Abstract
Background
Clinical treatment of cardiac arrhythmia by radiofrequency ablation (RFA) currently lacks quantitative and precise visualization of lesion formation in the myocardium during the procedure. This study aims at evaluating thermal dose (TD) imaging obtained from real-time magnetic resonance (MR) thermometry on the heart as a relevant indicator of the thermal lesion extent.
Methods
MR temperature mapping based on the Proton Resonance Frequency Shift (PRFS) method was performed at 1.5 T on the heart, with 4 to 5 slices acquired per heartbeat. Respiratory motion was compensated using navigator-based slice tracking. Residual in-plane motion and related magnetic susceptibility artifacts were corrected online. The standard deviation of temperature was measured on healthy volunteers (N = 5) in both ventricles. On animals, the MR-compatible catheter was positioned and visualized in the left ventricle (LV) using a bSSFP pulse sequence with active catheter tracking. Twelve MR-guided RFA were performed on three sheep in vivo at various locations in left ventricle (LV). The dimensions of the thermal lesions measured on thermal dose images, on 3D T1-weighted (T1-w) images acquired immediately after the ablation and at gross pathology were correlated.
Results
MR thermometry uncertainty was 1.5 °C on average over more than 96% of the pixels covering the left and right ventricles, on each volunteer. On animals, catheter repositioning in the LV with active slice tracking was successfully performed and each ablation could be monitored in real-time by MR thermometry and thermal dosimetry. Thermal lesion dimensions on TD maps were found to be highly correlated with those observed on post-ablation T1-w images (R = 0.87) that also correlated (R = 0.89) with measurements at gross pathology.
Conclusions
Quantitative TD mapping from real-time rapid CMR thermometry during catheter-based RFA is feasible. It provides a direct assessment of the lesion extent in the myocardium with precision in the range of one millimeter. Real-time MR thermometry and thermal dosimetry may improve safety and efficacy of the RFA procedure by offering a reliable indicator of therapy outcome during the procedure.
Electronic supplementary material
The online version of this article (doi:10.1186/s12968-017-0323-0) contains supplementary material, which is available to authorized users.
Keywords: MR thermometry, Radiofrequency ablation, Catheter, Arrhythmia, Cardiac, Electrophysiology
Background
Radiofrequency ablation (RFA) is a well-established procedure to treat cardiac arrhythmias by inducing small thermal lesions in the myocardium. When the area responsible for the arrhythmia is isolated or destructed, durable sinus rhythm is restored and the mechanical function of the heart may even improve in some cases. To prevent recurrence of the arrhythmia, created lesions must be transmural and permanent. Success rate of this procedure is currently limited by a lack of visualization of lesion formation during the procedure [1, 2]. To date, X-Ray fluoroscopy remains the reference imaging modality for navigating catheters in cardiac chambers for electrical mapping and delivery of therapeutic RFA. However, the intrinsic poor soft tissue contrast of this modality does not allow visualization of myocardial injury. Catheter tip temperature rise, changes in tissue electrical impedance and delivered power showed poor correlation with the actual lesion size [3]. On the other hand, an excessive temperature exposure may translate into collateral extra cardiac damages in organs such as lungs, bronchi, phrenic nerves or esophagus [4]. Therefore, there is a need to develop new strategies to offer a more reliable monitoring lesion formation for increased patient safety and improvement of the therapeutic efficiency.
Magnetic Resonance (MR) thermometry using the Proton Resonance Frequency Shift (PRFS) method has been shown to be a relevant technique for the monitoring of the thermal treatment in a number of organs including liver [5, 6], prostate [7], uterus [8] and brain [9, 10]. In addition, it has been demonstrated that MR thermometry-derived cumulative thermal dose (TD) mapping provides a good correspondence between temperature increases for a given time of exposure and the resulting thermal lesion dimensions, using a lethal TD threshold equivalent to a constant heating at 43 °C for 240 min [11].
Recent studies demonstrated the feasibility of cardiovascular MR (CMR) thermometry with electrocardiogram (ECG) and echo-navigator triggering to address challenges inherent to heart contraction and respiratory motion [12–14]. Correction strategies were proposed to compensate for susceptibility artifacts associated with magnetic field changes due to breathing [15, 16]. A spatial resolution of 1.6 × 1.6 × 3 mm3 was recently reported with an imaging time of ~100 ms per slice [14]. However, visualization of temperature increase remained limited to a small number of pixels and in vivo data were reported on a limited number of preclinical experiments. To our knowledge, no studies reported correlation between the TD maps derived from MR thermometry and the RFA lesion sizes in the context of the treatment of cardiac arrhythmia.
This study aims at demonstrating the feasibility of simultaneous RF thermal ablation and real-time CMR TD mapping for predicting ablated lesion extent.
Methods
Volunteer study
CMR thermometry was evaluated on 5 healthy volunteers under free-breathing conditions. Each volunteer was informed about the protocol and consented to participate to this study.
Animal study
Animal experiments were conducted in vivo on sheep (N = 3) weighting 46–54 kg. Anesthesia was induced and maintained with an intravenous injection of ketamine (40 mg/kg/h, IM, Virbac, Carros, France) and 2 mg/kg/h midazolam (Mylan, Canonsburg, PA, U.S.A.). Two MR-compatible catheters were positioned using X-ray angiography system (Toshiba InfiniX, Toshiba Medical, Nasu, Japan). An eight-French MR-compatible catheter (Biosense Webster, Diamond Bar, CA, U.S.A.) was inserted into the right ventricle (RV) for cardiac pacing. A ten-French MR-compatible catheter (MRI Interventions, Irvine, CA, U.S.A.) was positioned into the LV for RFA. This catheter was equipped with 4 MR micro-coils to allow active device tracking under CMR (Fig. 1a red arrows, and b) and with 2 electrophysiological (EP) electrodes (Fig. 1a, blue arrows). The animals were installed in a supine position in the MR scanner and were ventilated using a MR-compatible ventilator (Aestivia, General Electric, Fairfield, CT, U.S.A.) with a respiratory rate of 15 breaths per minute (bpm). Vital signs including cardiac rhythm, intra-arterial pressure, capnometry and rectal temperature, were monitored during all the experiments (Carescape, General Electric, Fairfield, CT, U.S.A.).
Radiofrequency ablation device
The LV catheter was connected to a clinical RF generator (Stockert Medical Solution, Freiburg, Germany) located outside the Faraday cage, using the ClearTrace system [17] (MRI Interventions, Irvine, CA, U.S.A.). A RF current at 500 kHz was sent between the catheter tip electrode and an adhesive return electrode (MONOPlate, ERBE, Tübingen, Germany) positioned on animal skin. Electrical isolation between RF generator and MR scanner was achieved by low pass filters introduced in the transmission line [13] to prevent RF artifacts on MR acquisitions due to possible interferences between RFA device and MR receivers. The catheter was irrigated with a saline solution (0.9% concentration) with a flow rate of 3 mL/min during all the procedure and 10–12 mL/min during RFA. The heart was paced with the catheter located into the RV to reduce the risk of inducing ventricular arrhythmia commonly observed in large animal models during RFA. EP signals were measured between the tip electrode and the 2nd electrode of catheters in both ventricles and displayed on a clinical EP acquisition system (Bard, LabSystem PRO, Boston Scientific, Marlborough, MA, U.S.A.). A numerical low-pass Butterworth filter (cutoff frequency = 30 Hz, 8th order) was applied to LV signal for denoising.
Imaging protocol
CMR was performed on a 1.5 T MR system (Avanto, Siemens Healthcare, Erlangen, Germany) using two clinically used 16-channel cardiac coils (Invivo Corporation, Gainesville, FL, U.S.A.). After scout views were acquired, the imaging protocol was set as follow:
In volunteers, the receiver coils were positioned anterior and posterior around the chest. CMR temperature images were acquired in short axis orientation under free breathing.
In animals, the receiver coils were positioned left and right to the chest. Catheter navigation was performed under CMR guidance to position the catheter at different locations within the LV. Then, CMR temperature images were acquired under assisted ventilation during RF ablation. 3D T1-weighted images were acquired after completion of the ablations to assess lesion sizes. Details on each acquisition sequence are provided below.
Catheter navigation
Real-time guidance of the catheter was performed using an interactive sequence made of a balanced steady state free precession (bSSFP) pulse sequence (TE/TR = 1.93/183 ms, resolution = 1.8 × 1.8 × 4 mm3, Field of view (FoV) = 225×225 mm2, Flip angle (FA) = 45°) interleaved with a catheter tracking module (BEAT IRTTT: Interactive Real Time Tip Tracking, Siemens Healthcare, Erlangen, Germany). This module consisted in a 3D encoding of the micro-coils position (Fig. 1c) along three successive gradient orientations with a total duration of 20 ms. The 3D position of the catheter coils was computed online and three slices were automatically repositioned accordingly to align the center of the slice on the catheter tip (Fig. 1d). Images were displayed on a remote monitor inside the CMR room to provide visual feedback to the operator.
Dynamic temperature mapping
A fat-saturated, single-shot Echo Planar Imaging (EPI) sequence was combined with GRAPPA parallel acquisition technique (acceleration factor of 2) to achieve a 1.6×1.6×3 mm3 spatial resolution, TE/TR/FA = 18–20 ms/110 ms/60°, bandwidth = 1576 Hz/px, FoV = 180×180 mm2, 6/8 partial Fourier. Up to five slices (spacing between slices = 0.3 mm) were acquired at each heartbeat during 250–300 consecutive cardiac cycles with ECG triggering (Fig. 1e). Saturation slabs were positioned along the FoV in the phase encoding direction to avoid aliasing and two additional saturation bands were set parallel to the imaging slices to reduce the signal of blood inside the cavities (Fig. 1f), as proposed in [18]. A crossed-pair navigator was positioned on the dome of the liver to adjust the slice position (with 0.6 tracking factor) in real-time for compensating through-plane respiratory motion (Fig. 1g) [19]. On volunteers, the group of slices was positioned in short axis orientation at the medium part of the heart to evaluate the precision of the method on both ventricles. On sheep, the group of slices was positioned accordingly to intersect the catheter tip, previously visualized during the MR-guided navigation. Before the RFA, a series of 70 repetitions was first acquired to assess the CMR thermometry stability in the myocardium surrounding the catheter tip. Then an identical EPI acquisition was run for 250–300 repetitions and RF energy was started at the 50th image.
MR thermometry pipeline
Acquired EPI raw data were streamed online via TCP/IP to a remote computer for image reconstruction using the Gadgetron framework [20]. Reconstruction pipeline included EPI ghost-correction using three central line of k-space, followed by GRAPPA parallel reconstruction [14]. Zero filling was applied prior to Fourier transform of data, resulting in a matrix size of 224×224 px and a reconstructed pixel size of 0.8×0.8 mm2 in plane.
The PRFS thermometry method computes temperature change ΔT from difference between phase image φ t and a reference phase image φ ref acquired prior to heating:
Where γ is the gyromagnetic ratio (=42.58 MHz/T), α is the PRF temperature coefficient (= − 0.0094 ppm/°C), B 0 is the magnetic field strength (=1.5 T) and TE is the echo time. The use of the PRFS technique on moving organs such as the heart requires motion and associated susceptibility artifacts correction.
Motion correction
Through-plane respiratory motion was compensated by a real-time update of slice position from the echo-navigator readings measured at the liver-lung interface. A Principal Component Analysis (PCA)-based optical-flow algorithm was used to estimate residual in-plane motion on magnitude images [21]. Phase images were registered at a fixed position using the calculated motion fields.
Correction of respiratory-induced phase variations
Due to magnetic field variations associated with respiration, susceptibility artifacts induce variations on phase images. Thus direct computation of temperature maps from registered phase images may lead to significant errors. A PCA-based method was performed to build a parameterized flow model that establishes a relationship between motions fields and phase variations over the 30 first repetitions of the time series (learning step). This model was then used to correct artifacts of the following repetitions during the interventional procedure. This correction method was introduced in an earlier study [22].
Phase drift compensation was also performed to minimize the effect of fluctuations of magnetic field induced by the heating of magnet components during EPI scanning [14, 23]. Finally, temperature maps were temporally filtered on a pixel-by-pixel basis with a low-pass Butterworth filter (cutoff frequency of 0.14 Hz), taking advantage of the high temporal resolution (<1 s) compared to the slow evolution of the temperature in tissue.
To assess the thermometry precision, the temporal average μT and the temporal standard deviation σT were computed on each pixel over ~2 min 30 s of acquisition without heating on healthy volunteers. Regions of interest (ROIs) were hand-drawn around the LV and the RV to perform statistical analysis.
Thermal dose mapping
First introduced by Sapareto et al. [11], cumulative TD was formulated in each pixel as follows:
Where T(t) = Tref + ΔT is the absolute temperature at the time t. Rectal temperature served as a reference temperature value Tref. The dose corresponding to a constant heating at 43 °C for 240 min was considered as the lethal threshold. Computed TD maps were corrected to take in account the uncertainty of the temperature measurement σT, using the following formula [24]:
Where σT is the temperature standard deviation map, calculated on each pixel before RFA over 20 repetitions (~20 s). Computed temperature and TD maps were sent in real-time to a dedicated thermometry software (Thermoguide, Image Guided Therapy, Pessac, France) for display and online monitoring of the ablation.
Post-ablation lesion visualization
After each RFA, a 3D navigator-gated Turbo Flash IR pulse sequence was performed with long inversion time (TI) for blood suppression in order to visualize the lesion without the use of contrast agents, as proposed by M. Guttman et al. [25]. The parameters of the sequence were: TI = 700 ms, resolution = 1.3×1.3×2.5 mm3, FoV = 360×360 mm2, matrix = 288×288 px, ECG-triggered every two heartbeats, FA = 11°, 31 segments, 60 slices. The contrast-to-noise ratio (CNR) was measured as the inner lesion signal intensity minus surrounding tissue signal intensity divided by the standard deviation of the noise measured in a ROI located outside the body (Syngo.via software package, Siemens Healthcare, Erlangen, Germany). Apparent lesion sizes measured on the resulting images will be referred as 3D T1-w dimensions.
Macroscopic examinations
After completion of the MR-guided RFA and 3D imaging, the animals were euthanized with a lethal intravenous dose of Dolethal (Vetoquinol, Lure, France) until complete cardiac arrest was attested. The heart was rapidly excised and dissected to expose each ablation lesion. Each lesion sample was stained at 37 °C with tetrazolium chloride (TTC) solution to assess the range of ablation area. Lesions were measured with a ruler and dimensions were correlated to dimensions observed on 2-dimensional (2D) TD maps and on 3D T1-w images.
Lesion size measurements
To compare 2D TD map and 3D T1-w lesion sizes, the equivalent 2D slice of the CMR thermometry sequence displaying the largest TD dimensions was reconstructed from the 3D T1-w volume. Largest and smallest dimensions of the apparent lesion were measured on both images. To compare 3D T1-w and macroscopic lesion sizes, 2 slices were reconstructed from the 3D volume to approach the orientation of the heart sample dissection. Three measurements were performed: largest and smallest dimensions from endocardium side and depth into the myocardium.
Results
Volunteer study
The rate and maximal amplitude of breathing were 24 ± 2 bpm and 9.2 ± 4.2 mm, respectively. The use of saturation slabs apart the imaging slice group resulted in effective blood suppression in three out of five slices, as shown in Fig. 2a on averaged magnitude images. Phase images showed significant phase wraps in all volunteers at the heart-liver-lung interface, depicted by the red arrows in Fig. 2b. The results of the evaluation of the accuracy σT and average of temperature μT in the myocardium of one typical volunteer are shown as an overlay on the averaged registered magnitude images in Fig. 2c and d, respectively. A reduced temperature precision was observed in the area with remaining phase wraps. The statistical analysis of temperature distribution over all volunteers is presented in Fig. 2 (e-f) using box and whisker plots for LV and RV ROIs. Over all the volunteers, the standard deviation of temperature σT for 2 min 30 s was 1.5 ± 0.4 °C in LV and 1.5 ± 0.5 °C in RV, excluding pixels where the thermometry performance was too much degraded (σT >7 °C), representing 3.7 ± 0.9% of the pixels of the myocardium (627 ± 240 pixels were excluded over a total of 16×104 ± 3.2×103 pixels). The temporal average was 0.3 ± 0.3 °C in the LV and 0.7 ± 0.5 °C in the RV, illustrating the efficiency of phase drift correction (up to 5 °C over 2–3 min of dynamic acquisition) of the MR thermometry pipeline. Thickness of heart walls measured on conventional cine images (resolution 1.5×1.5×7 mm3, 25 phases/cardiac cycle with retrospective reconstruction) at the medium part of the heart were 4.3 ± 0.7/2.7 ± 1.3 mm (mean ± sd in Systolic/Diastolic phases) in RV and 11.6 ± 3.4/7.4 ± 2.1 mm in LV.
In vivo radiofrequency ablations monitoring
For each animal, the MR-compatible catheter could be successfully navigated and visualized at a frame rate of 3 Hz, with automatic realignment of the imaging stack on the catheter tip. The catheter could be repositioned to a new location in the LV in less than 5 min. 12 RF procedures were performed at different locations in the endocardium of the 3 sheep. The standard deviation of the temperature σT before RFA was 1 °C or below on the LV. Catheter appeared as a hypointense area with a larger diameter (6 mm) than the actual catheter dimension (3 mm). However, thermometry remained precise enough to monitor the temperature evolution in surrounding myocardial tissue.
Figure 3 presents typical CMR thermometry results of a radiofrequency application at 70 W for 40 s. The local temperature evolution around the catheter tip showed a progressive rise during energy delivery with a maximal relative temperature increase of 75 °C on the central slice, followed by tissue cooling after completion of RF delivery (Fig. 3a, zoom on heating zone in Fig. 3b). TD maps were computed online with the lethal dose colored in red (Fig. 3c, the lethal threshold corresponds to 1). The temperature increase in 5×5 pixels centered on heating spot is displayed in Fig. 3d.
ECG recording from the ablation catheter could be performed during the procedure. As displayed on Fig. 4, changes in ECG signals acquired post-ablation (Fig. 4b) differed from the signals observed before the procedure (Fig. 4a). During the ablation, the simultaneous CMR thermometry acquisition caused significant artifacts and ECG signals were barely readable during the RFA.
Table 1 reports CMR thermometry results of 12 RF procedures performed on 3 sheep. The 12 RF procedures resulted in induction of a thermal lesion and were successfully monitored by CMR thermometry and dosimetry. Remaining spatial phase wraps at the apex prevented efficient CMR thermometry for one ablation that was not reported on Table 1. An average of 59.2 ± 7.9 W of RF power was delivered for 72.5 ± 18 s, leading to a total RF energy delivery ranging from 2.8 to 7.0 kJ (mean ± sd = 4.3 ± 1.3 kJ).
Table 1.
Lesion sizes in 2D | |||||||
---|---|---|---|---|---|---|---|
Sheep # | RFA # | Power (W) | Dur. (s) | Energy (J) | Temp max (°C) | TD map (mm2) | T1-w (mm2) |
1 | 1 | 60 | 60 | 3600 | 30 °C | 9×4 | 9×5 |
2 | 60 | 90 | 5400 | 55 °C | 6.5×4 | 7×4 | |
3 | 70 | 100 | 7000 | 30 °C | 8×3 | 9×3 | |
4 | 70 | 90 | 6300 | 40 °C | 7.5×5 | 6.5×4 | |
2 | 1 | 60 | 60 | 3600 | 40 °C | 11×4 | 11×5 |
2 | 60 | 60 | 3600 | 45 °C | 12.5×6 | 11×5.5 | |
3 | 50 | 90 | 4500 | 35 °C | 5×8 | 6×6 | |
4 | 60 | 60 | 3600 | 45 °C | 10×6 | 9×5 | |
3 | 1 | 50 | 90 | 4500 | 55 °C | 12×10 | 10×7 |
2 | 50 | 60 | 3000 | 30 °C | 8×8 | 8×6 | |
3 | 50 | 70 | 3500 | 55 °C | 10.5×7 | 10×7 | |
4 | 70 | 40 | 2800 | 75 °C | 10×7 | 10×7 |
Lesion extent maximal dimensions were measured in 2D on the cumulative TD maps and on the equivalent slice of the 3D T1-w volume
Post-ablation lesion size measurements
Figure 5a displays typical lesion visualization on 2D slice reconstructed from the 3D T1-w volume after a RFA, together with the cumulative TD map (Fig. 5b) obtained in real-time during the ablation. The 3D T1-w images obtained with an IR pulse sequence with a TI of 700 ms allowed visualization of lesion core in hyper signal with a CNR of 88.8 ± 38.3 with the surrounding tissue. The volume was acquired after each new ablation with an acquisition time in the range of 15 min to 25 min, depending on the acceptance rate of the echo-navigator and the heart cycle duration. The dimensions of this lesion performed at 60 W for 60 s were 9x5 mm2 on T1-w image and 10x6 mm2 on TD map. Correlation between ablation size on T1-w images and TD maps is shown in Fig. 6a with a Pearson’s coefficient R = 0.87 and a linear regression slope of 1.01 and an off-set of 0.58 mm. In Fig. 5 are presented T1-w volume (Fig. 5c) and macroscopic views (Fig. 5d) of the same representative lesion for purpose of comparison. The three measured dimensions were 8×6×3 mm3 and 7×6×3 mm3, respectively. Correlation study between T1-w volume and macroscopic measurements is presented Fig. 6 b with a Pearson’s coefficient R = 0.89 and a linear regression slope of 0.92 and an off-set of 0.00 mm. The Bland-Altman analysis displayed in Figs. 6c and d showed no bias depending on the ablation size. Only five measurements were found outside the 95% confidence interval. 2D lesion size estimation from TD maps was found larger than T1-w images measurements: 0.63 ± 1.3 mm (Fig. 6c) and 3D lesion size measurements on T1-w volume was found larger than macroscopic measurement: 0.46 ± 1.2 mm (Fig. 6 d). No correlation analysis could be performed between 2D TD maps and macroscopic measurements since no reliable 3D correspondence could be performed between orientation/position of 2D CMR slice and gross pathology dissection.
Discussion
One major challenge of RFA for the treatment of cardiac arrhythmia is the ability to visualize the lesion formation in the myocardium in real-time. Objectives of this mini-invasive therapy are to achieve a transmural ablation of the arrhythmogenic substrate while preserving the surrounding healthy tissues and to create a continuous electrical isolation through multiple adjacent RFA performed sequentially. This study demonstrates that i) CMR thermometry using our pipeline allows real-time tissue temperature monitoring, including during RF delivery, ii) thermal dosimetry predicts myocardial lesion size. To achieve this goal, we developed a rapid (~110 ms/image), multi-slice echo-navigated CMR thermometry method with a spatial resolution in the range of one millimeter, including real-time motion correction and compensation of associated susceptibility artifacts on phase images.
Compared to the recent paper from Ozenne et al. [14], an echo-navigated slice following technique was implemented to compensate for respiratory motion in free-breathing conditions, providing full flexibility in the slice orientation. This improvement allowed for the evaluation of CMR thermometry in five volunteers in short axis orientation and resulted in an uncertainty of approximately 1.5 °C in the myocardium over 2.5 min of acquisition. Although the wall thickness is thinner in the RV than in the LV, similar temperature precision could be achieved in both ventricles. To our knowledge, this is the first study to report temperature measurements in the RV. The echo-navigator based slice following approach was first reported by Denis de Senneville et al. [13] for CMR thermometry. However, the proposed acquisition protocol suffered from an insufficient spatial resolution (voxel size of 2.5x2.5x6 mm3) to precisely monitor RFA and temperature measurements in the RV. Moreover, only two RFA were performed in vivo in animals.
In this work, lesion sizes measured on TD, 3D T1-w imaging and gross pathology from 12 RFA performed in 3 sheep were compared. One procedure did not produce a lesion that was attributed to a poor electrode-tissue contact. The results showed an excellent correlation (slopes close to unity and off-set lower than 1 mm) between apparent lesion sizes measured in 2D on TD maps vs T1-w images and in 3D on T1-w images vs macroscopic measurements. Differences between lesion sizes remained in the range of image spatial resolution (0.8 mm). Thus, TD mapping can be used as a reliable surrogate for assessing lesion size in real-time during RFA. However, direct comparison between TD maps and gross pathology was not performed since the slice thickness of thermometry sequence (3 mm) was considered too large to assess the TD deposition in 3D with a sufficient precision, regarding the lesion size dimensions reported in Table 1.
Interestingly enough, this study shows that identical RFA parameters (Table 1, duration and power of RF delivery) can lead to different thermal lesion sizes. This may be explained by a number of factors such as the quality of the contact between the catheter tip and the tissue, potential local differences in tissue electrical and thermal conductivities, influence of blood flow, and efficiency of catheter cooling. This has been suspected clinically but the current method now provides an accurate and reliable way to directly measure the actual temperature in the myocardium and accumulated thermal dose, and thereby guide RF delivery in a much more effective way to achieve the desired endpoint [26]. This strategy should be compared to indirect means such as lesion index algorithm using contact force information.
Kolandaivelu et al. [12] proposed to use an isotherm lethal threshold of 50 °C assuming no time-temperature relation. They studied the correlation of the transmurality of the lesions (depth in short axis orientation) between MR thermometry, LGE MRI and gross pathology over 6 RF ablations in dogs. They reported that lesion transmurality by thermography was within 20% of that measured by pathology and LGE MRI. However, the insufficient temporal resolution of 10 to 20 s per image associated with breath-holding prevented an accurate and continuous monitoring of the temperature evolution. Moreover, such an approach remains hardly applicable to patients. Visualization of lesion using LGE CMR was found to overestimate the pathological lesion size [12, 27]. In the present study, we performed 3D IR T1-w imaging to visualize the lesions after the procedure without injection of contrast agent. The lesions appeared as an area of hyperintensity, well demarcated (CNR = 88.8 ± 38.3) from the surrounding edema appearing as hypointense [28] and healthy tissue. Image contrast did not change significantly (data not shown) over the time course of the experiment (>3 h). These findings were consistent with results obtained by other groups with the same CMR acquisition pulse sequence [25, 29]. This method offers the advantage of providing a high CNR without injection of a contrast agent, allowing to visualize previously created lesions at once. This is of particular interest to detect potential gaps between successive RFA that can result in arrhythmia recurrence. Redo procedure to create a continuous lesion can therefore be facilitated without requiring to wait for contrast agent elimination [27]. However, the sequence remained relatively time consuming (~15-25 min depending on the heart rate and the breathing cycle duration) and should not be used too often to avoid increasing total procedure time. Recent improvement in acceleration techniques, such as iterative reconstruction [30, 31], could help in reducing the acquisition time for potential clinical practice.
Limitations
This study aimed to investigate if the thermal dose can predict the actual lesion size. However, it was not designed to demonstrate that CMR thermometry can guide RF delivery in order to achieve transmural and contiguous lesions that are mandatory for efficient curative treatment of arrhythmia. As a consequence and considering this study as a first step, we chose to perform several ablations at distinct locations within the LV, to allow to characterize each RF delivery without bias from overlapping lesions.
Although reliable thermometry stability was achieved in 96% of the voxels located on the myocardium, the region near the liver-heart-lung interface remained prone to stronger spatial and temporal susceptibility variations that lead to higher temperature standard deviation σT (>7 °C). Such an observation was already reported in previous studies [13–15]. Further improvement in image processing might compensate for this artifact to retrieve accurate temperature measurements in this area.
The proposed CMR thermometry method needs to be evaluated on patients suffering from cardiac arrhythmia. Under these conditions, synchronization between the image acquisition and the ECG may be prone to errors and increased temperature uncertainty. Unobserved cardiac contraction pattern during the learning step may be difficult to compensate and justify further investigations and optimization of correction algorithms.
Echo-navigator based slice tracking was assumed to fully correct for through-plane respiratory motion. However, free-breathing patients could present more erratic respiratory motion than healthy volunteers and mechanically ventilated animals. The performance of the slice following should be evaluated under such conditions. The implementation of simultaneous active catheter tracking [32, 33] and RFA could address this limitation with a slice always positioned on the catheter tip during CMR thermometry.
The first results of thermometry stability on the RV of healthy volunteers are promising for future monitoring of RFA in this ventricle. However, the spatial resolution remains insufficient for atrial RFA considering the thickness of the atrial wall (~2-3 mm). Recently, the use of endovascular coils embedded in the catheter was proposed and may be explored to achieve high spatial resolution [34, 35] for monitoring temperature change in the atria.
The current MR-compatible catheters did not allow precise EP recording during RFA during CMR, due to perturbation induced by CMR gradient switching. Only two contact EP electrodes were available close to the catheter tip due to the presence of MR receiver coils, providing limited contact EP measurements. Thus, improvement of catheter design and associated instrumentation is also mandatory before clinical evaluation.
Conclusion
This study demonstrates that reliable thermometry can be achieved online on both ventricles using a multi-slice fast imaging sequence combined with slice tracking for respiratory motion compensation. In addition, the thermal dose mapping is highly correlated with post-ablation 3D T1-w images that were also correlated with actual lesion sizes measured at gross pathology. This real-time thermal dosimetry may provide guidance during RF delivery and therefore improve safety and efficacy of catheter ablation.
Acknowledgement
Authors gratefully acknowledge Virginie Loyer and Delphine Vieillot for assistance in animal experimentation, Marion Constantin for assistance in tissue analysis and Siemens Healthcare for advices in pulse sequences implementation.
Funding
This work received the financial support from the French National Investments for the Future Programs ANR-10-IAHU-04 (IHU Liryc), Laboratory of Excellence ANR-10-LABX-57 (TRAIL) and the research programs ANR-11-TecSan-003-01 (TACIT) and Equipex ANR-11-EQPX-0030 (MUSIC).
Availability of data and materials
Not applicable.
Authors’ contributions
ST conducted the study, analyzed the data and did the primary writing of the manuscript. ST, PB, MLC and RS participated to the design of MR thermometry pulse sequence and to data acquisition. ST, VO and BDS were involved in the design of the image processing pipeline including compensation of motion and associated artifacts. ST and MLC optimized the imaging protocol. ST, PB, VO, MLC, AC, FS and BQ carried out animal experiments. FS, AC and PJ carried out catheter positioning in animal and provided valuable contribution as electrophysiologists. AV was involved with image visualization tool development. HC had the idea for 3D post ablation MR imaging of the acute RF lesions. BQ conceived of the study, and participated in its design and coordination. ST, PJ, HC, MLC, VO, PB, BDS and BQ drafted the manuscript. All authors read and approved the final manuscript.
Authors’ information
Not applicable.
Competing interests
ST, MLC, AV and RS are employees of Siemens Healthcare. No other authors have competing interests to declare.
Consent for publication
Not applicable.
Ethics approval and consent to participate
Each volunteer was informed about the protocol and consented to participate to this study. The experimental protocol was approved by the local ethics committee called CCPRB (Comité de Protection des Personnes Sud-Ouest et Outre Mer III). The experimental protocol in animal was approved by the local ethics committee.
Abbreviations
- 2D
2-dimensional
- 3D
3-dimensional
- bSSFP
Balanced steady state free precession
- CMR
Cardiovascular magnetic resonance
- CNR
Contrast to noise ratio
- ECG
Electrocardiogram
- EP
Electrophysiology
- EPI
Echo Planar Imaging
- FA
Flip angle
- FoV
Field of view
- IR
Inversion recovery
- IRTTT
Interactive real time tip tracking
- LGE
Late Gadolinium Enhancement
- LV
Left Ventricle
- MR
Magnetic resonance
- MRI
Magnetic resonance imaging
- PCA
Principal component analysis
- PRFS
Proton resonance frequency shift
- RFA
Radiofrequency ablation
- ROI
Region of interest
- RV
Right ventricle
- T1-w
T1 weighted
- TCP/IP
Transmission Control Protocol/Internet Protocol
- TD
Thermal dose
- TE
Echo time
- TI
Inversion time
- TR
Repetition time
- TTC
Tetrazolium chloride
Additional file
Contributor Information
Solenn Toupin, Phone: +33 5 35 38 19 52, Email: toupin.solenn@gmail.com.
Pierre Bour, Email: pierre.bour@ihu-liryc.fr.
Matthieu Lepetit-Coiffé, Email: matthieu.lepetit-coiffe@siemens.com.
Valéry Ozenne, Email: valery.ozenne@ihu-liryc.fr.
Baudouin Denis de Senneville, Email: b.desenneville@gmail.com.
Rainer Schneider, Email: schneider.rainer@siemens.com.
Alexis Vaussy, Email: alexis.vaussy@siemens.com.
Arnaud Chaumeil, Email: arnaud.chaumeil@ihu-liryc.fr.
Hubert Cochet, Email: hubert.cauchet@ihu-liryc.fr.
Frédéric Sacher, Email: frederic.sacher@ihu-liryc.fr.
Pierre Jaïs, Email: pierre.jais@ihu-liryc.fr.
Bruno Quesson, Email: bruno.quesson@u-bordeaux.fr.
References
- 1.Tanner H, Hindricks G, Volkmer M, Furniss S, KüHlkamp V, Lacroix D, De Chillou C, Almendral J, Caponi D, Kuck K-H, Kottkamp H. Catheter Ablation of Recurrent Scar-Related Ventricular Tachycardia Using Electroanatomical Mapping and Irrigated Ablation Technology: Results of the Prospective Multicenter Euro-VT-Study. J Cardiovasc Electrophysiol. 2010;21:47–53. doi: 10.1111/j.1540-8167.2009.01563.x. [DOI] [PubMed] [Google Scholar]
- 2.Weerasooriya R, Khairy P, Litalien J, Macle L, Hocini M, Sacher F, Lellouche N, Knecht S, Wright M, Nault I, Miyazaki S, Scavee C, Clementy J, Haissaguerre M, Jais P. Catheter Ablation for Atrial Fibrillation. J Am Coll Cardiol. 2011;57:160–166. doi: 10.1016/j.jacc.2010.05.061. [DOI] [PubMed] [Google Scholar]
- 3.Eick OJ. Temperature controlled radiofrequency ablation. Indian Pacing Electrophysiol J. 2002;2:66. [PMC free article] [PubMed] [Google Scholar]
- 4.Cappato R, Calkins H, Chen S-A, Davies W, Iesaka Y, Kalman J, Kim Y-H, Klein G, Natale A, Packer D, Skanes A, Ambrogi F, Biganzoli E. Updated Worldwide Survey on the Methods, Efficacy, and Safety of Catheter Ablation for Human Atrial Fibrillation. Circ Arrhythm Electrophysiol. 2010;3:32–38. doi: 10.1161/CIRCEP.109.859116. [DOI] [PubMed] [Google Scholar]
- 5.Seror O, Lepetit-Coiffé M, Le Bail B, de Senneville BD, Trillaud H, Moonen C, Quesson B. Real time monitoring of radiofrequency ablation based on MR thermometry and thermal dose in the pig liver in vivo. Eur Radiol. 2008;18:408–416. doi: 10.1007/s00330-007-0761-4. [DOI] [PubMed] [Google Scholar]
- 6.Holbrook AB, Santos JM, Kaye E, Rieke V, Pauly KB. Real-time MR thermometry for monitoring HIFU ablations of the liver. Magn Reson Med. 2010;63:365–373. doi: 10.1002/mrm.22206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Ramsay E, Mougenot C, Köhler M, Bronskill M, Klotz L, Haider MA, Chopra R. MR thermometry in the human prostate gland at 3.0 T for transurethral ultrasound therapy: Prostate MR Thermometry at 3 T. J Magn Reson Imaging. 2013;38:1564–1571. doi: 10.1002/jmri.24063. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.McDannold N, Tempany CM, Fennessy FM, So MJ, Rybicki FJ, Stewart EA, Jolesz FA, Hynynen K. Uterine leiomyomas: MR imaging–based thermometry and thermal dosimetry during focused ultrasound thermal ablation 1. Radiology. 2006;240:263–272. doi: 10.1148/radiol.2401050717. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.de Senneville BD, Mougenot C, Quesson B, Dragonu I, Grenier N, Moonen CTW. MR thermometry for monitoring tumor ablation. Eur Radiol. 2007;17:2401–2410. doi: 10.1007/s00330-007-0646-6. [DOI] [PubMed] [Google Scholar]
- 10.Lipsman N, Schwartz ML, Huang Y, Lee L, Sankar T, Chapman M, Hynynen K, Lozano AM. MR-guided focused ultrasound thalamotomy for essential tremor: a proof-of-concept study. Lancet Neurol. 2013;12:462–468. doi: 10.1016/S1474-4422(13)70048-6. [DOI] [PubMed] [Google Scholar]
- 11.Sapareto SA, Dewey WC. Thermal dose determination in cancer theray. Int Contact Lens Clin. 1989;16:340–346. doi: 10.1016/0892-8967(89)90015-1. [DOI] [Google Scholar]
- 12.Kolandaivelu A, Zviman MM, Castro V, Lardo AC, Berger RD, Halperin HR. Noninvasive assessment of tissue heating during cardiac radiofrequency ablation using MRI thermography. Circ Arrhythm Electrophysiol. 2010;3:521–529. doi: 10.1161/CIRCEP.110.942433. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.de Senneville BD, Roujol S, Jaïs P, Moonen CTW, Herigault G, Quesson B. Feasibility of fast MR-thermometry during cardiac radiofrequency ablation: Feasibility Of Fast Mr-Thermometry During Cardiac Rf Ablation. NMR Biomed. 2012;25:556–562. doi: 10.1002/nbm.1771. [DOI] [PubMed] [Google Scholar]
- 14.Ozenne V, Toupin S, Bour P, de Senneville BD, Lepetit-Coiffé M, Boissenin M, Benois-Pineau J, Hansen MS, Inati SJ, Govari A, others: Improved cardiac magnetic resonance thermometry and dosimetry for monitoring lesion formation during catheter ablation. Magn Reson Med 2016. Advance online publication.doi:10.1002/mrm.26158. [DOI] [PubMed]
- 15.Grissom WA, Rieke V, Holbrook AB, Medan Y, Lustig M, Santos J, McConnell MV, Pauly KB. Hybrid referenceless and multibaseline subtraction MR thermometry for monitoring thermal therapies in moving organs. Med Phys. 2010;37:5014. doi: 10.1118/1.3475943. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.de Senneville BD, Ries M, Maclair G, Moonen C. MR-Guided Thermotherapy of Abdominal Organs Using a Robust PCA-Based Motion Descriptor. IEEE Trans Med Imaging. 2011;30:1987–1995. doi: 10.1109/TMI.2011.2161095. [DOI] [PubMed] [Google Scholar]
- 17.Vergara GR, Vijayakumar S, Kholmovski EG, Blauer JJE, Guttman MA, Gloschat C, Payne G, Vij K, Akoum NW, Daccarett M, McGann CJ, MacLeod RS, Marrouche NF. Real-time magnetic resonance imaging–guided radiofrequency atrial ablation and visualization of lesion formation at 3 Tesla. Heart Rhythm. 2011;8:295–303. doi: 10.1016/j.hrthm.2010.10.032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Hey S, Cernicanu A, de Senneville BD, Roujol S, Ries M, Jaïs P, Moonen CTW, Quesson B. Towards optimized MR thermometry of the human heart at 3 T: CARDIAC MR THERMOMETRY AT 3 T. NMR Biomed. 2012;25:35–43. doi: 10.1002/nbm.1709. [DOI] [PubMed] [Google Scholar]
- 19.Köhler MO, Denis de Senneville B, Quesson B, Moonen CTW, Ries M. Spectrally selective pencil-beam navigator for motion compensation of MR-guided high-intensity focused ultrasound therapy of abdominal organs. Magn Reson Med. 2011;66:102–111. doi: 10.1002/mrm.22784. [DOI] [PubMed] [Google Scholar]
- 20.Hansen MS, Sørensen TS. Gadgetron: An open source framework for medical image reconstruction: Gadgetron. Magn Reson Med. 2013;69:1768–1776. doi: 10.1002/mrm.24389. [DOI] [PubMed] [Google Scholar]
- 21.Denis de Senneville B, El Hamidi A, Moonen C. A Direct PCA-Based Approach for Real-Time Description of Physiological Organ Deformations. IEEE Trans Med Imaging. 2015;34:974–982. doi: 10.1109/TMI.2014.2371995. [DOI] [PubMed] [Google Scholar]
- 22.Roujol S, Ries M, Quesson B, Moonen C, Denis de Senneville B. Real-time MR-thermometry and dosimetry for interventional guidance on abdominal organs. Magn Reson Med. 2010;63:1080–1087. doi: 10.1002/mrm.22309. [DOI] [PubMed] [Google Scholar]
- 23.Bing C, Staruch RM, Tillander M, Köhler MO, Mougenot C, Ylihautala M, et al. Drift correction for accurate PRF-shift MR thermometry during mild hyperthermia treatments with MR-HIFU. Int J Hyperthermia. 2016;32:673–87. [DOI] [PMC free article] [PubMed]
- 24.Weidensteiner C, Quesson B, Caire-Gana B, Kerioui N, Rullier A, Trillaud H, Moonen CT. Real-time MR temperature mapping of rabbit liver in vivo during thermal ablation. Magn Reson Med. 2003;50:322–330. doi: 10.1002/mrm.10521. [DOI] [PubMed] [Google Scholar]
- 25.Guttman M, Kolandaivelu A, Fink S, Halperin H, Herzka DA. Towards MRI-guided cardiac ablation procedures with no contrast agent: safety and efficacy considerations. J Cardiovasc Magn Reson. 2016;18(Suppl 1):213. doi: 10.1186/1532-429X-18-S1-P213. [DOI] [Google Scholar]
- 26.Ikeda A, Nakagawa H, Lambert H, Shah DC, Fonck E, Yulzari A, Sharma T, Pitha JV, Lazzara R, Jackman WM. Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine Heart: Electrogram Amplitude, Impedance, and Electrode Temperature Are Poor Predictors of Electrode-Tissue Contact Force and Lesion Size. Circ Arrhythm Electrophysiol. 2014;7:1174–1180. doi: 10.1161/CIRCEP.113.001094. [DOI] [PubMed] [Google Scholar]
- 27.Dickfeld T, Kato R, Zviman M, Lai S, Meininger G, Lardo AC, Roguin A, Blumke D, Berger R, Calkins H, Halperin H. Characterization of Radiofrequency Ablation Lesions With Gadolinium-Enhanced Cardiovascular Magnetic Resonance Imaging. J Am Coll Cardiol. 2006;47:370–378. doi: 10.1016/j.jacc.2005.07.070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Celik H, Ramanan V, Barry J, Ghate S, Leber V, Oduneye S, Gu Y, Jamali M, Ghugre N, Stainsby JA, et al. Intrinsic contrast for characterization of acute radiofrequency ablation lesions. Circ Arrhythm Electrophysiol. 2014;7:718–727. doi: 10.1161/CIRCEP.113.001163. [DOI] [PubMed] [Google Scholar]
- 29.Kholmovski E, Ranjan R, Angel N, Vijayakumar S, Marrouche NF. Visibility of RF ablation lesions in native T1-weighted MRI reduces with time after ablation. J Cardiovasc Magn Reson. 2016;18(Suppl 1):196. doi: 10.1186/1532-429X-18-S1-P196. [DOI] [Google Scholar]
- 30.Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182–1195. doi: 10.1002/mrm.21391. [DOI] [PubMed] [Google Scholar]
- 31.Forman C, Piccini D, Grimm R, Hutter J, Hornegger J, Zenge MO. Reduction of respiratory motion artifacts for free-breathing whole-heart coronary MRA by weighted iterative reconstruction: Weighted Iterative Reconstruction for CMRA. Magn Reson Med. 2015;73:1885–1895. doi: 10.1002/mrm.25321. [DOI] [PubMed] [Google Scholar]
- 32.Dumoulin CL, Souza SP, Darrow RD. Real-time position monitoring of invasive devices using magnetic resonance. Magn Reson Med. 1993;29:411–415. doi: 10.1002/mrm.1910290322. [DOI] [PubMed] [Google Scholar]
- 33.Wang P, Unal O. Motion-compensated real-time MR thermometry augmented by tracking coils: Real-time MR Thermometry. J Magn Reson Imaging. 2015;41:851–857. doi: 10.1002/jmri.24574. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Volland NA, Kholmovski EG, Parker DL, Hadley JR. Initial feasibility testing of limited field of view magnetic resonance thermometry using a local cardiac radiofrequency coil: Limited FOV MR Thermometry with Local Coil. Magn Reson Med. 2013;70:994–1004. doi: 10.1002/mrm.24534. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Ertürk MA, Sathyanarayana Hegde S, Bottomley PA. Radiofrequency Ablation, MR Thermometry, and High-Spatial-Resolution MR Parametric Imaging with a Single, Minimally Invasive Device. Radiology. 2016;281:927–32. [DOI] [PMC free article] [PubMed]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Not applicable.