Abstract
Objective
To investigate the value of different kVp images of dual-energy CT (DECT) for the detection of liver metastases.
Methods
20 Patients with uveal melanoma were investigated with DECT of the liver. In each patient contrast-enhanced DECT imaging with arterial delay was performed. Number and size of metastases were documented on arterial phase 80-kVp images, virtual 120-kVp images and following angiographic images (DSA) as part of hepatic chemoperfusion. Attenuation of metastases and several anatomic regions, subjective (image noise, image quality) and objective (SNR, CNR) parameters were documented.
Results
The mean number of liver metastases detected was significant higher on 80-kVp images than on virtual 120-kVp/DSA images (5.6 ± 2.1 vs. 4.1 ± 1.8/4.3 ± 1.6); (p < 0.001). All lesions sizes were significant better detected with 80 kVp images than with virtual 120 kVp and DSA-Images (80 kVp vs. 120 kVp: <10 mm: 34 vs. 19, p < 0.05; 10–20 mm:, 33 vs. 25, p < 0.05; >20 mm: 56 vs. 42, p < 0.05/80 kVp vs. DSA: <10 mm: 34 vs. 18 p < 0.05; 10–20 mm: 33 vs. 24, p < 0.05; >20 mm: 56 vs. 41, p < 0.05). Number of detected small lesions <10 mm with 120 kVp compared to DSA-images were significant higher (19 vs. 13; p < 0.05), lesions 10–20 mm and >20 mm were measured statistically equally. Noise, SNR and CNR of 80 kVp images were higher compared to 120 kVp images. Image quality of 120 kVp images was higher compared to 80 kVp images.
Conclusion
Low-kVp images of DECT datasets are more sensitive in detecting liver metastases of patients with uveal melanoma than virtual 120 kVp- and DSA images.
Keywords: Dual energy CT, Liver, Uveal melanoma, Angiography, Staging
1. Introduction
Uveal melanoma (UM) is a rare disease arising from the pigmented uveal tract of the eye. The incidence in Europe is 4.4 cases per million, varying between 2 in the south to 8 in the north of Europe [1]. Around 5% of patients with UM have distant metastases at diagnosis. Metastases appear usually within a median of 3 years after the treatment of the primary tumor with a range between 1 and 10 years. There is an unexplained hepato-tropism with the liver being the first site of metastases in up to 90% of patients and is the only site in 46% of them [2], [3]
A recent development in CT has been the introduction of dual-source technology [4].
One important advantage of dual-source CT compared with a single-source system is the option to use the two tubes at different tube currents offering differentiation of materials of non-equal density. Previous study demonstrated the feasibility of contrast-enhanced dual-energy CT for the detection of hypervascular liver lesions in patients with hepatocellular carcinoma [5]. Higher detection rates were explained by higher attenuation of iodine in the low-kVp images in comparison to 140-kVp and virtual 120-kVp images.
FDG-PET-CT and MRI were described as reliable and accurate diagnostic methods in previous studies [6], [7], [8], [9]. So far however, there is no study that investigated the sensitivity of the detection of hepatic metastases of uveal melanoma in DECT.
The aim of our study was to investigate to what extent the low kVp images of DECT are suitable compared to the virtual 120 kVp and DSA images for detection of liver metastases in patients with uveal melanoma. DECT was performed in each patient as part of the regular staging and planning examination before and 3 anf 6 monh after transarterial hepatic chemoperfusion. Angiographic image data sets were consulted as the reference imaging technique for comparison [10], [11].
2. Materials and methods
2.1. Patient population
Between September 2009 and November 2011, 20 consecutive patients with uveal melanoma (13 men, 7 women; mean age 61 years ± 17 years) underwent dual-energy CT (DECT) for regular evaluation before transarterial chemoperfusion and 3 anf 6 months after therapy as regular follow up. The study was approved by our institutional review board and all patients provided written informed consent before participation (Fig. 1, Fig. 2).
2.2. Imaging protocoll and parameters
All CT examinations were performed with a dual-source multi-detector CT (Somatom Definition™ Dual Source; Siemens Medical Solutions, Forchheim, Germany). This system consists of two x-ray tubes mounted on one gantry at a 90° angle [4] and two corresponding 64-section detectors. One detector (detector A) covers a 50-cm field of view and the other detector (detector B) covers a 26-cm field of view [4]. Patients were positioned supine on the table. Patients were positioned slightly off centre to the left to ensure complete coverage of the liver by the smaller field of view of detector B. After intravenous injection of a non-ionic contrast agent (1.5 ml per kilogram of body weight, Xenetix 300™, Guerbet, Sulzbach, Germany) bolus tracking was started in the abdominal aorta at the level of the coeliac trunk. Dual-energy datasets were acquired in the arterial and portal venous (bolus tracking) phases from the dome of the liver to the iliac crest by operating tube A at 140 kVp and a reference value of 96 mAs and tube B at 80 kVp and a reference value of 404 mAs. For both tubes, an online dose modulation (Care DOSE 4D™, Siemens Medical Solutions) was used. For the dual-energy CT, collimation was set to 14 × 1.2 mm on both detectors. For all data sets, the gantry rotation speed was 0.5 s.
DSA (TM Philips Allura, Philips Healthcare, Best, The Netherlands and Toshiba Infinix DP-i, Toshiba Medical Systems, Tokyo, Japan) was performed via a transfemoral arterial access the day after DECT. The celiac trunk and the superior mesenteric artery were explored utilizing a macro catheter (5 French Sidewinder-1-Catheter, 5 French Cobra Catheter, Terumo Europe, Leuven, Belgium). After completing overviews of the hepatic arteries a selective angiography of the left and right liver lobe was performed using a microcatheter (Rebar Microcatheter, 0.027 in; EV3, Plymouth, MN, USA) and a contrast power injection (Flow 3 ml/s, volume 12 ml, Xenetix 300™, Guerbet, Sulzbach, Germany).
2.3. Postprocessing and image reconstruction
Axial post-contrast images were reconstructed by using a section thickness and an increment of 3 mm. The dual-energy CT generates three different series of images: 80-kV images, 140-kV images and weighted-average images, which are based on attenuation information on images obtained from both detectors, using 70% information from the high-kVp and 30% from the low-kVp imaging. Weighted-average images are similar to 120-kVp imaging of the abdomen. Images were loaded onto a dedicated dual-energy postprocessing workstation (syngo MMWP; Siemens Medical Solutions, Forchheim, Germany).
2.4. Measurements of metastases and image quality
Two experienced (CT-experience 15 years and DSA-experience 10 years for both) radiologists read all the cases together. On a dedicated workstation (syngo MMWP, Siemens Medical Solutions), 3-mm CT and angiography images were displayed side by side (window width, 350 HU; window level, 50 HU). CT attenuation was determined in 7 regions using regions of interest of 0.2–1.0 cm2 in size: hepatic metastases, normal hepatic parenchyma; aorta at the level of the coeliac trunk; psoas muscle; perihepatic fat; and air outside the patient body. For both CT image sets, the radiologists rated image noise on a five-point scale: 1, none; 2, minimal; 3, mild; 4, moderate; and 5, severe. Overall image quality was rated with a different five-point scale as follows: 1, excellent; 2, good; 3, fair; 4, poor; and 5, not diagnostic. Both readers reported all hypervascular liver metastases together for 80 kVp-, virtual-120-kVp and DSA images according to size (<10 mm; 10–20 mm; >20 mm). Follow up scans 3 and 6 months after therapy were consulted to verify and differentiate metastases against hypervasculare non metastases e.g. haemangioma.
2.5. Statistical analysis
CT-Attenuation, number of detected lesions and image quality parameters were tested for differences by using the Wilcoxon test for paired samples. All calculations were performed by using statistical software (SPSS, version 21, SPSS Inc., Chicago, USA). Means and standard deviations were calculated.
3. Results
3.1. Measurements of metastases
The mean attenuation values of 80 kVp- and virtual 120 kVp images for hypervascular hepatic metastases were 121.4 ± 13.3 HU and 96.7 ± 18.5 HU, respectively (p < 0.001). All other anatomical regions (liver, fat, aorta, muscle, air) differed significantly when the two datasets were compared (Table 1).
Table 1.
Anatomical Region | 80 kVp | virtual 120 kVp | p-value |
---|---|---|---|
Liver metastases | 121.4 ± 13.3 | 96.7 ± 18.5 | <0.05 |
Liver | 70.1 ± 11.6 | 61.4 ± 7.6 | <0.05 |
Muscle | 47.4 ± 16.2 | 42.4 ± 11.0 | <0.05 |
Fat | −111.3 ± 18.5 | −92.2 ± 16.2 | <0.05 |
Air | −957.4 ± 145.2 | −982.5 ± 99.3 | <0.05 |
Aorta | 367.9 ± 73.6 | 233.4 ± 54.6 | <0.05 |
The mean (Table 2). number of detected lesions of all 20 patiens on 80-kVp, virtual 120-kVp and DSA images was 5.6 ± 2.1, 4.1 ± 1.8, and 4.3 ± 1.6, respectively. The differences between 80 kVp-images and virtual 120 kVp-/DSA- images were significant (p < 0.05/p < 0.05).
Table 2.
80 kVp | virtual 120 kVp | DSA | |
---|---|---|---|
HLM (mean per liver) | 5.6 ± 2.1 | 4.1 ± 1.8 | 4.3 ± 1.6 |
HLM (overall) (min.-max. size/mean size in mm) |
123 (3–53/2.5 ± 2) |
88 (4–53/2.6 ± 1.9) |
84 (4–54/2.5 ± ) |
All lesions sizes were significant better detected with 80 kVp images than with virtual 120 kVp and DSA-Images (80 kVp vs. 120 kVp: <10 mm: 34 vs. 19, p < 0.05; 10–20 mm:, 33 vs. 25, p < 0.05; >20 mm: 56 vs. 42, p < 0.05/80 kVp vs. DSA: <10 mm: 34 vs. 18 p < 0.05; 10–20 mm: 33 vs. 24, p < 0.05; >20 mm: 56 vs. 41, p < 0.05). Number of detected small lesions <10 mm with 120 kVp compared to DSA-Images were significant higher (19 vs.13; p < 0.05), lesions 10–20 mm and >20 mm were measured statistically equal (Table 3).
Table 3.
Tumor Size (mm) | 80 kVp images | virtual 120 kVp images | DSA |
---|---|---|---|
<10 | 34 | 19 | 18 |
10–20 | 33 | 25 | 24 |
>20 | 56 | 42 | 41 |
3.2. Image quality
The background noise, subjective noise rating, signal-to-noise ratio, contrast-to-noise ratio and image quality of 80-kVp and virtual 120-kVp images are summarised in Table 4. Image noise, SNR and CNR were higher on 80 kVp images compared to 120 kVp images, image quality was higher on virtual 120 kVp images.
Table 4.
Variable | 80 kVp | virtual 120 kVp | p-value |
---|---|---|---|
BN | 13.6 ± 4.3 | 9.7 ± 5.4 | <0.05 |
Subjective noise rating | 3.3 ± 1.5 | 1.5 ± 0.4 | <0.05 |
SNR | 11.4 ± 2.6 | 7.7 ± 3.7 | <0.05 |
CNR | 7.5 ± 3.2 | 4.8 ± 2.9 | <0.05 |
Image quality | 2.8 ± 0.7 | 1.6 ± 0.9 | <0.05 |
4. Discussion
The present study demonstrates that dual energy computed tomography (DECT) has great potential for detection of hepatic metastases in patients with uveal melanoma. The ability of DECT to extract basic information regarding iodine concentration and material composition has led to practical abdominal DECT applications that include the triage of patients with nephrolithasis based on stone composition, quantification of liver iron stores, evaluation of tissue perfusion, potential reduction in radiation dose by elimination of a separate unenhanced CT acquisition, and improved detection of pathologic hyperenhancement [5], [12], [13], [14], [15], [16], [17].
We could show that significant more and smaller metastases can be detected with the help of low kVp images compared to the virtual 120 kVp and DSA images. In a previous study we investigated the diagnostic value of DECT for detecting HCC-lesions [5]. Low kVp images in this study were also more sensitive regarding small HCC tumors. When compared with 120 kVp data, pure 80 kVp data acquired from a dual source dual energy MDCT scanner demonstrates greater attenuation differences and improved contrast to noise between metastatic disease and normal liver in an other study. In contrast to previous studies we have consulted the angiography as a sensitive investigation for comparison.
The poorer image quality of low kVp images caused by the increased image noise had no negative influence on the recognizability of metastases because sensitivity was increased by the higher SNR and CNR. In this context Hur et al. showed that the iterative reconstruction in image space algorithm yielded a significant improvement in image quality for MDCT images at low tube voltage (80 kVp) and intermediate tube current (340 mAs) setting and further may be effective for reducing patient radiation dose and for improving the conspicuity of hyperenhancing HCCs in thin adults when compared with standard 120 kVp CT images using the filtered back-projection reconstruction algorithm [18].
We could also verify a good diagnostic agreement between the angiography (DSA) of the liver and the regular used 120 kVp images of computed tomography as it has been done before by other authors [11]. But our results suggests that there is also an diagnostic advantage of the low kVp-images of DECT compared to the regular 120 kVp-images.
Cross-sectional imaging methods are the standard methods for staging of advanced melanoma. The former time-consuming and expensive multimodality approach is increasingly replaced by novel whole-body (WB) staging methods, such as 18F-fluorodeoxyglucose positron emission tomography/computed tomography (18F-FDG-PET-CT) and whole-body magnetic resonance imaging (WBMRI) because they offer a complete head-to-toe coverage of the patient in a single examination with an accurate and sensitive detection of tumor spread. The preliminary results indicate a high overall diagnostic accuracy of both methods; however, these methods differ in organ-based detection rates: PET-CT was more accurate in N-staging and detection of lung and soft tissue metastases whereas WB-MRI was superior in detecting liver, bone and brain metastases [19].
Compared to PET/CT and PET/MRI DECT is less expensive and time consuming focused on detection of liver metastases, but for whole body staging DECT is inferior to the PET-methods. Another advantage of DECT compared to regular computed tomography is the decreased radiation dose because of the missing native scans.
Wagner et al. aimed to assess the sensitivity of diffusion-weighted (DW) magnetic resonance (MR) imaging for the detection of pathologically confirmed uveal melanoma liver metastases (UMLM). They showed that the addition of DW imaging to morphologic-dynamic images does not significantly increase MR sensitivities for UMLM detection [20].
As a limitation of this study histopathology was not available for the detected hypervascular liver lesions rated as metastases. However diagnosis was carried out by follow-up studies.
In conclusion, low-kVp images of DECT datasets are more sensitive in detecting liver metastases of patients with uveal melanoma than virtual 120 kVp- and DSA images.
References
- 1.Virgili G., Gatta G., Ciccolallo L. Incidence of uveal melanoma in Europe. Ophthalmology. 2007;114(12):2309–2315. doi: 10.1016/j.ophtha.2007.01.032. [DOI] [PubMed] [Google Scholar]
- 2.Diener-West M., Reynolds S.M., Agugliaro D.J. Development of metastatic disease after enrollment in the COMS trials for treatment of choroidal melanoma: collaborative: ocular melanoma study group report No 26. Arch. Ophthalmol. 2005;123(12):1639–1643. doi: 10.1001/archopht.123.12.1639. [DOI] [PubMed] [Google Scholar]
- 3.Leyvraz S., Piperno-Neumann S., Suciu S. Hepatic intra-arterial versus intravenous fotemustine in patients with liver metastases from uveal melanoma (EORTC 18021): a multicentric randomized trial. Ann. Oncol. 2014;25(3):742–746. doi: 10.1093/annonc/mdt585. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Flohr T.G., McCollough C.H., Bruder H. First performance evaluation of a dual-source CT (DSCT) system. Eur. Radiol. 2006;16(2):256–268. doi: 10.1007/s00330-005-2919-2. [DOI] [PubMed] [Google Scholar]
- 5.Altenbernd J., Heusner T.A., Ringelstein A., Ladd S.C., Forsting M., Antoch G. Dual-energy-CT of hypervascular liver lesions in patients with HCC: investigation of image quality and sensitivity. Eur. Radiol. 2011;21(4):738–743. doi: 10.1007/s00330-010-1964-7. [DOI] [PubMed] [Google Scholar]
- 6.Duran R., Chapiro J., Frangakis C. Uveal melanoma metastatic to the liver: the role of quantitative volumetric contrast-enhanced MR imaging in the assessment of early tumor response after transarterial chemoembolization. Transl. Oncol. 2014;7(4):447–455. doi: 10.1016/j.tranon.2014.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Matsuo T., Ogino Y., Ichimura K., Tanaka T., Kaji M. Clinicopathological correlation for the role of fluorodeoxyglucose positron emission tomography computed tomography in detection of choroidal malignant melanoma. Int. J. Clin. Oncol. 2014;19(2):230–239. doi: 10.1007/s10147-013-0538-5. [DOI] [PubMed] [Google Scholar]
- 8.Orcurto V., Denys A., Voelter V. (18)F-fluorodeoxyglucose positron emission tomography/computed tomography and magnetic resonance imaging in patients with liver metastases from uveal melanoma: results from a pilot study. Melanoma Res. 2012;22(1):63–69. doi: 10.1097/CMR.0b013e32834d3dcb. [DOI] [PubMed] [Google Scholar]
- 9.Servois V., Mariani P., Malhaire C. Preoperative staging of liver metastases from uveal melanoma by magnetic resonance imaging (MRI) and fluorodeoxyglucose-positron emission tomography (FDG-PET) Eur. J. Surg. Oncol. 2010;36(2):189–194. doi: 10.1016/j.ejso.2009.08.010. [DOI] [PubMed] [Google Scholar]
- 10.Young K., Fidelman N., Yao F.Y. Implications of discordant findings between hepatic angiography and cross-sectional imaging in transplant candidates with hepatocellular carcinoma. Liver Transpl. 2015 doi: 10.1002/lt.24090. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Zheng X.H., Guan Y.S., Zhou X.P. Detection of hypervascular hepatocellular carcinoma: comparison of multi-detector CT with digital subtraction angiography and Lipiodol CT. World J. Gastroenterol. 2005;11(2):200–203. doi: 10.3748/wjg.v11.i2.200. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Yeh B.M., Shepherd J.A., Wang Z.J., Teh H.S., Hartman R.P., Prevrhal S. Dual-energy and low-kVp CT in the abdomen. AJR Am. J. Roentgenol. 2009;193(1):47–54. doi: 10.2214/AJR.09.2592. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Graser A., Johnson T.R., Chandarana H., Macari M. Dual energy CT: preliminary observations and potential clinical applications in the abdomen. Eur. Radiol. 2009;19(1):13–23. doi: 10.1007/s00330-008-1122-7. [DOI] [PubMed] [Google Scholar]
- 14.Fletcher J.G., Takahashi N., Hartman R. Dual-energy and dual-source CT: is there a role in the abdomen and pelvis? Radiol. Clin. North Am. 2009;47(1):41–57. doi: 10.1016/j.rcl.2008.10.003. [DOI] [PubMed] [Google Scholar]
- 15.Chu A.J., Lee J.M., Lee Y.J., Moon S.K., Han J.K., Choi B.I. Dual-source, dual-energy multidetector CT for the evaluation of pancreatic tumours. Br. J. Radiol. 2012;85(1018):e891–e898. doi: 10.1259/bjr/26129418. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Dai X., Schlemmer H.P., Schmidt B. Quantitative therapy response assessment by volumetric iodine-uptake measurement: initial experience in patients with advanced hepatocellular carcinoma treated with sorafenib. Eur. J. Radiol. 2013;82(2):327–334. doi: 10.1016/j.ejrad.2012.11.013. [DOI] [PubMed] [Google Scholar]
- 17.De Cecco C.N., Darnell A., Rengo M. Dual-energy CT: oncologic applications. AJR Am. J. Roentgenol. 2012;199(5 Suppl):S98–S105. doi: 10.2214/AJR.12.9207. [DOI] [PubMed] [Google Scholar]
- 18.Hur S., Lee J.M., Kim S.J., Park J.H., Han J.K., Choi B.I. 80-kVp CT using Iterative Reconstruction in Image Space algorithm for the detection of hypervascular hepatocellular carcinoma: phantom and initial clinical experience. Korean J. Radiol. 2012;13(2):152–164. doi: 10.3348/kjr.2012.13.2.152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Pfannenberg C., Schwenzer N. Whole-body staging of malignant melanoma: advantages, limitations and current importance of PET-CT, whole-body MRI and PET-MRI. Radiologe. 2015;55(2):120–126. doi: 10.1007/s00117-014-2762-z. [DOI] [PubMed] [Google Scholar]
- 20.Wagner M., Mariani P., Bidard F.C. Diffusion-weighted MRI for uveal melanoma liver metastasis detection. Eur. Radiol. 2015 doi: 10.1007/s00330-015-3662-y. [DOI] [PubMed] [Google Scholar]