Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2017 May 1.
Published in final edited form as: IEEE J Sel Top Quantum Electron. 2015 Nov 2;22(3):6800911. doi: 10.1109/JSTQE.2015.2493962

Computed Optical Interferometric Imaging: Methods, Achievements, and Challenges

Fredrick A South 1, Yuan-Zhi Liu 2, P Scott Carney 3, Stephen A Boppart 4
PMCID: PMC5082437  NIHMSID: NIHMS763456  PMID: 27795663

Abstract

Three-dimensional high-resolution optical imaging systems are generally restricted by the trade-off between resolution and depth-of-field as well as imperfections in the imaging system or sample. Computed optical interferometric imaging is able to overcome these longstanding limitations using methods such as interferometric synthetic aperture microscopy (ISAM) and computational adaptive optics (CAO) which manipulate the complex interferometric data. These techniques correct for limited depth-of-field and optical aberrations without the need for additional hardware. This paper aims to outline these computational methods, making them readily available to the research community. Achievements of the techniques will be highlighted, along with past and present challenges in implementing the techniques. Challenges such as phase instability and determination of the appropriate aberration correction have been largely overcome so that imaging of living tissues using ISAM and CAO is now possible. Computed imaging in optics is becoming a mature technology poised to make a significant impact in medicine and biology.

Index Terms: Adaptive optics, biomedical optical imaging, computed imaging, high-resolution imaging, optical coherence tomography, synthetic aperture

I. Introduction

In recent decades, optical imaging has become an invaluable tool in medicine and biology. In particular, optical coherence tomography (OCT) has found widespread application in ophthalmology, where it is now the standard-of-care for retinal imaging [1, 2]. OCT is the optical analogue of ultrasound imaging, and uses backscattered light to generate three-dimensional structural maps of scattering samples of interest [3, 4]. In addition to ophthalmology, OCT has been successfully applied in various other clinical settings as well as non-biological materials imaging [5, 6]. Following a historical trend seen in other modalities such as X-ray imaging, magnetic resonance imaging, and radar, optical imaging is becoming increasingly computational. These computational optical imaging methods have recently become capable of reconstructing in vivo data, providing high-resolution aberration-free imaging of living biological tissues beyond what was previously achievable by standard OCT imaging alone [7, 8].

The use of computed imaging techniques has been inspired by two inherent limitations ubiquitous in optical imaging— restricted depth-of-field and wavefront aberrations. Limited depth-of-field reduces the usable imaging range, while optical aberrations of both the imaging system and the sample itself limit the achievable resolution and signal-to-noise ratio (SNR).

Various hardware solutions have been proposed to overcome these challenges. In particular, the trade-off between resolution and depth-of-field has been addressed in hardware by focus shifting [9] and multiple beam imaging [10]. Alternatively, a non-Gaussian imaging beam, such as a Bessel beam, can be used to extend the depth-of-field [11, 12]. Inspired by developments in astronomy [13, 14], hardware adaptive optics (HAO) has been used to correct aberrations in microscopy through the use of a wavefront sensor and deformable mirror to sense and negate wavefront distortions [15, 16]. However, HAO systems have not seen widespread use due to their inherent cost and complexity.

It is desirable to achieve high-resolution aberration-free volumetric imaging without the need for additional hardware. Interferometric detection provides access to both the amplitude and the phase of the backscattered light. Access to this phase information provides computational control of the full complex wavefront, enabling refocusing and correction of higher order optical aberrations. Computational depth-of-field solutions include numerical refocusing [1719], depth-encoded synthetic aperture microscopy [20, 21], holoscopy [22], and interferometric synthetic aperture microscopy [23]. Aberration correction can also be performed through manipulation of the spatial frequency data. This has been previously achieved in digital holographic microscopy for imaging of thin samples [24, 25], and recently in thick tissue samples using computational digital adaptive optics [26, 27].

Interferometric synthetic aperture microscopy (ISAM) is a solution to the inverse problem for optical microscopy that provides depth-independent transverse resolution, overcoming the traditional depth-of-field limitation without the need for additional hardware [23, 2830]. ISAM reconstruction achieves this through a Fourier domain resampling which provides three-dimensional defocus correction. ISAM is a point-scanning method, which avoids the cross-talk artifacts inherent in related full-field techniques [19, 22, 31]. It has been shown to be robust with respect to motion through the use of high scanning speeds or motion correction algorithms, and is parallelizable for real-time processing on a graphics processing unit (GPU) [7, 32]. Together, these features have contributed to the successful in vivo application of the ISAM technique.

Computational adaptive optics (CAO) extends the ISAM model to include optical aberrations which can be corrected through manipulation of the complex interferometric data [26]. This is achieved by multiplication with a phase-only aberration correction filter which negates the wavefront aberrations. In this way, volumetric aberration-free imaging can be achieved. This flexible processing method can be used to correct for arbitrary aberrations of high order [8, 33], and can be optimized for selected sample regions to overcome spatially varying aberrations [34]. CAO has recently been applied to in vivo imaging of the human retina for visualization of the photoreceptor mosaic [8].

This paper will explain both the theory and practical implementation of the ISAM and CAO methods. In addition, the various technical challenges encountered will be addressed along with proposed methods for overcoming these challenges. Key results will be summarized throughout the review, ending with discussion of current and future directions for these computational techniques.

II. Interferometric Synthetic Aperture Microscopy

This section provides an overview of the ISAM model as well as the necessary steps for implementing the reconstruction. Stability requirements and methods for overcoming phase instability are also discussed. Lastly, the capability of in vivo ISAM reconstruction is demonstrated.

A. Theoretical Model

ISAM is performed using low coherence interferometry. In summary, broadband near-infrared light is focused into the sample of interest, and the reflected light is interferometrically detected in the spectral domain. Point scanning in the transverse dimension provides a three-dimensional complex dataset S(x, y; k) which is a function of transverse position (x, y) and wavenumber k. Under the first Born approximation, the measured signal is given by the convolution of the sample scattering potential η(x, y, z) with the system point-spread function h(x, y, z; k) as

S(x,y;k)=h(xx,yy,z;k)η(x,y,z)dxdydz. (1)

The point-spread function of the system is defined as

h(x,y,z;k)=μrk2|P(k)|2g2(x,y,z;k), (2)

where μr is the interferometric power splitting ratio, |P(k)|2 is the power spectral density, and g2 (x, y, z; k) is the product of the identical Gaussian illumination and detection beams. The axial origin of the coordinate system is set as the focal plane z = 0 without loss of generality. Using the convolution theorem, Equation (1) can be rewritten in the spatial frequency domain as

S˜(Qx,Qy;k)=H(Qx,Qy,z;k)η˜(Qx,Qy,z)dz. (3)

where (~) represents the 2D Fourier transform and Q denotes spatial frequency.

In general, the frequency domain response is depth-dependent. However, using asymptotic approximations for both the near-focus and far-from-focus cases, Equation (3) can be simplified to the forward model

S˜(Qx,Qy;k)=H(Qx,Qy;k)η(Qx,Qy,Qz), (4)

where () represents the 3D Fourier transform and

Qz=4k2Qx2Qy2 (5)

The system response H(Qx, Qy; k) takes slightly different forms for the near- and far-from-focus cases, and in theory it must be inverted to fully recover the scattering potential (a detailed treatment of the system response function can be found in Ref. [30]). However, the system response is generally slowly varying within the pass-band and does not introduce significant image distortions [7, 30, 32]. Therefore, Equation (4) can be inverted through a Fourier domain resampling of the data according to the relationship

k=12Qx2+Qy2+Qz2. (6)

This recovers the unfiltered solution of the scattering potential in the frequency domain

η+(Qx,Qy,Qz)=S˜(Qx,Qy;k)|k=12Qx2+Qy2+Qz2, (7)

which can be returned to the spatial domain by the 3D inverse Fourier transform

η+(x,y,z)=F1{η+(Qx,Qy,Qz)}. (8)

The coordinate resampling step in Equation (7) represents the re-phasing of the three-dimensional spatial frequency content, and is comparable to Stolt mapping in synthetic aperture radar [35]. In addition, a helpful analogy can be drawn between ISAM and diffraction tomography [30, 36]. The measured spatial frequency vector Q is given by the difference of the output and input k-vectors kout and kin,

Q=koutkin. (9)

In the far-from-focus approximation, kout = −kin and therefore Q = −2kin. Considering the entire range of illumination angles within the imaging beam, the measured spatial frequencies trace out a wedge of radius 2k on the Ewald sphere [37]. Calculation of the square magnitude of Q gives the relationship

Qx2+Qy2+Qz2=4k2, (10)

which simplifies immediately to the ISAM coordinate resampling as in Equation (6).

B. Implementation

The ISAM imaging system is essentially identical to that used in OCT imaging, achieving resolutions of 1–10 µm and imaging depths of 1–3 mm in tissue. The key difference is that the numerical aperture (NA) may be increased without sacrificing depth-of-field, and further, a higher NA is often used to achieve higher transverse resolution. The NA used in ISAM typically ranges from 0.05 to 0.1 [7, 23]. Recently, ISAM was demonstrated at 0.6 NA for optical coherence microscopy, removing the need for z-stacking [38].

Detailed steps for the implementation of the ISAM reconstruction are listed in Table 1. The ISAM simulation for a point-scattering sample shown in Fig. 1 illustrates the data resampling process. The initial curvature of the data in the spatial frequency domain is corrected through appropriate resampling, resulting in a reconstruction with depth-independent resolution. Note that ISAM processing is in addition to any spectral-domain OCT processing steps such as k-linearization, apodization, and dispersion correction.

TABLE I.

3D ISAM Processing Steps

1. Begin with the dataset S(x, y; k).
2. Calculate the 1D inverse Fourier transform to retrieve the complex dataset
S(x, y, z), keeping only the positive depths z > 0.
3. If desired, zero-pad the signal along depth z to improve the subsequent
frequency-domain resampling.
4. Circularly shift along depth z to place the focal plane at z = 0.
5. Calculate the 3D Fourier transform to retrieve (Qx, Qy; k).
6. Resample (Qx, Qy; k) along k according to Equation (6) to retrieve the
frequency-domain scattering potential η͌+(Qx, Qy, Qz).
7. Calculate the 3D inverse Fourier transform of η͌+(Qx, Qy, Qz) to retrieve
the reconstructed scattering potential η+(x, y, z).

Fig. 1.

Fig. 1

Simulation of in focus and far-from-focus scattering particles. (a) Cross-section of the standard OCT reconstruction showing strong defocus. (b) ISAM reconstruction showing high-resolution throughout depth. (c) Frequency-domain representation of original dataset. Black line illustrates ISAM resampling curve. (d) Resampled frequency-domain data. Adapted from [26].

Alternatively, ISAM can be decomposed into two orthogonal 2D reconstructions for improved processing speed, which enables real-time reconstruction using a graphics processing unit (GPU) [7, 32]. In this scenario, each cross-sectional frame along the fast scanning axis is immediately reconstructed when acquired. Following acquisition of the entire volume, the cross-sectional frames are processed along the slow scanning axis for a complete 3D reconstruction.

ISAM has been validated in both scattering phantoms and biological tissue. OCT and ISAM reconstructions of a TiO2 scattering phantom are shown in Figs. 2(a) and 2(b). The limited depth-of-field is clear in the OCT rendering, while ISAM shows high-resolution throughout the volume. A plot of measured transverse resolution over depth is shown in Fig. 2(c). An en face plane approximately eight Rayleigh ranges (450 µm) above the focus was selected for comparison. The beam focus was then moved to the plane of interest and an additional OCT reconstruction was generated. The ISAM reconstruction of the out-of-focus plane was shown to match the structure of the in-focus OCT measurement [39].

Fig. 2.

Fig. 2

Comparison of OCT and ISAM reconstructions of TiO2 scattering phantoms imaged with an 800 nm, 0.05 NA OCT system. Transverse dimensions of the rendered volume are 360 µm by 360 µm. (a) 3D OCT reconstruction. (b) 3D ISAM reconstruction. (c) Transverse resolution throughout depth for OCT and ISAM experimentally demonstrating that ISAM achieves extended depth-of-field. Adapted from [23, 39].

ISAM also provides an accurate reconstruction of biological tissue structure. This was demonstrated in ex vivo human breast tissue where OCT and ISAM reconstructions were compared to the corresponding histology. As a result of high transverse resolution throughout the entire imaging depth, ISAM showed a stronger correspondence to histology [23]. These results illustrate the improved capability of ISAM to perform volumetric high-resolution imaging in biological tissue.

C. Phase stability

ISAM is a phase-sensitive technique which re-phases optical wavefronts via a coordinate resampling. Like other phase-sensitive imaging techniques, including CAO, a certain level of phase stability is required for the reconstruction to succeed [40]. To ensure phase stable data, it is necessary to both measure and correct the phase instabilities [41, 42].

Phase instabilities can be caused by motion of the sample or instability of the imaging system itself. Mechanical instability of the reference arm or scanning mirrors can introduce motion artifacts, and sample motion is often unavoidable in living tissues. In either case, it is necessary that the phase be stable over the interrogation time or interrogation length. The interrogation time is the duration for which the scanning Gaussian beam illuminates or “interrogates” a scattering particle. The interrogation time will be greater for particles far-from-focus, as the beam diameter is larger than at the focus. Therefore, phase stability requirements become stricter farther from focus.

Motion can occur along both the axial and transverse directions. The requirements for the axial dimension are defined in terms of the standard deviation of the phase in units of radians per frame. Requirements in the transverse dimension are defined in terms of transverse resolution elements per frame. The effect of phase instability on computational reconstruction is shown in Fig. 3 for both ex vivo tissue and scattering phantom data. For high scanning speeds, the phase variation is well below the threshold of approximately 0.3 radians per second. However, at the slowest scanning speed, the axial phase instability is too large to achieve an accurate reconstruction. This demonstrates that motion can be largely mitigated using high scanning speeds.

Fig. 3.

Fig. 3

ISAM reconstructions of en face planes and corresponding stability analysis at various imaging speeds for ex vivo tissue and a scattering phantom. Images acquired on an 860 nm, 0.6 NA OCT system. Graphs on the right show calculated stability for each imaging speed (horizontal lines) in comparison to previously determined stability requirements (black squares). The approximate interrogation length is indicated by the vertical black line. The axial motion at 2.6 FPS is above the required threshold, and therefore the corresponding ISAM reconstruction fails. Scalebars represent 85 µm for the tissue and 10 µm for the phantom. Reproduced from [41].

Both simulation and experiment have demonstrated that computational techniques are most sensitive to phase fluctuations along the axial dimension. The steps for measurement and correction of axial motion are described in Table 2 (see Ref. [42] for a detailed discussion of transverse motion correction.) Calculation of an en face axial motion phase map A(x, y) is used to negate the axial motion through complex conjugate multiplication with the complex data S(x, y, z). The result is phase stable data to which various computed imaging techniques can be applied. An representative motion map is shown in Fig. 4.

TABLE 2.

Axial Motion Correction Processing Steps

1. Begin with the complex OCT dataset S(x, y, z).
2. Calculate the axial phase shift across the slow axis (y) by complex
conjugate multiplication of adjacent frames S(x, yi, zS*(x, yi+1, z).
3. Depth-average the resulting complex signal over the depth (z) containing
the imaging sample, resulting in a two-dimensional axial motion map
A(x, y).
4. Set the phase difference to zero for the first frame. Calculate the cumulative
sum along the slow axis (y) to convert incremental phase changes to
cumulate phase changes.
5. Smooth the axial phase map using a mean filter along the fast axis (x).
7. Apply the motion map correction by complex conjugate multiplication at
each depth A*(x, yS(x, y, zi).

Fig. 4.

Fig. 4

Axial motion map resulting from the steps outlined in Table 2 showing smoothed cumulative phase variations. Adapted from [42].

Sample motion can be reduced through physically stabilizing the sample using mounting hardware. For skin imaging, this can be accomplished with an objective mount requiring physical contact between the sample arm hardware and the skin [7, 38, 41]. Unfortunately, this type of solution is not feasible for some important imaging applications such as retinal imaging. In these cases motion correction solutions must be applied.

Low SNR does not have the same destructive effect as phase instability. This has been verified in simulation and experiment where ISAM did not fail despite the effective SNR being decreased by as much as 18 dB [40].

D. In vivo imaging

By careful consideration of phase stability requirements it is possible to perform in vivo imaging [7, 42]. Figure 5 illustrates the results of computational reconstruction with and without motion correction. The sample was mounted separately from the imaging hardware and was therefore free to move. Without motion correction, the reconstruction clearly fails. After implementing three-dimensional motion correction, the reconstruction succeeds. These results suggest that the motion correction algorithms are robust enough to enforce phase stability in a variety of living tissues.

Fig. 5.

Fig. 5

Image reconstruction of in vivo human skin showing the spiral sweat duct structure acquired using a 1300 nm, 0.075 NA OCT system. Prior to phase correction the refocusing fails drastically. Following phase correction the refocused data shows enhanced resolution of the tissue structure. Transverse dimensions of the rendered volume are 250 µm × 250 µm. Adapted from [42].

A detailed comparison of in vivo OCT and ISAM imaging is shown in Fig. 6 showing skin images in the fingerprint region of the human index finger. This data was acquired under physical contact with an objective mount at approximately 100 frames per second, which mitigated sample motion. Therefore, no phase stabilization was required. Using the two-dimensional ISAM processing scheme summarized in section II-A, the data was processed and displayed in real-time using a GPU.

Fig. 6.

Fig. 6

Visualization of the in vivo sub-surface human fingerprint using OCT and ISAM acquired using a 1300 nm, 0.1 NA OCT system. Blue and yellow markers indicate the zoomed regions for OCT and ISAM, respectively. Top row: 3D rendering of spiral sweat duct structures. Middle row: En face plane showing a representative cross section of the sweat ducts. The measured diameter of the ISAM reconstruction is more anatomically correct. Bottom row: En face plane of the superficial dermis 780 µm into tissue. Scale bars represent 500 µm. Adapted from [7].

Anatomical structure was more accurately resolved throughout depth in the ISAM reconstruction. In addition to providing depth-independent resolution, ISAM also provided improvement in other image metrics. ISAM showed improved SNR away from focus when compared to OCT. Because depth-of-field was not a concern, the optical focus was placed deep into tissue, providing increased signal collection at greater depths. This improved both anisotropy [43] and image contrast. The high-quality volumetric visualization of tissue microstructure provided by ISAM could be of key clinical importance for diagnosis of skin disease [44, 45].

III. Computational Adaptive Optics

This section provides an overview of CAO theory as an extension of the ISAM model, including the typical CAO processing steps. Methods of determining the appropriate aberration correction are addressed. The combination of CAO and ISAM is demonstrated in both scattering phantoms and biological tissues.

A. Theory

Significant aberrations exist in many imaging scenarios due to imperfections of the imaging optics or the sample itself. These imperfections distort the phase of the optical wavefront and significantly degrade imaging quality. Examples of commonly encountered aberrations include astigmatism and spherical aberration.

Aberrations may be modeled as a phase-only linear filter which modifies the objective lens pupil function [26, 36]. Within the volume of stationarity V(x0, y0, z0) centered at position (x0, y0, z0), a particular wavefront description is valid. The corresponding aberration phase filter can be written as H(Qx, Qy; k)|V(x0, y0, z0). This filter can be applied to the previously derived ISAM model, and Equation (4) can be modified to

S˜A(Qx,Qy;k)|V(x0,y0,z0)=HA(Qx,Qy;k)|V(x0,y0,z0)×H(Qx,Qy;k)η(Qx,Qy,Qz). (11)

The aberration effects can be inverted through phase conjugation using the inverse filter

HAC(Qx,Qy;k)|V(x0,y0,z0)=HA*(Qx,Qy;k)|V(x0,y0,z0) (12)

which results in the aberration-free signal

S˜AC(Qx,Qy;k)|V(x0,y0,z0)=HAC(Qx,Qy;k)|V(x0,y0,z0)×S˜A(Qx,Qy;k)|V(x0,y0,z0) (13)

If the aberrations can be considered invariant over the transverse field-of-view, the aberration and correction filters can be simplified to HA(Qx, Qy; k)|V(z0) and HAC(Qx, Qy; k)|V(z0), respectively.

B. Implementation

The aberration correction filter HAC (Qx, Qy; k)|V(z0) is related to both the spatial frequency and spectral domains, so it can correct both monochromatic as well as chromatic aberrations [30, 33]. Under the assumption of an achromatic system, the k-dependence of the 3D aberration correction filter can be simplified to the monochromatic aberration function, HAC(Qx, Qy)|V(z0). The steps for applying CAO processing are summarized in Table 3. It is noted that if the aberrations vary considerably within the transverse field-of-view at a given depth, one can divide the image field into sub-regions and implement processing steps 2–5 on each sub-region. The aberration corrected sub-regions can then be recombined to retrieve the composite image for each depth [33]. In some special cases where the dominant artifacts are space-invariant along depth, the same aberration correction filter can be applied to the entire 3D volume to speed up the computation.

TABLE 3.

CAO Processing Steps

1. Begin with the complex OCT or ISAM dataset SA (x, y, z).
2. Calculate the 2D Fourier transform of en face image fields at each depth zi,
A(Qx, Qy, zi).
3. Calculate the aberration correction filter HAC(Qx, Qy)|zi for the image field.
4. Apply the aberration correction in the Fourier plane by complex
multiplication at each depth
AC (Qx, Qy, zi) = HAC (Qx, Qy)|zi × A (Qx, Qy, zi).
5. Calculate the 2D inverse Fourier transform of AC (Qx, Qy, zi) to retrieve the
aberration-free image field, SAC (x, y, zi).
6. Stack the en face image fields along the depth to obtain the aberration-free
volume SAC (x, y, z)

Figure 7 demonstrates aberration correction by applying CAO. The dataset was measured from a silicone phantom with subresolution TiO2 microparticles illuminated by a highly astigmatic beam. En face planes following standard OCT processing show the presence of two orthogonal line foci associated with astigmatism, as well as the plane of least confusion located halfway in between. After correcting the astigmatism, the plane of least confusion was restored as the focal plane and the line foci showed a Gaussian-like point-spread function (PSF). ISAM was then applied, reconstructing the entire volume with high-resolution. Imaging with the astigmatic beam also gave improved SNR far-from-focus as a result of the dual line foci. This suggests that imaging with an aberrated beam may actually be beneficial when the PSF can be computationally corrected.

Fig. 7.

Fig. 7

Demonstration of aberration correction in a scattering phantom using CAO. Images acquired on a 1300 nm, 0.1 NA OCT system. Select en face planes have been rendered to highlight the strong astigmatism. CAO restores the focal plane while ISAM improves the depth-of-field. Transverse dimensions of the rendered volume are 256 µm × 256 µm. Reproduced from [26].

Many biological samples have been successfully imaged using CAO such as ex vivo human breast tissue, in vitro 3D cell cultures, and various ex vivo animal tissues. As with ISAM, CAO can also be applied to in vivo imaging when stability concerns are considered. Figure 8 demonstrates this capability for three-dimensional in vivo imaging of human skin. Together, the ISAM and CAO methods complement one another to provide three-dimensional high-resolution images of biological tissue without the need for depth scanning or additional imaging hardware [26, 33, 38].

Fig. 8.

Fig. 8

High-resolution imaging of in vivo human skin acquired using an 860 nm, 0.6 NA OCM system. Arrows indicate (a, f) boundary of the stratum corneum and epidermis, (b, g) granular cell nuclei, (c, h) dermal papillae, (d, i) basal cells, and (e, j) connective tissue. Scale bar indicates 40 µm and applies to all images. Reproduced from [38].

C. Determining the correction filter

Because the optical aberrations of a particular imaging system are generally unknown, it can be a significant challenge to determine the appropriate correction filter. One solution is similar to sensor-less HAO techniques [46]. In sensor-less HAO, a sequence of aberration corrections

A technique termed subaperture correlation has also been proposed to digitally correct the aberrations in OCT data [27]. While CAO mimics the behavior of a deformable mirror, the subaperture technique simulates the operation of the Shack-Harman wavefront sensor by dividing the Fourier plane into several subapertures. The wavefront slopes of each subaperture are calculated from the cross-correlation of the reconstructed subaperture images to the center subaperture reference image, as illustrated in Fig. 9. This method has shown image improvement for scattering samples that have a uniform Fourier spectrum, although only low-order aberrations can be corrected.

Fig. 9.

Fig. 9

Illustration of the subaperture method for estimation of the aberration correction filter. Subapertures in the spatial-frequency domain are compared to the central reference subaperture to determine the slope of the wavefront. Reproduced from [27].

If the sample contains point-like structures, the so-called “guide-star” method can be used in CAO to correct not only the high-order monochromatic but also the chromatic aberrations [33]. A spatial window is used to capture the entire aberrated PSF of the guide-star, and a sub-resolution window is used to crop the guide-star to the desired target PSF. By isolating the signal from a guide-star, the aberrations can be sensed from the phase difference of the measured and desired wavefront in the Fourier domain. The aberration correction filter can then be digitally conjugated in the Fourier domain and applied to the whole image or 3D volume of stationarity containing the guide-star. Finally, after the inverse Fourier transform, the aberration-free data can be achieved.

IV. Current research

This section discusses current research directions in computed optical interferometric imaging. Efforts are focused on in vivo imaging of the human retina, which is currently the most widespread application of optical coherence tomography and adaptive optics. Imaging of the photoreceptors aids the diagnosis and understanding of diseases such as macular degeneration [48, 49]. Although it is possible to image the photoreceptors without adaptive optics provided the subject’s eye has very low aberrations [50, 51], in general this is not the case [52]. As the numerical aperture of the imaging beam is increased, aberrations play an increased role and distort the point-spread function [53, 54]. Because of this, HAO solutions have been used to achieve diffraction-limited imaging of the retina [16]. However, the cost and complexity of HAO has restricted its clinical application. A computational alternative, therefore, is highly desirable.

The eye exhibits frequent and unavoidable motion [55]. This motion poses a significant challenge for computational techniques and must be carefully considered. Unlike the skin, direct physical contact of the imaging hardware with the eye is not clinically feasible. Therefore, high scanning speeds and motion correction techniques must be implemented. Transverse motion correction can be avoided by imaging at cross-sectional frame rates greater than 1.5 kHz [56].

The first CAO images of the living human retina were recently reported using an en face OCT imaging system [8]. The use of en face OCT allowed for a higher effective frame rate in comparison to typical spectral-domain methods. The system operated at 4,000 en face lines per second, which for stability purposes is equivalent to a spectral-domain OCT system operating at a frame rate of 4 kHz. Even at this high imaging speed axial motion correction was required due to phase instabilities from the acoustic optic modulators and sample motion. The axial motion correction algorithm outlined in section II-C was used, slightly modified to work with en face data.

The subaperture and guide-star methods discussed in section III-C were implemented sequentially to determine the appropriate aberration correction filter for imaging the photoreceptor layer. The resulting aberration correction is shown in Fig. 10. The subaperture method was used to determine a bulk correction. The individual photoreceptors served as natural guide-stars to fine-tune the correction, modifying the Zernike polynomial terms of the bulk correction by up to 40%. The resulting photoreceptor mosaic is shown in Fig. 11. The CAO result clearly resolves the individual photoreceptors, gleaning valuable biological information from previously unintelligible data.

Fig. 10.

Fig. 10

Example wavefront correction for the human eye calculated using the subaperture and guide-star methods. Adapted from [8].

Fig. 11.

Fig. 11

Aberration correction in the living human retina using computational adaptive optics. (a) Scanning laser ophthalmoscope image of the retinal surface. N: nasal, S: superior. OCT images taken in the boxed region. (b) Uncorrected en face OCT data shows no recognizable features. (c) Aberration corrected data showing individual photoreceptors. (d) Simultaneously acquired cross-sectional OCT data showing the various retinal layers. The en face photoreceptor mosaic was acquired at the IS/OS junction. RNFL: retinal nerve fiber layer, GCL/IPL: ganglion cell layer/inner plexiform layer, INL: inner nuclear layer, OPL: outer plexiform layer, ONL: outer nuclear layer, IS/OS: inner segment/outer segment, RPE: retinal pigment epithelium. Colored boxes correspond to the matching zoomed regions. Scale bar represents 1° in (a) and 0.25° in (b–d). Reproduced from [8].

Visualization of other retinal layers also benefits from improved resolution. The health of the retinal nerve fiber layer (RNFL) is an important marker for diseases such as glaucoma [57, 58] and multiple sclerosis [59, 60]. These fibers are densely packed and difficult to resolve without aberration correction. CAO imaging of retinal fiber structure is shown in Fig. 12, acquired using the same en face imaging system. The aberration correction was manually adjusted to maximize the anisotropy image metric [43]. The previously hidden structure was clearly resolved.

Fig. 12.

Fig. 12

Computational adaptive optics of retinal fiber structure. (a) Scanning laser ophthalmoscope image of the retinal surface. The boxed region indicates the approximate OCT imaging location. (b) Uncorrected en face OCT data. (c) Data corrected using CAO showing improved resolution of the fiber structure. Scale bar represents approximately 2° in (a) and 0.2° in (b, c). Data previously unpublished.

Because HAO physically modifies the optical wavefront, the signal collected from the retina is greatly increased. However, CAO is a post-processing technique which can only redistribute the collected energy. Therefore, in subjects with severe eye disease where there is insufficient signal to generate an initial image, a purely computational solution may not be sufficient. In this case CAO could be supplemented with a low-order hardware correction to provide the necessary signal strength.

V. Conclusion

Together, ISAM and CAO are complementary computational methods which achieve high-resolution volumetric optical imaging in living tissue, overcoming both limited depth-of-field and optical aberrations. Moving forward, a primary goal as well as a key challenge will be the translation of these computed imaging techniques to the clinical setting. Computational techniques were previously restricted by phase stability requirements. Recently, however, the challenge of phase stabilization has been largely overcome through motion correction techniques. Implementation of these techniques, however, may vary across clinical imaging scenarios and existing clinical systems may or may not be suitable for phase-sensitive imaging. Aberration correction encounters the additional challenge of determining the appropriate correction filter. Several methods have been proposed, yet it remains to be determined which method will prove most useful for real-time applications.

As with any medical tool, the capability of the technology to improve the performance of clinical tasks must be determined. Since many features in the eye cannot be seen using the current standard-of-care, the impact of computational correction is expected to be significant. Application to other clinical areas such as dermatology and oncology may also benefit from the improved image quality. Future studies will involve the imaging of disease in collaboration with medical specialists to assess the true clinical impact of the computational techniques.

In summary, recent research results have demonstrated that computed optical interferometric imaging has matured to supplement, and in some applications, potentially even replace traditional optical tomography in medicine and biology.

Acknowledgments

The authors thank former members of the Biophotonics Imaging Laboratory who have helped pioneer, develop, and demonstrate these computed imaging techniques including Tyler Ralston, Dan Marks, Bryn Davis, Adeel Ahmad, Steven Adie, and Nathan Shemonski. The human subjects research was performed under protocols approved by the Institutional Review Board at the University of Illinois at Urbana-Champaign. Stephen Boppart and P. Scott Carney are co-founders of Diagnostic Photonics, Inc., which is licensing intellectual property from the University of Illinois at Urbana-Champaign related to ISAM. Stephen Boppart also received royalties from MIT related to patents related to OCT.

This work was supported in part by grants from the National Institutes of Health (1 r01 eb013723 and 1 r01 ca166309) and the National Science Foundation (CBET 14-45111 and 14-03660). p. s. Carney acknowledges funding in part with Federal funds from the National Cancer Institute, National Institutes of Health, Department of Health and Human Services, under Contract No. hhsn261201400044c. Additional information can be found at http://biophotonics.illinois.edu.

Biographies

graphic file with name nihms763456b1.gif

Fredrick A. South received the B.S. and M.S. degrees in electrical and computer engineering from the University of Illinois at Urbana-Champaign in 2010 and 2013, respectively. He is currently working toward the Ph.D. degree in electrical and computer engineering at the University of Illinois at Urbana-Champaign.

In 2010 he was an engineering intern at Siemens Healthcare in Hoffman Estates, IL where he focused on electrical engineering of SPECT and PET molecular imaging systems. Since 2011, he has been a research assistant in the Biophotonics Imaging Laboratory at the Beckman Institute for Advanced Science and Technology in Urbana, IL. His research interests include the development of novel optical imaging modalities for application in medicine and biology.

graphic file with name nihms763456b2.gif

Yuan-Zhi Liu received the B.S. degree in physics, in 2007, and the Ph.D. degree in optics, in 2012, both from Sun Yat-sen University, Guangzhou, China.

He was a Post-Doctoral researcher at Ohio State University, Columbus, Ohio, in 2012. Since 2013, he has been a Post-Doctoral Research Associate in the Biophotonics Imaging Laboratory, Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign. His current research interests include biomedical imaging, especially the development and applications of optical coherence tomography, interferometric synthetic aperture microscopy, hardware/computational adaptive optics, and wavefront shaping.

graphic file with name nihms763456b3.gif

Paul Scott Carney received the B.S. degree in engineering physics from the University of Illinois at Urbana-Champaign, Urbana, in 1994, and the Ph.D. degree from the University of Rochester, Rochester, NY, in 1999.

He was a Post-Doctoral Associate at Washington University, St. Louis, MO, from 1999 to 2001. He then joined the faculty of the Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign. His current research interests include inverse problems, imaging, coherence theory, and other branches of optical physics.

graphic file with name nihms763456b4.gif

Stephen A. Boppart (S’90–M’90–SM’06–F’11) was born in Harvard, IL, in 1968. He received the B.S. degree in electrical and bioengineering and the M.S. degree in electrical engineering from the University of Illinois at Urbana-Champaign, Urbana, IL, in 1990 and 1991, respectively, the Ph.D. degree in electrical and medical engineering from the Massachusetts Institute of Technology, Cambridge, in 1998, and the M.D. degree from Harvard Medical School, Boston, MA, in 2000.

He was a Research Scientist with the Air Force Laser Laboratory, Brooks Air Force Base, San Antonio, TX, where he was engaged in research on developing national (ANSI) and Air Force laser safety standards. Since 2000, he has been with the University of Illinois at Urbana-Champaign, Urbana, from where he completed residency training in internal medicine in 2005. He is currently an Able Bliss Professor of Engineering in the departments of Electrical and Computer Engineering, Bioengineering, and Medicine, the Head of the Biophotonics Imaging Laboratory, Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, and also the Director of Imaging at Illinois. He has authored or coauthored more than 300 invited and contributed publications, and more than 670 invited and contributed presentations. He holds more than 40 patents, filed or pending. His research interests include the development of novel optical imaging technologies for biological and medical applications, with particular emphasis on translating these to clinical applications.

Dr. Boppart is a Fellow of IEEE, AAAS, OSA, SPIE, and AIMBE. He is also a member of the Biomedical Engineering Society, the American Association for Cancer Research, and the American Medical Association. He was named one of the top 100 innovators in the world by the Technology Review Magazine for his research in medical technology, and received the IEEE Engineering in Medicine and Biology Society Early Career Achievement Award. He received the Paul F. Forman Engineering Excellence Award from the Optical Society of America for dedication and advancement in undergraduate research education, and recently, the international Hans Sigrist Prize for his work in diagnostic laser medicine.

Contributor Information

Fredrick A. South, Email: south1@illinois.edu, Beckman Institute for Advanced Science and Technology, also with the Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, IL 61801 USA.

Yuan-Zhi Liu, Email: yzliu@illinois.edu, Beckman Institute for Advanced Science and Technology, also with the Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, IL 61801 USA.

P. Scott Carney, Email: carney@illinois.edu, Beckman Institute for Advanced Science and Technology, also with the Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, IL 61801 USA.

Stephen A. Boppart, Email: boppart@illinois.edu, Beckman Institute for Advanced Science and Technology, also with the Departments of Electrical and Computer Engineering, Bioengineering, and Internal Medicine, University of Illinois at Urbana-Champaign, Urbana, IL 61801 USA.

References

  • 1.Huang D, Swanson EA, Lin CP, Schuman JS, Stinson WG, Chang W, Hee MR, Flotte T, Gregory K, Puliafito CA, Fujimoto JG. Optical coherence tomography. Science. 1991;254:1178–1181. doi: 10.1126/science.1957169. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Drexler JG, Fujimoto W. State-Of-The-Art Retinal Optical Coherence tomography. Prog. Retin. Eye. Res. 2008;27:45–88. doi: 10.1016/j.preteyeres.2007.07.005. [DOI] [PubMed] [Google Scholar]
  • 3.Schmitt JM. Optical coherence tomography (OCT): a review. IEEE J. Sel. Topics. Quantum Electron. 1999;5:1205–1215. [Google Scholar]
  • 4.Tomlins RK, Wang PH. Theory, developments and applications of optical coherence tomography. J. Phys. D: Appl. Phys. 2005;38:2519–2539. [Google Scholar]
  • 5.Zysk AM, Nguyen FT, Oldenburg AL, Marks DL, Boppart SA. Optical coherence tomography: a review of clinical development from bench to bedside. J. Biomed. Opt. 2007;12:051403. doi: 10.1117/1.2793736. [DOI] [PubMed] [Google Scholar]
  • 6.Stifter D. Beyond biomedicine: a review of alternative applications and developments for optical coherence tomography. Appl. Phys. B. 2007;88:337–357. [Google Scholar]
  • 7.Ahmad A, Shemonski ND, Adie SG, Kim HS, Hwu WMW, Carney PS, Boppart SA. Real-time in vivo computed optical interferometric tomography. Nature Photon. 2013;7:445–449. doi: 10.1038/nphoton.2013.71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Shemonski ND, South FA, Liu Y-Z, Adie SG, Scott Carney P, Boppart SA. Computational high-resolution optical imaging of the living human retina. Nature Photon. 2015;9:440–443. doi: 10.1038/NPHOTON.2015.102. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Lee K-S, Zhao H, Ibrahim SF, Meemon N, Khoudeir L, Rolland JP. Three-dimensional imaging of normal skin and nonmelanoma skin cancer with cellular resolution using Gabor domain optical coherence microscopy. J. Biomed. Opt. 2012;17:126006. doi: 10.1117/1.JBO.17.12.126006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Beau AS, Kenneth KCL, Adrian M, Nigel RM, Michael KKL, Victor XDY, Vitkin IA. In vivo endoscopic multi-beam optical coherence tomography. Phys. Med. Biol. 2010;55:615. doi: 10.1088/0031-9155/55/3/004. [DOI] [PubMed] [Google Scholar]
  • 11.Ding ZH, Ren HW, Zhao YH, Nelson JS, Chen ZP. High-resolution optical coherence tomography over a large depth range with an axicon lens. Opt. Lett. 2002;27:243–245. doi: 10.1364/ol.27.000243. [DOI] [PubMed] [Google Scholar]
  • 12.Blatter C, Weingast J, Alex A, Grajciar B, Wieser W, Drexler W, Huber R, Leitgeb RA. In situ structural and microangiographic assessment of human skin lesions with high-speed OCT. Biomed. Opt. Express. 2012;3:2636–2646. doi: 10.1364/BOE.3.002636. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Beckers JM. Adaptive optics for astronomy: principles, performance, and applications. Annu. Rev. Astron. Astrophys. 1993;31:13–62. [Google Scholar]
  • 14.Davies R, Kasper M. Adaptive optics for astronomy. Annu. Rev. Astron. Astrophys. 2012;50:305–351. [Google Scholar]
  • 15.Liang J, Williams DR, Miller DT. Supernormal vision and high-resolution retinal imaging through adaptive optics. J. Opt. Soc. Am. A. 1997;14:2884–2892. doi: 10.1364/josaa.14.002884. [DOI] [PubMed] [Google Scholar]
  • 16.Zhang Y, Cense B, Rha J, Jonnal RS, Gao W, Zawadzki RJ, Werner JS, Jones S, Olivier S, Miller DT. High-speed volumetric imaging of cone photoreceptors with adaptive optics spectral-domain optical coherence tomography. Opt. Express. 2006;14:4380–4394. doi: 10.1364/OE.14.004380. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Yasuno Y, Sugisaka J-i, Sando Y, Nakamura Y, Makita S, Itoh M, Yatagai T. Non-iterative numerical method for laterally superresolving Fourier domain optical coherence tomography. Opt. Express. 2006;14:1006–1020. doi: 10.1364/oe.14.001006. [DOI] [PubMed] [Google Scholar]
  • 18.Yu L, Rao B, Zhang J, Su J, Wang Q, Guo S, Chen Z. Improved lateral resolution in optical coherence tomography by digital focusing using two-dimensional numerical diffraction method. Opt. Express. 2007;15:7634–7641. doi: 10.1364/oe.15.007634. [DOI] [PubMed] [Google Scholar]
  • 19.Kumar A, Drexler W, Leitgeb RA. Numerical focusing methods for full field OCT: a comparison based on a common signal model. Opt. Express. 2014;22:16061–16078. doi: 10.1364/OE.22.016061. [DOI] [PubMed] [Google Scholar]
  • 20.Mo J, de Groot M, de Boer JF. Focus-extension by depth-encoded synthetic aperture in optical coherence tomography. Opt. Express. 2013;21:10048–10061. doi: 10.1364/OE.21.010048. [DOI] [PubMed] [Google Scholar]
  • 21.Mo J, de Groot M, de Boer JF. Depth-encoded synthetic aperture optical coherence tomography of biological tissues with extended focal depth. Opt. Express. 2015;23:4935–4945. doi: 10.1364/OE.23.004935. [DOI] [PubMed] [Google Scholar]
  • 22.Hillmann D, Franke G, Lührs C, Koch P, Hüttmann G. Efficient holoscopy image reconstruction. Opt. Express. 2012;20:21247–21263. doi: 10.1364/OE.20.021247. [DOI] [PubMed] [Google Scholar]
  • 23.Ralston TS, Marks DL, Carney PS, Boppart SA. Interferometric synthetic aperture microscopy. Nature Phys. 2007;3:129–134. doi: 10.1038/nphys514. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Colomb T, Montfort F, Kühn J, Aspert N, Cuche E, Marian A, Charrière F, Bourquin S, Marquet P, Depeursinge C. Numerical parametric lens for shifting, magnification, and complete aberration compensation in digital holographic microscopy. J. Opt. Soc. Am. A. 2006;23:3177–3190. doi: 10.1364/josaa.23.003177. [DOI] [PubMed] [Google Scholar]
  • 25.Miccio L, Alfieri D, Grilli S, Ferraro P, Finizio A, De Petrocellis L, Nicola SD. Direct full compensation of the aberrations in quantitative phase microscopy of thin objects by a single digital hologram. Appl. Phys. Lett. 2007;90:041104. [Google Scholar]
  • 26.Adie SG, Graf BW, Ahmad A, Carney PS, Boppart SA. Computational adaptive optics for broadband optical interferometric tomography of biological tissue. Proc. Natl. Acad. Sci. U.S.A. 2012;109:7175–7180. doi: 10.1073/pnas.1121193109. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Kumar A, Drexler W, Leitgeb RA. Subaperture correlation based digital adaptive optics for full field optical coherence tomography. Opt. Express. 2013;21:10850–10866. doi: 10.1364/OE.21.010850. [DOI] [PubMed] [Google Scholar]
  • 28.Ralston TS, Marks DL, Carney PS, Boppart SA. Inverse scattering for optical coherence tomography. J. Opt. Soc. Am. A. 2006;23:1027–1037. doi: 10.1364/josaa.23.001027. [DOI] [PubMed] [Google Scholar]
  • 29.Ralston TS, Marks DL, Boppart SA, Carney PS. Inverse scattering for high-resolution interferometric microscopy. Opt. Lett. 2006;31:3585–3587. doi: 10.1364/ol.31.003585. [DOI] [PubMed] [Google Scholar]
  • 30.Davis BJ, Schlachter SC, Marks DL, Ralston TS, Boppart SA, Carney PS. Nonparaxial vector-field modeling of optical coherence tomography and interferometric synthetic aperture microscopy. J. Opt. Soc. Am. A. 2007;24:2527–2542. doi: 10.1364/josaa.24.002527. [DOI] [PubMed] [Google Scholar]
  • 31.Marks DL, Ralston TS, Boppart SA, Carney PS. Inverse scattering for frequency-scanned full-field optical coherence tomography. J. Opt. Soc. Am. A. 2007;24:1034–1041. doi: 10.1364/josaa.24.001034. [DOI] [PubMed] [Google Scholar]
  • 32.Ralston TS, Marks DL, Carney PS, Boppart SA. Realtime interferometric synthetic aperture microscopy. Opt. Express. 2008;16:2555–2569. doi: 10.1364/oe.16.002555. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Adie SG, Shemonski ND, Graf BW, Ahmad A, Scott Carney P, Boppart SA. Guide-star-based computational adaptive optics for broadband interferometric tomography. Appl. Phys. Lett. 2012;101:221117. doi: 10.1063/1.4768778. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Kumar A, Kamali T, Platzer R, Unterhuber A, Drexler W, Leitgeb RA. Anisotropic aberration correction using region of interest based digital adaptive optics in Fourier domain OCT. Biomed. Opt. Express. 2015;6:1124–1134. doi: 10.1364/BOE.6.001124. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Davis BJ, Marks DL, Ralston TS, Carney PS, Boppart SA. Interferometric synthetic aperture microscopy: computed imaging for scanned coherent microscopy. Sensors. 2008;8:3903. doi: 10.3390/s8063903. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Adie SG, Shemonski ND, Ralston TS, Carney PS, Boppart SA. Interferometric Synthetic Aperture Microscopy (ISAM) In: Drexler W, Fujimoto JG, editors. Optical Coherence Tomography: Technology and Applications. 2nd. Springer; 2015. [Google Scholar]
  • 37.Born M, Wolf E. Principles of Optics: Electromagnetic Theory of Propagation, Interference and Diffraction of Light. 7th. Cambridge, UK: Cambridge University Press; 1999. [Google Scholar]
  • 38.Liu Y-Z, Shemonski ND, Adie SG, Ahmad A, Bower AJ, Carney PS, Boppart SA. Computed optical interferometric tomography for high-speed volumetric cellular imaging. Biomed. Opt. Express. 2014;5:2988–3000. doi: 10.1364/BOE.5.002988. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Ralston TS, Adie SG, Marks DL, Boppart SA, Carney PS. Cross-validation of interferometric synthetic aperture microscopy and optical coherence tomography. Opt. Lett. 2010;35:1683–1685. doi: 10.1364/OL.35.001683. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Shemonski ND, Adie SG, Liu Y-Z, South FA, Carney PS, Boppart SA. Stability in computed optical interferometric tomography (Part I): Stability requirements. Opt. Express. 2014;22:19183–19197. doi: 10.1364/OE.22.019183. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Shemonski ND, Ahmad A, Adie SG, Liu Y-Z, South FA, Carney PS, Boppart SA. Stability in computed optical interferometric tomography (Part II): in vivo stability assessment. Opt. Express. 2014;22:19314–19326. doi: 10.1364/OE.22.019314. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Shemonski ND, Ahn SS, Liu Y-Z, South FA, Carney PS, Boppart SA. Three-dimensional motion correction using speckle and phase for in vivo computed optical interferometric tomography. Biomed. Opt. Express. 2014;5:4131–4143. doi: 10.1364/BOE.5.004131. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Gabarda S, Cristóbal G. Blind image quality assessment through anisotropy. J. Opt. Soc. Am. A. 2007;24:B42–B51. doi: 10.1364/josaa.24.000b42. [DOI] [PubMed] [Google Scholar]
  • 44.Welzel J. Optical coherence tomography in dermatology: a review. Skin Res. Technol. 2001;7:1–9. doi: 10.1034/j.1600-0846.2001.007001001.x. [DOI] [PubMed] [Google Scholar]
  • 45.Gambichler T, Moussa G, Sand M, Sand D, Altmeyer P, Hoffmann K. Applications of optical coherence tomography in dermatology. J. Dermatol. Sci. 2005;40:85–94. doi: 10.1016/j.jdermsci.2005.07.006. [DOI] [PubMed] [Google Scholar]
  • 46.Hofer H, Sredar N, Queener H, Li CH, Porter J. Wavefront sensorless adaptive optics ophthalmoscopy in the human eye. Opt. Express. 2011;19:14160–14171. doi: 10.1364/OE.19.014160. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Fienup JR, Miller JJ. Aberration correction by maximizing generalized sharpness metrics. J. Opt. Soc. Am. A. 2003;20:609–620. doi: 10.1364/josaa.20.000609. [DOI] [PubMed] [Google Scholar]
  • 48.Duncan JL, Zhang YH, Gandhi J, Nakanishi C, Othman M, Branham KEH, Swaroop A, Roorda A. High-resolution imaging with adaptive optics in patients with inherited retinal degeneration. Invest. Ophthalmol. Vis. Sci., IOVS. 2007;48:3283–3291. doi: 10.1167/iovs.06-1422. [DOI] [PubMed] [Google Scholar]
  • 49.Zayit-Soudry S, Duncan JL, Syed R, Menghini M, Roorda AJ. Cone structure imaged with adaptive optics scanning laser ophthalmoscopy in eyes with nonneovascular age-related macular degeneration. Invest. Ophthalmol. Vis. Sci., IOVS. 2013;54:7498–7509. doi: 10.1167/iovs.13-12433. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Miller DT, Williams DR, Morris GM, Liang J. Images of cone photoreceptors in the living human eye. Vision Res. 1996;36:1067–1079. doi: 10.1016/0042-6989(95)00225-1. [DOI] [PubMed] [Google Scholar]
  • 51.Pircher M, Baumann B, Götzinger E, Hitzenberger CK. Retinal cone mosaic imaged with transverse scanning optical coherence tomography. Opt. Lett. 2006;31:1821–1823. doi: 10.1364/ol.31.001821. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Thibos LN, Hong X, Bradley A, Cheng X. Statistical variation of aberration structure and image quality in a normal population of healthy eyes. J. Opt. Soc. Am. A. 2002;19:2329–2348. doi: 10.1364/josaa.19.002329. [DOI] [PubMed] [Google Scholar]
  • 53.Liang J, Williams DR. Aberrations and retinal image quality of the normal human eye. J. Opt. Soc. Am. A. 1997;14:2873–2883. doi: 10.1364/josaa.14.002873. [DOI] [PubMed] [Google Scholar]
  • 54.Williams DR. Imaging single cells in the living retina. Vision Res. 2011;51:1379–1396. doi: 10.1016/j.visres.2011.05.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Martinez-Conde S, Macknik SL, Hubel DH. The role of fixational eye movements in visual perception. Nat. Rev. Neurosci. 2004;5:229–240. doi: 10.1038/nrn1348. [DOI] [PubMed] [Google Scholar]
  • 56.Shemonski ND, Adie SG, Liu Y-Z, South FA, Carney PS, Boppart SA. A computational approach to high-resolution imaging of the living human retina without hardware adaptive optics. Proc. SPIE 9307, Ophthalmic Technologies XXV. 2015:930710. [Google Scholar]
  • 57.Kocaoglu OP, Cense B, Jonnal RS, Wang Q, Lee S, Gao W, Miller DT. Imaging retinal nerve fiber bundles using optical coherence tomography with adaptive optics. Vision Res. 2011;51:1835–1844. doi: 10.1016/j.visres.2011.06.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Takayama K, Ooto S, Hangai M, Ueda-Arakawa N, Yoshida S, Akagi T, Ikeda HO, Nonaka A, Hanebuchi M, Inoue T, Yoshimura N. High-resolution imaging of retinal nerve fiber bundles in glaucoma using adaptive optics scanning laser ophthalmoscopy. Am. J. Ophthalmol. 2013;155:870–881. doi: 10.1016/j.ajo.2012.11.016. [DOI] [PubMed] [Google Scholar]
  • 59.Fisher JB, Jacobs DA, Markowitz CE, Galetta SL, Volpe NJ, Nano-Schiavi ML, Baier ML, Frohman EM, Winslow H, Frohman TC, Calabresi PA, Maguire MG, Cutter GR, Balcer LJ. Relation of visual function to retinal nerve fiber layer thickness in multiple sclerosis. Ophthalmology. 2006;113:324–332. doi: 10.1016/j.ophtha.2005.10.040. [DOI] [PubMed] [Google Scholar]
  • 60.Saidha S, Syc SB, Ibrahim MA, Eckstein C, Warner CV, Farrell SK, Oakley JD, Durbin MK, Meyer SA, Balcer LJ, Frohman EM, Rosenzweig JM, Newsome SD, Ratchford JN, Nguyen QD, Calabresi PA. Primary retinal pathology in multiple sclerosis as detected by optical coherence tomography. Brain. 2011;134:518–533. doi: 10.1093/brain/awq346. [DOI] [PubMed] [Google Scholar]

RESOURCES