Abstract
Endovascular stents are the mainstay of interventional cardiovascular medicine. Technological advances have reduced biological and clinical complications but not mechanical failure. Stent strut fracture is increasingly recognized as of paramount clinical importance. Though consensus reigns that fractures can result from material fatigue, how fracture is induced and the mechanisms underlying its clinical sequelae remain ill-defined. In this study, strut fractures were identified in the prospectively maintained Food and Drug Administration's (FDA) Manufacturer and User Facility Device Experience Database (MAUDE), covering years 2006–2011, and differentiated based on specific coronary artery implantation site and device configuration. These data, and knowledge of the extent of dynamic arterial deformations obtained from patient CT images and published data, were used to define boundary conditions for 3D finite element models incorporating multimodal, multi-cycle deformation. The structural response for a range of stent designs and configurations was predicted by computational models and included estimation of maximum principal, minimum principal and equivalent plastic strains. Fatigue assessment was performed with Goodman diagrams and safe/unsafe regions defined for different stent designs. Von Mises stress and maximum principal strain increased with multimodal, fully reversed deformation. Spatial maps of unsafe locations corresponded to the identified locations of fracture in different coronary arteries in the clinical database. These findings, for the first time, provide insight into a potential link between patient adverse events and computational modeling of stent deformation. Understanding of the mechanical forces imposed under different implantation conditions may assist in rational design and optimal placement of these devices.
Keywords: Stent fracture, Arterial deformation, Finite element analysis
Introduction
Stents now account for greater than 95% of all endovascular interventions. Advances in materials science, pharmacology and controlled drug delivery have nearly eliminated stent restenosis from tissue hyperplasia. Yet, while biological and pathological complications are reduced there remains mechanical device failure, especially strut fracture. With the loss of overriding tissue from the reduction in intimal hyperplasia offered by controlled drug delivery, exposed and penetrating struts take on ever more important impact on the biological response to these implants. Stent strut fracture is an increasingly recognized phenomenon, with rates ranging from 2.6%, for severe, transectional disruption on angiography,1 to 29% when all types of fracture are included at autopsy.24,37 Clinical investigations have demonstrated that fracture rates rise with stent and lesion length,23,24,37,42,43 multiple, overlapping stents,10,17,20,23,37,43 sirolimus release,39 calcified complex lesions,20,30 relative vessel angulation and “hinge” motion,24,42 higher balloon inflation pressures,35,45 and right coronary artery (RCA) implantation,1,39 in particular in the ostium.42 When fracture occurs it is associated with higher rates of in-stent restenosis,10,24,39,44,51 thrombosis,7,37,40 and target vessel revascularization10,24,39,42,51—outcomes which may be exacerbated by more severe, transectional fractures.37
Though there is a growing consensus that a proportion of these fractures result from mechanical fatigue,20 what remains ill-defined is how these observational associations lead to increased mechanical stresses and strains as well as the mechanisms underlying fracture initiation of clinical relevance. The stresses and strains encountered in native and stented arteries are poorly understood and are only now being characterized.9,15,29,53 Understanding of all these issues would allow for more precise delineation of rates of fracture, more focused evaluation and assessment of secondary intervention and allow for primary interventions to minimize and reduce the occurrence of clinically significant outcomes associated with stent strut fracture.
The structural performance of stents can be assessed virtually using finite element analysis (FEA). This performance is generally assessed in terms of two major phases: the implantation, short term behavior, and the fatigue response of the device, long term behavior. Considerable attention has been given to implantation behavior and multiple assessments of mechanical characteristics of crimping, deliverability, scaffolding, radial strength, conformability have been reported. The majority of studies focused on deployment and modeling the immediate tissue interactions therein.11,12,19,26,27 Few have concentrated on why devices fail long after initial implantation, the role of mechanical material fatigue and integrating the highly dynamic in vivo environment in which these devices reside. Further, those studies that assess the fatigue response of stents, coronary or peripheral, tend to be limited to consideration of isolated loading modes.
Studies of peripheral stents have focused on long term fatigue response under isolated bending and compression and have largely avoided the clinically observed torsional component. Dordoni et al.16 examined fatigue of a Nitinol superficial femoral arterial (SFA) stent subjected to axial compression and bending with varied stenosis geometries—each cyclic loading mode was examined separately. The FEA results were compared with the Nitinol fatigue strain limit curve, derived from the experimental work of Pelton et al.41 The predictions did not indicate fatigue failure would occur under the simulated loading conditions, in contrast to clinically observed fracture rates of ∼ 38% in Nitinol SFA stents.23,43 Early and Kelly18 also examined the effect of vessel bending and compression, applied separately, on the fatigue life of stainless steel and Nitinol stents. Goodman analysis performed for the former devices plotted the stress amplitude and mean stress for each integration point of the stent elements and indicated a propensity for mechanical failure. In contrast, Nitinol stents were not predicted to fail when a factor of safety was calculated by comparing maximum strain amplitude from FEA with an experimental strain amplitude.41 Hsiao and Ying22 used a similar approach where different Nitinol peripheral stent designs under pulsatile pressure loadings were modeled and analyzed with a Goodman analysis, in terms of a fatigue safety factor (FSF). Their computational predictions showed improved FSFs with design alterations but again did not predict fatigue failure. Meoli et al.33 used FEA and Goodman analysis to model previously published in vitro fatigue tests of two peripheral stent designs, subjected to either axial compression or bending in air or a constraining tube, and identified relative device performance and predicted fatigue failure at 107 cycles, less than a few months of service time at physiological frequencies.
Coronary stents are for the most part stainless steel or chrome alloys and studies of these devices have largely been restricted to analyses of expansion and radial pulsation. Marrey et al.31 used a fracture mechanics based approach to predict the fatigue life of cobalt chromium coronary stents, virtually subjected to a pulsatile radial loading. Though a global stent model post-processed using Goodman analysis did not predict fatigue failure, a local “submodel” of a stent region with postulated flaws analyzed using fracture mechanics techniques defined life of the device in terms of initial flaw size. Li et al.28 reported FEA and experimental results of fatigue tests under pulsatile pressure. While high conformity was achieved between the experimental results and FEA predictions, fatigue failure of the device was not shown. Argente dos Santos et al.2 similarly used a two-scale plasticity damage model with coronary stents virtually subjected to pulsatile pressure loading. Microcracks were predicted to initiate after ∼53–64 million cycles. Barrera et al.5 presented a fatigue life assessment approach, based on the Dang Van high cycle criterion for coronary stents subjected to cyclic pressure, and even the most critically stressed element did not reach fatigue failure throughout the loading cycle. In a similar study, Azaouzi et al.4 reported fatigue life of pressure loaded coronary stents and post-processed using a Goodman analysis. Fatigue failure was predicted in the extreme case of some 50% expansion of the surrounding idealized artery. Sweeney et al.47,50 used crystal plasticity constitutive theory in a micromechanical fatigue assessment of cobalt chromium stents developing on the work of McGarry et al.32 2D and ultimately 3D representative unit cells allowed examination of grain size and orientation on the fatigue life of the device for low cycle49 and high cycle fatigue.48 Morlacchi et al.36 investigated the effects of local calcifications on the coronary stent fatigue failure and were the first to include cardiac wall movement which showed that inclusion of cardiac wall movement significantly affected the Goodman analysis results, when compared to pressurization alone. Recently, Auricchio et al.3 presented a computational framework for the lifetime prediction of coronary stents following stent implantation due to pulsation and minor bending angles. The fatigue assessment performed showed that bending significantly reduced predicted lifetime assessment of the device.
All of these works have significantly contributed to our understanding of fatigue behavior of endovascular stents, but they have not explained the high rates of clinically observed fractures, or realistically represented the complex dynamic loading environment in which these devices reside. This work seeks to link the observed complex loading environment with the mechanical state experienced by the stent. The objective is to gain insight into stent fracture through analysis of the arterial locations and device features most often associated with fracture in patients through the use of predictive finite element models of device-level stress and strain under representative multimodal and multi-cycle deformations.
Methods
Prospective Analysis of Stent Strut Fracture Adverse Events in Patients
Adverse event reports of stent fracture were extracted from the FDA MAUDE database over years 2006–2011. The MAUDE database represents prospectively collected reports from clinicians, manufacturers, and patients of adverse events that meet reportability criteria defined as death and serious injury caused or contributed to by a medical device.6 An initial search of coronary bare metal stent (BMS) and drug eluting stent (DES) records for the terms “strut” and “fracture”, followed by manual sorting to remove duplicates and inappropriate results returned 28 BMS and 481 DES reports as valid incidences of stent strut fracture (Table 1). Each report was subsequently analyzed for device and procedural information, which may or may not have been included in the text of the adverse event report (e.g., information was excluded or provided in attached article), including: patient age, implantation duration, target lesion, lesion dimensions, brand name, stent diameter and length, number stents implanted, number fractures, fracture location(s) and severity, and presence and length of overlap. Reports were compared against sex-matched controls selected from adverse event reports from 2006 to 2011 that were excluded as cases of strut fracture.
Table 1. Baseline demographic characteristics and implantation conditions associated with stent fracture.
Characteristic | Fx+ (n = 509) | Fx− (n = 506) | p value |
---|---|---|---|
Demographics | |||
Age, mean ± SE | 61.4 ± 0.74 | 66.7 ± 0.77 | <0.001 |
Women [n (%)] | 80 (23.7%) | 81 (28.8%) | 0.17 |
Stent dimensions, mean ± SE | |||
Stent diameter, mm | 3.0 ± 0.02 | 2.9 ± 0.02 | <0.001 |
Stent length, mm | 24.2 ± 0.36 | 19.5 ± 0.31 | <0.001 |
Stents implanted, mean ± SE | 2.0 ± 0.06 | 1.1 ± 0.04 | <0.001 |
Stents in vessel [n (%)] | |||
1 | 255 (50.3%) | 331 (67.4%) | <0.001 |
> 1 | 252 (49.7%) | 68 (13.8%) | <0.001 |
Overlapped stents [n (%)]* | 176 (69.8%) | 11 (16.2%) | <0.001 |
Stent-in-stent [n (%)]* | 35 (13.9%) | 10 (14.7%) | 0.71 |
Vessels stented, mean ± SE | 1.2 ± 0.02 | 0.9 ± 0.02 | <0.001 |
Target vessel [n (%)] | |||
RCA | 263 (52.2%) | 125 (28.4%) | <0.001 |
LAD | 150 (29.8%) | 143 (32.5%) | 0.40 |
LCX | 34 (6.7%) | 80 (18.2%) | <0.001 |
LM | 4 (0.8%) | 13 (3.0%) | 0.01 |
Diagonal (LAD) | 7 (1.4%) | 19 (4.3%) | 0.008 |
Marginal (LCX) | 8 (1.6%) | 24 (5.5%) | 0.002 |
Other | 38 (7.5%) | 36 (8.2%) | 0.72 |
Values are mean ± SE or n(%). Fx+ = Patients with adverse events and stent fracture, Fx− = Patients with adverse events not including stent fracture.
Overlapped and stent-in-stent values reported as a percentage of those reports where multiple stents were employed. LAD Left coronary artery, LCX left circumflex, LM left marginal, RCA right coronary artery.
Image Acquisition and Deformation Analysis
Cardiac-gated coronary arterial trees were imaged with 64-slice computed tomography (CT) (Light-Speed® VCT 64, GE Healthcare, WI, USA) in seven anonymized patients with prior stent placement, and two patients with native arteries. Patients were retrospectively selected from existing case library and data sets with artifacts at the stented segment that precluded full imaging were excluded from this study. In some patients with stents, in-stent restenosis was observed with calcifications at the stent edges. The average spatial resolution was 0.395 × 0.395 × 0.625 mm3 (matrix size: 512 × 512 × 177). Diastole and systole were visually selected by evaluation of ventricular volumes from nine cardiac frames. Vessel shortening (compression), bending, and torsion were extracted in native vessels. Bending deformation was measured by computing the change in curvature of each coronary section over the cardiac cycle. The curvature was measured by computing a best-fit torus to the surface model of the coronary artery.8 Briefly, a centerline path was first constructed by collecting centroids of cross-sectional lumen along an initial geodesic path on the surface model based on the constraints of minimum cross-sectional area. Then torus parameters (center, radius of torus plane, radius of torus ring, normal vector of torus plane) were optimally computed by minimizing the standard deviation of distances from the surface mesh to the centerline of the torus. For initial values of optimization, the radius and normal plane of the circumscribed circle over the stented region were utilized.9
Finite Element Models and Goodman Fatigue Analysis
A full description of the finite element and Goodman Fatigue analysis is provided in Supplementary Information. Briefly, 3D stent geometries for delta wing, corrugated ring and multilink designs were created in straight deployed states as well as increasingly long stents, overlapped and stent-in-stent configurations utilizing the delta wing design. The stent material was modeled as solid, homogeneous, elastoplastic 316L stainless steel with strain hardening and dimension-dependent plastic strain.
Deformations were acquired from cardiac-gated CT images described above and supplemented with published strain data for stented coronary,9,15,29,53 mid LAD and ostial RCA segments34 (Table 2). Boundary conditions were applied using multipoint constraints (MPCs), using rigid beam elements as per the approach of Ref. 52 and as illustrated in Fig. 1 for the delta wing design. Deflections representative of a generic artery case, ostial right (oRCA) and mid-proximal (mpLAD) coronary arteries were chosen.
Table 2. Native and stented coronary artery displacements.
Unstented | Stented | |||||
---|---|---|---|---|---|---|
|
|
|||||
Parameter | LAD | RCA | Ref. | LAD | RCA | Ref. |
Shortening (%) | 4.2 | 2.9 | Present study | |||
6.7 | a | 9 | 5.0 | 7.7 | 53 | |
8.1 | a | 15 | ||||
Bending (°) | 7.9 | 6.8 | Present study | 10.6 | 7.1 | Present study |
25 | 25 | 29 | ||||
Torsion | ||||||
(°) | 7.0 | 5.7 | Present study | |||
(°) | 6.0 | a | 9 | a | a | |
(cm−1) | 8.6 | 4.5 | 15 |
None available.
Following FEA, a Goodman fatigue analysis was performed extracting the maximum principal stresses in the maximally loaded and unloaded states. The mean and alternating stress were calculated, plotted for each of the three stent designs and compared against a critical failure line defined by the fatigue strength and ultimate tensile strength.
Results are presented in terms of contour plots from FEA illustrating spatially the von Mises stress distributions and also the ranges of particular strain variables at the final loading state for each case. The temporal development of this stress state is shown in supplementary material for each of the three stent designs. In addition, the Goodman analysis was spatially plotted to demonstrate predicted unsafe areas.
Statistical Methodology
Categorical variables, presented as frequencies and percentages, were compared with Fisher's exact test. Continuous variables, expressed as mean ± standard error, were compared with Student's paired t test with unequal variance. All p values were 2 sided, and p <0.05 was considered statistically significant. Minitab software (State College, PA, USA, v16.2.2) was used for the statistical analyses.
Results
Stent Implantation Parameters Associated with Stent Fracture
Amongst demographics (Table 1) only younger age was associated with fracture (61.4 vs. 66.7 years, p < 0.001), suggesting a role for physical activity and cardiac cycling in development of stent fracture. In concert with previous reports,37,38 DES fractures more often involved Cypher stents (Cordis™, Johnson & Johnson, NJ, USA), though the bare metal backbone Bx Velocity stent (Cordis™, Johnson & Johnson, NJ, USA) had fracture occurrence on par with other BMS designs (Supplementary Table 1). Differences in stent diameter were trivial between the two patient groups (Table 1); however, patients with stent fracture had longer stents. Further, patients with fracture had both more stents implanted in toto and multiple stents in the target vessel. Interestingly, many of these cases (69.8%, when multiple stents implanted) represented implantation of stents in overlapping fashion (Fig. 2). Several patients had fractures in other configurations such as stent-in-stent and end-to-end stent placement, though these associations did not reach significance in our data set.
The location of fracture along the stent length varied with implantation conditions. For single stents implanted, the majority of fractures occurred at the mid stent (Fig. 2). However, in overlapped stents, fracture most often localized to the stent region adjacent to the overlap site. In stent-in-stent geometries, the majority of fractures localized to the mid stent, as in the case of single stent implantation. With stent-in-stent geometries, the total of all fractures adjacent to the overlap region (e.g., proximal and distal overlap) were equivalent to those in the mid stent region, indicating that stent in stent implantation cases may behave as either a single stent case or an overlap case.
Arterial Locations Associated with Stent Fracture
Adverse events and stent fractures were associated with stent implantation in the RCA (Fig. 3). Though previous reports associated fractures with the left anterior descending (LAD) position, we found a statistically identical rate of non-fracture complications in the LAD, perhaps representing the frequency of intervention in this artery. At the same time, perhaps reflecting the different bending stresses on the different coronary arteries, the left circumflex (LCX), left main and branch vessels of the LAD and LCX appeared to be less often associated with fracture in this population.
To assess the stress and strain microenvironment, we stratified the location of implantation along the vessel length (Fig. 3). In adverse event reports, patients with fracture were associated with stent implantation in the oRCA, mid proximal LCX and proximal LCX as compared to controls. The association of fractures with oRCA may be indicative of the constrained position at the arterial take off at the base of the heart and focal high curvature evident in this arterial sub-segment in some patients, as described below. Interestingly, patients with stents implanted in the distal RCA had fewer reports of adverse events and fracture compared to controls.
Coronary Artery Curvature and Anatomic Variation
Unstented and stented coronary arteries experience significant cyclic shortening, bending, and torsional deflections (Table 2). In general, there is wide variation from one arterial segment to another and from patient to patient. On a per artery basis, stents in the RCA experience significant shortening and bending, in particular. In our analysis, which stratified stent location by vessel subsegment, the smallest radius of curvature (ROC) on EKG-gated CT images, corresponding to the most severe bending angle, was 2.82 cm in the mid proximal RCA, closely followed by the proximal RCA and proximal LAD. Of note, our patient cohort of arterial deflections did not have cases of stent implantation in the oRCA segment. These cases also exhibited an increase in ROC upon systole, perhaps reflecting their relative proximal location (Supplementary Table 2).
Finite Element Models of Arterial Segments Associated with Fracture
Stented arterial segments experience significant dynamic deformation (Table 2) and to our knowledge this is the first time multimodal displacements incorporating combined bending, torsion, and compression have been represented in a finite element model. When uniformly applied boundary conditions, simulating the relatively unconstrained mid LAD segment were applied, the highest von Mises stresses and maximum principal strains were localized to the mid stent region (Fig. 4). In contrast, application of a proximal stent constraint with distal displacement resulted in a high ROC as evident in the oRCA position34 with high von Mises stress and maximum principal strain in the proximal stent segment.
Finite Element Models of Stent Implantation Geometries Associated with Fracture
We evaluated the effect of increasing stent length on the development of stresses and strains when subjected to displacements representative of those observed in stented arterial segments (Fig. 5). Increasing stent length resulted in a relatively higher proportion of elements experiencing elevated material strain, as well as strains above that associated with the development of fatigue-type fractures. That longer stents are more prone to fracture, has been reported in the clinical literature. But this is to our knowledge the first report of mechanical evidence for this behavior. Similar to clinical observations, the finite element analyses predicted that for single stents the location of fracture, due to elevated maximum principal strains, localized to the mid stent region and this failure mode was preserved for all stent lengths.
In total, strains developed in overlapped stents exceeded those in single stents. When overall deformation behavior was considered, both overlapped and stent-in-stent geometries predicted focal deformation adjacent to the overlap sites (Fig. 5), which complements the clinical observations of fracture location for these implantation conditions. Von Mises stresses exceeding UTS are also noted in the sites adjacent to overlap but as these predictions are in the absence of tissue it is likely that these numbers are overestimates and that the presence of tissue would be associated with less stress in the device.
Finite Element Models of Multimodal Deformation of Stents of Varied Design
Representative symmetric multimodal arterial deformations were applied to stents of varied design, including delta-wing, corrugated ring, and multilink designs, representative of commercially clinically available devices (Fig. 6). Interestingly, the multi-cycle deformation behavior and localization of elevated von Mises stress and maximum principal strain were remarkably consistent across stent designs. Examining the strain ranges for each design (Table 3) similar ranges are predicted for all three designs in terms of maximum and minimum principal strains. However, high equivalent plastic strain accumulation is predicted for the delta-wing design, consistent with its relatively closed unit cell design. Slight variations in the intra-cyclic deflection behavior, as well as von Mises stress distribution, were observed that were unique to each stent design modeled (Supplementary Videos 1–3).
Table 3. Predicted strain for three uniformly deformed designs.
PEEQ (%) | Max principal ε (%) | Min principal ε (%) | ||||
---|---|---|---|---|---|---|
|
|
|
||||
Design | Max | Min | Max | Min | Max | Min |
Delta wing | 307.2 | 0 | 48.2 | −0.1 | 0 | 51.0 |
Corrugated wing | 95.5 | 0 | 49.1 | 0 | 0 | −97.1 |
Open cell | 92.7 | 0 | 61.5 | 0 | 0 | −73.1 |
Values represent strain range at the end of each simulation following four fully reversed cycles of simultaneous bending, torsion, and compression.
Reported variables are equivalent plastic strain (PEEQ), maximum principal strain and minimum principal strain.
Goodman Fatigue Analysis
Goodman analysis for the three stent designs subjected to multimodal deformations (Fig. 7) confirmed that failure is expected for all three designs in a significant percentage of elements, 10.1% (7074 elements) for the delta-wing, 6.8% (4426 elements) for the corrugated wing and 2.1% (2622 elements) for the open cell design. The spatial locations for the “unsafe” regions (Fig. 7) vary with design. The variation of results with each subsequent loading cycle remained stable after 4 loading cycles for the boundary conditions and material model used (Supplementary Fig. 1). While the majority of points fall within the “safe” zone of the Goodman diagram, those falling in the “unsafe” zone are of concern, suggesting that, for those boundary conditions examined, high stress and strain are at the level which may result in material failure.
Discussion
Despite increases in adverse event reporting and advances in imaging and monitoring,25,42 the causes and mechanical predictors of stent fracture remain poorly understood. Evaluation and understanding of the clinical phenomenon of stent fracture requires integration of data from adverse events, and finite element models of material fatigue and realistic input parameters and boundary conditions for implanted stent stress and strain. Such integration is understandably challenging and has to date largely been addressed in isolated, single mode analysis, modeling radial pulsation only.
Analysis of patient adverse event reports revealed the artery and device specific pre-implantation conditions associated with the clinical outcome of stent fracture. Importantly, this analysis allows us to deconstruct the specific conditions associated with device dysfunction such that they may be later tested for the true impact of these conditions in representative computational models that probe certain physical conditions of interest. Drug eluting stents represented the majority of adverse events in the MAUDE database reflecting either a preference in use or reporting, or perhaps a different integrated chemical-biological–mechanical phenomenon underlying fracture events in DES. Our results suggest the former, as adverse event reports secondary to fracture had similar DES use to non-fracture events. However, future work should investigate the potential for differential arterial response secondary to strut fracture when drugs are eluted from stents and possible pharmacological, biological and physical interplay in fatigue failure. Clinical fracture reports were also associated with stent length and the use of multiple, overlapping stents (Table 1; Fig. 2). The end of stents closest to the coronary ostium may be more prone to fracture compared to the segments farther away (Fig. 2), due to the apex-to-base motion of the heart and tethering of the coronary artery by the aorta. Our results further suggest that large stresses and strains result from dynamic changes in curvature, which result in stress and strain concentrations. Load concentration can also explain the predilection for clinical fractures when long and/or multiple stents are used. Long stents will be subject to a greater bending deflection for a given radius of curvature and thus experience higher material strains (Fig. 5), and the stent region adjacent to the overlap site in overlapped and stent-in-stent implantation geometries represents a relative discontinuity in longitudinal stiffness that concentrates stress and strain which further localizes fracture to the regions adjacent to overlap.
Specific arterial location and stent geometrical parameters identified in adverse outcomes correlated well with mechanical stress and strain concentrations when recapitulated in finite element models with representative arterial deformations. The RCA, in particular the ostial RCA, mid proximal LCX, and proximal LCX were associated with clinically evident fracture (Fig. 3). These regions are especially subjected to high, dynamic changes in curvature. Though the individual geometries and deflections are patient-specific and span a wide range across the physiologic spectrum, it is assumed generalizable to the population at large where such conditions exist. Finite element models showed material stress and strain concentrations along the length of the stent in good agreement with the same regions implicated in clinical reports of fracture (Fig. 4). While we did not explicitly model stent fracture initiation or propagation, we do provide evidence for the underlying mechanical force. For the first time finite element predictions showed how multimodal deformations and acute changes in curvature, as evident in the ostial RCA for example, can result in strain concentrations at the proximal stent region. Additionally, when a relatively unconstrained case with multimodal displacement is considered, material stress and strain localizes to the mid stent region. Overlap and stent-in-stent implantation cases revealed similar regions of concern that correlated with those regions implicated in clinical observations of the same.
Finally, the regions of high von Mises stress, maximum and minimum principal strains, and equivalent plastic strain, in delta-wing, corrugated ring, and multilink designs transcend design in part when cyclic deformations were kept constant (Fig. 6; Table 3; Supplementary Videos 1–3). Each individual design experienced slight differences in magnitude and distribution of material strains, likely due to differences in unit cell, strut thickness, and number of connectors and links along the radial and longitudinal directions but overall patterns were similar. Thus, though there are unique features evident in each arterial and device specific system, these differences can be realistically modeled using advanced computational methods. Further, stress and strain concentration predictions can be used to assess for potentially unsafe regions such as in the Goodman analysis (Fig. 7), which indicated a significant percentage of elements fall within the “unsafe” region for all three designs. These results highlight the need for consideration of multimodal deformations in assessing stent performance and for design of next generation devices.
Clinical observational studies have assessed the clinical associations of stent fracture,1,42,43 but these have largely been separate from mechanistic computational studies and previous reports cannot comment on how severity or nature of fracture might contribute to extent of clinical consequence. Our findings support the findings of previous reports but also extend and most importantly begin to explain them. In this way we can now direct attention to clinically meaningful and mechanistically robust analysis of specific fracture modalities and areas of the arterial tree to more closely scrutinize or more precisely rule out strut fracture as a potential contribution or complication in specific patient subsets or clinical scenarios. Thus, with a more inclusive perspective of the range of clinical fractures, we can begin to tease out the spectrum of clinical consequences and variability therein based not only on the type of fracture evident but the potential for more severe fractures to propagate thereafter.
Also to be acknowledged are the limitations of this work. The FDA MAUDE database is a user-initiated, prospectively maintained reporting system that though may not represent the full burden of clinical cases, given the significant incidence of stent fracture reported it was deemed indicative of the problem and analyzed here. The applied boundary conditions were derived from a finite number of patients—a larger data set may provide slight differences in numerical values. Also, because these values represent a spectrum of patients, and were not all collected from similar locations along the length of the artery, we did not seek to replicate any exact set of these conditions. Rather, we chose representative values for compression, bending, and torsion that were within the range of what might be expected in a dynamically strained coronary artery.
Yet, this study does demonstrate the effect of representative deformations, which admittedly may vary slightly per design, as a first step towards a more thorough examination of device performance. Though the arterial segment cases presented rely on representative boundary conditions and our understanding of coronary dynamics, they provide an interesting first look at the potential deformation mechanisms that may underlie the elevated stresses and strains in the regions associated with fracture in patients. Thus, the current analysis is limited to a representative set boundary conditions that does not attempt to fully characterize what indeed are a wide array of possible in vivo deformations. Additionally, finite element stent materials were represented as biomedical grade stainless steel 316L but other alloys could have been modeled such as cobalt chromium. The selection of 316L was deemed acceptable for comparison purposes. The effects of crimping and deployment were not included in the finite element models and ongoing work is examining the combined effects of deployment and subsequent multimodal loading. Additionally, though contact between stent elements in a single stent were considered, contact between overlapping stents was not. None-the-less, this study shows the significant effect that multimodal loading has on device performance and this is the first time such an approach has been taken. The finite element models presented predict multimodal stent deformations without an artery present. The effect of stent oversizing and arterial pathology is therefore not assessed. We expect that these components will likely add to the material stress and strain states predicted, though additional work is needed. As the use of MPCs necessarily introduce point discontinuities in applied and observed deformations, addition of a surrounding vessel would also allow for application of a non-uniform deformation gradient along the length of the stent, as in reality the whole stent body will be subject to non-uniform loading and this will have an impact on the stresses generated.
Also to be acknowledged are the limitations of the Goodman approach due to its base formulation, it is derived from the overall macroscale behavior of engineering components and applied here at the scale of individual finite elements. Finally, though our work represents a first step toward more complex multi-modal modeling, future work would benefit from more complex computational modeling of fatigue, fracture, and material evolution which should utilize more microscale-based and physically-based prediction methodologies.
Our results suggest that though there are unique and appreciable differences in arterial deflections and stent design for any given patient and procedure, there are underlying mechanistic concepts driving regional development of device stress and strain that can 1 day be leveraged to predict the localization of fracture in patients. Our results provide the first associations of stent and artery conditions of interest identified in fracture adverse events in United States' patients with computational mechanical models that can incorporate the multimodal, multi-cycle deformation behavior of multiple stent designs with specific implantation conditions associated with clinically significant stent fracture.
Supplementary Material
Acknowledgments
This work was funded in part by Grants from FDA Office of Regulatory Affairs, R01 GM49039 (ERE) from the National Institute of Health, and appointment to the Research Participation Program at FDA administered by Oak Ridge Institute for Science and Education (CC).
Footnotes
Electronic supplementary material: The online version of this article (doi:10.1007/s10439-015-1476-3) contains supplementary material, which is available to authorized users.
Disclosures: Drs Choi and Taylor are employees of and have equity interest in HeartFlow, Inc. Other authors report no conflicts.
Contributor Information
Kay D. Everett, Email: kfurman@mit.edu.
Gerard J. Desany, Email: Gerard.Desany@fda.hhs.gov.
Brian L. Baker, Email: Brian.Baker@fda.hhs.gov.
Gilwoo Choi, Email: giroo@heartflow.com.
Charles A. Taylor, Email: taylor@heartflow.com.
Elazer R. Edelman, Email: ere@mit.edu.
References
- 1.Aoki J, Nakazawa G, Tanabe K, Hoye A, Yamamoto H, Nakayama T, et al. Incidence and clinical impact of coronary stent fracture after sirolimus-eluting stent implantation. Catheter Cardiovasc Interv. 2007;15(69):380–386. doi: 10.1002/ccd.20950. [DOI] [PubMed] [Google Scholar]
- 2.Argente dos Santos HAF, Auricchio F, Conti M. Fatigue life assessment of cardiovascular balloon-expandable stents: a two-scale plasticity—damage model approach. J Mech Behav Biomed Mater. 2012;15:78–92. doi: 10.1016/j.jmbbm.2012.06.011. [DOI] [PubMed] [Google Scholar]
- 3.Auricchio F, Constantinescu A, Conti M, Scalet G. A computational approach for the lifetime prediction of cardiovascular balloon-expandable stents. Int J Fatigue. 2015;75:69–79. [Google Scholar]
- 4.Azaouzi M, Makradi A, Petit J, Belouettar S, Polit O. On the numerical investigation of cardiovascular balloon-expandable stent using finite element method. Comput Mater Sci. 2013;79:326–335. [Google Scholar]
- 5.Barrera O, Makradi A, Abbadi M, Azaouzi M, Belouettar S. On high-cycle fatigue of 316L stents. Comput Methods Biomech Biomed Eng. 2014;17:239–250. doi: 10.1080/10255842.2012.677442. [DOI] [PubMed] [Google Scholar]
- 6.CFR—Code of Federal Regulations Title 21. [cited 2014 Jun 4]; [Internet]. http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/cfrsearch.cfm?fr=803.19.
- 7.Choe H, Hur G, Doh JH, Namgung J, Lee SY, Park KT, et al. A case of very late stent thrombosis facilitated by drug eluting stent fracture: comparative images before and after stent fracture detected by 64-multidetector computed tomography. Int J Cardiol. 2009;17(133):e125–128. doi: 10.1016/j.ijcard.2008.11.040. [DOI] [PubMed] [Google Scholar]
- 8.Choi G. In vivo quantification of arterial deformation due to pulsatile and non-pulsatile forces: implications for the design of stents and stent-grafts. [cited 2014 Jun 4];2009 Internet. http://gradworks.umi.com/33/82/3382704.html.
- 9.Choi G, Cheng CP, Wilson NM, Taylor CA. Methods for quantifying three-dimensional deformation of arteries due to pulsatile and nonpulsatile forces: implications for the design of stents and stent grafts. Ann Biomed Eng. 2009;37:14–33. doi: 10.1007/s10439-008-9590-0. [DOI] [PubMed] [Google Scholar]
- 10.Chung WS, Park CS, Seung KB, Kim PJ, Lee JM, Koo BK, et al. The incidence and clinical impact of stent strut fractures developed after drug-eluting stent implantation. Int J Cardiol. 2008;25(125):325–331. doi: 10.1016/j.ijcard.2007.02.033. [DOI] [PubMed] [Google Scholar]
- 11.Conway C, McGarry JP, McHugh PE. Modelling of atherosclerotic plaque for use in a computational test-bed for stent angioplasty. Ann Biomed Eng. 2014;11(42):2425–2439. doi: 10.1007/s10439-014-1107-4. [DOI] [PubMed] [Google Scholar]
- 12.Conway C, Sharif F, McGarry J, McHugh P. A computational test-bed to assess coronary stent implantation mechanics using a population-specific approach. Cardiovasc Eng Technol. 2012;3:1–14. [Google Scholar]
- 13.Crossland B. Proceedings of International Conference on Fatigue of Metals, Institution of Mechanical Engineering. Vol. 138. London: 1956. Effect of large hydrostatic pressures on the torsional fatigue strength of an alloy steel. [Google Scholar]
- 14.Dang-Van K. Macro-micro approach in high-cycle multi-axial fatigue. In: McDowell DL, Ellis R, editors. Advances in multiaxial fatigue. Philadelphia: ASTM International; 1993. pp. 120–130. [Google Scholar]
- 15.Ding Z, Zhu H, Friedman MH. Coronary artery dynamics in vivo. Ann Biomed Eng. 2002;1(30):419–429. doi: 10.1114/1.1467925. [DOI] [PubMed] [Google Scholar]
- 16.Dordoni E, Meoli A, Wu W, Dubini G, Migliavacca F, Pennati G, et al. Fatigue behaviour of Nitinol peripheral stents: the role of plaque shape studied with computational structural analyses. Med Eng Phys. 2014;36:842–849. doi: 10.1016/j.medengphy.2014.03.006. [DOI] [PubMed] [Google Scholar]
- 17.Duda SH, Pusich B, Richter G, Landwehr P, Oliva VL, Tielbeek A, et al. Sirolimus-eluting stents for the treatment of obstructive superficial femoral artery disease six-month results. Circulation. 2002;17(106):1505–1509. doi: 10.1161/01.cir.0000029746.10018.36. [DOI] [PubMed] [Google Scholar]
- 18.Early M, Kelly DJ. The consequences of the mechanical environment of peripheral arteries for nitinol stenting. Med Biol Eng Comput. 2011;49:1279–1288. doi: 10.1007/s11517-011-0815-2. [DOI] [PubMed] [Google Scholar]
- 19.Early M, Lally C, Prendergast PJ, Kelly DJ. Stresses in peripheral arteries following stent placement: a finite element analysis. Comput Methods Biomech Biomed Eng. 2009;12:25–33. doi: 10.1080/10255840903065043. [DOI] [PubMed] [Google Scholar]
- 20.Halwani DO, Anderson PG, Brott BC, Anayiotos AS, Lemons JE. The role of vascular calcification in inducing fatigue and fracture of coronary stents. J Biomed Mater Res B Appl Biomater. 2012;1(100B):292–304. doi: 10.1002/jbm.b.31911. [DOI] [PubMed] [Google Scholar]
- 21.Harewood FJ, McHugh PE. Modeling of size dependent failure in cardiovascular stent struts under tension and bending. Ann Biomed Eng. 2007;35:1539–1553. doi: 10.1007/s10439-007-9326-6. [DOI] [PubMed] [Google Scholar]
- 22.Hsiao HM, Yin MT. An intriguing design concept to enhance the pulsatile fatigue life of self-expanding stents. Biomed Microdev. 2014;16:133–141. doi: 10.1007/s10544-013-9813-5. [DOI] [PubMed] [Google Scholar]
- 23.Iida O, Nanto S, Uematsu M, Morozumi T, Kotani J, Awata M, et al. Effect of exercise on frequency of stent fracture in the superficial femoral artery. Am J Cardiol. 2006;15(98):272–274. doi: 10.1016/j.amjcard.2006.01.091. [DOI] [PubMed] [Google Scholar]
- 24.Ino Y, Toyoda Y, Tanaka A, Ishii S, Kusuyama Y, Kubo T, et al. Predictors and prognosis of stent fracture after sirolimus-eluting stent implantation. Circ J. 2009;73:2036–2041. doi: 10.1253/circj.cj-09-0343. [DOI] [PubMed] [Google Scholar]
- 25.Jaff M, Dake M, Pompa J, Ansel G, Yoder T. Standardized evaluation and reporting of stent fractures in clinical trials of noncoronary devices. Catheter Cardiovasc Interv. 2007;1(70):460–462. doi: 10.1002/ccd.21240. [DOI] [PubMed] [Google Scholar]
- 26.LaDisa JF, Olson LE, Douglas HA, Warltier DC, Kersten JR, Pagel PS. Alterations in regional vascular geometry produced by theoretical stent implantation influence distributions of wall shear stress: analysis of a curved coronary artery using 3D computational fluid dynamics modeling. Biomed Eng OnLine. 2006;16(5):40. doi: 10.1186/1475-925X-5-40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Lally C, Dolan F, Prendergast PJ. Cardiovascular stent design and vessel stresses: a finite element analysis. J Biomech. 2005;38:1574–1581. doi: 10.1016/j.jbiomech.2004.07.022. [DOI] [PubMed] [Google Scholar]
- 28.Li J, Luo Q, Xie Z, Li Y, Zeng Y. Fatigue life analysis and experimental verification of coronary stent. Heart Vessels. 2010;1(25):333–337. doi: 10.1007/s00380-009-1203-9. [DOI] [PubMed] [Google Scholar]
- 29.Liao R, Chen SYJ, Messenger JC, Groves BM, Burchenal JEB, Carroll JD. Four-dimensional analysis of cyclic changes in coronary artery shape. Catheter Cardiovasc Interv. 2002;1(55):344–354. doi: 10.1002/ccd.10106. [DOI] [PubMed] [Google Scholar]
- 30.Ling AJ, Mwipatayi P, Gandhi T, Sieunarine K. Stenting for carotid artery stenosis: fractures, proposed etiology and the need for surveillance. J Vasc Surg. 2008;1(47):1220–1226. doi: 10.1016/j.jvs.2008.01.043. [DOI] [PubMed] [Google Scholar]
- 31.Marrey RV, Burgermeister R, Grishaber RB, Ritchie RO. Fatigue and life prediction for cobalt-chromium stents: a fracture mechanics analysis. Biomaterials. 2006;27:1988–2000. doi: 10.1016/j.biomaterials.2005.10.012. [DOI] [PubMed] [Google Scholar]
- 32.McGarry JP, O'Donnell BP, McHugh PE, McGarry JG. Analysis of the mechanical performance of a cardiovascular stent design based on micromechanical modelling. Comput Mater Sci. 2004;31:421–438. [Google Scholar]
- 33.Meoli A, Dordoni E, Petrini L, Migliavacca F, Dubini G, Pennati G. Computational modelling of in vitro set-ups for peripheral self-expanding nitinol stents: the importance of stent-wall interaction in the assessment of the fatigue resistance. Cardiovasc Eng Technol. 2013;1(4):474–484. [Google Scholar]
- 34.Messenger JC, Chen SY, Carroll JD, Burchenal JE, Kioussopoulos K, Groves BM. 3D coronary reconstruction from routine single-plane coronary angiograms: clinical validation and quantitative analysis of the right coronary artery in 100 patients. Int J Card Imaging. 2000;16:413–427. doi: 10.1023/a:1010643426720. [DOI] [PubMed] [Google Scholar]
- 35.Min PK, Yoon YW, Moon Kwon H. Delayed strut fracture of sirolimus-eluting stent: a significant problem or an occasional observation? Int J Cardiol. 2006;26(106):404–406. doi: 10.1016/j.ijcard.2004.12.087. [DOI] [PubMed] [Google Scholar]
- 36.Morlacchi S, Pennati G, Petrini L, Dubini G, Migliavacca F. Influence of plaque calcifications on coronary stent fracture: a numerical fatigue life analysis including cardiac wall movement. J Biomech. 2014;3(47):899–907. doi: 10.1016/j.jbiomech.2014.01.007. [DOI] [PubMed] [Google Scholar]
- 37.Nakazawa G, Finn AV, Vorpahl M, Ladich E, Kutys R, Balazs I, et al. Incidence and predictors of drug-eluting stent fracture in human coronary artery: a pathologic analysis. J Am Coll Cardiol. 2009;17(54):1924–1931. doi: 10.1016/j.jacc.2009.05.075. [DOI] [PubMed] [Google Scholar]
- 38.Park MW, Chang K, Her SH, Lee JM, Choi YS, Kim DB, et al. Incidence and clinical impact of fracture of drug-eluting stents widely used in current clinical practice: comparison with initial platform of sirolimus-eluting stent. J Cardiol. 2012;60:215–221. doi: 10.1016/j.jjcc.2012.07.011. [DOI] [PubMed] [Google Scholar]
- 39.Park KW, Park JJ, Chae IH, Seo JB, Yang HM, Lee HY, et al. Clinical characteristics of coronary drug-eluting stent fracture: insights from a two-center des registry. J Korean Med Sci. 2011;26:53–58. doi: 10.3346/jkms.2011.26.1.53. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Park JS, Shin DG, Kim YJ, Hong GR, Cho IH. Acute myocardial infarction as a consequence of stent fracture and plaque rupture after sirolimus-eluting stent implantation. Int J Cardiol. 2009;15(134):e79–81. doi: 10.1016/j.ijcard.2007.12.095. [DOI] [PubMed] [Google Scholar]
- 41.Pelton AR, Schroeder V, Mitchell MR, Gong XY, Barney M, Robertson SW. Fatigue and durability of Nitinol stents. J Mech Behav Biomed Mater. 2008;1:153–164. doi: 10.1016/j.jmbbm.2007.08.001. [DOI] [PubMed] [Google Scholar]
- 42.Popma JJ, Tiroch K, Almonacid A, Cohen S, Kandzari DE, Leon MB. A qualitative and quantitative angiographic analysis of stent fracture late following sirolimus-eluting stent implantation. Am J Cardiol. 2009;1(103):923–929. doi: 10.1016/j.amjcard.2008.12.022. [DOI] [PubMed] [Google Scholar]
- 43.Scheinert D, Scheinert S, Sax J, Piorkowski C, Bräunlich S, Ulrich M, et al. Prevalence and clinical impact of stent fractures after femoropopliteal stenting. J Am Coll Cardiol. 2005;18(45):312–315. doi: 10.1016/j.jacc.2004.11.026. [DOI] [PubMed] [Google Scholar]
- 44.Serikawa T, Kawasaki T, Koga H, Orita Y, Ikeda S, Goto Y, et al. Late catch-up phenomenon associated with stent fracture after sirolimus-eluting stent implantation: incidence and outcome. J Interv Cardiol. 2011;24:165–171. doi: 10.1111/j.1540-8183.2010.00614.x. [DOI] [PubMed] [Google Scholar]
- 45.Sianos G, Hofma S, Ligthart JMR, Saia F, Hoye A, Lemos PA, et al. Stent fracture and restenosis in the drug-eluting stent era. Catheter Cardiovasc Interv. 2004;1(61):111–116. doi: 10.1002/ccd.10709. [DOI] [PubMed] [Google Scholar]
- 46.Sines G, Ohgi G. Fatigue criteria under combined stresses or strains. J Eng Mater Technol. 1981;1(103):82–90. [Google Scholar]
- 47.Sweeney CA, McHugh PE, McGarry JP, Leen SB. Micromechanical methodology for fatigue in cardiovascular stents. Int J Fatigue. 2012;44:202–216. [Google Scholar]
- 48.Sweeney CA, O'Brien B, Dunne FPE, McHugh PE, Leen SB. Micro-scale testing and micromechanical modelling for high cycle fatigue of CoCr stent material. J Mech Behav Biomed Mater. 2015;46:244–260. doi: 10.1016/j.jmbbm.2015.02.011. [DOI] [PubMed] [Google Scholar]
- 49.Sweeney CA, O'Brien B, Dunne FPE, McHugh PE, Leen SB. Strain-gradient modelling of grain size effects on fatigue of CoCr alloy. Acta Mater. 2014;1(78):341–353. [Google Scholar]
- 50.Sweeney CA, O'Brien B, McHugh PE, Leen SB. Experimental characterisation for micromechanical modelling of CoCr stent fatigue. Biomaterials. 2014;35:36–48. doi: 10.1016/j.biomaterials.2013.09.087. [DOI] [PubMed] [Google Scholar]
- 51.Umeda H, Kawai T, Misumida N, Ota T, Hayashi K, Iwase M, et al. Impact of sirolimus-eluting stent fracture on 4-year clinical outcomes. Circ Cardiovasc Interv. 2011;4:349–354. doi: 10.1161/CIRCINTERVENTIONS.110.958306. [DOI] [PubMed] [Google Scholar]
- 52.Wu W, Yang DZ, Qi M, Wang WQ. An FEA method to study flexibility of expanded coronary stents. J Mater Process Technol. 2007;12(184):447–450. [Google Scholar]
- 53.Zhu H, Warner JJ, Gehrig TR, Friedman MH. Comparison of coronary artery dynamics pre- and post-stenting. J Biomech. 2003;36:689–697. doi: 10.1016/s0021-9290(02)00447-5. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.