regression model taking into account Langewouter’s aortic compliance [43], mean arterial pressure (MAP), along with variance, skewness, and kurtosis of the arterial pressure wave [43, 44]. A greater number of hyperdynamic and vasodilated patients were incorporated into the algorithm database, and additional physiologically based variables were added to the algorithm’s vascular tone k factor in order to adjust automatically for hyperdynamic and vasodilated patients in the third-generation software update. This has increased the accuracy of SV measurement in vasodilated patients [4547].

Other measurements

The presence of any arrhythmias was recorded. Respiratory data was collated regarding the mode of ventilation, presence of spontaneous respiratory efforts, peak inspiratory pressures, plateau pressures, PEEP value and tidal volume. Dosages of any vasoactive drugs were recorded.

Study design

The study design is illustrated in Fig. 1. The following peak haemodynamic parameters were all recorded at each of the four stages in the protocol: heart rate (HR), mean arterial pressure (MAP), central venous pressure (CVP), cardiac index (CI), CO, SV, stroke volume index (SVI) and stroke volume variation (SVV). At the first baseline stage, measurements were taken while the patient was 45° semi-recumbent. During the next stage (PLR), the remote control of the bed was used to tilt the trunk to a horizontal position, and the legs were tilted 45° upwards. The patient was then returned to the 45° semi-recumbent position until the haemodynamic parameters stabilised to similar values as the original baseline. These values were recorded as representing the second baseline stage. Following this, a VE of 250 ml of GelofusineR was administered at the maximum rate allowed by the volumetric pump (within 15 min). The final stage, denoted as post volume expansion (PVE), occurred immediately following the fluid challenge, when all haemodynamic parameters were recorded once again.

Fig. 1.

Fig. 1

Study design. Patient positioning during the four stages of measurement as described

Statistical analysis

Continuous data are reported as mean and standard deviation or median and inter-quartile range. A change in CO from baseline of >15 and <15 % following the volume challenge was used to classify the patients as ‘responders’ and ‘non-responders’, respectively [3]. Measured variables were then compared between responders and non-responders using t tests, chi-squared tests or Fisher’s exact test for low counts, as appropriate for the data type, and reported as p values. The positively skewed variables of time to admission, tidal volume and norepinephrine dose were analysed after log-transforming the original data values. A receiver operating characteristic (ROC) curve was used to measure the sensitivity and specificity of the PLR test to predict responders to the fluid challenge.

p values of less than 0.05 were considered to be statistically significant. Statistical analysis was performed using R version 2.15.2 [48].

Results

Patient characteristics

Thirty-seven patients with CO monitoring in situ and meeting clinical criteria for a fluid challenge were initially included and four were subsequently excluded. Three had different diagnoses (cardiogenic shock with blood volume overload which was an exclusion criteria), and one had missing baseline and PLR CO data. The data from the remaining 33 patients who received a fluid challenge were analysed. Patient characteristics are summarised in Table 1 and were similar between groups. The majority (80 %), overall and in each group, were on a controlled mode of ventilation, with similar tidal volumes between groups. Cardiac arrhythmias were present in 9 % of all patients and balanced between the groups. The aetiology of circulatory insufficiency was septic shock in the majority (82 %) and severe systemic inflammatory response syndrome (SIRS) in the remainder.

Table 1.

Patient characteristics

All Responders Non-responders
n = 33 n = 10 n = 23 p
Age 60.0 (13.6) 58.4 (16.2) 60.7 (12.7) 0.70
Sex
 Male 15 (45 %) 4 (40 %) 11 (48 %) 0.97
 Female 18 (55 %) 6 (60 %) 12 (52 %)
Weight (kg) 72.1 (12.9) 75.0 (14.1) 70.9 (12.4) 0.44
APACHE II score 19.9 (6.9) 19.0 (5.8) 20.4 (7.5) 0.58
Died in hospital
 Yes 16 (48 %) 7 (70 %) 9 (39 %) 0.21
 No 17 (52 %) 3 (30 %) 14 (61 %)
Admission to PLR (h) 26 [18, 52] 38 [21,102] 26 [16, 42] 0.25
Arrhythmia
 Present 3/32 (9 %) 1/9 (11 %) 2/23 (9 %) 1.00
 Absent 29/32 (91 %) 8/9 (89 %) 21/23 (91 %)
Tidal volume (ml) 510 [463, 563] 540 [473, 660] 491 [460, 549] 0.39
Plateau pressure (cmH2O) 26.6 (7.1) 26.4 (5.5) 26.7 (7.7) 0.90
PEEP (cmH2O) 8.5 (2.5) 8.9 (2.8) 8.4 (2.4) 0.60
Respiratory rate (/min) 16.8 (4.2) 15.5 (2.4) 17.3 (4.7) 0.15
Spontaneous breathing
 Yes 6/30 (20 %) 2/9 (22 %) 4/21 (19 %) 1.00
 No 24/30 (80 %) 7/9 (78 %) 17/21 (81 %)
Arterial line site:
Femoral 4 (12 %) 1 (10 %) 3 (13 %) 1.00
Other (radial and brachial) 29 (88 %) 9 (90 %) 20 (87 %)
Diagnostic groups:
Septic shock 27 (82 %) 7 (70 %) 20 (87 %) 0.34
 – Endocarditis 1 (3 %) 0 (0 %) 1 (4 %)
 – Occult 7 (21 %) 2 (20 %) 5 (22 %)
 – Peritonitis 8 (24 %) 2 (20 %) 6 (26 %)
 – Pneumonia 11 (33 %) 3 (30 %) 8 (35 %)
SIRS 6 (18 %) 3 (30 %) 3 (13 %)
 – Ischaemic bowel 2 (6 %) 1 (10 %) 1 (4 %)
 – Occult 1 (3 %) 0 1 (4 %)
 – Pancreatico-duodenectomy 1 (3 %) 0 1 (4 %)
 – Pancreatitis 2 (6 %) 2 (20 %) 0

Data are shown as mean (standard deviation), median (inter-quartile range), or number (percentage) with p value comparing responders to non-responders for certain characteristics. Sample sizes are as in column headings unless stated (some variables had small numbers of missing values)

Baseline haemodynamic variables and vasoactive infusions

There were no statistically significant differences in baseline haemodynamic variables (Table 2). Vasoactive drug infusions (Table 3) were similar between groups, and only two patients did not receive any vasoactive infusions at the time of the PLR test. Thirty (91 %) patients received a norepinephrine infusion with a mean dose of 0.30 (0.17) mcg/kg/min. Seven patients received Dobutamine, and this was split between responders and non-responders. Vasopressin infusions were used in two patients. It was combined with norepinephrine in one patient and used as sole vasopressor in the other.

Table 2.

Haemodynamic variables at baseline

All Responders Non-responders p
n = 33 n = 10 n = 23
HR (/min) 100 (19.5) 106 (19.9) 98 (19.2) 0.29
CVP (mmHg) 10.6 (5.7) 9.9 (6.1) 10.9 (5.7) 0.66
MAP (mmHg) 69.5 (10.7) 65.1 (8.8) 71.5 (11.0) 0.09
PP (mmHg) 52.6 (13.1) 51.6 (17.8) 52.9 (11.2) 0.85
SV (ml/beat) 60.8 (14.5) 56.7 (16.3) 62.7 (13.7) 0.33
SVI (ml/m2/beat) 33.4 (7.8) 30.9 (8.8) 34.4 (7.4) 0.28
CO (L/min) 6.1 (1.5) 6.0 (1.6) 6.1 (1.5) 0.84
CI (L/min/m2) 3.3 (0.8) 3.3 (1.0) 3.4 (0.8) 0.81
SVV (%) 14.2 (11.0) 18.4 (10.5) 12.3 (11.0) 0.15
SVR (dynes s/cm5) 842 (347) 917 (504) 810 (259) 0.54

Data are shown as mean (standard deviation) or number (percentage) with p value comparing responders to non-responders.

HR heart rate, CVP central venous pressure, MAP mean arterial pressure, PP pulse pressure, SV stroke volume, SVI stroke volume index, CO cardiac output, CI cardiac index, SVV stroke volume variation, SVR systemic vascular resistance

Table 3.

Vasoactive drug infusions

All Responders Non-responders p
Norepinephrine (mcg/kg/min) 0.28 (0.20, 0.36) (n = 30) 0.31 (0.16, 0.37) (n = 10) 0.28 (0.21, 0.34) (n = 20) 0.92
Vasopressin (mcg/min) 4.5 (n = 1) 3.0 (n = 1)a
Dobutamine (mcg/kg/min) 4.6 (1.2) (n = 7) 4.4 (1.3) (n = 4) 4.8 (1.2) (n = 3)

Data are shown as mean (standard deviation) or median (inter-quartile range), and number of patients (n) (p value compares dose between responders and non-responders in the case of norepinephrine). In the case of vasopressin, the only dose is given, as there was one patient only in each group

aThe only patient requiring vasopressor support that did not include norepinephrine

Effects of PLR and VE on CO and other haemodynamic variables

The peak increase in CO occurred within 120 s in all cases. CO change during PLR was positively related to subsequent VE (correlation = 0.65) (Fig. 2). Spontaneous breathing and/or arrhythmia was only present in a few patients in each group thus not allowing any meaningful analysis of the influence of these two factors on the association (Table 1).

Fig. 2.

Fig. 2

Relationship between percentage increase in cardiac output from baseline 1 after PLR and PVE after baseline 2 and line of best fit (correlation = 0.65). Open circles are patients with spontaneous breathing efforts; triangles are patients with arrhythmias; crosses are patients with both spontaneous breathing efforts and arrhythmias, and black circles are patients with neither spontaneous breathing efforts nor arrhythmias

Effects of PLR and VE on changes in CO

The changes in CO induced by PLR were significantly greater in responders than in non-responders (p = 0.02). In fluid responders, CO increased by an average of 1.43 l/min (95 % CI 0.53 to 2.33) from baseline (average 5.98) to during PLR (average 7.41), corresponding to a 24 % increase from baseline (95 % CI 8 to 39 %), and by an average of 1.94 l/min (95 % CI 0.99 to 2.89) from before VE (average 5.39) to after VE (average 7.33), corresponding to a 35 % increase from baseline (95 % CI 24 to 47 %).

In fluid non-responders, CO increased by an average of 0.24 l/min (95 % CI 0.09 to 0.39) from baseline (average 6.10) to during PLR (average 6.34), corresponding to a 3 % increase from baseline (95 % CI 1 to 6 %), and by an average of 0.45 l/min (95 % CI 0.31 to 0.59) from before VE (average 5.94) to after VE (average 6.39), corresponding to a 7 % increase from baseline (95 % CI 5 to 10 %).

Seven patients had an insignificant reduction in their CO during PLR (Fig. 2).

Prediction of fluid responsiveness

The ROC curve (Fig. 3) shows the varying predictive performance of the PLR test as the cut point changes. The optimal cut point on the ROC curve for this dataset was approximately 9 %, i.e. an increase in CO of 9 % or greater during PLR predicts fluid responsiveness with the greatest accuracy generating a sensitivity of 80 % (95 % CI 44 to 96 %) and a specificity of 91 % (95 % CI 70 to 98 %).

Fig. 3.

Fig. 3

ROC curve demonstrating predictive performance after PLR on cardiac output response to PVE (AUC 0.85). Figures on the curve indicate the relevant cut point of cardiac output response (% change from baseline)

Although there was little distinction between a range of 10 to 15 % (for a 15 % cut point the sensitivity and specificity was 60 % (95 % CI 27 to 86 %) and 96 (95 % CI 76 to 99.8 %), the positive and negative predictive values were 86 % (695 % CI 42 to 99 %) and 85 % (95 % CI 64 to 95 %), respectively).

The area under the empirical ROC curve was 0.85 (95 % CI 0.63 to 1.00).

Using the increase in SVV during the PLR test as the predictor provides no better predictive performance than would be expected by chance with an area under the ROC curve of 0.56 (95 % CI 0.34 to 0.77).

An SVV cut point of 9.6 % at baseline, rather than change during PLR, as the predictor of fluid responsiveness, gives an area under the ROC curve of 0.74 (95 % CI 0.53 to 0.91), with 70 % sensitivity and 57 % specificity.

Discussion

Our study shows that changes in CO measured using the Vigileo™ monitor during a PLR manoeuvre is a useful predictor of fluid responsiveness in mixed critically ill patients with vasopressor-dependent circulatory shock. An increase in CO of ≥ 9 % during the PLR predicted an increase in CO of ≥15 % following subsequent volume expansion with good sensitivity (80 %) and specificity (91 %) resulting in an AUC of 0.85 (95 % CI 0.63 to 1.00). The high specificity has the potential to avoid the deleterious effects of unnecessary volume expansion in this patient group [4, 6, 7, 11].

Our results were consistent with previous validation studies of the PLR manoeuvre [2138, 49] which are encapsulated in three recent systematic reviews [3941]. These included nine, 21 and 23 studies assessing a total of 353, 991 and 1013, patients respectively. Our results were similar to the only other study using the Vigileo FloTrac device conducted by Biais et al. [14]. They reported a sensitivity of 85 % and a specificity of 90 % with a sample size of 34 patients and a peak change in flow within 120 s. Their study differed from our study in the following respects. They used second-generation FloTrac software whereas we used third-generation FloTrac software. Almost all the patients in the Biais study were surgical as opposed to an even split between medical and surgical in our study. The patients in our sample were much sicker, i.e. only 65 % were invasively ventilated in their study and all were breathing spontaneously with none receiving vasoactive drugs. That is in contrast with our study where all the patients were invasively ventilated and receiving vasoactive drugs (mean norepinephrine dosage was 0.3 ± 0.17mcg/kg/min). We reported a high APACHE2 score and hospital mortality.

Our study had the following limitations. Firstly, we identified less than one third of our sample as fluid responders (ten out of 33 subjects), whereas 50 % of patients were identified as responders in the systematic review by Michard et al. [3] and similarly in the PLR systematic reviews [3941]. This may have weakened our calculated sensitivity and specificity. Responders were defined using a cut off of 15 %, as this definition was used consistently throughout the other published PLR studies [2141, 49, 50], and we used the recommended PLR manoeuvre, i.e. patients started from a semi-recumbent position. This has been indentified as essential to ensure that splanchnic blood redistribution occurs thereby ensuring an adequate self-volume challenge [49].The small sample size and the volume chosen for our fluid challenges may have contributed to this lower rate of fluid responders. However, the sample size was still larger than half of all other PLR studies published [2138, 49]. We chose 250 ml of Gelofusine for our fluid challenges as this was a pragmatic study which followed the local clinical practice at that time. Although most of the studies in the systematic reviews [3941] used volumes of 500 ml, the study by Kang et al. [31] used 250 ml, and the study by Boulain et al. [21] used 300 ml of Gelatin. Both of these studies reported the usual rate of fluid responders and the Kang et al. study reported very high sensitivity and specificity. Boulain et al. discussed the rationale for choosing that volume, i.e. it is roughly equal to the volume of blood redistributed by the PLR manoeuvre. The volume and type of fluid we used were also consistent with the 2008 Surviving Sepsis Campaign Guidelines [51], a review by Vincent et al. [52] on circulatory shock and Trof et al. showed a more rapid change in SV with colloid boluses [53]. We delivered our volume challenge within fifteen minutes which is in keeping with the other studies.

Secondly, we measured CO using a single non-calibrated device and did not include a second calibrated device. However, the study by Biais et al. [28] had simultaneously measured haemodynamic changes during PLR with the Vigileo™ monitor and transthoracic echocardiography and found that the changes induced by volume expansion correlated well between these two devices (r 2 = 0.77, p < 0.0001). This was despite Biais et al. using the less accurate second-generation Vigileo software (version 1.14). In addition, several studies have validated the Vigileo third-generation software in the patient population we recruited. Meng et al. [46] assessed the trending ability of the Vigileo third-generation software using directly measured Oesophageal Doppler blood flow as the comparison device and found that the two devices showed 96 % concordance when measuring a change in preload induced by whole body tilting. De Backer et al. [45] reported acceptable accuracy using the third-generation software in patients with vasoplegia (septic shock and liver failure) as did Slagt et al. [47] in a systematic review comparing the different generations of Vigileo software. Despite the absence of a second calibrated device in our study, we derived a similar optimal cut point as the other published PLR studies [2141, 49, 50] and specifically in the subgroups that also used pulse contour methods [28, 3033].

Thirdly, in the majority of our patients (88 %), measurement was via a peripheral artery which may be less accurate than via a central artery in vasodilated patients [54]. The number receiving a central (femoral) arterial catheter was too few to analyse for this effect.

Finally, 39 % of the patients included in our study were surgical and therefore had the potential for intra-abdominal hypertension which has been shown by Mahjoub et al. [50] to reduce the accuracy of PLR for predicting fluid responsiveness. We did not have sufficient data on intra-abdominal pressures to analyse for this effect.

Conclusions

We have demonstrated that changes in cardiac output measured using the Vigileo™ monitor with third-generation software during a PLR test were predictive of fluid responsiveness in both medical and surgical patients with vasopressor-dependent circulatory shock. As the EV1000™ monitoring system, which has replaced the Vigileo™ monitor, continues to use the third-generation software for the FloTrac device these findings remain valid.

Acknowledgements

Lynne Bullock and Donna Harrison-Briggs participated in the data collection. This study received no financial support. The Vigileo™ monitor (Vigileo™; FloTrac; Edwards™; Lifesciences, Irvine, CA, USA) was part of our routine flow monitoring on our critical care unit for 3 years prior to this study.

Funding

This study was undertaken by the authors and the acknowledged members of the clinical team without any external funding.

Availability of data and material

The dataset supporting the conclusions of this article is included within the article (and its Additional file 1).

Authors’ contributions

AK conceived and designed the study, participated in the data collection and drafted the manuscript. MB participated in the data collection and contributed to drafting the manuscript. TF provided the statistical analysis and contributed to drafting the manuscript. AK has full access to all the study data and takes responsibility for data integrity and accuracy of data analysis. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Not applicable.

Ethics approval

The UK NRES Bolton Research Ethics Committee (REC reference number 08/H1009/61) approved the study.

Abbreviations

AUC

Area under the receiving operating characteristic curve

CO

Cardiac output

PLR

Passive leg raise

PVE

Post volume expansion

SV

Stroke volume

TTE

Transthoracic echocardiography

VE

Volume expansion

Additional file

Additional file 1: (8.1KB, csv)

Study dataset. (CSV 8 kb)

Contributor Information

Anton Krige, Email: anton.krige@elht.nhs.uk.

Martin Bland, Email: martin.bland@elht.nhs.uk.

Thomas Fanshawe, Email: thomas.fanshawe@phc.ox.ac.uk.

References

  • 1.Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, Peterson E, Tomlanovich M. Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med. 2001;345:1368–1377. doi: 10.1056/NEJMoa010307. [DOI] [PubMed] [Google Scholar]
  • 2.Kumar A, Anel R, Bunnell E, Habet K, Zanotti S, Marshall S, Neumann A, Ali A, Cheang M, Kavinsky C, Parillo J. Pulmonary artery occlusion pressure and central venous pressure fail to predict ventricular filling volume, cardiac performance, or the response to volume infusion in normal subjects. Crit Care Med. 2004;32(3):691–699. doi: 10.1097/01.CCM.0000114996.68110.C9. [DOI] [PubMed] [Google Scholar]
  • 3.Michard F, Teboul JL. Predicting fluid responsiveness in ICU patients: a critical analysis of the evidence. Chest. 2002;121:2000–2008. doi: 10.1378/chest.121.6.2000. [DOI] [PubMed] [Google Scholar]
  • 4.Alsous F, Khamiees M, DeGirolamo A, Amoateng-Adjepong Y, Manthous CA. Negative fluid balance predicts survival in patients with septic shock. Chest. 2002;117:1749–1754. doi: 10.1378/chest.117.6.1749. [DOI] [PubMed] [Google Scholar]
  • 5.Holte K, Kehlet H. Fluid therapy and surgical outcomes in elective surgery: a need for reassessment in fast-track surgery. J Am Coll Surg. 2006;202:971–989. doi: 10.1016/j.jamcollsurg.2006.01.003. [DOI] [PubMed] [Google Scholar]
  • 6.Wiedemann HP, Wheeler AP, Bernard GR, et al. Comparison of two fluid-management strategies in acute lung injury. N Engl J Med. 2006;354:2564–2575. doi: 10.1056/NEJMoa062200. [DOI] [PubMed] [Google Scholar]
  • 7.Boyd JH, Forbes J, Nakada TA, et al. Fluid resuscitation in septic shock: a positive fluid balance and elevated central venous pressure are associated with increased mortality. Crit Care Med. 2011;39:259–265. doi: 10.1097/CCM.0b013e3181feeb15. [DOI] [PubMed] [Google Scholar]
  • 8.Nisanevich V, Felsenstein I, Almogy G, et al. Effect of intraoperative fluid management on outcome after intra-abdominal surgery. Anesthesiology. 2005;103:25–32. doi: 10.1097/00000542-200507000-00008. [DOI] [PubMed] [Google Scholar]
  • 9.Bundgaard-Nielsen M, Secher NH, Kehlet H. ‘Liberal’ vs. ‘restrictive’ perioperative fluid therapy—a critical assessment of the evidence. Acta Anaesthesiol Scand. 2009;53:843–851. doi: 10.1111/j.1399-6576.2009.02029.x. [DOI] [PubMed] [Google Scholar]
  • 10.Holte K, Klarskov B, Christensen DS, et al. Liberal versus restrictive fluid administration to improve recovery after laparoscopic cholecystectomy: a randomized, double-blind study. Ann Surg. 2004;240:892–899. doi: 10.1097/01.sla.0000143269.96649.3b. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Weidemann HP, Wheeler AP, Bernard GR, Thompson BT, Hayden D, deBoisblanc B, Connors AF, Jr, Hite RD, Harabin AL. Comparison of two fluid management strategies in acute lung injury. N Engl J Med. 2006;354:2564–2575. doi: 10.1056/NEJMoa062200. [DOI] [PubMed] [Google Scholar]
  • 12.Cavallaro F, Sandroni C, Antonelli M. Functional hemodynamic monitoring and dynamic indices of fluid responsiveness. Minerva Anestesiol. 2008;74:137–143. [PubMed] [Google Scholar]
  • 13.Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37:2642–2647. doi: 10.1097/CCM.0b013e3181a590da. [DOI] [PubMed] [Google Scholar]
  • 14.Michard F, Teboul JL. Using heart-lung interactions to assess fluid responsiveness during mechanical ventilation. Crit Care. 2000;4:282–289. doi: 10.1186/cc710. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Perner A, Faber T. Stroke volume variation does not predict fluid responsiveness in patients with septic shock on pressure support ventilation. Acta Anaesthesiol Scand. 2006;50:1068–1073. doi: 10.1111/j.1399-6576.2006.01120.x. [DOI] [PubMed] [Google Scholar]
  • 16.Hofer CK, Senn A, Weibel L, Zollinger A. Assessment of stroke volume variation for prediction of fluid responsiveness using the modified FloTrac and PiCCOplus system. Crit Care. 2008;12:R82. doi: 10.1186/cc6933. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Soubrier S, Saulnier F, Hubert H, Delour P, Lenci H, Onimus T, Nseir S, Durocher A. Can dynamic indicators help the prediction of fluid responsiveness in spontaneously breathing critically ill patients? Intensive Care Med. 2007;33:1117–1124. doi: 10.1007/s00134-007-0644-9. [DOI] [PubMed] [Google Scholar]
  • 18.Rutlen DL, Wackers FJ, Zaret BL. Radionuclide assessment of peripheral intravascular capacity: a technique to measure intravascular volume changes in the capacitance circulation in man. Circulation. 1981;64:146–152. doi: 10.1161/01.CIR.64.1.146. [DOI] [PubMed] [Google Scholar]
  • 19.Gaffney FA, Bastian BC, Thal ER, et al. Passive leg raising does not produce a significant or sustained autotransfusion effect. J Trauma. 1982;22:190–193. doi: 10.1097/00005373-198203000-00003. [DOI] [PubMed] [Google Scholar]
  • 20.Wong DH, O’Connor D, Tremper KK, et al. Changes in cardiac output after acute blood loss and position change in man. Crit Care Med. 1989;17:979–983. doi: 10.1097/00003246-198910000-00002. [DOI] [PubMed] [Google Scholar]
  • 21.Boulain T, Achard JM, Teboul JL, Richard C, Perrotin D, Ginies G. Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest. 2002;121:1245–1252. doi: 10.1378/chest.121.4.1245. [DOI] [PubMed] [Google Scholar]
  • 22.Lafaneche’re A, P’ene F, Goulenok C, Delahaye A, Mallet V, Choukroun G, Chiche JD, Mira JP, Cariou A. Changes in aortic blood flow induced by passive leg raising predicts fluid responsiveness in critically ill patients. Crit Care. 2006;10:R132–R140. doi: 10.1186/cc5044. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Monnet X, Rienzo M, Osman D, Anguel N, Richard C, Pinsky MR, Teboul JL. Passive leg raising predicts fluid responsiveness in the critically ill. Crit Care Med. 2006;34:1402–1407. doi: 10.1097/01.CCM.0000215453.11735.06. [DOI] [PubMed] [Google Scholar]
  • 24.Thiel SW, Kollef MH, Isakow W. Non-invasive stroke volume measurement and passive leg raising predicts volume responsiveness in medical ICU patients: an observational cohort study. Crit Care. 2009;13:R111–R120. doi: 10.1186/cc7955. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Lamia B, Ochagavia A, Monnet X, Chemla D, Richard C, Teboul JL. Echocardiographic prediction of volume responsiveness in critically ill patients with spontaneous breathing activity. Intensive Care Med. 2007;33:1133–1138. doi: 10.1007/s00134-007-0642-y. [DOI] [PubMed] [Google Scholar]
  • 26.Maizel J, Airapetian N, Lorne E, Tribouilloy C, Massy Z, Slama M. Diagnosis of central hypovolemia by using passive leg raising. Intensive Care Med. 2007;33:1133–1138. doi: 10.1007/s00134-007-0642-y. [DOI] [PubMed] [Google Scholar]
  • 27.Caille V, Jabot J, Belliard G, Charron C, Jardin F, Vieillard-Baron A. Hemodynamic effects of passive leg raising: an echocardiographic study in patients with shock. Intensive Care Med. 2008;34:1239–1245. doi: 10.1007/s00134-008-1067-y. [DOI] [PubMed] [Google Scholar]
  • 28.Biais M, Vidil L, Sarrabay P, Cottenceau V, Revel P, Sztark F. Changes in stroke volume induced by passive leg raising in spontaneously breathing patients: comparison between echocardiography and Vigileo TM/FloTracTM device. Crit Care. 2009;13:R195. doi: 10.1186/cc8195. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Pr’eau S, Saulnier F, Dewavrin F, Durocher A, Chagnon JL. Passive leg raising is predictive of fluid responsiveness in spontaneous breathing patients with severe sepsis or acute pancreatitis. Crit Care Med. 2010;38:819–825. doi: 10.1097/CCM.0b013e3181c8fe7a. [DOI] [PubMed] [Google Scholar]
  • 30.Lakhal K, Ehrmann S, Runge I, et al. Central venous pressure measurements improve the accuracy of leg raising-induced change in pulse pressure to predict fluid responsiveness. Intensive Care Med. 2010;36:940–948. doi: 10.1007/s00134-010-1755-2. [DOI] [PubMed] [Google Scholar]
  • 31.Kang WS, Kim SH, Kim SY, Oh CS, Lee SA, Kim JS. The influence of positive end-expiratory pressure on stroke volume variation in patients undergoing cardiac surgery: an observational study. J Thorac Cardiovasc Surg. 2014;148:3139–3145. doi: 10.1016/j.jtcvs.2014.07.103. [DOI] [PubMed] [Google Scholar]
  • 32.Geerts B, de Wilde R, Aarts L, et al. Pulse contour analysis to assess hemodynamic response to passive leg raising. J Cardiothorac Vasc Anesth. 2011;25:48–52. doi: 10.1053/j.jvca.2010.09.013. [DOI] [PubMed] [Google Scholar]
  • 33.Dong ZZ, Fang Q, Zheng X, et al. Passive leg raising as an indicator of fluid responsiveness in patients with severe sepsis. World J Emerg Med. 2012;3:191–196. doi: 10.5847/wjem.j.issn.1920-8642.2012.03.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Benomar B, Ouattara A, Estagnasie P, et al. Fluid responsiveness predicted by noninvasive bioreactance-based passive leg raise test. Intensive Care Med. 2010;36:1875–1881. doi: 10.1007/s00134-010-1990-6. [DOI] [PubMed] [Google Scholar]
  • 35.Marik PE, Levitov A, Young A, et al. The use of bioreactance and carotid Doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest. 2013;143:364–370. doi: 10.1378/chest.12-1274. [DOI] [PubMed] [Google Scholar]
  • 36.Duus N, Shogilev DJ, Skibsted S, et al. The reliability and validity of passive leg raise and fluid bolus to assess fluid responsiveness in spontaneously breathing emergency department patients. J Crit Care. 2015;30:217.e1–217.e5. doi: 10.1016/j.jcrc.2014.07.031. [DOI] [PubMed] [Google Scholar]
  • 37.Monge García MI, Gil Cano A, Gracia Romero M, et al. Non-invasive assessment of fluid responsiveness by changes in partial end-tidal CO2 pressure during a passive leg-raising maneuver. Ann Intensive Care. 2012;2:9. doi: 10.1186/2110-5820-2-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Fellahi JL, Fischer MO, Dalbera A, et al. Can endotracheal bioimpedance cardiography assess hemodynamic response to passive leg raising following cardiac surgery? Ann Intensive Care. 2012;2:26. doi: 10.1186/2110-5820-2-26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Cavallaro F, Sandroni C, Marano C, La Torre G, Chiara A, De Waure C, Bello G, Maviglia R, Antonelli M. Diagnostic accuracy of passive leg raising for prediction of fluid responsiveness in adults: systematic review and meta-analysis of clinical studies. Intensive Care Med. 2010;36:1475–1483. doi: 10.1007/s00134-010-1929-y. [DOI] [PubMed] [Google Scholar]
  • 40.Monnet X, Marik P, Teboul J-L: Passive leg raising for predicting fluid responsiveness: a systematic review and meta-analysis. Intensive Care Med. published online 20th Jan 2016 [DOI] [PubMed]
  • 41.Chernapath TGV, Hirsch A, Geerts BF, et al. Predicting fluid responsiveness by passive Leg raising: a systematic review and meta-analysis of 23 clinical trials. Crit Care Med. 2016;44(5):981–91. doi: 10.1097/CCM.0000000000001556. [DOI] [PubMed] [Google Scholar]
  • 42.Cheng L, Lin F-q, Shu-kun F, et al. Stroke volume variation for prediction of fluid responsiveness in patients undergoing gastrointestinal surgery. Int J Med Sci. 2013;10(2):148–155. doi: 10.7150/ijms.5293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Langewouters GJ, Wesseling KH, Goedhard WJ. The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component model. J Biomech. 1985;18:613–620. doi: 10.1016/0021-9290(85)90015-6. [DOI] [PubMed] [Google Scholar]
  • 44.Pratt B, Roteliuk L, Hatib F, Frazier J, Wallen RD. Calculating arterial pressure-based cardiac output using a novel measurement and analysis method. Biomed Instrum Technol. 2007;41:403–411. doi: 10.2345/0899-8205(2007)41[403:CAPCOU]2.0.CO;2. [DOI] [PubMed] [Google Scholar]
  • 45.De Backer D, Marx G, Tan A, et al. Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients. Intensive Care Med. 2011;37:233–240. doi: 10.1007/s00134-010-2098-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Meng L, Tran PS, Brenton AS, Lanning K, et al. The impact of phenyephrine, ephedrine and increased preload on the third-generation Vigileo-Flotrac and esophageal Doppler cardiac output measurements. Anesth Analg. 2011;113:751–7. doi: 10.1213/ANE.0b013e31822649fb. [DOI] [PubMed] [Google Scholar]
  • 47.Slagt C, Malagon I, Groeneveld ABG. Systematic review of uncalibrated arterial pressure waveform analysis to determine cardiac output and stroke volume variation. Br J Anaesth. 2014;112(4):626–37. doi: 10.1093/bja/aet429. [DOI] [PubMed] [Google Scholar]
  • 48.R Development Core Team (2008). R: a language and environment for statistical computing. R Foundation for Statistical Computing. Vienna, Austria. ISBN 3-900051-07-0, URLhttp://www.R-project.org.
  • 49.Jabot J, Teboul JL, Richard C, Monnet X. Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009;35:85–90. doi: 10.1007/s00134-008-1293-3. [DOI] [PubMed] [Google Scholar]
  • 50.Mahjoub Y, Touzeau J, Airapetian N, Lorne E, Hijazi M, Zogheb E, Tinturier F, Slama M, Dupont H. The passive leg-raising maneuver cannot accurately predict fluid responsiveness in patients with intra-abdominal hypertension. Crit Care Med. 2010;38(9):1824–1829. doi: 10.1097/CCM.0b013e3181eb3c21. [DOI] [PubMed] [Google Scholar]
  • 51.Dellinger RP, Levy MM, Carlet JM, et al. Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock. Crit Care Med. 2008;36:296–327. doi: 10.1097/01.CCM.0000298158.12101.41. [DOI] [PubMed] [Google Scholar]
  • 52.Vincent JL, Weil MH. Fluid challenge revisited. Crit Care Med. 2006;34:1333–1337. doi: 10.1097/01.CCM.0000214677.76535.A5. [DOI] [PubMed] [Google Scholar]
  • 53.Trof RJ, Sukul SP, Twisk JW, et al. Greater cardiac response of colloid than saline fluid loading in septic and non-septic critically ill patients with clinical hypovolaemia. Intensive Care Med. 2010;36:697–701. doi: 10.1007/s00134-010-1776-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Dorman T, Breslow MJ, Lipsett PA, Rosenberg JM, Balser JR, Almog Y, Rosenfeld BA. Radial artery pressure monitoring underestimates central arterial pressure during vasopressor therapy in critically ill surgical patients. Crit Care Med. 1998;26:1646–1649. doi: 10.1097/00003246-199810000-00014. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Intensive Care are provided here courtesy of BMC

RESOURCES