Summary
Microbial water quality monitoring for the presence of faecal indicator bacteria (FIB) is a mandatory activity in many countries and is key in public health protection. Despite technological advances and a need for methodological improvements, chromogenic and fluorogenic enzymatic techniques remain the mainstays of water quality monitoring for both public health agencies and regulated utilities. We demonstrated that bioelectroanalytical approaches to FIB enumeration are possible and can be achieved using commercially available enzyme‐specific resorufin glycosides, although these are expensive, not widely available or designed for purpose. Following this, we designed two naphthoquinone glycosides which performed better, achieving Escherichia coli detection in the range 5.0 × 102 to 5.0 × 105 CFU ml−1 22–54% quicker than commercially available resorufin glycosides. The molecular design of the naphthoquinone glycosides requires fewer synthetic steps allowing them to be produced for as little as US$50 per kg. Tests with environmental samples demonstrated the low tendency for abiotic interference and that, despite specificity being maintained between β‐glucuronidase and β‐galactosidase, accurate enumeration of E. coli in environmental samples necessitates development of a selective medium. In comparison to a commercially available detection method, which has U.S. Environmental Protection Agency (EPA) approval, our approach performed better at high organism concentrations, detecting 500 organisms in 9 h compared with 13.5 h for the commercial method. Bioelectroanalytical detection is comparable to current approved methods and with further development could result in improved detection times. A recent trend for low‐cost open‐source hardware means that automated, potentiostatically controlled E. coli detection systems could be constructed for less than US$100 per channel.
Introduction
There is a link between exposure to human waste and the transmission of infectious diseases like cholera, typhoid fever, shigellosis and acute gastroenteritis (Lee et al., 2002; Ashbolt, 2004; U.S. EPA, 2012). Considerable effort is expended to detect, quantify and track faecal contamination of water to ensure drinking water quality meets sanitary guidelines (Cabral, 2010; Nshimyimana et al., 2014) and that recreational waters remain safe for primary and secondary contact activities such as swimming and boating. Usually, faecal indicator bacteria (FIB) are monitored as a proxy for pathogenic faecal contamination of water in conjunction with risk assessment techniques that correlate the frequency of a specific hazard (such as the occurrence of gastroenteritis) with a given level of FIB exposure.
FIB monitoring commonly involves enumerating total coliforms (TC) and faecal coliforms (FC), also referred to as thermotolerant coliforms (TTC), which includes Escherichia coli. Faecal enterococci are the preferred indicator for marine waters in the United States (U.S. EPA, 2012) and have more recently been listed as admissible for freshwater alongside E. coli (U.S. EPA, 2012). Alternative indicators, such as Clostridium perfringens and Bacillus spp., have been used as FIB in an attempt to definitively identify human faecal contamination of water (Kemmochi et al., 2007; Douterelo et al., 2014). In addition, cultivation‐independent methods targeting the order Bacteroidales have used 16S rRNA‐based genetic markers with faecal host‐associated distribution and geographic stability (Seurinck et al., 2005; Wuertz et al., 2011; Liang et al., 2012; Reischer et al., 2013; Odagiri et al., 2015). Similarly, immunological tests like ELISA exist, but have not been routinely applied (Wang et al., 2007).
Regulatory resistance to emerging FIB enumeration technologies remains, even though the range of indicator proxies and potential enumeration technologies has improved (Lazcka et al. 2007). Consequently, few regulatory guidelines exist prescribing standard protocols for molecular approaches to FIB monitoring. An exception is the U.S. EPA guidelines permitting the application of qPCR for enumerating Enterococcus faecalis in recreational water quality monitoring. Despite drawbacks, such as their indirect nature, time required for growth and assumptions underlying FIB selection, FIB still remain important tools in assessing microbial water quality. Along with traditional culture techniques, enzymatic enumeration of FIB is favoured for evaluating microbial water quality under most regulated jurisdictions (Douterelo et al., 2014; TMR, 2015).
Bioelectroanalytical approaches are emerging as a method of scrutinizing microbial systems (Seviour et al., 2015). Although specialist microbes can reduce electrodes directly, it is a near universal property of microbes to indirectly reduce an electrode via soluble redox mediators (Rabaey et al., 2005, 2007; Mathis et al., 2008). Although more commonly applied for biocatalysis (Rabaey and Rozendal, 2010) or electricity production (Logan and Regan, 2006), cell–electrode interactions can be used to qualitatively or quantitatively describe features of living microbial systems by monitoring the redox processes they mediate. Furthermore, the advent of low cost, open source and electrochemical components, such as the CheapStat potentiostat, has enabled inexpensive hand‐held, multi‐analyte detection systems to be developed, reportedly for as little as US$30 (Rowe et al., 2011; Nemiroski et al., 2014). With regards to analysis of environmental samples, there are additional benefits to bioelectroanalytics, including increased durability and less interference from deeply coloured or fluorescent matrices (Nijak et al. 2012; Hata et al. 2014; Kapoor et al. 2015). Furthermore, bioelectronalytical systems are inherently scalable, a property which has been extensively discussed in the literature in the context of microbial fuel cells (Logan, 2009a,b).
Enzymatic detection of FIB usually involves a β‐glucuronidase‐ or β‐galctosidase‐specific monosaccharide conjugated to a chromogenic or fluorogenic compound (aglycon). When cleaved by the corresponding enzyme, the aglycon is released into the medium and detected optically and the presence of the target organism is inferred (Manafi et al., 1991; Nelis and Van Poucke, 2000). Because E. coli is not strongly electrogenic, incorporating redox‐active aglycons into the enzymatic substrate would enable targeted E. coli bioelectroanalysis and would represent a natural extension to the chemistry underlying optical techniques. Such an approach would find utility in bioelectroanalytical approaches to FIB enumeration while simultaneously satisfying industrial and regulatory preference for enzymatic techniques. Previous electrochemical attempts at E. coli enumeration have combined techniques to separate the medium from the growing culture for analysis in a separate electrochemical procedure (Perez et al., 2001). Perez et al. (2001) achieved electrochemical FIB detection using 4‐aminophenol‐β‐D‐galactopyranoside. However, with their experimental set‐up, pre‐filtration and flow injection analysis (FIA) were required. We contend that specifically designed electrochemically active glycosides, which mediate the reversible transfer of electrons between E. coli and an electrode, are key to bioelectranalytical approaches (Fig. 1). By exploiting the natural redox cycling capabilities of target microbes, bioelectrochemical detection of E. coli can be achieved without sample preparation in a single pot analysis for the majority of environmental samples.
Figure 1.

Mechanism of bioelectroanalytical detection technique and the detection compounds used here (from top to bottom): resorufin‐β‐d‐galactopyranoiside; resorufin‐β‐d‐glucuronide and novel peracetylated methyl‐esters of 2‐hydroxy‐1,4‐napthoquinone‐β‐d‐galactopyranoside and 2‐hydroxy‐1,4‐napthoquinone‐β‐d‐glucuronide.
Resorufin can be reversibly reduced and oxidized by microorganisms and is used as a redox indicator in anaerobic microbiology (Candeias et al., 1998). Resorufin glycosides make excellent candidates for bioelectroanalytical E. coli enumeration, although this has not been previously demonstrated. Resorufin glycosides are commercially available but because resorufin is not available in bulk quantities, it is expensive as are its derivatives (Magro et al., 2014). The aim of this study is to demonstrate that the rational design of redox‐active glycosides can be applied to achieve performance benefits in electroanalytical enumeration of E. coli over resorufin glycosides. We present two novel naphthoquinone glycosides specifically designed for the bioelectroanalytical enumeration of FIB. The candidate aglycons are analogues of the redox component of menaquinone, a naturally occurring, membrane‐bound charge carrier in E. coli that can be readily conjugated to pyranose rings (Unden, 1988; Qiao et al., 2008). Here, we demonstrate the feasibility of the bioelectranalytcial approach using resorufin glycosides, we expound on the development of naphthoquinone glycosides, demonstrate their improved performance and finally we explore the potential utility of redox‐active glycosides in the eventual bioelectroanalytical detection of FIB. To our knowledge, this is the first study that considers FIB‐specific glycosides that have been specifically selected and rationally designed for bioelectroanalytical enumeration.
Materials and methods
Microorganism and growth conditions
Escherichia coli strain BL21 was selected to represent a FIB of human origin that contains the uidA and lacZ genes, i.e. a TTC. Aeromonas hydrophila (DSM 6173), on the other hand, was chosen as an environmental organism that has the lacZ gene but which is not considered a coliform. All organisms were cultured aerobically for 20 h. E. coli was grown in LB broth (Lennox) (37°C), and A. hydrophila in tryptone soy broth at 30°C in shaking incubator. Overnight cultures were diluted in sterile‐modified M9 medium [Na2HPO4·7H2O (6.8 g l−1); KH2PO4 (3.2 g l−1); NH4Cl (1 g l−1); NaCl (0.5 g l−1)]; 1 ml l−1 vitamin stock solution, 0.2 ml l−1 micronutrients stock solution (Mathews et al., 2010); tryptone (12.5 g l−1); HEPES (6 g l−1). The carbon source was either D‐galacturonic acid or D‐glucuronic acid (3.75 g l−1) to achieve final inoculum densities in the range of 1.0 × 100, 5.0 × 100, 5.0 × 101, 5.0 × 102, 5.0 × 103, 5.0 × 104 and 5.0 × 105 CFU ml−1.
Mediators and glycosides
Resorufin (TCI‐Europe, Zwijndrecht, Belgium), 2‐hydroxy‐1,4‐naphthoquinone (2H14NQ; TCI‐Europe), 5‐hydroxy‐1,4‐naphthoquinone (5H14NQ; TCI‐Europe) and 5,8‐dihydroxy‐1,4‐naphthoquinone (58H14NQ; Sigma‐Aldrich, Singapore) were obtained commercially as were resorufin glycosides (resorufin‐β‐d‐galactopyranoside (ResGal) and resorufin‐β‐d‐glucuronide (ResGlu); Sigma‐Aldrich), whereas 2‐hydroxy‐1,4‐naphthoquinone‐β‐d‐galactopyranoside (2H14NQGal) and 2‐hydroxy‐1,4‐naphthoquinone‐β‐d‐glucuronide (2H14NQGlu) were custom synthesized by Sussex Research (Ottawa, ON, Canada). All unconjugated mediators and conjugated glycosides were added to achieve a final concentration of 50 μM. Stock concentrations (1 mM) of 2H14NQ and 5H14NQ were prepared by first dissolving in 10 μl of 2 M NaOH and then making up to volume in ultrapure water (Veolia, Paris, France), whereas 58H14NQ (1 mM) was prepared in 1:1 DMSO:ultrapure water aliquots. Both stock solutions were stored at −20°C until needed. Glycosides were maintained as 1 mM stock solutions in ultrapure water and stored at 5°C.
Environmental samples
Three environmental samples were collected from a private lake on Nayang Technological University campus on March 10th (Day 1), 17th (Day 2) and 25th (Day 3), 2016. A 1000 ml grab sample was collected from the bank. The water temperature was consistently 33°C, whereas the air temperature ranged between 31.5 and 36.5°C.
The sample taken on Day 1 was filter sterilized and stored at 4°C until use, at which point it was warmed and spiked with 5 × 103 CFU ml−1 of E. coli BL21 prior to duplicate bioelectranalytical analysis at 37°C with ResGlu to detect E. coli only. The samples taken on Day 2 were used for TC and E. coli enumeration, and were analysed immediately after transport to the laboratory using both the bioelectroanalytical approach (in duplicate) and Colifast® Quanti‐Tray 2000® according to the manufacturer's instructions. On Day 3, the bioelectroanalytical approach was used to enumerate TC only. Incubation for both methods was at 37°C. For the enumeration using bioelectroanalytical approach, autoclaved 10× modified M9 (as described above) was added to environmental water samples to achieve a working volume of 10 ml and a glycoside concentration of 50 μM. TC were estimated from non‐weighted exponential regression of the data presented in Table 1 for ResGal at 37°C. The standard curve, the regression coefficient and the equation are presented in Fig. S1. As the detection times observed for E. coli were out of the range of the standard curve, no estimation of CFUs in environmental samples was made.
Table 1.
Mean enumeration times for Escherichia coli with inoculum sizes of 5.0 × 102, 5.0 × 103, 5.0 × 104 and 5.0 × 105 CFU ml−1 achieved with two commercially available glycosides (ResGlu and ResGal) and two compounds described for the first time in this study (2H14NQGal and 2H14NQGlu). Detection time is calculated from chronoamperometry performed in ECs and expressed as the mean detection time in minutes. n = 3 except * where n = 1
| Cell count (CFU ml−1) | E. coli enumeration time (h) | |||||
|---|---|---|---|---|---|---|
| ResGal (30°C) | 2H14NQGlu (30°C) | ResGal (37°C) | ResGlu (37°C) | 2H14NQGal (37°C) | 2H14NQGlu (37°C) | |
| 5 × 102 | 8.6 (2.03)a | 13.9* | 12.2 (3.13) | 7.2 (0.50) | 6.9 (0.23) | 5.7 (0.70) |
| 5 × 103 | 6.2 (1.38) | 9.6* | 7.6 (0.38) | 6.6 (0.88) | 5.9 (0.22) | 3.1 (0.53) |
| 5 × 104 | 3.6 (0.52) | n/db | 5.1 (0.92) | 5.2 (0.56) | 2.4 (0.29) | 2.4 (0.20) |
| 5 × 105 | 2.5 (0.93) | 2.8* | 4.3 (0.92) | 4.3 (0.32) | 1.8 (0.14) | 2.0 (0.35) |
One standard deviation in parentheses.
Not determined.
Construction of electrochemical cells
Conical ECs with a working volume of 10 ml and equipped with carbon‐felt electrodes were used. The ECs were fitted with a customized Teflon cap to accommodate a three‐electrode configuration and gas fittings. The carbon‐felt working electrodes (1 × 1 × 0.318 cm; VWR, Singapore), pre‐treated for 24 h in 1 M HCl and stored in deionized water until use, were connected to titanium wire (0.25 mm diameter; Sigma‐Aldrich) using a nylon nut and bolt. A coiled titanium wire counter electrode and an Ag/AgCl reference electrode (78 mm × 6 mm outer diameter; Biologic, Claix, France) were connected using a 1 M KCl in 1.5% agar salt bridge ending in a 4 mm glass frit (CoralPor™; Biologic). The entire assembly except the Ag/AgCl reference electrode and 1.5% 1 M KCl salt bridge was autoclaved (121°C for 15 min). Nine millilitres of degassed modified M9 medium were added to the ECs and supplemented with mediator to a final concentration of 50 μM and inoculated with the test organism to achieve a final working volume of 10 ml. The headspace was flushed continuously with sterile, humidified N2 to maintain a moist anaerobic atmosphere. The ECs were operated at 30, 37 or 44.5°C in the dark to prevent photo‐bleaching of the light‐sensitive mediators. Stirring was maintained continuously with a magnetic stirrer.
Electrochemical data acquisition
Electrochemical cells were connected to a multichannel potentiostat (Biologic), and cyclic voltammetry (CV) and differential pulse voltammetry (DPV) were performed immediately but without data collection for diagnostic purposes and to discharge the electrode. Data were recorded by EC‐Lab software (Biologic). Chronoamperometry (CA) was applied for up to 24 h to detect the activity of E. coli via redox mediators. A detection event was calculated when the slope of the line exceeded four times the standard deviation of the baseline and detection times are expressed as the mean of triplicate analyses unless stated otherwise. A sample raw electrochemical output can be seen in Fig. S2.
The parameters for the electrochemical techniques were chosen accordingly: CA: Eapplied = 200 mV versus Ag/AgCl; CV: equilibrium time of 5 s; scan rate of 10 mV/s; Ei = −700 mV versus Ag/AgCl; Ef = 500 mV versus Ag/AgCl; DPV: Ei = −700 mV versus Ag/AgCl; Ef = 500 mV versus Ag/AgCl; pulse height of 50 mV; pulse width of 200 ms; step height of 2 mV; step time of 400 ms and scan rate of 5 mV/s.
Results and discussion
Bioelectroanalytical enumeration with resorufin glycosides
Enumeration of E. coli was achieved with resorufin‐β‐d‐galactopyranoside (ResGal) at 37°C in ECs in 4–12 h for inoculum sizes ranging between 5.0 × 105 and 5.0 × 102 CFU ml−1 with good reproducibility across triplicate analysis (Table 1). For resorufin‐β‐d‐glucuronide (ResGlu), enumeration time was achieved more quickly at 37°C (between 4.3 and 7.2 h) for a similar inoculum size (Table 1). Abiotic controls show the absence of current (Fig. S3) and the detection time seems to correlate well with the time when cultures enter log phase (Fig. S4).
Electrochemical enumeration of E. coli using chromogenic glycosides has been reported previously (Perez et al., 2001), where an enumeration time of 6.5 h was documented for an inoculum size of 5.0 × 103 CFU ml−1 E. coli in samples not induced with isopropyl‐β‐d‐1‐thiogalactopyranoside. This was faster than the enumeration time of 7.6 h reported here for ResGal at 37°C (Table 1), but comparable with the time of 6.6 h that was achieved for ResGlu at 37°C (Fig. 2A).
Figure 2.

A. Detection time for different inoculum sizes of Escherichia coli incubated at 37°C with a commercially available detection compound, ResGlu, and also the novel 2H14NQGlu designed in this study (error bars indicate standard deviation, n = 3).
B. Duplicate current profiles of candidate redox‐active aglycons in ECs. Note that 2H14NQ outperforms the other candidate compounds (5H14NQ and 58H14NQ) in terms of onset time and maximum signal intensity, which is sustained over longer periods, despite only minor differences in structure. Note the absence of significant current in the controls incubated with E. coli only.
Enumeration times for colorimetric β‐glucuronidase assays have been reported to be as low as 0.5 h with high sensitivity (1 CFU ml−1), but these techniques require several pre‐treatment steps, such as filtration, incubation and staining in addition to specialist detection equipment. The time saving is offset by a concomitant increase in required manpower or complexity, effectively making these approaches labour intensive and laboratory based (Fiksdal et al., 1994; Nelis and Van Poucke, 2000; Heery et al., 2016). Whereas the sensitivity of these rapid techniques is undeniably high (detecting 4‐APGal in quantities as low as 45 ng l−1), the quantification step usually requires alkaline conditions to maximize the optical properties of the aglycon, further complicating the procedure although they have utility in field detection particularly when FIB concentrations are expected to be high. Despite these advances, the detection time for market‐leading, U.S. EPA approved enzymatic E. coli detection methods is in the range 15–28 h for drinking water applications (TMR, 2015, Colifast, 2016, Idexx, 2016).
Similarly, the method reported by Perez et al. (2001) required a filtration step and, in addition, a combination of FIA and a potentiostatic technique to achieve electrochemical detection of 4‐aminophenyl (4‐AP) that had been cleaved by E. coli from 4‐aminophenyl‐β‐d‐galactopyranoside (4‐APGal) glycoconjugate. The complexity and low throughput of this method make it less amenable to automated or remote applications when compared with the documented durability and low cost of bioelectroanalytical methods that do not require pre‐treatment and which could be combined with automated sampling devices (Logan, 2009a,b; Nemiroski et al., 2014). Furthermore, 4‐APGal and 4‐nitrophenol‐β‐d‐glucuronidase (4‐NPglu) have been designed for colorimetric and fluorimetric assays. Although they have some electrochemical activity, their redox chemistry is not fully reversible by E. coli under physiological conditions the redox kinetics are slow. Because E. coli is not strongly electrogenic in character (Roller et al., 1984; Choi et al., 2003; Wang et al., 2013), the utility of non‐reversible chromogenic substrates in bioelectroanalytical methods is restricted to techniques, such as voltammetry, that do not rely on or fully exploit microbially mediated electron transfer that is achievable in bioelectroanalytical systems.
Low‐cost bioelectroanalytical methods would be amenable to automation and portable applications. The approach presented here is unsophisticated and unoptimized using low‐cost carbon‐felt electrodes to directly measure an actively respiring culture without pre‐treatment. The scope for refinement and to improve enumeration times remains considerable assuming that the cost of the electrode can be offset by the design of cost‐effective glycosides.
Experimental rationale for napthoquinone glycosides
Although ResGal can be used to quantify E. coli, it is not available in bulk and is, therefore, expensive (Biosynth, 2012, Magro et al., 2014). Our source cost more than US$ 3000 per gram and export records suggest prices in excess of US$ 1000 per kg are typical for resorufin (Zauba, 2016). Naphthoquinones are cheap, readily available redox‐active molecules, which are compatible with E. coli (Park and Zeikus, 2000; Rau et al., 2002). Because naphthoquinones possesses a hydroxyl group, they can be readily conjugated to pyranose rings (Magro et al., 2014). Their ability to mediate electron transfer between E. coli and an electrode in ECs shows evidence of their potential as candidate redox‐active E. coli detection compounds (Fig. 2B). Furthermore, they can be acquired in bulk for as little as US$ 50 per kg making the raw materials 20 times cheaper than resorufin (Zauba, 2016).
Glycoside synthesis requires that the carboxyl and hydroxyl groups of the sugar be protected during synthesis with a methyl and an acetyl group, respectively. Deprotection yields can be as low as 5% (Shen et al., 2003; Biosynth, 2012). Peracetylated methyl‐esters conjugated to cheap redox mediators make good candidate glycosides and may offset the cost of the electrode in bioelectroanalytical E. coli detection. Synthesis of the resulting glycoside would require no deprotection step, resulting in increased yields and therefore reduced production costs compared with fully deprotected glycosides. Because E. coli cells express native esterase, they are biochemically equipped to convert the peracetylated methyl‐esters into the acid form in situ if this is necessary for subsequent electrochemical activity (Antonczak et al., 2009). If the deprotection step, with yields between 5% and 50%, is omitted, then naphthoquinone peracatelyated methyl‐ester glycosides could be as much as 40–800 times cheaper than resorufin glycosides (Biosynth, 2012). Magro et al. (2014) estimate the cost of chromogenic glycosides to be in the range US$ 500–1000 per kg. The naphthoquinones are in the same price range as chromogenic substrates leading to a cost estimate for naphthoquinone glycosides in the range US$ 50–500 kg. On a per test basis this could be as little as 0.005 cents assuming on current concentrations of 50 μM (Biosynth, 2012, Magro et al., 2014; Zauba, 2016).
Of three candidate naphthoquinones: 2‐hydroxy‐1,4‐naphthoquinone (2H14NQ), 5‐hydroxy‐1,4‐naphthoquinone (5H14NQ) and 5,8‐dihydroxy‐1,4‐naphthoquinone (58H14NQ) tested, 2H14NQ outperformed both 5H14NQ and 58H14NQ in ECs (Fig. 2B). The maximum recorded current in ECs incubated with 5.0 × 105 CFU ml−1 E. coli and using 2H14NQ as the redox mediator was ~100 μA (Fig. 2B), which was ~14 times the maximum current achieved with 5H14NQ (~7 μA) and ~25 times that observed with 58H14NQ (~4 μA). Accordingly, 2H14NQ was selected over 5H14NQ and 58H14NQ for conjugation to both galacturonic acid and glucuronic acid on the basis of its sound electrochemical performance and its potential to enable targeted bioelectroanalytical enumeration of E. coli. It has been previously reported that electron transfer kinetics are dependent on the stability of intramolecular hydrogen bonds of the various naphthoquinones (Frontana and González, 2005). There are two types of hydrogen bond systems in the naphthoquinones studied here: 2H14NQ has an α‐hydroxy system (where the interaction between the carbonyl group and the hydroxy group is supported on the same ring), whereas 5H14NQ and 58H14NQ have β‐hydroxy systems (where the intramolecular interaction occurs between adjacent rings). Electron transfer kinetics is affected by the β‐hydroxy systems and likely underlies the poor bioelectrochemical performance of 5H14NQ and 58H14NQ. The suspected incompatibility of β‐hydroxy systems in bioelectroanalytical E. coli enumeration is interesting and offers valuable insights into the future selection and design of candidate glycosides (Frontana and González, 2005).
Performance of novel glycosides
Both 2H14NQGal and 2H14NQGlu can be used to enumerate E. coli, confirming sound rationale in molecule selection. The enumeration time for E. coli using 2H14NQGlu ranges between 2 and 5.7 h (±7.7–16.6%, n = 3) for inoculum sizes between 5.0 × 105 and 5.0 × 102 CFU ml−1 (Fig. 2A). The enumeration time achieved with 2H14NQGlu is quicker than that observed for ResGlu (4.3–7.2 h) (Table 1) and compares well with the enumeration time of ~6.5 h for 5.0 × 103 CFU ml−1 reported by Perez et al. (2001). However, the value reported by Perez et al. (2001) was for a galactopyranoside conjugate. The galactopyranoside equivalent (2H14NQGal) used here resulted in only a slightly slower enumeration time of ~6.9 h (±3.3%, n = 3) versus 6.5 h reported by Perez et al. (2001) for an equivalent inoculum size. In this study, the naphthoquinone glycosides achieved faster enumeration times than their resorufin counterparts, demonstrating that peracetelylated methyl‐ester glycosides offer both cost savings and performance benefits in the context of bioelectroanalytical E. coli detection. Variations in glycoconjugate performance have previously been attributed to transport phenomena, particularly cellular uptake (Magro et al., 2014). It may be, therefore, that the relatively small size of the naphthoquinone glycoconjugate compared with the resorufin glycosides allow for improved transport across the cell membrane and, hence, better performance.
Applicability of bioelectroanalytical approach to environmental TC detection
A remote biosensor would ideally have as few components as possible. We explored the possibility of omitting the heating component of a detection unit and if TC monitoring at ambient Singapore temperatures (~30°C) was achievable. The inoculum density of E. coli correlated well with the electrochemical response of ResGal at 30°C with delays in enumeration time of up to 60%. The enumeration times with 2H14NQGlu were more negatively impacted by temperature (Table 1). Even though detection at ambient temperatures is possible, faster detection times could be realized with incubation at 37 or 44.5°C.
Environmental samples may present a number of challenges for bioelectralanytical detection, particularly abiotic interference from redox‐active compounds and the possibility of interference from non‐target bacteria. To rule out the importance of abiotic interference, a filter sterilized lake water sample was spiked with a known number of E. coli and analysed using the bioelectroanalytical technique with 2HN14QGlu. The detection time was longer than expected and, by extrapolation from the data in Table 1 for ResGal at 37°C, a sample of 5 × 103 CFU ml−1 was estimated to contain < 500 CFU ml−1 (Table 2). This likely reflects the documented loss of viability observed in laboratory strains when introduced to environmental samples rather than electrochemical interference, which would be expected to speed up the analysis time (Bissonnette et al., 1975). The natural environmental concentration of compounds that could cause interference such as humics is typically over 100 times lower (< 0.22 mg l−1) than the redox mediator added here (≈ 24 mg l−1) and are not likely to interfere with the analysis (Vuorenmaa et al., 2006).
Table 2.
The analysis time for a filter sterilized environmental sample spiked with known number of Escherichia coli and a comparison of a colorimetric reference method with the proposed bioelectroanalytical approach. Total coliforms and E. coli were estimated in environmental water samples collected over three different days. Bile salts were included in one treatment on day 3 to demonstrate the need for selective media. For the reference method, n = 1, whereas for the bioelctroanalytical method n = 2, ±1 S.D in parentheses
| Method | Time to detection (h) | Total coliforms (CFU ml−1) | E. coli (CFU ml−1) |
|---|---|---|---|
| Day 1 | |||
| Bioelectroanalytical (spiked) | 8.46 (±0.54) | n/a | < 500b |
| Day 2 | |||
| Quanti‐Tray 2000 | 24 | 36.5 | 0.031 |
| Bioelectroanalytical | 7.43 (± 0.36) | 1030a (± 374) | < 500b |
| Day 3 | |||
| Quanti‐Tray 2000 | 24 | 92.8 | 0.26 |
| Bioelectroanalytical | 5.47 (± 0.10) | 3628a (± 620) | n/a |
| Bioelectroanalytical (with bile salts) | 7.27 (± 0) | < 500b (± 0) | n/a |
Detection time calculated from standard curve for ResGal (Fig. S4).
Where the detection time lays outside the calibration range, the value has been reported as < 500 CFU.
Enzymatic methods typically exploit the uidA gene, which encodes for the β‐glucuronidase enzyme, and is found in almost 98% of E. coli (Manafi et al., 1991; Martins et al., 1993). Similarly, the lacZ gene, which encodes for β‐galactosidase, is indicative of TC such as Klebsiella spp., Citrobacter spp. as well as E. coli (Bej et al., 1991; Manafi et al., 1991; Cabral, 2010). However, although conferring some degree of specificity, enzymatic tests in themselves are not definitively differential because a number of non‐target organisms are positive for either β‐glucuronidase or β‐galactosidase. In samples, particularly of environmental origin, that are expected to contain a variety of organisms, measures that prevent the growth of non‐target organisms accompany enzymatic tests, such as the inclusion of a selective medium and or elevated incubation temperatures (often 44.5°C). In addition, agents that help in the recovery of environmentally stressed cells such as pyruvate are added to the medium to improve performance (Kapuscinski and Mitchell 1981; Khaengraeng and Reed 2005). Organisms known to contribute to false positives in β‐glucuronidase and β‐galactosidase enzymatic tests include members from the genera: Staphylococus, Sphingomonas, Bacillus, Aerococcus, Entrobacter, Pseudomonas and, notably, Aeromonas which is often considered to be a nuisance in E. coli monitoring.
Tests with environmental samples confirmed interference from non‐target organisms. When compared with a standard reference method there was a 30‐fold overestimation of TC using our approach (1030 CFU ml−1) compared with the reference method (36.54 CFU ml−1) and an almost 200‐fold difference in E. coli estimation (5.84 CFU ml−1 versus 0.031 CFU ml−1). This points to the lack of selective agents in our approach compared with the reference method and that, although a degree of selectivity between galactosidase and glucuronidase positive organisms has been maintained, the accuracy is low due to inability to discriminate between target and non‐target organisms. To demonstrate the importance of a selective medium, we ran additional tests with and without bile salts. In this test, the reference method estimated TC as 92 CFU ml−1, whereas the bioelectroanlaytical approach with and without bile salts estimated the concentration as 3660 (±620) and 408 (±0) CFU ml−1, respectively (Table 2). Bile salts are commonly used to inhibit some non‐target microbes demonstrating the importance of selective agents for the accurate estimation of E. coli in environmental samples.
The ubiquitous aquatic organism A. hydrophila is a common false positive in water quality monitoring methods that rely on the enzymatic testing of β‐galactosidase activity because it is of telluric rather than enteric origin and thus not considered a coliform (Waltman et al., 1982; Cabral, 2010). A. hydrophila was detected electrochemically with ResGal at 30 and 37°C (Fig. 3) and because it is reportedly electrogenic (Pham et al., 2003; Logan, 2009a,b), its ability to interfere with bioelectroanalytical TC enumeration was investigated in more detail to further support the notion of biological interference over abiotic environmental effects.
Figure 3.

Detection time of a pure culture of Aeromonas hydrophila at 30 and 37°C as a function of inoculum size (filled and unfilled circles, respectively), and of a mixed culture of A. hydrophila and Escherichia coli at 30°C as a function of A. hydrophila inoculum size but with a constant E. coli inoculum of 5.0 × 105 CFU ml−1 (crossed empty circles). The glycoside is the commercially available ResGal. In addition, the detection time for a mixed culture of A. hydrophila and E. coli at 37°C (black triangles) and 44.5°C (crossed circles) as a function of A. hydrophila inoculum size is included. The E. coli inoculum is kept constant at 5.0 × 105 CFU ml−1 and the glycoside is the novel 2H14NQGlu (error bars represent 1 standard deviation, n = 3).
In a test with a constant inoculum size of E. coli (5.0 × 105 CFU ml−1), but with an increasing A. hydrophila concentration from 5.0 × 102 to 5.0 × 105 CFU ml−1 (equivalent to 1000:1, 100:1, 10:1 and 1:1), a decrease in enumeration time from ~7.5 to ~4.7 h was observed, indicating that A. hydrophila strongly influences the detection signal in a binary culture of E. coli and A. hydrophila at 30°C (Fig. 3). Furthermore, enumeration time is likely to be dominated by A. hydrophila as the combined incubation is similar to that of a pure culture of A hydrophila at 30°C (Fig. 3), indicating suppression of E. coli by A. hydrophila or that its electrogenic character is dominant (Pham et al., 2003). Although the inability to differentiate coliforms, such as Kleibsella spp. and Citrobacter spp., is expected for galactosidase enzymatic tests, the inability to differentiate between E. coli and A. hydrophila means that this technique will overestimate E. coli or TCs at ambient temperatures in the presence of hydromonads. Interference from hydromonads is a well‐documented shortcoming of enzymatic β‐galactosidase detection and four commercially available methods exhibited this problem upon testing (Manafi, 2000). Hydromonad interference can be mitigated by the incorporation of selective inhibitors into the medium but is beyond the scope of this contribution (Geissler et al., 2000; Manafi, 2000). However, in principle, bioelectroanalytical TC quantification is possible and could be achieved at ambient tropical temperatures with shortcomings similar to what would be expected using established enzymatic β‐galactosidase detection methods and with only minor impact on enumeration times if an appropriate selective medium was incorporated into the tests.
Applicability of bioelectroanalytical approach to TTC detection
Enzymatic tests at 44.5°C are sometimes used to select for E. coli. It is preferably referred to as the TTC test because some Shigella spp. and Salmonella spp. also detect positive using this method. This apparent lack of specificity has its origins in the historical assignment of these organisms to different genera based on the pathology of the diseases they cause. However, with modern molecular insights, these taxa would be consigned to the same genus (Lan and Reeves, 2002). In a test with a constant inoculum size of E. coli (5.0 × 105 CFU ml−1) but an increasing inoculum size of A. hydrophila (from 5.0 × 102 to 5.0 × 105 CFU ml−1, equivalent to 1000:1, 100:1, 10:1 and 1:1) and using 2H14NQGlu as a detection compound, the enumeration time remained reasonably constant (~2–2.9 h) which was not observed at 37°C, indicating that the contribution of A. hydrophila to the signal was minor at 44.5°C but had a significant impact at 37°C (Fig. 4). However, there was some attenuation of the enumeration time when the ratio of A. hydrophila to E. coli was high (0.1:1 and 1:1, Fig. 3) suggesting a competitive effect between A. hydrophila and E. coli. In a real situation, this would manifest as a slight overestimation of the total number of E. coli in the sample. As this effect is only apparent at both high inoculum density and when the relative concentration of A. hydrophila compared to E. coli is high (1:1), it would not pose significant problems in field applications, nor would the extent of interference be above that which would be expected with chromogenic β‐glucuronidase enzymatic tests. The electrochemical activity of A. hydrophila, thought to be achieved by direct electron transfer via c‐type cytochromes, was previously shown to be dependent on growth conditions, it is a noteworthy observation that there appears to be a temperature dependence too (Pham et al., 2003).
Figure 4.

Detection time of Escherichia coli at 44.5°C with 2H14NQGlu at initial inoculum sizes of 10, 50, 500 and 5000 CFU (equivalent to 1, 5, 50 and 500 CFU ml−1) compared with a commercial detection system of 10, 100, 1000, and 1000 CFU (data from James et al., 2010). Error bars represent 1 standard deviation with n = 3 except for the commercial system where n = 5.
It has been demonstrated that the utility of the electrochemical glycosides is equivalent to their commercially available chromogenic counterparts. Nonetheless, it is clear that the presence of environmental organisms will have an effect on the accuracy of the enumeration times in field settings. This problem is shared by chromogenic, fluorogenic and bioelectroanalytical enzymatic tests alike, which rely on glycosides to detect the presence of β‐glucuronidase or β‐galactosidase. Interference by non‐target organisms is usually overcome by inclusion of a selective medium. The most common component of selective media for Enterobacteriacea is bile salts whose utility has been shown here. To completely suppress all non‐target organisms that interfere with enzymatic coliform or E. coli detection assays requires the incorporation of a mixture of specific inhibitors into the medium. Examples include antifungals such as amphotericin; rosolic acid to suppress organisms that grow at 44.5°C (e.g. Staphylococcus aureus) and cephalosporins such as cefsulodin that selectively inhibit Aeromonads (Presswood and Strong, 1978; Geissler et al., 2000; Manafi, 2000). The ideal medium for bioelectroanalytical detection of E. coli that suppresses non‐target fungi and bacteria should be the subject of further investigations.
Performance against commercial system
Inoculum sizes in the range 5.0 × 102–5.0 × 105 CFU ml−1 are useful to demonstrate proof of principle, but we acknowledge that this is not an ideal range for commercial purposes. U.S. EPA drinking water standards recommended samples are free from E. coli, whereas recreational water quality standards recommend a threshold of 100 CFU E. coli in 100 ml of water with a statistical threshold value of 320 CFU in 100 ml, which must not be exceeded by 10% of samples taken (U.S. EPA, 2012). The market leading enzymatic tests achieve detection of 1 CFU E. coli in 100 ml in 12–48 h (Colifast, 2016, Idexx, 2016).
Bearing in mind that the set‐up described here has not been optimized, we decided to extend the range down to 1 CFU ml−1 to investigate the current limits of the technique. To benchmark our technique against commercially available system that has a different reactor volume, the absolute number of organisms in the sample rather than the inoculum density has been used. At 44.5°C, the enumeration time for 1 CFU ml−1 (10 CFU) of E. coli in an electrochemical system was ~19 h and progressed in inverse fashion to less than 10 h for an inoculum size of 500 CFU ml−1 (5000 CFU). However, at low inoculum density, the times are less reproducible (Fig. 4). Poor reproducibility at low inoculum density is to be expected as differences in inoculum size begin to disproportionately contribute to the error of the technique. With the small volume ECs, achieving commercial detection targets (0.01 CFU ml−1) is difficult as the average number of cells per reactor would be less than 1 CFU (effectively 0 CFU in 9 reactors of 10). To overcome this limitation would require larger volume ECs and the application of an iterative statistical approach, which is impractical with the current system. However, with some design modifications larger‐scale reactors could be developed to bring this method in line with commercial and regulatory standards.
A commercially available, U.S. EPA approved, automated E. coli detection system that uses a chromogenic method and a similar time‐to‐detection quantification procedure as used here can quantify 10 (0.1 CFU ml−1) organisms in around 15 h (James et al., 2010) compared to 19 h achieved with our system (Fig. 4). However, we achieved similar enumeration times of ~12.5 h for 500 (50 CFU ml−1) organisms and faster enumeration times of ~9 h for 5000 (500 CFU ml−1) organisms compared to 13.5 h for the commercial technique (James et al., 2010). Because of differences in reactor size and the precise conditions of the commercial technique (e.g. detection compound and concentration, and enumeration method) remain unknown, such a comparison is superficial but nonetheless illustrative. Whereas recognizing that additional optimization, medium development and validation is needed for the bioelectroanalytical approach, the results are very promising for an unoptimized system. Although the performance reported for the bioelectroanalytical detection of E. coli is not yet sufficient to assess drinking water quality, the scope for improved enumeration times and its current utility for environmental monitoring, subject to the inclusion of a selective medium, have been demonstrated.
Conclusions
A bioelectroanalytical approach to E. coli enumeration using redox‐active glycosides has been demonstrated and the rational design of redox‐active glycosides, 2H14NQGal and 2H14NQGlu, led to both cost and performance improvements compared with commercially available compounds. The potential application of these novel glycosides in bioelectroanalytical detection of E. coli is feasible opening the door for further development towards alternative, low‐cost bioelectroanalytical detection systems. It is clear, however, that the development of a selective medium that hastens the recovery of environmentally stressed cells along with scale up to 100 ml volumes is a research priority which will allow validation of this method against U.S. EPA standards. We contend that bioelectroanalytical approaches offer much scope for future development including better performing glycosides, novel detection media and functionalized electrodes and that, potentially, a number of performance benefits such as more rapid detection in a better range of matrices could be realized. Importantly, the approach opens up a new research space in an area that has not, commercially at least, reflected the innovative trends apparent in research literature.
Conflicts of interest
The authors declare no competing financial interest.
Microbial Biotechnology (2016) 9(6), 746–757
Funding Information SCELSE is funded by the National Research Foundation, Ministry of Education, Nanyang Technological University (NTU) and National University of Singapore (NUS) and hosted by NTU in partnership with NUS. This work was enabled by a start‐up grant from the Sustainable Earth office at NTU under the Sustainable Earth programme. Special thanks to Brady Clark of Sussex Research Ltd for discussions on glycoconjugate synthesis and Jean Pierre Nshimyimana for assistance with the reference method for enumerating E. coli in environmental samples. We acknowledge with gratitude the important improvements suggested by three anonymous reviewers.
Contributor Information
Jamie Hinks, Email: jhinks@ntu.edu.sg.
Stefan Wuertz, Email: swuertz@ntu.edu.sg.
References
- Antonczak, A.K. , Simova, Z. , and Tippmann, E.M. (2009) A critical examination of Escherichia coli esterase activity. J Biol Chem 284: 28795–28800. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ashbolt, N.J. (2004) Microbial contamination of drinking water and disease outcomes in developing regions. Toxicology 198: 229–238. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bej, A.K. , DiCesare, J.L. , Haff, L. , and Atlas, R.M. (1991) Detection of Escherichia coli and Shigella spp. in water by using the polymerase chain reaction and gene probes for uid. Appl Environ Microbiol 57: 1013–1017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Biosynth (2012) Method for detection of beta‐d‐glucuronidase. Google Patents.
- Bissonnette, G.K. , Jezeski, J.J. , McFeters, G.A. , and Stuart, D.G. (1975) Influence of environmental stress on enumeration of indicator bacteria from natural Waters. J Appl Microbiol 29: 186–194. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cabral, J.P. (2010) Water microbiology. Bacterial pathogens and water. Int J Environ Res Public Health 7: 3657–3703. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Candeias, L.P. , MacFarlane, D.S. , McWhinnie, S.W. , Maidwell, N. , Roeschlaub, C. , Sammes, P. , and Whittlesey, R. (1998) The catalysed NADH reduction of resazurin to resorufin. J Chem Soc Perkin Trans 2: 2333–2334. [Google Scholar]
- Choi, Y. , Kim, N. , Kim, S. , and Jung, S. (2003) Dynamic behaviors of redox mediators within the hydrophobic layers as an important factor for effective microbial fuel cell operation. Bull Korean Chem Soc 24: 437–440. [Google Scholar]
- Colifast (2016). Norway. www.colifast.no
- Douterelo, I. , Boxall, J.B. , Deines, P. , Sekar, R. , Fish, K.E. , and Biggs, C.A. (2014) Methodological approaches for studying the microbial ecology of drinking water distribution systems. Water Res 65: 134–156. [DOI] [PubMed] [Google Scholar]
- Fiksdal, L. , Pommepuy, M. , Caprais, M.P. , and Midttun, I. (1994) Monitoring of fecal pollution in coastal waters by use of rapid enzymatic techniques. Appl Environ Microbiol 60: 1581–1584. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Frontana, C. , and González, I. (2005) The role of intramolecular hydrogen bonding in the electrochemical behavior of hydroxy‐quinones and in semiquinone stability. J Braz Chem Soc 16: 299–307. [Google Scholar]
- Geissler, K. , Manafi, M. , Amoros, I. , and Alonso, J. (2000) Quantitative determination of total coliforms and Escherichia coli in marine waters with chromogenic and fluorogenic media. J Appl Microbiol 88: 280–285. [DOI] [PubMed] [Google Scholar]
- Hata, A. , Katayama, H. , and Furumai, H. (2014) Application of organic analysis for characterization of substances which interfere with RT‐qPCR based virus detection in water samples. Appl Environ Microbiol. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Heery, B. , Briciu‐Burghina, C. , Zhang, D. , Duffy, G. , Brabazon, D. , O'Connor, N. , and Regan, F. (2016) ColiSense, today's sample today: a rapid on‐site detection of β‐d‐Glucuronidase activity in surface water as a surrogate for E. coli . Talanta 148: 75–83. [DOI] [PubMed] [Google Scholar]
- Idexx (2016) www.idexx.com/water/water-testing-solutions.html.
- James, R. , Lorch, D. , Pala, S. , Dindal, A. and McKernan, B.J. (2010) Environmental Technology Validation Report. Pathogen Detection Systems, Inc. Automated microbiolgy platform. In: Environ. Technol.: EPA.
- Kapoor, V. , Pitkänen, T. , Ryu, H. , Elk, M. , and Wendell, D. (2015) Santo Domingo JW. Distribution of Human‐Specific Bacteroidales and Fecal Indicator Bacteria in an Urban Watershed Impacted by Sewage Pollution, Determined Using RNA‐ and DNA‐Based Quantitative PCR Assays. Appl Environ Microbiol. 81(1), 91–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kapuscinski, R.B. , and Mitchell, R. (1981) Solar radiation induces sublethal injury in Escherichia coli in seawater. Appl. Environ. Microbiol. 41, 670–674. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kemmochi, Y. , Tsutsumi, K. and Onda, K. (2007) Method and apparatus for detecting bacteria. Google Patents.
- Khaengraeng, R. , and Reed, R.H. (2005) Oxygen and photoinactivation of Escherichia coli in UVA and sunlight. J Appl Microbiol. 99, 39–50. [DOI] [PubMed] [Google Scholar]
- Lan, R. , and Reeves, P.R. (2002) Escherichia coli in disguise: molecular origins of Shigella . Microb Infect 4: 1125–1132. [DOI] [PubMed] [Google Scholar]
- Lazcka, O. , Del Campo, F.J. , and Munoz, F.X. (2007). Pathogen detection: A perspective of traditional methods and biosensors. Biosensors and Bioelectronics. 22(7), 1205–17. [DOI] [PubMed] [Google Scholar]
- Lee, S.H. , Levy, D.A. , Craun, G.F. , Beach, M.J. , and Calderon, R.L. (2002) Surveillance for waterborne‐disease outbreaks–United States, 1999‐2000. MMWR Surveill Summ 51: 1–47. [PubMed] [Google Scholar]
- Liang, Z. , He, Z. , Zhou, X. , Powell, C.A. , Yang, Y. , Roberts, M.G. , and Stoffella, P.J. (2012) High diversity and differential persistence of fecal Bacteroidales population spiked into freshwater microcosm. Water Res 46: 247–257. [DOI] [PubMed] [Google Scholar]
- Logan, B.E. (2009a) Scaling up microbial fuel cells and other bioelectrochemical systems. Appl Microbiol Biotechnol 85: 1665–1671. [DOI] [PubMed] [Google Scholar]
- Logan, B.E. (2009b) Exoelectrogenic bacteria that power microbial fuel cells. Nat Rev Micro 7: 375–381. [DOI] [PubMed] [Google Scholar]
- Logan, B.E. , and Regan, J.M. (2006) Microbial fuel cells‐challenges and applications. Environ Sci Technol 40: 5172–5180. [PubMed] [Google Scholar]
- Magro, G. , Bain, R.E.S. , Woodall, C.A. , Matthews, R.L. , Gundry, S.W. , and Davis, A.P. (2014) Synthesis and application of resorufin β‐d‐glucuronide, a low‐cost chromogenic substrate for detecting Escherichia coli in drinking water. Environ Sci Technol 48: 9624–9631. [DOI] [PubMed] [Google Scholar]
- Manafi, M. (2000) New developments in chromogenic and fluorogenic culture media. Int J Food Microbiol 60: 205–218. [DOI] [PubMed] [Google Scholar]
- Manafi, M. , Kneifel, W. , and Bascomb, S. (1991) Fluorogenic and chromogenic substrates used in bacterial diagnostics. Microbiol Rev 55: 335–348. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Martins, M.T. , Rivera, I.G. , Clark, D.L. , Stewart, M.H. , Wolfe, R.L. , and Olson, B.H. (1993) Distribution of uidA gene sequences in Escherichia coli isolates in water sources and comparison with the expression of beta‐glucuronidase activity in 4‐methylumbelliferyl‐beta‐d‐glucuronide media. Appl Environ Microbiol 59: 2271–2276. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mathews, J. , Li, Q. , and Wang, G. (2010) Characterization of hydrogen production by engineered Escherichia coli strains using rich defined media. Biotechnol Bioprocess Eng 15: 686–695. [Google Scholar]
- Mathis, B. , Marshall, C. , Milliken, C. , Makkar, R. , Creager, S.E. , and May, H. (2008) Electricity generation by thermophilic microorganisms from marine sediment. Appl Microbiol Biotechnol 78: 147–155. [DOI] [PubMed] [Google Scholar]
- Nelis, H. and Van Poucke, S. (2000) Enzymatic detection of coliforms and Escherichia coli within 4 hours In Environmental Challenges: Springer, pp. 43–52. [Google Scholar]
- Nemiroski, A. , Christodouleas, D.C. , Hennek, J.W. , Kumar, A.A. , Maxwell, E.J. , Fernández‐Abedul, M.T. , and Whitesides, G.M. (2014) Universal mobile electrochemical detector designed for use in resource‐limited applications. Proc Natl Acad Sci U S A 111: 11984–11989. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nijak, G.M., Jr , Geary, J.R. , Larson, S.L. , and Talley, J.W. (2012). Autonomous, wireless in‐situ sensor (AWISS) for rapid warning of Escherichia coli outbreaks in recreational and source waters. Environ. Eng. Sci. 29, 64–69. [Google Scholar]
- Nshimyimana, J.P. , Ekklesia, E. , Shanahan, P. , Chua, L.H.C. , and Thompson, J.R. (2014) Distribution and abundance of human‐specific Bacteroides and relation to traditional indicators in an urban tropical catchment. J Appl Microbiol 116: 1369–1383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Odagiri, M. , Schriewer, A. , Hanley, K. , Wuertz, S. , Misra, P.R. , Panigrahi, P. , and Jenkins, M.W. (2015) Validation of Bacteroidales quantitative PCR assays targeting human and animal fecal contamination in the public and domestic domains in India. Sci Total Environ 502: 462–470. [DOI] [PubMed] [Google Scholar]
- Park, D.H. , and Zeikus, J.G. (2000) Electricity generation in microbial fuel cells using neutral red as an electronophore. Appl Environ Microbiol 66: 1292–1297. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Perez, F. , Tryland, I. , Mascini, M. , and Fiksdal, L. (2001) Rapid detection of Escherichia coli in water by a culture‐based amperometric method. Anal Chim Acta 427: 149–154. [Google Scholar]
- Pham, C.A. , Jung, S.J. , Phung, N.T. , Lee, J. , Chang, I.S. , Kim, B.H. , et al (2003) A novel electrochemically active and Fe (III)‐reducing bacterium phylogenetically related to Aeromonas hydrophila, isolated from a microbial fuel cell. FEMS Microbiol Lett 223: 129–134. [DOI] [PubMed] [Google Scholar]
- Presswood, W.G. , and Strong, D.K. (1978) Modification of M‐FC medium by eliminating rosolic acid. Appl Environ Microbiol 36: 90–94. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Qiao, Y. , Li, C.M. , Bao, S.‐J. , Lu, Z. , and Hong, Y. (2008) Direct electrochemistry and electrocatalytic mechanism of evolved Escherichia coli cells in microbial fuel cells. Chem Commun 11: 1290–1292. [DOI] [PubMed] [Google Scholar]
- Rabaey, K. , and Rozendal, R.A. (2010) Microbial electrosynthesis—revisiting the electrical route for microbial production. Nat Rev Microbiol 8: 706–716. [DOI] [PubMed] [Google Scholar]
- Rabaey, K. , Boon, N. , Höfte, M. , and Verstraete, W. (2005) Microbial phenazine production enhances electron transfer in biofuel Cells. Environ Sci Technol 39: 3401–3408. [DOI] [PubMed] [Google Scholar]
- Rabaey, K. , Rodriguez, J. , Blackall, L.L. , Keller, J. , Gross, P. , Batstone, D. , et al (2007) Microbial ecology meets electrochemistry: electricity‐driven and driving communities. ISME J 1: 9–18. [DOI] [PubMed] [Google Scholar]
- Rau, J. , Knackmuss, H.‐J. , and Stolz, A. (2002) Effects of different quinoid redox mediators on the anaerobic reduction of azo dyes by bacteria. Environ Sci Technol 36: 1497–1504. [DOI] [PubMed] [Google Scholar]
- Reischer, G.H. , Ebdon, J.E. , Bauer, J.M. , Schuster, N. , Ahmed, W. , Åström, J. , et al (2013) Performance characteristics of qPCR assays targeting human and ruminant‐associated Bacteroidetes for microbial source tracking across sixteen countries on six continents. Environ Sci Technol 47: 8548–8556. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Roller, S.D. , Bennetto, H.P. , Delaney, G.M. , Mason, J.R. , Stirling, J.L. , and Thurston, C.F. (1984) Electron‐transfer coupling in microbial fuel cells: 1. comparison of redox‐mediator reduction rates and respiratory rates of bacteria. J Chem Technol Biotechnol 34: 3–12. [Google Scholar]
- Rowe, A.A. , Bonham, A.J. , White, R.J. , Zimmer, M.P. , Yadgar, R.J. , Hobza, T.M. , et al (2011) CheapStat: an open‐source, “do‐it‐yourself” potentiostat for analytical and educational applications. PLoS ONE 6: e23783. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Seurinck, S. , Defoirdt, T. , Verstraete, W. , and Siciliano, S.D. (2005) Detection and quantification of the human‐specific HF183 Bacteroides 16S rRNA genetic marker with real‐time PCR for assessment of human faecal pollution in freshwater. Environ Microbiol 7: 249–259. [DOI] [PubMed] [Google Scholar]
- Seviour, T. , Doyle, L. , Lauw, S. , Hinks, J. , Rice, S. , Nesatyy, V. , et al (2015) Voltammetric profiling of redox‐active metabolites expressed by Pseudomonas aeruginosa for diagnostic purposes. Chem Commun 51: 3789–3792. [DOI] [PubMed] [Google Scholar]
- Shen, G.G.Y. , Oh, C.S. , Yancey, S.D. and Adams, C.W. (2003) Synthesis of chlorophenol red glucuronic acid. Google Patents.
- TMR (2015) E. coli Testing (Environmental Testing and Clinical Testing) Market ‐ Global Industry Analysis, Size, Share, Growth, Trends and Forecast 2014 ‐ 2022. Pune, India: TMR. [Google Scholar]
- Unden, G. (1988) Differential roles for menaquinone and demethylmenaquinone in anaerobic electron transport of E. coli and their fnr‐independent expression. Arch Microbiol 150: 499–503. [DOI] [PubMed] [Google Scholar]
- U.S. EPA (2012) Recreational Water Quality Criteria. EPA 820‐F‐12‐058.
- Vuorenmaa, J. , Forsius, M. , and Mannio, J. (2006) Increasing trends of total organic carbon concentrations in small forest lakes in Finland from 1987 to 2003. Sci Total Environ 365: 47–65. [DOI] [PubMed] [Google Scholar]
- Waltman, W. , Shotts, E. , and Hsu, T. (1982) Enzymatic characterization of Aeromonas hydrophila complex by the API ZYM system. J Clin Microbiol 16: 692–696. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wang, N. , He, M. , and Shi, H.‐C. (2007) Novel indirect enzyme‐linked immunosorbent assay (ELISA) method to detect total E. coli in water environment. Anal Chim Acta 590: 224–231. [DOI] [PubMed] [Google Scholar]
- Wang, V.B. , Du, J. , Chen, X. , Thomas, A.W. , Kirchhofer, N.D. , Garner, L.E. , et al (2013) Improving charge collection in Escherichia coli‐carbon electrode devices with conjugated oligoelectrolytes. Phys Chem Chem Phys 15: 5867–5872. [DOI] [PubMed] [Google Scholar]
- Wuertz, S. , Wang, D. , Reischer, G. and Farnleitner, A. (2011) Library‐independent bacterial source tracking methods In Microbial Source Tracking: Methods, Applications, and Case Studies Hagedorn C., Blanch A.R. and Harwood V.J. (eds). New York: Springer, pp. 61–112. [Google Scholar]
- Zauba (2016). Online: Zauba Technologies & Data Services Pvt Ltd 2013. Napthoquinone and resorufin prices.