Abstract
High-speed countercurrent chromatography with a spiral tube assembly can retain a satisfactory amount of stationary phase of polymer phase systems used for protein separation. In order to improve the partition efficiency a simple tool to modify the tubing shapes was fabricated, and the following four different tubing modifications were made: intermittently pressed at 10 mm width, flat, flat-wave, and flat-twist. Partition efficiencies of the separation column made from these modified tubing were examined in protein separation with an aqueous-aqueous polymer phase system at flow rates of 1–2 ml/min under 800 rpm. The results indicated that the column with all modified tubing improved the partition efficiency at a flow rate of 1 ml/min, but at a higher flow rate of 2 ml/min the columns made of flattened tubing showed lowered partition efficiency apparently due to the loss of the retained stationary phase. Among all the modified columns, the column with intermittently pressed tubing gave the best peak resolution. It may be concluded that the intermittently pressed and flat-twist improve the partition efficiency in a semi-preparative separation while other modified tubing of flat and flat-wave configurations may be used for analytical separations with a low flow rate.
Keywords: Spiral high-speed countercurrent chromatography, tubing shape modification, protein separation, polymer phase system
1. Introduction
Countercurrent chromatography (CCC) uses two-phase solvent systems without solid support [1–5] and it has an advantage over the conventional column chromatography for eliminating the risk of sample loss on the solid support matrix used in liquid chromatography. Since 1970 various CCC systems have been developed and examined their partition efficiency in terms of theoretical plate number, peak resolution and retention of the stationary phase. Among those CCC systems high speed countercurrent chromatography (HSCCC) has been most widely used for the separation of natural and synthetic products [6–8]. This system uses a type-J planetary centrifuge which rotates a coiled separation column in such a way that the column synchronously counter-rotates about its own axis while revolving around the central axis of the centrifuge. The use of the coiled column in the rotating centrifugal force field generates an Archimedean screw effect to mix the two phases while retaining the stationary phase in the column. The separation column is fabricated by winding a long piece of PTFE (polytetra-fluoroethylene) tubing coaxially around the column holder making multiple layers. This multilayer coil separation column can retain a satisfactory amount of the stationary phase for moderately polar organic/aqueous two-phase solvent systems and yields excellent peak resolution of various natural products. However, this conventional multilayer coil fails to retain a satisfactory amount of the stationary phase of polar aqueous/aqueous polymer phase systems [9] which are useful for the separation of bioactive macromolecules such as proteins and nucleic acids. In order to improve the stationary phase retention of these viscous and low interfacial tension polymer phase systems, column configuration was made in a spiral shape so that it utilizes the radially formed centrifugal force gradient to hold the lighter stationary phase in the periphery and the heavier stationary phase in the inner spiral channel to improve the retention of the stationary phase. Two types of spiral columns have been introduced, i.e. spiral disk assembly [10, 11] and spiral tube assembly [11, 12]. The spiral disk assembly is composed of several inert plastic disks (Kel-F, polyethylene and polypropylene) each having four interwoven spiral channels that are connected in series with narrow channels made on the other side of the disk. It can make various channel shapes and also accommodate mixing glass beads in the channel to improve the partition efficiency [13]. The spiral tube assembly is made from a plastic disk with a set of deep spiral grooves to accommodate the tubing. It has an advantage over the spiral disk assembly by eliminating the risk of leakage through the seal, but it cannot easily insert the mixers to improve the partition efficiency as in the spiral disk assembly. Instead, the performance of the column can be improved by modifying the tube geometry to enhance the mixing and interrupt with the laminar flow of the mobile phase which would produce sample band broadening. In the past several tubing modifications such as cross-pressing [14] and flat-twisting [15] were manually made and tested. Although these modifications substantially improved the partition efficiency, it takes a long tedious process for column fabrication and lacks accuracy and reproducibility. Recently, Englert and Vetter have made a crimping tool to modify the tubing shape by drawing the tubing through a pair of wheels equipped with extrusions at regular intervals. They reported positive results with the modified tubing in the separation of 6 alkyl p-hydroxybenzoates with a multilayer coil assembly [16]. We have constructed a similar devise to form various tubing shapes such as intermittently pressed at various widths (10-0 mm), flat, flat-wave and flat-twist [17]. In the present study, performances of these four differently modified tubing geometries were examined in separation of a set of stable protein samples with a spiral tube assembly by HSCCC.
2. Materials and Methods
2.1. HSCCC Apparatus
High-speed CCC separation was performed by a type-J coil planet centrifuge (P.C. Inc., Potomac, MD, USA). It holds a single separation column and its counter-weight symmetrically on the rotary frame at a distance of 10 cm from the central axis of the apparatus. The separation column assembly was made from a spiral tube support (STS) custom-made with a laser sintering method from Nylon powder through CCBiotech, Rockville, MD, USA. It is 16 cm in diameter and 5 cm in height and has four deep interwoven spiral grooves to accommodate a single piece of PTFE tubing of 1.6 mm ID (SW 14, Zeus Industrial Products, Orangeburg, SC, USA) modified into various shapes (intermittently pressed, flattened, flat-waving and flat-twisting) by the tubing modifier. Each column has 8 – 9 spiral layers with the total capacity of about 80 ml. For comparative studies, the same STS was filled with non-pressed tubing of the same ID which formed 10 spiral layers with a total capacity of 114 ml. The column was rotated at 800 rpm with a speed controller. The effluent from the column was directly monitored with a UV detector (LKB Uvicord IIs, LKB Instruments, Bromma, Sweden) at 280 nm and the chromatogram was recorded with a strip-chart recorder (Millipore, Billerica, MA, USA).
2.2. Design of the tubing modifier
The basic design of the tubing modifier has been described previously (Fig. 1a) [17]. It is composed of a pair of identical gears of 7.6 cm (3 inch) pitch diameters and approximately 2 mm thick with 57 spur teeth. Each gear has a coaxial pressing wheel of 5 mm thick and ca 7.5 cm diameter with three different designs, i.e. intermittently pressing (IP 10), flat, flat-wave as shown in Figure 1b (flat-twist configuration was made by manually twisting the flat tubing along its axis). Intermittently pressing wheels is indented every 1 cm to compress the tubing in 1 cm width at 1 cm intervals. The wheel for flattening tubing has no indentation with smooth surface, and flat-waving wheel is waving surface at 1 cm pitch A long handle is attached to one of the gears to manually rotate the pair of pressing wheels each in the opposite direction. The tubing is inserted from a side hole to reach the pressing wheels, and the rotation of the gears will compress and automatically send out the modified tubing continuously until the rotation is stopped. This pressing tool can be applied to the tubing with the wall thickness of around 0.4 mm and is suitable to modify the PTFE (polytetrafluoroethylene) tubing of SW20 (0.85 mm ID, analytical), SW14 (1.6 mm ID, semi-preparative) and SW 10 (2.6 mm ID, preparative).
Figure 1a.

Tubing modifier
Figure 1b.

Standard and four differently modified tubing
A: standard; B: IP 10 mm; C: Flat; D: Flat-wave; E: Flat-twist
2.3. Reagents
Polyethylene glycol 1000 (PEG 1000), dibasic potassium phosphate, cytochrome c, myoglobin and lysozyme were all purchased from Sigma Chemicals, St. Louis, MO, USA. Water used for preparation of the two-phase system was obtained from reverse osmosis system (Hydro Service and Supplies, Gaithersburg, MD, USA).
2.4. Preparation of two-phase solvent system and sample solution
The two-phase solvent system for proteins separation was prepared by dissolving PEG 1000 and dibasic potassium phosphate each at 12.5 % (w/w) in ionized water. The solvent mixture was thoroughly equilibrated in a separatory funnel and the two phases were separated shortly before use. Sample solution was prepared by dissolving suitable amounts of test protein mixture in the stationary upper phase used for separation.
2.5. HSCCC separation procedure
In each separation the column was first completely filled with the upper stationary phase of the two-phase solvent system followed by sample charge (1 ml volume containing 1 mg of cytochrome c, 10 mg of myoglobin and 10 mg of lysozyme). Then the column was rotated at 800 rpm while the mobile phase was introduced into the internal terminal of the spiral channel at a desired flow rate (1–2 ml/min) in a tail to head elution mode. The effluent from the outlet of the column was continuously monitored with a UV detector at 280 nm. In order to improve the elution curve, the effluent was diluted with deionized water through a dialyser (Spectrum, California, USA) inserted on-line near the inlet of the UV monitor at a flow rate of 1/10 of the mobile phase flow. After three peaks were eluted, the rotation was stopped and the column contents were pushed out by pressurized N2 into a graduated cylinder to determine the retention of the stationary phase. All separations were performed at room temperature.
2.6. Analysis of HSCCC fractions
The partition efficiency of each separation was determined from the chromatogram in terms of theoretical plate number (N) and peak resolution (Rs) according to the conventional formulae as follow:
| (1) |
| (2) |
where tR denotes the retention time and W, the peak width at base where the subscripts indicate peak numbers. The stationary phase retention (Sf) was calculated by the volume of the retained stationary phase divided by the total column capacity and expressed in percentage. The efficiency values obtained from the non-processed and four differently modified columns were all normalized to a 60 ml column volume for fair comparison (Table 1). The method of normalization is as follows:
| (3) |
| (4) |
where V is the column capacity.
Table 1.
Partition efficiencies of standard and 4 modified columns by normalizing in column capacity.
| Column | 1 ml/min | 2 ml/min | ||||
|---|---|---|---|---|---|---|
| Nnorm | Rsnorm | Sf | Nnorm | Rsnorm | Sf | |
| Standard | 108/11/19 | 1.05/0.90 | 78% | 455/35/27 | 1.09/1.15 | 63% |
| IP 10 mm | 460/46/39 | 1.55/1.27 | 65% | 294/45/43 | 1.17/1.21 | 56% |
| flat | 266/120/37 | 1.21/0.92 | 40% | 777/189/45 | 0.92/1.02 | 38% |
| flat-wave | 644/176/77 | 1.36/1.35 | 38% | 340/72/40 | 0.86/0.75 | 26% |
| flat-twist | 404/96/80 | 1.82/1.43 | 40% | 718/111/72 | 0.97/1.02 | 38% |
N: Theoretical plate number (peaks 1/2/3); Rs (peaks 1&2/2&3): Peak resolution; Sf; Percentage retention of the stationary phase.
Note: All separations were performed at 800 rpm and monitored uv absorbance at 280 nm.
3. Results
Figure 2 shows a set of chromatograms obtained from separation of three test proteins with a two-phase solvent system composed of PEG1000 and dibasic potassium phosphate each at 12.5% (w/w) in deionized water. All separations were performed at flow rates of 1 & 2 ml/min under 800 rpm. Partition efficiency and retention of the stationary phase of these separations are all normalized to a 60 ml column capacity for fair comparison (Table 1). At 1 ml/min flow rate, all the modified columns show higher partition efficiency over the standard column. Among the four modified columns, IP 10 (intermittently pressed at 10 mm width at 10 mm interval) and flat-twisted tubing show the best results. At 2 ml/min flow rate, however, the partition efficiencies of modified columns were all sharply decreased associated with the lowered stationary phase retention especially in flat-waved tubing, while the IP 10 column still maintained higher peak resolution (1.17/1.21) than the non-pressed standard column (1.09/1.15). It is interesting to observe that the standard column and IP 10 could maintain higher retention level of the stationary phase than three other flat pressed columns which restrict the flow of the mobile phase through narrow opening of the channel.
Figure 2.

Set of HSCCC chromatograms for protein separation using standard and 4 modified tubing. Experimental conditions: Apparatus: type-J coil planet centrifuge with 10 cm revolution radius; column: spiral tube assembly with 4 interwoven spiral tubes containing standard and 4 processed tubing; volume for each column is as follows: standard: 60 ml; IP 10: 52 ml; flat: 45 ml; flat-wave: 45 ml; and flat-twist: 38 ml: solvent system: polymer phase system composed of PEG1000 and dibasic potassium phosphate each 12.% in deionized water; sample: cytochrome c (1 mg) (peak I), myoglobin (10 mg)(peak II) and lysozyme (10 mg)(peak III) in 1 ml of upper phase; flow rate: 1 – 2 ml/min; mobile phase: upper phase; elution mode; tail to head elution; revolution speed: 800 rpm; detection: 280 nm. Stationary phase retention: see Table 1. IP 10 mm indicates intermittently pressed at 10 mm interval in 10 mm width 38 ml.
4. Discussion
As aforementioned, all the modified tubing showed higher peak resolution over the non-processed tubing. The mechanism of improved partition efficiency by modification in the intermittently pressed tubing was fully discussed in our previous paper [17]. Shortly, the flattened portion of the tubing gives larger interface and thin layer of each phase to accelerate the mass transfer rate of the analytes. The flattened portion of the tubing will interfere with the laminar flow of the mobile phase which would cause the sample band broadening to decrease the peak resolution. Also the flatten portion gives higher flow of the mobile phase which efficiently mixed with the stationary phase in the non-flattened portion of the column. In the present studies, tubing modification is further extended to flattened, flat-waving and flat-twisting. Although flattening tubing has been made simply by extruding the tubing through a narrow slit as previously performed [15], the degree of the flattening depends on the opening of the slit and the tubing tends to be stretched according to the applied force and the pulling speed, hence it lacks accuracy and reproducibility. In contrast, positively pushing out the tubing through rotating pressing wheels used in the present study totally eliminates these problems. Among three types of flattened tubing, the flat-twisting tubing gave the highest partition efficiency as shown in Table 1. The partition efficiency of the columns made of flattened and flat-waved tubing will be increased by decreasing the flow rate and/or increasing the revolution speed, but with an increased separation time. These tubing can be accommodated in the STS with much narrower channels of a sharper spiral pitch and used for analytical separations at a low flow rate of the mobile phase.
Footnotes
Conflict of Interest: The authors declare there is no conflict of interest.
Contributor Information
Yoichiro Ito, Email: itoy2@mail.nih.gov.
Robert Clary, Email: claryrob@od.nih.gov.
References
- 1.Ito Y, Bowman RL. Countercurrent chromatography: liquid–liquid partition chromatography without solid support. Science. 1970;167:281–283. doi: 10.1126/science.167.3916.281. [DOI] [PubMed] [Google Scholar]
- 2.Mandava NB, Ito Y. Countercurrent Chromatography: Theory and Practice. Marcel Dekker; New York: 1988. [Google Scholar]
- 3.Conway WD. Countercurrent Chromatography: Apparatus, Theory & Applications. VCH; Weinheim: 1990. [Google Scholar]
- 4.Berthod A. Countercurrent Chromatography: The Support-Free Liquid Stationary Phase. Wilson & Wilsons, Elsevier; New York: 2002. [Google Scholar]
- 5.Menet J-M, Thiébaut D. Countercurrent Chromatography, Chromatographic Science Series. Vol. 82. Marcel Dekker; New York: 1999. [Google Scholar]
- 6.Ito Y, Conway WD. High-speed countercurrent chromatography. Wiley-Interscience; New York: 1996. [Google Scholar]
- 7.Ito Y. High-speed countercurrent chromatography. Crit Rev Anal Chem. 1986;17:65–143. doi: 10.1021/ac00113a049. [DOI] [PubMed] [Google Scholar]
- 8.Ito Y. High-speed countercurrent chromatography. Nature. 1987;326:419–420. doi: 10.1038/326419a0. [DOI] [PubMed] [Google Scholar]
- 9.Albertsson PÅ. Partition of cell particles and macromolecules. Wiley-Interscience, John Wiley & Sons; New York: 1986. [Google Scholar]
- 10.Ito Y, Yang FQ, Fitze PE, Powell J, Ide D. Improved spiral disk assembly for high-speed countercurrent chromatography. J Chromatogr A. 2003;1017:71–81. doi: 10.1016/j.chroma.2003.08.011. [DOI] [PubMed] [Google Scholar]
- 11.Ito Y. Spiral column design for separation of proteins by high-speed countercurrent chromatography. Chem Eng Proc. 2010;49(7):782–792. doi: 10.1016/j.cep.2009.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Ito Y, Clary R, Powell J, Knight M, Finn MT. Spiral tube assembly for high- speed countercurrent chromatography: Choice of the elution modes for four typical two-phase solvent systems. J Liq Chrom, & Rel Technol. 2009;32:2013–2029. doi: 10.1080/10826070802019913. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Ito Y, Clary R, Sharpnack F, Metger H, Powell J. Mixer-settler counter-current chromatography with multiple spiral disk assembly. J Chromatog A. 2007;1172:151–157. doi: 10.1016/j.chroma.2007.09.078. [DOI] [PubMed] [Google Scholar]
- 14.Ito Y, Clary R, Powel J, Knight M, Finn TM. Spiral tube assembly for high-speed countercurrent chromatography: Choice of elution modes for four typical two-phase solvent systems. J Liq Chrom & Rel Technol. 2009;32:2013–2029. doi: 10.1080/10826070802019913. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Tong SQ, Ito Y, Ma Y. Enantioseparation of DL-tryptophan by spiral tube assembly counter-current chromatography and evaluation of mass transfer rate for enantiomers. J Chromatogr A. 2014;374:77–84. doi: 10.1016/j.chroma.2014.11.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Englert M, Vetter W. Tubing modifications for countercurrent chromatography (CCC): Stationary phase retention and separation efficiency. Anal Chim Acta. 2015;884:114–123. doi: 10.1016/j.aca.2015.04.055. [DOI] [PubMed] [Google Scholar]
- 17.Ito Y, Ma XF, Clary R. A simple tool for tubing modification to improve the spiral high-speed counter-current chromatography. Current Chromatography. doi: 10.2174/2213240603666160618073940. in press. [DOI] [PMC free article] [PubMed] [Google Scholar]