. 2016 Oct 24;54(11):2701–2706. doi: 10.1128/JCM.01475-16

Rapid Antimicrobial Susceptibility Testing Using Forward Laser Light Scatter Technology

Randall T Hayden a,, Lani K Clinton b, Carolyn Hewitt a, Terri Koyamatsu c, Yilun Sun d, Ginger Jamison a, Rosalie Perkins a, Li Tang d, Stanley Pounds d, Matthew J Bankowski c,e
Editor: R Patelf
PMCID: PMC5078546  PMID: 27558176

Abstract

The delayed reporting of antimicrobial susceptibility testing remains a limiting factor in clinical decision-making in the treatment of bacterial infection. This study evaluates the use of forward laser light scatter (FLLS) to measure bacterial growth for the early determination of antimicrobial susceptibility. Three isolates each (two clinical isolates and one reference strain) of Staphylococcus aureus, Escherichia coli, and Pseudomonas aeruginosa were tested in triplicate using two commercial antimicrobial testing systems, the Vitek2 and the MicroScan MIC panel, to challenge the BacterioScan FLLS. The BacterioScan FLLS showed a high degree of categorical concordance with the commercial methods. Pairwise comparison with each commercial system serving as a reference standard showed 88.9% agreement with MicroScan (two minor errors) and 72.2% agreement with Vitek (five minor errors). FLLS using the BacterioScan system shows promise as a novel method for the rapid and accurate determination of antimicrobial susceptibility.

INTRODUCTION

Rapid, accurate identification and antimicrobial susceptibility testing (AST) is necessary for optimal clinical decision-making and resource utilization. Timely antibiotic therapy is particularly critical to outcomes in patients with sepsis (1), because each hour of delay within the first 6 h for patients with septic shock results in a 7.6% decrease in survival (2). Long turnaround times (TAT) for either identification or AST result in empirical rather than pathogen-specific therapy. As a result, patients are often given empirical broad-spectrum antibiotics based on clinical suspicion and epidemiology.

Delays in both pathogen identification and AST have led to inappropriate or unnecessary antibiotic use in 20 to 50% of patients in acute-care hospitals in the United States (38). Such care provides uncertain clinical benefit and may increase the risk for unexpected side effects, such as Clostridium difficile infection, hepatotoxicity, cardiac arrhythmias, or hemolysis (912). Standard identification and susceptibility methods based on traditional phenotype testing in the past typically required 24 to 72 h, with an average TAT of 40 h as reported by Kerremans et al. (13). Fortunately, TATs have decreased due to advances in technology. For example, in a recent report by Machen et al., the average TAT for both identification and AST was 11.4 h using the lysis-filtration method for both Vitek MS and VITEK2 compared with 56.3 h for conventional methods (14).

Molecular-based platforms are an increasingly popular option that offers rapid TAT for the detection of genes such as mecA, which confers antibiotic resistance to methicillin-resistant Staphylococcus aureus (MRSA). These assays may provide results in less than 2 h (1517). However, molecular detection of the gene targets is far from a perfect solution, targeting only select genetic loci, typically responsible for only a fraction of important resistance patterns. In addition, findings are not always specific. For example, methicillin resistance by genotyping does not always indicate phenotypic resistance. This is evidenced by the finding of “mecA dropout” or “empty cassette” strains (17, 18).

There is a widespread effort to improve pathogen-specific antibiotic use. In order to support this goal, the Centers for Disease Control and Prevention (CDC) have recommended that all acute-care hospitals implement antibiotic stewardship programs (ASPs) (8). The goal of ASPs is to optimize appropriate antimicrobial treatment, thereby reducing adverse outcomes (1921) and preventing further antimicrobial resistance. Technologies have evolved to address these needs, and they include not only molecular methods but also methods based on the use of microfluidics, cell lysis detection, mass spectrometry, rapid cytometry, isothermal microcalorimetry, and magnetic bead rotation, among others (2225). Forward laser light scatter (FLLS) is another promising emerging technology that depends on electro-optical technology to measure bacterial growth prior to visual assessment. It has the potential of offering the sensitivity of phenotypic methodology with the rapid TAT of a molecular test. This study provides initial proof-of-principle data regarding the use of FLLS technology for AST testing, comparing accuracy and TAT to those of conventional AST systems.

MATERIALS AND METHODS

Study design.

Three isolates each of Staphylococcus aureus, Escherichia coli, and Pseudomonas aeruginosa were used, including two delinked clinical isolates and one reference strain. All isolates were coded and tested blindly in triplicate using FLLS and in singlet using two commercial AST systems. The commercial systems used were the automated Vitek2 (bioMérieux, Marcy l'Etoile, France) and the manual MicroScan (Siemens, Erlangen, German). The commercial AST systems were compared to FLLS using the BacterioScan 216R system (BacterioScan, Inc., St. Louis, MO). Errors were characterized using standard definitions as described in the CLSI M23-A3 document as very major, major, and minor (26).

Cation-adjusted Mueller-Hinton broth (CAMHB) was used to dilute the stock solution of each antimicrobial using a 2-fold increment dilution scheme with concentrations based on CLSI guidelines (M100-S25) (27). Bacterial inocula were prepared as follows. Isolated colonies were picked from an 18- to 24-h agar plate and suspended in saline to achieve a turbidity equivalent to a 0.5 McFarland standard. Within 15 min of preparation, this suspension was further diluted 1:150 in saline to give an approximate concentration of 1 × 106 CFU/ml of organism. A final 2-fold dilution occurred when 1 ml of the adjusted inoculum was added to each tube containing 1 ml of antimicrobial agent in the dilution series. The final volume for each of the bug-drug combination dilutions was 2 ml. The 2-ml aliquot was then used for inoculation of the BacterioScan system. Broth microdilution using the MicroScan panel system and MIC testing using the Vitek2 system were performed at Diagnostic Laboratory Services (DLS). BacterioScan FLLS system testing was performed at the St. Jude Children's Research Hospital.

Forward laser light scatter system.

There is a long history of using laser light scatter instrumentation to study microbiological samples, including the effects of antibiotics (2831). The BacterioScan system uses a laser light source to measure a liquid sample's optical density (OD) as well as the scattered intensity in a direction that passes near the laser beam. This low-angle forward scattered signal allows the instrument to approximate the sample's OD value to levels significantly lower (1 or 2 orders of magnitude) than those obtainable using a simple ratiometric transmittance type measurement. The BacterioScan system uses a low-power laser (typically in the visible wavelength range, i.e., 400 to 700 nm) passing through approximately 25 mm of a liquid sample in a specially designed cuvette to minimize noise signals from various sources. Since typical bacterial samples have sizes in the 100- to 10,000-nm range, the sample size is comparable to the light wavelength; thus, Mie scattering theory is appropriate. Mie scattering theory is named after its developer, German physicist Gustav Mie, and refers to a series solution of Maxwell's equations for scattering by spheres or by infinite cylinders or other geometries where one can write separate equations for the radial and angular dependence of solutions. The formalism is generally used to calculate either how much light is scattered, the total optical cross section, or where it goes in the form factor. The laser beam transmitted through the sample and the low-angle forward-scattered signals are captured on a CMOS 2-dimensional camera sensor. The instrument mathematically processes the sensor image to minimize background noise and clutter, calibrates the measured values against predetermined baselines, and generates values for organism density. Multiple measurements over time are compared to calculate growth rates and generation times.

The BacterioScan model 216 tabletop instrument used for this study uses light scattering to measure the concentration of particles in a liquid sample and can be used to estimate the density of microorganisms in a liquid sample. Initially developed for measurement of actively growing organisms in urine, the instrument performs reliably down to a minimum density of 10,000 CFU/ml, which is commonly considered a diagnostic threshold for bacteriuria. Therefore, the laser microbial growth monitor (LMGM) technology has potential for other applications, such as AST. The system capacity accommodates up to 16 samples, which can be run simultaneously with a volume of 2 to 3 ml each. Samples are maintained at a thermostatically controlled elevated temperature to promote growth (onboard incubation), and repeated measurements are automatically collected at approximately 3-min increments to provide real-time continuous measurement of organism growth.

In the present study, each bacterial strain was tested using serial dilutions of three antimicrobials. Antimicrobials tested by FLLS depended on the organism: for S. aureus, clindamycin (0.25 to 4 μg/ml), moxifloxacin (0.5 to 4 μg/ml), and oxacillin (0.25 to 4 μg/ml); for E. coli, cefepime (2 to 32 μg/ml), ciprofloxacin (0.5 to 4 μg/ml), and gentamicin (1 to 16 μg/ml); for P. aeruginosa, cefepime (2 to 32 μg/ml), ciprofloxacin (0.5 to 4 μg/ml), and gentamicin (1 to 16 μg/ml). After inoculation, growth was measured by FLLS for 24 h at 36.7 to 37.3°C, with light scatter plotted against incubation time. The MIC of each antimicrobial corresponded to the lowest concentration at which all three replicate inocula showed growth inhibition, as defined below (see the Statistical Analysis section). Interpretation of MIC results as susceptible (S), susceptible-dose dependent (SDD), intermediate (I), and resistant (R) were based on CLSI guidelines (27).

Categorization of errors.

According to standard terminology listed in CLSI M23-A3, errors were classified as very major, major, and minor. A very major error (VME) occurred when the reference method result was resistant and the test method result was susceptible. A major error (ME) occurred when the reference method result was susceptible and the test method result was resistant. A minor error (mE) occurred when the reference method result was either susceptible or resistant with a test method result of intermediate, or the reference method result was intermediate with a test method result of susceptible or resistant (26).

Statistical analysis.

For each sample (representing one replicate of a given bacterial strain at a single concentration of antimicrobial), a dense series of OD measurements was obtained over time. The area under the curve (AUC) of the OD measurements, as a function of time, next was computed for each sample. For each treated sample, the relative area under the curve (rAUC) was computed as the ratio of the complete AUC (cAUC) of the treated sample to the cAUC of the control sample. For each set of three replicates of a particular organism strain treated with a particular agent, bacteria were defined to be inhibited if all three rAUC values were less than or equal to 0.10. In this way, for each organism-agent pair, a series of inhibition calls was obtained as a function of agent dose. The MIC was defined as the lowest dose at which an inhibition call was made. For each treated sample among a set of three replicates declared as inhibited, the partial area under the curve until each time point, pAUC(t), was calculated. For each of these treated samples, the relative partial area under the curve from each time point t, rpAUC(t), was further determined as the ratio of the pAUC(t) for the treated sample to the untreated sample. Time to inhibition was defined as the earliest time point, T, at which all calculated rpAUC(t) values of the set of three replicates were less than 0.10 for all subsequent time points (tT). Figures S1 and S2 in the supplemental material illustrate the calculation of the variables used in the statistical analysis. All analyses were performed using SAS, Windows version 9.3.

RESULTS

There were 360 experimental samples and 120 control samples collected for 27 drug-bacterium pairs. Strong bacterial growth was observed for all 120 control samples. One experimental sample was dropped from analysis for E. coli 9992 with cefepime at a concentration of 4 μg/ml, due to insufficient length (duration) of data collection. A total of 27 drug-bacterium pairs were investigated, and 19 (70%) revealed inhibition for at least one tested concentration of antimicrobial (Table 1). The remaining 8 drug-bacterium pairs gave a MIC greater than the highest tested concentration of antimicrobial. Data quality was adequate for interpretation for all drug-bacterium pairs tested. The MICs for S. aureus, E. coli, and P. aeruginosa were similar for cefepime, ciprofloxacin, and gentamicin on all three platforms, BacterioScan, Vitek, and MicroScan (Table 1). The expected MICs on Vitek and MicroScan are shown for the ATCC reference strains. Between Vitek and MicroScan, the only disparate results (showing >2 dilution difference) were seen for the P. aeruginosa isolate 27853 (cefepime) and for the S. aureus isolate 29213 (moxifloxacin).

TABLE 1.

MIC from BacterioScan, MicroScan, and Vitek

Bacterium, ID no., and antibiotic MIC in μg/ml (result) by:
AST interpretive criteria (μg/ml)
Quality control rangeb (MIC, in μg/ml)
BacterioScan MicroScan Viteka S SDD I R
E. coli (ESBL)
    3267
        Cefepime 32 (R) 8 (SDD) No MIC (R) ≤2 4–8 ≥16
        Ciprofloxacin >8 (R) >2 (R) ≥4 (R) ≤1 2 ≥4
        Gentamicin ≤4 (S) 2 (S) ≤1 (S) ≤4 8 ≥16
    9992
        Cefepime >64 (R) >16 (R) No MIC (R) ≤2 4–8 ≥16
        Ciprofloxacin >8 (R) >2 (R) ≥4 (R) ≤1 2 ≥4
        Gentamicin ≤4 (S) ≤1 (S) ≤1 (S) ≤4 8 ≥16
P. aeruginosa
    2700
        Cefepime 32 (R) >16 (R) ≥64 (R) ≤8 16 ≥32
        Ciprofloxacin ≤1 (S) ≤0.5 (S) No MIC (I) ≤1 2 ≥4
        Gentamicin ≤4 (S) 4 (S) ≤1 (S) ≤4 8 ≥16
    9018
        Cefepime 64 (R) 16 (I) 16 (I) ≤8 16 ≥32
        Ciprofloxacin 2 (I) 2 (I) ≥4 (R) ≤1 2 ≥4
        Gentamicin 32 (R) >8 (R) 8 (I) ≤4 8 ≥16
S. aureus (MRSA)
    3032
        Clindamycin >8 (R) ≥4 (R) ≥8 (R) ≤0.5 1–2 ≥4
        Moxifloxacin >8 (R) 4 (R) ≥8 (R) ≤0.5 1 ≥2
        Oxacillin >8 (R) ≥2 (R) ≥4 (R) ≤2 ≥4
    6172
        Clindamycin >8 (R) ≥4 (R) ≥8 (R) ≤0.5 1–2 ≥4
        Moxifloxacin 4 (R) 2 (R) 1 (I) ≤0.5 1 ≥2
        Oxacillin >8 (R) ≥2 (R) ≥4 (R) ≤2 ≥4
E. coli (ATCC)
    25922
        Cefepime ≤4 ≤2 ≤1 0.015–0.12
        Ciprofloxacin ≤1 ≤0.5 ≤0.25 0.004–0.015
        Gentamicin ≤4 ≤1 ≤1 0.25–1.0
P. aeruginosa (ATCC)
    27853
        Cefepime ≤4 4 ≤1 0.5–4.0
        Ciprofloxacin ≤1 ≤0.5 ≤0.25 0.25–1.0
        Gentamicin ≤4 2 ≤1 0.5–2.0
S. aureus (ATCC)
    29213
        Clindamycin ≤1 0.5 ≤0.25 0.06–0.25
        Moxifloxacin ≤1 ≤2 ≤0.25 0.015–0.12
        Oxacillin ≤1 ≤0.25 0.5 0.12–0.5
a

No MIC refers to an expert interpretation using the Vitek2 instrument.

b

Ranges for cation-adjusted Mueller-Hinton broth (see Table 5A in reference 27).

A comparison of the interpretative results across the platforms is included in Tables 1 and 2. The overall agreement between BacterioScan and Vitek was 72.2% (13/18), and the overall agreement between BacterioScan and MicroScan was 88.9% (16/18). There was one minor error (mE) for the E. coli 3267 extended-spectrum beta-lactam (ESBL), where BacterioScan indicated that the organism was resistant to cefepime but MicroScan indicated it was susceptible, dose dependent. There was also one mE with cefepime for P. aeruginosa 9018 where the MicroScan result was intermediate and the BacterioScan result was resistant. There was a high degree of concordance for bacterial resistance with Vitek with only five mEs (Tables 1 and 2). These mEs occurred with ciprofloxacin for P. aeruginosa 2700 and 9018, with cefepime and gentamicin for P. aeruginosa 9018, and with moxifloxacin for S. aureus 6172.

TABLE 2.

Summary of minor errorsa

Test comparison ID no. Bacterium Antibiotic
BacterioScan vs. MicroScan 3267 E. coli Cefepime
BacterioScan vs. MicroScan 9018 P. aeruginosa Cefepime
BacterioScan vs. Vitek 9018 P. aeruginosa Cefepime
BacterioScan vs. Vitek 9018 P. aeruginosa Gentamicin
BacterioScan vs. Vitek 2700 P. aeruginosa Ciprofloxacin
BacterioScan vs. Vitek 9018 P. aeruginosa Ciprofloxacin
BacterioScan vs. Vitek 6172 S. aureus Moxifloxacin
a

No very major or major errors were seen.

Optical density plots for E. coli 3267 are shown in Fig. 1 for both gentamicin (susceptible) and cefepime (resistant). The triplicate growth curves for cefepime at 4 μg/ml were identical to the control with no antimicrobial agent. It should be noted that as the dose of cefepime increased, the growth curves shifted to the right until growth was inhibited with a cefepime dose of 32 μg/ml. The triplicate growth curves for all concentrations of gentamicin show no growth, while the control with no antimicrobial agent showed rapid growth. The remainder of the optical density plots can be seen in the supplemental material (see Fig. S3).

FIG 1.

FIG 1

Optical density (OD) plots for treated replicates and controls for E. coli 3267 with gentamicin (A) and cefepime (B).

The distribution of time to inhibition for E. coli, P. aeruginosa, and S. aureus is shown in Fig. 2. The minimum time to inhibition was 32 min for S. aureus, and the maximum time to inhibition was 1,014 min for P. aeruginosa. Note that 80% of organisms were inhibited within 346 min (less than 6 h), and 95% of organisms were inhibited within 598 min (less than 10 h). There was more variability in the time to inhibition for P. aeruginosa with more than 20% of P. aeruginosa organisms inhibited beyond 350 min (Fig. 2 and Table 3).

FIG 2.

FIG 2

Distribution of time to inhibition.

TABLE 3.

Summary of starting times for inhibition (minutes)

Bacterium Starting time (min) by % inhibition
0% (minimum) 50% (median) 80% 90% 95% 100% (maximum)
Overall 32 211 346 429 598 1,014
E. coli 112 164 233 262 270 271
P. aeruginosa 141 231 494 678 858 1,014
S. aureus 32 232 288 346 388 388

DISCUSSION

This study demonstrates proof of principle for the use of forward laser light scatter to rapidly determine MICs for several bacterial isolates of high clinical importance. Representative Gram-positive and Gram-negative pathogens, tested against a variety of antimicrobial agents, demonstrated close agreement with results achieved by more conventional testing modalities, with 7 categorical errors (all minor errors). Results were obtained for most organism-drug pairs in less than 6 h. These results suggest promise for this technology to markedly reduce the time to susceptibility testing results in the clinical laboratory.

While a fairly stringent metric was used to determine MIC, in most cases the drop-off at MIC was marked and could be determined by casual visual inspection of the light scatter curves, or by using a much more liberally defined cutoff, without affecting results. However, there were some cases where this was not observed. Moreover, the rapidity of time to result was also not completely uniform. Although most isolates produced rapid results, 5% required greater than 10 h and rarely up to 12 h. Pseudomonas aeruginosa appeared to take the longest to produce a result, with one bug-drug combination requiring nearly 17 h to generate a final MIC. The latter finding may represent a weakness of the system, in that a potentially significant proportion of isolates may require read time approaching that of conventional methods. However, most results easily challenged the longer conventional MIC turnaround time (TAT). Given that accuracy and reproducibility were maintained and no read times exceeded those of conventional methods, one could envision using such a system for a majority of testing and reserving conventional MIC testing for use as a reflex method.

Several methods for rapid AST have been described in the literature (2225). Those in clinical use are primarily genotypic and sensitive and offer a rapid TAT. However, correlation between genotype and phenotype is imperfect, and as genotypic mechanisms of resistance continue to rapidly evolve, it is difficult for manufacturers to keep pace with new assays. Moreover, such modalities target relatively few genes in a well-defined subset of bacteria for which clear genotypic mechanisms are responsible for a large number of cases of clinically significant antibiotic resistance. Therefore, it remains unlikely that even a majority of bacterium-drug-mechanism combinations can be detected through clinical genotypic tests. On the other hand, rapid phenotypic tests have been increasingly seen as having the potential to decrease TAT while not compromising sensitivity of resistance detection or clinical correlation. FLLS is one of the newest entries in this field, and the data here suggest that this technology will prove to be a flexible and accurate alternative to broth microdilution or other traditional methods.

This pilot study was limited by the narrow selection of bacteria and antibiotics used and the fact that only a few clinical strains of each bacterial species were included in the testing. Further studies with a broader range of bacterial strains, antimicrobials, and previously defined genotypes will be necessary to assess the applicability of this method, relative error rates, and the potential advantage in TAT. Unlike some methods, this technique is not inherently limited by the number of optical channels, the production and validation of specific probes, or the definition of proteomic profiles or other surrogate markers of metabolism or growth. It may well be amenable to improving the speed of MIC for determination in a wide variety of clinical settings. This would include routine Gram-positive and -negative isolates as well as even more challenging bacteria, such as nontuberculous mycobacteria (NTM).

We used AUC metrics to quantitatively characterize the observed growth patterns in this study. Future research that explores the utility of other quantitative metrics of the growth patterns may yield algorithms for calling resistance or sensitivity that improve clinical performance. A practical limitation of the AUC metric is that it cannot be computed until after the entire series has been acquired. A more practical metric for real-time utilization may be the time until the OD measurements exceed a specific threshold. Future research should seek to determine the best threshold value.

The speed and accuracy of this innovative system holds great potential for clinical use in expediting AST results, improving the choice of appropriate antimicrobial treatment, and supporting antibiotic stewardship. Findings reported in this study suggest a broader evaluation of clinical isolates and antimicrobials to further explore the correlation between current methods of MIC determination and rapid phenotypic testing by FFLS. In addition, clinical outcome studies will be critical to demonstrate the potential clinical impact of such methods on clinical decision making, resource utilization, and clinical outcome.

Supplementary Material

Supplemental material

ACKNOWLEDGMENTS

We thank BacterioScan, Inc. (St. Louis, MO), for providing instrumentation and reagents for this study.

This work was supported in part by ALSAC.

We also thank the clinical and molecular staff of the Microbiology Department at Diagnostic Laboratory Services, Inc. (The Queen's Medical Center), for their superior technical support.

Footnotes

Supplemental material for this article may be found at http://dx.doi.org/10.1128/JCM.01475-16.

REFERENCES

  • 1.Ibrahim EH, Sherman G, Ward S, Fraser VJ, Kollef MH. 2000. The influence of inadequate antimicrobial treatment of bloodstream infections on patient outcomes in the ICU setting. Chest 118:146–155. doi: 10.1378/chest.118.1.146. [DOI] [PubMed] [Google Scholar]
  • 2.Kumar A, Roberts D, Wood KE, Light B, Parrillo JE, Sharma S, Suppes R, Feinstein D, Zanotti S, Taiberg L, Gurka D, Kumar A, Cheang M. 2006. Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med 34:1589–1596. doi: 10.1097/01.CCM.0000217961.75225.E9. [DOI] [PubMed] [Google Scholar]
  • 3.Camins BC, King MD, Wells JB, Googe HL, Patel M, Kourbatova EV, Blumberg HM. 2009. Impact of an antimicrobial utilization program on antimicrobial use at a large teaching hospital: a randomized controlled trial. Infect Control Hosp Epidemiol 30:931–938. doi: 10.1086/605924. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Ingram PR, Seet JM, Budgeon CA, Murray R. 2012. Point-prevalence study of inappropriate antibiotic use at a tertiary Australian hospital. Int Med J 42:719–721. doi: 10.1111/j.1445-5994.2012.02809.x. [DOI] [PubMed] [Google Scholar]
  • 5.Levin PD, Idrees S, Sprung CL, Weissman C, Weiss Y, Moses AE, Benenson S. 2012. Antimicrobial use in the ICU: indications and accuracy–an observational trial. J Hosp Med 7:672–678. doi: 10.1002/jhm.1964. [DOI] [PubMed] [Google Scholar]
  • 6.Patel SJ, Oshodi A, Prasad P, Delamora P, Larson E, Zaoutis T, Paul DA, Saiman L. 2009. Antibiotic use in neonatal intensive care units and adherence with Centers for Disease Control and Prevention 12 step campaign to prevent antimicrobial resistance. Pediatr Infect Dis J 28:1047–1051. doi: 10.1097/INF.0b013e3181b12484. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Dellit TH, Owens RC, McGowan JE Jr, Gerding DN, Weinstein RA, Burke JP, Huskins WC, Paterson DL, Fishman NO, Carpenter CF, Brennan PJ, Billeter M, Hooton TM, Infectious Diseases Society of America, Society for Healthcare Epidemiology of America. 2007. Infectious Diseases Society of America and the Society for Healthcare Epidemiology of America guidelines for developing an institutional program to enhance antimicrobial stewardship. Clin Infect Dis 44:159–177. doi: 10.1086/510393. [DOI] [PubMed] [Google Scholar]
  • 8.Fridkin S, Baggs J, Fagan R, Magill S, Pollack LA, Malpiedi P, Slayton R, Khader K, Rubin MA, Jones M, Samore MH, Dumyati G, Dodds-Ashley E, Meek J, Yousey-Hindes K, Jernigan J, Shehab N, Herrera R, McDonald CL, Schneider A, Srinivasan A, Centers for Disease Control and Prevention. 2014. Vital signs: improving antibiotic use among hospitalized patients. MMWR Morb Mortal Wkly Rep 63:194–200. [PMC free article] [PubMed] [Google Scholar]
  • 9.Alshammari TM, Larrat EP, Morrill HJ, Caffrey AR, Quilliam BJ, LaPlante KL. 2014. Risk of hepatotoxicity associated with fluoroquinolones: a national case-control safety study. Am J Health Syst Pharm 71:37–43. doi: 10.2146/ajhp130165. [DOI] [PubMed] [Google Scholar]
  • 10.Boggs SR, Cunnion KM, Raafat RH. 2011. Ceftriaxone-induced hemolysis in a child with Lyme arthritis: a case for antimicrobial stewardship. Pediatrics 128:e1289–1292. doi: 10.1542/peds.2010-1570. [DOI] [PubMed] [Google Scholar]
  • 11.Hensgens MP, Goorhuis A, Dekkers OM, Kuijper EJ. 2012. Time interval of increased risk for Clostridium difficile infection after exposure to antibiotics. J Antimicrob Chemother 67:742–748. doi: 10.1093/jac/dkr508. [DOI] [PubMed] [Google Scholar]
  • 12.Lapi F, Wilchesky M, Kezouh A, Benisty JI, Ernst P, Suissa S. 2012. Fluoroquinolones and the risk of serious arrhythmia: a population-based study. Clin Infect Dis 55:1457–1465. doi: 10.1093/cid/cis664. [DOI] [PubMed] [Google Scholar]
  • 13.Kerremans JJ, Verboom P, Stijnen T, Hakkaart-van Roijen L, Goessens W, Verbrugh HA, Vos MC. 2008. Rapid identification and antimicrobial susceptibility testing reduce antibiotic use and accelerate pathogen-directed antibiotic use. J Antimicrob Chemother 61:428–435. [DOI] [PubMed] [Google Scholar]
  • 14.Machen A, Drake T, Wang YF. 2014. Same day identification and full panel antimicrobial susceptibility testing of bacteria from positive blood culture bottles made possible by a combined lysis-filtration method with MALDI-TOF VITEK mass spectrometry and the VITEK2 system. PLoS One 9:e87870. doi: 10.1371/journal.pone.0087870. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Wang XP, Ginocchio CC. 2009. Automation of the BD GeneOhm methicillin-resistant Staphylococcus aureus assay for high-throughput screening of nasal swab specimens. J Clin Microbiol 47:1546–1548. doi: 10.1128/JCM.00114-09. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Paule SM, Hacek DM, Kufner B, Truchon K, Thomson RB Jr, Kaul KL, Robicsek A, Peterson LR. 2007. Performance of the BD GeneOhm methicillin-resistant Staphylococcus aureus test before and during high-volume clinical use. J Clin Microbiol 45:2993–2998. doi: 10.1128/JCM.00670-07. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Grobner S, Dion M, Plante M, Kempf VA. 2009. Evaluation of the BD GeneOhm StaphSR assay for detection of methicillin-resistant and methicillin-susceptible Staphylococcus aureus isolates from spiked positive blood culture bottles. J Clin Microbiol 47:1689–1694. doi: 10.1128/JCM.02179-08. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Deresinski S. 2011. Missing mec. Clin Infect Dis 53:iii–iv. doi: 10.1093/cid/cir648. [DOI] [PubMed] [Google Scholar]
  • 19.Davey DD, Neal MH, Wilbur DC, Colgan TJ, Styer PE, Mody DR. 2004. Bethesda 2001 implementation and reporting rates: 2003 practices of participants in the College of American Pathologists Interlaboratory Comparison Program in Cervicovaginal Cytology. Arch Pathol Lab Med 128:1224–1229. [DOI] [PubMed] [Google Scholar]
  • 20.Davey P, Brown E, Charani E, Fenelon L, Gould IM, Holmes A, Ramsay CR, Wiffen PJ, Wilcox M. 2013. Interventions to improve antibiotic prescribing practices for hospital inpatients. Cochrane Database Syst Rev 4:CD003543. [DOI] [PubMed] [Google Scholar]
  • 21.Malani AN, Richards PG, Kapila S, Otto MH, Czerwinski J, Singal B. 2013. Clinical and economic outcomes from a community hospital's antimicrobial stewardship program. Am J Infect Control 41:145–148. doi: 10.1016/j.ajic.2012.02.021. [DOI] [PubMed] [Google Scholar]
  • 22.van Belkum A, Dunne WM Jr. 2013. Next-generation antimicrobial susceptibility testing. J Clin Microbiol 51:2018–2024. doi: 10.1128/JCM.00313-13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Pulido MR, Garcia-Quintanilla M, Martin-Pena R, Cisneros JM, McConnell MJ. 2013. Progress on the development of rapid methods for antimicrobial susceptibility testing. J Antimicrob Chemother 68:2710–2717. doi: 10.1093/jac/dkt253. [DOI] [PubMed] [Google Scholar]
  • 24.Huang TH, Ning X, Wang X, Murthy N, Tzeng YL, Dickson RM. 2015. Rapid cytometric antibiotic susceptibility testing utilizing adaptive multidimensional statistical metrics. Anal Chem 87:1941–1949. doi: 10.1021/ac504241x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Mohan R, Mukherjee A, Sevgen SE, Sanpitakseree C, Lee J, Schroeder CM, Kenis PJ. 2013. A multiplexed microfluidic platform for rapid antibiotic susceptibility testing. Biosens Bioelectron 49:118–125. doi: 10.1016/j.bios.2013.04.046. [DOI] [PubMed] [Google Scholar]
  • 26.CLSI. 2008. Development of in vitro susceptibility testing criteria and quality control parameters; approved guideline, 3rd ed CLSI document M23-A3. Clinical and Laboratory Standards Institute, Wayne, PA. [Google Scholar]
  • 27.CLSI. 2016. Performance standards for antimicrobial susceptibility testing; twenty-fourth informational supplement. CLSI document M100-S26. Clinical and Laboratory Standards Institute, Wayne, PA. [Google Scholar]
  • 28.Murray J, Evans P, Hukins DW. 1980. Light-scattering methods for antibiotic sensitivity tests. J Clin Pathol 33:995–1001. doi: 10.1136/jcp.33.10.995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Conville PS, Witebsky FG, MacLowry JD. 1994. Antimicrobial susceptibilities of mycobacteria as determined by differential light scattering and correlation with results from multiple reference laboratories. J Clin Microbiol 32:1554–1559. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Berkman RM, Wyatt PJ, Phillips DT. 1970. Rapid detection of penicillin sensitivity in Staphylococcus aureus. Nature 228:458–460. doi: 10.1038/228458a0. [DOI] [PubMed] [Google Scholar]
  • 31.Minton AP. 2016. Recent applications of light scattering measurement in the biological and biopharmaceutical sciences. Anal Biochem 501:4–22. doi: 10.1016/j.ab.2016.02.007. [DOI] [PMC free article] [PubMed] [Google Scholar]