Skip to main content
Wiley Open Access Collection logoLink to Wiley Open Access Collection
. 2016 Dec 28;30(2):e3684. doi: 10.1002/nbm.3684

Proton and phosphorus magnetic resonance spectroscopy of the healthy human breast at 7 T

Wybe JM van der Kemp 1,, Bertine L Stehouwer 1, Vincent O Boer 1, Peter R Luijten 1, Dennis WJ Klomp 1, Jannie P Wijnen 1
PMCID: PMC5248643  PMID: 28032377

Abstract

In vivo water‐ and fat‐suppressed 1H magnetic resonance spectroscopy (MRS) and 31P magnetic resonance adiabatic multi‐echo spectroscopic imaging were performed at 7 T in duplicate in healthy fibroglandular breast tissue of a group of eight volunteers. The transverse relaxation times of 31P metabolites were determined, and the reproducibility of 1H and 31P MRS was investigated. The transverse relaxation times for phosphoethanolamine (PE) and phosphocholine (PC) were fitted bi‐exponentially, with an added short T 2 component of 20 ms for adenosine monophosphate, resulting in values of 199 ± 8 and 239 ± 14 ms, respectively. The transverse relaxation time for glycerophosphocholine (GPC) was also fitted bi‐exponentially, with an added short T 2 component of 20 ms for glycerophosphatidylethanolamine, which resonates at a similar frequency, resulting in a value of 177 ± 6 ms. Transverse relaxation times for inorganic phosphate, γ‐ATP and glycerophosphatidylcholine mobile phospholipid were fitted mono‐exponentially, resulting in values of 180 ± 4, 19 ± 3 and 20 ± 4 ms, respectively. Coefficients of variation for the duplicate determinations of 1H total choline (tChol) and the 31P metabolites were calculated for the group of volunteers. The reproducibility of inorganic phosphate, the sum of phosphomonoesters and the sum of phosphodiesters with 31P MRS imaging was superior to the reproducibility of 1H MRS for tChol. 1H and 31P data were combined to calculate estimates of the absolute concentrations of PC, GPC and PE in healthy fibroglandular tissue, resulting in upper limits of 0.1, 0.1 and 0.2 mmol/kg of tissue, respectively.

Keywords: 1H, 31P, 7 T, breast, MRSI, total choline, transverse relaxation time


Abbreviations

AFP

adiabatic full passage

AHP

adiabatic half passage

AMESING

adiabatic multi‐echo spectroscopic imaging

AMP

adenosine monophosphate

ATP

adenosine triphosphate

FID

free induction decay

FOCI

frequency offset corrected inversion

GOIA

gradient‐modulated offset independent adiabaticity

GPC

glycerophosphocholine

GPE

glycerophosphoethanolamine

GPtC

(diacyl‐)glycerophosphatidylcholine

GPtE

(diacyl‐)glycerophosphatidylethanolamine

MPL

membrane phospholipid

MRS

magnetic resonance spectroscopy

MRSI

magnetic resonance spectroscopic imaging

NMP

nucleoside monophosphate

NTP

nucleoside triphosphate

PC

phosphocholine

PCr

phosphocreatine

PDE

phosphodiester

PE

phosphoethanolamine

Pi

inorganic phosphate

PME

phosphomonoester

RF

radiofrequency

SAR

specific absorption rate

SNR

signal‐to‐noise ratio

tChol

‘total choline’ (also including ethanolamine)

VAPOR

variable power and optimized relaxation delays

1. INTRODUCTION

Elevated concentrations of the phosphomonoesters (PMEs) phosphocholine (PC) and phosphoethanolamine (PE), which are reaction intermediaries in membrane phospholipid synthesis, are a metabolic hallmark of cancer.1, 2, 3, 4 Intermediaries of phospholipid breakdown are the phosphodiesters (PDE) glycerophosphocholine (GPC) and glycerophosphoethanolamine (GPE).

Cell membrane metabolism can be monitored in vivo by proton (1H) and phosphorus (31P) magnetic resonance spectroscopy (MRS), which may be used clinically for cancer diagnosis and treatment monitoring.5 Unfortunately, the spectral resolution of in vivo 1H spectroscopy is not sufficient to show the individual PME and PDE resonances. These resonances are within the small chemical shift range of 3.2 and 3.3 ppm, and overlap with other metabolites. Particularly in organs that have tissue compartments with substantially different susceptibility, such as lipids and glandular tissue in breast, even higher field strengths will not sufficiently improve the spectral resolution. Moreover, fluctuations caused by breathing, heartbeat and subject motion complicate in vivo shimming of the B 0 field, resulting in further compromises in spectral resolution. Nevertheless, a total choline (tChol) signal can be observed by in vivo 1H MRS in breast tumors and also in healthy fibroglandular breast tissue. The tChol signal comprises choline, PC, PE, GPC and GPE, but may potentially also be contaminated by signals from β‐glucose, taurine and myo‐inositol as a result of poor spectral resolution in the breast at 7 T. It should be noted that the term ‘total choline’ also includes the PEs.

The application of 1H spectroscopy in breast cancer diagnostics and treatment response has been the subject of several reviews,6, 7, 8, 9, 10, 11, 12 in which extensive entries to the literature can be found. In vivo tChol 1H MRS in the human breast is challenging because of the high water and (potentially) high fat signals, which are several orders of magnitude larger than the tChol signal. Appropriate placement of the MRS voxel, excluding as much fatty tissue as possible, is therefore important. In addition, robust water and fat suppression is required,12 for which we have chosen a VAPOR (variable power and optimized relaxation delays)13 scheme (water suppression) prior to the sequence and two chemical shift‐selective radiofrequency (RF) pulses surrounded by crusher gradients applied within the TE of the MRS sequence (MEGA14 scheme fat suppression). At high magnetic field, the most accurate localization technique available so far is a single‐voxel semi‐LASER sequence,15 as it uses adiabatic high‐bandwidth RF pulses that have small chemical shift displacement artifacts and slice profiles with sharp transitions. In the case of MR coils and systems with limited B 1, the use of a semi‐LASER with FOCI (frequency offset corrected inversion) or GOIA (gradient‐modulated offset independent adiabaticity) pulses could provide a better alternative to minimize the chemical shift artifact, although the use of these pulses leads to longer TEs and higher specific absorption rate (SAR). In order to effectively incorporate single‐shot frequency alignments to mitigate field fluctuations during the scan time, metabolite cycling can be applied,16, 17 in which the large signal of water can be removed by subtraction, but used for accurate frequency and phase alignments. When residual water or lipid signals remain visible, Bolan et al18 have demonstrated the use of TE averaging to remove potential artifacts originating from side bands caused by gradient‐induced vibrations. Finally, to ensure that cardiac‐induced rapid field fluctuations remain outside of the TE of the sequence, cardiac triggering can be applied as demonstrated by Andreychenko et al.19 All of these techniques were incorporated in one scan sequence to enable robust 1H MRS in the breast.

Despite the technical advances in robust 1H MRS, in vivo 1H tChol MRS has the severe limitation that it only reflects tChol and not the individual PMEs and PDEs. Although less sensitive than 1H MRS, in vivo 31P MRS offers a more complete window on cell membrane metabolism. It can detect the individual PMEs and PDEs, but also the phosphorus species involved in energy metabolism, e.g. adenosine triphosphate (ATP), inorganic phosphate (Pi) and phosphocreatine (PCr). It has been shown that the PC/GPC ratio increases with increasing aggressiveness of breast cancer cells,2 and such a ratio can only be measured in vivo with 31P MRS and not with 1H MRS. Likewise, the chemotherapy treatment response is often accompanied by decreasing PMEs,1, 20, 21, 22 which can only be measured with 31P MRS. The decrease in tChol on response, which can be measured with 1H MRS, may be less specific.10 However, the intrinsic sensitivity of 1H MRS is much higher than that of 31P MRS.

Recent advances in 31P MRS, as reviewed by Khlebnikov et al,23 have resulted in an increased sensitivity. Moreover, the general formula of the sensitivity ratio between 1H and 31P, as known from the NMR field (γ1H/γ31P)3,24 is not what is practically encountered when applied in the human breast in vivo. First, in the regime of tissue load dominance of the receive coil, i.e. if noise from the sample dominates thermal noise from the coil, which is the case in human in vivo applications, the sensitivity ratio is reduced by approximately γ1H/γ31P to (γ1H/γ31P)2, as the thermal noise of a conducting sample is proportional to the resonance frequency.24 Second, when in the regime in which T 2* substantially dominates over T 2 (i.e. as a result of susceptibility differences between lipids and glandular tissue in the breast: T 2* < < T 2), the sensitivity ratio between 1H and 31P MRS is reduced by approximately the square root of the ratio γ1H/γ31P to (γ1H/γ31P)3/2. This includes two counteracting effects. The sensitivity of 31P is increased by γ1H/γ31P, because of a sharper line width (peak height) for 31P, as a result of the lower resonance frequency of the 31P nucleus relative to the 1H nucleus, but decreased by the square root of γ1H/γ31P, because of the longer sampling time of this sharper peak. (If T 2* < < T 2, then T 2*(1H)/T 2*(25P) ≈ (γ1H/γ31P)−1. The effect of the longer sampling time for 31P is analogous to decreasing the number of sampling averages, and varies with the reciprocal square root (γ1H/γ31P)–1/2.24 It should be noted that all of these sensitivity comparisons only hold under the assumption of RF tissue load dominance and susceptibility dominance.) Moreover, as 31P signals cannot be detected from adipose lipids, voxel selection in 31P MRS becomes substantially less critical. Third, when using multi‐echo sequences, such as AMESING (adiabatic multi‐echo spectroscopic imaging), the sensitivity of 31P MRS can be increased.26 Finally, polarization transfer (not used here) can potentially increase the sensitivity by a factor γ1H/γ31P.27, 28, 29, 30

In this study, we compare state‐of‐the‐art 1H MRS, incorporating all of the techniques mentioned, with AMESING 31P MRS to assess the reproducibility of detection of choline levels in the human breast in a group of eight healthy volunteers. Finally, we combine 1H and 31P MRS data to make an estimation of the concentrations of PMEs and PDEs in healthy fibroglandular tissue.

2. METHODS

2.1. Experimental details

A group of eight healthy female volunteers (21–30 years) underwent two MRI scan sessions of the right breast with a dual‐tuned 1H–31P quadrature coil (MR Coils BV, Drunen, the Netherlands)21 on a 7‐T scanner (Philips Healthcare, Cleveland, OH, USA). Each volunteer was scanned twice on the same day to minimize any effects of hormonal fluctuations/physiological variations on the measurements. The protocol consisted of imaging (scout, fat‐suppressed T 1‐weighted imaging), a B 0 map used for shimming, 31P magnetic resonance spectroscopic imaging (MRSI) and metabolite‐cycled 1H semi‐LASER spectroscopy. 1H spectroscopy was focused on the measurement of the tChol signal and was performed with a water‐ and fat‐suppressed semi‐LASER proton sequence15 with metabolite cycling16, 17 of the choline resonance at 3.2 ppm and a voxel of 15 × 15 × 15 mm3 carefully placed in fibroglandular tissue. The semi‐LASER sequence was performed with offset independent trapezoid pulses with a duration of 3.92 ms at a maximum γB 1 of 1065 Hz and had a 95% inversion bandwidth of 6 kHz. During the first scan session, a screen print was made of the voxel placement (in three directions) and, during the second scan session, this screen print was used as a map to visually place the voxel at approximately the same position for the second scan session. As water was used as an internal reference for quantification, any variation in fibroglandular tissue content in the voxel would be accounted for and would not lead to different tChol concentrations as long as the tChol concentration is homogeneous over the fibroglandular tissue. Water suppression was performed by a VAPOR scheme13 and fat suppression by a MEGA scheme14 with narrow‐band adiabatic full passage (AFP) pulses. Metabolite cycling for tChol was performed by measuring a dynamic series of 2 × 14 scans for each TE (118, 119, 120 and 121 ms) and selectively inverting (with a narrow‐band AFP pulse with a frequency sweep of 300 Hz) the signal at 3.2 ppm in the even scans, and subsequently subtracting even from odd scans after phasing the residual water signal, which was a few per cent of the unsuppressed water signal. 1H spectroscopy was cardiac triggered with a TR of three heart beats for the choline scan (at four different TEs: 118, 119, 120 and 121 ms) and nine heart beats for the water reference scan with TE = 26 ms. The total scan time for the 1H spectroscopy protocol at an average heart rate of 60 beats/min was 6 min and 12 s.

31P MRSI was performed by AMESING26 with the following parameters: TR = 6 s; TE = 45 ms; matrix, 8 × 8 × 8; 4 × 2 × 4 cm3 (RL × AP × FH) voxels; spherical k‐space sampling; acquiring one free induction decay (FID) and five full echoes; a bandwidth of 8200 Hz and 256 data points for the echoes. The adiabatic pulses were driven with γB 1 = 1700 Hz, and the pulse durations of the adiabatic half passage (AHP) and BIR‐4 refocusing pulses were 2 and 8 ms, respectively. Their respective 95% excitation/refocusing bandwidths were 1400 Hz and 1100 Hz. The total scan time of the 31P spectroscopy protocol was 25 min and 36 s. No proton decoupling was applied, as this contributes merely to SAR and not to the signal‐to‐noise ratio (SNR) in vivo in the breast at 7 T, because of the spectral line widths, which are several times the coupling constants between 1H and 31P. All volunteers gave written informed consent to participate in this study.

2.2. Post‐processing of 31P MRSI data and 1H semi‐LASER data

31P MRSI data were spatially Hanning filtered and subsequently apodized by 10 Hz with a Lorentzian line shape and zero filled to 2048 data points in the time domain. After zero‐ and first‐order phasing, the baselines of the FID spectra were corrected by fitting a second‐order spline. All spectral FIDs and echoes were frequency aligned to Pi. The voxel with the highest SNR was chosen for further analysis by maximizing the amount of glandular tissue in the selected voxel by voxel shifting.

1H spectra at the four different TEs were phased and frequency aligned on the residual water signal, and averaged subsequently. The averaged spectra from the even scans were subtracted from the averaged odd scans in order to obtain the tChol signal.

2.3. Spectral fitting of 31P data

The acquired data were analyzed on a group level and on an individual level. Spectral fitting was performed with Lorentzian line shapes in JMRUI31 using the AMARES25 algorithm, employing a priori constraints for the chemical shift difference between PE and PC of 0.5 ppm32 and 0.56 ppm for the chemical shift difference between (diacyl‐)glycerophosphatidylethanolamine (GPtE) + GPC and (diacyl‐)glycerophosphatidylcholine (GPtC).33 In addition, a 10‐Hz additional line width in the FID spectra for the PDEs, relative to Pi and PMEs, was set to compensate for the much shorter T 2 times of the mobile phospholipid PDE signals in the breast.22, 34 Fibroglandular tissue also contains small amounts of adenosine monophosphate (AMP),35 with a chemical shift in between PE and PC at physiological pH.36 The signal of GPE in the breast overlaps partially with the GPtE‐MPL (membrane phospholipid) resonance, and concentrations of GPE in the breast are too low to give sufficient signal intensity for quantification. Line widths for PE, PC and Pi were kept free but identical, and line widths for γ‐nucleoside triphosphate (γ‐NTP) and α‐NTP were also kept free but identical. As the spectral resolution in the breast at 7 T is insufficient to distinguish multiplets from J coupling with protons, multiplets were not considered and fitted as single resonances. Spectral fitting with these a priori constraints was used to obtain the values of the line widths and chemical shifts for all metabolites per volunteer.

2.4. Spectral fitting of 1H data and calculation of tChol concentrations

Spectral fitting was performed with Lorentzian line shapes in JMRUI31 using the AMARES25 algorithm. The frequency of the singlet peak from tChol was constrained between 3.1 and 3.3 ppm and the line width was not constrained. In a second analysis, we verified the influence of constraint of the tChol line width to the line width of the water peak in the sum of even and odd spectra (i.e. the summed spectrum in which tChol should cancel out and residual water and fat peaks should not).

The quantification of tChol was based on a comparison with the reference water scan of the same voxel volume. Longitudinal and transverse relaxation times of water at 7 T were taken from Haddadin et al,9 and were 2.27 s and 36 ms, respectively. For tChol, the T 1 value was taken to be equal to that of water and a T 2 value of 200 ms was assumed (see Discussion).

2.5. T 2 analyses of 31P metabolites

Metabolite T 2 values were determined based on the weighted averaged group spectra (one FID and five echoes) of the eight volunteers. Each volunteer was measured twice, and so the group‐averaged FID is the Pi‐weighted average of 16 FID spectra. Weighted averaging was chosen to maximize SNR of the group spectra. Likewise, the group‐averaged echo spectra were also weighted with the Pi intensity of the FID. Weighted averaged spectra were subsequently spectrally fitted for PE, PC, Pi, GPC + GPtE, GPtC, PCr, γ‐NTP, α‐NTP and NAD in JMRUI31 with the AMARES25 algorithm. Spectral peak areas of all metabolites in the FID and echoes were fitted mono‐exponentially as a function of TE. The spectral amplitudes of PE, PC and GPC + GPtE were also fitted bi‐exponentially. For the PMEs, it was assumed that, at short TE, the signal shows contributions from nucleoside monophosphate, for which a T 2 similar to that of γ‐NTP was chosen. The GPC + GPtE resonance was fitted assuming a T 2 of GPtE similar to GPtC.

2.6. Reproducibility

All volunteers were measured twice on the same day (to minimize physiological influences) and the spectral fitting of these two datasets for each volunteer was used to calculate a standard deviation and a coefficient of variation for each metabolite in all volunteers. The standard deviation of the fitted relative peak area (with respect to the total 31P signal) A m,n of a metabolite m in volunteer n can be written as:

sdAm,n=A¯m,nAm,n,12+A¯m,nAm,n,222 (1)

where A¯m,nis the average of the fitted peak area of metabolite m over the two measurements of volunteer n. Subsequently, the coefficient of variation CoV in the fitted peak area of a metabolite m in volunteer n can be written as:

CoVAm,n=100sdAm,nA¯m,n (2)

Analogously, we can define the average standard deviation and the average coefficient of variation for the peak area of a metabolite m over the whole group of eight volunteers as:

sd¯Am=n=1n=8sdAm,n28 (3)

and

CoV¯Am=100sd¯AmA¯m (4)

respectively.

3. RESULTS AND DISCUSSION

3.1. T 2 analysis of 31P metabolites in fibroglandular breast tissue

For 31P spectroscopy, the measured line widths are in the range 35–50 Hz (0.3–0.4 ppm). The group‐averaged FID and echo spectra (echo spacing, 45 ms) based on the spectra of the eight volunteers are shown in Figure 1. It should be noted that the peak labeled GPtC is no longer visible from Echo 2 onwards, comparable with the rapid disappearance of γ‐NTP. The short apparent T 2 of γ‐NTP is affected by J modulation because of homonuclear 31P coupling.37 In addition, it should be noted that the peak labeled GPC + GPtE loses substantial intensity from FID to the first echo (GPtE has a short T 2), after which the decrease in intensity is much slower, leaving only the slowly decaying GPC signal. The short T 2 PDE signals (GPtC and GPtE) are thought to originate from highly mobile membrane phospholipids.34, 38, 39 In the FID spectrum, there appears to be some signal between the peaks labeled PE and PC that seems to disappear in the echoes; this signal could be nucleoside monophosphate (NMP).35

Figure 1.

Figure 1

Averaged free induction decay (FID) and echo 31P magnetic resonance spectra (echo spacing, 45 ms) for a group of eight healthy volunteers scanned twice. AMP, adenosine monophosphate; ATP, adenosine triphosphate; GPC, glycerophosphocholine; GPE, glycerophosphoethanolamine; GPtC, (diacyl‐)glycerophosphatidylcholine; GPtE, (diacyl‐)glycerophosphatidylethanolamine; PC, phosphocholine; PCr, phosphocreatine; PE, phosphoethanolamine; Pi, inorganic phosphate

The β‐NTP signal could not be observed above noise level and therefore the spectra are only shown between 10 and −10 ppm. The reason for the apparent absence of the β‐NTP signal in the FID spectrum may be two‐fold. First, the 95% excitation bandwidth of the AHP pulse is only 1400 Hz and the transmitter offset is approximately 2500 Hz away from the β‐NTP resonance. Second, the β‐NTP resonance is dependent on the Mg2 + concentration, which may be low in healthy breast fibroglandular tissue.

The group‐averaged spectra depicted in Figure 1 allow for a pH estimation of fibroglandular tissue based on the chemical shift of Pi. As the small PCr signal most probably originates from voxel bleeding from the pectoralis muscle (which may have a B 0 offset because shimming was aimed at the glandular tissue), it is safer to use the signal from GPC and/or α‐NTP as chemical shift reference value and take, for these resonances, chemical shift values that have been measured in vivo, in the presence of PCr, which is by definition 0.0. Here, we adopt for GPC and α‐NTP, values of 2.97 ppm and −7.57 ppm, respectively, as measured in mammalian brain.40 For the FID spectrum, we use the α‐NTP resonance as it is most clear, whereas the GPC signal in FID is affected by mobile phospholipid; for the echo spectra, we take GPC as chemical shift reference. In this way, we have six measurements (one FID and five echoes) for the chemical shift of Pi which is, on average, 5.31 ± 0.05 ppm, corresponding to a pH value of 7.51 ± 0.07.

Figure 2 shows the T 2 fits for the different 31P metabolites. It should be noted that the FID signal amplitudes of PE + NMP, PC + NMP and GPC + GPtC all appear well above the fitted mono‐exponential curve. The additional bi‐exponential fit for these resonances shows a better agreement with the data. As the T 2 values of the NMP and GPtC components are not known, we assume a similar T 2 for NMP as for the apparent T 2 of γ‐NTP that we measured and, likewise, a similar T 2 of GPtC as for the GPtE component that we measured. An overview of the fitted T2 data is shown in Table 1.

Figure 2.

Figure 2

Transverse relaxation times T 2 fitted for various 31P metabolites in fibroglandular tissue, derived from the weighted average group spectra [one free induction decay (FID) and five echoes] of eight healthy volunteers measured twice. (a1) Mono‐exponential T 2 fit for the combined resonance PE + NMP. (a2) Bi‐exponential T 2 fit for the combined resonance PE + NMP. (b) (Mono‐exponential) T 2 fit for Pi. (c1) Mono‐exponential T 2 fit for the combined resonance PC + NMP. (c2) Bi‐exponential T 2 fit for the combined resonance PC + NMP. (d) (Mono‐exponential) T 2 fit for GPtC. (e1) Mono‐exponential T 2 fit for the combined resonance GPC + GPtE. (e2) Bi‐exponential T 2 fit for the combined resonance GPC + GPtE. (f) (Mono‐exponential) T 2 fit for γ‐NTP. GPC, glycerophosphocholine; GPtC, (diacyl‐)glycerophosphatidylcholine; GPtE, (diacyl‐)glycerophosphatidylethanolamine; NMP, nucleoside monophosphate; NTP, nucleoside triphosphate; PC, phosphocholine; PE, phosphoethanolamine; Pi, inorganic phosphate

Table 1.

Fitted T 2 values of 31P metabolites in healthy fibroglandular breast tissue

Metabolite T 2 (ms) Metabolite T 2 (ms)
PE* 199 ± 8 GPC* 177 ± 6
PC* 239 ± 14 GPtC 20 ± 4
Pi 180 ± 4 γ‐NTP 19 ± 3
*

Fitted bi‐exponentially with an additional short T 2 compound of 20 ms (NMP for PE and PC; GPtE for GPC).

GPC, glycerophosphocholine; GPtC, (diacyl‐)glycerophosphatidylcholine; GPtE, (diacyl‐)glycerophosphatidylethanolamine; NMP, nucleoside monophosphate; NTP, nucleoside triphosphate; PC, phosphocholine; PE, phosphoethanolamine; Pi, inorganic phosphate.

The relative metabolite abundances expressed as a percentage of the total 31P signal, for this group of volunteers measured twice, as obtained from the data in Figures 1 and 2, are: PE = 8%; AMP = 5%; PC = 4%; Pi =29%; GPE ~ 0%; GPC = 4%; GPtE =10%; GPtC =17%; γ‐NTP = 13%; α‐NTP = 8%; NADP =2%.

3.2. Reproducibility of 31P measurements

The average relative errors (Equation (4)) in the amplitudes of the different metabolite signals, derived from spectral fitting of the FID spectra and the T 2‐weighted (T2w) average spectra, are shown in Figure 3a,b. Errors on the individual monoester signals are largest because these signals have, on average (see Figure 1), the lowest SNR, overlap partially and are also influenced by the presence of a disturbing signal from NMP. The sum of monoesters, however, shows only small variability. Note the reduction in the relative error and standard deviation of the relative error when making use of T2w spectra, which combine FID and echo spectra, relative to FID spectra, which results in increased SNR and more reliable fitting.

Figure 3.

Figure 3

Coefficient of variation in metabolite peak amplitudes (Equation (4)), as fitted with JMRUI. A, 31P magnetic resonance spectroscopy (MRS) derived from free induction decay (FID) data. B, 31P MRS derived from T 2‐weighted (T2w) data. C, 1H MRS total choline (tChol), with free line width (LW) and fixed LW for tChol. GPC, glycerophosphocholine; GPtC, (diacyl‐)glycerophosphatidylcholine; GPtE, (diacyl‐)glycerophosphatidylethanolamine; PC, phosphocholine; PDE, phosphodiester PE, phosphoethanolamine; Pi, inorganic phosphate; PME, phosphomonoester

3.3. 1H spectroscopy

Figure 4 shows an example of the voxel placement for the semi‐LASER 1H MRS in a single volunteer for the duplicate measurements. The line width of water in the small 1H MRS voxel is in the range 20–50 Hz (0.07–0.17 ppm). The tChol concentrations, determined in duplicate, for a voxel of fibroglandular tissue of the eight volunteers are shown in Table 2. In the case of volunteers 3 and 5, one of the two determinations gave insufficient signal to determine a tChol concentration. Spectral fitting was performed with an unconstrained line width for tChol and also with a fixed line width for tChol, equal to the line width of the water signal, whilst adding the (frequency aligned) odd and even scans (i.e. tChol cancels out and residual water and fat peaks do not). The group‐averaged tChol concentration of the data analyzed with a free line width for tChol, as well as with a fixed line width, from Table 2 is 0.54 mM (or, more precisely, mmol/kgwater). Correcting for the water content of fibroglandular tissue being approximately 60%,41 this would reduce to 0.32 mM.

Figure 4.

Figure 4

Example of voxel placement in three different planes of view for duplicate 1H semi‐LASER magnetic resonance spectroscopy in a volunteer, the inner square is the selected voxel. Top row, first measurement series. Bottom row, voxel placement for second measurement series

Table 2.

Total choline (tChol) concentration of fibroglandular tissue of a group of eight healthy volunteers. Data are calculated as choline, i.e. assuming that all signal between 3.1 and 3.3 ppm is choline based (nine protons). T 1 and T 2 values for water were taken from Haddadin et al.9 Errors are based on Cramer–Rao lower bounds from JMRUI. Analyses were performed with a free line width for tChol and with a fixed line width equal to the line width of the water signal

Volunteer Measurement 1 tChol (mM) Measurement 2 tChol (mM)
Free line width Fixed line width Free line width Fixed line width
1 0.4 ± 0.1 0.7 ± 0.1 0.4 ± 0.1 0.5 ± 0.1
2 1.0 ± 0.3 1.1 ± 0.3 1.1 ± 0.3 1.0 ± 0.3
3 (0) (0) 0.6 ± 0.1 0.5 ± 0.1
4 0.5 ± 0.1 0.6 ± 0.1 1.0 ± 0.2 0.6 ± 0.1
5 0.7 ± 0.2 0.5 ± 0.2 (0) (0)
6 0.6 ± 0.1 0.5 ± 0.1 0.6 ± 0.1 0.5 ± 0.1
7 0.6 ± 0.2 0.6 ± 0.2 0.3 ± 0.1 0.6 ± 0.2
8 0.4 ± 0.1 0.4 ± 0.1 0.4 ± 0.1 0.3 ± 0.1

The calculated tChol concentration depends on the T 1 and T 2 values that were assumed for tChol. As these values are not known in the breast at 7 T, they had to be estimated. For the T 1 of tChol, we assumed a value similar to the T 1 of water (2.27 s) at 7 T, as tabulated by Haddadin et al.9 For the T 2 of tChol, we assumed a value of 200 ms. Reported T 2 values of tChol in the breast at 1.5 T range between 181 and 360 ms, whereas, at 4 T, a value of 399 ± 133 ms was measured, as reviewed by Haddadin et al.9 A value of 178 ± 28 ms for tChol was measured at 9.4 T in rat brain.42 Therefore, a value in the range 150–250 ms for T 2 of tChol at 7 T seems plausible. A value of T 2 = 150 ms for tChol would lead to approximately 30% higher tChol concentrations than at T 2 = 200 ms, and a value of 250 ms would lead to approximately 20% lower concentrations than reported here.

The relative error in the 1H tChol determination, calculated analogously to the 31P data (Equation (4)), as shown in Figure 3c, is 34% for the spectral fitting with free line width and 26% for the spectral fitting with fixed line width for tChol.

The interpretation of the relative errors depicted in Figure 3 should also take into consideration the difference in voxel sizes used for 31P and 1H spectroscopy. Here, we used 4 × 2 × 4 cm3 (nominal resolution) voxels for 31P MRSI of healthy glandular tissue, whereas, in a clinical setting with breast cancer patients, we would use smaller voxels of 2 × 2 × 2 cm3. Considering the acquisition‐weighted k‐space sampling and the applied Hanning filter in the MRSI sequence, the true voxel size in our study is even larger: an ellipsoid of approximately 82 cm3. In a worst case, this would mean an increase in the coefficient of variation for 31P spectroscopy of a factor of four on using the 2 × 2 × 2 cm3 nominal voxel size. As the larger voxel size does not imply that these voxels are completely filled with glandular tissue, they may contain fatty tissue (without 31P signal) as well, such that the factor of four in the coefficient of variation is the limiting case.

Although the intrinsic sensitivity of 31P MRS remains less than that of 1H MRS, the absence of disturbing signals from lipids and water and the insensitivity of the AMESING sequence to B 0 inhomogeneities provide substantial advantages of 31P MRS over 1H MRS in the robust detection of phospholipid metabolites. With the larger voxel sizes, longer scan times and absence of artifacts, 31P MRS could detect the phospholipid metabolite signals with an order of magnitude improved coefficient of variation (Figure 3) than with 1H MRS.

Figure 5 shows an example of 31P MR spectra and metabolite cycled 1H MR tChol spectra for volunteer 2 measured in duplicate. Although the noise level is expected to be the same in all 31P acquisitions, the noise pattern of the 31P spectrum in the top right of Figure 5 seems to include a higher spectral noise density than in the remaining spectra. This may be caused by truncation artifacts as a result of the relatively short acquisition window of 16 ms of FID acquisition, which affects the spectral region most strongly around sharp and high‐intensity peaks.

Figure 5.

Figure 5

Duplicate 31P spectra [free induction decay (FID) and T 2‐weighted] and 1H total choline (tChol) spectra for volunteer 2. GPC, glycerophosphocholine; GPtC, (diacyl‐)glycerophosphatidylcholine; GPtE, (diacyl‐)glycerophosphatidylethanolamine; NTP, nucleoside triphosphate; PC, phosphocholine; PE, phosphoethanolamine; Pi, inorganic phosphate

Estimates of the average PME and PDE concentrations in healthy fibroglandular tissue based on a combination of 31P and 1H MRS, using the 31P data shown in Figures 1 and 2, and the average tChol value of Table 1 of 0.54 mM, are PE = 0.4 mM (two protons), PC = 0.2 mM (nine protons) and GPC = 0.2 mM (nine protons), where it was assumed that the free choline concentration can be neglected. The estimated concentrations of PMEs and PDEs are therefore an upper limit. Correction for 60% water content of fibroglandular tissue39 would lead to approximate values, expressed in mmol/kg of tissue (assuming a tissue density of approximately 1 kg/L), which are 60% of the values expressed in mM. An assessment by liquid chromatography‐mass spectrometry of three samples of healthy breast tissue by Mimmi et al43 led to values for PC in the range 0.03–0.21 mmol/kg; this corresponds well with our value for PC expressed in the same units (mmol/kg of tissue) of 0.1 mmol/kgtissue.

This study shows the feasibility of combining 31P MRSI with localized 1H MRS in vivo in the breast at 7 T. The combination of these techniques may be of value in monitoring the response to chemotherapy in breast cancer patients, where it may possibly better discriminate between responders and non‐responders than 31P or 1H MRS alone. From ex vivo analysis43, 44 of breast cancer samples, it is known that the free choline concentrations are small, compared with PC; therefore, combined 31P and 1H MRS in vivo in breast cancer will enable, to a good approximation, absolute concentration measurements of key metabolites of cell membrane metabolism.

4. CONCLUSIONS

In vivo 31P MRS offers a better window on cell membrane metabolism and energy metabolism than does tChol derived from 1H MRS. The 31P MRS reproducibility of the individual PMEs was similar to that of 1H tChol MRS. The reproducibility of Pi, the sum of PMEs and the sum of PDEs with 31P MRSI was superior to the reproducibility of 1H MRS for tChol. Upper limits for the PME and PDE concentrations in healthy fibroglandular breast tissue derived from this study are [PC] = 0.1 mmol/kg, [GPC] = 0.1 mmol/kg and [PE] = 0.2 mmol/kg.

ACKNOWLEDGEMENTS

We acknowledge our sponsors Alpe d'Huzes UU2013‐6302.

van der Kemp WJM, Stehouwer BL, Boer VO, Luijten PR, Klomp DWJ, Wijnen JP. Proton and phosphorus magnetic resonance spectroscopy of the healthy human breast at 7 T. NMR in Biomedicine. 2017;30:e3684. doi: 10.1002/nbm.3684.

REFERENCES

  • 1. Negendank W, Li CW, Padavic‐Shaller K, Murphy‐Boesch J, Brown TR. Phospholipid metabolites in 1H‐decoupled 31P MRS in vivo in human cancer: implications for experimental models and clinical studies. Anticancer Res. 1996;16:1539–1544. [PubMed] [Google Scholar]
  • 2. Aboagye EO, Bhujwalla ZM. Malignant transformation alters membrane choline phospholipid metabolism of human mammary epithelial cells. Cancer Res. 1999;59:80–84. [PubMed] [Google Scholar]
  • 3. Katz‐Brull R, Lavin PT, Lenkinski RE. Clinical utility of proton magnetic resonance spectroscopy in characterizing breast lesions. J Natl Cancer Inst. 2002;94:1197–1203. [DOI] [PubMed] [Google Scholar]
  • 4. Glunde K, Bhujwalla ZM, Ronen SM. Choline metabolism in malignant transformation. Nat Rev Cancer. 2011;11:835–848. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Korteweg MA, Veldhuis WB, Visser F, et al. Feasibility of 7 Tesla breast magnetic resonance imaging determination of intrinsic sensitivity and high‐resolution magnetic resonance imaging, diffusion‐weighted imaging, and (1)H‐magnetic resonance spectroscopy of breast cancer patients receiving neoadjuvant therapy. Invest Radiol. 2011;46:370–376. [DOI] [PubMed] [Google Scholar]
  • 6. Tse GM, Yeung DK, King AD, Cheung HS, Yang WT. In vivo proton magnetic resonance spectroscopy of breast lesions: an update. Breast Cancer Res Treat. 2007;104(3):249–255. [DOI] [PubMed] [Google Scholar]
  • 7. Bartella L, Huang W. Proton (1H) MR spectroscopy of the breast. Radiographics. 2007;27(Suppl. 1):S241–S252. [DOI] [PubMed] [Google Scholar]
  • 8. Sardanelli F, Fausto A, Podo F. MR spectroscopy of the breast. Radiol Med. 2008;113:56–64. [DOI] [PubMed] [Google Scholar]
  • 9. Haddadin IS, McIntosh A, Meisamy S, et al. Metabolite quantification and high‐field MRS in breast cancer. NMR Biomed. 2009;22:65–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Sharma U, Baek HM, Su MY, Jagannathan NR. In vivo 1H MRS in the assessment of the therapeutic response of breast cancer patients. NMR Biomed. 2011;24:700–711. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Begley JK, Redpath TW, Bolan PJ, Gilbert FJ. In vivo proton magnetic resonance spectroscopy of breast cancer: a review of the literature. Breast Cancer Res. 2012;14(2):207 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Bolan PJ. Magnetic resonance spectroscopy of the breast: current status. Magn Reson Imaging Clin N Am. 2013;21:625–639. [DOI] [PubMed] [Google Scholar]
  • 13. Tkác I, Starcuk Z, Choi IY, Gruetter R. In vivo 1H NMR spectroscopy of rat brain at 1 ms echo time. Magn Reson Med. 1999;41:649–656. [DOI] [PubMed] [Google Scholar]
  • 14. Mescher M, Tannus A, O'Neil Johnson M, Garwood M. Solvent suppression using selective echo dephasing. J Magn Reson Ser A. 1996;123:226–229. [Google Scholar]
  • 15. Scheenen TW, Klomp DW, Wijnen JP, Heerschap A. Short echo time 1H‐MRSI of the human brain at 3 T with minimal chemical shift displacement errors using adiabatic refocusing pulses. Magn Reson Med. 2008;59(1):1–6. [DOI] [PubMed] [Google Scholar]
  • 16. Dreher W, Leibfritz D. New method for the simultaneous detection of metabolites and water in localized in vivo 1H nuclear magnetic resonance spectroscopy. Magn Reson Med. 2005;54:190–195. [DOI] [PubMed] [Google Scholar]
  • 17. Hock A, MacMillan EL, Fuchs A, et al. Non‐water‐suppressed proton MR spectroscopy improves spectral quality in the human spinal cord. Magn Reson Med. 2013;69(5):1253–1260. [DOI] [PubMed] [Google Scholar]
  • 18. Bolan PJ, DelaBarre L, Baker EH, et al. Eliminating spurious lipid sidebands in 1H MRS of breast lesions. Magn Reson Med. 2002;48:215–222. [DOI] [PubMed] [Google Scholar]
  • 19. Andreychenko A, Boer VO, Arteaga de Castro CS, Luijten PR, Klomp DW. Efficient spectral editing at 7 T: GABA detection with MEGA‐sLASER. Magn Reson Med. 2012;68:1018–1025. [DOI] [PubMed] [Google Scholar]
  • 20. Leach MO, Verrill M, Glaholm J, et al. Measurements of human breast cancer using magnetic resonance spectroscopy: a review of clinical measurements and a report of localized 31P measurements of response to treatment. NMR Biomed. 1998;11:314–340. [DOI] [PubMed] [Google Scholar]
  • 21. Klomp DW, van de Bank BL, Raaijmakers A, et al. 31P MRSI and 1H MRS at 7 T: initial results in human breast cancer. NMR Biomed. 2011;24:1337–1342. [DOI] [PubMed] [Google Scholar]
  • 22. van der Kemp WJ, Stehouwer BL, Luijten PR, van den Bosch MA, Klomp DW. Detection of alterations in membrane metabolism during neoadjuvant chemotherapy in patients with breast cancer using phosphorus magnetic resonance spectroscopy at 7 Tesla. Springerplus. 2014;3:634 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Khlebnikov V, Wijnen J, van der Kemp WJM, Klomp DWJ. 31P MRSI studies in patients with cancer. Annu Rep NMR Spectrosc. 2016;87:319–368. [Google Scholar]
  • 24. De Graaf RA. In Vivo NMR, Principles and Techniques. Chichester: Wiley; 2007. [Google Scholar]
  • 25. Vanhamme L, van den Boogaart A, Van Huffel S. Improved method for accurate and efficient quantification of MRS data with use of prior knowledge. J Magn Reson. 1997;129:35–43. [DOI] [PubMed] [Google Scholar]
  • 26. van der Kemp WJM, Boer VO, Luijten PR, Stehouwer BL, Veldhuis WB, Klomp DWJ. Adiabatic Multi‐Echo P Spectroscopic ImagiNG (AMESING) at 7 tesla for measuring transverse relaxation times and regaining sensitivity in tissues with short T2* values. NMR Biomed. 2013;26:1299–1307. [DOI] [PubMed] [Google Scholar]
  • 27. Morris G, Freeman R. Enhancement of nuclear magnetic resonance signals by polarisation transfer. J Am Chem Soc. 1979;101:760–762. [Google Scholar]
  • 28. Burum DP, Ernst RR. Net polarization transfer via a J‐ordered state for signal enhancement of low‐sensitivity nuclei. J Magn Reson. 1980;39:163–168. [Google Scholar]
  • 29. Merkle H, Wei H, Garwood M, Uğurbil K. B1‐insensitive heteronuclear adiabatic polarization transfer for signal enhancement. J Magn Reson. 1992;99:480–494. [Google Scholar]
  • 30. van der Kemp WJ, Boer VO, Luijten PR, Klomp DW. Increased sensitivity of 31P MRSI using direct detection integrated with multi‐echo polarization transfer (DIMEPT). NMR Biomed. 2014;27:1248–1255. [DOI] [PubMed] [Google Scholar]
  • 31. Naressi A, Couturier C, Devos JM, et al. Java‐based graphical user interface for the MRUI quantitation package. MAGMA. 2001;12:141–152. [DOI] [PubMed] [Google Scholar]
  • 32. Payne GS, Troy H, Vaidya SJ, Griffiths JR, Leach MO, Chung YL. Evaluation of 31P high‐resolution magic angle spinning of intact tissue samples. NMR Biomed. 2006;19:593–598. [DOI] [PubMed] [Google Scholar]
  • 33. Jensen JE, Drost DJ, Menon RS, Williamson PC. In vivo brain 31P MRS: measuring the phospholipid resonances at 4 Tesla from small voxels. NMR Biomed. 2002;15:338–347. [DOI] [PubMed] [Google Scholar]
  • 34. van der Kemp WJM, Stehouwer BL, Runge JH, et al. AMESING and BINEPT 31P MRS at 7 T distinguishes glycerophosphatidylcholine from glycerophosphocholine. Proceedings of the 22nd Annual Meeting ISMRM, Milan, Italy, 2014; 4073.
  • 35. Smith TA, Glaholm J, Leach MO, et al. A comparison of in vivo and in vitro 31P NMR spectra from human breast tumours: variations in phospholipid metabolism. Br J Cancer. 1991;63:514–516. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Robitaille PL, Robitaille PA, Brown GG Jr, Brown GG. An analysis of the pH‐dependent chemical‐shift behaviour of phosphorus containing metabolites. J Magn Reson. 1991;92:73–84. [Google Scholar]
  • 37. Jung WI, Straubinger K, Bunse M, et al. A pitfall associated with determination of transverse relaxation times of the 31P NMR signals of ATP using the Hahn spin‐echo. Magn Reson Med. 1993;30:138–141. [DOI] [PubMed] [Google Scholar]
  • 38. van der Kemp WJM, Wijnen JP, Luijten PR, Klomp DWJ. Saturation‐transfer effects and longitudinal relaxation times of 31P metabolites in fibroglandular breast tissue at 7 tesla. Magn Reson Med. 2016;76:402–407. [DOI] [PubMed] [Google Scholar]
  • 39. van der Kemp WJM, Stehouwer BL, Runge JH, et al. Glycerophosphocholine and glycerophosphoethanolamine are not the main sources of the in vivo 31P MRS phosphodiester signals from healthy fibroglandular breast tissue at 7 T. Front Oncol. 2016;6:29. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Cady EB. In vivo cerebral 31P magnetic resonance spectroscopy. In: Choi I‐Y, Gruetter R, eds. Neural metabolism in vivo. Adv Neurobiol. 2012;4:149–179. [Google Scholar]
  • 41. Brooksby B, Pogue BW, Jiang S, et al. Imaging breast adipose and fibroglandular tissue molecular signatures by using hybrid MRI‐guided near‐infrared spectral tomography. Proc Natl Acad Sci U S A. 2006;103:8828–8833. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Xin L, Gambarota G, Mlynárik V, Gruetter R. Proton T2 relaxation time of J‐coupled cerebral metabolites in rat brain at 9.4 T. NMR Biomed. 2008;21:396–401. [DOI] [PubMed] [Google Scholar]
  • 43. Mimmi MC, Finato N, Pizzolato G, et al. Absolute quantification of choline‐related biomarkers in breast cancer biopsies by liquid chromatography electrospray ionization mass spectrometry. Anal Cell Pathol (Amst). 2013;36:71–83. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Mimmi MC, Picotti P, Corazza A, et al. High‐performance metabolic marker assessment in breast cancer tissue by mass spectrometry. Clin Chem Lab Med. 2011;49:317–324. [DOI] [PubMed] [Google Scholar]

Articles from Nmr in Biomedicine are provided here courtesy of Wiley

RESOURCES