Fluid-structure interaction study of transcatheter aortic valve dynamics using smoothed particle hydrodynamics

Wenbin Mao 1, Kewei Li 1,1, Wei Sun 1,*
PMCID: PMC5289304  NIHMSID: NIHMS829925  PMID: 27844463

Abstract

Computational modeling of heart valve dynamics incorporating both fluid dynamics and valve structural responses has been challenging. In this study, we developed a novel fully-coupled fluid-structure interaction (FSI) model using smoothed particle hydrodynamics (SPH). A previously developed nonlinear finite element (FE) model of transcatheter aortic valves (TAV) was utilized to couple with SPH to simulate valve leaflet dynamics throughout the entire cardiac cycle. Comparative simulations were performed to investigate the impact of using FE-only models versus FSI models, as well as an isotropic versus an anisotropic leaflet material model in TAV simulations. From the results, substantial differences in leaflet kinematics between FE-only and FSI models were observed, and the FSI model could capture the realistic leaflet dynamic deformation due to its more accurate spatial and temporal loading conditions imposed on the leaflets. The stress and the strain distributions were similar between the FE and FSI simulations. However, the peak stresses were different due to the water hammer effect induced by the flow inertia in the FSI model during the closing phase, which led to 13%–28% lower peak stresses in the FE-only model compared to that of the FSI model. The simulation results also indicated that tissue anisotropy had a minor impact on hemodynamics of the valve. However, a lower tissue stiffness in the radial direction of the leaflets could reduce the leaflet peak stress caused by the water hammer effect. It is hoped that the developed FSI models can serve as an effective tool to better assess valve dynamics and optimize next generation TAV designs.

Keywords: Transcatheter aortic valve, Fluid-structure interaction, Smoothed particle hydrodynamics, Finite element method, Hemodynamics, Bioprosthetic heart valve

1. Introduction

Computational analysis of bio-prosthetic heart valve (BHV) dynamics under pulsatile loading conditions has been challenging primarily due to technical difficulties involved in the modeling of nonlinear large deformation of the leaflets, unsteady flow transition to turbulence, and nonlinear interactions between the leaflets and the surrounding fluid. Many decoupled finite element (FE) and computational fluid dynamics (CFD) simulations have been performed to investigate valve leaflet dynamics (Kim et al. 2008; Koch et al. 2010; Li and Sun 2010; Loerakker et al. 2013; Mohammadi et al. 2009; Morganti et al. 2015; Saleeb et al. 2013; Sun et al. 2010), however, structural responses and hemodynamics of a BHV cannot be accurately evaluated using either FE models or CFD models alone due to strong coupling effects between the blood flow and very flexible and stretchable leaflets. In recent years, FSI simulations of BHV opening and closing dynamics have been actively pursued (Marom 2014). Most of these FSI models adopted mesh-based methods, which can be broadly classified into two categories: a) boundary conforming methods, in which the grid moves with the boundary; and b) non-boundary conforming methods, in which the structure is immersed in the fixed background fluid grid. In boundary conforming methods, the arbitrary Lagrangian-Eulerian (ALE) approach is probably the most popular one reported in the literature (Katayama et al. 2008). A major limitation of the ALE approach lies in its time-consuming re-meshing process, which is usually required to maintain a high quality mesh throughout the simulation (large deformation of valve leaflets can severely distort the fluid mesh, leading to bad quality meshes and convergence issues). The immersed boundary method (IBM) (Peskin 1977) is one of the most popular non-boundary conforming methods. The IBM method has been extended to second-order accurate, adaptive grids, and sharp interface (Borazjani 2013; Flamini et al. 2015; Griffith 2012; Griffith et al. 2009; Hsu et al. 2015; Kamensky et al. 2015). Other methods, such as the fictitious domain method (Astorino et al. 2009; Shadden et al. 2010), and the operator splitting method used by LS-Dyna (Ranga et al. 2006; Sturla et al. 2013; Sundaram et al. 2015; Wu et al. 2015) have been used to simulate FSI of BHVs. These approaches are clearly promising in simulating flow through BHVs, facilitating device design and providing clinical implications. However, there are several limitations associated with these methods, including time-consuming re-meshing, numerical instabilities, high computational cost, and challenges dealing with valve closing, which remain to be addressed (Marom 2014).

Recently, there has been a trend toward mesh-free methods that have the potential to overcome the challenges in mesh-based methods (Marom 2014). An example of a mesh-free method is the smoothed particle hydrodynamics (SPH) method (Liu and Liu 2003; Monaghan 1992). SPH is a meshless, Lagrangian particle-based method, in which a continuum medium, such as fluid, is discretized as a set of particles distributed over the solution domain without the need of a spatial mesh (SIMULIA 2010). The method’s Lagrangian nature, associated with the absence of a fixed mesh, is its main strength. Difficulties associated with fluid flow and structural problems involving large deformations and free surfaces can be resolved by SPH in a relatively natural way. Therefore, it has emerged as a new FSI method to assess the hemodynamic responses of BHVs and blood flow in the left ventricle (Shahriari et al. 2012a; Shahriari et al. 2012b; Toma et al. 2016a; Toma et al. 2016b; Yuan et al. 2013).

In this study, we developed a novel, fully-coupled FSI model using SPH that could capture the hemodynamics and structural responses of a BHV during the full cardiac cycle. A previously developed FE model of transcatheter aortic valves (TAV) (Li and Sun 2010; Sun et al. 2010) was utilized and coupled with SPH in ABAQUS/Explicit 6.13 (Dassault Systèmes SIMULIA Corp., Johnston, RI) (SIMULIA 2010). With the developed FSI TAV model, we further investigated the differences in simulation results between a structural FE-only model and a full-coupled FSI model. For instance, the “water-hammer” effect, which is a pressure surge when the blood is forced to stop suddenly at the valve closure, can be naturally evaluated in a FSI study, but not in a FE study. In addition, we utilized the developed computational models to study the impact of leaflet material anisotropy on the TAV leaflet dynamics.

2. Methods

2.1 Smoothed particle hydrodynamics formulation

In this study, the SPH method was used to model the blood flow passing through a TAV. The SPH methodology is thus briefly introduced as follows. As illustrated by a 2D example in Figure 1, each SPH particle has its own physical properties, A, which are interpolated by a kernel function, W, using the properties of its neighboring particles.

Figure 1.

Figure 1

A 2D illustration of the basic principle of the smoothed particle hydrodynamics methodology. The property A of particle ‘a’ is determined by the properties of its neighboring particles, for example, ‘b’, based on an interpolating kernel function (W) which is a function of the smoothing length, h, and the distance between particle ‘a’ and its neighboring particle ‘b’.

This concept is interpreted numerically using a summation (Monaghan 1992):

A(ra)=bmbAbρbW(ra-rb,h) (1)

where Ab denotes any physical property at particle ‘b’ within the neighboring domain (limited by the influence length h of the kernel) of particle ‘a’ at position ra. Particle ‘b’ has mass mb, position ra, and density ρb. In this study, a cubic spline kernel function was adopted. Using this equation and its derivatives, the governing equations of fluid flow (conservation of mass, momentum and energy) can be rewritten under the form of SPH formulation.

The time derivative form of the conservation of mass gives

dρadt=bmbVab·aWab. (2)

Here ∇aWab is the gradient of the kernel function regarding the coordinates of given particle ‘a’ and Vab = VaVb denotes the relative velocity vector between particles ‘a’ and ‘b’.

Similarly, the conservation of momentum under the SPH scheme can be written as

dVadt=-bmb(Pa+Pbρaρb)aWab+bmb(μa+μb)Vabρaρbrab2rab·aWab (3)

where P is pressure and μ is the dynamic viscosity of the fluid. The movement of the particles is governed by

dradt=Va (4)

In ABAQUS, the pressure is related to the density by the Mie-Grüneisen equation of state (SIMULIA 2010). In this work, the linear UsUp Hugoniot form is used, given by

p=ρ0c02η(1-sη)2(1-Γ0η2)+Γ0ρ0Em (5)

where Γ0 is a material constant, c0is the artificial speed of sound, η = 1 − ρ0 is the nominal volumetric compressive strain, ρ is the current density, and Em is internal energy per unit mass. It is commonly assumed that blood is isothermal, incompressible and Newtonian (Marom 2014). Thus, in this study, we set a reference density of ρ0 = 1056 kg/m3, and a dynamic viscosity of μ = 0.0035 Pa·s. For the blood flow in the aorta, the speed of sound is high compared to the bulk velocity of the blood, therefore an artificial sound speed of c0 = 145 m/s was employed to avoid very small computational time steps while keeping density fluctuations within a small range to maintain incompressible flow behavior. Other material parameters were chosen as s = 0 and Γ0 = 0.

2.2 Finite Element Modeling of TAV

TAV leaflets are generally fabricated from either thin (~0.25 mm) glutaraldehyde-treated bovine pericardium (GLBP) or porcine pericardium (Zajarias and Cribier 2009). In this study, GLBP was selected as a representative valve leaflet material. GLBP was assumed to be an incompressible, anisotropic, nonlinear, hyperelastic material (Sun and Sacks 2005); thus the strain energy function, W, can be expressed by a fiber-reinforced hyperelastic material model (MHGO) based on the work of Holzapfel, Gasser and Ogden (Gasser et al. 2006; Holzapfel et al. 2000),

W=C10{exp[C01(I¯1-3)]-1}+k12k2i=12{exp[k2(I¯4i-1)2]-1}+1D(J-1)2, (6)

where C10, C01, k1, k2 and D are material constants, Ī1 and Ī4i are the deviatoric strain invariants. C10 and C01 are used to describe the matrix material. D is a material constant to impose incompressibility, and J is the determinant of the deformation gradient. k1 is a positive constant with the dimension of stress to describe the fiber material and k2 is a dimensionless parameter. The fiber orientations is defined by θ, which is the angle between the fiber orientation and the circumferential direction (Figure 2b).

Figure 2.

Figure 2

(a) GLBP equi-biaxial response from Sun’s thesis (Sun 2004) (open circles), with anisotropic MHGO model fit (black lines), and the averaged response (red squares) with isotropic Ogden model fit (red line). (b) Diagram of the leaflet material orientations and the fiber orientations, m01 and m02, drawn on the 2D leaflet schematic.

In the heart valve industry, it is well known that leaflets made of pericardial tissues can have various degrees of anisotropy, however, the impact of leaflet material anisotropy on the valve stress distribution and hemodynamics has not been investigated using a fully-coupled FSI model. Thus, in this study, an incompressible, isotropic, hyperelastic Ogden model (Ogden 1972) was used for comparison to the anisotropic MHGO model. The Ogden strain energy function is given by

W=i=1N2μiai2(λ¯1ai+λ¯2ai+λ¯3ai-3) (7)

where μi and ai are material constants, and λ̄i are the modified principal stretches.

The FE model of a generic size 23 TAV developed previously (Li and Sun 2010) was used in this study. The TAV leaflets were discretized using 7896 large-strain brick (C3D8R) elements with a thickness of 0.25 mm. For an anisotropic MGHO model, local material orientations were defined for the leaflets at each element. The material properties defined by the constitutive law of Eq. (6) were incorporated into ABAQUS via a user subroutine UMAT (Sun et al. 2008). The contact between each leaflet was modeled using a general contact algorithm in ABAQUS. The TAV stent was not included in the simulations because the deformation of the metallic stent is assumed to be minimal. Thus, the stent-attachment nodes were fixed in space to mimic attachment to the non-deformable stent. The GLBP leaflet material parameters were determined by fitting the biaxial testing data of GLBP leaflets presented by Sun (Sun 2004). The constitutive model was able to capture the response very well (Figure 2a). The two fiber families were assigned identical material properties and oriented symmetrically about the leaflet material axis in the radial direction illustrated in Figure 2b. The averaged material response from the biaxial testing data was fitted using a four-parameter isotropic Ogden model. The material parameters of both models are given in Table 1. The mass density of the leaflets was defined as 1100 kg/m3.

Table 1.

GLBP material properties.

Modified-Holzapfel model C10 (kPa) C01 k1 (kPa) k2 θ(°) D (kPa−1)
30.03 3.47 74.5 63.19 43.11 1.00e–5
Isotropic Ogden model μ1(kPa) a1 μ2(kPa) a2
19.58 67.74 260.56 27.47

2.3 FE-SPH model setup and boundary conditions

The aortic valve flow condition was modeled following the ISO 5840-3:2013 testing conditions by adopting a 36.8 mm diameter tubular structure, as illustrated in Figure 3. The tubes upstream and downstream from the valve were long enough to eliminate the boundary effect on the region of interest. The length of downstream SPH particles was chosen as 110 mm, which corresponds to a typical, normal human ascending aorta length (Sugawara et al. 2008). (Note that the downstream length in our study includes sinus height). To accommodate a size 23 TAV, rigid gaskets and skirts were introduced to the model (Figure 3). The tube, gaskets, and skirts were all fixed in space. The mesh sensitivity of SPH particles was checked on the FSI-Ogden model. Three different mesh densities were used with spatial resolutions of 0.9mm, 0.7mm, and 0.5mm, respectively. This leads to approximately 370,300 one-node PC3D elements, 790,200 elements, and 2,171,800 elements in the fluid domain, respectively. The relative error of the peak velocity through the valve was 16.5% between the coarse and fine meshes, and 4.1% between the medium and fine meshes. A reasonable grid convergence was obtained on the medium mesh density (for further details refer to Appendix A.1). Therefore, the results presented in this study employed the medium mesh density. Two rigid plates were used to apply pressure boundary conditions on the blood volume. This setting implicitly assumes that the inlet and outlet velocities have flat profiles as the particles adjacent to the plates are driven by the flat plates, which is commonly adopted for FSI simulations (Marom 2014; Sturla et al. 2013). A nearly-closed configuration of the leaflets (see the inset in Figure 6) was chosen as the stress-free initial configuration. Physiological pressure waveforms at a heart rate of 75 bpm (Figure 4a) were used to obtain the pressure gradient between the left ventricle and aorta (Figure 4b), which was applied in the simulation. The beginning of the ejection phase was selected as the starting point of the simulation. Two cardiac cycles were conducted and the results from the second cycle were analyzed. We found that the difference between the first and second cycle was within 10% error. Simulations were performed on an Intel Xeon E5-2670 cluster with 64 cores and required 155 hours for the FSI model using the medium mesh in one cardiac cycle. In the FE-only simulations, three leaflets were simulated with the stent-attachment nodes fixed in space. The same pressure gradient waveform was uniformly applied to the surface of the leaflets to deform the leaflets.

Figure 3.

Figure 3

Three-dimensional FE-SPH model for TAV opening and closing simulations. The model consists of a 36.8 mm diameter rigid tubular structure, a rigid gasket and skirt to seal the gap between the tube and valve, two rigid plates at the aortic and ventricular sides respectively, blood particles (half SPH particles are shown in the fluid domain for clarity), and three flexible TAV leaflets.

Figure 6.

Figure 6

A series of the time-dependent leaflet deformed shape (the red curve in the center inset) is depicted with respect to the perpendicular bisection plane of the leaflet. The red curves represent the opening profiles, and the blue curves represent the closing profiles. (a) FSI-Ogden model. (b) FSI-MHGO model. (c) FE-Ogden model. (d) FE-MHGO model.

Figure 4.

Figure 4

(a) Wiggers diagram representing physiological pressure waveforms at a heart rate of 75 bpm (Wiggers). (b) Pressure boundary conditions applied to the plates.

3. Results

3.1 TAV leaflet kinematics

Leaflet kinematics were evaluated during the TAV opening and closing phases. Leaflet motion was monitored through the time-dependent radial position of the midpoint of the leaflet free edge (see red dot in Figure 2b). Since the three leaflets move similarly during the opening and closing phases, the averaged radial motion of the midpoints is shown in Figure 5a. Several qualitative differences in the leaflet motion between the FE-only and FSI simulations were observed. Firstly, the opening process in the FE-only models began immediately once the pressure load was applied; while in the FSI models, there was approximately 20 ms delay before the free edges of the leaflets began to open. The total valve opening time in the FE models was short, about 5 ms, compared to about 40 ms in the FSI models. A larger radial excursion of the free edge was observed for the FE models than that of the FSI models.

Figure 5.

Figure 5

(a) Time-dependent radial position of the midpoint of leaflet free edge during a cardiac cycle. (b) Time-dependent GOA of TAV during a cardiac cycle. Note for the sake of clarity, a full cardiac cycle was truncated to show the variation from 0 s to 0.4 s.

The geometric orifice area (GOA) of each TAV was calculated during a cardiac cycle from our simulations. An in-house MATLAB (The MathWorks, Inc., Natick, MA) code was implemented to obtain the GOA based on images of the deformed valve from the top-view. The time-dependent GOA curves in Figure 5b showed a similar behavior. The results from the FSI models had smaller GOAs than those from the FE models. The closing process in the FE models was faster than that in the FSI models. The FE results had higher frequency fluctuations in the leaflet motion during the opening and ejection phases.

A series of the time-dependent leaflet deformed shape, represented by a middle-line at the perpendicular bisection plane of the leaflet, was shown in Figure 6 for all models. The red curves represent the opening profiles, and the blue curves are the closing profiles. As shown in Figure 6, at t=0 ms, all models started from the same initial configurations. For the FSI models, at the beginning of the opening process, the leaflet deformation was initiated at the bottom of the valve near the attachment edge and the belly region of the leaflets (see leaflet profiles at 20 ms in Figure 6a, 6b), then expanded towards the free edge region as the fluid pushed through the valve. However, for the FE model, at t = 5 ms, the valves almost reached the maximum opening orifice area. During the closing phase, the coaptation of the free edges occurred earlier in the FE models (at approximately t = 240 ms) than in the FSI models (at approximately t = 270 ms). The coaptation of the leaflets was achieved in all models before the maximum transvalvular pressure gradient was reached. The fully-closed leaflet shapes were almost the same in both FE and FSI models. At other phases, however, relatively larger discrepancies in leaflet deformed curvature were observed between the FE-only and FSI simulations.

The material models also had a noticeable impact on the leaflet kinematics. The isotropic Ogden model showed a slightly stiffer leaflet behavior compared to the anisotropic MHGO model in the radial direction, since the radial position of free edge midpoint was lower during the ejection phase for both the FE and FSI models. In the FSI-MHGO model, large magnitude and low frequency fluctuations of the leaflets during the ejection phase were observed. Similarly, the fully-closed configuration in the MHGO models (see Figure 6) was more concave than that of the Ogden models.

3.2 Structural stress and strain responses

Maximum principal true strain (LE) contour plots of the deformed leaflets are shown in Figure 7, 8, 9 for the FE-MHGO, FSI-MHGO, and FSI-Ogden models, respectively. In the FE-MHGO model (Figure 7), during the ejection phases at t= 5 ms, the peak strain was observed below the commissure region along the leaflet-stent attachment line with a peak value about 0.22. During the valve closing phase, large strain was concentrated in the commissure and belly regions. A peak value of 0.185 was found at the stent-leaflet attachment hinge regions in the fully-closed configuration. Compared to the results of the FE-MHGO model, in the FSI-MHGO model (Figure 8), the peak value of strain was much smaller (i.e., about 0.112 at t=40 ms) during the opening phase and mainly concentrated below the commissure region along the leaflet-stent attachment line. In the fully-closed configuration, a similar strain pattern was observed with a peak strain of 0.185.

Figure 7.

Figure 7

Opening and closing of TAV leaflets colored by the maximum principal strain from the FE-MHGO model. Note the different scale for each time instance.

Figure 8.

Figure 8

Opening and closing of TAV leaflets colored by the maximum principal strain from the FSI-MHGO model. Note the different scale for each time instance.

Figure 9.

Figure 9

Opening and closing of TAV leaflets colored by the maximum principal strain from the FSI-Ogden model. Note the different scale for each time instance.

Comparing the FSI-Ogden and FSI-MHGO models, a similar pattern of strain distribution was found during the opening phase, see Figures 8 and 9. However, the strain distribution was different in the fully-closed configuration. A peak strain value of 0.144 was observed in the FSI-Ogden model and concentrated in the commissure regions.

The maximum principal stress distribution on the leaflets in the fully open and closed configurations are depicted in Figure 10. The stress distribution and peak value were very similar in the fully-closed phase between the FE-only and FSI models. However, in the full open phase, the peak stress value was much smaller in the FSI models compared to the FE models. Differences between the isotropic and anisotropic models were observed. The peak stress value at the commissure in the fully-closed configuration reached 1.13 MPa in the FSI-MHGO model compared to the value of 1.42 MPa in the FSI-Ogden model. The peak stress in the belly region was 0.74 MPa and 0.59 MPa for the FSI-MHGO and FSI-Ogden models, respectively. Meanwhile, the high stress regions in the MHGO models shifted downwards from the commissure region and distributed along the attachment line.

Figure 10.

Figure 10

Contour plots of the maximum principal stress (in MPa) for each valve in the fully-open (t = 40 ms) and fully-closed (t = 300 ms) configurations. (a) FE-Ogden model. (b) FSI-Ogden model. (c) FE-MHGO model. (d) FSI-MHGO model.

3.3 Hemodynamics

The TAV hemodynamics were evaluated based on the FSI-MHGO and FSI-Ogden models. The waveforms of the transvalvular pressure drop are plotted in Figure 11a. The transvalvular pressure was recorded one diameter upstream and three diameters downstream from the annulus of the valve following the ISO-5840 guideline. The two pressure curves were very similar during systole. The FSI-MHGO model had a slightly lower peak pressure drop (11.7 mmHg at t = 25 ms) compared to the FSI-Ogden model (14.5 mmHg at t = 25 ms). Note that the transvalvular pressure was not the same as the pressure boundary conditions applied on the plates. Especially during the valve closing phase, large oscillations can be seen due to the strong effect of water inertia. The axial hydrodynamic force acting on the leaflets in a cardiac cycle is shown in Figure 11b. In systole, this force is equal to the pull-out force on the TAV that may cause device migration under the forward blood flow. The peak values of pull-out force are 0.26 N and 0.20 N for the FSI-Ogden and FSI-MHGO models, respectively. In the diastole, the peak hydrodynamic forces reach up to −6.9 N and −6.1 N at t = 270 ms for the FSI-Ogden and FSI-MHGO models, respectively. The maximum static pressure force of −5.4 N on the valve in diastole, represented as the dotted line in Figure 11b.

Figure 11.

Figure 11

Comparison of (a) transvalvular pressure gradient, (b) hydrodynamic force acting on the leaflets (the dotted line represent the maximum hydrostatic pressure force in diastole), (c) maximum velocity at 15 mm downstream the TAV annulus (basal ring), (d) flow rate through the TAV in a cardiac cycle between the FSI-Ogden and FSI-MHGO models. For the sake of clarity, only a partial cardiac cycle from 0 s to 0.4 s is shown.

The maximum flow velocity at an axial distance of 15 mm downstream of the TAV annulus was monitored throughout the cardiac cycle (see Figure 11c). The velocity reached a peak value of 1.40 m/s and 1.36 m/s for the FSI-Ogden and FSI-MHGO models at t = 80 ms, respectively. The time-dependent flow rate, which was averaged among several cross-sectional planes starting from the upstream to the downstream of the valve, is depicted in Figure 11d. The peak flow rates in the FSI-Ogden and FSI-MGHO models were 13.1 L/min and 14.7 L/min, respectively. Other hemodynamic parameters, such as mean systolic pressure difference (MSPD), peak systolic pressure difference (PSPD), effective orifice area (EOA), regurgitant volume (RV) are listed in Table 2.The EOA was calculated according to the formula defined as (Yoganathan et al. 2004): EOA(cm2)=Qrms51.6ΔP, where Qrms is the root mean square systolic flow rate (cm3/s), and ΔP is the mean systolic pressure drop (mmHg).

Table 2.

Hemodynamic parameters calculated from the results of FSI simulations.

Model MSPD (mmHg) PSPD (mmHg) EOA (cm2) RV (ml)
FSI-Ogden 5.3 14.5 1.22 3.2
FSI-MGHO 5.0 11.7 1.50 3.1

MSPD: Mean systolic pressure drop; PSPD: Peak systolic pressure drop; EOA: Effective orifice area; RV: Regurgitant volume.

Flow velocity vector field were assessed on a cross-section through the symmetry plane of the valve at six time instances throughout the cardiac cycle (Figure 12). During the opening phase (t = 25 ms and 40 ms), the forward flow developed along the axial direction with an almost circular jet profile. At t = 100 ms, a wide central jet with a maximum velocity of 1.40 m/s was observed immediately downstream of the valve. Vortices were identified beside the central jet downstream the tube expansion. These vortices developed progressively downstream throughout the ejection phase (t = 230 ms). During the closing phase (t = 270 ms), due to the closing of the leaflets and the reverse pressure gradient, a large reverse flow was generated through the central gap between the leaflets. The regurgitant jet only existed for a very short time due to the rapid closing, thus did not result in large regurgitant volumes (see Table 2). After valve closure, the flow was nearly hydrostatic with negligible velocities at t = 300 ms.

Figure 12.

Figure 12

Blood velocity vector fields at different time instances for the FSI-Ogden (the left hand side) and FSI-MHGO (the right hand side) models. Note the color legend on the top applies to the first four time instances, while the bottom legend corresponds to the last two instances.

Additionally, we calculated the rapid valve opening and closing times and ejection time defined as the time from the initial opening to complete closure of the valve. According to references (Leyh et al. 1999; Sturla et al. 2013; Wu et al. 2015), rapid valve opening times can be calculated when the radial displacement curves of the nodes have a high positive and high negative slope, respectively. All numerical predictions for these quantities are listed in Table 3.

Table 3.

RVOT, RVCT, and ET calculated from our models and measured in vivo (Leyh et al. 1999).

FSI-Ogden FSI-MHGO FE-Ogden FE-MHGO In vivo data
RVOT (ms) 40 40 5 6 57.5±11.1
RVCT (ms) 35 35 8 9 39.5±5
ET (ms) 270 270 240 240 329±63

RVOT: rapid valve opening time; RVCT: rapid valve closing time; ET: ejection time.

4. Discussion

In this study, we developed, to our knowledge, the first SPH-based FSI model to investigate TAV leaflet dynamics throughout the whole cardiac cycle. In the simulations, we also utilized nonlinear hyperelastic material models to describe the experimentally derived leaflet material properties. Comparative simulations were performed to study the effects of using FE-only models versus FSI models, as well as the isotropic Ogden material model versus the anisotropic MHGO material model.

Valve leaflet dynamics were only recently evaluated in vivo using 4-dimensional (4D), volume-rendered CT scans (Holmes and Mack 2015; Makkar et al. 2015). Notably, a report by Makkar et al (Makkar et al. 2015) demonstrated that by using 4D CT scans, reduced TAV leaflet motion was observed in patients who had a stroke after TAV implantation. However, such 4D CT scans are not commonly available yet, and post-procedural CT scans are not part of the standard of care in TAV patients. To date, ex vivo bench testing probably remains to be the best approach to quantify BHV leaflet dynamics for the device design evaluation and optimization. However, the experimental studies of valve deformation are relatively expensive and difficult to perform because of practical limitations in measurements very close to the leaflets and valve housing. FE-only models, due to an inaccurate pressure field applied to the leaflets and the lack of fluid damping as illustrated in this study, led to inaccurate simulations of leaflet dynamics; consequently, a match between FE result and experiment data is usually hard to achieve (Kim et al. 2008; Saleeb et al. 2013).

From the results of this study, it can be seen that FE-only models tend to overestimate the pressure force exerted on the leaflets in systole, leading to abnormally fast opening of the valve, whereas the results from the FSI models were comparable with in vivo measurement (Leyh et al. 1999), as demonstrated in other FSI results as well (Piatti et al. 2015; Sturla et al. 2013). Some FE-only studies used a numerical damping force to account for the effect of surrounding fluid (Kim et al. 2008; Marom et al. 2012; Saleeb et al. 2013). Since the damping coefficient is not known a priori, it can only be chosen by a trial-and-error process (Hsu et al. 2015). Moreover, the valve orifice area was larger in the FE models in the fully open configuration than the FSI models, as observed in (Sturla et al. 2013), which is consistent with our findings.

In the closing phase, by comparison of the hydrostatic pressure force with the hydrodynamic force on the leaflet in Figure 11b, it can be seen that the large fluctuation in the closing phase was not captured by the hydrostatic force. The peak hydrodynamic forces reach up to −6.9 N and −6.1 N at t = 270 ms for the FSI-Ogden and FSI-MHGO models, respectively. These forces would exceed the maximum hydrostatic force on the valve (Fp = ΔPmax·A = 5.4 N) in the closed configuration by about 13%–28%. Thus, when considering the dynamic pull-out force on TAVs, the fluid inertial force due to the water hammer effect in the closing phase should be considered. Moreover, the larger peak hydrodynamic force obtained in the FSI-Ogden model may be attributed to the relatively stiffer material response in the radial direction than that of the FSI-MHGO model.

Our FSI simulations replicated similar leaflet dynamics during the cardiac cycle as observed in the experiments (Dellimore et al. 2012; Vismara et al. 2010). Noticeably, the leaflets were pushed open by the fluid flow from the belly of the valve first and progressively to the free edge of the leaflets. Thus, the fully-opened leaflets appears to be convex in the FSI simulations, whereas in the FE-only simulations, the free edge of the leaflets were deformed instantaneously once the uniform pressure loading is applied, resulting in a concave shape of the fully-opened leaflets.

The peak flow rates obtained from the FSI-Ogden and FSI-MGHO models were 13.1 L/min and 14.7 L/min, respectively. These values agree with the findings in (Sturla et al. 2013) (with a peak flow rate of 15.7 L/min) using a similar pressure drop waveform as the boundary condition. Other FSI studies (Hsu et al. 2015; Piatti et al. 2015) obtained higher peak flow rates possibly due to the larger transvalvular pressure drop used in their simulations. The cardiac output from our simulations is smaller than the physiological measurement. This may be related to two simplifications of our FSI model setups. Firstly, the straight tube geometry with sudden contraction and expansion used in our models would cause a larger pressure drop compared to the anatomical aortic root geometry with sinuses. Secondly, rigid aortic wall used in our models is not physiological, which may cause additional errors. Nonetheless, other hemodynamic parameters, such as EOA, were comparable to the published clinical data (Bauer et al. 2004; Stühle et al. 2011).

There are several limitations of the study, which should be considered when interpreting our results. Firstly, we used a straight and rigid tube connection downstream from the aortic annulus, neglecting the anatomic geometry of the sinuses and the compliant tissue behavior of the aortic root. This simplification of a straight tube may affect the kinematics of the leaflets and associated stress distribution (Katayama et al. 2008; Leyh et al. 1999). The influence of arterial wall elasticity in heart valve simulations was studied by Hsu et al. (Hsu et al. 2014). They found that the major difference between rigid and elastic models occurred immediately following valve closure. The rigid wall model results had larger and longer oscillations in the flow rate and the valve movement compared to those of the compliant arterial wall model. Therefore, the water hammer impact in our model may be overestimated due to the absence of aortic compliance. Secondly, the SPH solver in ABAQUS has no turbulence model, and the no-slip wall boundary condition is not fully constrained. This omission is likely to affect the blood flow in the boundary layers region. Thirdly, the stent of TAV was ignored in our model. It may affect the hemodynamics of the valve. However, it is unlikely to deteriorate our conclusions due to the side-by-side comparison between the FSI and FE models.

In conclusion, the major novelty of the present study is the development of a SPH-FE coupled FSI model that is capable of modeling the overall structural and hemodynamic characteristics of TAVs throughout the whole cardiac cycle. Through a comparative study using FE-only models, the FSI models were able to capture the leaflet kinematics during the valve opening and closing phases due to more realistic spatial and temporal loading conditions. The stress and the strain distributions were similar between the FE and FSI simulations. However, the peak stresses were different due to the water hammer effect induced by the flow inertia in the FSI model during the closing phase, which led to 13%–28% lower peak stresses in the FE-only model compared to that of the FSI model. The simulation results also indicated that tissue anisotropy had a minor impact on hemodynamics of the valve. However, a lower tissue stiffness in the radial direction of the leaflets could reduce the leaflet peak stress caused by the water hammer effect. It is hoped that the developed FSI models can serve as an effective tool to better assess valve dynamics and optimize next generation TAV designs.

Acknowledgments

Research for this work was funded in part by NIH HL104080 and HL108240 grants. Wenbin Mao was also supported by an American Heart Association Post-doctoral Fellowship 15POST25910002.

Appendix

A.1 Mesh sensitivity of fluid domain

The mesh sensitivity of SPH particles was analyzed by using the FSI-Ogden model. Three different mesh densities were used with spatial resolutions of 0.9mm (coarse mesh), 0.7mm (medium mesh), and 0.5mm (fine mesh) respectively. This leads to approximately 370,300 one-node PC3D elements, 790,200 elements, and 2,171,800 elements in the fluid domain, respectively. The relative error of the peak velocity through the valve was 16.5% between the coarse and fine meshes, and 4.1% between the medium and fine meshes. A reasonable grid convergence was obtained on the medium mesh density. Therefore, the results presented in this study employed the medium mesh density.

The results of the corresponding FSI analyses were compared in terms of (a) transvalvular pressure gradient, (b) hydrodynamic force acting on the leaflets, (c) maximum velocity at 15 mm downstream the TAV annulus, (d) flow rate through the TAV.

graphic file with name nihms829925u1.jpg

The comparison of the respective results showed differences in peak values within 5% between the medium and the fine mesh, resulting in acceptable level of grid convergence.

A.2 Fiber orientation on the leaflet

The circumferential and radial fiber directions of the FSI-MHGO model at two time instants are shown. The fiber orientation in open (t = 100 ms) and closed (t = 300 ms) configurations: C-direction in (a), and R-direction in (c) at t = 100 ms; C-direction and R-direction in (b) and (d), respectively, at t = 300 ms.

For ease in visualization, the fiber directions are projected onto the undeformed leaflet configuration. The C-direction fibers in the belly region, where the leaflets subject to high strain and stress, undergo significant rotation in closed configuration compared to the open configuration, whereas the R-direction fibers undergo large rotation in the free edge region. The differences in the C- and R-direction fiber orientation signifies the presence of anisotropy in BAV tissues.

graphic file with name nihms829925u2.jpg

Footnotes

Conflict of Interest

The authors declare that they have no conflict of interest.

Statement of human studies

No human studies were carried out by the authors for this article.

Statement of animal studies

No animal studies were carried out by the authors for this article.

References

  1. Astorino M, Gerbeau J-F, Pantz O, Traoré K-F. Fluid–structure interaction and multi-body contact: application to aortic valves. Computer Methods in Applied Mechanics and Engineering. 2009;198:3603–3612. [Google Scholar]
  2. Bauer F, et al. Acute improvement in global and regional left ventricular systolic function after percutaneous heart valve implantation in patients with symptomatic aortic stenosis. Circulation. 2004;110:1473–1476. doi: 10.1161/01.CIR.0000134961.36773.D6. [DOI] [PubMed] [Google Scholar]
  3. Borazjani I. Fluid–structure interaction, immersed boundary-finite element method simulations of bio-prosthetic heart valves. Computer Methods in Applied Mechanics and Engineering. 2013;257:103–116. [Google Scholar]
  4. Dellimore K, Kemp I, Rodriguez R, Scheffer C. In vitro characterization of an aortic bioprosthetic valve using Doppler echocardiography and qualitative flow visualization. Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE; 2012; IEEE; pp. 6641–6644. [DOI] [PubMed] [Google Scholar]
  5. Flamini V, DeAnda A, Griffith BE. Immersed boundary-finite element model of fluid-structure interaction in the aortic root. arXiv preprint arXiv:150102287. 2015 doi: 10.1007/s00162-015-0374-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Gasser TC, Ogden RW, Holzapfel GA. Hyperelastic modelling of arterial layers with distributed collagen fibre orientations. Journal of the royal society interface. 2006;3:15–35. doi: 10.1098/rsif.2005.0073. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Griffith BE. Immersed boundary model of aortic heart valve dynamics with physiological driving and loading conditions. International Journal for Numerical Methods in Biomedical Engineering. 2012;28:317–345. doi: 10.1002/cnm.1445. [DOI] [PubMed] [Google Scholar]
  8. Griffith BE, Luo X, McQueen DM, Peskin CS. Simulating the fluid dynamics of natural and prosthetic heart valves using the immersed boundary method. International Journal of Applied Mechanics. 2009;1:137–177. [Google Scholar]
  9. Holmes DR, Mack MJ. Uncertainty and Possible Subclinical Valve Leaflet Thrombosis. The New England journal of medicine. 2015;373:2080–2082. doi: 10.1056/NEJMe1511683. [DOI] [PubMed] [Google Scholar]
  10. Holzapfel GA, Gasser TC, Ogden RW. A new constitutive framework for arterial wall mechanics and a comparative study of material models. Journal of elasticity and the physical science of solids. 2000;61:1–48. [Google Scholar]
  11. Hsu M-C, Kamensky D, Bazilevs Y, Sacks MS, Hughes TJ. Fluid–structure interaction analysis of bioprosthetic heart valves: significance of arterial wall deformation. Computational mechanics. 2014;54:1055–1071. doi: 10.1007/s00466-014-1059-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Hsu M-C, et al. Dynamic and fluid–structure interaction simulations of bioprosthetic heart valves using parametric design with T-splines and Fung-type material models. Computational Mechanics. 2015:1–15. doi: 10.1007/s00466-015-1166-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Kamensky D, et al. An immersogeometric variational framework for fluid–structure interaction: Application to bioprosthetic heart valves. Computer methods in applied mechanics and engineering. 2015;284:1005–1053. doi: 10.1016/j.cma.2014.10.040. [DOI] [PMC free article] [PubMed] [Google Scholar]
  14. Katayama S, Umetani N, Sugiura S, Hisada T. The sinus of Valsalva relieves abnormal stress on aortic valve leaflets by facilitating smooth closure. The Journal of thoracic and cardiovascular surgery. 2008;136:1528–1535. e1521. doi: 10.1016/j.jtcvs.2008.05.054. [DOI] [PubMed] [Google Scholar]
  15. Kim H, Lu J, Sacks MS, Chandran KB. Dynamic simulation of bioprosthetic heart valves using a stress resultant shell model. Annals of Biomedical Engineering. 2008;36:262–275. doi: 10.1007/s10439-007-9409-4. [DOI] [PubMed] [Google Scholar]
  16. Koch T, Reddy B, Zilla P, Franz T. Aortic valve leaflet mechanical properties facilitate diastolic valve function. Computer methods in biomechanics and biomedical engineering. 2010;13:225–234. doi: 10.1080/10255840903120160. [DOI] [PubMed] [Google Scholar]
  17. Leyh RG, Schmidtke C, Sievers H-H, Yacoub MH. Opening and closing characteristics of the aortic valve after different types of valve-preserving surgery. Circulation. 1999;100:2153–2160. doi: 10.1161/01.cir.100.21.2153. [DOI] [PubMed] [Google Scholar]
  18. Li K, Sun W. Simulated thin pericardial bioprosthetic valve leaflet deformation under static pressure-only loading conditions: implications for percutaneous valves. Annals of biomedical engineering. 2010;38:2690–2701. doi: 10.1007/s10439-010-0009-3. [DOI] [PubMed] [Google Scholar]
  19. Liu G-R, Liu MB. Smoothed particle hydrodynamics: a meshfree particle method. World Scientific; 2003. [Google Scholar]
  20. Loerakker S, Argento G, Oomens CW, Baaijens FP. Effects of valve geometry and tissue anisotropy on the radial stretch and coaptation area of tissue-engineered heart valves. Journal of biomechanics. 2013;46:1792–1800. doi: 10.1016/j.jbiomech.2013.05.015. [DOI] [PubMed] [Google Scholar]
  21. Makkar RR, et al. Possible Subclinical Leaflet Thrombosis in Bioprosthetic Aortic Valves. The New England journal of medicine. 2015;373:2015–2024. doi: 10.1056/NEJMoa1509233. [DOI] [PubMed] [Google Scholar]
  22. Marom G. Numerical methods for fluid–structure interaction models of aortic valves. Archives of Computational Methods in Engineering. 2014:1–26. [Google Scholar]
  23. Marom G, Haj-Ali R, Raanani E, Schäfers H-J, Rosenfeld M. A fluid–structure interaction model of the aortic valve with coaptation and compliant aortic root. Medical & biological engineering & computing. 2012;50:173–182. doi: 10.1007/s11517-011-0849-5. [DOI] [PubMed] [Google Scholar]
  24. Mohammadi H, Boughner D, Millon L, Wan W. Design and simulation of a poly (vinyl alcohol)—bacterial cellulose nanocomposite mechanical aortic heart valve prosthesis Proceedings of the Institution of Mechanical Engineers, Part H. Journal of Engineering in Medicine. 2009;223:697–711. doi: 10.1243/09544119JEIM493. [DOI] [PubMed] [Google Scholar]
  25. Monaghan JJ. Smoothed particle hydrodynamics. Annual review of astronomy and astrophysics. 1992;30:543–574. [Google Scholar]
  26. Morganti S, Auricchio F, Benson D, Gambarin F, Hartmann S, Hughes T, Reali A. Patient-specific isogeometric structural analysis of aortic valve closure. Computer Methods in Applied Mechanics and Engineering. 2015;284:508–520. [Google Scholar]
  27. Ogden R. Proceedings of the Royal Society of London A: Mathematical, Physical and Engineering Sciences. Vol. 1567. The Royal Society; 1972. Large deformation isotropic elasticity-on the correlation of theory and experiment for incompressible rubberlike solids; pp. 565–584. [Google Scholar]
  28. Peskin CS. Numerical analysis of blood flow in the heart. Journal of computational physics. 1977;25:220–252. [Google Scholar]
  29. Piatti F, et al. Hemodynamic and thrombogenic analysis of a trileaflet polymeric valve using a fluid–structure interaction approach. Journal of biomechanics. 2015;48:3641–3649. doi: 10.1016/j.jbiomech.2015.08.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Ranga A, Bouchot O, Mongrain R, Ugolini P, Cartier R. Computational simulations of the aortic valve validated by imaging data: evaluation of valve-sparing techniques. Interactive cardiovascular and thoracic surgery. 2006;5:373–378. doi: 10.1510/icvts.2005.121483. [DOI] [PubMed] [Google Scholar]
  31. Saleeb A, Kumar A, Thomas V. The important roles of tissue anisotropy and tissue-to-tissue contact on the dynamical behavior of a symmetric tri-leaflet valve during multiple cardiac pressure cycles. Medical engineering & physics. 2013;35:23–35. doi: 10.1016/j.medengphy.2012.03.006. [DOI] [PubMed] [Google Scholar]
  32. Shadden SC, Astorino M, Gerbeau J-F. Computational analysis of an aortic valve jet with Lagrangian coherent structures Chaos: An Interdisciplinary. Journal of Nonlinear Science. 2010;20:017512. doi: 10.1063/1.3272780. [DOI] [PubMed] [Google Scholar]
  33. Shahriari S, Kadem L, Rogers B, Hassan I. Smoothed particle hydrodynamics method applied to pulsatile flow inside a rigid two-dimensional model of left heart cavity. International journal for numerical methods in biomedical engineering. 2012a;28:1121–1143. doi: 10.1002/cnm.2482. [DOI] [PubMed] [Google Scholar]
  34. Shahriari S, Maleki H, Hassan I, Kadem L. Evaluation of shear stress accumulation on blood components in normal and dysfunctional bileaflet mechanical heart valves using smoothed particle hydrodynamics. Journal of biomechanics. 2012b;45:2637–2644. doi: 10.1016/j.jbiomech.2012.08.009. [DOI] [PubMed] [Google Scholar]
  35. SIMULIA. ABAQUS Analysis user’s Manual Dassault Systemes. Simulia Corp; Providence, RI: 2010. [Google Scholar]
  36. Stühle S, et al. In-vitro investigation of the hemodynamics of the Edwards Sapien™ transcatheter heart valve. Journal of Heart Valve Disease. 2011;20:53. [PubMed] [Google Scholar]
  37. Sturla F, Votta E, Stevanella M, Conti CA, Redaelli A. Impact of modeling fluid–structure interaction in the computational analysis of aortic root biomechanics. Medical engineering & physics. 2013;35:1721–1730. doi: 10.1016/j.medengphy.2013.07.015. [DOI] [PubMed] [Google Scholar]
  38. Sugawara J, Hayashi K, Yokoi T, Tanaka H. Age-associated elongation of the ascending aorta in adults. JACC: Cardiovascular Imaging. 2008;1:739–748. doi: 10.1016/j.jcmg.2008.06.010. [DOI] [PubMed] [Google Scholar]
  39. Sun W. Biomechanical simulations of heart valve biomaterials 2004 [Google Scholar]
  40. Sun W, Chaikof EL, Levenston ME. Numerical approximation of tangent moduli for finite element implementations of nonlinear hyperelastic material models. Journal of biomechanical engineering. 2008;130:061003. doi: 10.1115/1.2979872. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Sun W, Li K, Sirois E. Simulated elliptical bioprosthetic valve deformation: implications for asymmetric transcatheter valve deployment. Journal of biomechanics. 2010;43:3085–3090. doi: 10.1016/j.jbiomech.2010.08.010. [DOI] [PubMed] [Google Scholar]
  42. Sun W, Sacks MS. Finite element implementation of a generalized Fung-elastic constitutive model for planar soft tissues. Biomechanics and modeling in mechanobiology. 2005;4:190–199. doi: 10.1007/s10237-005-0075-x. [DOI] [PubMed] [Google Scholar]
  43. Sundaram GBK, Balakrishnan KR, Kumar RK. Aortic Valve Dynamics using a Fluid Structure Interaction Model-The Physiology of opening and closing. Journal of Biomechanics. 2015 doi: 10.1016/j.jbiomech.2015.05.012. [DOI] [PubMed] [Google Scholar]
  44. Toma M, Bloodworth CH, IV, Einstein DR, Pierce EL, Cochran RP, Yoganathan AP, Kunzelman KS. High-resolution subject-specific mitral valve imaging and modeling: experimental and computational methods. Biomechanics and modeling in mechanobiology. 2016a:1–12. doi: 10.1007/s10237-016-0786-1. [DOI] [PubMed] [Google Scholar]
  45. Toma M, Einstein DR, Bloodworth CH, Cochran RP, Yoganathan AP, Kunzelman KS. Fluid–structure interaction and structural analyses using a comprehensive mitral valve model with 3D chordal structure. International Journal for Numerical Methods in Biomedical Engineering. 2016b doi: 10.1002/cnm.2815. [DOI] [PMC free article] [PubMed] [Google Scholar]
  46. Vismara R, Fiore GB, Mangini A, Contino M, Lemma M, Redaelli A, Antona C. A novel approach to the in vitro hydrodynamic study of the aortic valve: mock loop development and test. ASAIO Journal. 2010;56:279–284. doi: 10.1097/MAT.0b013e3181d9c295. [DOI] [PubMed] [Google Scholar]
  47. Wiggers CJ. https://en.wikipedia.org/wiki/Carl_J._Wiggers.
  48. Wu W, et al. Fluid–Structure Interaction Model of a Percutaneous Aortic Valve: Comparison with an In Vitro Test and Feasibility Study in a Patient-Specific Case. Annals of biomedical engineering. 2015:1–14. doi: 10.1007/s10439-015-1429-x. [DOI] [PubMed] [Google Scholar]
  49. Yoganathan AP, He Z, Casey Jones S. Fluid mechanics of heart valves. Annu Rev Biomed Eng. 2004;6:331–362. doi: 10.1146/annurev.bioeng.6.040803.140111. [DOI] [PubMed] [Google Scholar]
  50. Yuan Q, Ye X, Ma H, Huang X. Fluid Structure Interaction Study of Bioprosthetic Heart Valve with FESPH Method. International Journal of Advancements in Computing Technology. 2013:5. [Google Scholar]
  51. Zajarias A, Cribier AG. Outcomes and safety of percutaneous aortic valve replacement. Journal of the American College of Cardiology. 2009;53:1829–1836. doi: 10.1016/j.jacc.2008.11.059. [DOI] [PubMed] [Google Scholar]

RESOURCES