Infection prevention and ultrasound probe decontamination practices in Europe: a survey of the European Society of Radiology

Christiane Marita Nyhsen 1,, Hilary Humphreys 2, Carlos Nicolau 3, Gerhard Mostbeck 4, Michel Claudon 5
PMCID: PMC5110482  PMID: 27778309

Abstract

Objectives

Although ultrasound (US) is considered one of the safest imaging modalities, concerns have been raised regarding potential infection transmission risks through US procedures. A survey was undertaken by the European Society of Radiology (ESR) to establish infection prevention and control measures in US and to highlight the importance of good medical practice.

Methods

An online survey was sent to all 22,000 full ESR members.

Results

The response rate of completed surveys was 4.3 % (946 practitioners, 97 % of which were radiologists, mostly working in larger hospital settings). Among respondents, 29 %, 11 % and 6 % did not disinfect the US probe after every patient when performing standard surface US, endo-cavity US and interventional procedures, respectively. Eleven percent did not always use probe covers for endo-cavity US; for interventional procedures, the proportion was 23 %. A minority used sterile gel sachets in direct patient contact for endo-cavity scans (30 %), and 77.5 % used sterile gel for interventional procedures.

Conclusions

The survey results highlight a wide range of practices throughout Europe and the need to raise awareness amongst practitioners regarding the importance of infection prevention and control measures. The development of European recommendations encompassing all US examinations, together with education is a priority.

Main Messages

Transmission of infection through ultrasound procedures is possible.

There is a wide range of ultrasound probe decontamination practices in Europe.

Not all practitioners use probe covers for endo-cavity or interventional ultrasound.

Not all practitioners use sterile gel for internal and invasive procedures.

Currently there are no European recommendations encompassing all US examinations.

Electronic supplementary material

The online version of this article (doi:10.1007/s13244-016-0528-z) contains supplementary material, which is available to authorized users.

Keywords: Ultrasound, Gels, Infection prevention and control, Decontamination, Disinfection

Introduction

Ultrasound (US) is generally considered one of the safest diagnostic modalities available. Neither professionals nor patients undergoing US imaging anticipate that US will cause significant harm. However, in recent years, several outbreaks of infection related to endoscopic procedures or trans-oesophageal US have been reported [15], resulting in the introduction of more stringent infection prevention measures [6, 7].

Concerns have been raised that transmission of infection may also pose a risk to patients in conventional US, in particular endo-cavity (i.e., trans-vaginal or trans-rectal) US, performing US near wounds/stomas, or interventional procedures involving US. Furthermore, there is a potential risk of transmitting infection through US gel, and the risk of microscopic or macroscopic probe cover perforations necessitates consideration.

It is very difficult to prove contamination and the subsequent transmission of pathogens through US examinations, but it should not be assumed that the risk does not exist. Surveys in the UK have already revealed that there is a wide range of practices, with the authors calling for national guidance [8, 9]. To our knowledge, only European guidelines relating to interventional US have been published [10]; no European guidance is yet available encompassing all US settings.

We report the results of a survey undertaken by the European Society of Radiology (ESR) to identify current European US infection prevention and control practices in place and to raise awareness of the importance of appropriate decontamination protocols.

Materials and methods

An online survey (using the SurveyMonkey software) was sent to all full ESR members in Europe (22,000) in September 2015.

A total of 22 questions were posed, initially regarding the country and city of work, the work setting (private centre versus hospital), size of the establishment and number of yearly US procedures undertaken. Thereafter, known incidents involving transmission of infection were queried, with free text available for further clarification.

The final part of the survey gathered detailed data on the US gel used, both in direct contact with the patient and inside probe covers. Decontamination procedures were surveyed over a wide range of US examinations: standard surface US, when scanning patients with known transmissible infections, endo-cavity (trans-vaginal and trans-rectal) US, and US used in an interventional setting (US-guided biopsy, drainage or US used in theatre). Free text sections were available throughout the survey (see Appendix 1).

Results

Of the 22,000 questionnaires sent, a total of 1088 replies (4.94 %) were received, of which 946 (4.3 %) were complete and subsequently analysed. Responses were gathered from most European countries, with the largest numbers originating from Italy, the UK, Spain and France.

The majority of survey respondents worked exclusively in a hospital setting (81 %); the remainder worked in private centres or offices (15 %), and 4 % indicated mixed/other work commitments. Most practitioners (67 %) worked in relatively large institutions (100–1000 beds), with 16 % working in very large hospitals of over 1000 beds. Twelve percent had no patient admission facilities. The majority (88 %) performed over 3000 US examinations per year, with 42 % of the institutions carrying out over 10,000 annual scans.

Almost all surveys were completed by radiologists (97 %). Eight responses in total were received from cardiologists, gynaecologists/obstetricians and urologists. Twenty replies stated mixed practice or work in other specialties such as breast services, paediatrics or emergency medicine.

Known cases of infection transmission through ultrasound

Of the 946 respondents, 26 (2.7 %) stated that they were aware of cases of infection transmission through US procedures. Details provided in the free text section mentioned insufficient infection prevention measures several times. Presumed transmitted organisms included methicillin-resistant Staphylococcus aureus (MRSA), Staphylococcus, cytomegalovirus and human papillomavirus, as well as cutaneous fungal infection and scabies. Infections were recorded after breast biopsy/drainage, following joint aspiration/injection, following trans-vaginal US and after trans-rectal biopsy (the latter of which is well recognised given the presence of bowel flora). A few cases described skin infections and complications when assessing wounds as well as gastroenteritis/gastroenterocolitis. Others mentioned MRSA without offering further details.

Reviewing this subgroup in comparison to overall results did not reveal relevant differences in infection prevention measures recorded.

Surface ultrasound on unbroken skin (i.e., presumed relatively low-risk procedures)

Most practitioners used either refill US gel bottles (520, 55 %) or single-use bottles (405, 43 %). Only 2 % used sterile sachets.

Almost two thirds of respondents decontaminated the US probe by wiping off the US gel first with subsequent disinfection of probes with foam or wipe after each patient (618, 65 %), whilst 276 (29 %) disinfected the probes only at the end of the list. Only 4 % used a dedicated washer to thoroughly decontaminate probes in this setting (see Fig. 1). A few responses detailed other protocols or gave insufficient information.

Fig. 1.

Fig. 1

US probe decontamination

In the free text section, some practitioners stated that they wiped the probe with a dry towel but that no subsequent disinfection measures were undertaken. A multitude of different wipes and foam brands were detailed, and two respondents mentioned the use of ultraviolet light sterilisation, but this question was not answered by 803 respondents.

Cases with known infections (surface ultrasound on unbroken skin)

When the patient was known to have a transmissible infection (infection control cases), 724 (77 %) wiped off the US gel and then used disinfecting wipes/foam after the procedure, and 174 (18 %) cleaned the probe in a dedicated washer. Forty-eight (5 %) detailed other procedures or gave insufficient information.

In the free text section, 60 respondents stated that in these cases, they would use probe covers as an additional preventative step, but not all disinfected the probe after removal of the protective sheath.

Endo-cavity ultrasound procedures (trans-vaginal or trans-rectal ultrasound)

Most practitioners stated that they used a probe cover for endo-cavity US scans at all times (697, 89 %), whilst 85 (11 %) did not (see Fig. 2). Unfortunately, 164 respondents did not answer this question, possibly because they did not perform these procedures.

Fig. 2.

Fig. 2

Use of US probe covers at all times

A minority (110 respondents, 14 %) used sterile gel inside the probe cover; most used either refill bottles (355, 46 %) or single-use bottles (304, 40 %). A total of 177 respondents skipped this question. Nearly a third, 233 (30 %), stated that they used sterile gel in direct patient contact for endo-cavity procedures (233, 30 %). Refill bottles were used by 269 (35 %), 271 (35 %) used gel from single-use bottles (see Fig. 3), and 173 skipped this question.

Fig. 3.

Fig. 3

US gel used

With regard to endo-cavity US probe decontamination procedures, the majority of respondents (69 %) chose “wiping off ultrasound gel and disinfection after each patient with foam/wipe”. A more thorough “3-step cleaning process” with high-level disinfectant wipes or “probe cleaning in dedicated washer” was chosen by 69 and 70 (9 %) practitioners, respectively. Just over one in ten (86, 11 %) cleaned the probe only after the list was finished and not after each patient (see Fig. 1). A few practitioners detailed other procedures, and 166 skipped this question.

Interventional procedures

Over two thirds of practitioners taking part in this survey stated that they performed interventional procedures (70 %). Of these 666 interventional practitioners, the majority used probe covers at all times (513, 77 %), whilst almost one quarter (153, 23 %) did not (see Fig. 2).

The US gel inside the probe cover used for interventional procedures was typically from either a refill bottle (269, 40.4 %) or a single-use bottle (229, 24.4 %), whilst 168 (25.2 %) used sterile sachets. US gel used in direct contact during interventional procedures was mostly from sterile sachets (516, 77.5 %). The remainder stated they used gel from refill bottles (70, 10.5 %) or single-use bottles (80, 12 %) as show in Fig. 3.

US probe decontamination procedures were similar to endo-cavity probe disinfection, with most practitioners wiping off the gel with subsequent disinfection after each patient (507, 76 %), whilst 47 (7 %) used a thorough three-step high-level disinfectant wipe cleaning process, and 66 (10 %) used a dedicated probe washer (see Fig. 1). A minority did not disinfect the probe after each patient, but only after the list was finished (40, 6 %). Again, some practitioners detailed other procedures which did not fit into the categories above or provided insufficient information. Several responses stated that probe covers were used but that the probes were not disinfected afterwards.

Discussion

Publications dating back to 1988 have raised concerns regarding cross-infection risks of US procedures [11]. A case report from the year 2000 describes the transmission of hepatitis C during US-guided procedures in the setting of assisted conception, the exact route of transmission being unclear [12]. Several other studies have highlighted that post-procedure US probe contamination is significant, and that transmission of infection is possible [8, 1318]. Reviewing the literature concerning US-guided procedures, an overall incidence of infectious complications of 0.1 % has been reported [19]. This rate may increase to 0.19 % after trans-bronchial endoscopic US-guided needle aspiration [20] and to 4.29 % after trans-rectal prostate biopsy [21].

Infections may be clinically silent and discovered only at a later date, potentially not connected to previous US examinations. Obtaining proof that an infection is due to previous US investigations can be difficult unless incidents emerge with a traceable originating source. Also, some incidents of infection transmission will not have been published.

The survey reported here confirms a very wide range of practices throughout Europe. Twenty-six practitioners stated that they knew of cases of infection transmission through US procedures, proving that this is a topic worth considering. The actual figure of US probe contamination and transmission of infection is unknown, and is potentially higher, as an estimated risk study from France shows [22], even though another French study did not find an increased risk of acquired HIV or hepatitis C infections following trans-vaginal US procedures [23]. Whatever the anticipated risk, known avoidable healthcare-acquired infections are likely to be reviewed with increasing scrutiny. In addition, all US practitioners have an ethical obligation to keep their patients as safe as possible, and hence should seek to minimise risk wherever possible.

Potential risks can be minimised by reducing transducer contamination, better US probe decontamination after every patient and by the use of appropriate US gel:

  1. Reduction of US probe contamination (use of probe covers)

    Basic hygiene protocols such as hand washing, the use of gloves and keeping surfaces clean are essential first infection prevention and control measures. In addition, the use of transducer covers can lower contamination levels, particularly in interventional or endo-cavity US and whenever skin is not intact (i.e., when scanning close to wounds or stomas).

    The survey results show that a significant minority did not always use probe covers for endo-cavity scans (11 %) or interventional procedures (23 %), possibly due to a lack of awareness.

    The use of sterile probe covers is strongly recommended in the recently published EFSUMB Guidelines on Interventional Ultrasound, which also emphasise that subsequent transducer decontamination after every patient is needed [10]. The latter is important, as varying rates of probe cover perforations have been reported, ranging from 2 % [24] to 9 % of cases [25], and even up to 81 % in one older study [26]. Less frequent perforations were reported when condoms were used for trans-vaginal US rather than US probe covers [27, 28], but published studies are somewhat dated. Visual inspection may significantly underestimate contamination levels and is therefore unreliable [29]. Even when performing low-level decontamination, e.g., single-step wiping of probes with a disinfectant, persistent contamination was found on trans-vaginal US probes after probe covers had been used [8, 16].

  2. Choice of US gel

    Gels used for medical purposes do not have bacteriostatic properties per se, and cannot be viewed as sterile unless specifically stated by the manufacturer. This was proven by studies in the 1990s [30, 31]. Contaminated US gel has been identified in routine cultures, and several reports of outbreaks of infection related to US gel have been published [18, 3235].

    In this survey, a minority of practitioners used sterile gel sachets in direct patient contact for endo-cavity scans (30 %), whilst the majority used sterile gel for interventional procedures (77.5 %). Again, it appears that increasing awareness is important. It is also noted that a minority used sterile gel inside transducer covers. Considering the significant reported probe cover perforation rates (see above), it is debatable whether the use of sterile gel would lower infection rates.

    Recommendations regarding the safe use of US gels, including gel warmers, were published by Oleszkowicz et al. in 2012 [36]. The newly published EFSUMB Guidelines on Interventional Ultrasound recommend the use of sterile gel in interventional US, and state that new sachets should be opened for every patient [10]. We would agree with this advice.

  3. US probe decontamination

    Complete decontamination of US transducers is challenging [8, 37]. A study testing the susceptibility of human papillomavirus, for example, showed that many disinfectants (including alcohol) may have limited to no effect [37], which is concerning and necessitates a general review of decontamination protocols used.

    Wiping probes with a dry towel resulted in persistent contamination [18]. Many potent and proven in vitro-effective chemical disinfectants, such as glutaraldehyde, can pose health hazards to staff in direct skin contact or through inhalation, as well as causing unhealthy work environments for staff and patients. The nature of the US probe surface must be considered as well, and its susceptibility to damage by chemicals—for example, alcohol [38]. Individual manufacturers should provide detailed information on which disinfectants are compatible, and should be prepared to provide further product testing upon request.

    Questionnaire responses reveal that not all practitioners disinfected the US transducers after every patient: 29 %, 11 % and 6 % of respondents did not disinfect the US probe after every patient when performing standard surface US, endo-cavity US and interventional procedures respectively. It seems that more needs to be done to emphasise the importance of adequate US transducer decontamination after every patient, and not just at the end of the US list.

    We acknowledge that implementing any additional infection prevention measures will not be cost-neutral. Capital investment for additional transducers and decontamination equipment may be needed, and there will be higher ongoing costs for probe covers and sterile gel. All staff working in US departments must be adequately trained in using the available resources as effectively as possible.

    Recommendations available to date include the European guidelines on interventional US mentioned above, which contain sections on infection prevention [10]. The British Medical Ultrasound Society (BMUS) published a survey and guidance in 2003 [9], but whilst very valuable at the time, this advice was never adopted UK-wide. NHS Health Scotland has just published a document, Guidance for Decontamination of Semi-Critical Ultrasound Probes; Semi-invasive and Non-invasive Ultrasound Probes [39], regarding recommended protocols. Several more recent publications from the USA and Canada relating to the safe use of medical gel and US probe decontamination are available, as well as guidance from Australasia and France [4045]. Many local infection prevention and control guidelines also exist. However, to our knowledge, there are no specific published European or international guidelines available to advise general US practitioners on infection prevention and control matters. Up-to-date European expert recommendations, after a thorough literature review, would be helpful for protecting patients and ensuring that US examinations are as safe as possible.

Limitations

The limitations of this survey include the relatively low response rate, with only 946 completed questionnaires (4.3 %). This may partly reflect the fact that not all radiologists performing US examinations (in 2015, just under half of the full ESR Radiologist Members chose US as one of their special interests), or in some departments, only the US lead radiologist may have completed the survey. There was unequal representation among European countries, with a larger number of responses from Italy, the UK, Spain and France. The vast majority of responses came from hospital settings, and a smaller number were received from private centres, which may not fully represent the second group. This survey focussed on infection prevention issues relating to US procedures. To keep the questionnaire as short as possible, the more wide-ranging basic infection prevention and control precautions such as hand washing, the use of gloves and protective drapes, and the cleaning of US machines (apart from transducers) and other surfaces were not explored. No question was included about the use of US gel warmers. Finally, we did not include options to allow for differences in the complexity of interventional procedures.

Conclusion

These survey results underscore the importance of raising awareness amongst clinicians of the risks of infection associated with US, in particular when transducers are in contact with mucous membranes and potentially infected bodily fluids. Expert European recommendations are needed to educate clinicians, guide best practices and ensure that safe patient care is provided.

Electronic supplementary material

Below is the link to the electronic supplementary material.

ESM 1 (99.1KB, pdf)

(PDF 99 kb)

Acknowledgments

The authors are members of the ESR Working Group on Ultrasound. This survey was carried out in collaboration with the ESR.

The authors would like to thank Mrs Catherine Fisher and her colleagues from City Hospitals Sunderland Library Services for their continuous excellent support and assistance.

Contributor Information

Christiane Marita Nyhsen, Phone: 00441915656256, Email: christiane.nyhsen@chsft.nhs.uk.

Hilary Humphreys, Phone: 01-8093710, Email: hhumphreys@rcsi.ie.

Carlos Nicolau, Phone: +34//932275735, Email: cnicolau@clinic.ub.es.

Gerhard Mostbeck, Phone: +43/1/49150 3101, Email: gerhard.mostbeck@wienkav.at.

Michel Claudon, Phone: +33/383/15-41-83, Email: m.claudon@chu-nancy.fr.

References

  • 1.Wendorf KA, Kay M, Baliga C, Weissman SJ, Gluck M, Verma P, et al. Endoscopic retrograde cholangiopancreatography-associated AmpC Escherichia coli outbreak. Infect Control Hosp Epidemiol. 2015;36(6):634–642. doi: 10.1017/ice.2015.66. [DOI] [PubMed] [Google Scholar]
  • 2.Kola A, Piening B, Pape UF, Veltzke-Schlieker W, Kaase M, Geffers C, et al. An outbreak of carbapenem-resistant OXA-48 - producing Klebsiella pneumonia associated to duodenoscopy. Antimicrob Resist Infect Control. 2015;4:8. doi: 10.1186/s13756-015-0049-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Centers for Disease Control and Prevention (CDC) (2014) Notes from the field: New Delhi metallo-β-lactamase-producing Escherichia coli associated with endoscopic retrograde cholangiopancreatography - Illinois, 2013. MMWR Morb Mortal Wkly Rep 62(51–52):1051. http://www.cdc.gov/mmwr/preview/mmwrhtml/mm6251a4.htm Accessed 3 Feb 2016 [PMC free article] [PubMed]
  • 4.Koo VS, O’Neill P, Elves A. Multidrug-resistant NDM-1 Klebsiella outbreak and infection control in endoscopic urology. BJU Int. 2012;110(11 Pt C):E922–E926. doi: 10.1111/j.1464-410X.2012.11556.x. [DOI] [PubMed] [Google Scholar]
  • 5.Seki M, Machida H, Yamagishi Y, Yoshida H, Tomono K. Nosocomial outbreak of multidrug-resistant Pseudomonas aeruginosa caused by damaged transesophageal echocardiogram probe used in cardiovascular surgical operations. J Infect Chemother. 2013;19(4):677–681. doi: 10.1007/s10156-012-0542-0. [DOI] [PubMed] [Google Scholar]
  • 6.CDC (Centers for Disease Control and Prevention) Statement (2015) Los Angeles County/UCLA investigation of CRE transmission and duodenoscopes. http://www.cdc.gov/hai/outbreaks/cdcstatement-LA-CRE.html Accessed 27 Jan 2016
  • 7.Department of Health (UK, 2013) Choice framework for local policy and procedures 01-06 – decontamination of flexible endoscopes: Operational management, https://www.gov.uk/government/uploads/system/uploads/attachment_data/file/148559/CFPP_01-06_Operational_mgmt_Final.pdf. Accessed 27 Jan 2016
  • 8.Gray RA, Williams PL, Dubbins PA, Jenks PJ. Decontamination of transvaginal ultrasound probes: review of national practice and need for national guidelines. Clin Radiol. 2012;67(11):1069–1077. doi: 10.1016/j.crad.2012.02.015. [DOI] [PubMed] [Google Scholar]
  • 9.Backhouse S (2003) Establishing a protocol for the cleaning and sterilisation/disinfection of ultrasound transducers. BMUS Bull 11(1)
  • 10.Lorentzen T, Nolsøe CP, et al. EFSUMB Guidelines on Interventional Ultrasound (INVUS), Part I. Ultraschall Med. 2015;36:E1–E14. doi: 10.1055/s-0035-1553792. [DOI] [PubMed] [Google Scholar]
  • 11.Spencer P, Spencer RC. Ultrasound scanning of postoperative wounds - the risks of cross-infection. Clin Radiol. 1988;39:245e6. doi: 10.1016/s0009-9260(88)80517-8. [DOI] [PubMed] [Google Scholar]
  • 12.Lesourd F, Izopet J, Mervan C, Payen JL, Sandres K, et al. Transmissions of hepatitis C virus during the ancillary procedures for assisted conception. Hum Reprod. 2000;15:1083–1085. doi: 10.1093/humrep/15.5.1083. [DOI] [PubMed] [Google Scholar]
  • 13.M’Zali F, Bounizra C, Leroy S, Mekki Y, Quentin-Noury C, Kann M. Persistence of microbial contamination on transvaginal ultrasound probes despite low-level disinfection procedure. PLoS One. 2014;9(4):e93368. doi: 10.1371/journal.pone.0093368. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Koibuchi H, Kotani K, Taniguchi N. Ultrasound probes as a possible vector of bacterial transmission. Med Ultrason. 2013;15(1):41–44. doi: 10.11152/mu.2013.2066.151.hk1upp2. [DOI] [PubMed] [Google Scholar]
  • 15.Leroy S. Infectious risk of endovaginal and transrectal ultrasonography: systematic review and meta-analysis. J Hosp Infect. 2013;83(2):99–106. doi: 10.1016/j.jhin.2012.07.014. [DOI] [PubMed] [Google Scholar]
  • 16.Ma ST, Yeung AC, Chan PK, Graham CA. Transvaginal ultrasound probe contamination by the human papillomavirus in the emergency department. Emerg Med J. 2012;30(6):472–475. doi: 10.1136/emermed-2012-201407. [DOI] [PubMed] [Google Scholar]
  • 17.Casalegno JS, Le Bail Carval K, Eibach D, Valdeyron ML, Lamblin G, Jacquemoud H, et al. High risk HPV contamination of endocavity vaginal ultrasound probes: an underestimated route of nosocomial infection? PLoS One. 2012;7(10):e48137. doi: 10.1371/journal.pone.0048137. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Mullaney PJ, Munthali P, et al. How clean is your probe? microbiological assessment of ultrasound transducers in routine clinical use, and cost-effective ways to reduce contamination. Clin Radiol. 2007;62(7):694–698. doi: 10.1016/j.crad.2007.01.002. [DOI] [PubMed] [Google Scholar]
  • 19.Cervini P, Hesley GK, Thompson RL, Sampathkumar P, Knudsen JM. Incidence of infectious complications after an ultrasound-guided intervention. AJR Am J Roentgenol. 2010;195(4):846–850. doi: 10.2214/AJR.09.3168. [DOI] [PubMed] [Google Scholar]
  • 20.Asano F, Aoe M, Ohsaki Y, Okada Y, Sasada S, Sato S, et al. Complications associated with endobronchial ultrasound-guided transbronchial needle aspiration: a nationwide survey by the Japan Society for Respiratory Endoscopy. Respir Res. 2013;14:50. doi: 10.1186/1465-9921-14-50. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Lodeta B, Trkulja V. Septic complications and hospital admissions after transrectal ultrasound-guided prostate biopsy: incidence rates and outcomes in 913 consecutive biopsies. Int Urol Nephrol. 2014;46(12):2335–2336. doi: 10.1007/s11255-014-0815-x. [DOI] [PubMed] [Google Scholar]
  • 22.Leroy S, M’Zali F, Kann M, Weber D, Smith D. Impact of vaginal-rectal ultrasound examinations with covered and low-level disinfected transducers on infectious transmissions in France. Infect Control Hosp Epidemiol. 2014;35(12):1497–1504. doi: 10.1086/678604. [DOI] [PubMed] [Google Scholar]
  • 23.Bénet T, Ritter J, Vanhems P. Risk of human immunodeficiency virus and hepatitis C virus infection related to endocavitary ultrasound probe exposure in France. Infect Control Hosp Epidemiol. 2014;35(11):1429–1431. doi: 10.1086/678421. [DOI] [PubMed] [Google Scholar]
  • 24.Milki AA, Fisch JD. Vaginal ultrasound probe cover leakage: implications for patient care. Fertil Steril. 1998;69(3):409–411. doi: 10.1016/S0015-0282(97)00571-2. [DOI] [PubMed] [Google Scholar]
  • 25.Masood J, Voulgaris S, Awogu O, Younis C, Ball AJ, Carr TW. Condom perforation during transrectal ultrasound guided (TRUS) prostate biopsies: a potential infection risk. Int Urol Nephrol. 2007;39(4):1121–1124. doi: 10.1007/s11255-007-9213-y. [DOI] [PubMed] [Google Scholar]
  • 26.Hignett M, Claman P. High rates of perforation are found in endovaginal ultrasound probe covers before and after oocyte retrieval for in vitro fertilization-embryo transfer. J Assist Reprod Genet. 1995;12(9):606–609. doi: 10.1007/BF02212583. [DOI] [PubMed] [Google Scholar]
  • 27.Amis S, Ruddy M, Kibbler CC, Economides DL, MacLean AB. Assessment of condoms as probe covers for transvaginal sonography. J Clin Ultrasond. 2000;28(6):295–298. doi: 10.1002/1097-0096(200007/08)28:6<295::AID-JCU5>3.0.CO;2-G. [DOI] [PubMed] [Google Scholar]
  • 28.Rooks VJ, Yancey MK, Elg SA, Brueske L. Comparison of probe sheaths for endovaginal sonography. Obstet Gynecol. 1996;87(1):27–29. doi: 10.1016/0029-7844(95)00336-3. [DOI] [PubMed] [Google Scholar]
  • 29.Storment JM, Monga M, Blanco JD. Ineffectiveness of latex condoms in preventing contamination of the transvaginal ultrasound transducer head. South Med J. 1997;90(2):206–208. doi: 10.1097/00007611-199702000-00007. [DOI] [PubMed] [Google Scholar]
  • 30.Muradali D, Gold WL, Phillips A, Wilson S. Can ultrasound probes and coupling gel be a source of nosocomial infection in patients undergoing sonography? An in vivo and in vitro study. Am J Roentgenol. 1995;164:1521–1524. doi: 10.2214/ajr.164.6.7754907. [DOI] [PubMed] [Google Scholar]
  • 31.Ohara T, Itoh Y, Itoh K. Ultrasound instruments as possible vectors of staphylococcal infection. J Hosp Infect. 1998;40:73–77. doi: 10.1016/S0195-6701(98)90028-7. [DOI] [PubMed] [Google Scholar]
  • 32.Chittick P, Russo V, Sims M, et al. Outbreak of Pseudomonas aeruginosa respiratory tract infections in cardiovascular surgery associated with contaminated ultrasound gel used for transesophageal echocardiography—Michigan, December 2011–January 2012. MMWR Morb Mortal Wkly Rep. 2012;61:262–264. [PubMed] [Google Scholar]
  • 33.Olshtain-Pops K, Block C, Temper V, et al. An outbreak of Achromobacter xylosoxidans associated with ultrasound gel used during transrectal ultrasound guided prostate biopsy. J Urol. 2011;185:144–147. doi: 10.1016/j.juro.2010.08.093. [DOI] [PubMed] [Google Scholar]
  • 34.Jacobson M, Wray R, Kovach D, Henry D, Speert D, Matlow A. Sustained endemicity of Burkholderia cepacia complex in a pediatric institution, associated with contaminated ultrasound gel. Infect Control Hosp Epidemiol. 2006;27:362–366. doi: 10.1086/503343. [DOI] [PubMed] [Google Scholar]
  • 35.Weist K, Wendt C, Petersen L, Versmold H, Ruden H. An outbreak of pyodermas among neonates caused by ultrasound gel contaminated with methicillin-susceptible Staphylococcus aureus. Infect Control Hosp Epidemiol. 2000;21:761–764. doi: 10.1086/501729. [DOI] [PubMed] [Google Scholar]
  • 36.Oleszkowicz SC, Chittick P, et al. Infections associated with use of ultrasound transmission gel. proposed guidelines to minimize risk. Infect Control Hosp Epidemiol. 2012;33(12):1235–1237. doi: 10.1086/668430. [DOI] [PubMed] [Google Scholar]
  • 37.Meyers J, Ryndock E, Conway MJ, Meyers C, Robison R. Susceptibility of high-risk human papillomavirus type 16 to clinical disinfectants. J Antimicrob Chemother. 2014;69(6):1546–1550. doi: 10.1093/jac/dku006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Koibuchi H, Fujii Y, Kotani K, et al. Degradation of ultrasound probes caused by disinfection with alcohol. J Med Ultrason. 2011;38:97–100. doi: 10.1007/s10396-010-0296-1. [DOI] [PubMed] [Google Scholar]
  • 39.Health Facilities Scotland Decontamination Services (2016) NHS scotland guidance for decontamination of semi-critical ultrasound probes; semi-invasive and non-invasive ultrasound probes. http://www.hps.scot.nhs.uk/documents/hai/infection-control/guidelines/NHSScotland-Guidance-for-Decontamination-of-Semi-Critical-Ultrasound-Probes.pdf Accessed 20 May 2016
  • 40.Health and Drug Alerts (2004) Medical gels and the risk of serious infection. JAMC 171(11) http://www.ncbi.nlm.nih.gov/pmc/articles/PMC527335/pdf/20041123s00035p1348.pdf. Accessed 27 Jan 2016 [DOI] [PMC free article] [PubMed]
  • 41.CHICA-Canada Standard and Guidelines Committee Position Statement: Medical Gels (Last revision 2005). http://www.ipac-canada.org/pdf/medgels.pdf. Accessed 27 Jan 2016
  • 42.Rutala WA, Weber DJ, and the Healthcare Infection Control Practices Advisory Committee (HICPAC) (2008) Guideline for disinfection and sterilization in healthcare facilities. Available: http://www.cdc.gov/hicpac/pdf/guidelines/disinfection_nov_2008.pdf. Accessed Jan 29 2016
  • 43.American Institute of Ultrasound in Medicine (aium, 2014) Guidelines for cleaning and preparing external- and internal-use ultrasound probes between patients. http://www.aium.org/officialstatements/57. Accessed 29 Jan 2016
  • 44.Australasian Society for Ultrasound in Medicine (revised November 2013) Policy on the Disinfection of Transducers http://www.asum.com.au/files/public/SoP/B2-Policy-on-the-Disinfection-of-Transducers.pdf. Accessed 3 Feb 2016
  • 45.Chalouhi GE, Salomon LJ, Marelle P, Bernard JP, Ville Y. Hygiène en échographie endocavitaire gynécologique et obstétricale en 2008 [Hygiene in endovaginal gynecologic and obstetrical ultrasound in 2008] J Gynecol Obstet Biol Reprod (Paris) 2009;38(1):43–50. doi: 10.1016/j.jgyn.2008.09.009. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ESM 1 (99.1KB, pdf)

(PDF 99 kb)


Articles from Insights into Imaging are provided here courtesy of Springer

RESOURCES