Abstract
With every breath, the airways within the lungs are strained. This periodic stretching is thought to play an important role in determining airway caliber in health and disease. Particularly, deep breaths can mitigate excessive airway narrowing in healthy subjects, but this beneficial effect is absent in asthmatics, perhaps due to an inability to stretch the airway smooth muscle (ASM) embedded within an airway wall. The heterogeneous composition throughout an airway wall likely modulates the strain felt by the ASM but the magnitude of ASM strain is difficult to measure directly. In this study, we optimized a finite element image registration method to measure the spatial distribution of displacements and strains throughout an airway wall during pressure inflation within the physiological breathing range before and after induced narrowing with acetylcholine (ACh). The method was shown to be repeatable, and displacements estimated from different image sequences of the same deformation agreed to within 5.3 μm (0.77%). We found the magnitude and spatial distribution of displacements were radially and longitudinally heterogeneous. The region in the middle layer of the airway experienced the largest radial strain due to a transmural pressure (Ptm) increase simulating tidal breathing and a deep inspiration (DI), while the region containing the ASM (i.e., closest to the lumen) strained least. During induced narrowing with ACh, we observed temporal longitudinal heterogeneity of the airway wall. After constriction, the displacements and strain are much smaller than the relaxed airway and the pattern of strains changed, suggesting the airway stiffened heterogeneously.
Keywords: Elastography, finite element method, image registration, airway smooth muscle, deep inspiration, bronchodilation
1. Introduction
The lungs and branching network of airways are periodically stretched with every breath. The luminal diameter of an airway, which determines the resistance to flow, is believed to be determined as a dynamic equilibrium of the forces which favor narrowing and those that oppose narrowing [1]. Bronchoconstriction, as occurred during an asthma attack, is a result of active force generated by airway smooth muscle (ASM) embedded with an airway wall in response to acetylcholine (ACh) released from parasympathetic nerves. The forces of tidal breathing and deep inspirations (DIs) may work to mitigate this airway narrowing in healthy subjects through straining of the ASM [2]. This beneficial effect of a DI is absent in asthmatics [3–6] and it has been hypothesized that this is because asthmatic are unable to sufficiently strain their ASM during a DI which is indicative of airway wall properties that are stiffer and that can potentiate more reactivity.
While most researchers have focused on the effect of stretching of strips of isolated ASM, the airway wall is a construct comprised of several different components which together create a nonlinear, heterogeneous, thick-walled cylinder. In contrast to the results from isolated ASM strip experiments, recent studies performed on healthy airway segments suggest that transmural pressures simulating tidal breathing and DIs are ineffective at preventing future airway narrowing and only moderately effective at reversing induced constriction [7–10]. The heterogeneous composition and mechanical properties throughout an airway wall likely modulate the strain felt by the ASM. Therefore, understanding the spatial distribution of displacements and strains throughout the thickness of an airway wall could lead to better understanding of airway narrowing by elucidating contributions of different tissue types to the mechanical behavior of the airway, and the role that breathing and DIs play in modulating airway narrowing.
Elastography is a technique to image tissue based on contrast in mechanical properties [11]. In quasistatic elastography, a tissue is imaged before and after a force is applied and the resulting images are compared to estimate tissue deformation. These deformations are then visualized directly (e.g., strain imaging) or used to infer the mechanical properties of the tissue via modulus reconstructions [11–13]. The technique has been applied in a variety of tissues using many different image modalities such as ultrasound [14], magnetic resonance [15], and optical coherence tomography [16]. In this study, ultrasound imaging was utilized due to its high axial resolution and phase information, high frame rate, and availability. Elastography has many different applications [17] such as detecting breast tumors [18] and predicting plaques vulnerable for rupture in atherosclerosis [19,20], but has never been used to study the spatial distribution of displacements throughout an airway wall.
Many different approaches have been developed to estimate tissue motion from ultrasound image sequences. Among the most common is block matching using normalized cross-correlation [11]. In this approach, the motion within each block of pixels is assumed to be uniform. While the block matching method is computationally inexpensive and easy to implement, it performs poorly at estimating displacements in regions of highly heterogeneous deformation or large strains [21,22], it has recognized biases when estimating sub-sample displacements [23,24], and incorporating prior information through regularization in a systematic way is cumbersome. In this study we utilized a finite element based image registration approach similar to the deformable grid methods [25–28]. These methods easily accommodate arbitrary geometries, allow for large local strain within an element, and accommodate prior information through regularization in a natural way. They have the drawback that they are comparatively computationally costly, which for the present application was not a concern.
In this study, we optimized our finite element image registration method to measure the spatial distribution of displacements and strains throughout an airway wall during pressure inflation within the physiological breathing range, as well as monitor tissue motions during induced narrowing. To measure large deformation over the entire physiological range of transmural pressure (Ptm), we incrementally match images captured over small changes in Ptm. We estimate the strain distributions throughout the airway walls which are used to gain insight into the relative moduli of the different airway wall components. We found the displacements and strain during inflation to be longitudinally and radially heterogeneous. The region in the middle layer of the airway experienced the largest radial strain due to a Ptm increase simulating tidal breathing and a DI, while the region containing the ASM (i.e., closest to the lumen) strained least. During induced narrowing with ACh, (10−3 mol/L), we observed temporal longitudinal heterogeneity of the airway wall. Equal pressure increments lead to much smaller strain and displacements in the constricting airway than in the relaxed airway. The spatial patterns of displacements and strains also changed, suggesting the components of the airway did stiffened heterogeneously.
2. Material and Methods
2.1. Intact Airway System
An intact airway (main stem bronchus, generations 4-10, 35 mm long) was dissected from a fresh bovine lung (Research 87, Boylston, MA) and side branches were ligated to form a leak-free airway. Cannulas were inserted into each end and the airway was mounted inside a tissue bath with heated (37°C, 5% CO2) Krebs solutions (Sigma, St. Louis, MO). The entire setup was secured to a floating optical table (Newport I-2000/S-2000) to isolate the system from building vibrations and increase displacement SNR. The airway was stretched to 120% of its resting length to mimic airway lengthening during breathing and its length was held fixed throughout the experiment [29]. The pressure inside the airway was controlled by modulating the height of fluid in a pressure column in series with the airway [30]. Following 60 min in the heated bath, tissue viability was confirmed with electric field stimulation (EFS) as previously described [8,30,31].
2.2. Ultrasound Data Acquisition
Ultrasound radio frequency (RF) data was collected and digitized to 16-bits at 40 MHz using an ultrasound system (SonixTablet, Ultrasonix Medical Corporation, Richmond, BC, Canada). A 128 element linear array transducer (L40-8/12, 12 mm width) was partially submerged in the tissue bath and mounted less than 7 mm above the outer edge of the airway and positioned parallel to the long axis of the airway segment (Fig. 1). The imaging plane of the transducer cuts through the middle of the airway along its length and the region was more than two diameters away from the proximal and distal cannulas to minimize end effects. This transducer orientation allowed for the dominant displacement component to be aligned with the direction of ultrasound propagation. The field of view in the ultrasound axial direction (hereafter the “axial” direction) and ultrasound lateral direction (hereafter the “lateral” direction) was 22 mm and 12 mm, respectively, and the axial and lateral pixel sizes were 18.7 μm and 46.9 μm, respectively. The resolution in the axial direction was approximately 103 μm. The transducer had a center frequency of 15 MHz and −6 dB bandwidth from 10.5-19.5 MHz. The RF data was upsampled by a factor of two using the fast Fourier transform interpolation method in order to improve the discrete integration of the image during processing.
Figure 1.
Intact airway mounted in the tissue bath with high frequency ultrasound transducer mounted above.
In order to measure the displacement distributions throughout an airway wall, we collected ultrasound RF image data as the airway Ptm was incremented throughout the entire physiological range (i.e., −10 to 25 cmH2O). Data was collected in increments of 0.5 cmH2O, except in the range of Ptm where airways are most compliant (i.e. −5 to 10 cmH2O), where data was collected in increments of 0.2 cmH2O. The airway was held for approximate 5 s at each Ptm before data was captured. Two full cycles were recorded to verify measurement repeatability. Before data collection, the Ptm was slowly cycled between −10 and 25 cmH2O at a constant rate of 1 cmH2O to normalize for volume history. We repeated this process after the airway was constricted with an ASM agonist (ACh, 10−3 mol/L) in order to assess the changes that occur in displacement distributions of the airway walls due to ASM activation.
To induce airway narrowing akin to what occurs during an asthma attack, the airway was exposed to cumulative doses of ACh (10−5, 10−4, and 10−3 mol/L) and allowed to narrow against a constant Ptm of 5 cmH2O for 10 min at each dose. During the constriction, ultrasound data was collected at a frame rate of 1.5 Hz in order to visualize narrowing dynamics.
During tidal breathing, Ptm fluctuate between approximately 5 cmH2O at end-expiration and 10 cmH2O at end-inspiration. During DIs, which people naturally take every 6 min [32], Ptm increases to approximately 25 cmH2O.
2.3. Displacement Estimation
2.3.1 Edge detection and mesh generation
An automated segmentation algorithm developed in MATLAB (MathWorks, Natick, MA, USA) was used to determine the location of the airway walls, as previously described [7]. RF data was first converted to B-Mode [33] through FIR filtering, envelope detection, non-linear compression, and scan conversion. The airway walls were then segmented from the Krebs solution by thresholding the image [34], followed by morphological operations (e.g. smoothing) to form contiguous walls with smooth boundaries.
An unstructured, quadrilateral finite element mesh was then generated from approximately equally spaced (in real space) nodes on the identified airway boundary for both airway walls [35].
2.3.2 Image registration using finite elements
The recorded RF ultrasound data, RFt(t,n), is a function of echo transit time, t, and the line number, n. This data is scan converted to physical Cartesian coordinates, based on the speed of sound, c, and the distance between RF lines, dy
| (1) |
Spatial deformation distributions through the airway walls were estimated using a finite element image registration technique adapted from the method described by Richards et al. [26]. The deformation required to map the reference image of the tissue, I1, and the image after a mechanical perturbation (i.e., a small change in Ptm) is assumed to result directly from underlying tissue motion, u(X). For clarity, we define X(x,y) as a two dimensional space position :
| (2) |
To find u(X), the displacement vector of the tissue originally at spatial location X, the following functional was minimized using the Gauss-Newton method
| (3) |
The first term of the functional, Fmismatch, minimizes the difference between motion compensated image pairs over the spatial domain of interest. The second term, Freg, is a H1-seminorm regularization term which penalizes large gradients in u(X). The target strain in the axial direction, εt, is a constant used to correct for the expected gradient of u(X). εt is estimated for each pair of images as the mean difference between the displacements of the outer and inner walls, divided by the mean wall thickness. The strength of regularization is determined by the parameter, α; the selection of α is described below.
The function u(X) was discretized using bi-linear finite element interpolation on quadrilateral finite elements. Though u(X) varies linearly over the element, the RF-image intensities may oscillate significantly over an element. Therefore, rather than standard low-order Gauss quadrature over each finite-element, the parent domain of each element was divided into sub-domains. Integration over each sub-domain was performed using 3x3 point Gaussian quadrature. RF images were evaluated at each Gauss integration point in the subdomains using cubic Lagrange polynomial interpolation of the image intensity. The number of element sub-domains was chosen to yield roughly one Gauss point per pixel and exact evaluation of the cubic interpolated image polynomials.
2.3.3 Initial Guess
Since the Gauss-Newton method is an iterative method, an initial guess is required. When matching RF images, we found the results significantly depended on the choice of the initial guess. We found the following two-step procedure gives reliably repeatable results for an image pair. For the first step, an initial guess for was chosen as the spatial linear interpolation of the mean displacements of the inner and outer edge detected walls. This displacement estimate is then refined by registering the envelope detected image data (i.e., (X) = abs(hilbert(RF(X)), where hilbert refers to the Hilbert Transform) according to equation (3). Matching the envelope detected data first helps to avoid local minima when matching RF data. The output of this process was then used as the initial guess for matching of RF(X).
2.3.4 Regularization Strength
The correct value for α depends on the images, SNR, and the magnitude of the displacements. We therefore used the L-curve method to determine the best value of α for our data [36,37]. Specifically, Fmismatch was plotted vs Freg on a log-log plot. The optimal α was chosen at the corner of the resulting L-curve. This value of α represents a balance between the image matching error and regularization (i.e., smoothing) error [38]. The image matching component of the functional was calculated after each iteration to ensure the algorithm converged.
2.3.5 Successive Image Matching and Lagrangian Displacement Estimation
A large deformation can be computed by matching successive pairs of images in a long sequence of images. If in the process of matching these images in the sequence, one uses the same finite element mesh (i.e., one keeps the finite element nodes fixed in space), then the resulting displacement field represents an Eulerian description of the displacement field. We sought to measure the Lagrangian description of the displacement distribution, however, over the full range of Ptm. To do so, we incrementally matched images captured at increasing Ptm. For each displacement increment, the finite element nodes were moved from their original positions to updated positions determined by the estimated displacement. The next two images were then matched on the updated mesh. Thus the finite element mesh followed individual tissue regions throughout the pressure increments. We then calculated the total displacement of each node on the original mesh by simply summing the incremental displacements nodally to thus obtain the Lagrangian description of the total displacement field. In a similar manner, airway narrowing was visualized by matching successive frames after the airway was activated.
2.3.6 Sign Convention and Strain Calculation
Radial outward displacement is defined as positive, while radial inward displacement is defined as negative. In order to gain insight into the mechanical property distribution throughout the airway wall, strain fields in the radial (i.e., y) direction were calculated as using the GradientOfUnstructuredDataSet filter in ParaView [39]. Here, ur is the displacement in the radial direction and R is the radial distance from the axis of the airway. Circumferential strain was calculated as ur/R. We defined compressive strain as negative and tensile strain as positive.
2.3.7 Quality Assurance
Two different metrics of quality assurance were applied to ensure the results were internally consistent. One is a metric that measures the quality of the image registration process which assures that the images are well matched. Having well matched images is necessary, but not sufficient, to having a good displacement field. Therefore, another test was performed to measure displacement accuracy and consistency.
To measure how well the images were matched, the quality of the image registration itself was quantified within each element. To do so, the quantity was computed within each element, e, as the normalized L2 norm of the difference in the motion compensated image pairs
| (4) |
A value of se of 0 corresponds to a perfect matching of the data within an element while values of greater than 1 correspond to poorly matching data. The mean value of se over all elements was confirmed to converge.
A second check is on self-consistency of the measured displacements. To test this, the displacements from matching data at all Ptm increments (i.e. increments of 0.2 cmH2O) were compared to matching every other Ptm increment (i.e. increments of 0.4 cmH2O) in the tidal breathing range. This test assures that registering different images in a sequence gives the same displacement fields.
2.4. Comparison to homogeneous thick-walled cylinder approximation
A stress-strain relationship was estimated from the inflation Ptm-diameter data assuming a homogeneous, thick-walled and incompressible cylinder, and following the technique described in [40]. The Cauchy shear stress (τ) was calculated as
| (5) |
where ri and ro are the inner and outer radii, respectively, at a given Ptm. The Cauchy shear strain (B) was defined as
| (6) |
where ri and ro are the inner and outer radii, respectively, at Ptm = 0, and Ri and Ro are the deformed inner and outer radii, respectively. The stretch ratio in the longitudinal direction, λZ, for an isotropic, incompressible material, is given as . Finally, an effective shear modulus was defined as the ratio of Cauchy shear stress to shear strain
| (7) |
After combining equations (6) and (7), we numerically solve for ri and calculate the radial displacement throughout the thickness of the airway wall, ur, as
| (8) |
where Ri ≤ R ≤ Ro.
To help visualize how the displacement varied throughout the thickness of the walls, each wall was binned into equally spaced distance from the inner wall to the outer wall. The radial displacements in each layer were averaged and compared to the approximations of a homogeneous, thick-walled cylinder.
3. Results
3.1. Quasistatic inflation
Figure 2a shows the mean luminal diameter calculated using the detected edges along the length of the airway as a function of Ptm in both the relaxed (black) and constricted (gray) state. In the relaxed curve, the solid symbols are data from the first recorded cycle while the open symbols are from the second cycle. The solid lines show averaged and smoothed curves. The relaxed airway is most compliant between Ptm of −5 and 10 cmH2O and then becomes very stiff. The constricted airway is much stiffer than the relaxed state between Ptm of −5 and 10 cmH2O, but continues to dilate above a Ptm of 10 cmH2O. Interestingly, the increased stiffness of the constricted airways prevents full closure even at negative Ptm. Fig. 2b shows the Cauchy shear stress as a function of the Cauchy-Green shear strain. The relaxed airway displays a nonlinear stress-strain relationship at high strains, while the constricted airway is stiffer and nearly linear over the range of strains studied. The airway is subjected to the same range of Ptm (i.e., −10 to 25 cmH2O) in both the relaxed and constricted states, but the shear stress (i.e., the non-pressure component of the stress) is much higher in the more compliant, relaxed state than in the stiffer, constricted state.
Figure 2.
(a) Quasistatic Ptm-diameter relationship of an intact airway segment. Ptm is slowly increased from −10 to 25 cmH2O and back to −10 cmH2O twice. Closed circles represent the first recorded loop, open circles the second loop, Gray circles depict the Ptm-diameter relationship after induced narrowing with 10−3 mol/L ACh. Red lines show binned average of the curves (b) Cauchy shear stress-strain relationship computed from the data in (a) for the relaxed (black) and constricted (gray) airway.
Figure 3a-d depicts B-mode ultrasound data collected at four notable Ptm: a. 0 cmH2O (undeformed state), b. 5 cmH2O (Ptm corresponding to functional residual capacity (FRC)), b. 10 cmH2O (Ptm corresponding to end inspiration during tidal breathing), and d. 15 cmH2O (at end inspiration of exaggerated breathing, after the plateau of Ptm-diameter relationship). Videos of B-mode images showing the inflation of the airway throughout the entire Ptm range in the relaxed and constricted states are available as supplementary content. The inner and outer wall edges are detected and overlaid. As the Ptm increases, the luminal diameter increases while the wall thickness decreases since the tissue is nearly incompressible.
Figure 3.
Top: B-mode image of an isolated airway segment at 0 (a), 5 (b), 10 (c), and 15 (d) cmH2O. Middle: Estimated Lagrangian radial displacement of the airway due to an increase in Ptm from 0 up to 5 (f), 10 (g), and 15 (h) cmH2O with deformed meshes (green) overlaid. Panel (e) depicts the initial finite element mesh at 0 cmH2O.
Figure 3e shows the initial mesh on the airway walls at = 0 cmH2O. The mesh used consisted of 250 quadrilateral finite elements in each wall. Each element contained between 50-250 RF pixels. Figure 3f-h show the Lagrangian radial displacement of the airway due to an increase in Ptm from 0 up to 5, 10, and 15 cmH2O, respectively. Overlaid on each image are the deformed meshes with each node moved by its estimated displacement, u(X). The displacements are longitudinally and radially heterogeneous, with the deformation greatest near the inner lumen and decreasing throughout the thickness of the wall. The pattern of deformation is qualitatively similar for the three magnitudes of Ptm.
Figure 4a-d show the radial strain maps as Ptm changes from 0 (Fig. 4a,e) up to 5 (Fig. 4b), 10 (Fig. 4c), and 15 (Fig. 4d). Interestingly, the middle layer of the airway wall appears to compress significantly more that the regions closer to the inner and outer walls, implying a more compliant and highly compressible material. The pattern of strains remains relatively constant at the three different Ptm levels. In the bottom wall, areas of low strain are visible which might correspond to cartilage plates within the wall. Fig. 4e-h show the circumferential strain resulting from the same increases in Ptm from 0 cmH2O. The circumferential strain is highest near the lumen and decreases toward the outside of the wall.
Figure 4.
Radial (top) and circumferential (bottom) strain fields for the undeformed airway (a, e) and due to an increase in Ptm from 0 to 5 (b, f), 10 (c, g), 15 (d, l) cmH2O.
Figure 5 quantifies the radial distribution of displacements throughout the top (black) and bottom (gray) airway walls and compares the results to those predicted by the thick-walled cylinder approximation (equation (8)) for a cylinder with = 0.8 kPa. As expected, the radial displacement decreases with 1/r from the inner to outer wall.
Figure 5.

Binned and averaged radial distribution of displacements throughout the top (black) and bottom (gray) airway walls compared to distribution predicted by a homogeneous thick-walled cylinder with the same (red).
3.2. Displacements and Strains During Breathing
In section 3.1 we calculated the Lagrangian displacements throughout the airway wall relative to the unstressed condition at Ptm = 0 cmH2O. In vivo, the airways are prestressed and breathing occurs above FRC which is typically assumed to be at Ptm = 5 cmH2O. We calculated the displacement and strains in the tidal breathing range from 5 to 10 cmH2O (Fig. 6a,c) and due to an increase in Ptm from 5 to 25 cmH2O simulating a DI (Fig. 6b,d). A video showing the cumulative displacement distributions with deformed meshes (green) overlaid as the Ptm is increased from 5 to 15 cmH2O is available as supplement content.
Figure 6.
Radial displacement fields of a relaxed airway during tidal breathing (i.e., increase in Ptm from 5 to 10 cmH2O) (a) and during a DI (i.e., increase in Ptm from 5 to 25 cmH2O) (b). Corresponding radial strain maps are shown in (c) and (d), respectively.
Figure 7 shows the corresponding displacements and strains of the airway following constriction with ACh (10−3 mol/L). After narrowing, the displacements are much smaller than the relaxed airway and the pattern of displacements has changed, suggesting the airway did not stiffen homogeneously.
Figure 7.
Radial displacement fields of an airway which has been previously constricted with ACh (10−3 mol/L). Displacement distributions resulting from an increase in Ptm corresponding to tidal breathing (i.e., 5 to 10 cmH2O) (a) and due to an increase in Ptm simulating a DI (i.e., 5 to 25 cmH2O) (b). Corresponding radial strain maps are shown in (c) and (d), respectively.
3.3. Airway narrowing
Figure 8a shows the mean luminal diameter of the airway while exposed to increasing doses of ACh. Figure 8b is a detailed view of the airway’s luminal diameter in response to 10−5 mol/L ACh where displacements were tracked. The white circles show frames which were matched which were chosen when the displacement of the luminal walls had moved by more than a chosen threshold (> 2 pixels). The red open circles highlight five time points of interest shown. Figure 9 displays the B-mode (a-e), initial mesh (f), and displacements (g-j) at each of these points of interest. The narrowing is heterogeneous along the length of the airway. In time point II (b,g), the proximal portion begins to narrow. Shortly after, at time point III (c,h), the distal portion of the airway narrows. As the narrowing continues, the constriction proceeds (IV, d,i) as a wave from the distal to proximal side and by time point V, the entire length of the airway narrowed.
Figure 8.
(a) Mean luminal diameter of an airway after exposed to three increasing does of ACh. The gray section in (a) is expanded in (b). Frames that were used for matching are indicated by white circles and red open circles indicate time points of interest shown in Fig. 9.
Figure 9.
B-mode (a-e), initial mesh (f), and displacements (g-j) at each of the points of interest from Fig. 8. The narrowing is heterogeneous along the length of the airway as the constriction proceeds.
3.4. Regularization tuning
Figure 10 shows the L-curve comparing Fmismatch and Freg when the data at 0 cmH2O was matched to 0.2 cmH2O for values of α ranging over three decades. The optimal value of α was chosen as 5.6 × 107. This value was found consistently to be the optimal value for matches at different Ptm and was used for all matches. The displacements calculated when using these values of α were also consistent with the optimal value determined by subjective visual inspection of the displacement maps. The regularization strength in the direction perpendicular to the ultrasound propagation (αx) was set to be 100 times larger than the regularization strength in the direction of propagation (αy) in order to suppress spurious noise in the lateral displacement estimates.
Figure 10.
L-curve analysis used to choose the optimal regularization strength, α. The value of the mismatch term of the function Fmismatch is compared to the regularization term Freg a log-log scale for a wide range of α. The optimal α (red) corresponds to the corner of the resulting curve. Note the reverse direction of the x-axis.
3.5. Quality Assurance
Figure 11 displays the progression of our method. In Fig. 11a, an initial guess based on the displacements of the detected edges is shown. The corresponding map of mean match quality, se (equation (4)), is shown in Fig. 11c, which is averaged for each element over all the matches from Ptm 5 to 10 cmH2O. The final estimate of displacements due to an increase in Ptm from 5 to 10 cmH2O is shown in Fig. 11b, with the corresponding map of se shown in Fig. 11d.
Figure 11.
Progression of image registration process of airway inflated from a Ptm of 5 to 10 cmH2O. The initial guess based on the detected edges (a) results in a very poor image match (i.e., high se) (c). The final displacement estimate field (b) matches the images very well (i.e., low se) (d).
Two tests were performed to ensure our measurements were repeatable. First, two full inflation-deflation loops were recorded for the relaxed airway and consistent results were obtained, as shown in Fig. 2.
Next, in the tidal breathing range of Ptm (5 to 10 cmH2O), ultrasound images were collected in increments of 0.2 cmH2O (25 images total, mean nodal displacement: 780 ± 377 μm). Total displacements were calculated separately by matching all frames (i.e. 5.0, 5.2, 5.4, …, 10.0 cmH2O), matching just the odd frames (i.e., 5.0, 5.4, 5.8, …, 10.0 cmH2O), and matching just the even frames (i.e., 5.0, 5.2, 5.6, …, 10.0 cmH2O). The median absolute difference in total displacements calculated using only odd frames from those calculated using only even frames was 2.8 μm (0.43%) with a range of 0.02 to 79.2 μm (0.001% to 11.0%). Similar results were found comparing the total displacements from all frames to only odd frames (median: 3.8 μm (0.49%)) and all frames to only even frames (median: 5.3 μm (0.77%)). The largest differences tended to be in the regions of lowest signal intensity.
4. Discussion
In this study, we optimized a finite element image registration technique to estimate tissue deformation within an airway wall. By incrementally matching ultrasound data collected over the physiological range of Ptm, the spatial distribution of displacements during normal tidal breathing and during a DI were estimated. The displacements calculated were heterogeneous, both longitudinally and radially through the thickness of the wall. Further, deformations during active airway narrowing were tracked which showed a large amount of temporal longitudinal heterogeneous deformation.
This study aimed to develop a robust approach for investigating of the complex way changes in Ptm imposed to a heterogeneous wall structure translate into strain of the individual components of the airway wall. Subsequent studies will examine the consistency of how airway wall constituents strain locally before and after induced narrowing for similar sized airways. These displacement distributions can then be used as an input to an inverse problem in order to estimate the spatial distribution of material parameters throughout an airway wall.
Of particular interest is the strain experienced by the ASM layer embedded within the wall, since stretching of an isolated strip of ASM causes a dramatic decrease in the amount of force it can generate. For our airway, we found that the region of the wall closest to the inner lumen displaced the most; the displacement magnitude gradually decreased toward the outer wall. This is qualitatively consistent with the prediction of a homogeneous thick walled cylinder (Fig. 5). We found that as Ptm increased, different regions and components of the airway wall (i.e., epithelium, ASM, cartilage plates) experienced different strains. Specifically, we found that a region in the middle layer experienced greater radial strain during inflation than the regions near the inner and outer walls, implying that part of the airway was more compliant than others. The ASM layer is located very close to the inner lumen, directly behind the basement membrane, so is likely not contained in this compliant region of the wall. After induced narrowing (ACh, 10−3 mol/L) the displacements and strains decreased non-uniformly, suggesting the airway stiffened heterogeneously.
We also observed and quantified the temporal longitudinal heterogeneity of the airway wall during induced narrowing (Figure 9). This behavior could be explained by variation in the amount of ASM along the length of the airway. Alternatively, the ASM might be equally distributed along the length but the mechanical properties and relative amounts of other airway components could vary along the length and account for this observation. It is also possible that the ACh is arriving at the ASM at different times along the length, although this is unlikely given the way the agonist is added to the bath.
As with any image registration technique, there are several sources of error which limit the accuracy of our displacement estimates. First, we assume in Equation (2) that the deformation required to map an undeformed image to the deformed image results from in plane tissue motion alone. Out of plane tissue motion and acoustic artifacts such as reverberations could violate this assumption. The steps taken to minimize these effects include proper orientation of the transducer such that most of the movement occurred in the plane, and by collecting data over many small increments of Ptm. The quality assurance results in section 3.5 demonstrate that these steps were adequate.
Next, the highly oscillatory nature of the RF ultrasound data makes finding local minima likely. To help avoid this, we used the movements of the inner and outer wall to aid in the initial guess. We then used the displacements estimated by matching the envelope-detected image data as an initial guess for matching the RF data. Low values of for all elements in the domain confirmed that our solutions avoided local minima.
The accuracy of our results depends on both the noise inherent in the imaging and measurement system, and error introduced by the image registration algorithm. Specifically, the ultrasound system is limited by its resolution and the digitization of the RF data to 16 bit integers. A high resolution transducer was utilized and ultrasound imaging parameters such as line density (256 lines) and foci location (two, one located near each wall) were optimized. We orientated the transducer such that the majority of displacement was in the direction of ultrasound propagation since the axial displacement resolution of ultrasound is significantly better than the lateral resolution. Another source of noise is environmental vibrations which we minimized by collecting data on a floating optical table.
The finite element based image registration technique utilized here is advantageous in that it allows for arbitrary mesh geometries and distortion of elements. The regularization term was necessary to make the optimization problem more well-posed and to help ensure more meaningful displacement estimates. H1 semi-norm regularization typically forces boundary displacement toward a zero strain condition, resulting in errors near the boundaries [41]. We mitigated this error by modifying the regularization term in Equation (3) to include an estimate of the target strain, .
A unique method we implemented in this study was calculating the Lagrangian displacements over a large range of Ptm by matching the ultrasound data over many small Ptm increments and moving the finite element nodes by the estimated displacements of the previous match. Calculations of these large displacements over many small increments significantly increased the SNR of results and may allow for the estimation of nonlinear mechanical properties in the future.
We ensured the quality of our displacement estimates using two metrics. First, we verified that, given the calculated displacement field, the RF data within each element was well-matched between the undeformed and deformed images (see Fig. 11). In addition, we calculated nearly identical results when matching only odd frames compared to matching only even frames over a wide range of Ptm (i.e., 5 to 10 cmH2O). Comparing odd to even frames is a particularly challenging test because almost no data is shared between the two calculations. In particular, only the end frames (i.e., Ptm = 5.0 and 10.0 cmH2O), out of the total of 26 frames are common between the two datasets.
While this study advances a powerful technique to measure the displacements throughout an airway wall, there are limitations that should be noted. First, the airway was slowly inflated in small increments while capturing the ultrasound data. In vivo, however, tidal breathing and DIs occur at a faster rate (i.e., 0.2-0.3 Hz). This different loading rate might result in different deformations due to the tissue’s viscoelastic and active properties. The frame rate of ultrasound is sufficiently high to quantify displacements of an airway during physiological strain rates in a future study.
In order to control for the viscoelastic effects of the tissue, ultrasound images were captured at fixed intervals of 5 s during the inflation of the airway. However, the airway is an active, living tissue. The discontinuities observed in Fig. 2 during the inflation limb of the Ptm-diameter curve are potentially attributable to the intrinsic tone of the ASM. This region of the curve potentially corresponds to the peak of the ASM’s force-length relationship [42].
We oriented the ultrasound transducer parallel to the long axis such that the majority of the displacement during inflation and narrowing was in the direction of ultrasound propagation. Over the large range of Ptm examined, however, some out of plane tissue motion occurred. The elevation focus of our transducer (i.e., in the out of plane direction) is fixed at a depth of 6 mm. We found that positioning the top edge of the airway at a depth slightly below this focus resulted in the best results by capturing a slightly thicker slice without compromising lateral resolution. To further address this limitation, this method can be extended naturally to three dimensions in order to track out of plane motion while also enforcing tissue incompressibility [26].
Finally, while the stiffness of the material can be inferred from strain maps (i.e., Fig. 4), solving the inverse problem with proper boundary conditions and constitutive model is thought to yield more meaningful and accurate results [12]. In a future study, we will formulate and solve this inverse problem in order to reconstruct the distribution of material parameters throughout the airway wall.
5. Conclusions
In this study, we successfully optimized a finite element image registration method to quantify the deformation of airway walls. We found the deformation to be heterogeneous, with the medial layer undergoing significantly higher strains than neighboring layers. The ASM layer, which is close to the lumen, strains less than other components of the airway wall. We found that narrowing of the airway wall with an ASM agonist leads to heterogeneous stiffening of the wall components. More broadly, this study demonstrates that spatial heterogeneity in airway deformation means that different tissue types in the airway experience different mechanical environments.
Supplementary Material
Highlights.
Displacement distribution within airway wall estimated using image registration
Deformation of airway wall during pressure changes of breathing is heterogeneous
Medial layer of airway wall experiences highest strains during breathing
Airway narrows and stiffens heterogeneously in response to an agonist
Acknowledgements
This project was funded by the National Institutes of Health Grants No. R01-CA-140271 and R01-HL-096797 as well as the National Science Foundation Grant No. 1148124.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- [1].Harvey BC, Lutchen KR. Factors Determining Airway Caliber in Asthma. Crit. Rev. Biomed. Eng. 2013;41:515–532. doi:10.1615/CritRevBiomedEng.2014010687. [PubMed] [Google Scholar]
- [2].Fredberg JJ, Inouye D, Miller B, Nathan M, Jafari S, Raboudi SH, Butler JP, Shore SA. Airway smooth muscle, tidal stretches, and dynamically determined contractile states. Am. J. Respir. Crit. Care Med. 1997;156:1752–9. doi: 10.1164/ajrccm.156.6.9611016. [DOI] [PubMed] [Google Scholar]
- [3].Brown RH, Scichilone N, Mudge B, Diemer FB, Permutt S, Togias A. High-resolution computed tomographic evaluation of airway distensibility and the effects of lung inflation on airway caliber in healthy subjects and individuals with asthma. Am. J. Respir. Crit. Care Med. 2001;163:994–1001. doi: 10.1164/ajrccm.163.4.2007119. [DOI] [PubMed] [Google Scholar]
- [4].Fish JE, Ankin MG, Kelly JF, Peterman VI. Regulation of bronchomotor tone by lung inflation in asthmatic and nonasthmatic subjects. J. Appl. Physiol. 1981;50:1079–86. doi: 10.1152/jappl.1981.50.5.1079. [DOI] [PubMed] [Google Scholar]
- [5].Jensen A, Atileh H, Suki B, Ingenito EP, Lutchen KR. Selected contribution: airway caliber in healthy and asthmatic subjects: effects of bronchial challenge and deep inspirations. J. Appl. Physiol. 2001;91:506–515. doi: 10.1152/jappl.2001.91.1.506. discussion 504–505. [DOI] [PubMed] [Google Scholar]
- [6].Kapsali T, Permutt S, Laube B, Scichilone N, Togias A. Potent bronchoprotective effect of deep inspiration and its absence in asthma. J. Appl. Physiol. 2000;89:711–20. doi: 10.1152/jappl.2000.89.2.711. [DOI] [PubMed] [Google Scholar]
- [7].Harvey BC, Parameswaran H, Lutchen KR. Can breathing-like pressure oscillations reverse or prevent narrowing of small intact airways? J. Appl. Physiol. 2015;119:47–54. doi: 10.1152/japplphysiol.01100.2014. doi:10.1152/japplphysiol.01100.2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Harvey BC, Parameswaran H, Lutchen KR. Can tidal breathing with deep inspirations of intact airways create sustained bronchoprotection or bronchodilation? J. Appl. Physiol. 2013;115:436–45. doi: 10.1152/japplphysiol.00009.2013. doi:10.1152/japplphysiol.00009.2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Noble PB, Jones RL, Needi ET, Cairncross A, Mitchell HW, James AL, McFawn PK. Responsiveness of the human airway in vitro during deep inspiration and tidal oscillation. J. Appl. Physiol. 2011;110:1510–1518. doi: 10.1152/japplphysiol.01226.2010. doi:10.1152/japplphysiol.01226.2010. [DOI] [PubMed] [Google Scholar]
- [10].Noble PB, Jones RL, Cairncross A, Elliot JG, Mitchell HW, James AL, McFawn PK. Airway narrowing and bronchodilation to deep inspiration in bronchial segments from subjects with and without reported asthma. J. Appl. Physiol. 2013;114:1460–71. doi: 10.1152/japplphysiol.01489.2012. doi:10.1152/japplphysiol.01489.2012. [DOI] [PubMed] [Google Scholar]
- [11].Ophir J, Céspedes I, Ponnekanti H, Yazdi Y, Li X. Elastography: a quantitative method for imaging the elasticity of biological tissues. Ultrason. Imaging. 1991;13:111–34. doi: 10.1177/016173469101300201. [DOI] [PubMed] [Google Scholar]
- [12].Barbone PE, Bamber JC. Quantitative elasticity imaging: what can and cannot be inferred from strain images. Phys. Med. Biol. 2002;47:2147–64. doi: 10.1088/0031-9155/47/12/310. [DOI] [PubMed] [Google Scholar]
- [13].Doyley MM. Model-based elastography: a survey of approaches to the inverse elasticity problem. Phys. Med. Biol. 2012;57:R35–73. doi: 10.1088/0031-9155/57/3/R35. doi:10.1088/0031-9155/57/3/R35. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Bamber J, Cosgrove D, Dietrich CF, Fromageau J, Bojunga J, Calliada F, Cantisani V, Correas J-M, D’Onofrio M, Drakonaki EE, Fink M, Friedrich-Rust M, Gilja OH, Havre RF, Jenssen C, Klauser AS, Ohlinger R, Saftoiu A, Schaefer F, Sporea I, Piscaglia F. EFSUMB guidelines and recommendations on the clinical use of ultrasound elastography. Part 1: Basic principles and technology. Ultraschall Med. 2013;34:169–84. doi: 10.1055/s-0033-1335205. doi:10.1055/s-0033-1335205. [DOI] [PubMed] [Google Scholar]
- [15].Mariappan YK, Glaser KJ, Ehman RL. Magnetic resonance elastography: a review. Clin. Anat. 2010;23:497–511. doi: 10.1002/ca.21006. doi:10.1002/ca.21006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [16].Kennedy BF, Kennedy KM, Sampson DD. A Review of Optical Coherence Elastography: Fundamentals, Techniques and Prospects. IEEE J. Sel. Top. Quantum Electron. 2014;20:272–288. doi:10.1109/JSTQE.2013.2291445. [Google Scholar]
- [17].Szabo TL. Diagnostic Ultrasound Imaging: Inside Out. Elsevier; 2014. doi:10.1016/B978-0-12-396487-8.00016-1. [Google Scholar]
- [18].Garra BS, Cespedes EI, Ophir J, Spratt SR, Zuurbier RA, Magnant CM, Pennanen MF. Elastography of breast lesions: initial clinical results. Radiology. 1997;202:79–86. doi: 10.1148/radiology.202.1.8988195. doi:10.1148/radiology.202.1.8988195. [DOI] [PubMed] [Google Scholar]
- [19].de Korte CL, Pasterkamp G, van der Steen a. F.W., Woutman H. a., Bom N. Characterization of Plaque Components With Intravascular Ultrasound Elastography in Human Femoral and Coronary Arteries In Vitro. Circulation. 2000;102:617–623. doi: 10.1161/01.cir.102.6.617. doi:10.1161/01.CIR.102.6.617. [DOI] [PubMed] [Google Scholar]
- [20].de Korte CL, Sierevogel MJ, Mastik F, Strijder C, Schaar JA, Velema E, Pasterkamp G, Serruys PW, van der Steen AFW. Identification of atherosclerotic plaque components with intravascular ultrasound elastography in vivo: a Yucatan pig study. Circulation. 2002;105:1627–30. doi: 10.1161/01.cir.0000014988.66572.2e. [DOI] [PubMed] [Google Scholar]
- [21].Zhu Y, Hall TJ. A modified block matching method for real-time freehand strain imaging. Ultrason. Imaging. 2002;24:161–76. doi: 10.1177/016173460202400303. [DOI] [PubMed] [Google Scholar]
- [22].Jiang J, Hall TJ. A parallelizable real-time motion tracking algorithm with applications to ultrasonic strain imaging. Phys. Med. Biol. 2007;52:3773–90. doi: 10.1088/0031-9155/52/13/008. doi:10.1088/0031-9155/52/13/008. [DOI] [PubMed] [Google Scholar]
- [23].Cespedes I, Huang Y, Ophir J, Spratt S. Methods for Estimation of Subsample Time Delays of Digitized Echo Signals. Ultrason. Imaging. 1995;17:142–171. doi: 10.1177/016173469501700204. doi:10.1177/016173469501700204. [DOI] [PubMed] [Google Scholar]
- [24].Adrian RJ, Westerweel J. Particle Image Velocimetry. Cambridge University Press; 2011. [Google Scholar]
- [25].Richards MS. Quantitative Three Dimensional Elasticity Imaging. Boston University; 2007. [Google Scholar]
- [26].Richards MS, Barbone PE, Oberai AA. Quantitative three-dimensional elasticity imaging from quasi-static deformation: a phantom study. Phys. Med. Biol. 2009;54:757–79. doi: 10.1088/0031-9155/54/3/019. doi:10.1088/0031-9155/54/3/019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [27].Hild F, Roux S. Digital Image Correlation: from Displacement Measurement to Identification of Elastic Properties - a Review. Strain. 2006;42:69–80. doi:10.1111/j.1475-1305.2006.00258.x. [Google Scholar]
- [28].Zhu Y, Chaturvedi P, Insana M. Strain imaging with a deformable mesh. Ultrason. Imaging. 1999;21 doi: 10.1177/016173469902100204. [DOI] [PubMed] [Google Scholar]
- [29].Noble PB, Sharma A, McFawn PK, Mitchell HW. Airway narrowing in porcine bronchi with and without lung parenchyma. Eur. Respir. J. 2005;26:804–11. doi: 10.1183/09031936.05.00065405. doi:10.1183/09031936.05.00065405. [DOI] [PubMed] [Google Scholar]
- [30].LaPrad AS, Szabo TL, Suki B, Lutchen KR. Tidal stretches do not modulate responsiveness of intact airways in vitro. J. Appl. Physiol. 2010;109:295–304. doi: 10.1152/japplphysiol.00107.2010. doi:10.1152/japplphysiol.00107.2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [31].LaPrad AS, West AR, Noble PB, Lutchen KR, Mitchell HW. Maintenance of airway caliber in isolated airways by deep inspiration and tidal strains. J. Appl. Physiol. 2008;105:479–485. doi: 10.1152/japplphysiol.01220.2007. doi:10.1152/japplphysiol.01220.2007. [DOI] [PubMed] [Google Scholar]
- [32].Bendixen HH, Smith GM, Mead J. Pattern of ventilation in young adults. J. Appl. Physiol. 1964;19:195–198. doi: 10.1152/jappl.1964.19.2.195. doi:10.1097/00132586-196504000-00012. [DOI] [PubMed] [Google Scholar]
- [33].Dau AHL. RF2B - RF to B-Mode Ultrasound. 2010 [Google Scholar]
- [34].Otsu N. A Threshold Selection Method from Gray-Level Histograms. IEEE Trans. Syst. Man. Cybern. 1979;9:62–66. doi:10.1109/TSMC.1979.4310076. [Google Scholar]
- [35].Geuzaine C, Remacle J-F. Gmsh: A 3-D finite element mesh generator with built-in pre- and post-processing facilities. Int. J. Numer. Methods Eng. 2009;79:1309–1331. doi:10.1002/nme.2579. [Google Scholar]
- [36].Hansen PC. Analysis of Discrete Ill-Posed Problems by Means of the L-Curve. SIAM Rev. 1992;34:561–580. doi:10.1137/1034115. [Google Scholar]
- [37].Hansen PC, O’Leary DP. The Use of the L-Curve in the Regularization of Discrete Ill- Posed Problems. SIAM J. Sci. Comput. 1993;14:1487–1503. doi:10.1137/0914086. [Google Scholar]
- [38].Vogel CR. Computational Methods for Inverse Problems. SIAM. 2002 [Google Scholar]
- [39].Ayachit U. The ParaView Guide: A Parallel Visualization Application. Kitware. 2015 [Google Scholar]
- [40].Costanzo F, Brasseur JG. The invalidity of the Laplace law for biological vessels and of estimating elastic modulus from total stress vs. strain: a new practical method. Math. Med. Biol. 2013;32:1–37. doi: 10.1093/imammb/dqt020. doi:10.1093/imammb/dqt020. [DOI] [PubMed] [Google Scholar]
- [41].Richards MS, Doyley MM. Non-rigid image registration based strain estimator for intravascular ultrasound elastography. Ultrasound Med. Biol. 2013;39:515–33. doi: 10.1016/j.ultrasmedbio.2012.09.023. doi:10.1016/j.ultrasmedbio.2012.09.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [42].Pratusevich VR, Seow CY, Ford LE. Plasticity in canine airway smooth muscle. J. Gen. Physiol. 1995;105:73–94. doi: 10.1085/jgp.105.1.73. doi:10.1085/jgp.105.1.73. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.










