Abstract
Background
The 2014 CDC 4th generation HIV screening algorithm includes an orthogonal immunoassay to confirm and discriminate HIV-1 and HIV-2 antibodies. Additional nucleic acid testing (NAT) is recommended to resolve indeterminate or undifferentiated HIV seroreactivity. HIV-2 NAT requires a second-line assay to detect HIV-2 total nucleic acid (TNA) in patients’ blood cells, as a third of untreated patients have undetectable plasma HIV-2 RNA.
Objectives
To validate a qualitative HIV-2 TNA assay using peripheral blood mononuclear cells (PBMC) from HIV-2–infected Senegalese study participants.
Study design
We evaluated the assay precision, sensitivity, specificity, and diagnostic performance of an HIV-2 TNA assay. Matched plasma and PBMC samples were collected from 25 HIV-1, 30 HIV-2, 8 HIV-1/-2 dual-seropositive and 25 HIV seronegative individuals. Diagnostic performance was evaluated by comparing the outcome of the TNA assay to the results obtained by the 4th generation HIV screening and confirmatory immunoassays.
Results
All PBMC from 30 HIV-2 seropositive participants tested positive for HIV-2 TNA including 23 patients with undetectable plasma RNA. Of the 30 matched plasma specimens, one was HIV non-reactive. Samples from 50 non-HIV-2 infected individuals were confirmed as non-reactive for HIV-2 Ab and negative for HIV-2 TNA. The agreement between HIV-2 TNA and the combined immunoassay results was 98.8% (79/80). Furthermore, HIV-2 TNA was detected in 7 of 8 PBMC specimens from HIV-1/HIV-2 dual-seropositive participants.
Conclusions
Our TNA assay detected HIV-2 DNA/RNA in PBMC from serologically HIV-2 reactive, HIV indeterminate or HIV undifferentiated individuals with undetectable plasma RNA, and is suitable for confirming HIV-2 infection in the HIV testing algorithm.
Keywords: HIV-2, diagnostic, qualitative, total nucleic acid, DNA/RNA, PCR
BACKGROUND
Accurate HIV testing is the cornerstone of HIV/AIDS diagnosis, prevention, care and treatment. In the current United States Centers for Disease Control and Prevention (CDC) 4th generation HIV-1/HIV-2 testing algorithm [1], the initial screening utilizes HIV-1/2 antigen(Ag)/antibody(Ab) combination immunoassays, and resulting reactive samples are further tested using an HIV-1/HIV-2 Ab differentiation immunoassay. Samples determined to be HIV non-reactive or HIV-1 Ab and/or HIV-2 Ab indeterminate are recommended for further testing with FDA-approved HIV-1 nucleic acid tests (NAT). Due to the absence of an FDA-approved HIV-2 NAT, HIV-1/HIV-2 undifferentiated (dually seropositive) and HIV-2 Ab indeterminate samples are not clearly addressed in the algorithm. Although detectable HIV-2 plasma RNA confirms HIV-2 infection, undetectable HIV-2 plasma viral loads (<10 RNA copies/mL) have been reported in 36–46% of antiretroviral (ART)-naïve HIV-2–infected patients [2–4]. A second-line NAT that specifically detects HIV-2 DNA/RNA in PBMC is therefore essential for accurate HIV-2 diagnosis and would fulfill the need for a confirmatory HIV-2 assay in the current CDC testing algorithm.
OBJECTIVES
Our objective was to develop and validate a qualitative HIV-2 total nucleic acid (TNA) assay using the Abbott m2000 platform for clinical and research use.
STUDY DESIGN
Clinical samples
Senegalese study participant data, plasma, and PBMC were collected at the Clinique Des Maladies Infectieuses Ibrahima DIOP Mar, Centre Hospitalier Universitaire de Fann, Universite Cheikh Anta Diop de Dakar, Senegal, as part of ongoing studies investigating HIV-2 and HIV-1/HIV-2 dual infection and antiretroviral therapy. Written informed patient consent was obtained, and the University of Washington (UW) Institutional Review Board and the Senegal Ethics Committee (CNERS) approved all aspects of the study. Whole blood was collected in EDTA vacutainer tubes and PBMC were isolated by Ficoll gradient centrifugation. Plasma and PBMC were saved in Senegal at −80°C, shipped in liquid nitrogen dry shippers or dry ice to Seattle and stored at −80°C at UW. HIV infection status was determined in Senegal from serum samples using the Alere Determine™ HIV-1/-2 rapid test and the ImmunoCombII HIV- 1 & 2 BiSpot assay (ImmunoCombII) (Alere). For genotyping, portions of the HIV-2 pol gene were PCR-amplified from PBMC and sequenced, and the data were subjected to phylogenetic analysis [5, 6]. PBMC and plasma from HIV-2 seropositive and HIV-1/-2 dual-positive participants, who at the time of collection were either ART-experienced or ART-naïve, were used for this study.
Paired PBMC and plasma samples from 25 HIV-1-infected and 25 HIV-negative individuals were provided by the UW Center for AIDS Research HIV Specimen Repository Program. All subsequent testing was performed in the CLIA-certified UW Retrovirology Laboratory using good laboratory practice.
Assay controls and standards
A molecular clone (plasmid pROD9) containing the full genome of the HIV-2 group A ROD strain was used to prepare positive controls and standards. The concentration of DNA in the pROD stock was measured using a NanoDrop 1000 spectrophotometer (Thermo Fisher Scientific, MA) to calculate the copy number of HIV-2 genomes in pROD9 working dilutions. All working dilutions of pROD9 DNA was linearized by digestion with XhoI.
The negative control for the assay contained 1.3 million human PBMC prepared from HIV negative donors; one million cells were subject to the assay extraction procedure described below. The positive control contained nominal 180 copies of linear pROD9 DNA/million PBMC. Each serial dilution of pROD9 DNA at nominal 46, 23, 12, 6 or 3, copies were added to one million PBMC to evaluate assay sensitivity. The probit analysis function of Minitab 17 (Minitab statistical software, PA) was used to determine the limit of detection (LOD) and 95% confidence interval of the HIV-2 TNA assay.
To evaluate the potential for anticoagulant interference, venous blood was drawn from an HIV negative individual into vacutainers containing EDTA, sodium citrate, heparin, or acid citrate dextrose (ACD). PBMC were isolated from each blood sample by Ficoll gradient centrifugation. Aliquots of 1.3 million cells were then pelleted by centrifugation for assay testing.
HIV-2 TNA assay
For each sample, a 1.3-million PBMC pellet was mixed with 50μL Abbott proteinase K (>6000 units/mL), and 150μL of bulk lysis buffer (Abbott Molecular, US), and agitated at 56°C for 30 minutes using an Eppendorf Thermomixer at 5000rpm. Then, an additional 1.1-mL of bulk lysis buffer was added to each cell lysate to yield a final volume of 1.3 mL/sample. One-mL of lysate was processed by the Abbott m2000sp instrument using the Abbott mSample DNA preparation kit and an Abbott proprietary customer-specific DNA/RNA extraction protocol. Both DNA and RNA were extracted and eluted in 210μL of buffer. Approximately 1000 copies of exogenous internal control RNA (IC) comprised of in vitro transcripts of a recombinant clone including fruit fly and bacteria genes [2]were dispensed to each sample during the extraction procedure, and subsequently co-amplified with HIV-2 DNA/RNA.
HIV-2 specific PCR primers and a fluorescent probe detecting both HIV-2 groups A and B were used as previously described [2]. The final concentrations of the primers and the probe in the RT-PCR step of this assay were 900nM and 250nM, respectively. Amplification with the same primers and a second HIV-2 probe (5′FAM-TGCTGGAGAGAACCTCCC-BHQPlus, Biosearch Technologies, CA) was applied to samples showing inefficient amplification by the first HIV-2 probe.
PCR fluorescence analysis and result reporting for this assay were designed using the Abbott laboratory-defined application program specific for the qualitative assay. The cut-off cycle number (CN) was determined by examining amplification graphs generated for the aforementioned low-copy pROD9 DNA standards used to evaluate assay sensitivity; the cut-off CN was near the largest HIV-2 CN reported for those standards and the level of assay detection limit. The Delta Cycle (DC) value for each sample was calculated as the cut-off CN minus the reported HIV-2 CN. Therefore, larger DC values corresponded to higher levels of HIV-2 TNA in the cell lysates. Samples yielding DC values between 0 and 1 were interpreted as inconclusive status and were subject to re-test. Samples testing negative were reported as “Not Detected”.
The endogenous genomic DNA control was used by amplifying human ribonuclease P gene (RPP30) to ensure adequate DNA was extracted from the approximately 1 million PBMC used in the assay-specific extraction protocol. Briefly, for the RPP30 qPCR assay, 5μL of extracted eluate was amplified along with human genomic DNA standards at 50ng, 5ng and 0.5ng/reaction (Human Taqman Control Genomic DNA, Thermo Fisher Scientific) using the Applied Biosystems 7900HT fast real-time PCR instrument. The primer/probe sequences and thermocycler settings were adopted from Hindson et al, [7].
HIV assays
The Abbott ARCHITECT HIV Ag/Ab Combo assay (Combo) determined HIV serological status and provided a signal-to-cutoff ratio (S/CO) for each sample. The Bio-Rad Multispot HIV-1/-2 Rapid Test (Multispot) was used to confirm and differentiate HIV-1/HIV-2 Ab. Plasma samples with dual-reactive results were diluted by 10-or 100-fold for further testing [8]. Samples that were Multispot dual-reactive following a 1:100 dilution were reported to be HIV reactive but undifferentiated. The Bio-Rad Geenius™ HIV 1/2 Supplemental Assay (Geenius) was used to retest Multispot HIV-undifferentiated specimens[9]. The UW HIV-2 RNA quantitative assay [2] was used to measure HIV-2 RNA levels in the plasma specimens. The UW HIV-1 TNA assay was a modification of the Abbott RealTime HIV-1 Qualitative assay [10] and used the aforementioned DNA/RNA extraction protocol.
RESULTS
Assay precision and sensitivity
To evaluate assay precision, a total of 24 replicates of positive and negative controls were tested daily over two days. Four sets of positive and negative controls were tested daily over eight additional days for the PBMC samples. There were no false-positive negative controls (N=56) detected by the HIV-2 TNA assay. The assay mean, total assay standard deviation, and coefficient of variance for the positive control (N=56) were 5.76 DC, 0.40 DC and 6.97%, respectively.
To determine assay sensitivity, nominal copies of linear pROD9 DNA (46, 23, 12, 6, or 3 copies) in 1 million of PBMC were extracted and amplified. Sixteen replicas were tested at each DNA concentration across five days, as summarized in Table 1. The LOD for the assay was determined to be 26 copies/million cells (95%CI, 18–55 copies/million cells).
Table 1.
Copies of HIV-2 pROD9 DNA/million cells | Number Tested | Number Detected | Percent Detected (%) |
---|---|---|---|
46 | 16 | 16 | 100 |
23 | 16 | 15 | 93.4 |
12 | 16 | 10 | 62.5 |
6 | 16 | 9 | 56.3 |
3 | 16 | 3 | 18.8 |
Assay specificity
PBMC samples from 25 HIV-negative, 25 HIV-1-infected, and 30 HIV-2-infected individuals were tested with assay positive and negative controls over eight different days. All PBMC from 25 HIV-negative and 25 HIV-1–infected individuals tested negative by the HIV-2 TNA assay. Thirty of thirty (100%; 95%CI lower bound 88.4%) PBMC from Senegalese HIV-2 seropositive participants tested positive; the median DC was 6.36 (IQR 5.01–7.74; range, 1.07–9.36). Twenty-three of the 30 (76.7%) tested HIV-2 seropositive participants had undetectable plasma RNA (Table 2). Twenty-three of the 30 samples were HIV-2 group A, 1 was HIV-2 group B and in 6 samples the HIV-2 group was not available. In addition, the HIV-2 TNA assay detected HIV-2 in PBMC from an additional 15 of 15 (100%) Senegalese HIV-2 seropositive participants whose matching plasma samples were unavailable.
Table 2.
Plasma Sample | PBMC Sample | ||||||
---|---|---|---|---|---|---|---|
HIV Ag/Ab Combo# | Multispot^ | HIV-2 RNA | HIV-2 TNA | HIV-1 TNA | |||
Specimen ID | Group | Results (S/CO+) | Interpretation | Results (c/mL) | Results | DC~ | Results |
H2A32 | A | 5.89 | HIV NEG | TND* | Detected | 4.99 | Not detected |
H2A76 | A | 22.07 | HIV-2 (wk) | TND | Detected | 8.24 | Not detected |
H2A24 | A | 54.78 | HIV-2-after 1:10 | 32 | Detected | 8.86 | Not detected |
H2A67 | NT$ | 66.66 | HIV-2 (wk) | TND | Detected | 1.07 | Not detected |
H2A72 | B | 99.79 | HIV-2 | TND | Detected | 7.24 | Not detected |
H2A25 | A | 122.55 | HIV-2 | <10, detected | Detected | 5.73 | NT$ |
H2A66 | NT$ | 123.75 | HIV-2 | TND | Detected | 6.17 | Not detected |
H2A64 | A | 139.49 | HIV-2-after 1:10 | TND | Detected | 6.92 | Not detected |
H2A26 | A | 142.28 | HIV-2 | TND | Detected | 5.03 | NT$ |
H2A53 | A | 149.89 | HIV-2-after 1:100 | TND | Detected | 6.39 | Not detected |
H2A84 | NT$ | 157.64 | HIV-2 | TND | Detected | 3.00 | Not detected |
H2A14 | A | 160.84 | HIV-2-after 1:100 | n/a | Detected | 8.34 | Not detected |
H2A78 | NT$ | 165.91 | HIV-2 | TND | Detected | 6.17 | Not detected |
H2A3 | A | 168.28 | HIV-2-after 1:100 | 115 | Detected | 7.97 | Not detected |
H2A30 | A | 179.93 | HIV-2-after 1:100 | TND | Detected | 9.39 | Not detected |
H2A10 | A | 186.91 | HIV-2 | TND | Detected | 6.58 | Not detected |
H2A34 | A | 189.62 | HIV-2 | <10, detected | Detected | 5.32 | Not detected |
H2A38 | NT$ | 192.39 | HIV-2-after 1:100 | <10, detected | Detected | 5.59 | Not detected |
H2A74 | A | 192.51 | HIV-2 | TND | Detected | 9.08 | Not detected |
H2A62 | A | 196.98 | HIV-2 | TND | Detected | 4.76 | Not detected |
H2A57 | A | 205.59 | HIV-2 | TND | Detected | 6.66 | Not detected |
H2A77 | A | 206.43 | HIV-2-after 1:10 | TND | Detected | 8.46 | Not detected |
H2A22 | A | 207.58 | HIV-2-after 1:10 | TND | Detected | 5.15 | Not detected |
H2A55 | A | 219.28 | HIV-2 | TND | Detected | 8.67 | Not detected |
H2A6 | A | 222.62 | HIV-2 | TND | Detected | 6.51 | Not detected |
H2A33 | A | 223.00 | HIV-2-after 1:100 | TND | Detected | 6.32 | Not detected |
H2A20 | A | 225.10 | HIV-2 | 16 | Detected | 3.92 | Not detected |
H2A82 | NT$ | 234.44 | HIV-2 | TND | Detected | 8.25 | Not detected |
H2A9 | A | 234.58 | HIV-2 | TND | Detected | 4.72 | Not detected |
H2A31 | A | 293.67 | HIV-2 | TND | Detected | 4.17 | Not detected |
Abbott ARCHITECT HIV Ag/Ab Combo assay
Multispot results interpreted without dilution are stated without further description. Samples determined to be HIV-2 seropositive after 1:10 or 1:100 dilutions as stated. Wk, weak positive.
NT, Not tested;
S/CO, Signal-to-cutoff ratio;
TND, Target not detected; ~DC, Delta Cycle.
Monitoring assay extraction procedure and potential inhibitors
To determine an acceptable range for RNA extraction performance, we analyzed the IC RNA CN for 64 controls and 80 PBMC samples tested over 8 assay runs. The mean (±SD) of IC CN for PBMC samples and controls prepared from EDTA blood was 32.95 (±0.41). Based on these results, samples with an IC CN >3-SD above the mean (i.e., CN> 34.18) were considered for re-testing.
To account for variation due to cell counting and DNA extraction, the extracted eluates were tested separately using the RPP30 qPCR assay. Analysis of RPP30 cycle number across the threshold (Ct) for the 64 controls and 80 tested samples generated a median of 23.10 and a range from 21.18 to 24.85. The acceptable RPP30 Ct for adequate DNA extraction was no more than 25 as determined by this validation study.
To test for potential interference, PBMC prepared from blood samples containing each of four different anti-coagulants were tested in quadruplicate. IC CN (mean ± SD) were 32.54±0.21, 32.83±0.26, 32.52±0.23, and 32.66±0.35 for PBMC pellets prepared with EDTA, sodium citrate, heparin, and ACD anti-coagulant, respectively, indicating no significant differences among the four groups (p=0.364, one-way ANOVA). The RPP30 Ct values were less than 25 for all sixteen PBMC pellets, indicating no interference on DNA extraction by sodium citrate, heparin, or ACD anti-coagulant.
Diagnostic performance compared to the CDC 4th generation testing algorithm
The assay’s diagnostic performance for HIV-2 infection status was evaluated by comparing the outcome of PBMC tested using the HIV-2 TNA assay with the serologic results obtained from the Combo and Multispot immunoassays using matching plasma samples. All 30 (100%) of the HIV-2 infected plasma samples were HIV reactive by the Combo (median S/CO=183.5; range, 5.89–293.67) and 29 of 30 (96.7%) samples were confirmed to be HIV-2 Ab reactive by the Multispot test. A single HIV-2 plasma sample (H2A32) was reactive by the Combo with an S/CO of 5.89 but was non-reactive by the Multispot test. This participant was previously determined to be HIV-2 infected by the ImmunoCombII assay. The H2A32 plasma sample also had undetectable HIV-2 RNA, while the corresponding PBMC was positive for HIV-2 TNA (DC=4.99) and negative for HIV-1 TNA. In addition, two participants (H2A67 and H2A76), whose plasma samples were weakly positive by the Multispot test, were positive for PBMC HIV-2 TNA (DC=1.07, 8.24, respectively; Table 2).
Plasma from 25 HIV-negative individuals tested non-reactive by the Combo assay. Plasma specimens from 25 HIV-1 infected individuals tested HIV reactive by Combo (median S/CO=525.1; range, 444.00–738.80) and were confirmed by the Multispot test. As mentioned above, the matched PBMC samples from 25 HIV-negative and 25 HIV-1 infected individuals were negative by the HIV-2 TNA assay. Therefore, the agreement between the HIV-2 TNA assay and the combined results of the Combo and Multispot immunoassays was 98.75% (79 of 80), as summarized in Table 3.
Table 3.
The matching plasma samples tested by HIV Ag/Ab Combo and Multispot assays | |||
---|---|---|---|
HIV-2 Ab reactive | HIV-2 Ab non-reactive | ||
PBMC tested by HIV-2 TNA assay | Detected | 29 | 1* |
Not-detected | 0 | 50+ |
HIV reactive by Abbott ARCHITECT HIV Ag/Ab Combo assay; HIV nonreactive by Multispot. Participant’s baseline HIV testing: HIV reactive by Alere Determine™ HIV-1/-2 rapid test and HIV-2 reactive by the ImmunoComb II HIV- 1 & 2 BiSpot assay; negative HIV-2 RNA in plasma.
The number is the combined result for 25 HIV-negative and 25 HIV-1 infected individuals.
HIV-1 and HIV-2 dually seropositive Senegalese
PBMC from eight HIV-1/-2 dually-infected participants were tested by both the HIV-1 and HIV-2 TNA assays (Table 4). PBMC from six participants were positive for both HIV-1 and HIV-2 TNA and two were positive for either HIV-1 or HIV-2 TNA. The HD10 PBMC sample negative for HIV-2 TNA likely had HIV-2 DNA/RNA below the assay LOD. All eight corresponding plasma samples diluted up to 100-fold remained positive for HIV Ab (undifferentiated) by the Multispot assay. When tested by the Geenius assay, undiluted plasma samples from the dually-infected participants were re-confirmed as having undifferentiated (untypable) HIV results.
Table 4.
Specimen ID | Plasma sample | PBMC sample | ||
---|---|---|---|---|
HIV-1 RNA copies/mL | HIV-2 RNA copies/mL | HIV-1 TNA | HIV-2 TNA | |
HD01 | TND# | TND | Detected | Detected |
HD02 | 85120 | 2592 | Detected | Detected |
HD06 | 510596 | 311 | Detected | Detected |
HD09 | TND | TND | Detected | Detected |
HD12 | TND | TND | Detected | Detected^ |
H2A036 | TND | 136 | Detected | Detected |
HD13 | Detected, <40c/ml | 28726 | Not detected | Detected |
HD10 | 13730 | TND | Detected | Not detected |
HIV Ab was not differentiated for the sample at a 1:100 dilution.
Target not detected.
Detected by a HIV-2 second probe.
DISCUSSION
We developed and validated a qualitative HIV-2 TNA diagnostic assay running on the Abbott m2000 platform. For clinical usage, endogenous and exogenous controls were implemented to ensure adequate extraction and amplification of DNA and RNA from PBMC specimens. With regard to diagnostic sensitivity, the HIV-2 TNA assay was able to detect HIV-2 DNA/RNA in PBMC samples when the corresponding HIV-2 seropositive plasma samples were undetectable for HIV-2 RNA. The assay diagnostic performance largely agreed with the serological determination by the Combo and Multispot assays, with the exception of one plasma sample (H2A32) with a relatively low S/CO of 5.68 and undetectable HIV-2 plasma RNA (Table 2). This case demonstrates the utility of the plasma HIV-2 RNA and PBMC HIV-2 TNA assays for plasma specimens that yield indeterminate and undifferentiated HIV-1/-2 Ab results when tested using the CDC 4th generation HIV screening algorithm.
The ability to confirm and distinguish HIV-1, HIV-2 and HIV-1/HIV-2 dual infections has important implications for clinical management and antiretroviral therapy [11–13]. Currently, per the CDC 4th generation HIV diagnostic algorithm, the FDA-approved assays to confirm and differentiate HIV-1 or/and HIV-2 serology are the Multispot [14–16] and the Geenius tests [17, 18]. The Multispot reports nonreactive, HIV-1 reactive, HIV-2 reactive, or HIV reactive (undifferentiated) results. The Geenius cassette contains two HIV-2 peptide bands and 4 HIV-1 peptide or protein bands. The Geenius Reader detects the presence or absence of the bands and provides assay results using a proprietary algorithm[9]. In addition to the aforementioned Multispot interpretations, Geenius reports HIV-2 indeterminate, HIV indeterminate and HIV-2 positive with HIV-1 cross-reactivity results. Several recent studies have reported HIV-2 indeterminate cases using the Geenius test, which led to a recommendation to use HIV-2 NAT to further determine HIV infection status [18–20]; as such, the HIV-2 TNA assay has the potential to address this need.
Several groups used “in-house” quantitative assays or a modified Roche Amplicor test to measure HIV-2 proviral DNA in PBMC [3, 21, 22], ranging from not-detectable up to 4 logs copies/million PBMC. Though many HIV-2 infected participants in these three studies were highlighted for having plasma RNA below assay detection limits—40, 100 or 250 copies/mL, some of these studied samples would have had detectable or quantifiable plasma RNA by more current HIV-2 RNA assays [2, 23, 24]. Lately, Styer et al. [25] has developed a droplet digital PCR assay to detect HIV-2 DNA in the whole blood and blood cell pellets. We demonstrated that our assay has improved sensitivity by detecting HIV-2 in PBMC from 23 HIV-2 infected participants with undetectable plasma RNA, <10 copies/mL, and by amplification of both cell-associated DNA and RNA in PBMC.
In conclusion, the described HIV-2 TNA assay is a sensitive NAT for detecting HIV-2 DNA/RNA in PBMC specimens and is suitable for the diagnosis of HIV-2 for patients with serologically indeterminate or undifferentiated HIV status and undetectable HIV-2 plasma RNA viral load. As such, this validated HIV-2 TNA assay addresses an important deficiency in the current CDC 4th generation HIV testing algorithm.
Highlights.
A qualitative assay detecting HIV-2 total nucleic acid in patients’ PBMCs
A second-line assay to resolve HIV-2 sero-indeterminate cases
Able to detect HIV-2 DNA and RNA in PBMC from patients with undetected plasma RNA
Acknowledgments
FUNDING: NIH/NIAID; AI060466 & AI120765
We thank the participants in Senegal; HIV-2 pROD9 was kindly provided by Michael Emerman, Fred Hutchinson Cancer Research Center, Seattle, Washington; the UW CFAR Clinical Research and Retrovirology Core Repository program for the provision of PBMC and plasma samples from HIV-1 infected patients and healthy individuals (the UW CFAR program is supported by NIAID, NCI, NIMH, NIDA, NICHD, NHLBI, NIA, NIGMS, NIDDK of the National Institutes of Health under award number P30AI027757), HVTN HIV Diagnostic Laboratory (UM-AI-068618), and the AIDS Clinical Trials Group Laboratory Center (UM1-AI-106701) for technical support. These studies were supported by grants to GSG from the National Institutes of Health/National Institute of Allergy and Infectious Diseases (NIH/NIAID; AI060466 & AI120765). The University of Washington-Dakar HIV-2 Study Group includes Moussa Seydi, Papa Salif Sow, Selly Ba, Fatima Sall, Fatou Traore, Khadim Faye, Ousseynou Ndiaye, Marie Pierre Sy, Bintou Diaw, Mbaye Ndoye, Amadou Bale Diop, Marianne Fadam Diome (Service des Maladies Infectieuses Ibrahima Diop Mar, Centre Hospitalier Universitaire de Fann, Universite Cheikh Anta Diop de Dakar, Dakar, Senegal); Alassane Niang, ElHadji Ibrahima Sall, Ousseynou Cisse, Ibrahima Tito Tamba, Jean Phillippe Diatta, Raphael Bakhoum, Jacque Francois Sambou, Juliette Gomis (Region Medicale de Ziguinchor, Ziguinchor, Casamance, Senegal); Stephen Hawes, Noelle Benzekri, Robert Coombs, Ming Chang, John Lin, Nancy Kiviat, James Mullins, Sally Leong, and Vincent Wu (University of Washington). We also thank Gavin Cloherty, Bryne Martinez, Danijela Lucic (Abbott Molecular) for advice and technical support.
Footnotes
COMPETING INTERESTS: None
ETHICAL APPROVAL: UW IRB and the Senegal EC (CNERS)
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Branson B, et al. Laboratory testing for the diagnosis of HIV infection: updated recommendations. 2014 http://www.cdc.gov/hiv/pdf/hivtestingalgorithmrecommendation-final.pdf.
- 2.Chang M, et al. Validation for clinical use of a novel HIV-2 plasma RNA viral load assay using the Abbott m2000 platform. J Clin Virol. 2012;55(2):128–33. doi: 10.1016/j.jcv.2012.06.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Gottlieb GS, et al. Equal plasma viral loads predict a similar rate of CD4+ T cell decline in human immunodeficiency virus (HIV) type 1- and HIV-2-infected individuals from Senegal, West Africa. J Infect Dis. 2002;185(7):905–14. doi: 10.1086/339295. [DOI] [PubMed] [Google Scholar]
- 4.Ekouevi DK, et al. Plasma HIV-2 RNA According to CD4 Count Strata among HIV-2-Infected Adults in the IeDEA West Africa Collaboration. PLoS One. 2015;10(6):e0129886. doi: 10.1371/journal.pone.0129886. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Gottlieb GS, et al. Emergence of multiclass drug-resistance in HIV-2 in antiretroviral-treated individuals in Senegal: implications for HIV-2 treatment in resouce-limited West Africa. Clin Infect Dis. 2009;48(4):476–83. doi: 10.1086/596504. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Gottlieb GS, et al. HIV-2 Integrase Variation in Integrase Inhibitor-Naive Adults in Senegal, West Africa. PLoS One. 2011;6(7):e22204. doi: 10.1371/journal.pone.0022204. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Hindson BJ, et al. High-throughput droplet digital PCR system for absolute quantitation of DNA copy number. Anal Chem. 2011;83(22):8604–10. doi: 10.1021/ac202028g. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Multispot HIV-1/HIV-2 Rapid Test [package insert] Bio-Rad Laboratories; Redmond, WA: 2008. [Google Scholar]
- 9.Geenius™ HIV 1/2 Supplemental Assay [package insert] Bio-Rad Laboratories; Redmond, WA: 2014. [Google Scholar]
- 10.Huang S, et al. A novel RealTime HIV-1 Qualitative assay for the detection of HIV-1 nucleic acids in dried blood spots and plasma. J Virol Methods. 2011;178(1–2):216–24. doi: 10.1016/j.jviromet.2011.09.015. [DOI] [PubMed] [Google Scholar]
- 11.Tchounga BK, Ekouevi DK, Eholie SP. Response to “Rapid tests for HIV type discrimination in West Africa may perform differently”. J Int AIDS Soc. 2015;18(1):19380. doi: 10.7448/IAS.18.1.19380. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Ekouevi DK, et al. Characteristics of HIV-2 and HIV-1/HIV-2 Dually Seropositive Adults in West Africa Presenting for Care and Antiretroviral Therapy: The IeDEA-West Africa HIV-2 Cohort Study. PLoS One. 2013;8(6):e66135. doi: 10.1371/journal.pone.0066135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Panel on Antiretroviral Guidelines for Adults and Adolescents. Guidelines for the use of antirectorviral agents in HIV-1-infected adults and adolescents. Department of Health and Human Services; [Section accessed on July 28, 2016]. Available at http://aidsinfo.nih.gov/contentfiles/lvguidelines/AdultandAdolescentGL.pdf. [Google Scholar]
- 14.Cardenas AM, Baughan E, Hodinka RL. Evaluation of the Bio-Rad Multispot HIV-1/HIV-2 Rapid Test as an alternative to Western blot for confirmation of HIV infection. J Clin Virol. 2013;58(Suppl 1):e97–e103. doi: 10.1016/j.jcv.2013.08.021. [DOI] [PubMed] [Google Scholar]
- 15.Pandori MW, et al. The Multispot rapid HIV-1/HIV-2 differentiation assay is comparable with the Western blot and an immunofluorescence assay at confirming HIV infection in a prospective study in three regions of the United States. J Clin Virol. 2013;58(Suppl 1):e92–6. doi: 10.1016/j.jcv.2013.10.006. [DOI] [PubMed] [Google Scholar]
- 16.Ramos EM, et al. Clinical performance of the Multispot HIV-1/HIV-2 rapid test to correctly differentiate HIV-2 from HIV-1 infection in screening algorithms using third and fourth generation assays and to identify cross reactivity with the HIV-1 Western Blot. J Clin Virol. 2013;58(Suppl 1):e104–7. doi: 10.1016/j.jcv.2013.08.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Malloch L, et al. Comparative evaluation of the Bio-Rad Geenius HIV-1/2 Confirmatory Assay and the Bio-Rad Multispot HIV-1/2 Rapid Test as an alternative differentiation assay for CLSI M53 algorithm-I. J Clin Virol. 2013;58(Suppl 1):e85–91. doi: 10.1016/j.jcv.2013.08.008. [DOI] [PubMed] [Google Scholar]
- 18.Delaney KES, Wesolowski L, Owen M, Branson B. CROI 2015. Seattle, Washington: 2015. Performance of the Geenius HIV-1/HIV-2 Assay in the CDC HIV Testing Algorithm. [Google Scholar]
- 19.Bennett SB, FS, Crowe S, Bajzik O, Stewart G, Ferret M, Cardona M. Comparative performance of the Geenius HIV-1/HIV-2 supplemental test in Florida’s public health testing population. 2016 IV Diagnostic Conference; 2016; Atlanta, GA. [Google Scholar]
- 20.Luo W, AS, Sullivan V, Peters PJ, Westheimer E, Cohen S, Gay C, Hightow-Weidman L, Owen SM, Masciotra S. Performance evaluation of FDA-approved HIV supplemental assays. 2016 HIV diagnostic conference; 2016; Atlanta, GA. [Google Scholar]
- 21.Popper SJ, et al. Low plasma human immunodeficiency virus type 2 viral load is independent of proviral load: low virus production in vivo. J Virol. 2000;74(3):1554–7. doi: 10.1128/jvi.74.3.1554-1557.2000. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Damond F, et al. Quantification of proviral load of human immunodeficiency virus type 2 subtypes A and B using real-time PCR. J Clin Microbiol. 2001;39(12):4264–8. doi: 10.1128/JCM.39.12.4264-4268.2001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Styer LM, Miller TT, Parker MM. Validation and clinical use of a sensitive HIV-2 viral load assay that uses a whole virus internal control. J Clin Virol. 2013;58(Suppl 1):e127–33. doi: 10.1016/j.jcv.2013.09.007. [DOI] [PubMed] [Google Scholar]
- 24.Avettand-Fenoel V, et al. New sensitive one-step real-time duplex PCR method for group A and B HIV-2 RNA load. J Clin Microbiol. 2014;52(8):3017–22. doi: 10.1128/JCM.00724-14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Styer LM, CN, Santos P, Stella JM, Parker MM. Validation of a Droplet Digital PCR Assay that Provides Sensitive Detection and Accurate Quantification of HIV-2 DNA in Whole Blood or Blood Cell Pellets. 2016 HIV Diagnostics Conference; 2016; Altanta, GA. [Google Scholar]