Hip and knee arthroplasty implants contraindicated in obesity

JD Craik 1, MD Bircher 1, M Rickman 1
PMCID: PMC5227026  PMID: 27023636

Abstract

Introduction

High patient weight is a risk factor for mechanical implant failure and some manufacturers list obesity as a contraindication for implant use. We reviewed data from the 2012–2013 UK National Joint Registry to determine whether surgical practice reflects these manufacturer recommendations.

Methods

The product literature for the most commonly used hip and knee implants was reviewed for recommendations against use in obese patients (body mass index [BMI] ≥ 30kg/m2). The total number of obese patients undergoing hip and knee arthroplasty was calculated, as was the proportion receiving implants against manufacturer recommendations.

Results

Out of 200,054 patient records, 147,691 (74%) had a recorded BMI. The mean BMI for patients undergoing primary total hip arthroplasty was 29kg/m2, compared with 31kg/m2 for total knee arthroplasty. Of the 25 components reviewed, 5 listed obesity as a contraindication or recommended against implant use in obese patients. A total of 10,745 patients (16% of all obese patients) received implants against manufacturer recommendations.

Conclusions

A high proportion of patients are receiving implants against manufacturer recommendations. However, there are limitations to using BMI for stratifying risk of implant fatigue failure and manufacturers should therefore provide more detailed guidelines on size specific implant load limits to facilitate surgical decisions.

Keywords: Hip, Knee, Arthroplasty, Obesity, Prosthesis, Implant


Obesity is a risk factor for developing osteoarthritis of the hip and knee.1 Defined as a body mass index (BMI) of ≥30kg/m2, the prevalence of obesity has doubled in many European countries in the last 20 years.2 In 2013, 26% of the UK adult population were classified as obese.2 This trend is reflected in National Joint Registry (NJR) figures where the mean patient BMI continues to rise.3

The longevity of total hip replacement (THR) and total knee replacement (TKR) is determined by a combination of patient factors, surgical technique and implant design. In the presence of obesity, the risks of surgery are known to be greater.4,5 Heavier patients also exert greater loads on implanted prostheses. It is therefore crucial that all orthopaedic implant designs are tested thoroughly under appropriate loading conditions prior to general market availability and that manufacturers inform surgeons of any foreseen limitations in the target population.

Some manufacturers list ‘obesity’ as a contraindication for implant use. This potentially has significant implications regarding the responsibility of surgeons in cases of mechanical implant failure if these implants are used in this patient population. We are aware of one current case of litigation concerning femoral stem fatigue failure in an obese patient where the manufacturer has denied liability on these grounds. In light of the potential magnitude of this problem, the aim of this study was to determine whether surgical practice reflects manufacturer recommendations in obese patients.

Methods

NJR data for 2012–2013 were analysed to identify the most commonly used primary THR and TKR components. The product literature (instructions for use) for the top five primary cemented and uncemented femoral stems and acetabular cups, together with the top five primary TKR components, was requested from the manufacturers and reviewed to ascertain the recommendations for component use in obese patients (BMI ≥ 30kg/m2).

Results

The most commonly implanted primary THR and TKR components are shown in Table 1. Out of 200,054 patient records, 147,691 (74%) had a recorded BMI. The mean BMI for patients undergoing primary THR was 29kg/m2, compared with a BMI of 31kg/m2 for primary TKR. Of the 25 components reviewed, 5 listed obesity as a contraindication or recommended against implant use in obese patients and 4 reported morbid or pathological obesity as a contraindication. Of the remaining implants, all advised caution or patient counselling with regard to the risks of implant use in heavy, overweight or obese patients. Of all NJR data records reviewed with a recorded BMI of ≥30kg/m2, 10,745 patients (16%) received implants against manufacturer recommendations (Table 2).

Table 1.

The most commonly used hip and knee arthroplasty components and manufacturer recommendations for patients with obesity/excessive weight

Implant Manufacturer Recommendation Instructions for use reference
Cemented femoral stems Exeter™ V40™ Stryker Warning in heavy/obese patients 96E112 Rev. G
CPT® Zimmer Warning in heavy patients 87–6203–912–22
C-Stem® AMT DePuy Caution in obesity or excessive patient weight IFU-78410023 Rev. E
C-Stem® DePuy Caution in obesity or excessive patient weight IFU-78410023 Rev. E
CPCS Smith & Nephew Contraindication – Morbid obesity (relative or absolute) 81078790 Rev. C
Uncemented femoral stems Corail® DePuy Caution in obesity or excessive patient weight W90942 Rev. E
Furlong® JRI Contraindication – Obesity 115–029 Issue 13
Taperloc® Biomet Relative contraindication – Morbid obesity 5401000158 Rev. 10
Accolade® Stryker Contraindication – Obesity 0095–3-200V
M/L Taper Zimmer Warning in heavy patients 87–6203–501–22 Rev. E
Cemented acetabular cups Contemporary Stryker Warning in heavy/obese patients 96E112 Rev. G
Exeter® RimFit™ Stryker Warning in heavy/obese patients 96E112 Rev. G
Marathon™ DePuy Caution in obesity or excessive patient weight IFU-0902–00–701 Rev. N
Elite Plus™ Ogee® DePuy Caution in obesity or excessive patient weight IFU-0902–00–701 Rev. N
Elite Plus™ DePuy Caution in obesity or excessive patient weight IFU-0902–00–701 Rev. N
Uncemented acetabular cups Pinnacle® DePuy Caution in obesity or excessive patient weight IFU-0902–00–701 Rev. N
Trident® Stryker Warning – Do not implant in obese patients QIN 4350 Rev. G
CSF Plus JRI Contraindication – Obesity 155–019 Issue 14
Exceed ABT™ Biomet Relative contraindication – Morbid obesity 5401000427 Rev. 07
Trilogy® Zimmer Counselling for heavy patients 87–6203–367–22
Total knee arthroplasty PFC® Sigma® DePuy Caution in obesity or excessive patient weight IFU-0902–00–252 Rev. K
NexGen® Zimmer Counselling in heavy patients 87–6203–453–23
Triathlon® Stryker Contraindication – Obesity QIN4376 Rev. G
Vanguard Biomet Contraindication – Pathological obesity 0902–00–077D
Genesis™ II Smith & Nephew Contraindicated in conditions that tend to place increased loads on implants such as weight, which are incompatible with a satisfactory long-term result 81074901 Rev. O

Table 2.

National Joint Registry figures for the most commonly used hip and knee arthroplasty components. Dark shaded rows indicate implants that list obesity as a contraindication or recommend against implant use in obese patients.

Implant Number of procedures Number with recorded BMI Number with BMI =30kg/m2 Mean BMI
Cemented femoral stems Exeter™ V40™ 30,762 23,120 8,654 28.5kg/m2
CPT® 5,574 4,038 1,536 28.6kg/m2
C-Stem® AMT 2,835 1,780 671 28.5kg/m2
C-Stem® 1,790 839 313 28.5kg/m2
CPCS 666 581 202 28.2kg/m2
Uncemented femoral stems Corail® 18,507 13,380 5,540 29.0kg/m2
Furlong® 3,565 2,621 1,117 29.2kg/m2
Taperloc® 3,250 2,372 1,001 29.1kg/m2
Accolade® 2,672 2,088 851 28.9kg/m2
M/L Taper 1,373 1,064 440 28.9kg/m2
Cemented acetabular cups Contemporary 10,327 8,048 2,971 28.4kg/m2
Exeter® RimFit™ 4,338 3,343 1,275 28.7kg/m2
Marathon™ 3,653 2,230 880 28.8kg/m2
Elite Plus™ Ogee® 3,309 2,462 902 28.4kg/m2
Elite Plus™ 1,682 1,274 467 28.3kg/m2
Uncemented acetabular cups Pinnacle® 17,701 12,555 5,153 29.0kg/m2
Trident® 10,861 7,987 3,146 28.7kg/m2
CSF Plus 3,759 2,710 1,132 29.1kg/m2
Exceed ABT™ 3,624 2,713 1,146 29.1kg/m2
Trilogy® 3,329 2,685 1,010 28.5kg/m2
Total knee arthroplasty PFC® Sigma® 26,899 19,921 11,384 31.0kg/m2
NexGen® 13,733 10,058 5,773 31.1kg/m2
Triathlon® 10,865 8,209 4,499 30.8kg/m2
Vanguard 8,000 6,171 3,423 31.0kg/m2
Genesis™ II 6,980 5,442 3,017 30.7kg/m2

BMI = body mass index

Discussion

There are conflicting conclusions in the literature regarding the effect of obesity on functional outcomes and complication rates following primary THR and TKR. Some authors have highlighted that obesity does not have a negative impact on functional improvement6–10 whereas others emphasise that obese patients have poorer postoperative function, quality of life and satisfaction.5,11–18 Similarly, although some studies have been unable to identify a higher rate of complications in obese patients,6,8,19–21 others have observed significant increases in operative time and postoperative infection,5,12,16,22–31 THR dislocation rates,5,13,25,26,32 TKR patellofemoral symptoms33,34 and venous thromboembolism.24,32,35 Obesity may also be associated with poorer long-term implant survival11,16,24,36 and greater overall cost.37,38

Fatigue failure of THR femoral stems is a rare complication, usually as a consequence of excessive stress on the tensile surface of the implant.39–42 Risk factors include high patient weight, greater level of activity, undersized implants, lack of proximal support, varus or retroverted orientation and the presence of metallurgic defects or other stress risers.39–41,43–46 Although metallurgic defects appear to be the main cause of reduced fatigue strength, cases of failure in metallurgic normal stems do occur, prompting some authors to warn against use of small implants in patients with a raised BMI.42,47,48

Of the 25 THR and TKR components reviewed in this study, 5 listed obesity as a contraindication or recommended against implant use in obese patients. With more than 10,000 patients receiving implants against manufacturer recommendations each year, these results suggest that surgeons are either unaware of the listed contraindications for some implants or knowingly use these implants against manufacturers’ advice. In doing so, the surgeon may be adopting responsibility for the consequences of mechanical implant failure. By using the word ‘caution’ or ‘warning’, 15 of the remaining 20 implants still imply that there is a specific risk associated with use in the presence of obesity. Only 5 of the 25 implants reviewed appear not to warn against use in ‘obese’ patients.

This paper serves to highlight a number of important issues. First, in relation to implant failure, the use of BMI (or any terms of reference based on it such as ‘obesity’) is inherently flawed. The fatigue strength of any metallic construct is determined by the magnitude of internal stresses and number of load cycles. The internal stresses of a femoral stem are influenced by both the magnitude of the force vector (ie body weight) and its distance from the centre of rotation (moment arm). A patient of 1.67m and weighing 84kg is obese according to BMI. However, a patient of 1.85m and weighing 100kg has a BMI of <30kg/m2 and is therefore not classified as obese. If the moment arm, implant material and dimensions are the same for each patient, then the stem internal stresses will clearly be greater for the taller patient despite not being obese.

Second, by using obesity as a warning or contraindication for implant use, should we question the standards against which available implants are being tested? It would appear that if a surgeon is going to abide by manufacturer recommendations, then very few commonly used implants are as safe in the overweight population as they are in the remainder2. Is this really true? Given the steadily increasing size of this at-risk group, should there be specific implants for patients with certain dimensions and how should this population be defined? Certainly not by BMI or by the use of the term ‘obese’.

Finally, should we as surgeons be informing patients of these warnings and the use of implants against manufacturer recommendations or should we simply not perform hip and knee arthroplasty in this population until better implants are available? It remains the case that implant selection is not straightforward, especially for new surgeons, and this choice is affected significantly by both training and the media.

With regard to the litigation risk associated with the use of these implants against manufacturer recommendations, this is different in the National Health Service (NHS) setting and in the private sector. In the NHS, there is no contractual relationship between patient and surgeon, with care being provided free as a public statutory obligation. The duty of care in that setting is based on common law, a breach of which gives rise to a claim of negligence. In the private sector, however, the patient has a contract for the provision of services and can therefore also sue for a breach of this contract. If a prosthesis has been used, there is a statutory implied term that the goods will be fit for purpose.

An area that is less clear is the recent emergence of ‘fee assured’ consultants, where patients are directed to surgeons via an insurance company. In this situation, the contractual arrangement is unclear and so much of the contractual liability may lie with the insurance company involved.

One important limitation of this study is the possibility of selection bias leading to overestimation of the proportion of patients who are obese, and therefore the mean BMI of patients undergoing hip and knee arthroplasty. The NJR editorial board acknowledges that ‘all BMI data has to be viewed with caution as surgeons may be more likely to enter BMI data when the BMI is high’.3 Nevertheless, given that only 74% of NJR records reviewed had a recorded BMI, the absolute number of obese patients receiving implants against manufacturer recommendations is likely to be greater. Further studies surveying surgeon awareness of these recommendations and investigating failure rates in relation to patient mass rather than BMI would be of interest.

Conclusions

Advances in lower limb arthroplasty for heavier patients should be a combined effort between surgeons and industry to ensure that treatments are robust and safe. Manufacturers update their instructions for use regularly and many acknowledge the reduction in fatigue strength associated with particular implant geometry, such as smaller femoral stem size and increasing offset. Implant failure will remain a potential complication following arthroplasty, influenced by numerous factors including patient habitus. However, BMI is not the optimal method for stratifying risk of implant fatigue failure, and we would welcome more detailed guidelines on size specific implant load limits to facilitate surgical decisions and preoperative patient counselling. Surgeons must consider the risks associated with arthroplasty procedures in larger patients, and we should ensure that patients are fully informed of all risks involved to protect ourselves from potential litigation.

Acknowledgement

This publication is based on data collected by or on behalf of the Healthcare Quality Improvement Partnership, which has no responsibility or liability for the accuracy, currency, reliability and/or correctness of this publication.

References

  • 1.Richmond SA, Fukuchi RK, Ezzat A et al. Are joint injury, sport activity, physical activity, obesity, or occupational activities predictors for osteoarthritis? A systematic review. J Orthop Sports Phys Ther 2013; : 515–B19. [DOI] [PubMed] [Google Scholar]
  • 2.Health and Social Care Information Centre Statistics on Obesity, Physical Activity and Diet: England, 2013. London: HSCIC; 2013. [Google Scholar]
  • 3.National Joint Registry for England and Wales 10th Annual Report. Hemel Hempstead: NJR; 2013. [Google Scholar]
  • 4.Wallace G, Judge A, Prieto-Alhambra D et al. The effect of body mass index on the risk of post-operative complications during the 6 months following total hip replacement or total knee replacement surgery. Osteoarthritis Cartilage 2014; : 918–927. [DOI] [PubMed] [Google Scholar]
  • 5.Liu W, Wahafu T, Cheng M et al. The influence of obesity on primary total hip arthroplasty outcomes: a meta-analysis of prospective cohort studies. Orthop Traumatol Surg Res 2015; : 289–296. [DOI] [PubMed] [Google Scholar]
  • 6.McLaughlin JR, Lee KR. The outcome of total hip replacement in obese and non-obese patients at 10- to 18-years. J Bone Joint Surg Br 2006; : 1,286–1,292. [DOI] [PubMed] [Google Scholar]
  • 7.Bourne RB, McCalden RW, MacDonald SJ et al. Influence of patient factors on TKA outcomes at 5 to 11 years followup. Clin Orthop Relat Res 2007; : 27–31. [DOI] [PubMed] [Google Scholar]
  • 8.Andrew JG, Palan J, Kurup HV et al. Obesity in total hip replacement. J Bone Joint Surg Br 2008; : 424–429. [DOI] [PubMed] [Google Scholar]
  • 9.Baker P, Petheram T, Jameson S et al. The association between body mass index and the outcomes of total knee arthroplasty. J Bone Joint Surg Am 2012; : 1,501–1,508. [DOI] [PubMed] [Google Scholar]
  • 10.Baker P, Muthumayandi K, Gerrand C et al. Influence of body mass index (BMI) on functional improvements at 3 years following total knee replacement: a retrospective cohort study. PLoS One 2013; : e59079. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Foran JR, Mont MA, Etienne G et al. The outcome of total knee arthroplasty in obese patients. J Bone Joint Surg Am 2004; : 1,609–1,615. [DOI] [PubMed] [Google Scholar]
  • 12.Mulhall KJ, Ghomrawi HM, Mihalko W et al. Adverse effects of increased body mass index and weight on survivorship of total knee arthroplasty and subsequent outcomes of revision TKA. J Knee Surg 2007; : 199–204. [DOI] [PubMed] [Google Scholar]
  • 13.Davis AM, Wood AM, Keenan AC et al. Does body mass index affect clinical outcome post-operatively and at five years after primary unilateral total hip replacement performed for osteoarthritis? J Bone Joint Surg Br 2011; : 1,178–1,182. [DOI] [PubMed] [Google Scholar]
  • 14.Zhang ZJ, Zhao XY, Kang Y et al. The influence of body mass index on life quality and clinical improvement after total hip arthroplasty. J Orthop Sci 2012; : 219–225. [DOI] [PubMed] [Google Scholar]
  • 15.Järvenpää J, Kettunen J, Soininvaara T et al. Obesity has a negative impact on clinical outcome after total knee arthroplasty. Scand J Surg 2012; : 198–203. [DOI] [PubMed] [Google Scholar]
  • 16.Kerkhoffs GM, Servien E, Dunn W et al. The influence of obesity on the complication rate and outcome of total knee arthroplasty. J Bone Joint Surg Am 2012; : 1,839–1,844. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Liljensøe A, Lauersen JO, Søballe K, Mechlenburg I. Overweight preoperatively impairs clinical outcome after knee arthroplasty. Acta Orthop 2013; : 392–397. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Murgatroyd SE, Frampton CM, Wright MS. The effect of body mass index on outcome in total hip arthroplasty: early analysis from the New Zealand Joint Registry. J Arthroplasty 2014; : 1,884–1,888. [DOI] [PubMed] [Google Scholar]
  • 19.Haverkamp D, de Man FH, de Jong PT et al. Is the long-term outcome of cemented THA jeopardized by patients being overweight? Clin Orthop Relat Res 2008; : 1,162–1,168. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Patel AD, Albrizio M. Relationship of body mass index to early complications in knee replacement surgery. Arch Orthop Trauma Surg 2008; : 5–9. [DOI] [PubMed] [Google Scholar]
  • 21.Suleiman LI, Ortega G, Ong’uti SK et al. Does BMI affect perioperative complications following total knee and hip arthroplasty? J Surg Res 2012; : 7–11. [DOI] [PubMed] [Google Scholar]
  • 22.Winiarsky R, Barth P, Lotke P. Total knee arthroplasty in morbidly obese patients. J Bone Joint Surg Am 1998; : 1,770–1,774. [DOI] [PubMed] [Google Scholar]
  • 23.Namba RS, Paxton L, Fithian DC, Stone ML. Obesity and perioperative morbidity in total hip and total knee arthroplasty patients. J Arthroplasty 2005; (Suppl 3): 46–50. [DOI] [PubMed] [Google Scholar]
  • 24.Amin AK, Clayton RA, Patton JT et al. Total knee replacement in morbidly obese patients. J Bone Joint Surg Br 2006; : 1,321–1,326. [DOI] [PubMed] [Google Scholar]
  • 25.Lübbeke A, Stern R, Garavaglia G et al. Differences in outcomes of obese women and men undergoing primary total hip arthroplasty. Arthritis Rheum 2007; : 327–334. [DOI] [PubMed] [Google Scholar]
  • 26.Chee YH, Teoh KH, Sabnis BM et al. Total hip replacement in morbidly obese patients with osteoarthritis. J Bone Joint Surg Br 2010; : 1,066–1,071. [DOI] [PubMed] [Google Scholar]
  • 27.Jämsen E, Nevalainen P, Eskelinen A et al. Obesity, diabetes, and preoperative hyperglycemia as predictors of periprosthetic joint infection: a single-center analysis of 7181 primary hip and knee replacements for osteoarthritis. J Bone Joint Surg Am 2012; : e101. [DOI] [PubMed] [Google Scholar]
  • 28.Wang JL, Gadinsky NE, Yeager AM et al. The increased utilization of operating room time in patients with increased BMI during primary total hip arthroplasty. J Arthroplasty 2013; : 680–683. [DOI] [PubMed] [Google Scholar]
  • 29.Raphael IJ, Parmar M, Mehrganpour N et al. Obesity and operative time in primary total joint arthroplasty. J Knee Surg 2013; : 95–99. [DOI] [PubMed] [Google Scholar]
  • 30.Liabaud B, Patrick DA, Geller JA. Higher body mass index leads to longer operative time in total knee arthroplasty. J Arthroplasty 2013; : 563–565. [DOI] [PubMed] [Google Scholar]
  • 31.Pruzansky JS, Bronson MJ, Grelsamer RP et al. Prevalence of modifiable surgical site infection risk factors in hip and knee joint arthroplasty patients at an urban academic hospital. J Arthroplasty 2014; : 272–276. [DOI] [PubMed] [Google Scholar]
  • 32.Haverkamp D, Klinkenbijl MN, Somford MP et al. Obesity in total hip arthroplasty – does it really matter? Acta Orthop 2011; : 417–422. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Stern SH, Insall JN. Total knee arthroplasty in obese patients. J Bone Joint Surg Am 1990; : 1,400–1,404. [PubMed] [Google Scholar]
  • 34.Healy WL, Wasilewski SA, Takei R, Oberlander M. Patellofemoral complications following total knee arthroplasty. Correlation with implant design and patient risk factors. J Arthroplasty 1995; : 197–201. [DOI] [PubMed] [Google Scholar]
  • 35.Mantilla CB, Horlocker TT, Schroeder DR et al. Risk factors for clinically relevant pulmonary embolism and deep venous thrombosis in patients undergoing primary hip or knee arthroplasty. Anesthesiology 2003; : 552–560. [DOI] [PubMed] [Google Scholar]
  • 36.McElroy MJ, Pivec R, Issa K et al. The effects of obesity and morbid obesity on outcomes in TKA. J Knee Surg 2013; : 83–88. [DOI] [PubMed] [Google Scholar]
  • 37.Kim SH. Morbid obesity and excessive hospital resource consumption for unilateral primary hip and knee arthroplasty. J Arthroplasty 2010; : 1,258–1,266. [DOI] [PubMed] [Google Scholar]
  • 38.Dowsey MM, Liew D, Choong PF. Economic burden of obesity in primary total knee arthroplasty. Arthritis Care Res 2011; : 1,375–1,381. [DOI] [PubMed] [Google Scholar]
  • 39.Woolson ST, Milbauer JP, Bobyn JD et al. Fatigue fracture of a forged cobalt-chromium-molybdenum femoral component inserted with cement. A report of ten cases. J Bone Joint Surg Am 1997; : 1,842–1,848. [DOI] [PubMed] [Google Scholar]
  • 40.Della Valle AG, Becksaç B, Anderson J et al. Late fatigue fracture of a modern cemented forged cobalt chrome stem for total hip arthroplasty: a report of 10 cases. J Arthroplasty 2005; : 1,084–1,088. [DOI] [PubMed] [Google Scholar]
  • 41.Lam LO, Stoffel K, Kop A, Swarts E. Catastrophic failure of 4 cobalt-alloy Omnifit hip arthroplasty femoral components. Acta Orthop 2008; : 18–21. [DOI] [PubMed] [Google Scholar]
  • 42.Ishaque BA, Stürz H, Basad E. Fatigue fracture of a short stem hip replacement: a failure analysis with electron microscopy and review of the literature. J Arthroplasty 2011; : 665.e17–665.e20. [DOI] [PubMed] [Google Scholar]
  • 43.Galante JO. Causes of fractures of the femoral component in total hip replacement. J Bone Joint Surg Am 1980; : 670–673. [PubMed] [Google Scholar]
  • 44.Chao EY, Coventry MB. Fracture of the femoral component after total hip replacement. An analysis of fifty-eight cases. J Bone Joint Surg Am 1981; : 1,078–1,094. [PubMed] [Google Scholar]
  • 45.Wroblewski BM. Fractured stem in total hip replacement. A clinical review of 120 cases. Acta Orthop Scand 1982; : 279–284. [DOI] [PubMed] [Google Scholar]
  • 46.Kishida Y, Sugano N, Ohzono K et al. Stem fracture of the cementless spongy metal Lübeck hip prosthesis. J Arthroplasty 2002; : 1,021–1,027. [DOI] [PubMed] [Google Scholar]
  • 47.Morgan-Hough CV, Tavakkolizadeh A, Purkayastha S. Fatigue failure of the femoral component of a cementless total hip arthroplasty. J Arthroplasty 2004; : 658–660. [DOI] [PubMed] [Google Scholar]
  • 48.Harvie P, Haroon M, Henderson N, El-Guindi M. Fracture of the hydroxyapatite-ceramic-coated JRI-Furlong femoral component. J Bone Joint Surg Br 2007; : 742–745. [DOI] [PubMed] [Google Scholar]