Does the DVR® plate restore bony anatomy following distal radius fractures?

S Patel 1,, PB Menéndez 1, FS Hossain 1, HB Colaço 1, MH Lee 1, ED Sorene 1, EJ Taylor 1
PMCID: PMC5137656  PMID: 24417831

Abstract

INTRODUCTION

Fractures of the distal radius are common. Malreduced fractures are associated with residual functional deficiency. There has been a trend over the last few years for using fixed angle volar locking plates to surgically stabilise this injury. Our unit uses the DVR® plate (DePuy, Warsaw, IN, US). Nevertheless, it is unknown whether the normal bony anatomy is recreated or merely restored to acceptable limits with its usage. The aim of this study was to evaluate the reduction achieved compared with an uninjured population and pre-existing quoted ‘normal’ values. Furthermore, we wanted to identify the percentage of cases that were reduced to acceptable limits, and determine whether the grade of the surgeon and fracture type was a confounding influence on this reduction.

METHODS

A retrospective review of the 3-month postoperative radiography of 48 eligible patients who underwent open reduction and internal fixation of a distal radius fracture with a DVR® plate was undertaken.

RESULTS

Volar tilt, radial length and inclination were different to quoted normal values (p<0.01). Despite this, these parameters fell within acceptable limits in 46 cases; this was not influenced by fracture type or grade of operating surgeon.

CONCLUSIONS

The DVR® plate restores the bony anatomy to within acceptable limits in the majority of patients who have sustained a fracture of the distal radius although of all parameters investigated, the widest variability is seen in volar tilt.

Keywords: Radius, Internal fracture fixation, Cartilage fractures, Radiology information systems


Fractures of the distal radius are common injuries with an estimated incidence of 71,000 cases in Britain every year.1 Patients of all ages can be affected but the elderly are more susceptible to injury.2 It has long been recognised that malreduced fractures are associated with poor long-term function.3 Consequently, there is an indication to reduce displaced fractures and surgically stabilise those that are unstable. The devices used for maintaining reduction surgically include percutaneous wires, external fixators, intramedullary nails and plates.

The early results of plate fixation were poor but improved dramatically following the introduction of precontoured locking plates. This has been associated with a rise in popularity with respect to usage. Recent randomised controlled trials have demonstrated better function in the early postoperative period with this fixation method than with other methods such as percutaneous pinning, external fixation and radial column plates although long-term function is comparable.4,5

The DVR® plate (DePuy, Warsaw, IN, US) is a volar locking plate that was introduced in 2001. It has been shown to be biomechanically stable and possibly more so than other similar volar locking plates by other manufacturers.68 The DVR® plate has undergone an evolution in design since first coming to the market with two rows of screw options now available (Fig 1) to improve subchondral support. Furthermore, the number of available sizes has been increased to seven so that it can be used in patients of variable bony anatomy and size.

Figure 1.

Figure 1

Standardised three-month postoperative posteroanterior (A) and lateral (B) radiography of a fractured distal radius that has been stabilised with a DVR® plate

The biomechanical merits of plate osteosynthesis, complications and functional outcome are well documented. There is, however, a paucity of evidence as to what the expected radiographic parameters are following this fixation modality. Our institution always performs open reduction and internal fixation rather than closed reduction and percutaneous wire fixation for all adult patients presenting with an unstable fracture of the distal radius within two weeks of the injury. We have used the DVR® plate with two rows of distal screw options since 2006 and we do not commonly image both wrists unless there is a suspicion of congenital abnormal anatomy.

‘Normal’ values for the volar tilt (VT), radial length (RL) and radial inclination (RI) of the distal radius were described originally in 1962 by Scheck as 11º, 12mm and 23º respectively.9 A subsequent pooled analysis of multiple studies has demonstrated the mean values to be 14.5º, 13.5mm and 25.4º respectively10 although it is the former set of values that continue to be quoted widely in the literature. The aims of this study were to: (1) evaluate the reduction at three months following surgery against that at the end of the procedure, against a group of uninjured wrists and against both sets of pre-existing quoted ‘normal’ values to determine whether these can be achieved using the DVR® plate; (2) identify the percentage of cases that were reduced to acceptable limits; and (3) determine whether the grade of surgeon and fracture complexity were a confounding influence on this reduction.

Methods

A retrospective analysis was undertaken of all patients with a fracture of the distal radius between 2008 and 2009 who underwent open reduction and internal fixation with a DVR® plate. We identified 60 cases but excluded those patients whose fractures were stabilised with supplementary fixation (n=2), those lost to follow-up by the evaluation stage (n=3) and those who had a previous distal radius fracture (n=3) or inadequate radiography (n=4).

This left 48 cases in 48 patients (19 male, 29 female, mean age: 51.2 years, age range: 19–85 years). All fracture patterns were described according to the Arbeitsgemeinschaft für Osteosynthesefragen (AO) classification (type A: 13 cases, mean age: 38.3 years, age range: 19–79 years; type B: 9 cases, mean age: 48.2 years, age range: 21–72 years; type C: 26 cases, mean age: 59.1 years, age range: 21–85 years) and all procedures were performed within two weeks of injury by either a consultant (19 cases), a registrar with consultant supervision (4 cases) or a registrar independently (25 cases). The uninjured population with which these patients were compared as part of the analysis consisted of 48 wrists in 46 patients who presented to our clinic with a history of minor trauma to the upper limb and in whom no radiological abnormality was identified.

Operative technique

All cases were operated on in a standardised manner under general anaesthesia and with a tourniquet inflated around the affected limb at 250mmHg for the duration of the operation. The distal radius was approached through the bed of the flexor carpi radialis tendon and the operation performed as per the recommended operative technique. The fracture was identified and reduced under fluoroscopy guidance. The plate was then positioned on the radial shaft, securing it initially using the central sliding hole. The decision to release the brachioradialis tendon to facilitate fracture reduction on to the plate was made by the operating surgeon. All available peg holes were then filled to maintain fracture reduction and prevent redisplacement. After securing the plate to the shaft with the final screws, the wound was closed and dressed in a bulky bandage.

Postoperative protocol

Clinical and radiographic follow-up occurred at 2, 6 and 12 weeks following surgery, and thereafter based on clinical need. For weeks 2–6, patients were given a Futuro® splint (3M, Bracknell, UK) for support that could be removed for exercises with our hand therapists. Discontinuation of the splint was advised routinely at week 6 and further supervised exercised was dependent on residual functional deficit.

Radiographic assessment

Posteroanterior and lateral radiography was used to assess VT, RL and RI, with all values determined by two observers (SP and PBM). Intraoperative radiography was performed and further images were collected at least three months postoperatively to allow for any potential loss of fracture position. The first ten cases of three-month postoperative radiography were used for interobserver variability, showing good agreement (kappa = 0.91).

VT was defined as the angle created between the articular surface of the distal radius and a line perpendicular to the long axis of the radius as witnessed on lateral radiography. RL was defined as the distance between the tip of the radial styloid process and the distal articular surface of the ulna in a direction perpendicular to the long axis of the radius. RI was defined as the angle created between a line joining the tip of the radial styloid and the ulnar corner of the articular surface, and a line perpendicular to the long axis of the radius.

Statistical analysis

Descriptive statistics were applied to describe the basic characteristics of the datasets. A two-sample unpaired t-test was used for comparing the uninjured population with the three-month postoperative radiography, a two-sample paired t-test for comparing the intraoperative radiography with the three-month postoperative radiography and a one-sample t-test for comparing the three-month postoperative radiographic parameters with normal values quoted in the literature to determine whether normal anatomy was recreated.

A one-way analysis of variance and covariance was used to determine whether these values differed depending on the grade and supervision of the operating surgeon, the type of the fracture as defined by the AO classification or the interaction between them. A chi-squared test was used to see whether complications varied according to the grade of the operating surgeon. All analyses were performed using the XLSTAT module (Addinsoft, Paris, France) for Excel® (Microsoft, Redmond, WA, US), with statistical significance set at p<0.05.

Results

Comparison with normal values

Radiographic assessment at three months demonstrated that the mean VT achieved was 8.8º (standard deviation [SD]: 5.5º, range: -6–20º), the mean RL was 11.0mm (SD: 1.7mm, range: 7–15mm) and the mean RI was 21.0º (SD: 3.4º, range: 13–27º) (Table 1). Comparison of all values with normal values demonstrated that the difference was statistically significant for all parameters (Table 2).

Table 1.

The number of patients and radiographic parameters (mean, range) by different grades of surgeon and fracture types

AO fracture type Operating surgeon Consultant Registrar with consultant supervision Registrar Overall
A n=3VT: 6.0º (0–9º)RL: 11.3mm (10–14mm)RI: 23.3º (21–27º) n=2VT: 3.5º (-6–13º)RL: 12.5mm (10–15mm)RI: 22.0º (21–23º) n=8VT: 9.8º (4–14º)RL: 11.2mm (9–13mm)RI: 21.9º (16–26) n=13VT: 8.0º (-6–14º)RL: 11.4mm (9–15mm)RI: 22.2º (16–27º)
B n=2VT: 7.0º (5–9º)RL: 11.5mm (11–12mm)RI: 22.0º (22–22º) n=1VT: 2.0º (N/A)RL: 11.0mm (N/A)RI: 15.0º (N/A) n=6VT: 7.8º (-5–13º)RL: 10.7mm (7–15mm)RI: 22.0º (18–26º) n=9VT: 7.0º (-5–13º)RL: 10.9mm (7–15mm)RI: 21.2º (15–26º)
C n=14VT: 10.5º (0–20º)RL: 10.8mm (8–13mm)RI: 19.6º (16–24º) n=1VT: 9.0º (N/A)RL: 14.0mm (N/A)RI: 24.0º (N/A) n=11VT: 8.9º (-2–16º)RL: 10.6mm (9–13mm)RI: 20.9º (13–25º) n=26VT: 9.8º (-2–20º)RL: 10.8mm (8–14mm)RI: 20.3º (13–25º)
Overall n=19VT: 9.4º (0–20º)RL: 11.0mm (8–14mm)RI: 20.4º (16–27º) n=4VT: 4.5º (-6–13º)RL: 12.5mm (10–15mm)RI: 20.8º (15–24º) n=25VT: 8.9º (-5–16º)RL: 10.8mm (7–15mm)RI: 21.5º (13–26º) n=48VT: 8.8º (-6–20º)RL: 11.0mm (7–15mm) RI: 21.0º (13–27º)

AO = Arbeitsgemeinschaft für Osteosynthesefragen; VT = volar tilt; RL = radial length; RI = radial inclination

Table 2.

Comparison of obtained radiographic parameters with ‘normal’ values

Study Volar tilt p-value Radial length p-value Radial inclination p-value
Scheck, 19629 11º 0.007 12mm <0.001 23º <0.001
Mann, 199210 14.5º <0.001 13.5mm <0.001 25.4º <0.001

Comparison with uninjured wrists

The uninjured group was noted to have a mean VT of 8.5º (SD: 5.8º; range: -5–20º), a mean RL of 11.3mm (SD: 1.8mm, range: 7–16mm) and a mean RI of 24.3º (SD: 5.6º, range: 14–28º), with VT and RL remaining statistically different (p=0.02 and p=0.04 respectively); no difference was noted for RI (p=0.07).

Comparison with intraoperative values

The mean intraoperative VT was 12.7º (SD: 0.8º, range: 8–17º), the mean RL was 13.1mm (SD: 2.7mm, range: 5–18mm) and the mean RI was 21.4º (SD: 3.6º, range: 13–28º). No statistical difference was noted when comparing these with the three-month postoperative values (p=0.87, p=0.91 and p=0.61 respectively).

Comparison with acceptable values

Comparison with the radiographic criteria for acceptable healing of a distal radial fracture11 demonstrated that VT and RL were corrected for all patients but two patients had RIs of 13.0º and 14.4º, which thus approached an acceptable value but did not reach it (Table 3).

Table 3.

The number and percentage of patients who had radiographic parameters within acceptable limits

Acceptable measurement Number of patients
Volar tilt 15º dorsal tilt – 20º volar tilt 48 (100%)
Radial length >8.5mm 48 (100%)
Radial inclination ≥15º 46 (95.8%)

Influence of fracture type and surgeon grade

Univariate and covariate analysis determined that neither fracture type, surgeon grade or the interaction between them affected VT (p=0.36, p=0.28 and p=0.67 respectively), RL (p=0.57, p=0.17 and p=0.71 respectively) or RI (p=0.25, p=0.63 and p=0.36 respectively).

Complications

Follow-up of patients at 12 months determined that 9 patients had complications attributable to surgery, with 3 in the group where the consultant was the primary surgeon, 2 where the surgeon was a registrar operating under consultant supervision and 4 in the group where the registrar was operating independently; this was not statistically significant (p=0.32). Two patients had chronic regional pain syndrome: one suffered with a palsy of the sensory branch of the median nerve that had reversed by three months and one with a superficial wound infection that was treated successfully with oral antibiotics. Five patients required removal of the plate: two for poor plate positioning causing either joint or soft tissue impingement, one for placement of an intra-articular screw and two for fracture collapse with secondary joint impingement.

Discussion

This study demonstrates that so-called normal values of distal radial anatomy are not replicated when using the DVR® plate to treat unstable fractures of the distal radius with respect to VT, RL or RI. Despite this, the reduction achieved fell within acceptable limits in most cases irrespective of fracture complexity and the DVR® plate can be used by surgeons of differing experience without compromising this. Of note, however, is that the complication rate in our series approached one in five cases, which reinforces the need for suitable training and patient selection.

This study was limited by its use of plain radiography as a measure of radiographic parameters. Although radiography was standardised, it is well recognised that rotation of the forearm, which could occur, may affect these parameters with pronation of 10º decreasing the apparent VT, RL and RI by 4.4º, 1.6mm and 2.8º respectively.12 There is consequently an argument for computed tomography assessment over plain radiography since it is more reliable for quantifying displacement.13 Unfortunately, our retrospective design did not allow for computed tomography in the present study. Furthermore, consideration would need to be given to the additional radiation exposure in any prospective study.

The second potential limitation of the study relates to the use of expected values against which radiographic parameters were compared rather than patients’ contralateral uninjured wrists. While it may be expected that a patient’s own anatomy would be a better comparison, the mean difference of 2.5º for VT, 1.5º for RI and 0.5mm for ulnar variance that has been shown to exist between the wrists of healthy subjects14 could limit its suitability. This is corroborated by Schuind et al, who compared the variability of right and left wrists on plain radiography with the variability of the distribution of those measurements from within the general population.15 It is noted that although the contralateral uninjured wrist should be used for assessing carpal measures, the normal side does not provide a better reference than normal values obtained from databases for VT, RI or ulnar variance.

The question of whether volar locking plates offer superior outcomes over other treatment options has been a recent topic for debate. Direct comparison with percutaneous wire fixation remains difficult as this method tends to be indicated only for extra-articular fractures whereas internal fixation methods such as volar locking plates can be used both for those that do and do not affect articular congruity.

In a specific cohort of patients older than 70 years, Arora et al demonstrated that radiographic parameters were significantly better in those treated with a locking plate than in those treated non-operatively.16 It is, however, worth noting that subjective and functional outcomes did not differ at a mean follow-up of 4 years and 7 months. When compared with external fixation, volar locking plates have been shown to have improved function at three months4 although this difference was no longer present subsequently, with either similar or better radiographic outcomes.4,17 These cumulative findings appear to support the use of volar locking plates for the treatment of fractures of the distal radius.

While our study demonstrated a disparity between normal values and the achieved reduction, the mean values fell within the accepted limits and this was not influenced by the grade of operating surgeon. However, since we now have discriminatory scoring and evaluation methods, we recommend these acceptable limits be investigated to see whether they still hold true. Although traditionally associated with outcome, there is recent evidence from a number of authors that in an elderly population aged >65–70 years, radiographic indices do not correlate with outcome.16,18,19 Taken together with the added cost effectiveness of percutaneous wires,20 this suggests that routine volar locking plate fixation in this group may not be justified.

The difficulty in discussing and evaluating volar locking plates comes from the wide variety of implants available, and to date, there are no reported human clinical studies comparing them. First generation DVR® plates, which have a single row of distal screw holes, have been shown to be biomechanically superior to non-locking devices such as percutaneous wires21 or simple plates6 although they are comparable biomechanically with locking plates from other manufacturers under physiological loads6,8,2224 and nonspanning external fixators.25 The addition of a second row of distal screw holes as found in the second generation DVR® plates does not improve this.7 It is nevertheless worth noting that this design modification was to prevent subchondral collapse, which may be important clinically rather than biomechanically.

The clinical results of the DVR® plate have been reported previously in 48 patients with AO type C fractures by Frattini et al.26 In this subgroup of patients, the radiographic outcomes were again within acceptable limits with a mean VT of 10.3º, a mean RL of 9.2mm and a mean RI of 23.2º. This is comparable with our patients with type C fractures. However, it is of note that we have shown acceptable radiographic parameters can be attained in the majority of patients irrespective of fracture type.

Conclusions

We have shown that that the DVR® plate, which is a fixed angle volar locking plate, is able to restore the bony anatomy to within acceptable limits in the majority of patients following an unstable fracture of the distal radius. The parameter that exhibits the widest variability is VT and while the rate of complication approached one in five cases, most were attributable to surgically related factors rather than failure of fixation, emphasising the need for suitable preoperative counselling of patients undergoing plate osteosynthesis of the distal radius irrespective of which implant is used.

References

  • 1.O’Neill TW, Cooper C, Finn JD et al Incidence of distal forearm fracture in British men and women. Osteoporos Int 2001; : 555–558. [DOI] [PubMed] [Google Scholar]
  • 2.Thompson PW, Taylor J, Dawson A. The annual incidence and seasonal variation of fractures of the distal radius in men and women over 25 years in Dorset, UK. Injury 2004; : 462–466. [DOI] [PubMed] [Google Scholar]
  • 3.Bacorn RW, Kurtzke JF. Colles’ fracture; a study of two thousand cases from the New York State Workmen’s Compensation Board. J Bone Joint Surg Am 1953; : 643–658. [PubMed] [Google Scholar]
  • 4.Wei DH, Raizman NM, Bottino CJ et al Unstable distal radial fractures treated with external fixation, a radial column plate, or a volar plate. A prospective randomized trial. J Bone Joint Surg Am 2009; : 1,568–1,577. [DOI] [PubMed] [Google Scholar]
  • 5.Rozental TD, Blazar PE, Franko OI et al Functional outcomes for unstable distal radial fractures treated with open reduction and internal fixation or closed reduction and percutaneous fixation. A prospective randomized trial. J Bone Joint Surg Am 2009; : 1,837–1,846. [DOI] [PubMed] [Google Scholar]
  • 6.Willis AA, Kutsumi K, Zobitz ME, Cooney WP. Internal fixation of dorsally displaced fractures of the distal part of the radius. A biomechanical analysis of volar plate fracture stability. J Bone Joint Surg Am 2006; : 2,411–2,417. [DOI] [PubMed] [Google Scholar]
  • 7.Levin SM, Nelson CO, Botts JD et al Biomechanical evaluation of volar locking plates for distal radius fractures. Hand 2008; : 55–60. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Sobky K, Baldini T, Thomas K et al Biomechanical comparison of different volar fracture fixation plates for distal radius fractures. Hand 2008; : 96–101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Scheck M. Long-term follow-up of treatment of comminuted fractures of the distal end of the radius by transfixation with Kirschner wires and cast. J Bone Joint Surg Am 1962; : 337–351. [PubMed] [Google Scholar]
  • 10.Mann FA, Wilson AJ, Gilula LA. Radiographic evaluation of the wrist: what does the hand surgeon want to know. ? Radiology 1992; : 15–24. [DOI] [PubMed] [Google Scholar]
  • 11.Graham TJ. Surgical correction of malunited fractures of the distal radius. J Am Acad Orthop Surg 1997; : 270–281. [DOI] [PubMed] [Google Scholar]
  • 12.Pennock AT, Phillips CS, Matzon JL, Daley E. The effects of forearm rotation on three wrist measurements: radial inclination, radial height and palmar tilt. Hand Surg 2005; : 17–22. [DOI] [PubMed] [Google Scholar]
  • 13.Cole RJ, Bindra RR, Evanoff BA et al Radiographic evaluation of osseous displacement following intra-articular fractures of the distal radius: reliability of plain radiography versus computed tomography. J Hand Surg Am 1997; : 792–800. [DOI] [PubMed] [Google Scholar]
  • 14.Hollevoet N, Van Maele G, Van Seymortier P, Verdonk R. Comparison of palmar tilt, radial inclination and ulnar variance in left and right wrists. J Hand Surg Br 2000; : 431–433. [DOI] [PubMed] [Google Scholar]
  • 15.Schuind F, Alemzadeh S, Stallenberg B, Burny F. Does the normal contralateral wrist provide the best reference for x-ray film measurements of the pathologic wrist. ? J Hand Surg Am 1996; : 24–30. [DOI] [PubMed] [Google Scholar]
  • 16.Arora R, Gabl M, Gschwentner M et al A comparative study of clinical and radiologic outcomes of unstable Colles type distal radius fractures in patients older than 70 years: nonoperative treatment versus volar locking plating. J Orthop Trauma 2009; : 237–242. [DOI] [PubMed] [Google Scholar]
  • 17.Wright TW, Horodyski M, Smith DW. Functional outcome of unstable distal radius fractures: ORIF with a volar fixed-angle tine plate versus external fixation. J Hand Surg Am 2005; : 289–299. [DOI] [PubMed] [Google Scholar]
  • 18.Jaremko JL, Lambert RG, Rowe BH et al Do radiographic indices of distal radius fracture reduction predict outcomes in older adults receiving conservative treatment. ? Clin Radiol 2007; : 65–72. [DOI] [PubMed] [Google Scholar]
  • 19.Grewal R, MacDermid JC. The risk of adverse outcomes in extra-articular distal radius fractures is increased with malalignment in patients of all ages but mitigated in older patients. J Hand Surg Am 2007; : 962–970. [DOI] [PubMed] [Google Scholar]
  • 20.Shyamalan G, Theokli C, Pearse Y, Tennent D. Volar locking plates versus Kirschner wires for distal radial fractures – a cost analysis study. Injury 2009; : 1,279–1,281. [DOI] [PubMed] [Google Scholar]
  • 21.Knox J, Ambrose H, McCallister W, Trumble T. Percutaneous pins versus volar plates for unstable distal radius fractures: a biomechanic study using a cadaver model. J Hand Surg Am 2007; : 813–817. [DOI] [PubMed] [Google Scholar]
  • 22.Liporace FA, Kubiak EN, Jeong GK et al A biomechanical comparison of two volar locked plates in a dorsally unstable distal radius fracture model. J Trauma 2006; : 668–672. [DOI] [PubMed] [Google Scholar]
  • 23.Chen L, Dai Q, Wongworawat MD. A biomechanical comparison between two volar locking plate systems for distal radius fractures. Orthopedics 2006; : 927–929. [DOI] [PubMed] [Google Scholar]
  • 24.McCall TA, Conrad B, Badman B, Wright T. Volar versus dorsal fixed-angle fixation of dorsally unstable extra-articular distal radius fractures: a biomechanic study. J Hand Surg Am 2007; : 806–812. [DOI] [PubMed] [Google Scholar]
  • 25.Strauss EJ, Banerjee D, Kummer FJ, Tejwani NC. Evaluation of a novel, nonspanning external fixator for treatment of unstable extra-articular fractures of the distal radius: biomechanical comparison with a volar locking plate. J Trauma 2008; : 975–981. [DOI] [PubMed] [Google Scholar]
  • 26.Frattini M, Soncini G, Corradi M et al Complex fractures of the distal radius treated with angular stability plates. Chir Organi Mov 2009; : 155–162. [DOI] [PubMed] [Google Scholar]

Articles from Annals of The Royal College of Surgeons of England are provided here courtesy of The Royal College of Surgeons of England