. 2016 Oct 5;36(5):625–634. doi: 10.1007/s40846-016-0165-5

Using Lempel–Ziv Complexity to Assess ECG Signal Quality

Yatao Zhang 1,2, Shoushui Wei 1,, Costanzo Di Maria 3,4, Chengyu Liu 1,3,
PMCID: PMC5083778  PMID: 27853413

Abstract

The poor quality of wireless electrocardiography (ECG) recordings can lead to misdiagnosis and waste of medical resources. This study presents an interpretation of Lempel–Ziv (LZ) complexity in terms of ECG quality assessment, and verifies its performance on real ECG signals. Firstly, LZ complexities for typical signals, namely high-frequency (HF) noise, low-frequency (LF) noise, power-line (PL) noise, impulse (IM) noise, clean artificial ECG signals, and ECG signals with various types of noise added (ECG plus HF, LF, PL, and IM noise, respectively) were analyzed. Then, the effects of noise, signal length, and signal-to-noise ratio (SNR) on the LZ complexity of ECG signals were analyzed. The simulation results show that LZ complexity for HF noise was obviously different from those for PL and LF noise. The LZ value can be used to determine the presence of HF noise. ECG plus HF noise had the highest LZ values. Other types of noise had low LZ values. Signal lengths of over 40 s had only a small effect on LZ values. The LZ values for ECG plus all types of noise increased monotonically with decreasing SNR except for LF and PL noise. For the test of real ECG signals plus three types of noise, namely muscle artefacts (MAs), baseline wander (BW), and electrode motion (EM) artefacts, LZ complexity varied obviously with increasing MA but not for BW and EM noise. This study demonstrates that LZ complexity is sensitive to noise level (especially for HF noise) and can thus be a valuable reference index for the assessment of ECG signal quality.

Keywords: Electrocardiography (ECG), Lempel–Ziv (LZ) complexity, Signal quality assessment, Signal-to-noise ratio (SNR)

Introduction

Electrocardiography (ECG) recordings are often contaminated by various types of noise, including movement artefacts, power-line (50 or 60 Hz) noise, and muscular electrical activity. The presence of this noise can render the signals unsuitable for clinical use, wasting the resources utilized for their acquisition [1]. Assessing the quality of ECG signals would be extremely helpful with this regard.

The increasing use of mobile devices for acquiring ECG signals has recently driven research interest towards the problem of assessing the signal quality of ECG recordings [25]. Existing methods are mostly based on the characterization of time or frequency features of the signals. Time-based methods aim to identify particular characteristics, such as RR time interval outliers [6], flat lines, baseline wander (BW), and steep slopes [7], which usually can compromise recordings. Frequency-based methods use, for example, the ratio between the low- and high-frequency power of the signal [8]. A method that combines time and frequency information has also been proposed [9, 10]. Non-linear methods, such as entropy measures, have also been explored for this application but need further investigation [11, 12].

Lempel–Ziv (LZ) complexity [1315] is a measure of the complexity of a signal. It has been applied to a variety of biomedical signals, including ECG from patients with ventricular tachycardia or atrial fibrillation [16, 17], heart sound signals from patients with cardiovascular disease [18], electroencephalograms (EEG) from patients with Alzheimer’s disease [19, 20], EEG sleep signals [21], and brain function [22]. Aboy et al. [23] analysed the LZ complexity of periodic signals, Gaussian white noise, and colored noise. However, how typical types of noise affect the LZ complexity of ECG recordings, and the relationship between the LZ complexity and noise level, have not yet been systematically studied.

The aim of this study was to characterize the LZ complexity of various ECG signals (noise-free and noisy) and to assess the effects of noise type on the LZ complexity of ECG signals.

Materials and Methods

Database

Artificial ECG Signals

Clean artificial ECG signals were generated using the open source ECGSYN software, as described by McSharry et al. [24]. The sample rate was set at 360 Hz. The heart rate was set to be within the range of 50–100 beats per minute (bpm). Four types of noise were separately added to the clean ECG signals: 50–180 Hz high-frequency (HF) noise; 50 Hz power-line (PL) noise; 0–0.5 Hz low-frequency (LF) noise; and impulse (IM) noise. HF noise was used to simulate muscular electrical activity and other HF noise. In order to avoid the overlap of the frequency range of HF noise and the main frequency range of ECG, 50–180 Hz was chosen (the upper end of the range was determined at a sample rate of 360 Hz). PL noise was used to simulate noise from the mains. LF noise was used to simulate BW because its frequency range overlaps with 0–0.5 Hz [25, 26]; it can approximately be regarded as an electrode motion artefact with a significant amount of BW [27]. IM noise was used to simulate the spikes with high amplitudes contained in ECG signals. To generate IM noise, the zero sequence was replaced by various percentages of random spikes. To generate HF noise, Gaussian noise was firstly generated and then filtered by a band-pass filter (50–180 Hz). To generate LF noise, Gaussian white noise was firstly generated and then filtered by a low-pass filter (0–0.5 Hz).

Real ECG Signals

Real ECG signals were selected from the Massachusetts Institute of Technology Beth Israel Hospital (now the Beth Israel Deaconess Medical Centre) (MIT/BIH) arrhythmia database [28, 29]. This database contains 48 ECG recordings, each with a duration of 30 min and a sample rate of 360 Hz. Baseline correction for removing the main LW noise was performed because the BW noise contained in the raw signals could lead to inaccurate results. The processed ECG signals were used for further study.

Real noise signals taken from the Noise Stress Test Database (NSTDB) [30] were used. NSTDB provides three types of noise that can be typically found in ambulatory ECG recordings: muscle artefacts (MA), electrode motion (EM), and BW. Because NSTDB does not include 50-Hz PL noise and IM noise, this study added these types of noise to the real ECG signals for testing.

Both the MIT-BIH database and NSTDB are publicly available through the Physionet website [29].

LZ Complexity

Before LZ complexity can be computed, the original signal must be coarse-grained, and then transformed into a symbols sequence for simplifying the computation. In previous works, the binary (two-state) sequence was demonstrated to adequately represent the LZ complexity of the original signal [16, 23, 31]. For generating the two-state sequence, the signal data were converted into a 0–1 sequence R by comparison with the threshold T h. The binary symbolic sequence R = {r(1), r(2),…, r(n)} was produced as follows:

r(i)=0,ifx(i)<Th1,ifx(i)Th,i=1,2n 1

where n is the length of x(n). Usually, the mean value of the sequence is used as the threshold T h [16, 32]. This was thus done for the coarse-graining process in this study.

Following the initial coarse-graining process, the LZ complexity c(n) for the symbol sequence R was computed. The whole binary sequence R is scanned from left to right, and the counter c(n) is increased by one unit when a new subsequence (a new pattern) of consecutive characters is encountered in the scanning process. The counter c(n) conforms to the following rules [16, 23, 29]:

  1. Let S and Q denote two strings, respectively, SQ be the concatenation of S and Q, string SQπ be derived from SQ after its last character is deleted (π means the operation to delete the last character in the string). Let v(SQπ) denote the vocabulary of all different substrings of SQπ. Initially, c(n) = 1, S = s 1, and Q = s 2, and so SQπ = s 1.

  2. In summary, S = s 1 s 2, …, s r, Q = s r+1, and so SQπ = s 1 s 2, …, s; if Q belongs to v(SQπ), then s r+1, that is, Q is a substring of SQπ, and so S does not change. Update Q to be s r+1 s r+2, and then judge whether Q belongs to v(SQπ). Repeat this process until Q does not belong to v(SQπ).

  3. Now, Q = s r+1 s r+2, …, s r+i, which is not a substring of SQπ = s 1 s 2, …, s r s r+1,…, s r+i-1, so increase c(n) by one.

  4. Thereafter, S is updated to be S = s 1 s 2, …, s r+i, and Q = s r+i+1.

Then, the procedure is repeated until Q is the last character. At this time, the counter c(n) is the number of different substrings contained in R, and it reflects the number of different patterns in a sequence. c(n) might vary with sequence length [17, 23]. Thus, in order to obtain a complexity measure independent of the sequence length, c(n) should be normalized [17, 23].

It has been proved that the upper bound of c(n) is:

c(n)<n(1-εn)logα(n) 2

where n is the length of the sequence and α is the number of different symbols in the symbol set. In this study, α was 2 because the coarse-grained sequence was a 0–1 sequence. ε n is a small quantity and ε n → 0 (n → ∞). In fact:

limnc(n)=b(n)=nlogα(n) 3

c(n) can be normalized as:

C(n)=c(n)b(n) 4

where C(n) is the normalized LZ complexity, and denotes the arising rate of new patterns within the sequence. A detailed LZ complexity analysis can be found elsewhere [23]. In this study, the normalized complexity C(n), rather than c(n), is regarded as the result of LZ complexity.

LZ Complexity Analysis for ECG Signals

LZ Analysis for Typical Signals

The LZ complexity was calculated separately for some typical signals, specifically various types of artificial signal (HF, LF, PL, and IM noise and clean ECG) and noisy ECG signals. In this test, IM noise was generated by replacing 10 % of the 40-s zero sequence with random spikes. For each type of signal, 50 repeats were used to reduce the effect of random factors. Each repeat lasted for 40 s. For the synthetic noisy ECG signal, the signal-to-noise ratio (SNR) is defined as:

SNR=10×log10PsignalPnoise 5

where P signal and P noise denote the power of the clean ECG and that of the noise, respectively.

In this test, the SNR of the synthetic ECG was set as 10 dB. By analysing the LZ values of these signals and their statistics (i.e., the mean and standard deviation), we tried to show typical LZ complexity values for the special types of signals.

Effect of Signal Length on LZ Complexity

The effect of signal length on LZ complexity was tested using the five types of signals: clean ECG signals and synthetic ECG signals plus HF, PL, LF, and IM noise, respectively. For each type of signal, the SNR of 10 dB was used and the signal length was varied from 5 to 120 s, in steps of 5 s. For each signal length, 50 repeats were performed and the mean values and standard deviations were determined for comparison. This effect analysis aimed to determine a suitable signal length to obtain stable LZvalues.

Effect of SNR on LZ Complexity

In this test, SNR was varied from −10 to 20 dB for each type of signal, in steps of 5 dB. For each SNR level, 50 repeats were produced and the mean values and standard deviations were calculated. To observe the mixed effect of the various types of noise, all noise types were added to the clean ECG signals. The mixed noise included HF, PL, LF, and IM noise, with the same proportion (25 %) for each noise type. For the SNR effect analysis, the optimal signal length from the effect analysis of signal length was used.

Figure 1 gives examples of the artificial ECG signals, namely the clean ECG signals and the synthetic noisy ECG signals at various SNR levels, as well as their LZ values. Figure 2 gives similar examples from the real ECG signals. It is can be seen that the identifiability of ECG waveforms decreases with decreasing SNR.

Fig. 1.

Fig. 1

Clean artificial ECG signals and these signals plus various types of noise at various SNR levels (−10 to 20 dB, in steps of 5 dB). Artificial ECG signals plus a HF, b PL, c LF and d IM noise. Corresponding LZ values are also shown

Fig. 2.

Fig. 2

Real ECG signals and these signals plus various types of noise at various SNR levels (−10 to 20 dB, in steps of 5 dB). Real ECG signals plus a MA, b PL, c BW, d EM and e IM noise. Corresponding LZ values are also shown

Receiver Operating Characteristic Curve for Verification of LZ Complexity

For verification, we used the receiver operating characteristic (ROC) curve to evaluate the performance of LZ complexity for signal quality assessment. According to pre-observations, the clean artificial ECG signals plus HF, LF, or PL noise were identified as too noisy (unacceptable) for clinical application when SNR was less than 5 dB. Similarly, the threshold for the clean artificial ECG signals plus IM noise was set as 0 dB. We set the threshold of 4.6 dB for classifying the clear artificial ECG plus mixed noise as “unacceptable” (SNR <= 4.6 dB) or “acceptable” (SNR > 4.6 dB) in this study. The reason for choosing these values is that the main waveform features (i.e., P, Q, and T waves) of the corrupted ECGs could be identified when SNR decreases below the specified thresholds. For the mixed-noise-corrupted ECG, we observed that an SNR value of between 4 and 5 dB can be used as the threshold. A threshold of 4.6 dB was thus chosen based on the ROC analysis.

Results

LZ Values for Typical ECG Signals

Figure 3 shows the LZ results of the typical signals (i.e., HF, LF, PL, and IM noise, the clean ECG, and the clean ECG plus HF, PL, LF, and IM noise, respectively) when performing 50 repeat calculations for each type of signal. HF noise showed the highest values of LZ complexity. LF noise had an LZ value of close to 0.1, and the clean ECG had a slightly lower LZ than that of LF noise. PL noise had the lowest LZ value.

Fig. 3.

Fig. 3

LZ complexity results from typical signals when performing 50 repeat calculations with signal length of 40 s and SNR of 10 dB

We also calculated the mean values and standard deviations of the 50 repeats for each type of signal with SNR set to 10 dB. These values are shown in Table 1.

Table 1.

Mean values and standard deviations of LZ complexity from 50 repeats for each type of signal

Signal type Mean Standard deviation
HF noise 0.8278 0.0082
LF noise 0.0994 0.0051
PL noise 0.0163 0.0000
IM noise 0.4425 0.0039
Clean ECG 0.0807 0.0124
Clean plus HF noise 0.7873 0.1010
Clean plus PL noise 0.2567 0.0249
Clean plus LF noise 0.1502 0.0122
Clean plus IM noise 0.0983 0.0245

Effect of Signal Length

Figure 4 shows the effects of signal length on LZ complexity for the artificial ECG signals. The LZ complexity decreased with increasing signal length when the signal length was shorter than 20 s. When the signal length was longer than 40 s, the LZ complexity remained stable for all of signal types. Therefore, a signal length of 40 s was used for the following analysis.

Fig. 4.

Fig. 4

Effect of signal length on LZ complexity for artificial noisy ECG signals

Effect of SNR

Figure 5 shows the effects of SNR on LZ complexity for both the artificial (Fig. 5a) and real (Fig. 5b) ECG signals. As shown in Fig. 5a, the LZ values of the clean ECG plus HF or IM noise increase quickly and monotonically with decreasing SNR. The LZ values of the clean ECG plus PL noise increase until SNR reaches 10 dB and then decrease. The LZ values from the clean ECG plus LF noise decrease when the SNR is below 0 dB. Figure 5b shows the change trend of the LZ values for the real ECG signals. The LZ values from the real ECG plus MA or IM noise increase monotonously with decreasing SNR, and the other types of signals first increase and then decrease.

Fig. 5.

Fig. 5

Effect of SNR on LZ complexity for a artificial and b real noisy ECG signals

In order to further explain the effect of SNR on LZ complexity, we analyzed the relationship between SNR and the number of new patterns for the clean artificial ECG signals plus five types of noise (i.e., HF, PL, LF, IM, and mixed noise). Figure 6 shows the results. It is clear that the number of new patterns from the ECG plus HF, IM, or mixed noise significantly increase with increasing signal length and decreasing SNR, whereas the other LZ values do not show obvious changes. The change trends of the number of new patterns of five synthetic signals with decreasing SNR are consistent with the LZ values of these signals. It is also worth to note that for the clean ECG plus LF or PL noise, the number of new patterns with an SNR of −10 dB is lower than that with an SNR of 0 dB. This is because many signal details are lost during the coarse-graining process and the number of new patterns begins to drop off when SNR decreases to a certain level.

Fig. 6.

Fig. 6

Number of new patterns of artificial ECG plus HF, PL, LF, IM, and mixed-type noise, respectively, at various signal length and SNR settings

The one-way analysis of variance (ANOVA) was also emplyed for analysing the effects of the mixed noise types at various SNR values (20, 15, 10, 5, 0, −5, and −10 dB) on LZ complexity. The ANOVA results (F = 243.709, P = 0.000) indicate that the LZ complexities of the synthetic ECG at various SNR values have significant differences.

Validation Using ROC Curve

The Youden index (YI) was employed for choosing the optimal threshold. It is defined as follows:

YI=Sensitivity+Specificity-1 6

The ROC curve of LZ complexity was used to evaluate classification performance. Figure 7 shows the ROC curve of the LZ values of four synthetic ECGs, i.e., the clean ECG plus HF, LF, PL, and IM noise, respectively. The optimal LZ threshold for the clean ECG plus HF noise is 0.875, those for the clean ECG plus LF and IM noise are 0.150, and that for the clean ECG plus PL noise is 0.225.

Fig. 7.

Fig. 7

ROC curve for artificial ECG plus a HF, b PL, c LF and d IM noise

Figure 8 shows the ROC curve of the mixed-noise-corrupted ECG. The area under the curve is equal to 0.979, which means that the classification performance of the LZ complexity is good. The optimal LZ complexity (cut-off value) is equal to 0.775 for the artificial synthetic ECG plus mixed noise. Table 2 shows the sensitivity, specificity, and YI values for given LZ complexity thresholds for the mixed-noise-corrupted ECG.

Fig. 8.

Fig. 8

ROC curve for artificial ECG plus mixed-type noise

Table 2.

Sensitivity, specificity, and YI values for given LZ complexity thresholds for mixed-noise-corrupted ECG

LZ threshold 0.000 0.025 0.050 0.075 0.100 0.125 0.150 0.175 0.200 0.225 0.250 0.275 0.300 0.325
Sensitivity 1.000 1.000 1.000 1.000 0.000 1.000 1.000 1.000 0.000 1.000 1.000 1.000 1.000 1.000
Specificity 0.000 0.000 0.000 0.000 0.000 0.000 0.000 0.013 0.020 0.033 0.033 0.040 0.073 0.106
YI 0.000 0.000 0.000 0.000 −1.000 0.000 0.000 0.013 −0.980 0.033 0.033 0.040 0.073 0.106
LZ threshold 0.350 0.375 0.400 0.425 0.450 0.475 0.500 0.525 0.550 0.575 0.600 0.625 0.650 0.675
Sensitivity 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000 1.000 0.995 0.995
Specificity 0.147 0.160 0.200 0.273 0.300 0.347 0.367 0.407 0.460 0.520 0.587 0.627 0.660 0.693
YI 0.147 0.160 0.200 0.273 0.300 0.347 0.367 0.407 0.460 0.520 0.587 0.627 0.655 0.688
LZ threshold 0.700 0.725 0.750 0.775 0.800 0.825 0.850 0.875 0.900 0.925 0.950 0.975 1.000
Sensitivity 0.980 0.960 0.945 0.930 0.905 0.800 0.710 0.505 0.155 0.000 0.000 0.000 0.000
Specificity 0.760 0.827 0.873 0.907 0.927 0.987 1.000 1.000 1.000 1.000 1.000 1.000 1.000
YI 0.740 0.787 0.818 0.837 0.832 0.787 0.710 0.505 0.155 0.000 0.000 0.000 0.000

Bold values indicate the highest YI index achieved

Discussion

In this study, we systematically characterized the values of LZ complexity when different types of noise affected the ECG recordings. Figure 3 shows that the different signal types have different LZ values. In general, the LZ complexity is stable, except for the clean ECG signal plus HF noise, which shows fluctuation. The LZ values for the typical signals indicate that the LZ complexity is not only closely associated with the periodicity or randomness of the signals, but is also significantly different between various types of noise. The LZ complexity is low when the signal has an obvious periodicity. The ROC curve analysis shows that the classification performance of the LZ complexity is good, especially for the HF-noise-corrupted ECG.

This study showed that the LZ complexity can indicate the noise level contained in ECG signals. LZ complexity can thus be applied as a metric for assessing the quality of ECG signals corrupted by various types of noise. For ECG signals corrupted by LF noise, we recommend that the baseline should be removed before determining LZ complexity. This study also tested the performance of LZ complexity for real ECG signals. The change trends were consistent with the results for the artificial ECG.

Conclusion

Application of Lempel–Ziv (LZ) complexity for ECG quality assessment was investigated in this study and it concluded that LZ complexity is sensitive to noise level (especially for HF noise) and can thus be a valuable reference index for the assessment of ECG signal quality.

Acknowledgments

This work was supported by the National Natural Science Foundation of China under grants 61473174 and 61201049, the Excellent Young Scientist Awarded Foundation of Shandong Province in China under Grant BS2013DX029, and the China Postdoctoral Science Foundation under Grant 2013M530323. We would like to thank MIT-BIH for their arrhythmia database, which provided invaluable data for our research.

Contributor Information

Shoushui Wei, Phone: +08618769197261, Email: sswei@sdu.edu.cn.

Chengyu Liu, Phone: +08618769197261, Email: bestlcy@sdu.edu.cn.

References

  • 1.Li Q, Clifford GD. Signal quality and data fusion for false alarm reduction in the intensive care unit. Journal of Electrocardiology. 2012;45:596–603. doi: 10.1016/j.jelectrocard.2012.07.015. [DOI] [PubMed] [Google Scholar]
  • 2.Li Q, Clifford GD. Dynamic time warping and machine learning for signal quality assessment of pulsatile signals. Physiological Measurement. 2012;33:1491–1501. doi: 10.1088/0967-3334/33/9/1491. [DOI] [PubMed] [Google Scholar]
  • 3.Clifford GD, Moody GB. Signal quality in cardiorespiratory monitoring. Physiological Measurement. 2012;33:E01–E04. [Google Scholar]
  • 4.Hayn D, Jammerbund B, Schreier G. QRS detection based ECG quality assessment. Physiological Measurement. 2012;33:1449–1462. doi: 10.1088/0967-3334/33/9/1449. [DOI] [PubMed] [Google Scholar]
  • 5.Johannesen L, Galeotti L. Automatic ECG quality scoring methodology: mimicking human annotators. Physiological Measurement. 2012;33:1479–1490. doi: 10.1088/0967-3334/33/9/1479. [DOI] [PubMed] [Google Scholar]
  • 6.Liu, C. Y, Li, P., Zhao, L. N., Liu, F. F., & Wang, R. X. (2011). Real-time signal quality assessment for ECGs collected using mobile phone. In 38th Annual Scientific Conference of Computing in Cardiology (Vol. 38, pp. 357–360).
  • 7.Langley, P., Di Marco, L.Y., King, S., Duncan, D., Di Maria, C., Duan, W. F., Bojarnejad, M., Zheng, D. C., Allen, J., & Murray, A. (2011). An algorithm for assessment of quality of ECGs acquired via mobile telephones. In 38th Annual Scientific Conference of Computing in Cardiology (Vol. 38, pp. 281–284).
  • 8.Zaunseder, S., Huhle, R., & Malberg, H. (2011). Assessing the usability of ECG by ensemble decision trees. In 38th Annual Scientific Conference of Computing in Cardiology (Vol. 38, pp. 277–280).
  • 9.Clifford GD, Behar J, Li Q, Reze I. Signal quality indices and data fusion for determining clinical acceptability of electrocardiograms. Physiological Measurement. 2012;33:1419–1434. doi: 10.1088/0967-3334/33/9/1419. [DOI] [PubMed] [Google Scholar]
  • 10.Zhang YT, Liu CY, Wei SS, Wei CZ, Liu FF. ECG quality assessment using a kernel support vector machine and genetic algorithm with a feature matrix. Journal of Zheijang University SCIENCE C. 2014;15:564–573. doi: 10.1631/jzus.C1300264. [DOI] [Google Scholar]
  • 11.Liu CY, Li P, Di Maria C, Zhao LN, Zhang HG, Chen ZQ. A multi-step method with signal quality assessment and fine-tuning procedure to locate maternal and fetal QRS complexes from abdominal ECG recordings. Physiological Measurement. 2014;35:1665–1684. doi: 10.1088/0967-3334/35/8/1665. [DOI] [PubMed] [Google Scholar]
  • 12.Liu, C.Y., & Li, P. (2013). Systematic methods for fetal electrocardiographic analysis: determining the fetal heart rate, RR interval and QT interval. In 40th Annual Scientific Conference of Computing in Cardiology (Vol. 40, pp. 309–312).
  • 13.Lempel A, Ziv J. On the complexity of finite sequences. IEEE Transactions on Information Theory. 1976;22:75–80. doi: 10.1109/TIT.1976.1055501. [DOI] [Google Scholar]
  • 14.Ziv J, Merhav N. Estimating the number of states of a finite-state source. IEEE Transactions on Information Theory. 1992;38:61–65. doi: 10.1109/18.108249. [DOI] [Google Scholar]
  • 15.Hu J, Gao JB, Príncipe JC. Analysis of biomedical signals by the Lempel-Ziv complexity: the effect of finite data size. IEEE Transactions on Biomedical Engineering. 2006;53:2606–2609. doi: 10.1109/TBME.2006.883825. [DOI] [PubMed] [Google Scholar]
  • 16.Zhang XS, Zhu YS, Thakor NV, Wang ZZ. Detecting ventricular tachycardia and fibrillation by complexity measure. IEEE Transactions on Biomedical Engineering. 1999;46:548–555. doi: 10.1109/10.759055. [DOI] [PubMed] [Google Scholar]
  • 17.Abásolo, D., Alcaraz, R., Rieta, J. J., Hornero, R. (2011). Lempel-Ziv complexity analysis for the evaluation of atrial fibrillation organization. In 8th IASTED International Conference on Biomedical Engineering (Vol. 11, pp. 30–35).
  • 18.Sekar BD, Dou JY, Fu BB, Fei XL, Dong MC. Complexity and similarity analysis of heart sound for cardiovascular disease detection. Biotechnology. 2011;10:316–322. doi: 10.3923/biotech.2011.316.322. [DOI] [Google Scholar]
  • 19.Abásolo D, Hornero R, Gómez C, García M, López M. Analysis of EEG background activity in Alzheimer’s disease patients with Lempel-Ziv complexity and central tendency measure. Medical Engineering & Physics. 2006;28:315–322. doi: 10.1016/j.medengphy.2005.07.004. [DOI] [PubMed] [Google Scholar]
  • 20.Dauwels J, Srinivasan K, Ramasubba Reddy M, Musha T, Vialatte FB, Latchoumane C, Jeong J, Cichocki A. Slowing and loss of complexity in Alzheimer’s EEG: two sides of the same coin? International Journal of Alzheimer’s Disease. 2011;1–10:2011. doi: 10.4061/2011/539621. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Li, L., & Wang, R. P. (2010). Complexity analysis of sleep EEG signal. In 4th International Conference on Bioinformatics and Biomedical Engineering (pp. 1–3).
  • 22.Wu X, Xu J. Complexity and brain function. Acta Biochimica et Biophysica Sinica. 1991;7:103–106. [Google Scholar]
  • 23.Aboy M, Hornero R, Abásolo D, Álvarez D. Interpretation of the Lempel-Ziv complexity measure in the context of biomedical signal analysis. IEEE Transactions on Biomedical Engineering. 2006;53:2282–2288. doi: 10.1109/TBME.2006.883696. [DOI] [PubMed] [Google Scholar]
  • 24.McSharry PE, Clifford GD, Tarassenko L, Smith LA. A dynamical model for generating synthetic electrocardiogram signals. IEEE Transactions on Biomedical Engineering. 2003;50:289–294. doi: 10.1109/TBME.2003.808805. [DOI] [PubMed] [Google Scholar]
  • 25.Kiran GH, Krishna BT. Noise cancellation of ECG signal using adaptive technique. International Journal of Research in Computer and Communication Technology. 2014;3:556–562. [Google Scholar]
  • 26.Bokde PR, Choudhari NK. Implementation of adaptive filtering algorithms for removal of noise from ECG signal. International Journal of Computer Technology & Applications. 2015;6:51–56. [Google Scholar]
  • 27.Moody, G. B., & Mark, R. G. (1990). The MIT-BIH arrhythmia database on CD-ROM and software for use with it. In 17th Annual Scientific Conference of Computing in Cardiology (Vol. 17, pp. 185–188).
  • 28.Moody GB, Mark RG. The impact of the MIT-BIH Arrhythmia Database. IEEE Engineering in Medicine and Biology Magazine. 2001;20:45–50. doi: 10.1109/51.932724. [DOI] [PubMed] [Google Scholar]
  • 29.Goldberger AL, Amaral LAN, Glass L, Hausdorff JM, Ivanov PC, Mark RG, Mietus JE, Moody GB, Peng CK, Stanley HE. PhysioBank, PhysioToolkit, and PhysioNet: Components of a new research resource for complex physiologic signals. Circulation. 2000;101:E215–E220. doi: 10.1161/01.CIR.101.23.e215. [DOI] [PubMed] [Google Scholar]
  • 30.Moody, G. B., Muldrow, W. E., & Mark, R. G. (1984). A noise stress test for arrhythmia detectors. In 17th Annual Scientific Conference of Computing in Cardiology (Vol. 11, pp. 381–384).
  • 31.Zhang XS, Zhu YS, Zhang XJ. New approach to studies on ECG dynamics: extraction and analyses of QRS complex irregularity time series. Medical & Biological Engineering & Computing. 1997;35:467–473. doi: 10.1007/BF02525525. [DOI] [PubMed] [Google Scholar]
  • 32.Zhang XS, Roy RJ, Jensen EW. EEG complexity as a measure of depth of anesthesia for patients. IEEE Transactions on Biomedical Engineering. 2001;48:1424–1433. doi: 10.1109/10.966601. [DOI] [PubMed] [Google Scholar]