Skip to main content
Laboratory Animal Research logoLink to Laboratory Animal Research
. 2016 Dec 23;32(4):241–248. doi: 10.5625/lar.2016.32.4.241

Changes of hematological references depends on storage period and temperature conditions in rats and dogs

Jung-Min Lee 1, Jin Seok Kang 2,
PMCID: PMC5206231  PMID: 28053618

Abstract

Because changes in rat and dog hematological parameters according to storage conditions have been poorly documented, we sought to examine such changes. Blood analysis was performed using two hematology analyzers (ADVIA 2120i and Sysmex XN-V) after storage at room temperature and in cold storage for 5, 24, and 48 h, respectively. Interassay coefficients of variation for hematological parameters analyzed with the ADVIA 2120i and the XN-V showed similar. The levels of hematocrit, mean corpuscular volume, mean corpuscular hemoglobin, mean corpuscular hemoglobin concentration, and platelet (PLT) showed significant variations with time in blood samples of rats and dogs. The leukocyte subpopulation showed high variation with storage conditions. The data for leukocyte differential counts obtained using the ADVIA 2120i, XN-V, and a manual differential counting procedure showed good agreement for neutrophils and lymphocyte counts, but monocytes, eosinophils, and basophils showed differences between the procedures. In conclusions, most rat and dog hematological parameters showed minimal changes; however, some showed high variation with storage time and temperature, especially PLT and leukocyte subpopulations. In conclusion, when performing hematological analysis in dogs and rats, it will be exactitude to analyze blood samples in fresh condition and at least within 24 h in the cold storage.

Keywords: Rat, dog, hematology, analysis, stability


It is recommended that hematological analyses be performed rapidly after taking the samples, or at least 24 h. However, in various situations, changes in experimental animal hematology measurements may occur when analyses are delayed for some reasons. Indeed, changes in animal hematology measurements may occur due to secondary factors, such as sample storage conditions or storage period [1].

Human blood specimens stored at room temperature for more than 1 day were found to be acceptable, with some limitations for the complete blood count (CBC), but not for the differential [2]. For rat blood samples, plasma should be run immediately after collection for reliable activated partial thromboplastin time results [3]. For dog blood samples, differences in white blood cell (WBC) and platelet (PLT) counts varied with the specimen [4]. As accurate automated measurements are very important for bioassays, these reports raise questions about the accuracy of measured variables and data validation after blood storage.

Reference ranges of standard experimental parameters are useful for comparisons in toxicology. Thus, historical control data of experimental animals would help interpreting the effects of test substances in routine toxicity and efficacy studies. With hematology, data may be influenced by environmental conditions, such as sample storage methods, and by animal factors, such as species, strain, age, and sex [5].

Although reports about hematological parameters according to the age of the rodent and some blood parameters according to storage conditions have been published, changes in all hematological parameters according to storage conditions have been poorly documented for rat and dog blood samples. Thus, in this study, we sought to examine changes in hematological parameters with different storage condition, analyzed using two representative hematology analyzers (Siemens ADVIA 2120i and Sysmex XN-V). Blood samples from rats and dogs were measured with these analyzers within 5 h of sampling, and after storage for 24 and 48 h at 4 ℃ or room temperature.

Materials and Methods

Sample collection and storage methods

The animals had been fasted for ~18 h before sampling. Rat and dog samples were collected from various studies. The samples were placed into dipotassium ethylenediaminetetraacetic acid (EDTA) and analyzed within 5 h of collection, except those used in stability studies. Collected whole blood specimens were stored at 4 ℃ or room temperature, and assessed after 5, 24, and 48 h. This study (NSU-16-7) was approved by the animal experiment committee of Namseoul University, based on the Animal Protection Act.

Measurement of hematological parameters

The following hematological tests were performed using an automated hematology analyzer: an ADVIA 2120i (Siemens Healthcare Diagnostics, Tarrytown, NY) or an XN-V (Sysmex, Kobe, Japan). Hematological parameters included CBC parameters and leukocyte subpopulations; erythrocyte count (RBC), hemoglobin, hematocrit, reticulocytes, mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), mean corpuscular hemoglobin concentration (MCHC), PLT, WBC, neutrophils, lymphocytes, monocytes, eosinophils, and basophils.

Interassay repeatability

Interassay repeatability was evaluated by calculating the coefficient of variation (CV) of 10 consecutive measurements of internal quality control material in three different controls (low, normal, and high). Coefficients of variation (%) were calculated as standard deviation (SD)/mean×100.

Changes of hematological parameters by storage time and temperature

Blood samples from rats (n=40) and dogs (n=20) were analyzed within 5 h of collection and then at 24 and 48 h after storage at 4 ℃ or room temperature using both the ADVIA 2120i, and the XN-V for hematological parameters. For both species, the differences between the first analysis and the subsequent analyzers for hematological parameters were examined.

Comparison of rat and dog hematological parameters analyzed by the ADVIA 2120i and XN-V

Hematological parameters analyzed by ADVIA 2120i and XN-V were compared for correlations at each time point.

Comparison of leukocyte differential counts between manual counting and automatic counting

The rat and dog blood samples showing high variation between ADVIA 2120i and XN-V were smeared on a microscopic slide and stained with Diff-Quick (Hemacolor, Merck). Leukocyte differential counts were carried out manually and these data were compared with those obtained with the ADVIA 2120i and XN-V.

Statistical analysis

Statistical analysis was performed with the SAS (software ver. 9.3, SAS Institute Inc., Cary, NC). Data were analyzed using Bartlett's test for homogeneity of variance. One-way analysis of variance was used for homogeneous data, and if significant, Dunnett's test was then used for multiple comparisons (significance levels: P<0.05 and P<0.01, two-tailed). The Kruskal-Wallis was sued with heterogeneous data, and then, if significant, Steel's test was applied for multiple comparisons (significance levels: P<0.05 and P<0.01, two-tailed).

Results

Interassay repeatability

Interassay CV values are shown in Tables 1 and 2 for the ADVIA 2120i and XN-V, respectively. In analyses using the ADVIA 2120i, CV values over 3% were seen in reticulocytes and PLT of low, normal, and high controls, and in hematocrit of the normal control. In analyses using the XN-V, CV values over 3% were seen in reticulocytes and PLT of both the low and the normal controls. Compared with the ADVIA 2120i, the CV value of PLT was lower than with the XN-V. The CV values in leukocyte subpopulations of three different controls analyzed with the ADVIA 2120i and XN-V showed high levels, ranging from 4.2 to 28.4%.

Table 1. Interassay repeatability of the Siemens ADVIA 2120i hematology analyzer performed with quality control materials at three different concentrations (low, normal, and high).

Low Normal High
Mean±SD CV% Mean±SD CV% Mean±SD CV%
RBC (×106 cells/µL) 2.35±0.03 1.1 4.27±0.08 2.0 5.19±0.06 1.1
Hgb (g/dL) 6.2±0.05 0.8 12.8±0.30 2.3 17.4±0.26 1.5
HCT (%) 16.9±0.33 2.0 34.7±1.09 3.1 46.0±1.12 2.4
Reti (%) 5.96±0.24 4.0 3.35±0.12 3.6 0.53±0.03 5.5
MCV (f/L) 72.6±0.33 0.5 82.9±0.38 0.5 89.8±0.51 0.6
MCH (pg) 26.2±0.28 1.1 30.3±0.53 1.7 33.5±0.36 1.1
MCHC (g/dL) 36.4±0.44 1.2 36.8±0.47 1.3 37.4±0.37 1.0
PLT (×106 cells/µL) 76±2.79 3.7 232±8.23 3.6 452±15.85 3.5
WBC (×103 cells/µL) 3.61±0.11 3.1 7.93±0.24 3.1 19.74±0.41 2.1
Neu (%) 55.4±1.71 3.1 61.7±1.89 3.1 70.9±0.98 1.4
Lym (%) 31.0±1.10 3.5 24.5±0.88 3.6 17.3±0.32 1.8
Mono (%) 8.3±1.40 16.8 7.9±0.85 10.8 6.5±1.12 17.2
Eos (%) 1.9±0.39 20.8 1.9±0.27 14.1 2.1±0.32 15.2
Baso (%) 0.4±0.10 26.8 0.3±0.05 21.1 0.2±0.05 28.4

Abbreviations: CV, coefficient of variation; RBC, erythrocyte count; Hgb, hemoglobin; HCT, hematocrit; Reti, reticulocytes; MCV, mean corpuscular volume; MCH, mean corpuscular hemoglobin; MCHC, mean corpuscular hemoglobin concentration; PLT, platelets; WBC, leukocytes; Neu, neutrophils; Lym, lymphocytes; Mono, monocytes; Eos, eosinophils; Baso, basophils

Table 2. Interassay repeatability of the Sysmex XN-V hematology analyzer performed with quality control materials in three different concentrations (low, normal, and high).

Low Normal High
Mean±SD CV% Mean±SD CV% Mean±SD CV%
RBC (×106 cells/µL) 2.32±0.02 1.0 4.38±0.04 1.0 5.19±0.05 0.9
Hgb (g/dL) 5.5±0.06 1.0 11.6±0.10 0.8 16.1±0.08 0.5
HCT (%) 16.1±0.16 1.0 33.9±0.33 1.0 45.9±0.39 0.8
Reti (%) 6.9±0.23 3.3 4.2±0.18 4.4 0.3±0.01 1.7
MCV (f/L) 69.3±0.09 0.1 77.4±0.11 0.1 88.5±0.08 0.1
MCH (pg) 23.8±0.25 1.0 26.6±0.13 0.5 31.1±0.28 0.9
MCHC (g/dL) 34.3±0.31 0.9 34.3±0.18 0.5 35.2±0.29 0.8
PLT (×106 cells/µL) 84±3.3 4.0 233±6.68 2.9 481±6.95 1.4
WBC (×103 cells/µL) 3.43±0.04 1.2 7.02±0.09 1.3 16.85±0.25 1.5
Neu (%) 59.3±1.02 1.7 61.6±0.79 1.3 67.0±0.77 1.1
Lym (%) 24.7±0.68 2.7 23.0±0.83 3.6 18.1±0.38 2.1
Mono (%) 8.1±0.34 4.2 7.9±0.49 6.3 8.6±0.43 5.0
Eos (%) 2.1±0.32 15.4 2.3±0.20 8.7 2.0±0.13 6.4
Baso (%) 0.2±0.05 28.4 0.3±0.05 21.1 0.1±0.00 0.0

Abbreviations: CV, coefficient of variation; RBC, erythrocyte count; Hgb, hemoglobin; HCT, hematocrit; Reti, reticulocytes; MCV, mean corpuscular volume; MCH, mean corpuscular hemoglobin; MCHC, mean corpuscular hemoglobin concentration; PLT, platelets; WBC, leukocytes; Neu, neutrophils; Lym, lymphocytes; Mono, monocytes; Eos, eosinophils; Baso, basophils

Changes in hematological parameters with storage time and temperature

The effects of sample storage times and temperatures prior to analysis with the ADVIA 2120i and XN-V were examined by evaluating changes in hematological parameters of rat and dog samples both fresh and after 24 and 48 h of storage at 4 ℃ and room temperature (Tables 3 and 4). The hematological parameters analyzed with the ADVIA 2120i and XN-V showed minimal changes in RBC, hemoglobin, reticulocytes, and WBC after 24 and 48 h of storage at 4 ℃ or room temperature.

Table 3. Changes in hematological parameters according to storage time and temperature (ADVIA 2120i).

Rat (n=40) Dog (n=20)
Fresh sample After storage for Difference (%) Fresh sample After storage for Difference (%)
24h 48h 24h 48h
RBC (×106 cells/µL)
 Refrigerated 7.84 7.81 7.84 0.00 (0.0) 6.96 6.98 7.09 0.13 (1.9)
 Room temperature 7.45 7.50 7.48 0.03 (0.4) 7.33 7.37 7.43 0.10 (1.4)
Hgb (g/dL)
 Refrigerated 16.5 16.3 16.8 0.3 (1.8) 16.0 15.9 16.4 0.4 (2.5)
 Room temperature 15.7 15.5 16.1 0.4 (2.5) 17.1 17.2 17.8 0.7 (4.1)
HCT (%)
 Refrigerated 43.6 44.4 44.8 1.2 (2.8) 45.4 43.8 44.7 −0.7 (−1.5)
 Room temperature 40.9 44.9** 45.3** 4.4 (10.7) 48.4 45.1 46.1 −2.3 (−4.8)
Reti (%)
 Refrigerated 2.28 2.06 2.40 0.12 (5.26) 0.58 0.57 0.55 −0.03 (−5.2)
 Room temperature 2.39 2.41 3.09 0.70 (29.3) 1.00 0.90 0.98 −0.02 (−2.0)
MCV (f/L)
 Refrigerated 55.7 56.9 57.3** 1.6 (2.9) 65.4 62.9 63.2 2.2 (3.4)
 Room temperature 55.2 59.8** 60.9** 5.7 (10.3) 66.1 61.2** 62.2** 3.9 (5.9)
MCH (pg)
 Refrigerated 21.2 20.8 21.4 0.2 (0.9) 23.0 22.8 23.2 0.2 (0.9)
 Room temperature 21.2 20.8 21.7 0.5 (2.4) 23.4 23.4 24.1** 0.7 (3.0)
MCHC (g/dL)
 Refrigerated 38.0 36.6** 37.4** −0.6 (−1.6) 35.2 36.3 36.7** 1.5 (4.3)
 Room temperature 38.4 34.8** 35.6** −2.8 (−7.3) 35.4 38.2 38.7 3.3 (9.3)
PLT (×106 cells/µL)
 Refrigerated 947 879 896 −51 (−5.4) 295 297 301 6 (2.0)
 Room temperature 906 738** 831 −75 (−8.3) 297 295 297 0 (0.0)
WBC (×103 cells/µL)
 Refrigerated 9.83 8.97 9.05 −0.78 (−7.9) 10.71 10.65 10.56 0.15 (1.4)
 Room temperature 8.50 7.59 7.44 −1.06 (−12.5) 11.77 11.32 11.34 0.43 (3.7)
Neu (%)
 Refrigerated 14.18 15.32 9.21 −4.97 (−35.0) 50.22 49.98 1.40** −48.82 (−97.2)
 Room temperature 12.63 13.76 15.29 2.66 (21.1) 53.17 53.03 12.22** −40.95 (−77.0)
Lym (%)
 Refrigerated 81.5 81.3 87.1 5.6 (6.9) 39.3 39.6 95.9** 56.6 (144.0)
 Room temperature 83.5 83.1 81.0 −2.5 (−3.0) 36.3 37.0 89.4** 53.1 (146.3)
Mono (%)
 Refrigerated 2.3 1.2** 0.7** 1.6 (69.6) 5.1 5.0 0.7** −4.4 (−86.3)
 Room temperature 2.0 1.1** 1.4** −0.6 (−30.0) 4.4 4.1 1.2** −3.2 (−72.7)
Eos (%)
 Refrigerated 1.2 1.4 2.3 1.1 (91.7) 3.7 3.9 0.7** −3.0 (−81.1)
 Room temperature 1.0 1.4* 1.7** 0.7 (70.0) 3.9 4.1 1.4* −2.5 (−64.1)
Baso (%)
 Refrigerated 0.4 0.3 0.4 0.0 (0.0) 1.3 1.0 1.1 −0.2 (−15.4)
 Room temperature 0.3 0.3 0.4** 0.1 (33.3) 1.8 1.3 1.4 −0.4 (-22.2)

Abbreviations: RBC, erythrocyte count; Hgb, hemoglobin; HCT, hematocrit; Reti, reticulocytes; MCV, mean corpuscular volume; MCH, mean corpuscular hemoglobin; MCHC, mean corpuscular hemoglobin concentration; PLT, platelets; WBC, leukocytes: Neu, neutrophils; Lym, lymphocytes; Mono, monocytes; Eos, eosinophils; Baso, basophils

*, ** Significantly different from control (P<0.05, P<0.01, respectively)

Table 4. Changes in hematological parameters according to storage time and temperature (XN-V).

Rat (n=40) Dog (n=20)
Fresh sample After storage for Difference (%) Fresh sample After storage for Difference (%)
24h 48h 24h 48h
RBC (×106 cells/µL)
 Refrigerated 7.88 7.80 7.74 −0.14 (−1.8) 7.01 7.00 7.06 0.05 (0.7)
 Room temperature 7.53 7.48 7.27 −0.26 (−3.5) 7.48 7.50 7.45 −0.03 (−0.4)
Hgb (g/dL)
 Refrigerated 15.5 15.2 15.4 −0.1 (−0.6) 16.3 16.4 16.5 0.2 (1.2)
 Room temperature 14.6 14.4 14.4 −0.2 (−1.4) 18.0 17.8 18.0 0.0 (0.0)
HCT (%)
 Refrigerated 44.3 45.1 44.9 0.6 (1.4) 46.3 45.2 45.7 −0.6 (−1.3)
 Room temperature 41.8 45.4** 44.9* 3.1 (7.4) 50.4 48.1* 48.3* −1.9 (−3.8)
Reti (%)
 Refrigerated 3.56 3.55 3.55 −0.01 (−0.3) 0.73 0.64 0.66 −0.07 (−9.6)
 Room temperature 3.80 3.61 3.72 −0.08 (−2.1) 1.29 1.03 1.24 −0.05 (−3.9)
MCV (f/L)
 Refrigerated 56.4 58.0* 58.1* 1.7 (3.0) 66.3 64.7 64.8 1.5 (2.3)
 Room temperature 55.9 61.1** 62.2** 6.3 (11.3) 67.5 64.2* 65.0 −2.5 (−3.7)
MCH (pg)
 Refrigerated 19.7 19.6 19.9 0.2 (1.0) 23.3 23.4 23.3 0.0 (0.0)
 Room temperature 19.6 19.5** 20.0** 0.4 (2.0) 24.1 23.8 24.1 0.0 (0.0)
MCHC (g/dL)
 Refrigerated 34.9 33.8** 34.2** −0.7 (−2.0) 35.1 36.1* 36.1* 1.0 (2.8)
 Room temperature 35.0 31.8** 32.1** −2.9 (−8.3) 35.7 37.1** 37.2** 1.5 (4.2)
PLT (×106 cells/µL)
 Refrigerated 1072 1016 1003 −69 (−6.4) 328 291 279 −49 (−14.9)
 Room temperature 1011 953* 923** −88 (−8.7) 446 295 294 −152 (−34.1)
WBC (×103 cells/µL)
 Refrigerated 10.22 9.78 9.80 −0.42 (−4.1) 11.09 11.03 11.02 −0.07 (−0.6)
 Room temperature 8.85 8.09 8.63 −0.22 (−2.5) 12.11 12.10 12.08 −0.03 (−0.2)
Neu (%)
 Refrigerated 13.85 15.78 18.38** 4.53 (32.7) 47.69 48.97 49.23 1.54 (3.2)
 Room temperature 12.48 15.74* 23.23** 10.75 (86.1) 51.41 53.35 55.54** 4.13 (8.0)
Lym (%)
 Refrigerated 78.0 77.1 75.2* −2.8 (−3.6) 38.4 37.3 37.3 −1.1 (−2.9)
 Room temperature 80.5 78.4 70.8** −9.7 (−12.0) 35.6 34.1 34.5 −1.1 (−3.1)
Mono (%)
 Refrigerated 7.0 5.4** 4.5** −2.5 (−35.7) 10.0 9.7 9.6 −0.4 (−4.0)
 Room temperature 6.0 4.1** 3.7** −2.3 (−38.3) 8.9 7.8 5.8** −3.1 (−34.8)
Eos (%)
 Refrigerated 1.0 1.5* 1.7** 0.7 (70.0) 3.5 3.5 3.3 −0.2 (−5.7)
 Room temperature 0.9 1.6** 2.1** 1.2 (133.3) 3.5 3.9 3.4 −0.1 (−2.9)
Baso (%)
 Refrigerated 0.2 0.2 0.2 0.0 (0.0) 0.4 0.6 0.6 0.2 (50.0)
 Room temperature 0.1 0.2 0.2** 0.1 (100.0) 0.6 0.9 0.8 0.2 (33.3)

Abbreviations: RBC, erythrocyte count; Hgb, hemoglobin; HCT, hematocrit; Reti, reticulocytes; MCV, mean corpuscular volume; MCH, mean corpuscular hemoglobin; MCHC, mean corpuscular hemoglobin concentration; PLT, platelets; WBC, leukocytes; Neu, neutrophils; Lym, lymphocytes; Mono, monocytes; Eos, eosinophils; Baso, basophils

*, ** Significantly different from control (P<0.05, P<0.01, respectively)

In rat samples, most CBC parameters, except hematocrit, MCV, MCHC, and PLT, showed changes <±5% between fresh and after 48 h at 4 ℃. In contrast, changes in most leukocyte subpopulations were >±5% between fresh samples and after 24 and 48 h of storage.

In dog samples, all CBC parameters except reticulocytes and PLT at 4℃ and PLT at room temperature showed changes <±5% between fresh and 48 h. Changes in most leukocyte subpopulations showed changes >10% after 48 h at 4 ℃ and room temperature compared with the results of fresh samples.

Compared with those of rats, leukocyte subpopulations in dogs showed smaller changes over 48 h at 4 ℃ and room temperature. Changes in hematological parameters of two species between fresh samples and after 24 and 48 h of storage using the ADVIA 2120i were similar to those with the XN-V, but some parameters showed changes >±5% between times versus results with the XN-V.

Comparison of rat and dog hematological parameters analyzed with ADVIA 2120i and XN-V

Hematological parameters analyzed by ADVIA 2120i and XN-V were shown in Table 5. There were good agreement and correlations between the data from the both analyzers for CBC parameters and leukocyte subpopulations. Correlation coefficients (r2) of most CBC parameters were >0.9; exceptions including PLT in rat and MCHC in dog.

Table 5. Comparison of rat and dog hematological parameters analyzed by the ADVIA 2120i and XN-V.

Rat Dog
ADVIA 2120i XN-V Correlation (r2) ADVIA 2120i XN-V Correlation (r2)
RBC (×106 cells/µL) 7.63 7.68 0.999 7.31 7.36 0.999
Hgb (g/dL) 15.96 14.97 0.999 16.86 17.58 0.996
HCT (%) 42.70 43.58 0.989 48.89 50.13 0.968
Reti (%) 2.03 3.48 0.998 0.74 1.08 0.997
MCV (f/L) 56.05 56.92 0.972 66.64 68.23 0.973
MCH (pg) 20.99 19.55 0.999 23.17 23.94 0.975
MCHC (g/dL) 37.45 34.37 0.989 34.76 35.09 0.604
PLT (×106 cells/µL) 868.3 1010.4 0.725 273.6 281.7 0.913
WBC (×103 cells/µL) 8.65 9.13 0.998 11.53 11.84 0.997
Neu (%) 12.60 12.22 0.979 52.37 51.24 0.946
Lym (%) 83.57 80.78 0.855 36.67 35.36 0.930
Mono (%) 1.75 5.82 0.986 4.86 9.37 0.944
Eos (%) 1.29 0.96 0.996 4.36 3.60 0.970
Baso (%) 0.29 0.15 0.999 1.82 0.48 0.993

Abbreviations: RBC, erythrocyte count; Hgb, hemoglobin; HCT, hematocrit; Reti, reticulocytes; MCV, mean corpuscular volume; MCH, mean corpuscular hemoglobin; MCHC, mean corpuscular hemoglobin concentration; PLT, platelets; WBC, leukocytes; Neu, neutrophils; Lym, lymphocytes; Mono, monocytes; Eos, eosinophils; Baso, basophils

Comparison of leukocyte differential counts between manual counting and automatic counting

Data for leukocyte differential counts obtained using the ADVIA 2120i, XN-V, and a manual differential counting procedure were showed in Table 6. There were good agreements for neutrophils and lymphocytes counts, but monocytes, eosinophils, and basophils showed difference among the three procedures.

Table 6. Comparison of leukocyte differential counts obtained with the ADVIA 2120i and XN-V and a manual differential counting procedure for rat and dog blood samples.

Parameter Species ADVIA 2120i XN-V Man-Diff
Neu (%) Rat 11.3±3.4a 11.8±3.3 12.0±4.0
Dog 53.2±5.6 52.2±5.4 54.0±6.0
Lym (%) Rat 84.4±4.0 82.5±4.2 84.0±4.0
Dog 35.4±5.7 34.5±5.7 36.0±7.0
Mono (%) Rat 1.4±0.4 5.2±1.6 4.0±4.0
Dog 5.0±1.3 9.5±1.9 8.0±2.0
Eos (%) Rat 1.5±1.0 1.0±0.4 0.0±1.0
Dog 4.1±1.2 3.3±1.3 1.0±1.0
Baso (%) Rat 0.3±1.0 0.2±0.2 0.0±0.0
Dog 2.0±0.6 0.4±0.2 0.0±0.0

Abbreviations: NEW, neutrophils; LYM, lymphocytes; MONO, monocytes; EOS, eosinophils; BASO, basophils

aMean±SD

Discussion

This study was performed to examine changes in rat and dog blood parameters according to storage period and temperature. Because it was assumed that blood storage at room temperature might cause moderate variations in some hematological parameters, we first examined changes in blood parameters upon room temperature storage. In rat samples, hematocrit increased at room temperature. MCV increased and MCHC decreased upon room temperature storage, and at 4 ℃. Generally, it has been reported that refrigerated storage is important for the stability of CBC in human samples [6]. As MCV had potential to increase after 24 h if EDTA-treated samples were stored at room temperature [7], it seemed that EDTA as anticoagulant did not allow the accurate determination of the CBC at 24 h with storage at room temperature. However, our study showed that there were increases in MCV even with storage at 4 ℃. We suggest that strict protocols should be established for the storage of rat blood samples to minimize variations in hematological parameters.

In this study, the storage of dog samples at room temperature caused a decrease in MCV and an increase in MCHC, when analyzed with both blood analyzers. It has been reported that, using dog blood samples, hematocrit, MCV, and MCHC were influenced by erythrocyte swelling during measurement [8]. It has also been reported that the storage of dog samples at room temperature caused little or no change in RBC, hemoglobin, or MCH, while there were moderate increases in hematocrit, MCV, and reticulocytes count and a moderate decrease in MCHC [1].

In the rat samples, monocytes were decreased with storage at room temperature, and at 4 ℃. However, eosinophils were increased with storage at room temperature, although there was no change in WBC counts. It seems that, because the populations of monocytes and eosinophils were low, the change in total WBC was minimal. In the dog samples, neutrophils decreased markedly and monocytes and eosinophils decreased with storage for 48 h at room temperature, and at 4 ℃. However, lymphocytes increased markedly with storage for 48 h at room temperature, and at 4 ℃. However, there was no change in WBC counts. These results indicated dog samples should not be stored for more than 24 h, because there were many variations in differential counts. It was reported that spuriously low WBC counts may be observed because of agglutination in the presence of EDTA [7]. Disodium, dipotassium, and tripotassium EDTA have been used for hematological testing [9]. In this study, dipotassium EDTA did not cause a change in WBC counts.

In this study, two representative hematology analyzers were shown to be reliable, good agreement and correlations between the data from the analyzers. However, the r2 value of PLT was 0.725 in rat samples and the r2 value of MCHC was 0.604 in dog samples. These values may have been due to the detection method. For PLT, a PLT-fluorescent method demonstrated good results for reproducibility in samples with platelet counts less than 50×10 9/L and could distinguish platelets from fragmented erythrocytes [10,11]. As there is little information for the detection of MCHC, further studies are needed to investigate the difference between the two analyzers.

We showed good agreement for neutrophils and lymphocytes counts. However, monocytes, eosinophils, and basophils showed difference among the three procedures. Notably, the manual monocytes count was similar to that with the XN-V, but higher than the value with the ADVIA 2120i in rat and dog blood samples. Manual eosinophils and basophils counts were lower than those with the XN-V. In addition, monocytes counts may be underestimated and eosinophils and basophils numbers may be overestimated with the ADVIA 2120i, but not the XN-V. It has been reported that the Siemens ADVIA 120 showed notably lower measurements of lymphocytes and monocytes with human blood samples [12]. In contrast, this study suggested that monocytes numbers may be overestimated with the ADVIA 2120i in rat and dog blood samples. Using human specimens, a higher percentage of lymphocytes and a lower percentage of monocytes originated from an erroneous classification of immature and small monocytes as lymphocytes [13]. In many situations, the WBC differential scatter gram is of great help in flagging. Flags generated depend on the software version in the hematology analyzer, and the performance in detecting the same anomalies may differ according to which analyzer is used, even those from the same manufacturer [14]. As no animal data about leukocyte subpopulations have been reported yet, this study provides fundamental animal data.

Taken together, most hematological parameters in rat and dog showed minimal changes, according to storage time and temperature; however, some parameters showed high variation, especially for PLT and leukocyte subpopulation. In conclusion, when performing hematological analysis in dogs and rats, it will be exactitude to analyze blood samples in fresh condition and at least within 24 h in the cold storage.

Footnotes

Conflict of interests: The authors declare that there is no financial conflict of interests to publish these results.

References

  • 1.Bourgès-Abella NH, Geffré A, Deshuillers PL, Braun JP, Trumel C. Changes in hematology measurements in healthy and diseased dog blood stored at room temperature for 24 and 48 hours using the XT-2000iV analyzer. Vet Clin Pathol. 2014;43(1):24–35. doi: 10.1111/vcp.12119. [DOI] [PubMed] [Google Scholar]
  • 2.Gulati GL, Hyland LJ, Kocher W, Schwarting R. Changes in automated complete blood cell count and differential leukocyte count results induced by storage of blood at room temperature. Arch Pathol Lab Med. 2002;126(3):336–342. doi: 10.5858/2002-126-0336-CIACBC. [DOI] [PubMed] [Google Scholar]
  • 3.Goyal VK, Kakade S, Pandey SK, Gothi AK, Nirogi R. Determining the effect of storage conditions on prothrombin time, activated partial thromboplastin time and fibrinogen concentration in rat plasma samples. Lab Anim. 2015;49(4):311–318. doi: 10.1177/0023677215597137. [DOI] [PubMed] [Google Scholar]
  • 4.Médaille C, Briend-Marchal A, Braun JP. Stability of selected hematology variables in canine blood kept at room temperature in EDTA for 24 and 48 hours. Vet Clin Pathol. 2006;35(1):18–23. doi: 10.1111/j.1939-165x.2006.tb00083.x. [DOI] [PubMed] [Google Scholar]
  • 5.Matsuzawa T. Present status of animal clinical pathology examinations in the Japanese Pharmaceutical Manufacturers Association. Toxicol Pathol. 1992;20(3 Pt 2):528–533. doi: 10.1177/0192623392020003214. [DOI] [PubMed] [Google Scholar]
  • 6.Wood BL, Andrews J, Miller S, Sabath DE. Refrigerated storage improves the stability of the complete blood cell count and automated differential. Am J Clin Pathol. 1999;112(5):687–695. doi: 10.1093/ajcp/112.5.687. [DOI] [PubMed] [Google Scholar]
  • 7.Zandecki M, Genevieve F, Gerard J, Godon A. Spurious counts and spurious results on haematology analysers: a review. Part II: white blood cells, red blood cells, haemoglobin, red cell indices and reticulocytes. Int J Lab Hematol. 2007;29(1):21–41. doi: 10.1111/j.1365-2257.2006.00871.x. [DOI] [PubMed] [Google Scholar]
  • 8.Bauer N, Nakagawa J, Dunker C, Failing K, Moritz A. Evaluation of the automated hematology analyzer Sysmex XT-2000iV ™ compared to the ADVIA ® 2120 for its use in dogs, cats, and horses. Part II: Accuracy of leukocyte differential and reticulocyte count, impact of anticoagulant and sample aging. J Vet Diagn Invest. 2012;24(1):74–89. doi: 10.1177/1040638711436243. [DOI] [PubMed] [Google Scholar]
  • 9.England JM, Rowan RM, van Assendelft OW, Bull BS, Coulter W, Fujimoto K, Groner W, Richardson-Jones A, Klee G, Koepke JA, Lewis SM, McLaren CE, Shinton NK, Tatsumi N, Verwilghen RL. Recommendations of the International Council for Standardization in Haematology for Ethylenediaminetetraacetic Acid Anticoagulation of Blood for Blood Cell Counting and Sizing. International Council for Standardization in Haematology: Expert Panel on Cytometry. Am J Clin Pathol. 1993;100(4):371–372. doi: 10.1093/ajcp/100.4.371. [DOI] [PubMed] [Google Scholar]
  • 10.Schoorl M, Schoorl M, Oomes J, van Pelt J. New fluorescent method (PLT-F) on Sysmex XN2000 hematology analyzer achieved higher accuracy in low platelet counting. Am J Clin Pathol. 2013;140(4):495–499. doi: 10.1309/AJCPUAGGB4URL5XO. [DOI] [PubMed] [Google Scholar]
  • 11.Wada A, Takagi Y, Kono M, Morikawa T. Accuracy of a New Platelet Count System (PLT-F) Depends on the Staining Property of Its Reagents. PLoS One. 2015;10(10):e0141311. doi: 10.1371/journal.pone.0141311. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Meintker L, Ringwald J, Rauh M, Krause SW. Comparison of automated differential blood cell counts from Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples. Am J Clin Pathol. 2013;139(5):641–650. doi: 10.1309/AJCP7D8ECZRXGWCG. [DOI] [PubMed] [Google Scholar]
  • 13.Tan BT, Nava AJ, George TI. Evaluation of the Beckman Coulter UniCel DxH 800, Beckman Coulter LH 780, and Abbott Diagnostics Cell-Dyn Sapphire hematology analyzers on adult specimens in a tertiary care hospital. Am J Clin Pathol. 2011;135(6):939–951. doi: 10.1309/AJCP1V3UXEIQTSLE. [DOI] [PubMed] [Google Scholar]
  • 14.Zandecki M, Genevieve F, Gerard J, Godon A. Spurious counts and spurious results on haematology analysers: a review. Part I: platelets. Int J Lab Hematol. 2007;29(1):4–20. doi: 10.1111/j.1365-2257.2006.00870.x. [DOI] [PubMed] [Google Scholar]

Articles from Laboratory Animal Research are provided here courtesy of Korean Association for Laboratory Animal Science

RESOURCES