Skip to main content
Dentomaxillofacial Radiology logoLink to Dentomaxillofacial Radiology
. 2015 Sep 1;45(1):20150129. doi: 10.1259/dmfr.20150129

Effect of display type, DICOM calibration and room illuminance in bitewing radiographs

Soili Kallio-Pulkkinen 1,2,, Sisko Huumonen 3,4, Marianne Haapea 1,2, Esa Liukkonen 1,2, Annina Sipola 1, Osmo Tervonen 1,2,5, Miika T Nieminen 1,2,5
PMCID: PMC5083887  PMID: 26234536

Abstract

Objectives:

To compare observer performance in the detection of both anatomical structures and caries in bitewing radiographs using consumer grade displays with and without digital imaging and communications in medicine (DICOM) calibration, tablets (third generation iPad; Apple, Cupertino, CA) and 6-megapixel (MP) displays under different lighting.

Methods:

30 bitewing radiographs were blindly evaluated on four displays under bright (510 lx) and dim (16 lx) ambient lighting by two observers. The dentinoenamel junction, enamel and dentinal caries, and the cortical border of the alveolar crests were evaluated. Consensus was considered as reference. Intraobserver agreement was determined. The proportion of equivalent ratings and weighted kappa were used to assess reliability.

Results:

The proportion of equivalent ratings with consensus differed significantly between uncalibrated and DICOM-calibrated consumer grade display in enamel caries in upper and lower molars in bright (p = 0.013 and p = 0.003) lighting, and in dentinal caries in lower molars in both bright (p = 0.022) and dim (p = 0.004) lighting. The proportion also differed significantly between DICOM-calibrated consumer grade and 6-MP display in dentinal caries in lower molars in bright lighting (p = 0.039), tablet and consumer grade display in enamel caries in upper molars (p = 0.017) in bright lighting, tablet and 6-MP display in dentinal caries in lower molars (p = 0.003) in bright lighting and in enamel caries in lower molars (p = 0.012) in dim lighting.

Conclusions:

DICOM calibration improves the detection of enamel and dentinal caries in bitewing radiographs, particularly in bright lighting. Therefore, a calibrated consumer grade display can be recommended as a diagnostic tool for viewing bitewing radiographs.

Keywords: digital radiography, display, calibration, ambient light, bitewing radiographs

Introduction

In clinical practice, bitewing radiographs are taken to detect and monitor the progression of caries and to assess periodontal status. Radiographs are also an important component in assisting with the appropriate treatment and follow-up of patients.1 These radiographs are typically viewed using consumer displays without greyscale calibration (i.e. γ 2.2), usually under high ambient lighting. There is limited knowledge of how observer performance is affected by the type of display, level of ambient light or greyscale calibration.2,3

The American Association of Physicists in Medicine4,5 and the American College of Radiology6 have provided national guidelines regarding acceptable illuminance levels (25–40 lx) in the room used for radiographic interpretation and greyscale calibration. The basic principle is that the displays show greyscale images over an extensive luminance range according to the digital imaging and communications in medicine (DICOM) part 14: greyscale standard display function (GSDF).6,7 This relationship ensures that differences in greyscale are shown optimally to the human eye.8 The sensitivity of the human eye top-detect contrast increases with brightness of pixel elements. The GSDF curve has been derived from Barten's experiments with human observers, determining their contrast thresholds over the complete greyscale range. The difference between two luminances that the human eye can barely detect is defined as just noticeable difference. For a luminance range 0.05–4000 cd m−2, there are 0–1023 just noticeable differences as defined by the DICOM PS3.14.7 In contrast to medical displays, consumer grade displays and tablets are not adjusted to GSDF, which can cause insufficiency in image interpretation. These guidelines rarely come true in dental clinical practice where bitewing radiographs are often interpreted using the same display that is used for maintaining patient records. Furthermore, a significantly higher room illuminance is recommended for clinical work.9

With new technology, potentially useful mobile devices and radiology applications, such as tablets, have become readily available for the imaging community, including dentists. The manufacturers of tablets claim that they are able to display images with high resolution, enabling assessment of different types of digital images. The size of the tablet display is much smaller (e.g. 9.7 inches) than medical displays (e.g. 23 inches), meaning that the image cannot be assessed in full size. The medical displays employ look-up tables calibrated for viewing images, and these displays are supported by the proper configuration and quality control tools. These tools are lacking on tablets or consumer grade displays. The US Food and Drug Administration (FDA) has indicated that these applications will allow clinicians to view medical images on mobile devices.10 Following the FDA report, tablets have recently come to be used in hospitals for different purposes, such as detection of cerebral infarction.11,12 Volonté et al13 considered a tablet device, a tool that is helpful for improving surgical performance as well as for teaching purposes. To our knowledge, only one study has evaluated the use of tablets as devices to detect interproximal caries.14

Several in vitro studies have investigated the effect of varying displays on detecting interproximal caries in digital intraoral radiographs.1421 In these studies, some technical parameters were investigated, i.e. the effect of DICOM calibration, ambient light, luminance and monitor contrast and brightness settings. Hellén-Halme et al15 concluded that the diagnostic accuracy of dentinal caries lesions was significantly higher in ambient light of <50 lx. Odlum et al3 reported that the use of the DICOM GSDF significantly improved the diagnostic accuracy in endodontic and periodontal diagnostics of intraoral radiographs. Ilgüy et al20 found that the accuracy of the medical display was significantly higher than that of the consumer grade display in detecting artificial caries. Araki et al21 found that the diagnostic accuracy of the iPad Air tablet (Apple, Cupertino, CA) to be lower than that of the DICOM-calibrated medical display or the consumer grade display calibrated according to American Association of Physicists in Medicine TG-18 QC, especially in the detection of enamel caries, and there were no differences in the detection of dentinal caries. Otherwise, in these and a few other studies, the authors have reported that overall accuracy of caries detection in intraoral images is not affected by display type, i.e. consumer grade, tablet or medical display.2224 Though, interpretation with consumer grade display and tablet is time-consuming compared with interpretation with medical display.25

So far, one clinical study has reported the effect of varying displays under bright indirect lighting in bitewing radiographs.14 Shintaku et al14 concluded that iPad 2 can effectively show images comparable with a liquid crystal display (LCD). The area under the receiver operating characteristic curve obtained for the iPad 2 and LCD was >0.70, which is considered reasonable for a clinical imaging system.26,27

The aim of this study was to compare observer performance in the detection of both anatomical structures and interpretation of caries in bitewing radiographs using various displays under different lighting.

Methods and materials

30 horizontal molar bitewing radiographs were selected by standard criteria, i.e. the structures to be evaluated had to be clearly visible. An oral and maxillofacial radiologist who did not participate in the evaluation selected, retrospectively, images taken during a 3-month period. All images were subjected to routine quality standards, e.g. limited overlap of the approximal surfaces of the teeth was allowed, and both upper and lower crowns and the maxillary and mandibular alveolar crests had to be seen. Ambient light level was 16 lx. Ambient light was measured from the surface of the display in the direction of the viewer using a luminance metre (Unfors RaySafe Xi, Billdal, Sweden). The selection was made using a 6-megapixel (MP) display (Table 1).

Table 1.

Technical specifications of two sets of four identical displays

Display group Manufacturer/year Model Look-up table Antireflective coating GSDF Frequency of digital imaging and communications in medicine GSDF testing Screen size (inches) Screen type Colour Resolution Brightness (lx) Error (%) Brightness (lx) Error (%) Luminance (cd m−2) Error (%)
Consumer Grade 1 Fujitsu (Kortrijk, Belgium)/2011 P23T6IPS N/A No No 1 year 23.0 LED Yes 1920 × 1080 510.2 0.00 16.2 1.22 206.2 0.34
Consumer Grade 2 Fujitsu/2011 P23T6IPS N/A No No 1 year 23.0 LED Yes 1920 × 1080 510.2   16.4   205.5  
iPad Apple/2012 MD368KS N/A No No No 9.7 LED Yes 2048 × 1536 513.7 0.57 16.2 0.61 344.8 3.55
iPad Apple/2012 MD368KS N/A No No No 9.7 LED Yes 2048 × 1536 510.8   16.3   357.5  
Consumer Grade 1 Fujitsu/2011 P23T6IPS N/A No Yes 1 year 23.0 LED Yes 1920 × 1080 510.0 0.22 16.2 0.61 180.0 0.45
Consumer Grade 2 Fujitsu/2011 P23T6IPS N/A No Yes 1 year 23.0 LED Yes 1920 × 1080 508.9   16.3   179.2  
6 MP1 Barco (Kortrijk, Belgium)/2010 MDCC6130DL 10/12 Yes Yes 6 months 30.4 CCFL Yes 3280 × 2048 509.5 0.02 16.3 1.88 391.7 4.54
6 MP2 Barco/2010 MDCC6130DL 10/12 Yes Yes 6 months 30.4 CCFL Yes 3280 × 2048 509.4   16.0   374.7  

CCFL, cold cathode fluorescent lamp; GSDF, greyscale standard display function; LED, light emitting diode; Look-up table, bit in/bit out; MP, megapixels; N/A, not available.

The images were taken as part of patient treatment at the Department of Diagnostic Radiology, Oulu University Hospital, Oulu, Finland. The bitewings were taken with a storage phosphor plate system (DürrDental AG, Bietingheim-Bissingen, Germany) on the Focus™ Intraoral X-ray unit (Instrumentarium DentalTM; Palodex Group Oy, Tuusula, Finland) by experienced radiographers. The KerrHawe Kwik-Bite® holder (KerrHawe SA, Bioggio, Switzerland) was used. The central beam of the X-ray was positioned to pass at a right angle to the long axis of the tooth and tangentially through the contact area. The parameters used were 70 kV, 7 mA and 0.32 s. The plates were read by the VistaScan Mini Plus image plate scanner (DürrDental AG). All images were stored as DICOM format files. The images matrix size was 1932 × 1496 pixels (0.02 × 0.02-mm pixel size at detector).

The digital archive system used was the neaPACS (Neagen Oy, Oulu, Finland); the case selection system, i.e. the radiology information system used was neaRIS (Neagen Oy) and ESKO Hospital information system (Oulu University Hospital, Oulu, Finland). The viewer used was an HTML 4/5 neaLink (Neagen Oy). A laptop PC (Lifebook S-761 VPro, integrated graphic card: Esprimo C5731E, Fujitsu, Tokyo, Japan) was connected to the consumer grade displays. A power computer (Fujitsu Celsius R570, Fujitsu, graphic card: Barco 5200, Kortrijk, Belgium) was connected to the 6-MP medical display.

Two sets of four identical displays were used for convenience and to save interpretation time. Before the study, luminance was adjusted on the consumer grade displays, tablet, and 6-MP displays using a luminance metre (Unfors RaySafe Xi). The tablet was fixed on the table before adjusting the luminance. The DICOM-calibrated consumer grade displays' luminance was set by displays' adjustments in DICOM-mode GSDF.11 The detailed specifications of the displays are given in Table 1.

The bitewing radiographs were independently observed in random order by two observers: an oral and maxillofacial radiologist with 8 years' experience (Observer 1) and a resident in oral and maxillofacial radiology with 2 years' experience (Observer 2). Rating was based on a three-point scaling system used in the analyses: 1 = definitely not a finding (or not visible), 2 = uncertain and 3 = definitely a finding (or visible).

Altogether ten structures were evaluated: the possible presence of enamel and dentinal caries in the first and second upper and lower permanent molars, and the visibility of the dentinoenamel junction in the first lower permanent molar and cortical border of the alveolar crests. Both anatomical structures and pathology were chosen because they provided low- and high-contrast details.

The observers evaluated 30 bitewing radiographs from four displays. The consumer grade display without DICOM calibration and the tablet were evaluated during the first evaluation, the DICOM-calibrated consumer grade display in the second and the 6-MP display during the third evaluation with 3 weeks' delay between the evaluation sessions. Evaluations were made under bright (510 lx) and dim (16 lx) lighting. There was a 1-h delay between the room illuminance settings. The radiographs were displayed in random order at each evaluation to minimize the memory effect. Evaluation time of 1 min per one radiograph was allowed during which all ten structures were evaluated.

A consensus between the two observers, which was considered as a reference, was made within each structure in all 30 bitewing radiographs 6 months after the initial viewing. In consensus reading session, the 6-MP display was used under ambient lighting of 16 lx (Table 1). Intraobserver agreement was determined by Observer 1 by re-evaluating 15 bitewing radiographs 6 months after the initial viewing using the 6-MP display under ambient lighting of 16 lx (Table 1).

Statistical methods

Agreement between the observers and the consensus was calculated as a proportion of equivalent ratings for all structures. Additionally, the number and proportion of positive ratings (when positive in consensus) and negative ratings (when negative in consensus) were calculated. Rating of the dentinoenamel junction and cortical border of the alveolar bone was dichotomized as visible or unable to evaluate (i.e. not visible or uncertain). In order to evaluate the reliability of the ratings, linearly weighted kappa (κw) was calculated for enamel and dentinal caries.28 Kappa statistics were interpreted as follows: <0, poor (less than change); 0.00–0.20, slight; 0.21–0.40, fair; 0.41–0.60, moderate; 0.61–0.80, substantial; 0.81–0.99, almost perfect agreement.29 McNemar's test was used to compare the proportions of equivalent ratings separately in bright and dim lighting between (1) uncalibrated and DICOM-calibrated consumer grade displays; (2) DICOM-calibrated consumer grade and 6-MP displays; (3) tablet and other displays; and (4) between bright and dim lighting for all displays. Only the radiographs that were evaluated successfully in all displays were selected for the analyses. p-values <0.05 were considered statistically significant. To analyse the intraobserver and interobserver reliabilities, the proportions of equivalent ratings and kappa statistics (κw) were calculated for enamel and dentinal caries, and the numbers of radiographs were rated as visible for dentinoenamel junction and cortical border of the alveolar bone. For intraobserver reliability, the initial viewing by Observer 1 using the 6-MP display under dim lighting and the reviewing by Observer 1 were used. For interobserver reliability, the viewings by Observers 1 and 2 using the 6-MP display under dim lighting were used. IBM SPSS Statistics®, v. 22.0 (IBM Corporation, Armonk, NY; formerly SPSS Inc., Chicago, IL) was used in the statistical analyses.

Results

The cortical border of alveolar crests was visible on all displays in both lighting. The proportion of equivalent ratings with consensus in visibility of the dentinoenamel junction was the lowest (65%) with consumer grade display in bright lighting and the highest (95%) with 6-MP display in both bright and dim lighting (Table 2). Out of 35 radiographs in which the dentinoenamel junction was rated as certainly visible in consensus, only 23 (66%) were rated as visible with uncalibrated consumer grade display, whereas 34 (97%) were rated as visible with 6-MP display (Table 2).

Table 2.

A proportion of equivalent ratings in visibility of the dentinoenamel junction, and numbers and proportions of positive ratings in different displays when rated in consensus (n = 35), separately in bright and dim lighting

Display Bright
Dim
%Eqv. #pos. (%) %Eqv. #pos. (%)
Tablet 83.8 29 (82.9) 81.8 29 (82.9)
Consumer grade 64.9 23 (65.7) 78.4 29 (82.9)
Digital imaging and communications in medicine-calibrated consumer grade 83.8 30 (85.7) 89.2 32 (91.4)
6 megapixels 94.6 34 (97.1) 94.6 34 (97.1)

%Eqv., proportion of equivalent ratings with consensus; #pos. (%), number and proportion of positive findings when in consensus.

The proportion of equivalent ratings between observers and consensus was statistically significantly lower in uncalibrated than in DICOM-calibrated consumer grade display in enamel caries in upper and lower molars in bright (p = 0.013 and p = 0.003, respectively), and in dentinal caries in lower molars in both bright (p = 0.022) and dim (p = 0.004) lighting (Table 3). DICOM-calibrated consumer grade display had a lower proportion of equivalent ratings than 6-MP display in dentinal caries in lower molars in bright lighting (p = 0.039).

Table 3.

Comparison of agreement between different displays in enamel and dentinal caries in upper (dd. 26, 27) and lower (dd. 37, 36) molars compared with the consensus rating, separately in bright and dim lighting

n = 74 Pathology/display Bright
Dim
p-valuec
%Eqv. κw (SE) p-valuea p-valueb %Eqv. κw (SE) p-valuea p-valueb
Enamel caries, upper molars (d26 and d27)
 Tablet 77.0 0.61 (0.09)     70.3 0.48 (0.10)     0.23
 Consumer grade 60.8 0.36 (0.09) 0.013 0.017 71.6 0.56 (0.09) 0.58 >0.99 0.08
 DICOM-calibrated consumer grade 75.7 0.57 (0.09) 0.17 >0.99 75.7 0.52 (0.10) 0.34 0.45 >0.99
 6 MP 85.1 0.70 (0.08)   0.18 81.1 0.58 (0.10)   0.08 0.55
Enamel caries, lower molars (d36 and d37)
 Tablet 85.1 0.71 (0.09)     82.4 0.54 (0.11)     0.79
 Consumer grade 75.7 0.49 (0.11) 0.003 0.09 79.7 0.53 (0.11) 0.11 0.77 0.51
 DICOM-calibrated consumer grade 90.5 0.83 (0.07) >0.99 0.39 87.8 0.73 (0.09) 0.18 0.42 0.63
 6 MP 91.9 0.77 (0.09)   0.23 94.6 0.84 (0.08)   0.012 0.63
Dentinal caries, upper molars (d26 and d27)
 Tablet 73.0 0.42 (0.11)     79.7 0.54 (0.11)     0.27
 Consumer grade 67.6 0.42 (0.09) 0.79 0.48 79.7 0.62 (0.08) 0.73 >0.99 0.035
 DICOM-calibrated consumer grade 70.3 0.43 (0.10) 0.42 0.80 77.0 0.54 (0.10) >0.99 0.75 0.27
 6 MP 75.7 0.50 (0.10)   0.79 77.0 0.48 (0.10)   0.77 >0.99
Dentinal caries, lower molars (d36 and d37)
 Tablet 74.3 0.46 (0.11)     75.7 0.39 (0.11)     >0.99
 Consumer grade 66.2 0.30 (0.10) 0.022 0.18 68.9 0.31 (0.11) 0.004 0.23 0.75
 DICOM-calibrated consumer grade 78.4 0.56 (0.08) 0.039 0.58 81.1 0.50 (0.11) >0.99 0.29 0.63
 6 MP 89.2 0.75 (0.08)   0.003 82.4 0.54 (0.11)   0.23 0.13

%Eqv., proportion of equivalent ratings; DICOM, digital imaging and communications in medicine; MP, megapixels; κw, linearly weighted kappa; SE, standard error.

The p-values <0.05 are in bold.

a

Significance from McNemar's test between consumer grade and DICOM-calibrated consumer grade (upper p-value), and DICOM-calibrated consumer grade and 6 MP (lower p-value).

b

Significance from McNemar's test between tablet and other displays.

c

Significance from McNemar's test between bright and dim lighting.

The proportion of equivalent ratings was higher in tablet than in consumer grade display in enamel caries in upper molars (p = 0.017) and lower than in 6-MP display in dentinal caries in lower molars (p = 0.003) in bright lighting and in enamel caries in lower molars (p = 0.012) in dim lighting (Table 3).

Uncalibrated consumer grade display worked poorly in interpreting the corresponding findings (Table 4). Overall, none of the displays had high proportions of interpreting dentinal caries.

Table 4.

Numbers and proportions of positive and negative findings in different displays in enamel and dentinal caries in upper (dd. 26, 27) and lower (dd. 37, 36) molars when positive or negative finding, respectively, in consensus, separately in bright and dim lighting

n = 74 Pathology/display Positive findings
Negative findings
#Cons. Bright
Dim
#Cons. Bright
Dim
#pos. (%) #pos. (%) #neg. (%) #neg. (%)
Enamel caries, upper molars (d26 and d27) 24     48    
 Tablet   13 (54.2) 12 (50.0)   43 (89.6) 38 (79.2)
 Consumer grade   7 (29.2) 12 (50.0)   36 (75.0) 39 (81.3)
 DICOM-calibrated consumer grade   15 (62.5) 16 (66.7)   39 (81.3) 38 (79.2)
 6 MP   22 (91.7) 17 (70.8)   40 (83.3) 41 (85.4)
Enamel caries, lower molars (d36 and d37) 16     56    
 Tablet   11 (68.8) 8 (50.0)   50 (89.3) 51 (91.1)
 Consumer grade   7 (43.8) 7 (43.8)   47 (83.9) 50 (89.2)
 DICOM-calibrated consumer grade   13 (81.3) 12 (75.0)   52 (92.9) 51 (91.1)
 6 MP   13 (81.3) 13 (81.3)   53 (94.6) 55 (98.2)
Dentinal caries, upper molars (d26 and d27) 9     60    
 Tablet   5 (55.6) 6 (66.7)   46 (76.7) 50 (83.3)
 Consumer grade   3 (33.3) 4 (44.4)   43 (71.7) 52 (86.7)
 DICOM-calibrated consumer grade   4 (44.4) 5 (55.6)   45 (75.0) 50 (83.3)
 6 MP   7 (77.8) 5 (55.6)   48 (80.0) 50 (83.3)
Dentinal caries, lower molars (d36 and d37) 13     54    
 Tablet   5 (38.5) 3 (23.1)   47 (87.0) 50 (92.6)
 Consumer grade   2 (15.4) 3 (23.1)   43 (79.6) 45 (83.3)
 DICOM-calibrated consumer grade   2 (15.4) 4 (30.8)   50 (92.6) 51 (94.4)
 6 MP   8 (61.5) 6 (46.2)   54 (100.0) 53 (98.1)

#Cons., number of positive/negative findings (i.e. certainly a finding/not a finding) in consensus; #neg. (%), number and proportion of negative findings when not a finding in the consensus; #pos. (%), number and proportion of positive findings when a finding in the consensus. DICOM, digital imaging and communications in medicine; MP, megapixels.

Consumer grade display was significantly better in dim than in bright lighting in detecting dentinal caries in the upper molars (p = 0.035).

The intraobserver reliability was fair in enamel caries in upper (κw = 0.36) and moderate in lower (κw = 0.44) molars, and substantial in upper and lower dentinal caries (κw = 0.66 and 0.64; Table 5). On initial viewing, the dentinoenamel junction and the border of alveolar crests were evaluated as visible in all 15 radiographs, whereas on reviewing, the dentinoenamel junction was once evaluated as an uncertain finding.

Table 5.

Intraobserver and interobserver reliability assessed using 6 megapixel display in dim lighting

Pathology Intraobserver reliability,a 30 structures
Interobserver reliability,b 56 structures
%Eqv. κw %Eqv. κw
Enamel caries, upper molars (d26 and d27) 86.7 0.36 66.1 0.29
Enamel caries, lower molars (d36 and d37) 90.0 0.44 87.5 0.63
Dentinal caries, upper molars (d26 and d27) 80.0 0.66 64.3 0.46
Dentinal caries, lower molars (d36 and d37) 93.3 0.64 85.7 0.60

%Eqv., proportion of equivalent ratings; κw, linearly weighted kappa.

a

Between the initial viewing of Observer 1 and reviewing of 15 radiographs.

b

Between the viewings of Observers 1 and 2 (28 radiographs successfully rated by both observers).

The interobserver reliability between the readers was fair at upper and substantial in lower enamel caries (κw = 0.29 and κw = 0.63, respectively), and moderate in upper and substantial in lower dentinal caries (κw = 0.46 and κw = 0.60, respectively; Table 5). The dentinoenamel junction was evaluated as visible in all 28 radiographs by Observer 1 and in 26 radiographs by Observer 2, and the border of alveolar crests was evaluated as visible in all 28 radiographs by both observers.

Discussion

In the present study, DICOM-calibrated consumer grade display was significantly better than uncalibrated consumer grade display, slightly better than tablet and almost equal to medical displays in the interpretation of enamel and dentinal caries in bitewing radiographs. Tablet was slightly worse than medical displays in both lighting. Overall, with consumer grade display the proportions of equivalent ratings were higher under dim than bright lighting and were statistically significantly better for the interpretation of dentinal caries in upper molars. High-contrast structures such as the dentinoenamel junction and the cortical border of alveolar crest were seen almost equally well in all displays under both lighting conditions in bitewing radiographs.

The ability to detect caries lesions on dental radiographs depends on lesion grade. In this study, to mimic clinical conditions, lesion grade was not validated. Deeper lesions are easier to detect than relatively superficial ones.30 Early radiographic detection of caries lesions in the enamel is important, as the probability of cavitation is low and the prospect of interceptive treatment is good. Hellén-Halme et al15 concluded that it was easier to detect dentinal caries in dim conditions than in bright conditions, whereas with enamel caries, there were no differences between the lighting when the display was adjusted for optimal brightness and contrast. Pakkala et al19 also concluded that different displays and room illuminance levels did not affect the overall accuracy of intraoral radiographs in enamel caries detection. In the present study, DICOM calibration significantly improved enamel caries interpretation as compared with uncalibrated display in bright lighting. 6-MP display was significantly better than tablet in detecting enamel caries in lower molars in dim lighting, whereas tablet was significantly better than consumer grade display in detecting enamel caries in upper molars in bright lighting.

The purpose of the GSDF calibration is to ensure the optimal presentation and perception of contrast in digital radiography and to ensure that radiographs are presented unaltered over time and different locations. This standard is commonly used in medical displays. These displays are expensive, and the replacement of medical displays with consumer grade displays could translate into cost savings. Many dentists view radiographs on ordinary consumer grade displays without calibration, probably owing to lack of knowledge of how calibration may affect the diagnostic accuracy of digital radiographs.

Previously, Hellén-Halme et al16 found that there was no difference between bright and dim lighting in the ability to detect caries lesions in digital intraoral radiographs when the display was calibrated to DICOM part 14 standards. McIlgorm et al31 concluded that standardizing consumer grade display devices to the DICOM part 14: GSDF can improve the presentation of dental radiological images. They also concluded that a DICOM-calibrated consumer grade monitor is capable of displaying an image quality that is equally preferred to a DICOM-calibrated medical grade monitor for 8-bit dental images.32 In these studies,31,32 the ambient light level was set between 25 and 40 lx. Recently, Kallio-Pulkkinen et al33 concluded that the DICOM-calibrated consumer grade display is adequate for the detection of anatomical structures and pathology in panoramic radiographs also in bright (510 lx) ambient light.33 In the present study, the DICOM-calibrated consumer grade display was significantly better for the interpretation of enamel caries in lower and upper molars and dentinal caries in lower molars in bright lighting and in dentinal caries in lower molars in both lighting conditions. It can be concluded that the DICOM calibration can improve dentinal caries detection especially in bright lighting.

According to the FDA, tablets are not intended to replace full workstations and are indicated for use only when there is no access to a workstation.10 All participants in the FDA Task Group agreed that the device was sufficient for diagnostic image interpretation under the recommended lighting.34 Recently, Shintaku et al14 concluded that the iPad 2 can effectively display images, comparably to the evaluated LCD monitor, for the evaluation of interproximal caries under bright light conditions. Furthermore, Shintaku et al concluded that image size did not affect the ability to identify dental caries on the iPad 2 compared with a 24-inch LCD monitor. Similar findings were made by Cederberg et al23 and Isidor et al.24 Cederberg et al23 compared three medical-grade cathode ray tube displays and a laptop for the detection of interproximal caries. The authors did not detect a significant difference in diagnostic accuracy and luminance values, monitor size or resolution between these displays. Isidor et al24 compared five LCD displays with different sizes and resolutions for the detection of caries. The authors did not detect statistically significant differences between the displays. In the present study, the resolution of the consumer grade display was lower than that of the tablet and the 6-MP display or the matrix size of the displayed bitewing radiographs. Thus, viewing images on the medical display may have benefitted from the higher resolution. However, the display resolution or matrix size of the images did not affect observer performance in the detection of anatomical structures or interpretation of caries in bitewing radiographs when comparing the calibrated and non-calibrated consumer grade display. Furthermore, the tablet was slightly worse than medical display and almost equal to the DICOM-calibrated display in interpreting of enamel and dentinal caries in bitewing radiographs in both lighting conditions.

There are several limitations pertaining to the present clinical investigation. First, instead of measuring the ability to make a correct diagnosis, a subjective method was used to evaluate the underlying differences in image quality between modalities. Second, a real golden standard method was not used; instead, consensus with two observers with the 6-MP display was used as reference. Previously, histology has been used as the gold standard in in vitro studies; however, such an approach is of limited value for a clinical setting.35

In conclusion, DICOM calibration improves the detection of enamel and dentinal caries in bitewing radiographs, particularly in bright lighting. Therefore, a DICOM-calibrated consumer grade display can be recommended as a diagnostic tool with bitewing radiographs. Surprisingly, a tablet display used to view bitewing radiographs can almost equal a DICOM-calibrated display in the detection of enamel and dentinal caries in both bright and dim lighting. Further research is needed on the clinical use of tablet devices.

Contributor Information

Soili Kallio-Pulkkinen, Email: soili.kallio-pulkkinen@oulu.fi.

Sisko Huumonen, Email: sihahu@utu.fi.

Marianne Haapea, Email: marianne.haapea@oulu.fi.

Esa Liukkonen, Email: esa.liukkonen@ppshp.fi.

Annina Sipola, Email: annina.sipola@ppshp.fi.

Miika T Nieminen, Email: miika.nieminen@oulu.fi.

References

  • 1.White SC, Pharoah MJ. Oral radiology: principles and interpretation. 6th edn. St. Louis, MI: Mosby; 2009. pp. 297–313. [Google Scholar]
  • 2.Hellén-Halme K, Nilsson M, Petersson A. Digital radiography in general dental practice: a field study. Dentomaxillofac Radiol 2007; 36: 249–55. doi: 10.1259/dmfr/95125494 [DOI] [PubMed] [Google Scholar]
  • 3.Odlum N, Spalla G, Van Assche N, Vandenberghe B, Jacobs R, Quirynen M, et al. Preliminary display comparison for dental diagnostic applications. Proc SPIE 8318; 2012. doi: 10.1117/12.917043 [DOI] [Google Scholar]
  • 4.American Association of Physicists in Medicine (AAPM). Task Group 18 assessment of display performance for medical imaging systems. [Cited June 2015]. Available from: http://www.duke.edu/samei/th18.html
  • 5.Samei E, Badano A, Chakraborty D, Compton K, Cornelius C, Corrigan K, et al. ; AAPM TG18 report. Assessment of display performance for medical imaging systems: executive summary of AAPM TG18 report. Med Phys 2005; 32: 1205–25. doi: 10.1118/1.1861159 [DOI] [PubMed] [Google Scholar]
  • 6.American College of Radiology (ACR). ACR technical standard for electronic practice of medical imaging. [Updated 2012, accessed 7 February 2013]. Available from: gm.acr.orgSecondaryMainMenuCategories/quality_safety/guidelines/med_phys/electronic_practice.aspx.
  • 7.Digital Imaging and Communication in Medicine. DICOM PS3.14 2014-Greyscale Standard Display Function. 2014 NEMA. [Cited 1 July 2014]. Available from: http://dicom.nema.org/MEDICAL/dicom/current/output/html/part14
  • 8.Barten PGJ. Contrast sensitivity of the human eye and its effects on image quality. Knegsel, Netherlands: HV Press; 1999. [Google Scholar]
  • 9.Swedish Standards Institute. Light and lighting—lighting of work places—Part 1: indoor work places. [in English]. Stockholm, Sweden: Swed St Ins7semi; 2011. SS-EN 12464–1. [Google Scholar]
  • 10.U.S. Food and Drug Administration. Press release: FDA clears first diagnostic radiology applications for mobile devices. [Updated 7 February 2013]. http://www.fda.gov/NewsEvents/Newsroom/PressAnnouncements/ucm242295.htm
  • 11.McLaughlin P, Neill SO, Fanning N, McGarrigle AM, Connor OJ, Wyse G, et al. Emergency CT brain: preliminary interpretation with a tablet device: image quality and diagnostic performance of the Apple iPad. Emerg Radiol 2012; 19: 127–33. doi: 10.1007/s10140-011-1011-2 [DOI] [PubMed] [Google Scholar]
  • 12.Yoshimura K, Nihashi T, Ikeda M, Ando Y, Kawai H, Kawakami K, et al. Comparison of liquid crystal display monitors calibrated with gray-scale standard display function and with γ 2.2 and iPad: observer performance in detection of cerebral infraction on brain CT. AJR Am J Roentgenol 2013; 200: 1304–9. doi: 10.2214/AJR.12.9096 [DOI] [PubMed] [Google Scholar]
  • 13.Volonté F, Robert JH, Ratib O, Triponez F. A lung segmentectomy performed with 3D reconstruction images available on the operating table with an iPad. Interact Cardiovasc Thorac Surg 2011; 12: 1066–8. doi: 10.1510/icvts.2010.261073 [DOI] [PubMed] [Google Scholar]
  • 14.Shintaku WH, Scarbecz M, Venturin JS. Evaluation of interproximal caries using the Tablet 2 and a liquid crystal display display. Oral Surg Oral Med Oral Pathol Oral Radiol 2012; 113: e40–4. doi: 10.1016/j.oooo.2011.11.008 [DOI] [PubMed] [Google Scholar]
  • 15.Hellén-Halme K, Petersson A, Warfvinge G, Nilsson M. Effect of ambient light and monitor brightness and contrast settings on the detection of approximal caries in digital radiographs. An in vitro study. Dentomaxillofac Radiol 2008; 37: 380–4. doi: 10.1259/dmfr/26038913 [DOI] [PubMed] [Google Scholar]
  • 16.Hellén-Halme K, Nilsson M, Petersson A, Wafvinge G, Nilsson M. Effect of monitors on approximal caries detection in digital radiographs-standard versus precalibrated DICOM Part 14 displays: an in vitro study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009; 107: 716–20. doi: 10.1016/j.tripleo.2008.12.011 [DOI] [PubMed] [Google Scholar]
  • 17.Hellén-Halme K, Lith A. Carious lesions: diagnostic accuracy using pre-calibrated monitor in various ambient light levels: an in vitro study. Dentomaxillofac Radiol 2013; 42: 1–7. doi: 10.1259/dmfr.20130071 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Hellén-Halme K, Lith A. Effect of ambient light level at the monitor surface on digital radiographic evaluation of approximal carious lesions: an in vitro study. Dentomaxillofac Radiol 2012; 41: 192–6. doi: 10.1259/dmfr/15422221 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Pakkala T, Kuusela M, Ekholm M, Wenzel A, Haiter-Neto F, Kortesniemi M. Effect of varying displays and room illuminance on caries diagnostic accuracy in digital dental radiographs. Caries Res 2012; 46: 568–74. doi: 10.1159/000341218 [DOI] [PubMed] [Google Scholar]
  • 20.Ilgüy M, Dincer S, Ilgüy D, Bayirli G. Detection of artificial occlusal caries in a phosphor imaging plate system with two types of LCD monitors versus different films. J Digit Imaging 2009; 22: 242–9. doi: 10.1007/s10278-008-9146-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Araki K, Fujikura M, Sano T. Effect of display monitor devices on intra-oral radiographic caries diagnosis. Clin Oral Investig Jan 2015. Epub ahead of print. doi: 10.1007/s00784-015-1401-z [DOI] [PubMed] [Google Scholar]
  • 22.Ludlow JB, Abreu M, Jr. Performance of film, desktop monitor and laptop displays in caries detection. Dentomaxillofac Radiol 1999; 28: 26–30. doi: 10.1038/sj.dmfr.4600400 [DOI] [PubMed] [Google Scholar]
  • 23.Cederberg RA, Frederiksen NL, Benson BW, Shulman JD. Influence of the digital image display monitor on observer performance. Dentomaxillofac Radiol 1999; 28: 203–7. doi: 10.1038/sj/dmfr/4600441 [DOI] [PubMed] [Google Scholar]
  • 24.Isidor S, Faaborg-Andersen M, Hintze H, Kirkevang LL, Frydenberg M, Haiter-Neto F, et al. Effect of monitor display on detection of approximal caries lesions in digital radiographs. Dentomaxillofac Radiol 2009; 38: 537–41. doi: 10.1259/dmfr/21071028 [DOI] [PubMed] [Google Scholar]
  • 25.Krupinski EA, Siddqui KM, Siegel E, Shretsa R, Grant E, Roehring H, et al. Influence of 8-bit vs. 11-bit digital displays on observer performance and visual search: a multi-center evaluation. J Soc Inform Disp 2007; 15: 385–90. doi: 10.1889/1.2749324 [DOI] [Google Scholar]
  • 26.Metz CE. Some practical issues of experimental design and data analysis in radiological ROC studies. Invest Radiol 1989; 24: 234–45. [DOI] [PubMed] [Google Scholar]
  • 27.Metz CE. Receiver operating characteristic analysis: a tool for the quantitative evaluation of observer performance and imaging systems. J Am Coll Radiol 2006; 3: 413–22. doi: 10.1016/j.jacr.2006.02.021 [DOI] [PubMed] [Google Scholar]
  • 28.Fleiss JL, Levin B, Paik MC. Statistical Methods for Rates and Proportions. 3rd edn. New York, NY: John Wiley & Sons Inc.; 2003. [Google Scholar]
  • 29.Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics 1977; 33: 159–74. doi: 10.2307/2529310 [DOI] [PubMed] [Google Scholar]
  • 30.Cederberg RA, Frederiksen NL, Benson BW, Schulman JD. Effect of different background lighting conditions on diagnostic performance of digital and film images. Dentomaxillofac Radiol 1998; 27: 293–7. doi: 10.1038/sj/dmfr/4600377 [DOI] [PubMed] [Google Scholar]
  • 31.McIlgorm DJ, Lawinski C, Ng S, McNulty JP. Could standardizing “commercial off-the-shelf” (COTS) monitors to the DICOM part 14: GSDF improve the presentation of dental images? A visual grading characteristics analysis. Dentomaxillofac Radiol 2013; 42: 20130121. doi: 10.1259/dmfr.20130121 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.McIlgorm DJ, McNulty JP. DICOM part 14: GSDF-calibrated medical grade monitor vs a DICOM part 14: GSDF-calibrated “commercial off-the-shelf” (COTS) monitor for viewing 8-bit dental images. Dentomaxillofac Radiol 2015; 44, 20140148. doi: 10.1259/dmfr.20140148 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Kallio-Pulkkinen S, Haapea M, Liukkonen E, Huumonen S, Tervonen O, Nieminen MT. Comparison between digital imaging and communication in medicine-calibrated and uncalibrated consumer grade and 6-MP displays under different lighting conditions in panoramic radiography. Dentomaxillofac Radiol 2015; 44: 20140365. doi: 10.1259/dmfr.20140365 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Samei E, Ranger NT, Delong DM. A comparative contrast-detail study of five medical displays. Med Phys 2008; 35: 1538–64. doi: 10.1118/1.2868780 [DOI] [PubMed] [Google Scholar]
  • 35.Wenzel A. A review of dentists' use of digital radiography and caries diagnosis with digital systems. Dentomaxillofac Radiol 2006; 35: 307–14. doi: 10.1259/dmfr/64693712 [DOI] [PubMed] [Google Scholar]

Articles from Dentomaxillofacial Radiology are provided here courtesy of Oxford University Press

RESOURCES