Development and validation of a novel LC–MS/MS method for simultaneous determination of abiraterone and its seven steroidal metabolites in human serum: Innovation in separation of diastereoisomers without use of a chiral column

Mohammad Alyamani a,e, Zhenfei Li a, Sunil K Upadhyay d, David J Anderson e, Richard J Auchus d, Nima Sharifi a,b,c,e,*
PMCID: PMC5124534  NIHMSID: NIHMS782839  PMID: 27063554

Abstract

Abiraterone acetate (AA), the prodrug of abiraterone, is FDA-approved for the treatment of castration-resistant prostate cancer. Abiraterone is metabolized in patients to a more potent analogue, D4A. However, we have recently reported that this analogue is further metabolized to additional metabolites in patients treated with AA. Here, we present a liquid chromatography-tandem mass spectrometry method developed to resolve and detect abiraterone and its seven metabolites in human serum using an AB Sciex Qtrap 5500 mass analyzer coupled with a Shimadzu Nexera UPLC station. Analytes and the internal standard (abiraterone-d4) were extracted from human serum using the liquid–liquid extraction procedure. The analytes were separated using a Zorbax Eclipse Plus C18 150 × 2.1 mm, 3.5 μm column at 40 °C and an isocratic mobile phase 35% A (0.1% formic acid in water), 65% B (0.1% formic acid in methanol: acetonitrile; 60:40). Electrospray ionization in positive mode was applied with multiple reaction monitoring in a total run time of 13 min. Abiraterone detection was linear in the range 2–400 ng/mL and all metabolites from 0.1–20 ng/mL. The method was validated following US FDA guidelines for bioanalytical method validation, and all the metabolite results were within the acceptance limits. Despite the similarity in structure and mass transition between the metabolites, the validated method separated all the metabolites, including diastereomers, to allow accurate identification and quantitation of each compound.

Keywords: Prostate cancer, Abiraterone, LC–MS/MS, Method validation, Metabolites

1. Introduction

Prostate cancer (PCa) progression depends on continued androgen receptor activity [1,2]. Androgen receptor activation occurs through the binding of the potent androgens, testosterone and dihydrotestosterone (DHT), which has been shown to promote malignant cell growth [3]. Androgen deprivation therapy (ADT), either by medical or surgical castration, is the treatment for patients with advanced disease [4]. However, in most cases, although the cancer initially responds to ADT, after time, it becomes resistant, and castration-resistant prostate cancer (CRPC) develops, which is the cause of nearly all PCa deaths [57]. In 2016 in the United States, the number of new PCa cases is estimated to be 180,000, with about 26,000 PCa-related deaths, making prostate cancer the most frequently diagnosed cancer and the second-leading cause of cancer death in men [8].

Abiraterone [abi; 17-(3-pyridyl)-androsta-5,16-dien-3β-ol] is a 17-heteroazole steroidal compound and a potent inhibitor of steroid 17α-hydroxylase/17,20-lyase (CYP17A1), an enzyme required for androgen synthesis [9]. Abi’s structure makes it a potent CYP17A1 inhibitor, and the double bond at C16 is necessary for functionally irreversible inhibition of the enzyme [1012]. Abi is administered as the prodrug, abiraterone acetate (AA), for the treatment of CRPC and prolongs survival for these patients [1315]. In 2011, the United States Food and Drug Administration (FDA) approved AA in the post-chemotherapy setting, and in December 2012 it approved the use of AA for patients with chemotherapy-naïve CRPC [16]. Prednisone is administered with AA to block mineralocorticoid excess that occurs with the simultaneous inhibition of cortisol synthesis [17,18].

It has been reported that the main circulating Abi metabolites are abiraterone sulfate and N-oxide abiraterone sulfate [19]. We have recently shown that Abi is converted by 3β-hydroxysteroid dehydrogenase (3β-HSD) to its Δ4, 3-keto congener (Δ4-abiraterone; D4A) [20] both in vitro and in vivo. The steroid A and B rings of D4A (Fig. 1) are identical to that of testosterone and androstenedione, which enables further metabolism of D4A [21]. Either the 5α-reductase or 5β-reductase enzymes could initiate the downstream metabolism of D4A. The 5α-reduced metabolites are 3-keto-5α-Abi, 3α-OH-5α-Abi and 3β-OH-5α-Abi, while the corresponding 5β-reduced metabolites are 3-keto-5β-Abi, 3α-OH-5β-Abi and 3β-OH-5β-Abi (Fig. 1).

Fig. 1.

Fig. 1

Steroidogenic enzyme metabolism of abiraterone and structure of the resultant metabolites. 3β-HSD: 3β-hydroxysteroid dehydrogenase, 5α-Reductase: steroid 5α-reductase, 5β-reductase: steroid 5β-reductase, 3α-HSD: 3α-hydroxysteroid dehydrogenase.

The separation of the Abi metabolites and the parent Abi compound is expected to be difficult, given their structural similarities. Here, we describe and validate a novel method for the determination of Abi and its 7 metabolites (Fig. 1) in human serum. Separation of all diastereomeric metabolites was achieved using a reversed-phase chromatographic technique, not requiring a chiral column. We demonstrated that all seven steroidal Abi metabolites are present in serum samples from patients with CRPC undergoing treatment with AA [21].

2. Experimetal

2.1. Materials

Abiraterone and its seven metabolites 3-keto-Δ4-Abi, 3-keto-5α-Abi, 3α-OH-5α-Abi, 3β-OH-5α-Abi, 3-keto-5β-Abi, 3α-OH-5β-Abi and 3β-OH-5β-Abi were synthesized as described previously [21]. The internal standard (abiraterone-d4) was purchased from Toronto Research Chemicals, (Toronto, Canada). Methanol, acetonitrile, water, and formic acid were LC–MS grade, methyl tertiary butyl ether (MTBE) was HPLC grade and all were from Fisher Scientific (Fair Lawn, NJ). Double charcoal-stripped human serum (Golden West Biological Inc., Temecula, CA) was used for all samples in the validation studies (controls, calibrators) which were processed by spiking with known concentrations of the analytes.

2.2. Preparation of standards and quality control (QC) samples

The stock solutions (1 mg/mL) of Abi and its metabolites were prepared by dissolving the synthesized powder in 100% methanol. These solutions were further diluted with methanol:H2O, 1:1, and then mixed in one flask to prepare the working standard at concentration of 5.0 μg/mL Abi and 0.25 μg/mL for each metabolite. Freshly prepared working standard was used to prepare the serum calibrators and serum quality control samples, all of which contained appropriate concentrations of all the analytes. Quality control (QC) samples were prepared at three levels, based on the lower limit of quantitation (LLOQ) and the upper limit of quantitation (ULOQ): QC Low (3x LLOQ), QC Mid (half ULOQ), and QC High, which is 80% of ULOQ. The final concentrations of the calibrators and quality control in serum are listed in Table 1. The internal standard stock (2 mg/mL) was prepared by dissolving the stock powder in 100% methanol and further diluting it with methanol:H2O, 1:1 to a final concentration of 1.25 μg/mL. Stock and working standard solutions were stored at −20 °C, and the calibrators with quality controls were freshly prepared through the validation.

Table 1.

Calibrators and quality control samples preparation and final concentration.

Sample Volume of Working
Standard (μL)
Final Volume (mL) Final Concentration in
Serum ng/mL
Abi Metabolites
Cal.1 20 10 2 0.1
Cal. 2 40 10 4 0.2
Cal. 3 120 10 12 0.6
Cal. 4 300 5 60 3
Cal. 5 600 5 120 6
Cal. 6 1400 5 280 14
Cal. 7 2000 5 400 20
QC Low 60 10 6 0.3
QC Mid 1000 5 200 10
QC High 1600 5 320 16

2.3. Sample preparation

Each calibrator, quality control and patient serum sample was taken through the following sample preparation steps. The analytes were extracted from human serum using a liquid–liquid extraction procedure. 100 μL of serum with spiked analytes was placed in a glass tube, and 20 μL of 1.25 μg/mL internal standard working solution was added. The samples were vortexed for 30 s. After addition of 2 mL MTBE, the samples were vortexed for 1 min. The samples were then centrifuged for 5 min at 4000 rpm at 4 °C. The organic layer was transferred to another tube and evaporated to dryness under nitrogen at 40 °C. The dried extract was then reconstituted with 300 μL 1:1 methanol:H2O, and 200 μL was transferred to an HPLC vial.

2.4. Instrumentation and data analysis

The LC–MS/MS system was an ultra-pressure liquid chromatography (UPLC) system (Shimadzu Corporation, Japan), consisting of a LC-30AD solvent delivery system, a DGU-20A5R vacuum degasser, a CTO-30A thermostated column oven, SIL-30AC auto-sampler, and a system CBM-20A controller, that was coupled with a Qtrap 5500 mass spectrometer (AB Sciex, Redwood City, CA). Data acquisition and processing were performed using Analyst software (version 1.6.2) from AB Sciex. The peak area ratio of the analyte over the internal standard was used for quantification purposes.

2.5. UPLC–MS/MS conditions

Separation of drug metabolites was achieved using a Zorbax Eclipse Plus C18 column 150 mm × 2.1 mm, 3.5 μm (Agilent, Santa Clara, CA) at 40 °C with an isocratic mobile phase consisting of 35% A (0.1% formic acid in water) and 65% B (0.1% formic acid in methanol:acetonitrile, 60:40), at a flow rate of 0.2 mL/min. Sample injection volume was 10 μL, and analytical run time was 13 min.

The mass spectrometer was operated in positive ion mode using an electrospray ionization (ESI) source. Tuning parameters were optimized for the Abi metabolites and internal standard by infusing a solution containing 200 ng/mL of each analyte. Nitrogen was used as the nebulizing (40 L/min) agent and drying gas (30 L/min). Ion spray voltage and nebulizer temperature were regulated at 2500 V and 500 °C, respectively. All the analytes have the same declustering potential (120 V), collision energy (60 V), entrance potential (10 V) and collision exit potential (13 V). The analytes were quantified using multiple reaction monitoring (MRM) with the mass transitions for each compound as listed in Table 2.

Table 2.

Mass transition parameters for abiraterone, its metabolites, and the internal standard.

Analyte Q1
(m/z)
Q3
(m/z)
Abiraterone 350.5 156.1
D4A 348.3 156.1
3-keto-5α-Abi 350.3 156.2
3α-OH-5α-Abi 352.4 156.2
3β-OH-5α-Abi 352.3 156.1
3-keto-5β-Abi 350.4 156.1
3α-OH-5β-Abi 352.4 156.4
3β-OH-5β-Abi 352.1 156.1
Abiraterone-d4 (IS) 354.4 160.1

IS = internal standard.

2.6. Method validation

The method was validated following the US FDA guidance for bioanalytical method validation [22]. The validation parameters evaluated were linearity, accuracy and precision, selectivity, recovery, matrix effect, and stability.

2.6.1. Linearity

Six calibration curves were constructed in same serum matrix for each analyte to assess the linear range. The ratio of the analyte peak area over the internal standard peak area was plotted versus the nominal concentration. The correlation was then fit by linear regression, applying a weighting factor of 1/x. To assess selectivity, blank and zero (blank spiked with internal standard) samples were prepared with each calibration curve. The calibration curve consisted of 7 non-zero points with single repeat for each point. Based on the FDA guidelines, the acceptable deviation from the nominal concentration is <20% for the LLOQ calibrator and <15% for the other calibrators. Validation of the calibration curve requires 67% of the calibrators, including LLOQ and ULOQ, being within these acceptance criteria.

2.6.2. Accuracy, precision, and LLOQ

In order to determine the accuracy and precision, QC samples (LLOQ, low, mid, and high) were run. Intra-day precision was determined with five replicates and inter-day precision was determined running one of each QC sample on 3 separate days. Accuracy was determined by how close the mean of the intra-day QC sample results were to the nominal value. The method is considered accurate if the measured concentration is within 85–115% (80–120% for LLOQ) of the expected value. The acceptable precision criterion is a coefficient of variation percent (CV%) of the QC sample analyte concentration being no more than 15% (LLOQ not more than 20%).

2.6.3. Selectivity

Single runs of six different serum batches were used to study the selectivity of the method. To insure that the samples were free of interference, LLOQ samples were prepared and compared to the corresponding blank samples. The requirement of LLOQ is to be at least 5 times the peak height (or peak area) of any peak detected in the blank samples at the same retention time of the analyte.

2.6.4. Recovery

Relative recovery was assessed by calculating the ratio of the analyte/internal standard peak area in spiked samples before extraction over the analyte/internal standard peak area in spiked samples after extraction. Three QC samples (low, mid, and high) were determined. Each of the three QC samples, including both the pre- and post-extraction samples were assayed in triplicate. The FDA criterion for acceptable recovery results need to be consistent, precise and reproducible, as assessed by the coefficient of variation for each QC sample type being no more than 15%. The CV% was calculated using this formula CV% = [(DX/X)2 + (DY/Y)2)]1/2, where DX is the standard deviation for the pre-extraction samples, X is the mean area ratio of pre-extraction samples, DY is the standard deviation for the post-extraction samples, and Y is the mean area ratio of post-extraction samples.

2.6.5. Matrix effect

Matrix effect, calculated as a matrix factor percent (MF%), was assessed as a ratio of peak area ratio of spiked analyte/internal standard determined for QC low sample after extraction (n = 18) over the peak area ratio of the spiked analyte/internal standard in methanol: H2O, 1:1 (n = 3). The QC sample was tested in triplicate on each of six different serum batches. CV% was calculated as mentioned in section 2.6.4.

2.6.6. Stability

Analyte stability was evaluated for analytes in both solution and serum. Diluted standard solution stability was evaluated after 6 h at room temperature and after 9 days storage at 4 °C for all eight analytes at LLOQ and ULOQ levels and for the internal standard working solution. The stability was assessed in triplicate by comparing peak area of the old (6 h and 9 days) solutions with freshly prepared from stock stored at −20 °C.

In order to ensure that the processed samples were stable during the analytical run, post-preparative stability was studied at the three QC levels for 43 h at the auto sampler temperature of 4 °C by analyzing five samples per QC level. Analyte stability in serum was determined by preparing samples at two QC levels (low and high) and analyzing six serum samples containing a known concentration of each analyte as follows: bench top stability at room temperature after 21 h, after three freeze-thaw cycles (24 h at −80 °C with thaw unassisted at room temperature), and long-term stability at −80 °C after 28 weeks of storage. In order to evaluate the stability, the concentrations in these samples were compared to the nominal values by running a freshly prepared calibration curves. The accuracy should fall within the 85–115% acceptance criteria and RSD not to exceed 15%

3. Results and discussion

3.1. Method development

In the development of a comprehensive pharmacokinetic technique determining parent drug and metabolites, the challenge is to separate all of these compounds, while achieving the highest sensitivity without compromising linearity. Several mobile phases were evaluated including different organic modifiers (methanol, acetonitrile and methanol/acetonitrile mixtures) at various concentrations in water, as well mobile phases containing ammonium formate and/or formic acid without organic modifier. The Prodigy Phenomenex C18 analytical columns were also evaluated. Optimization of sample preparation according to recovery was also performed, varying volumes of MTBE, as well as evaluating acidic and neutral reconstitution solutions. The best method was then selected based on the separation of all metabolites as well as based on the analytical qualities of linearity and sensitivity.

3.2. Optimization of the mass analyzer and chromatographic conditions

Because these metabolites are steroidal compounds containing a pyridyl moiety, the mass analyzer was operated with electrospray ionization in the positive mode. The parameters were optimized to ensure that the highest sensitivity possible would be achieved. The settings of the mass spectrometric parameters used in the optimized method are listed in Section 2.5, and the analyte mass transitions are given in Table 2.

3.3. Optimization of the chromatographic conditions

Many of the metabolites are isomeric compounds, and thus it is not possible to distinguish between them based on their MRM transitions. Therefore, we separated the metabolites (as well as the parent compound) by chromatography. Optimization of isocratic chromatographic conditions was done investigating the effect of various mobile phase components at different concentrations. A mobile phase of 35% water mixed with 65% methanol/acetonitrile (60:40) (both containing 0.1% formic acid) resulted in the optimized chromatographic performance, separating the parent compound and metabolites, as shown in Fig. 2. It should be noted that the method was able to resolve diastereomeric metabolites with a C18 column, not requiring chiral columns for separation of these stereoisomers. The employment of a methanol/acetonitrile mixture was critical to achieve this separation, as methanol or acetonitrile as sole organic modifiers in the mobile phase did not resolve the chiral compounds.

Fig. 2.

Fig. 2

Chromatogram for abiraterone and its metabolites. The results were obtained by injecting 10 μL of 50 ng/mL standard solutions for A, Abi; B, D4A; C, 3-keto-5α-Abi; D, 3α-OH-5α-Abi; E, 3β-OH-5α-Abi; F, 3-keto-5β-Abi; G, 3α-OH-5β-Abi; and H, 3β-OH-5β-Abi.

3.4. Method validation

The method was validated according to FDA guidelines. As discussed below, all criteria of the guidelines were met for the present method. All the analytes were stable in solution and serum, meeting the stability criterion. The criteria for linearity, accuracy and precision fell within the acceptance criteria. The method gave excellent recovery without matrix effects or interference.

3.4.1. Linearity

Six sets of calibrators in the same serum matrices were used to generate calibration plots to evaluate the linearity for each analyte. Each calibration curve consisted of seven non-zero points. The plot of the response ratio (analyte peak area over internal standard peak area) versus the analyte concentration was linear, the R2 values for all analytes were ≥0.9979. The mean values for slope, intercept and R2 values for each analyte are listed in Table 3.

Table 3.

Mean values for slope, intercept and R2 values obtained from six different calibration curves.

Analyte Slope Intercept R2
Abiraterone 1.80E-02 8.17E-03 0.9990
D4A 2.47E-02 3.00E-06 0.9996
3-keto-5α-Abi 2.00E-02 1.04E-04 0.9995
3α-OH-5α-Abi 2.16E-02 −6.33E-05 0.9995
3β-OH-5α-Abi 2.56E-02 1.21E-04 0.9996
3-keto-5β-Abi 1.15E-02 3.80E-05 0.9995
3α-OH-5β-Abi 1.26E-02 −3.67E-05 0.9995
3β-OH-5β-Abi 2.03E-03 5.00E-06 0.9991

3.4.2. Accuracy, precision, and LLOQ

The results for inter- and intra-day precision and accuracy determination are given in Table 4. Five replicates of each QC sample were prepared in three different days, to test the method accuracy and precision. The CV% of the 5 replicates determining intra-day precision and the CV% of the 15 replicates determining inter-day precision for the eight analytes ranges between 0.99–10.75% and 2.84–12.18%, respectively. The accuracy values for intra-day and inter-day fall between 91.7–107.4% and 93.8–103.5%, respectively. Five LLOQ samples were prepared in the first batch, and the intraday accuracy mean values were 85.3–111.2% and CV% 2.63–16.89%. All the results fall within the FDA acceptance criteria.

Table 4.

Intra- and inter-day accuracy and precision for abiraterone and its metabolites in LLOQ and QC samples.

Analyte Intra-day (n = 5)
Inter-day (n = 15)
Concentrationa
Mean ± SD
Accuracy
%
RSD
%
Concentrationa
Mean ± SD
Accuracy
%
RSD
%
Abirateroneb 2.05 ± 0.06 102.7 2.89 n.a n.a n.a
6.40 ± 0.17 106.8 2.65 6.21 ± 0.31 103.5 4.95
203.8 ± 4.2 102.0 2.06 206.3 ± 5.9 103.3 2.84
316 ± 12 98.7 3.86 317 ± 22 99.1 6.80
D4Ac 0.111 ± 0.004 111.2 3.89 n.a n.a n.a
0.307 ± 0.024 102.3 7.76 0.297 ± 0.020 98.9 6.68
10.02 ± 0.19 100.2 1.86 9.74 ± 0 0.38 97.4 3.94
16.4 ± 0.5 102.3 3.02 16.0 ± 1.0 100.1 6.35
3-keto-5α-Abic 0.111 ± 0.004 111.0 3.43 n.a n.a n.a
0.322 ± 0.005 107.4 1.56 0.301 ± 0.02 100.1 7.86
10.32 ± 0.29 103.2 2.86 9.87 ± 0.5 98.7 5.03
16.8 ± 0.8 104.9 4.68 16.5 ± 1.2 103.1 7.56
3α-OH-5α-Abic 0.106 ± 0.010 106.2 9.23 n.a n.a n.a
0.314 ± 0.020 104.6 5.96 0.304 ± 0.030 101.5 9.45
9.91 ± 0.44 99.1 4.41 9.39 ± 0.59 93.9 6.33
14.8 ± 0.7 92.6 4.97 15.1 ± 1.2 94.5 7.61
3β-OH-5α-Abic 0.103 ± 0.003 103.0 2.63 n.a n.a n.a
0.305 ± 0.010 101.5 4.21 0.301 ± 0.01 100.3 4.66
9.92± 0.20 99.2 2.07 9.86 ± 0.39 98.6 3.99
16.0 ± 0.7 99.9 4.63 16.1 ± 1.0 100.6 6.51
3-keto-5β-Abic 0.106 ± 0.004 106.4 4.13 n.a n.a n.a
0.316 ± 0.008 105.4 2.56 0.303 ± 0.03 101.0 8.77
9.95 ± 0.27 99.5 2.70 9.56 ± 0.44 95.6 4.61
15.1 ± 0.5 94.2 3.61 15.5 ± 1.2 96.5 7.80
3α-OH-5β-Abic 0.108 ± 0.004 108.4 4.26 n.a n.a n.a
0.308 ± 0.020 102.6 6.74 0.301 ± 0.02 100.4 7.55
9.74 ± 0.29 97.4 2.98 9.37 ± 0.40 93.8 4.26
15.0 ± 0.7 93.8 4.65 15.3 ± 1.1 95.8 7.46
3β-OH-5β-Abic 0.085 ± 0.010 85.3 16.89 n.a n.a n.a
0.275 ± 0.030 91.7 10.75 0.29 ± 0.03 97.6 12.18
9.68 ± 0.09 96.8 0.99 9.76 ± 0.36 97.6 3.67
14.9 ± 0.8 93.3 5.60 16.0 ± 1.5 99.7 9.66
a

Measured Concentration in ng/ml.

b

Nominal Concentrations: (LLOQ = 2, QC Low = 6, Mid = 200, and High = 320ng/mL).

c

Nominal Concentrations: (LLOQ = 0.t QC Low = 0.3, Mid = 10, and High = 16 ng/mL).

3.4.3. Selectivity

Six LLOQ samples, each prepared in different serum batches, were compared to the blank samples showed that the serum was free from interference. The MRM chromatograms for the six representative blank samples compared to the MRM chromatograms of LLOQ samples of Abi and its 7 metabolites are shown in Fig. 3. The analyte peak in the spiked LLOQ samples ranged from a factor 6.7–592.6 greater than the blank serum matrix with largest interference peak at the same retention time, meeting the LLOQ requirement being a factor greater than 5 for the technique to be judged free of interference.

Fig. 3.

Fig. 3

Representative chromatogram of selectivity studies comparing blank samples (left chromatograms) matched to LLOQ (right chromatograms) for A, Abi; B, D4A; C, 3-keto-5α-Abi; D, 3-keto-5β-Abi; E, 3α-OH-5α-Abi; F, 3α-OH-5β-Abi; G, 3β-OH-5α-Abi; and H, 3β-OH-5β-Abi.

3.4.4. Recovery

Three QC levels were used to study the relative recovery, which is determined by a percent ratio of the analyte/internal standard peak area of a spiked sample before extraction over the analyte/internal standard peak area of a spiked sample after extraction. The results should be consistent, precise, and reproducible. The recovery was calculated as the mean of three triplicates for each analyte at each QC level. Recovery results were good, ranging from 85.1–105.2% (Table 5). CV% given in Table 5 ranged from 1.09–13.47% for all QC samples, meeting the 15% CV% criterion for acceptable recovery.

Table 5.

Matrix effect and recovery for abiraterone and its metabolites.

MF% (n = 18)
Recovery% (n = 3)
QC low
QC low
QC Mid
QC High
Mean (%) RSD (%) Mean (%) RSD (%) Mean (%) RSD
(%)
Mean (%) RSD (%)
Abi 114.7 5.04 95.7 6.54 91.7 1.49 93.4 4.65
D4A 88.5 5.99 105.2 4.78 99.0 3.39 94.6 7.81
3-keto-5α-Abi 96.1 5.22 91.6 7.47 89.1 7.62 97.7 9.12
3α-OH-5α-Abi 92.0 5.05 91.2 7.48 85.1 4.51 95.1 5.61
3β-OH-5α-Abi 99.4 4.57 88.5 6.87 86.9 2.12 92.1 5.78
3-keto-5β-Abi 93.2 8.31 91.0 13.47 87.1 6.33 96.3 6.39
3α-OH-5β-Abi 92.0 6.74 86.0 8.47 85.1 1.09 91.1 5.64
3β-OH-5β-Abi 100.0 11.13 96.6 11.50 86.3 4.28 91.4 8.16

RSD% - relative standard deviation percent which is the same as CV%.

3.4.5. Matrix effect

The matrix effect is quantified as a matrix factor percent (MF%) by calculating the ratio the peak area ratio of analyte/internal standard of a QC low sample (post-extraction addition of analyte) over the peak area ratio of analyte/internal standard in a methanol/water solution. Matrix effect results are given in Table 5, showing that the technique is essentially free of interference, with results ranging from 88.5–114.7% for QC low samples.

3.4.6. Stability

Analytes were stable in the diluted solution stored at room temperature for 6 h 98.5–115.0% and at 4 °C for 9 days 83.2–109.5% as given in Table 6. Results of serum-based stability studies employing QC samples are given in Table 7. Post-preparative stability in which the QC samples were stored in the autosampler at 4 °C for 43 h were found to be acceptable 96.1–103.3%. Short-term (21 h) stability of all analytes in the prepared QC samples at each concentration were within the acceptance criteria 88.2–114.8% for the determined mean concentration values. Even though all analytes did show a decrease in concentration when subjected to three freeze-thaw cycles 86.3–96.7% and long term stability 86.7–95.6%, the results were less than 15%, which is within the criteria for accuracy.

Table 6.

Stock stability for abiraterone, its metabolites and the internal standard.

6 hours at RT (n = 3)
9 days at 4 °C (n = 3)
LLOQ (%) ULOQ (%) LLOQ (%) ULOQ (%)
Abi 98.5 103.5 109.5 98.4
D4A 100.2 102.4 91.1 86.9
3-keto-5α-Abi 112.1 103.7 101.6 88.4
3α-OH-5α-Abi 103.8 102.7 97.9 85.7
3β-OH-5α-Abi 105.2 102.1 101.7 89.3
3-keto-5β-Abi 107.6 103.7 97.5 83.2
3α-OH-5β-Abi 104.1 103.3 105.0 85.4
3β-OH-5β-Abi 115.0 104.6 95.4 89.7
Internal standard 105.7 86.2

RT, Room temperature.

Table 7.

Stability of abiraterone and its metabolites after preparation and in serum.

Post-Preparative
42 h
Short-term
21 h
Freeze- thaw
3 cycles
Long-term stability 28 weeks
QC Low QC Mid QC High QC Low QC High QC Low QC High QC Low QC
High
Abi1 Concentration 5.9 203.4 314.6 5.3 293.8 5.4 284.9 5.7 290.7
Accuracy% 97.8 101.7 98.3 88.2 91.8 89.5 89.0 95.5 90.8
% RSD 3.26 0.75 3.04 6.03 1.97 11.38 9.45 0.82 2.28
D4A2 Concentration 0.3 9.6 16.0 0.3 18.4 0.3 15.3 0.3 14.2
Accuracy% 96.7 96.1 100.3 100.0 114.8 86.7 95.8 90.0 88.6
% RSD 3.68 2.87 2.67 3.03 1.86 8.82 11.71 3.64 1.42
3-keto-5α-Abi2 Concentration 0.3 10.0 16.3 0.3 15.7 0.3 14.7 0.3 14.4
Accuracy% 100.0 100.2 101.6 93.3 98.1 86.7 91.6 86.7 90.1
% RSD 4.49 1.25 3.95 8.04 2.57 12.84 14.11 5.90 5.59
3α-OH-5α-Abi2 Concentration 0.3 10.1 16.3 0.3 15.8 0.3 14.1 0.3 14.2
Accuracy% 103.3 100.7 102.0 96.7 98.7 86.7 88.2 86.7 88.6
% RSD 3.48 2.02 3.89 7.16 3.40 9.82 14.26 7.32 4.34
3β-OH-5α-Abi2 Concentration 0.3 9.9 15.7 0.3 15.2 0.3 13.9 0.3 14.1
Accuracy% 100.0 98.7 98.1 90.0 95.1 86.7 86.7 86.7 87.9
% RSD 3.16 1.37 4.20 10.28 2.22 14.55 11.27 4.58 2.50
3-keto-5β-Abi2 Concentration 0.3 10.2 16.3 0.3 16.3 0.3 14.2 0.3 14.4
Accuracy% 103.3 101.7 102.0 100.0 101.8 93.3 88.9 86.7 90.1
% RSD 6.40 1.62 4.03 7.58 3.66 14.57 13.44 4.92 4.26
3α-OH-5β-Abi2 Concentration 0.3 9.6 15.6 0.3 16.1 0.3 13.8 0.3 15.0
Accuracy% 100.0 96.2 97.5 96.7 100.8 86.7 86.3 90.0 93.5
% RSD 4.81 2.06 4.20 7.77 2.29 10.66 12.19 3.65 3.53
3β-OH-5β-Abi2 Concentration 0.3 9.8 16.1 0.3 16.5 0.3 14.4 0.3 14.7
Accuracy% 103.3 98.2 100.4 93.3 103.0 96.7 89.8 93.3 91.9
% RSD 7.82 1.27 3.19 10.44 5.30 14.41 13.84 5.21 3.56

RSD% = relative standard deviation% is the same as CV%;

1

Nominal Concentrations: (QC Low = 6, Mid = 200, and High = 320 ng/mL);

2

Nominal Concentrations: (QC Low = 0.3, Mid = 10, and High = 16 ng/mL).

4. Conclusion

We have developed and validated an LC–MS/MS MRM method for the determination and accurate quantification of Abi and its seven steroidal metabolites in human serum. To our knowledge, this is the first published report of a method to determine Abi metabolites that result from steroidogenic metabolism. The validated LC–MS/MS method resolved and quantitated all the metabolites despite the similarity in their structures, including resolving diastereomers, which precludes analysis of co-eluting isomers based solely on their MRM transitions. Reversed-phase chromatographic conditions were identified to accomplish the separation of all metabolites and their subsequent accurate quantification. This validated method can be applied to determine Abi and the aforementioned metabolites in human serum in clinical trials in which patients are treated with AA.

Acknowledgments

This work has been supported in part by funding from a Howard Hughes Medical Institute Physician-Scientist Early Career Award (to N.S.), a Prostate Cancer Foundation Challenge Award (to N.S.), an American Cancer Society Research Scholar Award (12-038-01-CCE; to N.S.), grants from the National Cancer Institute (R01CA168899, R01CA172382, and R01CA190289; to N.S.), a grant from the US Army Medical Research and Materiel Command (PC121382 to Z.L.), a Prostate Cancer Foundation Young Investigator Award to Z.L.

References

  • [1].Attard G, Parker C, Eeles RA, Schroder F, Tomlins SA, Tannock I, Drake CG, de Bono JS. Prostate cancer, Lancet (London, England) 2016;387:70–82. doi: 10.1016/S0140-6736(14)61947-4. [DOI] [PubMed] [Google Scholar]
  • [2].Miyamoto DT, Lee RJ, Stott SL, Ting DT, Wittner BS, Ulman M, Smas ME, Lord JB, Brannigan BW, Trautwein J, Bander NH, Wu CL, Sequist LV, Smith MR, Ramaswamy S, Toner M, Maheswaran S, Haber DA. Androgen receptor signaling in circulating tumor cells as a marker of hormonally responsive prostate cancer. Cancer Discov. 2012;2:995–1003. doi: 10.1158/2159-8290.CD-12-0222. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [3].Deslypere JP, Young M, Wilson JD, McPhaul MJ. Testosterone and 5 alpha-dihydrotestosterone interact differently with the androgen receptor to enhance transcription of the MMTV-CAT reporter gene. Mol. Cell. Endocrinol. 1992;88:15–22. doi: 10.1016/0303-7207(92)90004-p. [DOI] [PubMed] [Google Scholar]
  • [4].Sharifi N, Gulley JL, Dahut WL. Androgen deprivation therapy for prostate cancer. JAMA. 2005;294:238–244. doi: 10.1001/jama.294.2.238. [DOI] [PubMed] [Google Scholar]
  • [5].Scher HI, Halabi S, Tannock I, Morris M, Sternberg CN, Carducci MA, Eisenberger MA, Higano C, Bubley GJ, Dreicer R, Petrylak D, Kantoff P, Basch E, Kelly WK, Figg WD, Small EJ, Beer TM, Wilding G, Martin A, Hussain M. Design and end points of clinical trials for patients with progressive prostate cancer and castrate levels of testosterone: recommendations of the Prostate Cancer Clinical Trials Working Group. J. Clin. Oncol. 2008;26:1148–1159. doi: 10.1200/JCO.2007.12.4487. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Scher HI, Sawyers CL. Biology of progressive, castration-resistant prostate cancer: directed therapies targeting the androgen-receptor signaling axis. J. Clin. Oncol. 2005;23:8253–8261. doi: 10.1200/JCO.2005.03.4777. [DOI] [PubMed] [Google Scholar]
  • [7].Sharifi N. Minireview: androgen metabolism in castration-resistant prostate cancer. Mol. Endocrinol. 2013;27:708–714. doi: 10.1210/me.2013-1007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [8].Siegel RL, Miller KD, Jemal A. Cancer statistics 2016. CA Cancer J. Clin. 2016;66:7–30. doi: 10.3322/caac.21332. [DOI] [PubMed] [Google Scholar]
  • [9].Potter GA, Barrie SE, Jarman M, Rowlands MG. Novel steroidal inhibitors of human cytochrome P45017 alpha (17 alpha-hydroxylase-c17,20-lyase): potential agents for the treatment of prostatic cancer. J. Med. Chem. 1995;38:2463–2471. doi: 10.1021/jm00013a022. [DOI] [PubMed] [Google Scholar]
  • [10].Jarman M, Barrie SE, Llera JM. The 16,17-double bond is needed for irreversible inhibition of human cytochrome p45017alpha by abiraterone (17-(3-pyridyl)androsta-5, 16-dien-3beta-ol) and related steroidal inhibitors. J. Med. Chem. 1998;41:5375–5381. doi: 10.1021/jm981017j. [DOI] [PubMed] [Google Scholar]
  • [11].Salvador JA, Pinto RM, Silvestre SM. Steroidal 5alpha-reductase and 17alpha-hydroxylase/17,20-lyase (CYP17) inhibitors useful in the treatment of prostatic diseases. J. Steroid Biochem. Mol. Biol. 2013;137:199–222. doi: 10.1016/j.jsbmb.2013.04.006. [DOI] [PubMed] [Google Scholar]
  • [12].Garrido M, Peng HM, Yoshimoto FK, Upadhyay SK, Bratoeff E, Auchus RJ. A-ring modified steroidal azoles retaining similar potent and slowly reversible CYP17A1 inhibition as abiraterone. J. Steroid Biochem. Mol. Biol. 2014;143:1–10. doi: 10.1016/j.jsbmb.2014.01.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Grist E, Attard G. The development of abiraterone acetate for castration-resistant prostate cancer. Urol. Oncol. 2015;33:289–294. doi: 10.1016/j.urolonc.2015.03.021. [DOI] [PubMed] [Google Scholar]
  • [14].de Bono JS, Logothetis CJ, Molina A, Fizazi K, North S, Chu L, Chi KN, Jones RJ, Goodman OB, Jr., Saad F, Staffurth JN, Mainwaring P, Harland S, Flaig TW, Hutson TE, Cheng T, Patterson H, Hainsworth JD, Ryan CJ, Sternberg CN, Ellard SL, Flechon A, Saleh M, Scholz M, Efstathiou E, Zivi A, Bianchini D, Loriot Y, Chieffo N, Kheoh T, Haqq CM, Scher HI. Abiraterone and increased survival in metastatic prostate cancer. N. Engl. J. Med. 2011;364:1995–2005. doi: 10.1056/NEJMoa1014618. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [15].Ryan CJ, Smith MR, de Bono JS, Molina A, Logothetis CJ, de Souza P, Fizazi K, Mainwaring P, Piulats JM, Ng S, Carles J, Mulders PF, Basch E, Small EJ, Saad F, Schrijvers D, Van Poppel H, Mukherjee SD, Suttmann H, Gerritsen WR, Flaig TW, George DJ, Yu EY, Efstathiou E, Pantuck A, Winquist E, Higano CS, Taplin ME, Park Y, Kheoh T, Griffin T, Scher HI, Rathkopf DE. Abiraterone in metastatic prostate cancer without previous chemotherapy. N. Engl. J. Med. 2013;368:138–148. doi: 10.1056/NEJMoa1209096. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Kluetz PG, Ning YM, Maher VE, Zhang L, Tang S, Ghosh D, Aziz R, Palmby T, Pfuma E, Zirkelbach JF, Mehrotra N, Tilley A, Sridhara R, Ibrahim A, Justice R, Pazdur R. Abiraterone acetate in combination with prednisone for the treatment of patients with metastatic castration-resistant prostate cancer: U. S. Food and Drug Administration drug approval summary. Clin. Cancer Res. Off. J. Am. Assoc. Cancer Res. 2013;19:6650–6656. doi: 10.1158/1078-0432.CCR-13-2134. [DOI] [PubMed] [Google Scholar]
  • [17].Attard G, Reid AH, Auchus RJ, Hughes BA, Cassidy AM, Thompson E, Oommen NB, Folkerd E, Dowsett M, Arlt W, de Bono JS. Clinical and biochemical consequences of CYP17A1 inhibition with abiraterone given with and without exogenous glucocorticoids in castrate men with advanced prostate cancer. J. Clin. Endocrinol. Metab. 2012;97:507–516. doi: 10.1210/jc.2011-2189. [DOI] [PubMed] [Google Scholar]
  • [18].Auchus RJ, Yu MK, Nguyen S, Mundle SD. Use of prednisone with abiraterone acetate in metastatic castration-resistant prostate cancer. Oncologist. 2014;19:1231–1240. doi: 10.1634/theoncologist.2014-0167. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Acharya M, Gonzalez M, Mannens G, De Vries R, Lopez C, Griffin T, Tran N. A phase I open-label, single-dose, mass balance study of 14C-labeled abiraterone acetate in healthy male subjects. Xenobiotica Fate Foreign Comp. Biol. Syst. 2013;43:379–389. doi: 10.3109/00498254.2012.721022. [DOI] [PubMed] [Google Scholar]
  • [20].Li Z, Bishop AC, Alyamani M, Garcia JA, Dreicer R, Bunch D, Liu J, Upadhyay SK, Auchus RJ, Sharifi N. Conversion of abiraterone to D4A drives anti-tumour activity in prostate cancer. Nature. 2015;523:347–351. doi: 10.1038/nature14406. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Li Z, Alyamani M, Li J, Rogacki K, Abazeed M, Upadhyay SK, Balk SP, Taplin M, Auchus RJ, Sharifi N. Redirecting abiraterone metabolism to fine tune prostate cancer antiandrogen therapy. Nature. 2016 doi: 10.1038/nature17954. (in press) [DOI] [PMC free article] [PubMed] [Google Scholar]