Abstract
Respiratory disease is the leading cause of death in the UK. Methods for assessing pulmonary function and chest wall movement are essential for accurate diagnosis, as well as monitoring response to treatment, operative procedures and rehabilitation. Despite this, there is a lack of low-cost devices for rapid assessment. Spirometry is used to measure air flow expired, but cannot infer or directly measure full chest wall motion. This paper presents the development of a low-cost chest wall motion assessment system. The prototype was developed using four Microsoft Kinect sensors to create a 3D time-varying representation of a patient’s torso. An evaluation of the system in two phases is also presented. Initially, static volume of a resuscitation mannequin with that of a Nikon laser scanner is performed. This showed the system has slight underprediction of 0.441 %. Next, a dynamic analysis through the comparison of results from the prototype and a spirometer in nine cystic fibrosis patients and thirteen healthy subjects was performed. This showed an agreement with correlation coefficients above 0.8656 in all participants. The system shows promise as a method for assessing respiratory disease in a cost-effective and timely manner. Further work must now be performed to develop the prototype and provide further evaluations.
Keywords: Thoracic wall, Chest wall, Thoracic surgery, Respiratory system diagnostic technique, Medical device design
Introduction
Respiratory disease, including lung cancer, is the leading cause of death in the UK, accounting for 920,000 disability-adjusted life years lost [10]. It is the most frequent cause of disease in primary care in all age groups and the second most common cause of chronic conditions.
For patients who report to A&E, a quick and low-cost method of assessment is required. One process currently used to assess respiratory disease in a low-cost and time-effective manner is spirometry [14]. Spirometry allows for the measurement of expired airflow from the lungs, enabling physicians to better characterise the cause of breathlessness and to assess progression of respiratory disease over time.
However, spirometry can have significant limitations. Firstly, forced spirometric efforts allow assessment of initial diaphragm/muscle strength and can only measure total airflow in and out of the lungs; it therefore provides a limited amount of feedback and does not allow physicians to identify motion at the chest and the relative contribution of different areas of each lung to the subjects’ respiratory function [16]. This is particularly important in subjects with more focal lung abnormalities, such as emphysematous bullae, collapsed lung segments and previous surgical lung resection. Secondly, the effect of chest wall abnormalities, such as respiratory muscle weakness and pectus excavatum, as well as diaphragm movement cannot be assessed by spirometry [21]. Thirdly, since it is an effort-dependent procedure, there is a potential for inaccurate results in subjects unable to reliably perform a forced expiratory manoeuvre [12] (e.g. children, the elderly and subjects with hearing impairments, learning difficulties or a language barrier). Fourthly, subjects with facial abnormalities or muscle weakness are often unable to form a tight seal around the mouthpiece, preventing spirometry being accurately performed [8].
As a result of these limitations, there has been increasing interest in the development of alternative methods of assessing respiratory function, including chest wall motion analysis. Systems such as magnetometers [17], respiratory inductance plethysmography [13], optoelectronic plethysmography [1, 4] (OEP) and structured light plethysmography [5] (SLP) have been demonstrated for this purpose; however, these technologies have been shown can be cumbersome, expensive, time consuming, difficult to interpret and not suited to the clinical environment. A more recent method for evaluating chest wall motion is through the use of the Microsoft Kinect technology (a low-cost motion tracking camera). This system is portable, low cost, non-invasive and has been suggested as an alternative for spirometry [3] with positive correlations being shown (Ye et al. 2012 show correlations of r = 0.966 [24]). However, research demonstrating the use of this technology utilises a one-camera version of the system [3, 7, 23, 24]. This method follows a process of monitoring the chest wall only and detecting changes in surface. The issue with this is that motion may affect the sample, particularly large movements during analysis. In addition, it is difficult to compartmentalise different parts of the torso and also calculate volumes of the whole or parts of the torso.
This paper reports the development of a low-cost and time-effective novel prototype for capturing dynamic chest wall motion using four Microsoft Kinect sensors. An initial evaluation of this prototype is also presented, involving healthy volunteers and adults with cystic fibrosis (an inherited condition causing progressive respiratory failure) [20].
Methods
System design
This research created a system which was capable of assessing respiratory motion using a low-cost and time-effective motion tracking device. The device to be used was the Microsoft Kinect, which is a human tracking peripheral used in the gaming industry for the Microsoft Xbox to provide low-cost 3D motion capture capabilities [6]. The device is composed of an infrared (IR) laser projector, an IR camera, a colour camera and a microphone array. The Kinect uses the IR projector to detect distance of an object from the sensor by emitting a single infrared beam which is split to create an invisible pattern of speckles [11]. This pattern is captured by the IR camera and compared against a reference pattern stored in the device to calculate distance from external objects. This study utilised four of these devices to create three-dimensional representations of a subject’s torso over time, so difference could be calculated by comparing the difference between frames.
System evaluation: static analysis methods
In order to evaluate the derived system, static testing was initially performed to evaluate overall accuracy of the system. A cardiopulmonary resuscitation (CPR) training mannequin was used as the evaluation object, of which volume could be calculated. This was a torso-type mannequin measuring 63.5 cm in height and weighing 3.9 kg.
Initially, the mannequin was recorded five times by the developed, Kinect-based system in a room with mainly artificial light. Each capture lasted approximately 15 s.
Next, a Nikon Metrology MCA 2400 M7 articulated arm laser scanner (Nikon, Tokyo, Japan) was used to capture complete detailed geometric scans. Single point accuracy of the scanner is 0.031 mm, and length accuracy is 0.042 mm with an uncertainty of ±2σ. A typical scan path for the laser device can capture up to 72,000 points per second.
Results from the two scans were then compared for static volume estimation.
To make a comparison between the Kinect-based system and Nikon laser scanner, initially a root mean square (RMS) error was calculated, including mean and mode. Probability density and cumulative density function are then also utilised to assess estimate errors and bias.
System evaluation: dynamic capture methods
Chest wall motion was measured with the developed, Kinect-based system, whilst participants simultaneously performed spirometry using a MicroLab Spirometer (CareFusion Corp, San Diego, USA).
Spirometry was performed according to ATS/ERS guidelines [14]. CareFusion Spirometry PC Software version 1.07 (CareFusion Corp, San Diego, USA) was used to store patient information and spirometry results.
Nine subjects with cystic fibrosis were recruited at the West Midlands Adult Cystic Fibrosis Centre and 13 healthy volunteers from staff at Heart of England NHS Foundation Trust.
The acquisition protocol involved capturing quiet breathing for 20 s, followed by a relaxed vital capacity (VC) manoeuver (maximum inspiration and expiration) and followed by 20 s of quiet breathing. The test was run three times per patient, as per ATS/ERS guidelines. Males were captured bare-chested, whilst females wore tight t-shirts for modesty. From total chest wall volumes acquired by the Kinect-based device, the following parameters were obtained: tidal volume (VT), as the total chest wall volume variation, respiratory rate (RR) and minute ventilation (VT × RR).
In order to process the spirometry readings for analysis, some data required additional processing. Initially, spirometry data and time stamps were imported into MATLAB. The time series captured needed a shape-preserving piecewise cubic interpolation procedure to ensure a constant sampling rate (5 Hz). A fourth-order zero-phase band-pass Butterworth filter [0.025–1 Hz] was applied.
The two volume time series were aligned by searching for the peak in their cross-correlation function. A measure of similarity was needed to compare the two time series over a window of interest. The range for comparison was set at −8 to 14 s relative to the maximum point on the relaxed VC measurement. This window length was the largest possible to ensure all subjects tested had sufficient overlap in the time series recorded. The Pearson product moment correlation coefficient was chosen as a simple measure of similarity:
where is covariance between the Kinect-based and spirometry-based volume estimates and the and σ S are their respective standard deviations. It takes values at ±1 and at its extremes this corresponds to data points lying perfectly on a line. A summary of all of the correlation coefficients obtained between the Kinect-based and spirometry measurements for the two groups (healthy subjects and CF patients) is presented. This is followed by a one-way ANOVA to compare sample means between the two groups. Finally, a TLS fitting procedure is used to compare results from the Kinect-based system and spirometry measurements.
The research study was approved by the Local Research Ethics Committee (reference number: 10/H1202/58). Informed consent was obtained from all involved participants, and the study conforms to the Declaration of Helsinki.
Results
System implementation
To enable chest wall motion analysis, a system comprising four Kinect sensors was developed and orientated as shown in Fig. 1. The Kinect sensors were placed around the subject, ensuring a minimum distance of 1 metre, following recommendations from Alnowami et al. [2].
Each Kinect sensor was connected to a PC using an independent and dedicated USB bus. The general settings of the system allowed for 30 frames per second to be captured in 320 × 240 pixels for each sensor.
A bespoke software application was designed, using the Kinect Software Development Kit (SDK) 1.6 (Microsoft, Redmond, USA) which assisted the process.
In order to calibrate the system, a static object was placed in the middle of the four Kinects. The bespoke software was used to take a scan using each Kinect individually, with the other cameras turned off, to avoid interference from competing infrared laser projectors. The four separate scans taken from each Kinect were then exported from the bespoke software and imported into the meteorology software Geomagic Studio 2012. This software allowed for the four images to be aligned using in-built functionality to map the four images, and form a point cloud. This resulted in a transformation matrix being created which detailed the transformations of each of the original four scans. These were then exported back to the bespoke software.
At this point, a patient could be placed in between the four cameras and a recording taken of their torso, using the bespoke software. The images at each frame, from each camera, were combined using the previously generated transformation matrix from the calibration step. This formed a set of 3D point clouds at each frame captured, from which change in volume could be calculated, by comparing the point cloud at each frame.
In order to perform this calculation, Geomagic Studio 2012 was utilised again, using a set macro which was created. This allowed for a repetition of the same task, which included loading each of the point clouds which were exported from the bespoke software, building a 3D mesh (triangles and normals), filling any holes in the mesh and using the mesh doctor to smooth the resulting object. The user is asked to manually define the limits of the analysis before this macro is run, placing a bounding box around an initial image to limit volume calculations to the torso only. These limits are then applied to all meshes within the analysis. This process also helps to eliminate noise from outside the bounding box. An example of a scan taken of a mannequin is shown in Fig. 2.
A 3D model is created for each subsequent frame, and this allows change in volume in the torso to be detected. An example of the change in volume in a model is shown in Fig. 3.
In order to achieve this, a ply file was stored for each point cloud. 3D volume data and the individual frame time stamps for all four Kinect sensors were exported from Geomagic and imported into MATLAB R2012B (version 8.0) (MathWorks, Natick, USA). Volumes were calculated using a fast algorithm in MATLAB [19]. Occasional misalignment with the four camera time stamps caused errors in the 3D reconstruction of the chest wall volume. This occurred as the clocks for the four Kinect sensors cannot be synchronised with the drivers supplied in the Microsoft SDK. An algorithm was implemented to detect sudden sharp changes in volume, making the assumption that these were artefacts. The resulting 3D volume time series did not have a constant sampling rate, due to both the artefact removal process and the way the Kinect sensor captures data. Therefore, a shape-preserving, piecewise cubic interpolation procedure [9] was used to construct from the original time series one with a constant sampling rate of 30 Hz. A fourth-order zero-phase band-pass Butterworth filter [22] [0.025–1 Hz] was used to remove noise and emphasise the frequency region of interest, and a further down-sampling procedure was applied to reduce the sampling rate to 5 Hz. This also removed the static volume of each subject under test and only estimated the dynamic volume.
System evaluation: static analysis results
The mannequin was successfully scanned by the Nikon Metrology laser scanner and the Kinect-based system.
The estimates from the Kinect-based system were processed, and the gold standard reference volume (22.751 l) from the Nikon Metrology scanner was subtracted. The root mean square (RMS) error for the Kinect-based estimate was litres to three decimal places, corresponding to an error of 0.441 %. The mean of the error is 0.0626 l and the mode (most common value) 0.0216 l.
Figure 4 shows the empirical probability density and cumulative density function (CDF) (blue curves) of the Kinect-based volume estimate errors. It can be seen that the error distribution is skewed with a long right tail (indicating underestimation) and has a nonzero mean and mode (indicating bias). Taken together, these make it more likely that the volume is underestimated with the Kinect sensors. A series of probability distributions were fitted, and the best was chosen based on Akaike information criterion (AIC). This corresponded to a generalised extreme value distribution, as it could adequately model the long right tail seen in the empirical results. Figure 4 also shows the fitted generalised extreme value distribution (red curve) and a fitted normal/Gaussian distribution (black curve) for reference.
The probability density function for the generalised extreme value distribution was given by:
where is a location parameter, a scale parameter and a shape parameter. The fitted parameter values and their lower and upper 95 % confidence intervals (CI) were found to be:
μ = 0.0252 [0.0220, 0.0284] (location);
σ = 0.0580 [0.0557, 0.060] (scale);
k = 0.0641 [0.0231, 0.1051] (shape).
The cumulative density function was given by:
In order to derive error bounds for the distribution, the CDF was inverted to give (i.e. the volume error for a given cumulative probability). Thus, the error , with a probability of (F=) 0.95, and the error , with a probability of 0.05, were obtained. It is thus possible to define the 95 % error bounds for static volume estimation as .
System evaluation: dynamic capture results
Scans using the developed system, and simultaneous spirometry readings, were successfully taken of nine subjects with cystic fibrosis and 13 healthy volunteers. Table 1 presents the demographics of the study participants.
Table 1.
CF volunteers, n = 9 | Healthy volunteers, n = 13 | |
---|---|---|
Age (years), mean (SD) | 32.8 (9.9) | 30.0 (8.1) |
Sex (male/female) | 6/3 | 7/6 |
Weight (kg), mean (SD) | 67.3 (15.1) | 66.6 (15.7) |
Height (cm), mean (SD) | 168.9 (10.8) | 167.5 (9.9) |
BMI (kg/m2), mean (SD) | 23.4 (3.3) | 23.6 (4.5) |
Figure 5 shows a representative example of the Kinect-based system (solid blue line) and spirometry (solid red line) measurements obtained during periods of VC measurement and relaxed breathing in an individual subject. Both data series show estimated 95 % CI with the shaded areas. The Kinect sensor CI was derived from the static mannequin results and is clearly asymmetric or skewed towards lower volumes. The spirometry CI was obtained from the CareFusion Microlab Spirometer specifications (www.carefusion.com/medical-products/respiratory/cardio-pulmonary-diagnostics/pulmonary-function-testing/spirometers/microlab.aspx, accessed 12 September 2014). Accuracy was reported to be ±3 % in line with ATS/ERS standards.
There was generally good agreement between Kinect-based measurements and spirometry data. Also shown in Fig. 5 is a black arrow indicating the region around the two VC measurements used to calculate the correlation coefficient, used as a similarity measure. This region was defined in all cases by −8 to +12 s, relative to the peak of the VC time course.
Figure 6 shows the aligned point for point comparison of the region of interest (−8 to +12 s window) around the peak in an exemplar relaxed VC measurement, for patient 1, run 2. Indicated in this figure is the estimated correlation coefficient, , with upper and lower bounds showing a significant (p < 0.001) correlation between the two time series. Also indicated in this figure by the red line is the total least squares (TLS) regression estimate [15] deemed to be more accurate than standard linear regression, treating the spirometry estimate as the regressor, as both the Kinect and spirometry estimates contain error. The TLS fitting produces an estimate of regression slope and intercept and their respective CI. In the ideal situation, the TLS fit should have+ a slope of 1 and an intercept of 0. In this case, the slope is 0.9601 [0.9580, 0.9621] and the intercept is 0.0167 [0.0147, 0.0187]. The residual standard error from the linear fit is 0.0809 l and can be interpreted as a further similarity measure—the lower the error, the better the fit.
Time series comparison of relaxed VC between Kinect-based and spirometry measurements, for patient 1, run 2.
Figure 7 shows all of the correlation coefficients obtained between the Kinect-based and spirometry measurements for all subjects and repeated measurements. The figure shows a box-whisker plot for both groups. The central mark indicates the group median values, the edges of the box for the 25th and 75th percentiles and the whiskers extend to the most extreme data points. Identified outliers are shown by the red crosses. All correlation coefficients not deemed outliers were above 0.8656 for the patient group and above 0.9226 for the healthy volunteer group. All correlation coefficients were shown to be significant at a CI of 99 % (p values <0.001), where they were computed from a t distribution, assuming underlying bivariate normal distribution of the data.
A one-way ANOVA (F[1,51] = 7.5783; p = 0.0082) indicated that the two groups’ sample means were significantly different. Thus, the CF subjects demonstrated slightly poorer correlation coefficients.
Measurements of dynamic lung volume for all subjects and repeated measurements.
It was observed, but not plotted here for brevity, that there was a small but significant (p < 0.001) negative correlation [] between the correlation coefficients and the residual error from the TLS fitting procedure. Both the TLS fit and correlation coefficient measures attempt to determine the degree of similarity between the Kinect-based and spirometry measurements. Larger TLS residual error occurs where the two estimates are not well represented by a linear relationship, also indicated by a low correlation coefficient.
Histograms of total least squares regression fitting parameters, intercept and slope, for all dynamic lung volume measurements.
Figure 8 shows histograms of all the fitted intercept and slope estimates from the TLS regression. Typically, the intercept (left) is grouped around 0. The slopes are skewed to slightly less than unity. As established in the static analysis, the error on the combined Kinect-based system estimates tended towards underestimation rather than overestimation. This is seen in the bias (nonzero mean and median) and the long tail (skewed nature of the extreme value distribution). The TLS slope estimates are less than unity, indicating tendency towards underestimation.
Discussion
This paper reports the development of a low-cost and portable method for analysing chest wall motion through the use of four Microsoft Kinect devices used in parallel. The system uses a combination of bespoke software and off-the-shelf software, in the form of Geomagic Studio, to evaluate a three-dimensional model of a patient’s torso taken over time and evaluate changes in volume in order to calculate chest wall motion. Static testing and testing on healthy volunteers and cystic fibrosis patients in comparison with spirometry show good correlation (albeit a small underprediction).
The system could prove useful for assessing the effects of respiratory interventions such as medications and physiotherapy techniques. The system also has potential in the pre- and post-operative assessment of thoracic surgical interventions to treat lung conditions including lung cancer, as well as structural abnormalities such as pectus excavatum and paralysed hemidiaphragm. In particular, there is potential for the use of the tool as a tele-assessment device, particularly as there is potential to use the system whilst the patient is seated, or standing.
An advantage of using data from four Kinect sensors is that this system is able to accurately assess chest wall motion even in moving subjects. This is potentially very useful for assessing changes in respiratory physiology that occur on exercise, such as dynamic hyperinflation. This is in contrast to previously described chest wall motion analysis systems that rely on data from one viewpoint [18]. This again allows for more effective use of the system for tele-assessment.
It was previously discussed in this paper that an advantage of using this system is the ability to measure regional motion. Although this was not fully evaluated in this initial testing, the system was designed with this in mind. It is possible, when the point clouds are exported from the bespoke software, to be able to split the point cloud in any way necessary. It is then possible to evaluate the volume of each individual component created, allowing for regional motion analysis.
The system is currently in a prototype stage and shows promising results. However, further development is required to create a fully integrated software which does not involve the use of Geomagic Studio. The hardware itself must also be adapted to be more suitable for the clinical environment (e.g. alleviation of excess wiring, producing single device). With a new version of the Microsoft Kinect being released, and future SDK updates, there is also scope for improvement in these results through the enhancement of the technology itself.
In terms of the accuracy of the system, the probability distribution of the error on the Kinect-based system highlighted a systematic, albeit small, underprediction of static volume. It appears from this testing that the system is accurate when measuring static objects. For the purposes of this testing, it was decided that this underprediction was small enough to conduct dynamic capture successfully. However, in future iterations of the prototype, these data will require further investigation. With further work on the calibration process, and refinement of the algorithms used for smoothing, this error may be reduced.
Full system testing with 13 healthy volunteers and nine CF was conducted to assess dynamic capture capabilities. Estimates of volume change showed a high correlation with that of spirometry in both groups. Through the total least squares regression, it is shown that Kinect-based measurements and spirometry are in good agreement, with a tendency for Kinect-based measurements to underpredict that of spirometry. This is further in agreement with the static analysis.
There were significant differences between the groups, due to airflow obstruction, characteristic of CF lung disease. This may be down to the original design stage of the system in which testing was performed on healthy volunteers. At this point, there may not have been enough focus on the movement of the diaphragm, and this is something that will need to be addressed in future versions.
Conclusion
In summary, these initial results suggest that the novel Kinect-based system is an accurate and affordable method of rapidly measuring chest wall motion. Some slight underprediction has been shown in the results, but in general, a promising correlation with spirometry in static objects, healthy volunteers and CF patients has been demonstrated, providing grounds for further work.
Therefore, the research team will be developing the system further, particularly through use of the new version of the Kinect and SDK, and evaluation of the system for testing regional motion and tele-assessments. The team will also be evaluating it in comparison with optoelectronic plethysmography. There will also be further work in CF as well as other respiratory conditions.
Acknowledgments
The authors wish to thank all those who were involved with the study, particularly the participants who gave their time so willingly. The authors would also like to thank all those involved with the study at each of the institutions involved, particularly Professor Christopher James at the University of Warwick who helped with the transitioning of the project into the Institute of Digital Healthcare.
Biographies
Dr. James M. Harte
is presently manager of the Interacoustics Research Unit, located at the Technical University of Denmark. Interacoustics is a world leading diagnostics solutions provider in the field of hearing and balance assessment. James Harte’s research interests lie in biomedical engineering, physiological modelling and digital signal processing. He has held various academic positions in the University of Warwick, Technical University of Denmark and the University of Southampton and has published many peer-reviewed journal and conference articles. His research has received significant funding from the Engineering and Physical Sciences Research Council (EPSRC), Technology Strategy Board (TSB), Oticon Foundation and the Danish Research council.
Dr. Christopher K. Golby
is a Research Fellow at the Institute of Digital Healthcare, based in WMG, University of Warwick, UK. Christopher has a B.Sc. in Computer Science from Keele University, UK, and a Ph.D. in Engineering (software engineering for health care) from Warwick University. His research interests are in the area of e-Health technologies, including the development of prototype devices. He has worked with different healthcare systems, developing prototype tools in stroke rehabilitation, post-operative breast cancer care and respiratory illness.
Johanna Acosta
is a computer scientist who obtained a Master Degree with distinction in Programme and Project Management from the University of Warwick in 2011. As MSc student, she participated in a project to develop an RFID Kanban-based system for the University Hospitals Coventry and Warwickshire NHS Trust to improve supply of medical materials at ward level. Consequently, she was appointed as a Knowledge Transfer Partnership (KTP) Associate to develop an integrated system for examination of respiratory function using Microsoft Kinect. She awarded the Society Thoracic Medal for the best presentation of a research paper called “Exercise Improvement After Pectus Excavatum Repair Is Not Related To Chest Wall Function” in the Society of Cardiothoracic Surgery (SCTS) Annual Meeting 2012 in Manchester, UK. Her research is centred on development of healthcare technologies that help to improve patience experience and service quality.
Dr. Edward F. Nash
is a Consultant Respiratory Physician in the West Midlands Adult Cystic Fibrosis Centre, Heart of England NHS Foundation Trust and Honorary Senior Clinical Lecturer at University of Birmingham, UK. His interests include clinical research in a variety of areas related to the care of adults with CF.
Ercihan Kiraci
is a project engineer of the Product Evaluation Technologies (PET) research group based in WMG, the University of Warwick. He is an experienced engineer and researcher with expertise in non-contact measurement and product quality assessment and continues improvement. He has successfully demonstrated these skills in a number of industries, in particular automotive industry and Premium Vehicle Customer Interface Technologies (PVCIT) project. He is currently working under Catapult High Value Manufacturing project. At WMG, he is also a lecturer in Dimensional Measurement and Management Module and supervises M.Sc. projects.
Prof. Mark A. Williams
is leader of the Product Evaluation Technologies (PET) research group based in WMG, the University of Warwick. His current research interests include Advanced Metrology Systems, Visualisation and the Human Machine Interface (HMI), publishing over 100 papers at international conferences and academic journals. He is currently lead investigator on a diverse portfolio of research projects funded by the Technology Strategy Board (TSB), European Regional Development Fund (ERDF) and EPSRC within the Automotive, Healthcare and Defence industries. Mark is also active within local government and is a Chartered Engineer (CEng) and Fellow of the Institution of Mechanical Engineers (FIMechE).
Prof. Theodoros N. Arvanitis
received his RT (B.Sc.) degree (Medical Radiological Technology) in 1990, from the Technological Educational Institute of Athens, Greece, and his D.Phil. (Biomedical Engineering) in 1997 from the University of Sussex, UK. He is currently a Professor of e-Health Innovation at The Institute of Digital Healthcare, WMG, University of Warwick, UK. Previously, he worked as a laboratory director/research fellow at the Trafford Centre for Medical Research (1995) and a full-time Lecturer in the School of Cognitive and Computer Sciences (1995–1998). In 1998, he joined the School of Electronic, Electrical and Computer Engineering, University of Birmingham, UK, as a full-time Lecturer (1998–2003), Senior Lecturer (2003–2007) and subsequently Reader in Biomedical Informatics, Signals and Systems (2007–2013). His research interests span the areas of biomedical engineering, neuroimaging and health informatics. Professor Arvanitis has a substantial academic publication record and has received research funding from national (UK), European and international governmental funding agencies, charities and industry. Theo has been involved in the UK (BSI) and EU (CEN) standardisation work, whilst he has contributed work in various working groups of the US National Cancer Institute Cancer Biomedical Informatics Grid (CaBIG) effort. He is a Charted Engineer (CEng), Fellow of the Royal Society of Medicine (FRSM), Fellow of the Society for the Internet in Medicine (FSIM) and a member of various scientific and professional groups (IET, IEEE, AMIA, NYAS, RCS, ISMRM).
Mr. Babu Naidu
is currently a Clinical Scientist in Cardiovascular and Respiratory Sciences at the University of Birmingham, UK. He is also an honorary Consultant Thoracic Surgeon in the Heart of England NHS foundation Trust, the largest thoracic surgical centre in England. Babu Naidu is a Regional Lead of the Thoracic Surgery Research Collaborative and a National Advisor on Enhanced Recovery in Thoracic Surgery to the UK Department of Health. He has worked as a surgeon since 1995 and has a multitude of journal papers and research grants in this field.
Footnotes
James M. Harte and Christopher K. Golby have contributed equally to this work.
References
- 1.Aliverti A, Pedotti A (2002) Opto-electronic plethysmography. In: Mechanics of breathing. Springer, New York, pp 47–59
- 2.Alnowami M et al (2012) A quantitative assessment of using the Kinect for Xbox360 for respiratory surface motion tracking. In: SPIE medical imaging. International society for optics and photonics
- 3.Aoki H et al (2012) Non-contact respiration measurement using structured light 3-d sensor. In: 2012 Proceedings of SICE annual conference (SICE). IEEE
- 4.Cala S, et al. Chest wall and lung volume estimation by optical reflectance motion analysis. J Appl Physiol. 1996;81(6):2680–2689. doi: 10.1152/jappl.1996.81.6.2680. [DOI] [PubMed] [Google Scholar]
- 5.de Boer W, Lasenby J, Cameron J, Wareham R, Ahmed S, Roach C, Hills W, Iles R (2010) A zero-contact non-invasive method for pulmonary function testing. In: Proceedings of the British machine vision conference, pp 85.81–85.12
- 6.Dutta T. Evaluation of the Kinect™ sensor for 3-D kinematic measurement in the workplace. Appl Ergon. 2012;43(4):645–649. doi: 10.1016/j.apergo.2011.09.011. [DOI] [PubMed] [Google Scholar]
- 7.Eastwood-Sutherland C et al (2014) Non-contact respiratory monitoring in neonates. In 2014 7th biomedical engineering international conference (BMEiCON). IEEE
- 8.Fiz JA, et al. Spirometry and maximal respiratory pressures in patients with facial paralysis. CHEST J. 1993;103(1):170–173. doi: 10.1378/chest.103.1.170. [DOI] [PubMed] [Google Scholar]
- 9.Fritsch FN, Carlson RE. Monotone piecewise cubic interpolation. SIAM J Numer Anal. 1980;17(2):238–246. doi: 10.1137/0717021. [DOI] [Google Scholar]
- 10.Green S, Miles R (2007) The burden of disease and illness in the UK: a preliminary assessment to inform the development of UK Health Research and Development Priorities, Department of Health
- 11.Khoshelham K, Elberink SO. Accuracy and resolution of Kinect depth data for indoor mapping applications. Sensors. 2012;12(2):1437–1454. doi: 10.3390/s120201437. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Krowka MJ, Enright PL, Rodarte JR, Hyatt RE. Effect of effort on measurement of forced expiratory volume in one second. Am Rev Respir Dis. 1987;136:829–833. doi: 10.1164/ajrccm/136.4.829. [DOI] [PubMed] [Google Scholar]
- 13.Levine S, et al. Use of a triaxial magnetometer for respiratory measurements. J Appl Physiol. 1991;70(5):2311–2321. doi: 10.1152/jappl.1991.70.5.2311. [DOI] [PubMed] [Google Scholar]
- 14.Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, Crapo R, Enright P, van der Grinten CPM, Gustafsson P, Jensen R, Johnson DC, MacIntyre N, McKay R, Navajas D, Pedersen OF, Pellegrino R, Viegi G, Wanger J. ATS/ERS task force: standardisation of lung testing, general considerations for lung function testing. Eur Respir J. 2005;136(4):319–338. doi: 10.1183/09031936.05.00034805. [DOI] [Google Scholar]
- 15.Markovsky I, Van Huffel S. Overview of total least-squares methods. Sig Process. 2007;87(10):2283–2302. doi: 10.1016/j.sigpro.2007.04.004. [DOI] [Google Scholar]
- 16.Matsuoka S, et al. Quantitative CT assessment of chronic obstructive pulmonary disease 1. Radiographics. 2010;30(1):55–66. doi: 10.1148/rg.301095110. [DOI] [PubMed] [Google Scholar]
- 17.Mead J, et al. Pulmonary ventilation measured from body surface movements. Science. 1967;156(3780):1383–1384. doi: 10.1126/science.156.3780.1383. [DOI] [PubMed] [Google Scholar]
- 18.MedGadget (2013) Xerox developing tech for non-contact collection of vital signs. In: Medgadget. http://www.medgadget.com/2013/03/xerox-developing-tech-for-non-contact-collection-of-patient-vitals.html. Accessed 06 Jan 2016.
- 19.Mirtich B. Fast and accurate computation of polyhedral mass properties. J Graph Tools. 1996;1(2):31–50. doi: 10.1080/10867651.1996.10487458. [DOI] [Google Scholar]
- 20.O’Sullivan BP, Freedman S. Cystic fibrosis. Lancet. 2009;373(9678):1891–1904. doi: 10.1016/S0140-6736(09)60327-5. [DOI] [PubMed] [Google Scholar]
- 21.O’Donnell DE. Assessment of bronchodilator efficacy in symptomatic COPDIs spirometry useful? CHEST J. 2000;117(2_suppl):42S–47S. doi: 10.1378/chest.117.2_suppl.42S. [DOI] [PubMed] [Google Scholar]
- 22.White J. High frequency techniques: an introduction to RF and microwave engineering. Hoboken: Wiley; 2004. [Google Scholar]
- 23.Xia J, Siochi RA. A real-time respiratory motion monitoring system using KINECT: proof of concept. Med Phys. 2012;39(5):2682–2685. doi: 10.1118/1.4704644. [DOI] [PubMed] [Google Scholar]