. 2017 Jan-Feb;51(1):103–106. doi: 10.4103/0019-5413.197557

Pseudotumor due to metallosis after total elbow arthroplasty

Roberto Rotini 1, Graziano Bettelli 1, Michele Cavaciocchi 1, Lucia Savarino 1,✉
PMCID: PMC5296839  PMID: 28216759

Abstract

The incidence of primary total elbow arthroplasty (TEA) in young patients is increasing. The indications for revision surgery are also rising. Here, we report a rare case of pseudotumor detected in a patient 16 years after TEA. Intraoperative findings revealed a necrotic mass characterized by a conspicuous metallosis in the soft tissues around the prosthesis, which caused ulnar nerve dislocation. Due to this anatomical change, a lesion of the nerve was accidentally produced during revision surgery. The case report emphasizes that the indications for elbow replacement, as well as the patient education about the permanent physical limitations, should be carefully considered. Moreover, the high risks of complications related to the revision procedure and pseudotumor removal need to be addressed before surgery. The technique should be done carefully and a preliminary thorough imaging should be performed, since a newly formed mass can cause significant distortion of the anatomy.

Keywords: Total elbow arthroplasty, elbow prosthesis, pseudotumor, metallosis, revision elbow arthroplasty

MeSH terms: Joint, elbow joint, arthroplasty, prosthesis, metals, revision

INTRODUCTION

Total elbow arthroplasty (TEA) is the elective surgery for elderly patients with severe elbow osteoarthritis or unreconstructible elbow fractures.1 Expanding these indications to a young and active population can significantly increase the risk of early wear and prosthesis loosening.2 Revision surgery is difficult and the outcome is less satisfactory compared to primary implants.3 We describe a unique case of pseudotumor in a young patient who underwent revision surgery 16 years after TEA.

CASE REPORT

A 56-year-old woman presented to us 16 years after left TEA with a Kudo prosthesis (Biomet UK Ltd., Bridgend, Wales) performed in another hospital for a complex fracture of the distal humerus [Figure 1], followed by plaster immobilization for 40 days with an outcome of joint stiffness. 18 months after surgery, manipulations under anesthesia were performed, a joint motion of 70–120° was obtained, and the patient resumed her normal activity level. 9 years after surgery, she returned to the same hospital complaining about pain and elbow swelling and underwent an exploratory surgery with removal of “a broken part.” Afterwards, she resumed the same activity levels as before.

Figure 1.

Figure 1

X-ray of the elbow joint anteroposterior and lateral views showing the distal humerus fracture in a 40 years old patient

When the patient came to our institute 16 years after the index procedure, she complained about elbow pain and instability. She had one posterior and one lateral surgical scar, joint deformity and a posterolateral soft swelling [Figure 2]. The range of movement in flexion and extension was 60–120°. No vascular-neurological symptoms were observed. Radiographs [Figure 3] revealed dislocation and loosening of the implant, surrounded by a slightly radio-opaque mass. Anti inflammatory therapy and restricted use of the arm were prescribed, and after few weeks she was admitted to our unit for revision surgery. Low infection probability was suggested by scintigraphy with labelled leukocytes and a needle aspirate was performed. The cultures resulted negative, but blood leucocytes were increased, and erythrocyte sedimentation rate and C-reactive protein showed borderline values. Since periprosthetic infection could not be ruled out, a two-stage surgery was planned. She was operated in supine position, maintaining the upper limb above the chest with the best possible alignment of the elbow joint that remained dislocated and unstable.

Figure 2.

Figure 2

Clinical photograph of the elbow showing the swelling, skin discoloration and two surgical scars

Figure 3.

Figure 3

X-ray of the elbow joint anteroposterior and lateral views showing loosening of both components of the Kudo elbow prosthesis after 16 years and one surgical revision. Soft tissue swelling with metal debris

Of the two surgical scars the lateral one was followed. As the ulnar nerve was expected to lie medially to the mass, the dissection was begun on the posterior midline. During the isolation of the huge pseudotumoral mass, an oblique division of the ulnar nerve was accidentally produced [Figure 4]. The nerve had been dislocated from its normal course by the scar tissue and by the mass, and actually lied in the posterior midline of the elbow. The two ends of the ulnar nerve were dissected and isolated, and then the dissection and excision of the mass was completed. Grossly the tissue was necrotic, bloodless and dark-grey to blackish, with the unequivocal aspect of metallosis, as a consequence of the metal-on-metal friction of the dislocated components. Histology was thus deemed not necessary. In order to definitively rule out infection, multiple tissue biopsies were harvested and sent for culture and no aerobic/anaerobic bacteria were isolated. The stems were well fixed in the medullary canals.

Figure 4.

Figure 4

Intraoperative picture showing the pseudotumor mass. The ulnar nerve has been accidentally divided (arrow); the stumps of the nerve have been temporarily put close

After implant removal, thorough debridment was performed and temporary spacers (one humeral and one ulnar) were built with tobramycin and vancomycin-loaded cement supported by metal rods [Figure 5]. Ulnar nerve paralysis was present. 10 days later, the second surgery was performed. Antibiotic-loaded cement spacers were removed and again samples for culture were retrieved. After thorough debridment, a Latitude (Tornier, Warsaw, IN) anatomic prosthesis was implanted with tobramycin-loaded cement. A proper nerve repair was performed by a neurosurgeon with 8/0 polypropylene suture and fibrin glue. The postoperative course was uneventful; the arm was protected in a brace and mild elbow mobilization started after 2 days. Cultures again were negative. The patient was instructed to use her arm only for mild activities, avoiding loads over 2–3 kg.

Figure 5.

Figure 5

X-ray of the elbow joint anteroposterior and lateral views showing temporary cement spacers

At 26 months followup [Figure 6], the elbow was stable with a range of motion from 30° to 130°. Ulnar nerve paralysis was persistent with only slight sensory recovery.

Figure 6.

Figure 6

X-ray of the elbow joint anteroposterior and lateral views at 2 years after revision with Latitude prosthesis showing implant in situ

DISCUSSION

Total elbow arthroplasty is a surgical procedure that should be reserved, as a general rule, for elderly patients. Expanding the indications to younger patients increases the already high risk of complications and implant failure,4,5 with consequent need of revision surgery. TEA revision is a difficult surgical act, frequently requiring bone augmentation6,7,8 and the survival rate of these implants can be as low as 64% at 5 years.9,10,11

In this patient, the decision to perform a TEA when she was 40 years old was certainly a hazardous choice. Every effort should be done in this age group to perform fixation even in case of complex fractures. On the contrary, in this case, the retrospective evaluation of the pre surgery X-rays of the elbow, taken in the hospital where the patient had undergone TEA 16 years before, showed that the fracture pattern was not particularly complex. The correct treatment indication was therefore an internal fixation through trans-olecranon approach. Another point to be stressed is that when no alternative to prosthesis is possible, it is mandatory to instruct the young patient about the permanent functional limitations to be observed in order to avoid abnormal wear.

Cystic or solid masses resulting from wear debris have been commonly reported in patients with metal-on-metal hip prostheses12,13,14 and have been termed “pseudotumors.”15 Differently from hip surgery, only rare reports exist in literature about metallosis as a consequence of instability or loosening after TEA.16,17 However, in case of painful/loose elbow prosthesis, the pseudotumor occurrence must be suspected, together with aseptic loosening and infection.

In this condition, the revision surgery can be extremely complex. The excision required for the mass removal bears unusual risks, since anatomic relations may have been distorted. In our case, considering the two surgical procedures, it seemed highly probable that the ulnar nerve had been translocated anteriorly or that it lied in its anatomical position. Anterior transposition during primary TEA is a common practice associated with a lower risk of neurological complications.3 Unexpectedly, the nerve had been dislocated posteriorly and laterally by the development of the pseudotumor, from which it could not be differentiated both visually and by palpation. This unpredictable setting created the conditions for its lesion, a complication in which we never incurred in our experience of several dozens of primary elbow prostheses and revisions.

In the preoperative workout, deep attention had been paid to the risk of infection. Additional imaging examinations could have been performed. A magnetic resonance imaging and a computed tomography scan of the elbow would have perhaps disclosed in higher detail the shape and size of the pseudotumor, but almost certainly would not have been able to show the relationships of the ulnar nerve with the mass.

To conclude, a soft swelling around unstable elbow prosthesis is not always due to infection, but can also be caused by a pseudotumor. Surgery in this condition has an additional risk of complications related to possible anatomy distortions.

REFERENCES

  • 1.Cooney WP, Morrey BF. Elbow arthroplasty: historical perspective and emerging concepts. In: Morrey BF, Sanchez Sotelo J, editors. The Elbow and its Disorders. Philadelphia: W.B. Saunders; 2009. pp. 705–19. [Google Scholar]
  • 2.Voloshin I, Schippert DW, Kakar S, Kaye EK, Morrey BF. Complications of total elbow replacement: A systematic review. J Shoulder Elbow Surg. 2011;20:158–68. doi: 10.1016/j.jse.2010.08.026. [DOI] [PubMed] [Google Scholar]
  • 3.Kim JM, Mudgal CS, Konopka JF, Jupiter JB. Complications of total elbow arthroplasty. J Am Acad Orthop Surg. 2011;19:328–39. doi: 10.5435/00124635-201106000-00003. [DOI] [PubMed] [Google Scholar]
  • 4.Throckmorton T, Zarkadas P, Sanchez-Sotelo J, Morrey B. Failure patterns after linked semiconstrained total elbow arthroplasty for posttraumatic arthritis. J Bone Joint Surg Am. 2010;92:1432–41. doi: 10.2106/JBJS.I.00145. [DOI] [PubMed] [Google Scholar]
  • 5.Krenek L, Farng E, Zingmond D, SooHoo NF. Complication and revision rates following total elbow arthroplasty. J Hand Surg Am. 2011;36:68–73. doi: 10.1016/j.jhsa.2010.09.036. [DOI] [PubMed] [Google Scholar]
  • 6.Loebenberg MI, Adams R, O’Driscoll SW, Morrey BF. Impaction grafting in revision total elbow arthroplasty. J Bone Joint Surg Am. 2005;87:99–106. doi: 10.2106/JBJS.B.00038. [DOI] [PubMed] [Google Scholar]
  • 7.Kosy JD, Walter RP, Conboy VB, Blake SM. Use of a femoral impaction grafting system in revision total elbow arthroplasty. Ann R Coll Surg Engl. 2011;93:416–7. doi: 10.1308/003588411X582717c. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Sanchez-Sotelo J. Total elbow arthroplasty. Open Orthop J. 2011;5:115–23. doi: 10.2174/1874325001105010115. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Sneftrup SB, Jensen SL, Johannsen HV, Søjbjerg JO. Revision of failed total elbow arthroplasty with use of a linked implant. J Bone Joint Surg Br. 2006;88:78–83. doi: 10.1302/0301-620X.88B1.16446. [DOI] [PubMed] [Google Scholar]
  • 10.van der Lugt JC, Rozing PM. Outcome of revision surgery for failed primary Souter-Strathclyde total elbow prosthesis. J Shoulder Elbow Surg. 2006;15:208–14. doi: 10.1016/j.jse.2005.07.009. [DOI] [PubMed] [Google Scholar]
  • 11.Shi LL, Zurakowski D, Jones DG, Koris MJ, Thornhill TS. Semiconstrained primary and revision total elbow arthroplasty with use of the Coonrad-Morrey prosthesis. J Bone Joint Surg Am. 2007;89:1467–75. doi: 10.2106/JBJS.F.00715. [DOI] [PubMed] [Google Scholar]
  • 12.Sayed-Noor AS, Sjödén GO. Severe metallosis after total elbow arthroplasty-A case report. Hand (N Y) 2010;5:86–9. doi: 10.1007/s11552-009-9201-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Glyn-Jones S, Pandit H, Kwon YM, Doll H, Gill HS, Murray DW. Risk factors for inflammatory pseudotumour formation following hip resurfacing. J Bone Joint Surg Br. 2009;91:1566–74. doi: 10.1302/0301-620X.91B12.22287. [DOI] [PubMed] [Google Scholar]
  • 14.Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94:895–900. doi: 10.1302/0301-620X.94B7.29122. [DOI] [PubMed] [Google Scholar]
  • 15.Matthies AK, Skinner JA, Osmani H, Henckel J, Hart AJ. Pseudotumors are common in well positioned low-wearing metal-on-metal hips. Clin Orthop Relat Res. 2012;470:1895–906. doi: 10.1007/s11999-011-2201-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, et al. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–51. doi: 10.1302/0301-620X.90B7.20213. [DOI] [PubMed] [Google Scholar]
  • 17.Degreef I, Sciot R, De Smet L. Metallosis in revision total elbow arthroplasty. Complications and staging method. Acta Orthop Belg. 2008;74:753–60. [PubMed] [Google Scholar]

Articles from Indian Journal of Orthopaedics are provided here courtesy of Indian Orthopaedic Association