Abstract
Magnetometers are essential for scientific investigation of planetary bodies and are therefore ubiquitous on missions in space. Fluxgate and optically pumped atomic gas based magnetometers are typically flown because of their proven performance, reliability, and ability to adhere to the strict requirements associated with space missions. However, their complexity, size, and cost prevent their applicability in smaller missions involving cubesats. Conventional solid-state based magnetometers pose a viable solution, though many are prone to radiation damage and plagued with temperature instabilities. In this work, we report on the development of a new self-calibrating, solid-state based magnetometer which measures magnetic field induced changes in current within a SiC pn junction caused by the interaction of external magnetic fields with the atomic scale defects intrinsic to the semiconductor. Unlike heritage designs, the magnetometer does not require inductive sensing elements, high frequency radio, and/or optical circuitry and can be made significantly more compact and lightweight, thus enabling missions leveraging swarms of cubesats capable of science returns not possible with a single large-scale satellite. Additionally, the robustness of the SiC semiconductor allows for operation in extreme conditions such as the hot Venusian surface and the high radiation environment of the Jovian system.
The planetary science community has identified a strong need for scientific instruments which address crosscutting themes of planetary science: building new worlds, planetary habitats, and workings of the solar system1. Addressing aspects of all of these themes, magnetometers remotely probe the interiors of solar system bodies without the need to invasively penetrate the bodies being investigated. Magnetic field measurements have been used to better understand the internal workings of the planetary objects and have also been used in conjunction with simulations and models to shed insight into the predictive formation and evolution of the planets, satellites, and even the solar system. They also allow for a better understanding of planetary atmospheres and their interaction with the solar wind which influences climate and the ability to harbor life. Magnetic field measurements have also been useful for indirect detection of water, a requirement for life as we know it. For example, the magnetometer aboard the Galileo spacecraft that passed by Jupiter in 1996 acquired data that corroborated the idea that liquid water is present beneath the ice sheet of Europa, one of Jupiter’s moons. Specifically, computer modeling has shown that a planetary-scale conductive liquid ocean is necessary to explain the complicated, time-varying magnetic measurements of the Jovian environment2,3,4. More recently in 2005, the magnetometer on the Cassini spacecraft which will soon begin its grand finale tour of Saturn’s innermost rings, helped determine that the Saturn moon Enceladus is also likely to have a subsurface liquid ocean. By measuring the gyration frequency of the ionized gas ejected from the moon’s southern plumes in the presence of Saturn’s magnetic field, it was determined that the outgassing material is consistent with ionized water vapor5,6,7. These magnetometer-aided scientific discoveries are exciting because they suggest the existence of possible life sustaining environments beyond Earth. So much interest has been generated around these two moons that future missions are currently being planned to enable more detailed scientific investigations. A dedicated mission to Europa has recently been developed toward a proposed launch sometime in the 2020’s which would be equipped with a suite of magnetometers (both flux gate and optically pumped helium gas) called the Interior Characterization of Europa using Magnetometry (ICEMAG)8,9. The objective of ICEMAG is to characterize the complex magnetic environment of the Jupiter system at Europa at multiple frequencies, which will allow for a better understanding of the moon and potential life harboring environments beneath the ice. ICEMAG plans to use magnetic field measurements to determine the induction response at Europa, leading to constraints on the thickness and salinity of Europa’s ocean as well as the thickness of the ice shell. An additional ICEMAG goal is to use magnetic field measurements to probe the composition of molecular species picked up from Europa’s exosphere, as well as from any potential plumes that may exist, by detecting ion cyclotron waves at characteristic frequencies. As our ability to glean more and more information about planetary bodies grows along with sophisticated modeling of their magnetic environments, so does the pressing need for more aggressively miniaturized, high sensitive and robust magnetometers. This is the motivation for the work described herein.
Previous Work
Even though an extensive set of magnetometers have been developed over the past few decades10,11, only a subset are suitable for measuring DC magnetic fields in space and around other planetary bodies12,13. Of this limited set, the most widely utilized in space missions are fluxgates and optically pumped atomic gas. Solid-state based magnetometers, such as Hall and magnetoresistive based sensors, have only recently been seriously considered for space missions due to the relative infancy of the technology and the uncertainty of performance in the harsh environments of space. However, they are gaining significant interest as the semiconductor industry is continually making large strides to improve material systems that are resilient to the highly varying temperatures and radiation environments typically encountered in space and other planetary systems.
Fluxgate magnetometers are most often used for near-zero DC magnetic field sensing in space applications where simplicity and cost are of the highest concern. A fluxgate magnetometer is composed of a ferromagnetic core wrapped with three sets of coils. The first set of coils is used to drive the core into and out of saturation, and the second set is used to sense the induced effect. Because the response of the ferromagnetic core is nonlinear, the induced signal will be rich in harmonics which can be used as a reference to null the external field with a third set of coils. Fluxgate magnetometers have flown on an extensive list of missions; some of those worth noting include the two Voyager missions on the Planetary Grand Tour encompassing Jupiter, Saturn, Uranus, Neptune, Pluto, now venturing into interstellar space, the Ulysses mission to study the Sun14, the Cassini mission to Saturn15, the MESSENGER mission to Mercury16, the BepiColombo Mercury Planetary Orbiter17, the MAVEN mission to Mars18, and the Juno mission to Jupiter which arrived on July 4th 201619. While not the most sensitive option for space bound magnetometers, fluxgate sensors are moderately stable and can exhibit high sensitivity . Sigma-delta ADC technology with improved sampling resolution has been leveraged to enhance sensitivity of this technology20,21. Even though significant progress has been made to miniaturize these magnetometers22, the incorporation of a magnetically susceptible core wrapped with driving, sensing, and nulling coils for each axis will always limit the size and weight, and will never lead to a purely electrical device.
For applications where higher sensitivities and better stability are required, optically pumped He magnetometers are the most common choice. These magnetometers have flown on many missions including the Pioneer 10–11 missions to Jupiter and Saturn23, Ulysses14, Cassini15,24, SWARM25, and Juno19. They will also be flown on the upcoming NASA mission to Europa8. While being the most sensitive and stable magnetometers used in space, the optically pumped helium magnetometers are more complex as their scalar based design (can also be configured in vectorized mode) requires two sources of RF excitation, an optical pumping source, and an optical sensor. The first RF source excites the He atoms from their ground state 11S0 to a metastable state 23S1. This transition forces the electrons within the atoms from their singlet state (net spin angular momentum of S = 0, ms = 0) to a triplet state (S = 1, ms = −1, 0, +1) which makes the system electron paramagnetic resonance (EPR) active. Optical pumping of the metastable He gas using right-hand circularly polarized light with λ = 1083 nm provides a source of angular momentum mJ = −1 to depopulate the ms = +1 state into the 23P0 excited state which results in a system with net magnetization due to the remaining electrons in the ms = −1 state. (Note that left-hand circularly polarized light with angular momentum of mJ = +1 could also be used to depopulate the mJ = −1 state.) After some time, the majority of the photons from the optical source will completely pass through the gas because absorption of the ms = −1, 0 states is forbidden due to conservation of angular momentum. An optical detector on the opposite side of the He gas monitors the optical pumping efficiency and controls the frequency v of a second RF source which maintains the EPR condition of the electrons in the ms = −1 state. The resonance condition is defined by hv = Δms gμBB, where h is Planck’s constant, Δms is the difference in spin angular momentum quantum numbers of the spin states, g is the electron’s g-factor, μB is the Bohr magneton, and B is the external magnetic field to be measured. When satisfied, the electron spins are able to flip which eventually equally populates the metastable states thereby destroying the magnetization. This allows for an increase in optical pumping efficiency to be measured by the optical detector. The frequency v is therefore a direct measure of the external magnetic field in which the He gas is immersed. (A more detailed analysis of this magnetometer can be found in refs 23 and 24). These magnetometers achieve the highest sensitivity and stability for space applications at the cost of using high frequency RF and optical circuitry. These components not only add cost, size, and complexity to the instrument, but they also require stable temperatures for operation.
Solid-state based magnetometers alternatively have very limited history in space. They fall into three main categories: Hall based sensors, magnetoresistance based sensors, and the new and emerging semiconductor defect magnetic resonance based approaches. Hall sensors function by measuring a voltage across an appropriately biased, precisely doped semiconductor. Hall sensors are not ideal for sensitive magnetic field measurements in space because they have a sensitivity proportional to magnetic field strength, making them ill-suited for near-zero field sensing. They are also prone to temperature drift as well as prone to radiation damage as most are typically made from silicon. However, SiC based Hall sensors have previously been developed for higher magnetic field sensing applications26. Magnetoresistance based devices usually involve magnetic field induced changes in resistance of complex layered ferromagnetic structures, some of which include ordinary (OMR), anisotropic (AMR), giant (GMR), and tunneling magnetoresistance (TMR). Although many forms of magnetoresistive sensors exist, only a few have been considered for space missions, those mainly being AMR27,28,29,30,31,32 and also TMR33. Though these sensors have shown much progress in the past few years (e.g. noise floor for AMR sensor28), there still remains reliability concerns with their overall stability, temperature stability, and radiation hardness. Additionally, even though these types of sensors can be made extremely small, because most of their responses are orientation dependent, multiple sensors are usually required for simultaneous measurement of three axes which can add complexity. The solid-state magnetic resonance based approaches, including EPR34, electrically detected magnetic resonance (EDMR)35,36,37 and optically detected magnetic resonance (ODMR)38,39,40,41,42,43, are based on detecting the resonances of unpaired electrons tied up in atomic scale defects within the semiconductor. Some of the more interesting work entail those that sense defects within the SiC material system as its wide bandgap allows for operation in high temperature and high radiation environments34,37,40,41,42,43. Even though a wide bandgap material is usually termed “rad-hard”, SiC still is prone to point defect creation in very high radiation environments. As a result, the tuned resonance condition of these magnetometers may be altered due to the creation of defects with g tensor components different than those intended to be sensed, thereby changing the anticipated response making the measurement unreliable. These approaches also rely on high frequency optical and radio circuitry which adds complexity, cost, and reliability concerns. As a result, a fully reliable solid-state magnetometer for space applications remains elusive to this day. The work outlined in this study is based on the latter two forms of solid-state magnetometers: it leverages atomic scale defects to sense magnetic fields; however, it measures the associated magneto response that allows for field measurements without any high frequency circuitry or optical components which simplifies the measurement.
Magnetometer Description
The technique described herein involves the detection of magnetic fields using the recently demonstrated zero-field spin dependent recombination (ZFSDR) in SiC electronics37. The ZFSDR phenomenon provides these solid-state devices with magnetoresistive properties which makes them ideal for the development of miniaturized and purely electrical based magnetometers for near-zero magnetic field sensing (|B| < 10 uT). Similar to fluxgates, the proposed SiC magnetometer (SiCMag) is inexpensive, relatively simple to implement, and does not include high frequency RF or optical components which are sensitive to slight changes in temperature. Additionally, because only a single microelectronic device (with sensing area of less than 0.01 mm2) is required for simultaneous measurement of three magnetic axes, not only is the technology applicable to large scale missions, but it is also more adaptable to smaller missions which leverage nano- and picosats where fluxgate and optically pumped based designs are too large for implementation. (These smaller satellites are categorized as having a mass in the range 0.1 to 10 kg). Although this defect related magneto-response has been observed in Si based electronics44 and organic based semiconductors45,46,47, we focus on SiC based sensors, as its semiconductor properties are significantly more robust than its counterparts. Although the SDR phenomenon can be observed at temperatures as low as −260 °C, we do not plan on operating the magnetometer at this temperature. Like most magnetometers flown in space, the sensor will be housed in a small enclosure which will be space heated, likely no warmer than −120 °C (similar to the lower limit of some fluxgate magnetometers). Additionally, although the SiC semiconductor can operate reliably in temperatures to those encountered on the Venusian surface (≈460 °C), the magnetometer’s maximum operating temperature limit will likely be determined by the robustness of the supporting SiC electronics that will be eventually integrated. Current work in this area shows much promise as SiC ICs were demonstrated to reliably operate at 500 °C for more than 1000 hours48.
SiC Sensor
At the heart of the proposed magnetometer lies a 4H-SiC pn junction. The junction’s sensitivity to magnetic fields arises from intrinsic, deep level, atomic scale defects that play a dominating role in SDR. When a device is biased to yield a recombination dominated current, semiconductor conduction electrons couple with electrons associated with deep level defects. These intermediate coupled pair states can be described by the singlet-triplet basis: the symmetric triplet states, T+ = |↑↑〉, T0 = (|↑↓〉 + |↓↑〉)/, and T− = |↓↓〉, each having spin angular momentum S = 1 with ms = +1, 0, −1 respectively or the anti-symmetric singlet state, S0 = (|↑↓〉 + |↓↑〉)/ which has spin angular momentum S = 0 with ms = 0. Because recombination conserves angular momentum, capture will only occur for singlet pairs whereas triplet pairs will dissociate after a given amount of time. The process is completed by subsequent electron-hole annihilation. (The order of electron/hole capture may of course be reversed).
In the absence of a magnetic field, the ratio of singlet to triplet pairings within the SiC will maintain a certain rate of recombination. However, the singlet to triplet population ratio can be altered by application of an external magnetic field which can be detected as a change in diode current. To understand how this occurs, first consider the Hamiltonian of the two electron spin system in the presence of neighboring magnetic nuclei immersed in a small external magnetic field directed in the z direction, i.e. ,
Here, g is a defect specific and (though not necessarily) an orientation dependent constant usually around 2 (assumed here to be the same for defect and charge carrier), μB is the Bohr magneton, S1 and S2 are the spin angular momentum operators of the two spins, Ij are the nuclear spin angular moment operators for the N neighboring nuclei, Ai,j the nuclear hyperfine parameters of electron i with nucleus j, J0 is the isotropic exchange constant, and D are the dipolar parameters. The magnetic isotopes in the SiC system are the host atoms of 29Si (I = 1/2, 4.7% abundant) and 13C (I = 1/2, 1.1% abundant) as well as the p-type dopant atom of 27Al (I = 5/2, 100% abundant). It is assumed for brevity that the hyperfine tensor is axially symmetric, and therefore has diagonalized elements Ai,x, Ai,y, and Ai,z for i = 1, 2. Consider the simplified case of a dangling bond deep level defect coupled to a 29Si atom within a forward biased SiC pn junction as illustrated in Fig. 1a. Although exaggerated in the figure, note that each electron will experience a slightly different local field BL,1 and BL,2 due to the differences in position of the two electrons surrounding the magnetic 29Si nucleus. The local field is simply the vector sum of the external field B and the nuclear field present at each of the spin sites, BN,1 and BN,2. In general, the nuclear fields are related to the hyperfine parameters by . Table 1 illustrates the Hamiltonian matrix elements of the singlet triplet pair in the presence of a single nucleus (N = 1). Here, the ↑ and ↓ arrows indicate the spin orientation of the spin ½ nucleus. ΣAx, ΣAy, and ΣAz are the sum of the hyperfine parameters at the two spin sites and ΔAx, ΔAy, and ΔAz are their differences. Also, D1 = 3Dz/2 and D2 = (Dx − Dy) which naturally evolve from the axial symmetric dipolar matrix D with diagonal values of Dx, Dy, and Dz. As illustrated in the table, the ΔAx and ΔAy hyperfine components mix the singlet states with the T+ and T− states while the ΔAz component mixes the singlet state with the T0 state.
Table 1. Hamiltonian matrix elements of singlet triplet pairs in the presence of a I = 1/2 nucleus.
|T+, ↑〉 | |T0, ↑〉 | |S0, ↑〉 | |T−, ↑〉 | |T+, ↓〉 | |T0, ↓〉 | |S0, ↓〉 | |T−, ↓〉 | |
---|---|---|---|---|---|---|---|---|
〈↑, T+| | 0 | 0 | D2 | 0 | 0 | |||
〈↑,T0| | 0 | ΔAz/4 | 0 | 0 | 0 | |||
〈↑, S0| | 0 | ΔAz/4 | 0 | 0 | 0 | |||
〈↑, T−| | D2 | 0 | 0 | 0 | 0 | |||
〈↓, T+| | 0 | 0 | 0 | 0 | D2 | |||
〈↓, T0| | 0 | 0 | 0 | 0 | ||||
〈↓, S0| | 0 | 0 | 0 | 0 | ||||
〈↓, T−| | 0 | 0 | D2 | 0 | 0 |
The low-field hyperfine mixing alters the ratio of singlet to triplet pairings thereby allowing changes in recombination current to be measured for changes in external magnetic field. However, because the mixing coefficients are small, only small changes in current (up to a few percent) can be measured in the response. Therefore, frequency and phase synchronous detection is achieved via magnetic field modulation to enhance the sensitivity and signal-to-noise ratio of the measurement. This method restricts the measurement to the axis of modulation and allows for a vectorized measurement to be made. As a result, the responses are typically reported as derivatives which exhibit a sharp zero crossing that is approximately linear through zero magnetic field. As the magnetic field is increased, the mixing is suppressed by the Zeeman energy splitting. However, by exposing the spin system to electromagnetic (EM) radiation with energy equal to the splitting of the defect states, one is also able to measure a change in recombination current due to the modified populations of singlet and triplet states that occurs. This is more commonly known as EDMR. The EM radiation provides a source of angular momentum to the spin pair, thereby allowing the electrons to change state (or flip spin) if angular momentum is conserved and the resonance condition is satisfied, that being when . (Note that resonance also occurs at magnetic fields corresponding to B ± BN for a nucleus with I = 1/2 such as 29Si, 13C, or 1H) Essentially, triplet pairs are able to be converted to singlet pairs, thereby increasing the rate of recombination current and hence, depending on the circumstances, either an increase or decrease in resistance that can be detected electrically49. Figure 1b illustrates the SDR response, acquired via high-field EDMR (B = 340.3 mT, v = 9.54 GHz), of the device used in our study. The measurement reveals that the dominant defect is a silicon vacancy variant. (A previous EDMR study acquired from a nearly identical device directly linked the dominant defect spectrum to that of a negatively charged silicon vacancy50). With the crystalline c-axis of the SiC aligned with the external magnetic field, the acquired spectrum consists of a very strong dominating central line (96% of total spectrum), with two very small, almost negligible, sets of side peaks (1.5% and 0.5% of the central line) due to the 1.1% magnetic abundance of 13C. The strong electron-nuclear hyperfine side peaks that are apparent in the spectrum can be attributed to the, almost 100% magnetic abundant, spin mI = ±1/2 1H atom (1.1 mT splitting) which is present near the SiC/SiO2 interface. The defect is isotropic as the acquired spectra are nearly identical when the crystalline c-axis is rotated about three axes. Therefore, as illustrated in Fig. 1b, the response can simply be modeled with a strong dominating central line (78% of the total spectrum) with two side peaks (each contributing 11% to the defect spectrum) equally spaced 0.55 mT from the central line. Note that all three peaks in the model are generated from a Gaussian distribution with standard deviation σ = 0.05 mT. The peaks are summed and the derivative is taken to mimic the effect of magnetic field modulation. As illustrated from the figure, there is very good agreement between the model and the EDMR data.
Figure 1c illustrates the comparison of the high-field EDMR and the ZFSDR magnetoresistive responses acquired from the same device biased with the same potential. Even though the ZFSDR spectrum appears broader, it has a very similar zero-crossing slope compared to that of the high-field response. This makes the ZFSDR response an excellent technique for sensing near-zero magnetic fields (|B| < 0.1 mT). Remarkably, the hyperfine side peaks are retained in this simple measurement. By inspection of the eigenvalues of the Hamiltonian matrix evaluated at different magnetic fields using the isotropic defect hyperfine parameters obtained in the high-field EDMR model (A1,(x,y,z)mI/gμB = 0.55 mT) and that of a conduction electron located twice as far away (A2,(x,y,z)mI/gμB = 0.55/23 mT), it is apparent from Fig. 1c that not only do the singlet and triplet pairs mix precisely at zero magnetic field (between the |T+, ↓〉 and |S0, ↑〉 states), but there is also a singlet-triplet intersystem crossing at B = ±0.825 mT which exactly corresponds to the location of the electron-nuclear hyperfine peaks observed in the ZFSDR spectrum. Because the hyperfine interactions are isotropic, it is precisely the mixing between the |T−, ↑〉 and |S0, ↓〉 states which allows the satellite peaks to be measured due to the preservation of the ΔAx + ΔAy matrix elements. (Note that the ΔAx − ΔAy and ΔAy − ΔAx elements cancel for isotropic defects). In general however, depending upon the strength and number of the hyperfine interactions, multiple satellite peaks may be detected in the ZFSDR response at fields relating to the hyperfine parameters themselves. This is clearly evident in more complicated defects as those reported in 4H-SiC BJTs37. Also, depending on the proximity of the two electron spins, dipolar and or exchange interactions may cause a split peak to present at zero magnetic field due to the singlet-triplet degeneracy removal that occurs from this interaction. These features are depicted in some of the spectra illustrated in latter sections of this report.
Sensitivity
Because the magnetometer relies on a magnetic field modulation scheme, the sensitivity of the instrument can be defined by the zero crossing slope of the response and the noise retained in the bandwidth of interest of the modulation frequency used. (A review of magnetic field modulation as it relates to EDMR and intersystem crossing can be found in the work by Lee51). As we are only interested in the local derivative at the zero crossing center line, a Gaussian line shape can be assumed,
Here, σ is the width (standard deviation) of the underlying response in units of Tesla, ΔI is the change in current in Amps, and B is the magnetic field in Tesla where the current is maximum at B = 0nT. Magnetic field modulation will induce a series of harmonics in the measured response. The fundamental will have an envelope that is proportional to the responses derivative if the modulation amplitude Bm is relatively small. This concept is illustrated in Fig. 2a. The measured data resulting from demodulating the fundamental is,
The uncertainty, or sensitivity, in the magnetic field δB is related to the uncertainty in the measured current response δId by,
Additionally, the field modulation scheme allows for a measurement to be made that is shot noise limited (far from the flicker noise elbow). Therefore, the uncertainty in current response δId is related to shot noise by, , where q is the electronic charge, I0 is the DC current which is responsible for the flicker noise, and Δf is the bandwidth of the measurement which is related to the time constant τ of the low pass filter that follows the mixer in the demodulator, Δf = 1/2πτ. Setting eq. 4 equal to the shot noise and assuming Bm = σ, rearranging yields the sensitivity of the instrument,
The same relation was used to calculate sensitivity limits by Baker et al. for an EDMR based magnetometer36. The change in current ΔI at B = 0 mT was extracted from the measured data by numerical integration. (Integration was performed I(B[i]) = Id(B[i])/(FsBm) + I(B[i − 1]) for i = 1:M, where i is the index of the array, M is the number of samples in the spectrum, and Fs is the sample rate in units of samples per unit magnetic field).
Figure 2b illustrates the raw spectra acquired at various modulation amplitudes with a forward junction bias of 2.3 V. The peak-to-peak amplitude of the measured data grows with increasing modulation amplitude as it represents dI(B) rather than dI(B)/dB. The relative sharpness of the spectrum is attributed to the presence of a dominant defect, a silicon vacancy variant, within the ordered crystalline environment of the SiC semiconductor. The magnetoresistive response within the 4H-SiC lattice will therefore naturally have a sharper response (<0.3 mT in this case) than that observed in amorphous materials such as organic semiconductors and/or dielectric tunneling junctions52. As described earlier, the crystalline nature of the semiconductor allows for the detection of the electron-nuclear hyperfine interactions which are depicted by the inflections spaced 1.65 mT apart, symmetric about zero magnetic field. Additionally, as described earlier, the narrow notch located precisely at zero magnetic field (highlighted in the inset of the panel) is attributed to spin-spin interactions of the defect electron and can only be resolved when leveraging a low amplitude modulation waveform and an elevated junction bias. Its presence can therefore be turned on or off and leveraged for self-calibration which is described in more detail later in this report. Figure 2c illustrates the integrated responses of the spectra depicted in Fig. 3b, each offset by 10 pA for clarity. Note that, as expected, the change in current ΔI at B = 0 mT in the integrated spectra are all the same for different modulation amplitudes.
The junction bias was varied to find the optimal tradeoff between signal amplitude and device noise, and hence, sensitivity. Figure 3a illustrates a series of measurements made over a 1 Hz (τ = 1/2π) bandwidth at different junction biases. As illustrated in panels 3b and 3c, the maximum change in current (ΔI/I = 0.37%) was extracted from the integrated data to yield an approximate sensitivity of when a forward bias of 2.35 V was applied to the junction. Although the sensitivity reported here makes for a good magnetic field sensor on the surface of Earth (BEarth ≈ 30,000 nT), its sensitivity is not ideal in its current state for planets with small magnetic fields. However, the sensor has great potential and a straightforward path toward substantial sensitivity improvement. The steps outlined to increase the sensitivity to a level competitive with state of the art planetary sensors are later described in detail.
Vectorized Measurement
The proposed magnetometer is based on a magnetic field cancellation scheme that maximizes the ZFSDR current in the pn junction by maintaining a local region of zero-magnetic field across the volume of the device. The device is housed within three sets of Helmholtz coils (one for each dimension) that are driven independently to provide a low-frequency (<10 Hz) cancellation field and a modulation field at audio frequencies. As the low-frequency driving current in these Helmholtz coils is directly proportional to the magnetic field it generates, its measure will serve as an indirect measure of the field being cancelled in each dimension. The block diagram of the magnetometer is illustrated in Fig. 4a and a picture of our 3D printed prototype of the coil system is illustrated in Fig. 4b.
The magnetic field induced change in SDR current of the SiC sensor is first fed through a high gain transimpedance amplifier with AC coupling before it is digitized for optimum sensitivity. (The amplified DC current will also be digitized in order to extract a crude measure of temperature. This will allow the instrument to optimally bias the SiC sensor for maximum recombination for any changes in temperature that may occur). The conditioned AC signal is then digitized and fed into three independent (one for each axis), user-configurable digital demodulators operating at different frequency bands. This allows for the three vectorized magnetic field components to be frequency division multiplexed onto a single channel. The demodulator involves an optional first stage of digital bandpass filtering, and then is followed by a mixer that performs a point-by-point multiplication of the incoming signal with a time-synchronized, user-configured (demodulating harmonic and demodulating phase) sinusoid. This operation allows one to measure the signal strength at any particular harmonic of the fundamental modulation frequency. A low-pass filter is used to remove the high-frequency mixing artifact, and an exponential averaging block is used to further reduce noise on the remaining signal based upon the user-configured bandwidth. A digital-based PI controller is implemented for each of the three demodulators for tracking the zero crossing center line of each current component. The error output of each controller is added to its corresponding modulation waveform in software prior to digital-to-analog conversion. This error is proportional to the magnetic field in each axis, and hence, is the current required to be passed through each Helmholtz coil in order to maintain a local region of zero magnetic field. In order to test our sensitivity metric, we recorded the current for each axis in the presence of an alternating axis, square wave magnetic field of amplitude of ±1500 nT within a mu-metal shielded chamber. As illustrated in Fig. 4c, the sensor can easily and simultaneously measure these weak magnetic field signatures using a single sensor.
Self-calibration
One of the biggest concerns with magnetometers, and scientific instruments on board satellites and landers in general, is remote calibration. SiCMag has the ability to self-calibrate, a significant advantage. SiCMag has two forms of self-calibration, both of which are illustrated in Fig. 5. The first and simpler self-calibration method entails leveraging the electron spin interactions that are observed in the ZFSDR spectrum. Figure 5a illustrates the 1st and 2nd harmonic spectra which clearly demonstrate the detection of the electron spin interactions discussed earlier. Because these interactions are virtually independent of temperature and also tremendously stable over long periods of time, their spacings will be maintained with extraordinary stability. (A previous work has demonstrated that the spacing of the narrow split peak observed in organic semiconductors remains constant from room temperature all the way down to 10 K46). By measuring the spacing of these magnetic field indicators as a function of the applied coil current, one is able to extract a constant-of-proportionality measure that can be used to calibrate the coil-driving system. This self-calibration mode can be turned on or off with the bias junction voltage and/or modulation amplitude as was shown earlier. The modulation amplitude essentially controls the minimum detectable feature size while the bias voltage not only controls sensitivity, but also controls the strength of the coupling between the two electrons spins. It should be noted that each axis will need to be calibrated independently to account for not only defect anisotropy but also due to differences in coil size and non-ideal characteristics of the electronics that drive each axis.
The second method of calibration entails performing low-field EDMR35,36,37,52 on the same device as illustrated in Fig. 5b. This method is appealing because the measurement itself becomes a form of absolute magnetometry, and can be used to validate the measured ZFSDR response. The method can be accommodated by incorporating a miniature RF excitation source which drives a small loop coil or strip line resonator on the PCB to provide the oscillating magnetic field required to drive the spin system into resonance. Although this adds a bit of complexity, the additional magnetic field measurement provides invaluable redundancy in the remote environment of space. It should be noted that this method shouldn’t be used when operating in a high energy radiation environment for prolonged periods of time. If the radiation induces a significant number of defects (with different g factors) due to prolonged exposure, the EPR condition of the defect electron outlined earlier may change. This adds uncertainty to the precise field/frequency relationship of the EPR condition, thereby making the measurement unreliable. However, the wide bandgap of the SiC semiconductor should ensure that measurement altering defects are not created, or at least not in an abundance that overwhelms the original defect spectrum. It should be noted that even if different defects are created in these harsh environments, it will not affect the magnetoresistive response associated with the device because the ZFSDR response is insensitive to defects with different g values. Thus, the accuracy of the measurement is preserved. However, the linewidth may be slightly compromised which results in a slight loss in sensitivity. Radiation tests will need to be performed in order to further quantify these cases.
Future Work and Summary
Although SiCMag, in its present form, is characterized as having a sensitivity, there are many ways to improve performance. First and foremost, we leveraged the pn junction within an experimental microelectronic device designed for high power applications, which was not optimized in any way for magnetometry. The next step in this work will involve developing a custom device designed to exploit the ZFSDR recombination phenomenon. Tradeoffs between geometry, sensing area, doping, and annealing will be made to find the optimum line shape and response. Once processed and fabricated, the sensitivity can be significantly enhanced by creating stable silicon vacancy defects by exposing the devices to high energy electron radiation as was done for optical magnetometry methods in SiC40,41,42,43. These defect engineering methods will help increase ΔI and improve sensitivity. Also, eventual device fabrication using isotopically pure SiC, that is SiC in which the crystal is depleted of 29Si or 13C, will sharpen the linewidth σ, which will push the sensitivity to its limits. Even though the hyperfine interactions are responsible for the magneto-response, we feel that the 1.1% magnetically abundant 13C and 4.7% magnetically abundant 29Si (and magnetic dopant atoms) contribute to the breadth of the signal and limit the ultimate sensitivity. Using an isotopically purer material will allows us to better control point defect creation and limit the amount of random electron nuclear hyperfine interactions, thereby sharpening the magneto-response. And finally, improved electronics which leverage low-noise ICs, layout practices, and sigma-delta ADC sampling should improve SNR, sensitivity, and stability of the measurement. Figure 6 illustrates the extrapolated sensitivity, given by eq. 5, as a function of percent change in current (100% xΔI/I0) for devices with linewidths of σ = 0.2 mT, 0.1 mT, 0.05 mT, and 0.025 mT and I0 = 1 mA DC current for all four cases. As illustrated in the figure, the is easily attainable using the plan outlined above; however, it is still unclear if the sensitivity can be pushed into the regime. We also plan on experimenting with diamond and GaN based semiconductor devices as they also have much promise in this application due to their relatively large bandgaps and crystalline structure. Diamond may eventually show to have the optimum response due to its sharp response attributed to the small number of hyperfine interactions from the 13C atoms.
The proposed magnetometer technology has enormous potential for the future of NASA and its space missions. As the magnetometer can be made extremely small, it has the potential to be manufactured into a microelectromechanical system (MEMS) device. Additionally, the magnetometer has the potential application to operate in high temperature and high radiation environments due to the wide bandgap and robustness of the SiC semiconductor. The magnetometer does not require any high frequency circuitry, thus reduces risk and eliminates the need to maintain strict temperatures for accurate and reliable magnetic field measurements. The combination of these features, along with being purely electrical and inexpensive, enables the technology to be used for a variety of magnetic field sensing applications, including planetary entry probes, landers, missions in extreme environments such as Venus and Jupiter, and in swarms of spacecraft significantly smaller than current nanosats.
Methods
In this study, we use the source/drain to substrate pn junction of a 4H-SiC n-channel lateral MOSFET. The device was fabricated on a p-type substrate which was Al doped to about 6 × 1015. The source and drain n-wells were created by P implantation. The device has a channel area (L × W) of 400 × 400 μm2 and the gate oxide is a 50 nm thermally grown ONO stack (10 nm SiO2/30 nm Si3N4/10 nm SiO2) which received a 5 minute Ar anneal at 1600 °C.
We leveraged a custom designed preamp for amplification and signal conditioning. A NI USB-6289 M Series DAQ card was used to digitize the amplified device current and also used to generate three modulation waveforms with configurable offset, frequency, and phase. These signals were then fed into custom designed analog buffers which drove each of the three coils independently. LabVIEW software run on a Dell Inspiron laptop with 6 GB RAM, was used for signal processing and waveform generation. All measurements were made at room temperature. The low-field spectra were acquired on a custom built low-field EDMR spectrometer while the vectorized measurement illustrated in Fig. 4 were made in a mu-metal shielded chamber with three axis Helmholtz coil magnetic field generating source located at the Jet Propulsion Laboratory.
Additional Information
How to cite this article: Cochrane, C. J. et al. Vectorized magnetometer for space applications using electrical readout of atomic scale defects in silicon carbide. Sci. Rep. 6, 37077; doi: 10.1038/srep37077 (2016).
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Acknowledgments
A portion of this research was supported by an appointment to the NASA Postdoctoral Program at the Jet Propulsion Laboratory, administered by Oak Ridge Associated Universities through a contract with NASA. The research described in this publication was carried out at the Jet Propulsion Laboratory, California Institute of Technology, under a contract with the National Aeronautics and Space Administration (NASA) under Center Innovation Funding (CIF) and Planetary Instrument Concepts for the Advancement of Solar System Observations (PICASSO). We would like to thank Gerardo Hernandez from University of California Riverside for the design and construction of the 3D printed magnet, Neil Murphy of JPL for his help and assistance with JPL magnetometer facilities, Aivars Lelis of the Army research lab for contributing the SiC device for study and helpful discussions about defects and device performance, and Sang-Yun Lee of Stuttgart University for helpful discussion about defects in SiC.
Footnotes
Author Contributions C.C. conceived the idea, built the instrument and acquired data, wrote the main text of the manuscript, and prepared all of the figures. J.B. assited in writing the manuscript and was involved in discussions of the results. M.A. and P.L. acquired the EDMR data and were involved in discussion of the results. All authors reviewed, provided comments, and approved the final manuscript.
References
- National Academy of Sciences/National Research Council (NRC) Decadal Survey of Planetary Sciences: Visions and Voyages for Planetary Science in the Decade 2013–2022 (2011).
- Kivelson M. G. et al. Europa’s Magnetic Signature: Report from Galileo’s Pass on 19 December 1996. Science 276, 1239–1241 (1997). [DOI] [PubMed] [Google Scholar]
- Khurana K. K. et al. Induced magnetic fields as evidence for subsurface oceans in Europa and Callisto. Nature 395, 777–780 (1998). [DOI] [PubMed] [Google Scholar]
- Kivelson M. G. et al. Europa and Callisto: Induced or intrinsic fields in a periodically varying plasma. J Geophysical Research 104, A3, 4609–4625 (1999). [Google Scholar]
- Dougherty M. K. et al. Cassini magnetometer observations during Saturn orbit insertion. Science 307, 1266–1270 (2005). [DOI] [PubMed] [Google Scholar]
- Dougherty M. K. et al. Identification of a dynamo atmosphere at Enceladus with the Cassini magnetometer. Science 311, 1406–1409 (2006). [DOI] [PubMed] [Google Scholar]
- Simon S., Saur J., Treeck S. C., Kriegel H. & Dougherty M. K. Discontinuities in the magnetic field near Enceladus. Geophysical Research Letters 41, 3359–3366 (2014). [Google Scholar]
- Pappalardo R. Europa Mission Science Overview. Paper presented at the Outer Planets Assessment Group, Laurel, MD USA. 2015, August 24–26. http://www.lpi.usra.edu/opag/meetings/aug2015/presentations/day-1/8_a_Pappalardo.pdf.
- Raymond C. ICEMAG: Interior Characterization of Europa Using Magnetometry. Paper presented at the Outer Planets Assessment Group, Laurel, MD USA. 2015, August 24–26. http://www.lpi.usra.edu/opag/meetings/aug2015/presentations/day-1/8_g_raymond.pdf.
- Lenz J. & Edelstein A. S. Magnetic Sensors and their applications. IEEE Sensors Journal 6(3), 631–649 (2006). [Google Scholar]
- Robbes D. Highly sensitive magnetometers – a review. Sensors and Actuators A 129, 86–93 (2006). [Google Scholar]
- Diaz-Michelena M. Small magnetic sensors for space applications. Sensors 9(4), 2271–2288 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Acuna M. H. Space-based magnetometers. Review of scientific instruments 73, 3717 (2002). [Google Scholar]
- Balogh A. et al. The magnetic field investigation on the Ulysses mission – instrumentation and preliminary scientific results. Astronomy and Astrophysics Supplemental Series 92(2), 221–236 (1992). [Google Scholar]
- Dunlop M. W., Dougherty M. K., Kellock S. & Southwood D. J. Operation of the dual magnetometer on Casini: science performance. Planetary and Space Science 47(10–11), 1389–1405 (1999). [Google Scholar]
- Anderson B. J. et al. The magnetometer instrument on MESSENGER. Space Science Reviews 131(1–4), 417–450 (2007). [Google Scholar]
- Glassmeier K. H. et al. The fluxgate magnetometer of the BepiColombo Mercury Planetary Orbiter. Planetary and Space Science 58, 287–299 (2010). [Google Scholar]
- Connerney J. E. P. et al. The MAVEN magnetic field investigation. Space Science Reviews 195, 257–291 (2015). [Google Scholar]
- Matousek S. The Juno new frontiers mission. Actga Astronautica 61(10), 932–939 (2007). [Google Scholar]
- Magnes W. et al. A sigma-delta fluxgate magnetometer for space applications. Measurement Science and Technology 14, 1003–1012 (2003). [Google Scholar]
- Iguchi K. & Matsuoka A. A digital-type fluxgate magnetometer using a sigma-delta digital-to-analog converter for a sounding rocket experiment. Measurement Science and Technology 25, 075803 (2014). [Google Scholar]
- Lu J. H., Chiu P., Chiu S. & Jeng J. High-sensitivity low-noise miniature fluxgate magnetometers using a flip chip conceptual design. Sensors 14, 13815–13829 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith E. J., Connor B. V. & Foster G. T. Measuring the magnetic fields of Jupiter and the outer solar system. IEEE Trans Magnetics Mag-11(4), 962–980 (1975). [Google Scholar]
- Smith E. J., Dougherty M. K., Russell C. T. & Southwood D. J. Scalar helium magnetometer observations at Cassini Earth swing-by. J Geophysical Research 106, A12, 30, 129–130, 139 (2001). [Google Scholar]
- Fratter I. et al. Swarm Absolute Scale Magnetometers first in-orbit results. Acta Astronautica 121, 76–87 (2016). [Google Scholar]
- Robert J.-L. et al. 4H-SiC: a material for high temperature Hall sensor. Sensors and Actuators A: Physical 97–98, 27–32 (2002). [Google Scholar]
- Brown P. et al. Magnetoresistive magnetometer for space science applications. Measurement Science and Technology 23, 025902 (2012). [Google Scholar]
- Brown P. et al. Space magnetometer based on an anisotropic magnetoresistive hybrid sensor. Rev Scientific Instruments 85, 125117 (2014). [DOI] [PubMed] [Google Scholar]
- Leitner S., Valavanoglou A., Brown P. & Hagen C. Design of the Magnetoresistive Magnetometer for ESA’s SOSMAG Project. IEEE Trans Magnetics 51, 1 (2015). [Google Scholar]
- Sordo-Ibáñez S. et al. A front-end ASIC for a 3-D magnetometer for space applications by using anisotropic magnetoresistors. IEEE Trans Magnetics 51, 1 (2015). [Google Scholar]
- Díaz-Michelena M., Sanz R., Cerdán M. F. & Fernández A. B. Calibration of QM_MOURA three-axis magnetometer and gradiometer. Geoscientific Instrumentation Methods and Data Systems 4, 1–8 (2015). [Google Scholar]
- Archer M. O. et al. The MAGIC of CINEMA: first in-flight science results from a miniaturized anisotropic magnetoresistive sensor. Annales Geophsicae 33(6), 725–735 (2015). [Google Scholar]
- Persson A., Nguyen H., Riddar F. & Thornell G. Micro and nanostructured magnetic field sensor for space applications. Paper presented at the Actuators and Microsystems Conference, Denver, CO USA. Proceedings of Solid-state sensors, pp: 1190–1193. 2009, June 21–25.
- Lee S.-Y., Niethammer M. & Wrachtrup J. Vector magnetometry based on S=3/2 electronic spins. Phys Rev B 92, 115201 (2015). [Google Scholar]
- Jander A. & Dhagat P. Solid state magnetometer using electrically detected magnetic resonance. J. Magnetism and Magnetic Mat. 332, 1639–1641 (2010). [Google Scholar]
- Baker W. J. et al. Robust absolute magnetometry with organic thin-film devices. Nature Comm. 3, 898 (2102). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cochrane C. J. & Lenahan P. M. Zero-field detection of spin dependent recombination with direct observation of electron nuclear hyperfine interactions in the absence of an oscillating electromagnetic field. J Applied Physics 112, 123714 (2012). [Google Scholar]
- Aman H. & Plakhotnik T. Accuracy in the measurement of magnetic fields using nitrogen-vacancy centers in nanodiamonds. J Optical Society of America B 33, 3 (2016). [Google Scholar]
- Budker D. & Romalis M. Optical Magnetometry. Nature Physics 3, 227–234 (2007). [Google Scholar]
- Kraus H. et al. Magnetic field and temperature sensing with atomic-scale defects in silicon carbide. Nature Sci Reps 4, 5303 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Simin D. et al. High-precision angle-resolved magnetometry with uniaxial quantum centers in silicon carbide. Physical Review Applied 4, 014009 (2015). [Google Scholar]
- Simin D. et al. All-optical DC nanotesla magnetometry using silicon vacancy fine structure in isotopically purified silicon carbide. Physical Review X 6(3), 031014 (2016). [Google Scholar]
- Niethammer M. et al. Vector magnetometry using silicon vacancies in 4H-SiC at ambient conditions. Accepted for publication in Cond-mat.mes-hall, submitted June 3, 2016.
- Cochrane C. J. & Lenahan P. M. Detection of interfacial Pb centers in Si/SiO2 metal-oxide-semiconducting field-effect transistors via zero-field spin dependent recombination with observation of precursor pair spin-spin interactions. Applied Physics Letters 103(5) (2013). [Google Scholar]
- Waters D. P. et al. The spin-Dicke effect in OLED magnetoresistance. Nature Physics 11(11), 910–914 (2015). [Google Scholar]
- Nguyen T. D., Gautam B. R., Ehrenfreund E. & Vardeny Z. V. Magnetoconductance Response in Unipolar and Bipolar Organic Diodes at Ultrasmall Fields. Physical Review Letters 105, 166804 (2010). [DOI] [PubMed] [Google Scholar]
- Kersten S. P., Schellekens A. J., Koopmans B. & Bobbert P. A. Magnetic-Field Dependence of the Electroluminescence of Organic Light-Emitting Diodes: A Competition between Exciton Formation and Spin Mixing. Physical Review Letters 106, 197402 (2011). [DOI] [PubMed] [Google Scholar]
- Spry D. et al. Prolonged 500 °C Demonstration of 4H-SiC JFET ICs With Two-Level Interconnect. IEEE Electron Device Letters 37(5), 625–628 (2016). [Google Scholar]
- Kaplan D., Solomon I., Mott N. F. Explanation of the large spin dependent recombination effect in semiconductors. J Phys Lett (Paris) 39(4), L51–L54 (1978). [Google Scholar]
- Cochrane C. J., Lenahan P. M. & Lelis A. J. Identification of a Si vacancy as an important defect in 4H SiC MOSFETs using spin dependent recombination. Applied Physics Letters 100(2), 023509 (2012). [Google Scholar]
- Lee S. Y., Paik S., McCamey D. R. & Boehme C. Modulation frequency dependence of continuous-wave optically/electrically detected magnetic resonance. Physical Review B 86, 115204 (2012). [Google Scholar]
- Cochrane C. J. & Lenahan P. M. Spin counting in EDMR via low-field defect state mixing. Applied Physics Letters 104(9), 093503 (2014). [Google Scholar]