Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2017 Aug 17.
Published in final edited form as: RSC Adv. 2016 Aug 17;6(84):80468–80484. doi: 10.1039/C6RA16403J

Biomimetic Cross-Reactive Sensor Arrays: Prospects in Biodiagnostics

J E Fitzgerald a, H Fenniri a,
PMCID: PMC5312755  NIHMSID: NIHMS813253  PMID: 28217300

Abstract

Biomimetic cross-reactive sensor arrays have been used to detect and analyze a wide variety of vapour and liquid components in applications such as food science, public health and safety, and diagnostics. As technology has advanced over the past three decades, these systems have become selective, sensitive, and affordable. Currently, the need for non-invasive and accurate devices for early disease diagnosis remains a challenge. This review provides an overview of the various types of Biomimetic cross-reactive sensor arrays (also referred to as electronic noses and tongues in the literature), their current use and future directions, and an outlook for future technological development.

1. Introduction

1.1. Importance of non-invasive, early diagnosis

Many conventional methods of diagnosis are costly, painful, and inaccurate in early stages.1 Thus, there remains a need for devices that target all of these facets to provide early diagnosis and analysis for each individual patient that is altogether more cost effective, accurate, and non-invasive.

For example, cancer detection and treatment remain a significant challenge that affects many Americans. In 2008 alone, of the 1.6 million new diagnoses of lung cancer in the United States, 1.4 million died of the disease;2 in that same year, 1.38 million women were diagnosed with breast cancer globally, comprising 23% of invasive cancer cases in females worldwide.3 For these and all other types of cancer, treatment and maintenance costs are high. In 2010, the national cost of cancer was estimated to be $124.57 billion.4 Recent research has shown, however, that early diagnosis is a key factor for cancer survival and treatment cost, with the five year survival rate increasing from 4% to 24%, to 52% for local, regional, and distant stage disease, and the overall cost of cancer decreases for those treated in the early stages.2,4

Another prevalent challenge in medical treatment methods is the increase in highly antibiotic-resistant bacteria that cause various types of infections for many people worldwide. In 2009, over 50% of healthcare associated infections (HAIs) were caused by a resistant strain. HAIs have been estimated to be the fifth-leading cause of death in acute care hospitals, with 5-15% of patients contracting HAI during their stay. HAIs contracted resistant strains cause, on average, a 9 day increase in hospital stay, and $29,379 in additional medical costs.5 Prolonged bacterial growth in the body leads to the formation of bacterial biofilms, which are increasingly resistant to antimicrobial agents.6 Thus, early diagnosis of these infections is necessary for optimal infection treatment.

Concerning the accuracy of diagnosis, there remains a need for analytical tools for patients with chronic illnesses, such as inflammatory bowel disease, asthma, and chronic mental illnesses. Mental illness specifically is a widespread issue in the United States. In 2012, the NIH reported that 43.8 million adults suffered from a mental illness.7 In 2006, 36.2 million people paid for mental health services totalling $57.5 billion.8 In 2010, asthma was the cause for 1 in every 250 deaths worldwide due mainly to suboptimal management and delay in obtaining help during the final attack.9 The cause of many of these illnesses remains unknown. Treatment is administered on a trial-and-error basis, which reduces effectiveness and raises the risk of serious side effects. An accurate device that can provide a unique patient profile would contribute significantly to the development of a specific treatment method based on the individual's needs, and could act as a tool to understanding the physiological processes involved for each illness.

1.2. Mammalian olfactory system as a model for biomimetic sensors

Over the past 3 decades, many research efforts have looked to the mammalian olfactory system as a model for the development of a medical diagnostic device that meets all of these requirements. In March of 2015, The Guardian newspaper published an article about a dog named Frankie who could detect thyroid cancer with 88% accuracy among patients. Thyroid cancer is notoriously difficult to detect by conventional methods, and it is hard to tell if the entire tumour has been removed post-surgery. Frankie, however, was trained to lay down after smelling patients' urine samples if he detected metastatic cancer.10 Researchers have been working to develop biomimetic devices, called “electronic/artificial noses” and “electronic tongues” to detect certain olfactory elements present in both vapours and liquids that can indicate a certain disease or illness.

It is estimated that the human nose can detect as many as 10,000 to 100,000 chemicals in the form of “odorants,” which are small volatile molecules.11 Upon entering the nasal cavity, the odorants reach the olfactory epithelium and interact with the olfactory sensory neurons by dissolving into the nasal mucus, and reaching cilia extended into the nasal lumen.11 This releases a signalling cascade, whereupon the activated neurons send information to the olfactory bulb section of the brain, which then sends this information to the olfactory cortex. The cortex processes the information, and sends it to appropriate other areas of the brain responsible for odour determination and the emotional and physiological responses associated with each odour.11 Importantly, each sensory neuron has odorant receptors (ORs), which are a type of G-protein coupled receptor (GPCR), and are encoded by a multi-gene family to detect a wide variety of different odorants. Upon binding to odorants, ORs undergo conformational change, thereby converting the received chemical information into electrical information for the sensory neurons.1114 Overall, there are about 1000 genes that encode ORs, and each OR has multiple sites for odorant binding, enabling detection of more than one odorant for each OR—a characteristic called cross-reactivity.11,14 Each combination of activated receptors creates a unique signalling code for a specific odorant, making it possible to distinguish between thousands.1113 This review aims to provide a brief history and examination of current technology and future directions of both electronic noses and tongues that aim to mimic the mammalian olfactory system. These devices have proven to be successful for a wide variety of applications, providing cost-effective, minimally invasive, and highly accurate vapour and fluid component analysis.

2. History and Background Technology

The first use of the term “electronic nose” was at a conference in 1987, with the first conference dedicated specifically to artificial olfaction in 1989.15 In a 1994 publication, Gardner and Bartlett defined an “electronic nose” as: “an instrument, which comprises of an array of electronic-chemical sensors with partial specificity and an appropriate pattern recognition system, capable of recognising simple or complex odours.”15

The earliest attempts at creating such a device involved passing an odorant over sets of unique active materials, each connected to its own sensor. Upon activation of its material, the sensor would transmit an electrical signal to a processor, which would then compile an array of activated sensors to alert the user of the odorants present by cross referencing a database of known “odours” (i.e. analytes).15,16 The earliest designs employed metal oxide semiconductors as an electrical sensor to detect gases such as NO2.17,18 This was based on the principle that the conductivity of semiconductor metals changes upon variance in the atmospheric gas surrounding the sensor.18 Additionally, some of the early sensors were metal oxide semiconducting field effect transistors (MOSFETs). MOSFETs are usually constructed from a SiO2 insulating layer, with a semiconductor metal deposited on top as the gate in the circuit. A voltage is applied to maintain a constant current, and as the gas adsorbs onto the gate, the conductance of the FET changes, thereby causing the voltage to change.19,20

As more became known about the physiology of the mammalian olfactory system, sensors advanced to take advantage of these new findings, from incorporating sensors that were able to detect multiple analytes,21 to the development of a biomimetic flow chamber to enhance analyte detection at low levels.22 This allowed for a great number of analytes to be recognized using a smaller array of multi-selective sensors.16,21 The types of sensors employed in these devices vary, including gravimetric, or mass,2332 electrical,3340 and optical4158 sensors for optimal analyte detection.

Within the last two decades, electronic tongues for liquid detection have also been developed, mimicking the olfactory signalling pathways. These devices use mass,5963 electrochemical,6474 and optical sensors7578 for chemical detection. Table 1 gives an overview of the different types of sensors in both electronic noses and tongues, and outlines some of the advantages and disadvantages of each one.

Table 1. Overview of electronic nose and tongue sensor types.

Sensor type Signal Transduction Mechanism Common Materials
Mass Change in resonant frequency upon analyte adsorption Piezoelectric crystals (quartz), thin films, microcantilevers
Electrical and Electrochemical Electrical response from sensor sensor upon analyte adsorption (e.g. change in resistance, impedance, current, electrical potential) Metal oxide semiconductors, conductive polymers, single walled carbon nanotubes, lipid bilayer, metal electrodes
Optical Sensors display a shift in emission or absorption of different types of radiations (e.g. photons or electrons) upon binding with a desired analyte. Dye-coated fibre optic polymers, microbeads, thin films, barcoded resin polymer microbeads, sol-gels

The subsequent sections describe both electronic nose and tongue technology using different types of sensors, explore their use in medical diagnostics and outline future directions in device development.

3. Cross-Reactive Sensor Arrays for Volatiles (CRSA-V)

3.1. Mass Sensors

The most common mass CRSA-V devices are piezoelectric (PZ) crystals, which are usually quartz crystals. The crystals are coated with various materials that attract specific analytes. These coating materials include chemicals such as amines,24,29 enzymes,29 proteins,25,28 and polymers.29 Coated crystals are arranged in a microbalance, and their resonant frequency is recorded during exposure to a desired vapour. As analytes adsorb to the surface of the PZ crystals, the added mass leads to a reduction in the crystals' resonant frequency. A drop in frequency may also be attributed to both viscoelastic and gravimetric effects.29 Once the frequency change is recorded, the mass adsorbed to the crystal surface is determined by:

ΔF=FΔm/(Art)25,29

Where ΔF is the change in resonant frequency (Hz), F is the initial crystal resonant frequency (MHz), A is the total surface area (cm2), r is the density of crystal (g/cm3), t is the thickness of the crystal (cm), and m is the mass adsorbed (g).25,29

Mass sensors have been used in a wide variety of applications. For example, Filipov, et. al. in 2007 used quartz PZ crystals coated with Cu(II) stearate and octadecylamine to attract and detect volatile organic compounds (VOCs), known to be produced by certain types of cancer.7982 As illustrated in Figure 1, they placed a coated quartz crystal microbalance (QCM) in a thermostatic chamber filled with purified air and sealed at a relative humidity of 45%. Next, a vapour saturated with VOCs was injected into the chamber and passed over the QCM. The QCM/chamber system was connected to a frequency counter, which in turn fed the output to a computer to analyze frequency change, with results shown in Figure 2. To better calibrate the sensor to account for frequency changes caused only by VOCs, properties of the coated crystals were studied in diffused, ambient conditions as a correction factor. They further optimized the system by testing different ratios of metal ion to fatty acid/amine concentrations to promote affinity of the coatings to VOCs.24

Fig. 1.

Fig. 1

Schematic diagram of the experimental setup 24.

Fig. 2.

Fig. 2

Histogram indicating sensitivities of different coatings: HSt (1), ODA (2), CuSt2 (3), Cu(ODA)2St2 (4), Cu(ODA)4St2 (5), Cu(ODA)6St2 (6), Cu(AcAc)2 on SiO2 (7), towards different analytes: methanol (1), ethanol (2), acetone (3), diethyl ether (4), ethyl acetate (5), benzene (6), chloroform (7), n-hexane (8), toluene (9).24

Park et al. have used OR proteins in crystal coatings to test the ligand specificity and dose dependence of individual ORs as biosensors.25,28 For these two studies, commercially available QCMs were purchased and functionalized with desired ORs. In 2004, Ligand specificity and dose dependence of rat ORs were analyzed. Heterologous cells were induced to express a specific type of OR on their surface, then seeded and dried on a gold substrate to be used as a coating for the QCM.25 In 2005, an OR expressed in C-elegans was grown on E. coli, and the membrane fraction containing the OR was harvested for use as a coating. This OR coating was tested for specificity to a variety of different analytes, and optimized using the membrane fraction that showed the most sensitive affinity for diacetyl.28 This group has successfully shown that the use of ORs as biosensors can be highly efficient and selective. Furthermore, they discuss the use of the cross-reactivity of the OR sensors to create a sensor array, capable of detecting a wide variety of analytes.25,28

Some mass devices use changes in microcantilever oscillations as an indicator of analyte adsorption. Microcantilevers, such as those on an Atomic Force Microscope, are fabricated as a bilayer system, typically composed of silicon or silicon nitride and a chemically selective layer.23,26,3032 These two layers have a difference in thermal expansion, which causes the cantilever to oscillate at a specific fundamental frequency, calculated using Newton's second equation of motion30:

fi=λi22πL2EIρA

The mass of the chemically selective layer changes upon analyte adsorption, which in turn causes a change in oscillation frequency, as shown in Figure 3. Though mass sensors have been proven to be successful, there are many limitations with this device setup. The sensors are prone to inaccuracy due to subtle changes in coating amount on the surface, humidity, temperature, etc., and thus need to be calibrated very often. The setup, preparation, and calibration process is delicate and time consuming.21,24,25,28,29

Fig. 3.

Fig. 3

Frequency shift (drift removed) for a microcantilever resonating at 16.9 kHz with a sensitive coating thickness of 21 μm, exposed to ethanol at different concentrations26.

3.2. Electrical Sensors

Electrically sensing e-noses use an electronic circuit connected to a network of sensory materials that provide an electrical response upon binding with a specific known analyte. In addition to the MOS and MOSFET electrical sensors, a popular and more recent sensory material is conductive polymer. There are two types of conductive polymer sensors: intrinsically conductive and composite conductive polymers. To create intrinsically conductive sensors, thin polymer films are electropolymerized across a narrow electrode gap,21,34 and their chemical composition tuned to attract a desired analyte.34 Composite conductive polymers are produced when an insulating polymer is doped with a conductive material, such as carbon black.38 The polymer sensors are then placed in an electrical circuit, and act as resistors, reflecting a decrease in their conductance upon binding with the analytes in the vapour, as shown in Figure 4.21,33,34 This decrease in conductance is most likely due to polymer swelling upon analyte binding–as the polymer swells, gaps between polymer chains increase, lowering conductivity.83 There are a wider variety of conductive polymers available for analyte binding compared to metal-oxides, and thus a larger catalogue of unique sensors for vapour analysis can be created.21 Additionally, the incorporation of several uniquely functionalized polymer sensors in a single circuit allows for the simultaneous detection of multiple vapour components. Data from these sensors can then be arranged in multiple matrices, which in turn creates a unique pattern for each type of vapour analysed.34

Fig. 4.

Fig. 4

(A) Schematic of a sensor array showing an enlargement of one of the modified ceramic capacitors used as sensing elements. The response patterns generated by the sensor array are displayed for acetone (B), benzene (C), and ethanol (D).34

Conductive polymer sensors have recently taken advantage of advances in microchip technology to decrease the size of circuits and increase the number of sensors. Beccherelli, et. al. have designed a large chemical sensor based on the olfactory system that has incorporated up to 65,536 resistive sensing elements. Using a modular approach, each module is capable of reading a 64 × 64 sensor array, and 16 modules work in parallel to analyze and report all data gathered. The chip utilizes an inkjet printer system to deposit conductive polymers on the chips with high accuracy and low cost; furthermore, this deposition technique allows for faster diffusion of vapour elements into and out of the film, affording faster response and recovery time for each sensor.33

Electrical sensors for electronic noses have also progressed to incorporate new materials as a replacement for conductive polymers. Single-walled carbon nanotubes (SWNTs) have recently been used as a type of FET, where either change in conductance or capacitance is measured.39 In 2006, Star, et. al. developed a sensor that utilizes single-walled carbon nanotubes (SWNTs) functionalized with metal nanoparticles as ligand-binding sites, as shown in Figure 5. The device created was a nanotube field effect transistor (NTFET). In short, SWNTs were grown on doped silicon wafers via chemical vapour deposition (CVD). Using photolithography, SWNTs were overlaid with electrical leads made of 30 nm thick Ti films capped with 120 nm thick Au layer. Each 2.54 × 2.54 cm2 wafer can hold about 1000 dyes, and each dye consists of randomly assorted SWNTs patterned into several devices in between two parallel electrodes. The NTFET devices were then decorated with a metal/metal oxide via electron beam (e-beam) deposition. E-beam deposition methods allowed for each NTFET device on a single chip to be paired with a different catalytic metal, if desired. The devices were then exposed to various combustible gases, H2, CO, and H2S and two toxic gases, NH3 and NO2. Response to these gases was recorded as a change in conductance versus gate voltage, shown in Figure 6. Each gas exposure resulted in a specific curve, showing the ability of the device to distinguish between gases.39 SWNTs can also be coated with various polymers to tune their sensitivities.36,40

Fig. 5.

Fig. 5

(A) Conceptual illustration of a carbon nanotube network connecting source (S) and drain (D) electrodes of a FET. SWNTs are decorated with metal nanoparticles (silver dots) for selective detection of analyte gases (red dots). (B) Electronic measurements, such as source-drain conductance (GSD), as a function of gate voltage (VG) before (bare) and after thermal evaporation of discontinuous layer of gold (Au evap).39

Fig. 6.

Fig. 6

Response of bare NTFET device to (A) NH3 and (B) NO2 gases; response of Pd-decorated NTFET device to (C) H2 and (D) H2S gas. Left: A single gas exposure consisted of (i-ii) a five-minute settling period, (ii-iii) five-minute exposure to gas followed by (iii-iv) a ten-minute recovery period. NTFET conductance (G at VG) 0, blue curve) is derived from the G-VG curves measured during the test. Right: G-VG curves recorded before and after gas introduction.39

3.3. Optical Sensors

Optical sensors in e-nose systems have shown much promise to provide a facile, cost-effective, and accurate way of identifying analytes in vapour. Optical sensors work by displaying a shift in emission or absorption of different types of radiations upon binding with a desired analyte. There are two popular means of detection: fluorescent sensors4152,55,57,58 and colorimetric sensors.53,54 Optical sensors offer significant benefits compared with the sensor systems mentioned above; they can provide multiple complex data types simultaneously, including changes in intensity, fluorescence lifetime, wavelength, and spectral shape.50 This increases the ratio of recognizable analytes to number of sensors used.

Fibre optic chemical sensors mimic both the temporal and spatial signalling pathways of the mammalian olfactory system. Walt et al. have developed an e-nose using these fibre optic bundles. Briefly, fibres are coated with a single solvatochromatic dye (Nile Red), which has been mixed with a different polymer species for each fibre. These polymers differ in pore size, baseline polarity, hydrophobicity, flexibility, and swelling tendency, which leads to a unique interaction of each fibre with the analytes presented. Nile Red shows large shifts in emission wavelength maximum with changes in local polarity; the unique analyte-dye/polymer interactions causes a specific local polarity change, resulting in different fluorescent responses for each sensor. These responses have both spatial and temporal components, making it possible to track intensity and spectral changes over time.50,52,5658

Walt et al. have also implemented fluorescent microsphere arrays in their e-nose devices. Similar to fibre optic arrays, microspheres with different chemical properties can be synthesized from various polymers. These microspheres are then dyed with Nile Red and arranged on glass coverslips.47,55 Microsphere arrays provide an advantage over other multi-sensor systems because billions of beads that produce an identical response can be made simultaneously, in comparison to many sensors for which the fabrication process is tedious.47,55 Imaging of these arrays produces a 3D video which tracks fluorescence, time, and spatial orientation of the sensors for decoding and pattern recognition.47 Each stock of unique sensors can be used in multiple arrays with high fidelity, creating a unique “fingerprint” for each analyte in vapour.55

In collaboration with other labs, Walt et. al. have also combined the fibre optic bundles with microspheres.41,4446,49 The fibre ends were etched to create microwells in which microspheres were placed. The microwells were etched into the distal end of the fibre by submerging the polished end into a buffered hydrofluoric acid solution for 70s; the fibre was then transferred to a beaker of DI water to stop the reaction, run under tap water, and sonicated to remove excess salts that may have formed during etching.49 The microspheres fabricated and suspended in droplets of solution were placed on a clamped imaging fibre, as shown in Figure 7.45,49 The beads then fill the microwells (one bead per well), and packing density (number of filled wells) is directly proportional to the bead solution concentration and the number of drops applied.49 Beads may also be loaded into microwells by a dry deposition method—creating a slurry of beads and rubbing them into the surface via a latex glove44,45–or a block and load method, which uses a razor blade to control spatial separation and orientation of different types of beads.41 Each type of bead has a distinct, intrinsic response to the vapours presented, which eliminates the need for additional encoding for bead identification, as shown in Figure 8. The fibre/microsphere composite bundles have also been incorporated in portable devices for applications in the field as well. These devices consist of batteries for a CPU laptop and the electronic nose, an imaging and processing system, and the etched fibres that are connected to an optical block and vapour-sensing unit. 41,46

Fig. 7.

Fig. 7

Schematic depiction of the self-encoded bead array (SEBA) concept. A mixture of sensor beads is prepared by combining aliquots from three stock solutions each containing a different type of polymer/dye sensor suspended in a Nanopure water/0.01% Tween solution. A drop of the final mixture is then placed onto the distal tip of an etched imaging fibre.49

Fig. 8.

Fig. 8

CCD images of each imaging fibre bundle. Bundle A-C (top) and bundle D-F (bottom) showing the fitted ROIs corresponding to each addressable bead sensor. A total of 2683 sensors were analyzed: (A) 814, (B) 286, (C) 301, (D) 692, (E) 294, and (F) 296 beads. The ROIs were automatically generated using Labview and IMAQ Vision software. Each imaging fibre's face is 1 mm2. Images were acquired with 100-ms exposure time, and excitation/emission filters were 560 (bp 40) and 630 (bp 20) nm, respectively.45

Suslick et al. has also employed optical sensors for their electronic nose device. Their approach uses immobilized dyes that have chemoselective reactions when exposed to a variety of different ligands in vapour, shown in Figure 9.53,54 Metalloporphyrines were used because they display both open coordination sites for axial ligation, large spectral shifts upon ligand binding, and intense coloration.53 In addition, the array can be expanded to include bis-pocketed Zn porphyrines to detect for the size and shape of the analyte, acid/base indicators to differentiate Brønsted basicity, and solvatochromic dyes to report on the polarity of the analyte.54 The use of these additional sensors made it possible to create a 36 sensor array, capable of distinguishing between similar classes of amines at very small concentrations (<1ppmv).54 Briefly, dies were deposited on a hydrophobic, benign surface such as reverse-phase silica gel, and the device placed on an ordinary flatbed scanner. Images were taken before and after exposure at 1200 dots per inch in RGB color mode. Color difference maps were then produced by subtracting the “before” image from the “after” image; a 314 pixel average was used in this process, taken from the center of each dye spot. Each analyte is represented as a multi-dimensional vector, i.e. 72 in a 24-sensor array, one each for the red green and blue differences of 24 dyes. Image processing and spectral analysis create a unique color map for each analyte, and while similar analyte types show a correlation in clusters present, it is still possible to identify individual analytes by their color maps.54

Fig. 9.

Fig. 9

Miniaturized array. The metalloporphyrines were dissolved in a polymer film (dibutylphthalate in polystyrene) and deposited on Teflon (A, showing metals only for clarity). The array before vapour exposure (B) and colour change profile after exposure to n-butylamine (C, 1 min, 9.3% in N2).53

4. Electronic Tongues

Along with the development of electronic noses, liquid sensors have been produced that work in a similar way. The main methods are electrochemical, mass, and optical electronic tongues.

4.2. Mass Sensors

Piezoelectric sensors, similar to those used for vapour analysis, have also been used in liquids. Sensors have been fabricated both from quartz84,85 and thin films,5963 and their frequency change can be analyzed using either surface acoustic wave (SAW)84,85 or bulk acoustic wave (BAW)5963 resonance. Recently, thin film BAW sensors have taken precedence over the common quartz crystal microbalance, as thin films are much smaller, can reach higher frequencies, and are more sensitive because frequency is directly related to film thickness.60 Resonance-active films are functionalized to bind to a desired analyte, and attached to the base of the fluid flow chamber. Most of the thin film resonators are operated in shear mode, with resonance-active films sitting in a plane orthogonal to the fluid flow direction.60,61,63,85 In 2008, Wingqvist et. al. showed that thin film BAW sensors have enhanced sensitivity upon the addition of a layer that has low acoustic impedance.62

4.2. Electrochemical

The earliest, and most popular type of electrochemical tongues used potentiometric sensors. Potentiometric sensors consist of ion selective electrodes, and detection of desired analyte is accomplished by tracking changes in the potentiometric electrode versus a control electrode in zero current conditions. The ion activity in the solution is functionally related to the potential of the sensor. Each electrode provides a unique response upon exposure to a specific analyte, creating “taste” profiles for each substance identified.

The first commercialized electronic tongue was created by Toko in 1996. His goal was to fabricate electrodes that closely mimicked human taste buds to detect the basic five tastes: bitter, sour, sweet, salty, and umami. Each of 8 electrodes was outfitted with a unique biomimetic lipid bilayer species, each of which reacted differently upon exposure to the same “flavour” profile—creating a unique fingerprint that was readily identifiable.72 Since Toko's device, other potentiometric sensors have been created to identify ions such as ammonium,64,65 potassium,65 and nitrite,66 and to analyze components of some beverages such as tea68 and wine.69 Some limitations of the potentiometric sensors include a narrow range of detectable ions, electronic noise, and adsorption of ions to the electrodes, causing interference.74,86

A second type of electrochemical tongue uses voltammetric sensors—which records the relationship of potential and current over series of pulses. For voltammetric sensors, there are three electrodes: working, auxiliary, and reference. The reference electrode remains at a constant potential, while current is generated between the working and auxiliary electrodes. The earliest voltammetric tongue was created by Winquist, et. al. in 1997 to analyze the aging process of milk and orange juice over time.74 The setup involved 2 working electrodes, gold and platinum, and a potentiostat enabled the system to change between electrodes for detection of different compounds as shown in Figure 10. Two types of pulse voltammetry were used: large amplitude pulse voltammetry (LAPV), and small amplitude pulse voltammetry (SAPV). Both voltammetry methods induce a series of potential pulses to the electrodes, and track the resultant current response over time. Each electroactive compound in the solution responds differently to the pulses, creating a unique “map” of current response for each compound, shown in Figure 11.

Fig. 10.

Fig. 10

Schematic diagram of the voltammetric electronic tongue setup.74

Fig. 11.

Fig. 11

Recordings from experiments carried out on various samples using the application measurement sequence. In this, LAPV followed by SAPV is carried out at the gold working electrode, where after the sequence is repeated at the platinum working electrode. The samples investigated were: (A) Phosphate buffer; (B) Citrate buffer; (C) Milk.74

One disadvantage of voltammetric sensors is that there are thousands of data points to be processed from the I/V curves generated. This demands the necessity of data processing techniques that can handle and make sense of this large amount of information.67,71,74 Winquist et. al. used principle component analysis (PCA) to process the obtained data; more recently, others have used both multifrequency LAPV71 and data compression methods67 to more easily and quickly distinguish between overlapping frequencies created by solutions with multiple components.

4.3. Optical Sensors

In addition to vapour analysis with optical sensors, Walt et. al. and Suslick et. al. have also contributed to the development of optical sensors for electronic tongue devices. The principles in vapour sensing are much the same as in liquid sensing. In 2005, Walt et. al. incorporated the optical fibres with microbead arrays in a microfluidic chamber to detect DNA via fluorescence of the sensors, shown in Figure 12.75 Using colorimetric methods, Suslick et. al. developed an electronic tongue device that holds chemisensitive dyes in a sol-gel as solution are run over them. This system has been used successfully for a variety of liquid analyses, including soft drinks78 and protein identification.76 Still another group has used spectrophotometric analysis to detect inorganic bromide in water using bromide reactive dyes.77

Fig. 12. Schematic diagram showing the microfluidic T-junction layout (A) and the optical imaging fibre bundle showing DNA hybridization in the microfluidic manifold (B)75.

Fig. 12

5. Technological Advances and Applications: Barcoded Resin Technology

5.1. Sensor Technology: Spectroscopic Barcoding

Electronic noses and tongues can be more finely tuned and broadly selective using new materials such as barcoded resins (BCRs) and principles such as surface enhanced Raman scattering (SERS). Each BCR has a unique Raman and infrared spectra, arising from its composition, which is transformed into a barcode for identification. Example barcodes are shown in Figure 13. Upon interaction with an analyte, the vibrational signatures of the polymer array change, resulting in slight but detectable spectral variations for each BCR. The collective response (or disease-specific patterns) can then be quantified using multivariate data analysis. The spectral changes reflected in each BCR—ca. 700 BCRs available thus far87—can then be used as a data point for pattern recognition, increasing the accuracy of symptom specific and disease-specific response for the e-nose device.8793

Fig. 13.

Fig. 13

Raman (left) and infrared (right) spectroscopic barcodes of BCRs.91

To synthesize the BCRs, sub-library copolymers is created via various combinations of the five main alkylated styrene monomers. A six digit binary code is assigned to each BCR, one digit for each monomer. For example, BCR12's code is 001010, indicating that it is made up of equal parts of monomers 001000 and 000010. The sub-library of copolymers can then be resynthesized to include one of the fluorinated monomers (A-I) to produce hundreds of different copolymers. The library of currently available monomers is shown in Figure 14. Each copolymer can then be scanned via Raman spectroscopy or Fourier Transform Infrared (FTIR) spectroscopy and a spectrum generated that is unique to each composition. Following preprocessing steps such as baseline subtraction and normalization, spectra can then be converted into barcodes; each wavelength maximum peak corresponds to a line in the barcode. The barcode identification method has made it possible to distinguish between as many as ca. 700 unique copolymers, offering a significant selectivity improvement to those currently used.

Fig. 14.

Fig. 14

Styrene monomers used to generate a library of 630 spectroscopically encoded polymers.87

5.2. BCR Sensor Array Fabrication

BCR sensor arrays can be fabricated either using bulk copolymers or beaded copolymers. Both are deposited into etched, patterned silicon wells. Both methods preserve spectroscopic barcodes of the BCRs, and provide a facile, reproducible and reusable array that can be implemented in an e-nose device.87

To create the etched silicon substrates, a combination of lithography and anisotropic dry reactive ion etching (RIE) was used on silicon substrates for micro-well formation as follows: the silicon substrates were piranha cleaned, HMDS primed and spin-coated with AZ P46 20 photoresists. Areas of the silicon were exposed by irradiating the film (l = 400 nm) through a lithography mask built in-house. RIE using a combination of SF6/C2F4, at pressures of 125 and 75 sccm, respectively, was used to etch the exposed silicon of the wafer. In this way, 10, 100, and 200 mm-deep wells were obtained SEM images of patterned wafers are shown in Figure 15A. The patterned wafer was then coated with a 200 mm thick copper film by RF sputtering. For deeper wells, a combination 504 photoresist SiO2 hard mask was used. In this approach, 4 μm oxide films were deposited by plasma enhanced chemical vapour deposition (PECVD) and the resulting oxide patterned with conventional 504 photoresist. The exposed oxide areas were etched with buffer oxide etch (BOE). The same RIE conditions were used to etch the oxide.94,95

Fig. 15.

Fig. 15

(A, left to right) RIE etching and SEM images of micropatterned silicon chips94. (B, left to right) beads deposited in microwells and close-up view of bead in well.94 (C) Setup for scanning sensor arrays after vapour exposure.

For sensor arrays that used bulk copolymers, each of the polymer resins was dissolved in chloroform (ca. 10 % W/W). One drop of the homogeneous solution was dispersed into a well on the patterned substrate using a capillary tube. The procedure was repeated for each polymer in the library producing an 8×8 array. The coated substrate was heated to 50°C for ca. 2 h on a hot plate to evaporate residual chloroform from the polymer films. The polymer samples in each well were identified by a number (representing a row) and a letter (representing a column).95

For the sensor arrays that used beaded copolymers, beads were deposited onto the micro-patterned silicon substrates by dusting an amount of beads on the surface of the silicon chip. A rubberized spatula was used to spread the powder in a circular motion over the chip to ensure the filling of the holes. SEM of the BCR arrays are shown in Figure 15B, it clearly shows that by selecting the right bead/well size combination, an ideal ensemble could be achieved. Excess beads were removed by wiping with a kimwipe or under a stream of nitrogen.94

5.3. Application Using BCRs as an E-nose

The BCRs can be used for both vapour and liquid sensing without loss of their identifying barcodes; this offers facile array fabrication because spatial orientation of each copolymer in an array does not need to be controlled. Though the copolymer types can be randomly mixed throughout an array, ordered arrangement of the polymer beads into rows and columns is desirable for fast spectroscopic imaging.

As an analyte-containing vapour is passed over the sensor array (Figure 15C), analytes interact with the polymers in a variety of ways, including being absorbed into the polymer network, polymer swelling, or adsorbing to the polymer surface. This, in turn, changes the chemical make-up of the polymer network, which then causes slight changes in the FTIR and Raman spectra. The changes are very subtle, such that the microbead is still identifiable by its barcode after a partial least square regression (PLS). Spectra for each of the polymers in the microwell are taken before and after analyte exposure, and the resultant spectral change is represented by an angle, θ calculated by PLS. Angle maps and histogram graphs representing the resultant changes in the composition matrix provide a response “fingerprint” for each of the copolymers in the sensor array. Each polymer reacts uniquely to each analyte, and it is therefore possible to collect an individual response pattern for analyte identification in vapours for which the analyte is not known. Because these polymers are non-selective in vapours containing multiple analytes (such as exhaled breath) the global response pattern of the sensor must be considered when comparing trends in sample composition.

The polymer sensor array was put inside a gas flow cell and screened by FT-IR and Raman spectroscopy prior to analyte vapour exposure. Saturated analyte vapour, generated by slowly bubbling nitrogen through a vial containing the liquid of interest, was then passed through the gas flow cell containing the sensor arrays as shown in Figure 15C.95 After analyte vapour exposure, the sensing system was allowed to reach equilibrium before the spectrum or image acquisition; equilibrium was verified to be the absence of changes in measured spectra. Prior to the acquisition of the FTIR spectra, several pre-processing steps were taken. To account for the atmospheric or analyte vapour (in air) effects, a background measured for each trial was subtracted from the response. Histograms and angle response maps are shown in Figure 16.

Fig. 16.

Fig. 16

Raman (upper left) and FTIR (upper right) angle maps (before versus after analyte vapour exposure) and histograms of the individual Raman (lower left) and FTIR (lower right) responses (angle value) of 64 BCRs (sensors) to hexane (analyte).95 The values in the histogram correspond to the values along the diagonal in the angle map.

BCRs have also been used as biosensors in liquids for both antibody-antigen binding96 and pathogen detection.97 Both of these methods use a fluorescent sandwich immunoassay antibody detection method: the microspheres were conjugated with an antibody, and the corresponding antigen was captured between the microsphere antibody conjugate and a second fluorescently labelled antibody. Upon binding, the beads became fluorescent, and identification of the bound antibody was possible by cross-referencing the Raman barcode of the microsphere (Figure 17).96,97

Fig. 17.

Fig. 17

(A) Electronic absorption and emission spectra of Cy3 (upper spectra) and Cy2 (lower spectra) dyes. Excitation laser lines at 532 nm (green arrow) and at 780 nm (red arrow) are shown. (B) From top to bottom are the electronic emission spectra of Ab1 and Ag1, the Raman spectra of BCR2 before and after conjugation with Ab1 (75 μg), and exposure to Ag1 (15 μg/mL). (C) From top to bottom are the Raman spectra acquired with the 780 nm laser line of BCR2 before and after conjugation with Ab1, and exposure to Ag1. (D) Optical image of a mixture of BCR1, BCR3, BCR4, and Ab1-BCR2 after exposure to Ag1. (E) Raman mapping at 500 cm-1 (λex)532 nm) showing the fluorescent Ab1-BCR2 conjugates. (F) Raman classification of all the BCRs (λex) 780 nm).96

5.4. SERS Active Particles for Signal Enhancement

There is also much potential for the incorporation of SERS active particles for use in electronic nose and tongue systems. The use of SERS greatly enhances the optical properties of the original material, which in turn increases accuracy of analyte detection.98 Fenniri, et. al. have explored different nanoparticles, such as gold octohedra,99 gold and silver bimetallic particles on a silicon substrate,100 and silver fractals101 to induce SERS for molecule identification. These SERS active materials have also been combined with polymer microspheres, including silver nanoparticle cross-linked polymer microspheres102 and gold polystyrene/polyethylene glycol microbeads, shown in Figure 18.103 The use of SERS active particles and/or barcoded resins in electronic nose and tongue systems can greatly enhance selectivity and sensitivity—two parameters that are essential for accurate and early medical diagnosis.

Fig. 18.

Fig. 18

SEM images of (A) Au nanoparticle/bead and (B) close-up showing Au nanostructures attached to the surface. (C) Cross-sectional TEM with XRD pattern of the bead crust.102

Conclusions

As advances in electronics, image processing, and material design continue, the electronic nose and tongue devices have adapted to become smaller, more selective, and more sensitive, with fast data processing time and facile readout of vapour and liquid analysis.104107 Each of the sensor types have various advantages and disadvantages; thus, it is important to consider the ultimate aim of analysis and type of sample to be characterized. Both mass and electrical/electrochemical sensors both provide sensitivity and selectivity in vapour and liquid analysis; however, these devices have a delicate setup, are unfortunately prone to drift due to changes in environment such as temperature and humidity, and require complex data analysis to extract the de-convolute the sensor response signal from the background noise. Optical sensors for provide a cost-effective, facile fabrication method while maintaining high throughput sample analysis; however, they also may require complex data analysis and provide more of a qualitative characterization than quantitative. While these systems have been used in a wide variety of applications from food and drinks, to explosives vapours, some of the most promising work has been done in medical diagnostics. These devices have already been implemented to detect diseases such as lung cancer,1,79,81,82,108111 asthma,112 and other diseases known to produce specific VOCs in exhaled breath.80 Bacteria presence and growth phase identification has also been possible.113,114 In addition to exhaled breath, disease biomarkers have also recently been identified through multiple “omics” studies in saliva,115117 sweat,105,106,118120 and urine;121123As the technology for more expansive and selective sensors continues, devices may further be developed to create unique patient profiles for a variety of diseases via analysis of the various body vapours and fluids.

In the future, electronic noses and tongues may also prove to be a useful tool in unravelling the pathophysiology of mental illnesses. Several studies have shown that anxiety and depression lead to changes in salivary cortisol levels.124129 In addition, alterations in the sympathetic nervous system in response to emotional stress produce noticeable changes in saliva,128 such as secretion of chromagranin A,129131 salivary alpha amylase,130136 and pro-inflammatory cytokines.131,137 Mental illnesses such as Alzheimer's disease and Post Traumatic Stress Disorder can also be signified by changes in skin odour and urine.138,139

Acknowledgments

We acknowledge the support of the National Institutes of Health, the National Research Council of Canada, the university of Alberta and Northeastern University.

Biographies

graphic file with name nihms813253b1.gif

Ms. Jessica Fitzgerald is a Bioengineering Ph.D. candidate in the Supramolecular Nanomaterials Laboratory at Northeastern University, Boston, MA, USA. She received a B.S. in Engineering Physics from Oral Roberts University in Tulsa, Oklahoma, USA. Her research interests include mainly disease diagnosis through exhaled biomarkers. Her specific efforts are directed toward the development of an electronic olfactory device to evaluate abnormal stress response for the diagnosis of anxiety spectrum disorders and Post-Traumatic Stress disorder.

graphic file with name nihms813253b2.gif

Dr. Fenniri received all his degrees from the Université de Strasbourg, France. After postdoctoral training at the Scripps Research Institute, he moved to Purdue University, where he initiated his independent academic career, and established the Purdue Laboratory for Chemical Nanotechnology (1999). In 2003, He joined the National Research Council and the University of Alberta (Edmonton, AB, Canada) as Professor of Chemistry (2003-2013). Dr. Fenniri is currently Professor of Chemical and Biomedical Engineering at Northeastern University, Boston, MA, USA. Dr. Fenniri's contributions appeared in over 220 publications, 20 patents and patent applications, and over 450 contributed national and international conference papers. Dr. Fenniri has also lectured extensively around the globe and he has been an invited professor at several institutes and universities.

Notes and references

  • 1.Walt DR, Stitzel SE, Aernecke MJ. Artificial noses. Am Sci. 2012;100:38–45. doi: 10.1146/annurev-bioeng-071910-124633. [DOI] [PubMed] [Google Scholar]
  • 2.Alberg AJ, Brock MV, Ford JG, Samet JM, Spivack SD. Epidemiology of lung cancer: Diagnosis and management of lung cancer, 3rd ed: American college of chest physicians evidence-based clinical practice guidelines. Chest. 2013;143:1–29. doi: 10.1378/chest.12-2345. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Youlden DR, Cramb SM, Dunn NAM, Muller JM, Pyke CM, Baade PD. The descriptive epidemiology of female breast cancer: An international comparison of screening, incidence, survival and mortality. Cancer Epidemiol. 2012;36:237–248. doi: 10.1016/j.canep.2012.02.007. [DOI] [PubMed] [Google Scholar]
  • 4.Mariotto AB, Robin Yabroff K, Shao Y, Feuer EJ, Brown ML. Projections of the cost of cancer care in the United States: 2010-2020. J Natl Cancer Inst. 2011;103:117–128. doi: 10.1093/jnci/djq495. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Mauldin PD, Salgado CD, Hansen IS, Durup DT, Bosso JA. Attributable hospital cost and length of stay associated with health care-associated infections caused by antibiotic-resistant gram-negative bacteria. Antimicrob Agents Chemother. 2010;54:109–115. doi: 10.1128/AAC.01041-09. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Costerton JW, Stewart PS, Greenberg EP. Bacterial biofilms: a common cause of persistent infections. Science. 1999;284:1318–1322. doi: 10.1126/science.284.5418.1318. [DOI] [PubMed] [Google Scholar]
  • 7.N.I. of M. Health, NIMH. [accessed March 20, 2015];Any Mental Illness (AMI) Among Adults. 2014 http://www.nimh.nih.gov/health/statistics/prevalence/any-mental-illness-ami-among-us-adults.shtml.
  • 8.N.I. of M. Health, NIMH. [accessed March 20, 2015];Mental Healthcare Cost Data for All Americans (2006) 2006 http://www.nimh.nih.gov/health/statistics/cost/mental-healthcare-cost-data-for-all-americans-2006.shtml.
  • 9.Follenweider LM, Lambertino A. Epidemiology of Asthma in the United States. Nurs Clin North Am. 2013;48:1–10. doi: 10.1016/j.cnur.2012.12.008. [DOI] [PubMed] [Google Scholar]
  • 10.Glenza J. [accessed March 20, 2015];Dog trained to detect thyroid cancer “with 88% accuracy” | US news | The Guardian, Guard. 2015 http://www.theguardian.com/us-news/2015/mar/09/dog-detects-thyroid-cancer-research.
  • 11.Buck LB. Unraveling the sense of smell (Nobel Lecture) Angew Chemie - Int Ed. 2005;44:6128–6140. doi: 10.1002/anie.200501120. [DOI] [PubMed] [Google Scholar]
  • 12.Malnic B, Hirono J, Sato T, Buck LB. Combinatorial receptor codes for odors. Cell. 1999;96:713–723. doi: 10.1016/S0092-8674(00)80581-4. [DOI] [PubMed] [Google Scholar]
  • 13.Mombaerts P, Wang F, Dulac C, Chao SK, Nemes A, Mendelsohn M, Edmondson J, Axel R. Visualizing an olfactory sensory map. Cell. 1996;87:675–686. doi: 10.1016/S0092-8674(00)81387-2. [DOI] [PubMed] [Google Scholar]
  • 14.Katada S, Hirokawa T, Oka Y, Suwa M, Touhara K. Structural basis for a broad but selective ligand spectrum of a mouse olfactory receptor: mapping the odorant-binding site. J Neurosci. 2005;25:1806–1815. doi: 10.1523/JNEUROSCI.4723-04.2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Gardner JW, Bartlett PN. A brief history of electronic noses. Sensors Actuators B Chem. 1994;18:210–211. doi: 10.1016/0925-4005(94)87085-3. [DOI] [Google Scholar]
  • 16.Pearce TC. Computational parallels between the biological olfactory pathway and its analogue “the electronic nose”: Part I. Biological olfaction. BioSystems. 1997;41:43–67. doi: 10.1016/S0303-2647(96)01661-9. [DOI] [PubMed] [Google Scholar]
  • 17.Inoue T, Ohtsuka K, Yoshida Y, Matsuura Y, Kajiyama Y. Metal oxide semiconductor NO2 sensor. Sensors Actuators B Chem. 1995;25:388–391. doi: 10.1016/0925-4005(95)85087-2. [DOI] [Google Scholar]
  • 18.Meixner H, Lampe U. Metal Oxide Sensors. Sensors Actuators B Chem. 1996;33 doi: 10.1109/SENSOR.1995.717380. [DOI] [Google Scholar]
  • 19.Cheng S, Lin P, Wang W, Chen J. Highly sensitive MOSFET gas sensors with porous platinum gate electrode. Electron Lett. 1997;33:535–536. [Google Scholar]
  • 20.Eklo T, Lundstro I. Gas mixture analysis using a distributed chemical sensor system. Sensors Actuators B. 1999;57:274–282. doi: 10.1016/S0925-4005(99)00145-8. [DOI] [Google Scholar]
  • 21.Dickinson TA, White J, Kauer JS, Walt DR. Current trends in “artificial-nose” technology. Trends Biotechnol. 1998;16:250–258. doi: 10.1016/S0167-7799(98)01185-8. [DOI] [PubMed] [Google Scholar]
  • 22.Stitzel SE, Stein DR, Walt DR. Enhancing vapour sensor discrimination by mimicking a canine nasal cavity flow environment. J Am Chem Soc. 2003;125:3684–3685. doi: 10.1021/ja028239y. [DOI] [PubMed] [Google Scholar]
  • 23.Adams JD, Rogers B, Manning L, Hu Z, Thundat T, Cavazos H, Minne SC. Piezoelectric self-sensing of adsorption-induced microcantilever bending. Sensors Actuators, A Phys. 2005;121:457–461. doi: 10.1016/j.sna.2005.03.058. [DOI] [Google Scholar]
  • 24.Filippov AP, Strizhak PE, Il'in VG. Quartz crystal microbalance modified with Cu(II) stearate and octadecylamine co-ordination chemical compounds for detection of volatile organic compounds. Sensors Actuators, B Chem. 2007;126:375–381. doi: 10.1016/j.snb.2007.03.020. [DOI] [Google Scholar]
  • 25.Ko HJ, Park TH. Piezoelectric olfactory biosensor: Ligand specificity and dose-dependence of an olfactory receptor expressed in a heterologous cell system. Biosens Bioelectron. 2005;20:1327–1332. doi: 10.1016/j.bios.2004.05.002. [DOI] [PubMed] [Google Scholar]
  • 26.Lochon F, Fadel L, Dufour I, Rebière D, Pistré J. Silicon made resonant microcantilever: Dependence of the chemical sensing performances on the sensitive coating thickness. Mater Sci Eng C. 2006;26:348–353. doi: 10.1016/j.msec.2005.10.079. [DOI] [Google Scholar]
  • 27.Shiokawa S, Kondoh J. Surface Acoustic Wave Sensors. Jpn J Appl Phys. 2004;43:2799–2802. doi: 10.1143/JJAP.43.2799. [DOI] [Google Scholar]
  • 28.Sung JH, Ko HJ, Park TH. Piezoelectric biosensor using olfactory receptor protein expressed in Escherichia coli. Biosens Bioelectron. 2006;21:1981–1986. doi: 10.1016/j.bios.2005.10.002. [DOI] [PubMed] [Google Scholar]
  • 29.Wu TZ. A piezoelectric biosensor as an olfactory receptor for odor detection: electronic nose. Biosens Bioelectron. 1999;14:9–18. doi: 10.1016/s0956-5663(98)00086-4. [DOI] [PubMed] [Google Scholar]
  • 30.Zhou J, Li P, Zhang S, Huang Y, Yang P, Bao M, Ruan G. Self-excited piezoelectric microcantilever for gas detection. Microelectron Eng. 2003;69:37–46. doi: 10.1016/S0167-9317(03)00227-2. [DOI] [Google Scholar]
  • 31.Zhou J, Li P, Zhang S, Long Y, Zhou F, Huang Y, Yang P, Bao M. Zeolite-modified microcantilever gas sensor for indoor air quality control. Sensors Actuators, B Chem. 2003;94:337–342. doi: 10.1016/S0925-4005(03)00369-1. [DOI] [Google Scholar]
  • 32.Zhou W, Khaliq A, Tang Y, Ji H, Selmic RR. Simulation and design of piezoelectric microcantilever chemical sensors. Sensors Actuators, A Phys. 2005;125:69–75. doi: 10.1016/j.sna.2005.07.009. [DOI] [Google Scholar]
  • 33.Beccherelli R, Zampetti E, Pantalei S, Bernabei M, Persaud KC. Design of a very large chemical sensor system for mimicking biological olfaction. Sensors Actuators, B Chem. 2010;146:446–452. doi: 10.1016/j.snb.2009.11.031. [DOI] [Google Scholar]
  • 34.Freund MS, Lewis NS. A chemically diverse conducting polymer-based “electronic nose”. Proc Natl Acad Sci U S A. 1995;92:2652–2656. doi: 10.1073/pnas.92.7.2652. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Gardner JW, Shin HW, Hines EL. Electronic nose system to diagnose illness. Sensors Actuators, B Chem. 2000;70:19–24. doi: 10.1016/S0925-4005(00)00548-7. [DOI] [Google Scholar]
  • 36.Qi P, Vermesh O, Grecu M, Javey A, Wang Q, Dai H, Peng S, Cho KJ. Toward large arrays of multiplex functionalized carbon nanotube sensors for highly sensitive and selective molecular detection. Nano Lett. 2003;3:347–351. doi: 10.1021/nl034010k. [DOI] [PubMed] [Google Scholar]
  • 37.Shevade AV, Ryan MA, Homer ML, Manfreda AM, Zhou H, Manatt K. Molecular modeling of polymer composite-analyte interactions in electronic nose sensors. Sensors Actuators, B Chem. 2003;93:84–91. doi: 10.1016/S0925-4005(03)00245-4. [DOI] [PubMed] [Google Scholar]
  • 38.Sisk BC, Lewis NS. Vapour sensing using polymer/carbon black composites in the percolative conduction regime. Langmuir. 2006;22:7928–7935. doi: 10.1021/la053287s. [DOI] [PubMed] [Google Scholar]
  • 39.Star A, Joshi V, Skarupo S, Thomas D, Gabriel JCP. Gas sensor array based on metal-decorated carbon nanotubes. J Phys Chem B. 2006;110:21014–21020. doi: 10.1021/jp064371z. [DOI] [PubMed] [Google Scholar]
  • 40.Wei C, Dai L, Roy A, Tolle TB. Multifunctional chemical vapour sensors of aligned carbon nanotube and polymer composites. J Am Chem Soc. 2006;128:1412–1413. doi: 10.1021/ja0570335. [DOI] [PubMed] [Google Scholar]
  • 41.Aernecke MJ, Guo J, Sonkusale S, Walt DR. Design, implementation, and field testing of a portable fluorescence-based vapour sensor. Anal Chem. 2009;81:5281–5290. doi: 10.1021/ac900505p. [DOI] [PubMed] [Google Scholar]
  • 42.Aernecke MJ, Walt DR. Optical-fibre arrays for vapour sensing. Sensors Actuators, B Chem. 2009;142:464–469. doi: 10.1016/j.snb.2009.06.054. [DOI] [Google Scholar]
  • 43.Aernecke MJ, Walt DR. Temporally Resolved Fluorescence Spectroscopy of a Microarray-Based Vapour Sensing System. 2009;81:6155–6162. doi: 10.1021/ac900589b. [DOI] [PubMed] [Google Scholar]
  • 44.Albert KJ, Walt DR. High-speed fluorescence detection of explosives-like vapours. Anal Chem. 2000;72:1947–1955. doi: 10.1021/ac991397w. [DOI] [PubMed] [Google Scholar]
  • 45.Albert KJ, Walt DR, Gill DS, Pearce TC. Optical multibead arrays for simple and complex odor discrimination. Anal Chem. 2001;73:2501–2508. doi: 10.1021/ac001137a. [DOI] [PubMed] [Google Scholar]
  • 46.Albert KJ, Myrick ML, Brown SB, James DL, Milanovich FP, Walt DR. Field-deployable sniffer for 2,4-dinitrotoluene detection. Environ Sci Technol. 2001;35:3193–3200. doi: 10.1021/es010829t. [DOI] [PubMed] [Google Scholar]
  • 47.Albert KJ, Walt DR. Information coding in artificial olfaction multisensor arrays. Anal Chem. 2003;75:4161–4167. doi: 10.1021/ac0264776. [DOI] [PubMed] [Google Scholar]
  • 48.Bencic-Nagale S, Walt DR. Extending the longevity of fluorescence-based sensor arrays using adaptive exposure. Anal Chem. 2005;77:6155–6162. doi: 10.1021/ac0505021. [DOI] [PubMed] [Google Scholar]
  • 49.Dickinson TA, Michael KL, Kauer JS, Walt DR. Convergent, self-encoded bead sensor arrays in the design of an artificial nose. Anal Chem. 1999;71:2192–2198. doi: 10.1021/ac981457i. [DOI] [PubMed] [Google Scholar]
  • 50.Dickinson TA, White J, Kauer JS, Walt DR. A chemical-detecting system based on a cross-reactive optical sensor array. Nature. 1996;382:697–700. doi: 10.1038/382697a0. [DOI] [PubMed] [Google Scholar]
  • 51.Epstein JR, Walt DR. Fluorescence-based fibre optic arrays: a universal platform for sensing. Chem Soc Rev. 2003;32:203–214. doi: 10.1039/b300617d. [DOI] [PubMed] [Google Scholar]
  • 52.Johnson SR, Sutter JM, Engelhardt HL, Jurs PC, White J, Kauer JS, Dickinson TA, Walt DR. Identification of Multiple Analytes Using an Optical Sensor Array and Pattern Recognition Neural Networks. Differences. 1997;69:4641–4648. doi: 10.1021/ac970298k. [DOI] [Google Scholar]
  • 53.Rakow NA, Suslick KS. A colorimetric sensor array for odour visualization. Nature. 2000;406:710–713. doi: 10.1038/35021028. [DOI] [PubMed] [Google Scholar]
  • 54.Rakow NA, Sen A, Janzen MC, Ponder JB, Suslick KS. Molecular recognition and discrimination of amines with a colorimetric array. Angew Chemie - Int Ed. 2005;44:4528–4532. doi: 10.1002/anie.200500939. [DOI] [PubMed] [Google Scholar]
  • 55.Stitzel SE, Cowen LJ, Albert KJ, Walt DR. Array-to-array transfer of an artificial nose classifier. Anal Chem. 2001;73:5266–5271. doi: 10.1021/ac010111w. [DOI] [PubMed] [Google Scholar]
  • 56.Walt DR, Dickinson T, White J, Kauer J, Johnson S, Engelhardt H, Sutter J, Jurs P. Optical sensor arrays for odor recognition. Biosens Bioelectron. 1998;13:697–699. doi: 10.1016/S0956-5663(98)00024-4. [DOI] [PubMed] [Google Scholar]
  • 57.White J, Dickinson TA, Walt DR, Kauer JS. An olfactory neuronal network for vapour recognition in an artificial nose. Biol Cybern. 1998;78:245–251. doi: 10.1007/s004220050430. [DOI] [PubMed] [Google Scholar]
  • 58.White J, Kauer JS, Dickinson TA, Walt DR. Rapid analyte recognition in a device based on optical sensors and the olfactory system. Anal Chem. 1996;68:2191–2202. doi: 10.1021/ac9511197. [DOI] [PubMed] [Google Scholar]
  • 59.Gallardo J, Alegret S, de Roman MA, Munoz R, Hernandez PR, Leija L, del Valle M. Determination of ammonium ion employing an electronic tongue based on potentiometric sensors. Anal Lett. 2003;36:2893–2908. doi: 10.1081/AL-120026410. [DOI] [PubMed] [Google Scholar]
  • 60.Gallardo J, Alegret S, Muñoz R, De-Román M, Leija L, Hernández PR, Del Valle M. An electronic tongue using potentiometric all-solid-state PVC-membrane sensors for the simultaneous quantification of ammonium and potassium ions in water. Anal Bioanal Chem. 2003;377:248–256. doi: 10.1007/s00216-003-2042-7. [DOI] [PubMed] [Google Scholar]
  • 61.Gallardo J, Alegert S, Del Valle M. A flow-injection electronic tongue based on potentiometric sensors for the determination of nitrate in the presence of chloride. Sensors Actuators, B Chem. 2004;101:72–80. doi: 10.1016/j.snb.2004.02.027. [DOI] [Google Scholar]
  • 62.Holmin S, Spångeus P, Krantz-Rülcker C, Winquist F. Compression of electronic tongue data based on voltammetry - A comparative study. Sensors Actuators, B Chem. 2001;76:455–464. doi: 10.1016/S0925-4005(01)00585-8. [DOI] [Google Scholar]
  • 63.Lvova L, Legin A, Vlasov Y, Cha GS, Nam H. Multicomponent analysis of Korean green tea by means of disposable all-solid-state potentiometric electronic tongue microsystem. Sensors Actuators, B Chem. 2003;95:391–399. doi: 10.1016/S0925-4005(03)00445-3. [DOI] [Google Scholar]
  • 64.Simões Da Costa AM, Delgadillo I, Rudnitskaya A. Detection of copper, lead, cadmium and iron in wine using electronic tongue sensor system. Talanta. 2014;129:63–71. doi: 10.1016/j.talanta.2014.04.030. [DOI] [PubMed] [Google Scholar]
  • 65.Song HS, Kwon OS, Lee SH, Park SJ, Kim UK, Jang J, Park TH. Human taste receptor-functionalized field effect transistor as a human-like nanobioelectronic tongue. Nano Lett. 2013;13:172–178. doi: 10.1021/nl3038147. [DOI] [PubMed] [Google Scholar]
  • 66.Tian SY, Deng SP, Chen ZX. Multifrequency large amplitude pulse voltammetry: A novel electrochemical method for electronic tongue. Sensors Actuators B Chem. 2007;123:1049–1056. doi: 10.1016/j.snb.2006.11.011. [DOI] [Google Scholar]
  • 67.Toko K. Taste sensor with global selectivity. Mater Sci Eng C. 1996;4:69–82. doi: 10.1016/0928-4931(96)00134-8. [DOI] [Google Scholar]
  • 68.Winquist F, Holmin S, Krantz-Rülcker C, Wide P, Lundström I. A hybrid electronic tongue. Anal Chim Acta. 2000;406:147–157. doi: 10.1016/S0003-2670(99)00767-9. [DOI] [Google Scholar]
  • 69.Winquist F, Wide P, Lundström I. An electronic tongue based on voltammetry. Anal Chim Acta. 1997;357:21–31. doi: 10.1016/S0003-2670(97)00498-4. [DOI] [Google Scholar]
  • 70.Gabl R, Feucht HD, Zeininger H, Eckstein G, Schreiter M, Primig R, Pitzer D, Wersing W. First results on label-free detection of DNA and protein molecules using a novel integrated sensor technology based on gravimetric detection principles. Biosens Bioelectron. 2004;19:615–620. doi: 10.1016/S0956-5663(03)00259-8. [DOI] [PubMed] [Google Scholar]
  • 71.Weber J, Albers WM, Tuppurainen J, Link M, Gabl R, Wersing W, Schreiter M. Shear mode FBARs as highly sensitive liquid biosensors. Sensors Actuators, A Phys. 2006;128:84–88. doi: 10.1016/j.sna.2006.01.005. [DOI] [Google Scholar]
  • 72.Wingqvist G, Bjurström J, Liljeholm L, Yantchev V, Katardjiev I. Shear mode AlN thin film electro-acoustic resonant sensor operation in viscous media. Sensors Actuators, B Chem. 2007;123:466–473. doi: 10.1016/j.snb.2006.09.028. [DOI] [Google Scholar]
  • 73.Wingqvist G, Yantchev V, Katardjiev I. Mass sensitivity of multilayer thin film resonant BAW sensors. Sensors Actuators, A Phys. 2008;148:88–95. doi: 10.1016/j.sna.2008.07.023. [DOI] [Google Scholar]
  • 74.Wingqvist G, Bjurstrom J, Hellgren A, Katardjiev I. Immunosensor utilizing a shear mode thin film bulk acoustic sensor. Sensors Actuators B Chem. 2007;127:248–252. doi: 10.1016/j.snb.2007.07.051. [DOI] [Google Scholar]
  • 75.Bowden M, Song L, Walt DR. Development of a microfluidic plaform with an optical imaging microarray capable of attomolar targe DNA detection. Anal Chem. 2005;77:5583–5588. doi: 10.1021/ac050503t. [DOI] [PubMed] [Google Scholar]
  • 76.Hou C, Dong J, Zhang G, Lei Y, Yang M, Zhang Y, Liu Z, Zhang S, Huo D. Colorimetric artificial tongue for protein identification. Biosens Bioelectron. 2011;26:3981–3986. doi: 10.1016/j.bios.2010.11.025. [DOI] [PubMed] [Google Scholar]
  • 77.Rocha DL, Machado MC, Melchert WR. A sensitive flow-based procedure for spectrophotometric speciation analysis of inorganic bromine in waters. Talanta. 2014;129:93–99. doi: 10.1016/j.talanta.2014.05.013. [DOI] [PubMed] [Google Scholar]
  • 78.Zhang C, Suslick KS. Colorimetric sensor array for soft drink analysis. J Agric Food Chem. 2007;55:237–242. doi: 10.1021/jf0624695. [DOI] [PubMed] [Google Scholar]
  • 79.Phillips M, Gleeson K, Hughes JMB, Greenberg J, Cataneo RN, Baker L, McVay WP. Volatile organic compounds in breath as markers of lung cancer: A cross-sectional study. Lancet. 1999;353:1930–1933. doi: 10.1016/S0140-6736(98)07552-7. [DOI] [PubMed] [Google Scholar]
  • 80.Shirasu M, Touhara K. The scent of disease: Volatile organic compounds of the human body related to disease and disorder. J Biochem. 2011;150:257–266. doi: 10.1093/jb/mvr090. [DOI] [PubMed] [Google Scholar]
  • 81.Chan HP, Lewis C, Thomas PS. Exhaled breath analysis: Novel approach for early detection of lung cancer. Lung Cancer. 2009;63:164–168. doi: 10.1016/j.lungcan.2008.05.020. [DOI] [PubMed] [Google Scholar]
  • 82.D'Amico A, Pennazza G, Santonico M, Martinelli E, Roscioni C, Galluccio G, Paolesse R, Di Natale C. An investigation on electronic nose diagnosis of lung cancer. Lung Cancer. 2010;68:170–176. doi: 10.1016/j.lungcan.2009.11.003. [DOI] [PubMed] [Google Scholar]
  • 83.Albert KJ, Lewis NS, Schauer CL, Sotzing GA, Stitzel SE, Vaid TP, Walt DR. Cross-reactive chemical sensor arrays. Chem Rev. 2000;100:2595–626. doi: 10.1021/cr980102w. [DOI] [PubMed] [Google Scholar]
  • 84.Ciosek P, Wróblewski W. Sensor arrays for liquid sensing--electronic tongue systems. Analyst. 2007;132:963–978. doi: 10.1039/b705107g. [DOI] [PubMed] [Google Scholar]
  • 85.Josse F, Bender F, Cernosek RW. Sensors for Chemical and Biochemical Detection in Liquids. Society. 2004;73:5937–5944. doi: 10.1021/ac010859e. [DOI] [PubMed] [Google Scholar]
  • 86.Thompson M, Kiplingt AL, Duncan-hewitts WC. Thickness-shear-mode Acoustic Wave. 1991;116 [Google Scholar]
  • 87.Fenniri H, Chun S, Terreau O, Bravo-Vasquez JP. Preparation and infrared/raman classification of 630 spectroscopically encoded styrene copolymers. J Comb Chem. 2008;10:31–36. doi: 10.1021/cc7001292. [DOI] [PubMed] [Google Scholar]
  • 88.Bravo-Vasquez JP, Alvarez-Puebla RA, Fenniri H. Self-encoded polymer beads for microarray technologies. Sensors Actuators, B Chem. 2007;125:357–359. doi: 10.1016/j.snb.2007.03.014. [DOI] [Google Scholar]
  • 89.Chun S, Xu J, Cheng J, Ding L, Winograd N, Fenniri H. Spectroscopically encoded resins for high throughput imaging time-of-flight secondary ion mass spectrometry. J Comb Chem. 2006;8:18–25. doi: 10.1021/cc050086y. [DOI] [PubMed] [Google Scholar]
  • 90.Fenniri H, Ding L, Ribbe AE, Zyrianov Y. Barcoded resins: A new concept for polymer-supported combinatorial library self-deconvolution [18] J Am Chem Soc. 2001;123:8151–8152. doi: 10.1021/ja016375h. [DOI] [PubMed] [Google Scholar]
  • 91.Fenniri H, Chun S, Ding L, Zyrianov Y, Hallenga K. Preparation, physical properties, on-bead binding assay and spectroscopic reliability of 25 barcoded polystyrene - poly(ethylene glycol) graft copolymers. J Am Chem Soc. 2003;125:10546–10560. doi: 10.1021/ja035665q. [DOI] [PubMed] [Google Scholar]
  • 92.Fenniri H, Hedderich HG, Haber KS, Achkar J, Taylor B. Towards the DRED of resin-supported combinatorial libraries: A non-invasive methodology based on bead self-encoding and multispectral imaging. Angew Chemie - Int Ed. 2000;39:4483–4485. doi: 10.1002/1521-3773(20001215)39:24&#x0003c;4483&#x02237;AID-ANIE4483&#x0003e;3.0.CO;2-4. [DOI] [PubMed] [Google Scholar]
  • 93.Fenniri H, Terreau O, Chun S, Oh SJ, Finney WF, Morris MD. Classification of spectroscopically encoded resins by Raman mapping and infrared hyperspectral imaging. J Comb Chem. 2006;8:192–198. doi: 10.1021/cc050128i. [DOI] [PubMed] [Google Scholar]
  • 94.Bravo-Vasquez JP, Zhu J, Fitzgerald J, Fenniri H. Fabrication of Cross-Reactive, Self-encoded Polymer Beads Array. Unpublished work n.d. [Google Scholar]
  • 95.Zhu J, Bravo-Vasquez JP, Fitzgerald J, Fenniri H. Cross-Reactive, Self-encoded Polymer Film Arrays for Sensor Applications. Unpublished Work. doi: 10.1007/978-1-4939-9616-2_1. n.d. [DOI] [PubMed] [Google Scholar]
  • 96.Raez J, Biais DR, Zhang Y, Alvarez-Puebla RA, Bravo-Vasquez JP, Pezacki JP, Fenniri H. Spectroscopically encoded microspheres for antigen biosensing. Langmuir. 2007;23:6482–6485. doi: 10.1021/la700701x. [DOI] [PubMed] [Google Scholar]
  • 97.Blais DR, Alvarez-Puebla RA, Bravo-Vasquez JP, Fenniri H, Pezacki JP. Multiplex pathogen detection based on spatially addressable microarrays of barcoded resins. Biotechnol J. 2008;3:949–953. doi: 10.1002/biot.200700236. [DOI] [PubMed] [Google Scholar]
  • 98.Alvarez-Puebla RA, Bravo-Vasquez JP, Cui B, Veres T, Fenniri H. SERS classification of highly related performance enhancers. Chem Med Chem. 2007;2:1165–1167. doi: 10.1002/cmdc.200700085. [DOI] [PubMed] [Google Scholar]
  • 99.Zhang J, Gao Y, Alvarez-Puebla RA, Buriak JM, Fenniri H. Synthesis and SERS properties of nanocrystalline gold octahedra generated from thermal decomposition of HAuCl4 in block copolymers. Adv Mater. 2006;18:3233–3237. doi: 10.1002/adma.200601368. [DOI] [Google Scholar]
  • 100.Alvarez-Puebla RA, Bravo-Vasquez JP, Cheben P, Xu DX, Waldron P, Fenniri H. SERS-active Ag/Au bimetallic nanoalloys on Si/SiOx. J Colloid Interface Sci. 2009;333:237–241. doi: 10.1016/j.jcis.2009.01.070. [DOI] [PubMed] [Google Scholar]
  • 101.Bravo-Vasquez JP, Fenniri H. Single-molecule SERRS detection platforms obtained via galvanic displacement on silver fractals. J Phys Chem C. 2009;113:12897–12900. doi: 10.1021/jp901469w. [DOI] [Google Scholar]
  • 102.Farah AA, Alvarez-Puebla RA, Fenniri H. Chemically stable silver nanoparticle-crosslinked polymer microspheres. J Colloid Interface Sci. 2008;319:572–576. doi: 10.1016/j.jcis.2007.11.030. [DOI] [PubMed] [Google Scholar]
  • 103.Farah AA, Bravo-Vasquez JP, Alvarez-Puebla RA, Cho JY, Fenniri H. Robust au-peg/ps microbeads as optically stable platforms for SERS. Small. 2009;5:1283–1286. doi: 10.1002/smll.200801398. [DOI] [PubMed] [Google Scholar]
  • 104.Gromski PS, Correa E, Vaughan AA, Wedge DC, Turner ML, Goodacre R. A comparison of different chemometrics approaches for the robust classification of electronic nose data. Anal Bioanal Chem. 2014;406:7581–7590. doi: 10.1007/s00216-014-8216-7. [DOI] [PubMed] [Google Scholar]
  • 105.Heikenfeld J. Bioanalytical devices: Technological leap for sweat sensing. Nature. 2016;529:475–476. doi: 10.1038/529475a. [DOI] [PubMed] [Google Scholar]
  • 106.Oncescu V, O'Dell D, Erickson D. Smartphone based health accessory for colorimetric detection of biomarkers in sweat and saliva. Lab Chip. 2013;13:3232–8. doi: 10.1039/c3lc50431j. [DOI] [PubMed] [Google Scholar]
  • 107.Patil SJ, Duragkar N, Rao VR. An ultra-sensitive piezoresistive polymer nano-composite microcantilever sensor electronic nose platform for explosive vapour detection. Sensors Actuators, B Chem. 2014;192:444–451. doi: 10.1016/j.snb.2013.10.111. [DOI] [Google Scholar]
  • 108.Peng H, Tisch G, Adams U, Hakim O, Shehada M, Broza N, Billan YY, Abdah-Bortnyak S, Kuten R, Haick A. Diagnosing lung cancer in exhaled breath using gold nanoparticles. Nat Nanotechnol. 2009;4:669–673. doi: 10.1038/nnano.2009.235. http://www.nature.com/nnano/journal/v4/n10/full/nnano.2009.235.html. [DOI] [PubMed] [Google Scholar]
  • 109.Peng H, Trock G, Haick E. Detecting Simulated Patterns of Lung Cancer Biomarkers by Random Network of Single-Walled Carbon Nanotubes Coated with Nonpolymeric Organic Materials. [accessed June 9, 2015];Nano Lett. 2008 8:3631–3635. doi: 10.1021/nl801577u. http://pubs.acs.org/doi/pdf/10.1021/nl801577u. [DOI] [PubMed] [Google Scholar]
  • 110.Mazzone PJ, Wang XF, Xu Y, Mekhail T, Beukemann MC, Na J, Kemling JW, Suslick KS, Sasidhar M. Exhaled breath analysis with a colorimetric sensor array for the identification and characterization of lung cancer. J Thorac Oncol. 2012;7:137–42. doi: 10.1097/JTO.0b013e318233d80f. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 111.Peng G, Hakim M, Broza YY, Billan S, Abdah-Bortnyak R, Kuten A, Tisch U, Haick H. Detection of lung, breast, colorectal, and prostate cancers from exhaled breath using a single array of nanosensors. Br J Cancer. 2010;103:542–551. doi: 10.1038/sj.bjc.6605810. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 112.Dragonieri S, Schot R, Mertens BJA, Le Cessie S, Gauw SA, Spanevello A, Resta O, Willard NP, Vink TJ, Rabe KF, Bel EH, Sterk PJ. An electronic nose in the discrimination of patients with asthma and controls. J Allergy Clin Immunol. 2007;120:856–862. doi: 10.1016/j.jaci.2007.05.043. [DOI] [PubMed] [Google Scholar]
  • 113.Sibila O, Garcia-Bellmunt L, Giner J, Merino JL, Suarez-Cuartin G, Torrego A, Solanes I, Castillo D, Valera JL, Cosio BG, Plaza V, Agusti A. Identification of airway bacterial colonization by an electronic nose in Chronic Obstructive Pulmonary Disease. Respir Med. 2014;108:1608–1614. doi: 10.1016/j.rmed.2014.09.008. [DOI] [PubMed] [Google Scholar]
  • 114.Kodogiannis VS. Point-of-care diagnosis of bacterial pathogens in vitro, utilising an electronic nose and wavelet neural networks. Neural Comput Appl. 2013:1–14. doi: 10.1007/s00521-013-1494-8. [DOI] [Google Scholar]
  • 115.Rathnayake N, Åkerman S, Klinge B, Lundegren N, Jansson H, Tryselius Y, Sorsa T, Gustafsson A. Salivary Biomarkers for Detection of Systemic Diseases. PLoS One. 2013;8 doi: 10.1371/journal.pone.0061356. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 116.Rathnayake N, Åkerman S, Klinge B, Lundegren N, Jansson H, Tryselius Y, Sorsa T, Gustafsson A. Salivary biomarkers of oral health - A cross-sectional study. J Clin Periodontol. 2013;40:140–147. doi: 10.1111/jcpe.12038. [DOI] [PubMed] [Google Scholar]
  • 117.Yoshizawa JM, Schafer CA, Schafer JJ, Farrell JJ, Paster BJ, Wong DTW. Salivary biomarkers: Toward future clinical and diagnostic utilities. Clin Microbiol Rev. 2013;26:781–791. doi: 10.1128/CMR.00021-13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 118.Mark H, Harding CR. Amino acid composition, including key derivatives of eccrine sweat: Potential biomarkers of certain atopic skin conditions. Int J Cosmet Sci. 2013;35:163–168. doi: 10.1111/ics.12019. [DOI] [PubMed] [Google Scholar]
  • 119.Raiszadeh MM, Ross MM, Russo PS, Schaepper MA, Zhou W, Deng J, Ng D, Dickson A, Dickson C, Strom M, Osorio C, Soeprono T, Wulfkuhle JD, Petricoin EF, Liotta LA, Kirsch WM. Proteomic analysis of eccrine sweat: Implications for the discovery of schizophrenia biomarker proteins. J Proteome Res. 2012;11:2127–2139. doi: 10.1021/pr2007957. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 120.Jadoon S, Karim S, Akram MR, Kalsoom Khan A, Zia MA, Siddiqi AR, Murtaza G. Recent developments in sweat analysis and its applications. Int J Anal Chem. 2015;2015 doi: 10.1155/2015/164974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 121.Parikh CR, Dahl NK, Chapman AB, Bost JE, Edelstein CL, Comer DM, Zeltner R, Tian X, Grantham JJ, Somlo S. Evaluation of urine biomarkers of kidney injury in polycystic kidney disease. Kidney Int. 2012;81:784–90. doi: 10.1038/ki.2011.465. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 122.Rigau M, Olivan M, Garcia M, Sequeiros T, Montes M, Colás E, Llauradó M, Planas J, de Torres I, Morote J, Cooper C, Reventós J, Clark J, Doll A. The present and future of prostate cancer urine biomarkers. Int J Mol Sci. 2013;14:12620–12649. doi: 10.3390/ijms140612620. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 123.Watson L, Tullus K, Pilkington C, Chesters C, Marks SD, Newland P, Jones CA, Beresford MW. Urine biomarkers for monitoring juvenile lupus nephritis: A prospective longitudinal study. Pediatr Nephrol. 2014;29:397–405. doi: 10.1007/s00467-013-2668-4. [DOI] [PubMed] [Google Scholar]
  • 124.Hek K, Direk N, Newson RS, Hofman A, Hoogendijk WJG, Mulder CL, Tiemeier H. Anxiety disorders and salivary cortisol levels in older adults: A population-based study. Psychoneuroendocrinology. 2013;38:300–305. doi: 10.1016/j.psyneuen.2012.06.006. [DOI] [PubMed] [Google Scholar]
  • 125.Pruessner M, Béchard-Evans L, Boekestyn L, Iyer SN, Pruessner JC, Malla AK. Attenuated cortisol response to acute psychosocial stress in individuals at ultra-high risk for psychosis. Schizophr Res. 2013;146:79–86. doi: 10.1016/j.schres.2013.02.019. [DOI] [PubMed] [Google Scholar]
  • 126.Zandstra ARE, Hartman CA, Nederhof E, Van den Heuvel ER, Dietrich A, Hoekstra PJ, Ormel J. Chronic stress and adolescents' mental health: Modifying effects of basal cortisol and parental psychiatric history. The TRAILS study. J Abnorm Child Psychol. 2015;43:1119–1130. doi: 10.1007/s10802-014-9970-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 127.Walker EF, Trotman HD, Pearce BD, Addington J, Cadenhead KS, Cornblatt BA, Heinssen R, Mathalon DH, Perkins DO, Seidman LJ, Tsuang MT, Cannon TD, McGlashan TH, Woods SW. Cortisol levels and risk for psychosis: Initial findings from the North American Prodrome Longitudinal Study. Biol Psychiatry. 2013;74:410–417. doi: 10.1016/j.biopsych.2013.02.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 128.Obayashi K. Salivary mental stress proteins. Clin Chim Acta. 2013;425:196–201. doi: 10.1016/j.cca.2013.07.028. [DOI] [PubMed] [Google Scholar]
  • 129.Akcali A, Huck O, Tenenbaum H, Davideau JL, Buduneli N. Periodontal diseases and stress: A brief review. J Oral Rehabil. 2013;40:60–68. doi: 10.1111/j.1365-2842.2012.02341.x. [DOI] [PubMed] [Google Scholar]
  • 130.Noto Y, Sato T, Kudo M, Kurata K, Hirota K. The relationship between salivary biomarkers and state-trait anxiety inventory score under mental arithmetic stress: A pilot study. Anesth Analg. 2005;101:1873–1876. doi: 10.1213/01.ANE.0000184196.60838.8D. [DOI] [PubMed] [Google Scholar]
  • 131.Slavish DC, Graham-Engeland JE, Smyth JM, Engeland CG. Salivary markers of inflammation in response to acute stress. Brain Behav Immun. 2015;44:253–269. doi: 10.1016/j.bbi.2014.08.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 132.Nater UM, Rohleder N. Salivary alpha-amylase as a non-invasive biomarker for the sympathetic nervous system: Current state of research. Psychoneuroendocrinology. 2009;34:486–496. doi: 10.1016/j.psyneuen.2009.01.014. [DOI] [PubMed] [Google Scholar]
  • 133.Rohleder N, Nater UM, Wolf JM, Ehlert U, Kirschbaum C. Psychosocial stress-induced activation of salivary alpha-amylase: An indicator of sympathetic activity? Ann N Y Acad Sci. 2004;1032:258–263. doi: 10.1196/annals.1314.033. [DOI] [PubMed] [Google Scholar]
  • 134.Schumacher S, Kirschbaum C, Fydrich T, Ströhle A. Is salivary alpha-amylase an indicator of autonomic nervous system dysregulations in mental disorders?-A review of preliminary findings and the interactions with cortisol. Psychoneuroendocrinology. 2013;38:729–743. doi: 10.1016/j.psyneuen.2013.02.003. [DOI] [PubMed] [Google Scholar]
  • 135.Ali N, Pruessner JC. The salivary alpha amylase over cortisol ratio as a marker to assess dysregulations of the stress systems. Physiol Behav. 2012;106:65–72. doi: 10.1016/j.physbeh.2011.10.003. [DOI] [PubMed] [Google Scholar]
  • 136.Campbell J, Ehlert U. Acute psychosocial stress: Does the emotional stress response correspond with physiological responses? Psychoneuroendocrinology. 2012;37:1111–1134. doi: 10.1016/j.psyneuen.2011.12.010. [DOI] [PubMed] [Google Scholar]
  • 137.Byrne ML, O'Brien-Simpson NM, Reynolds EC, Walsh KA, Laughton K, Waloszek JM, Woods MJ, Trinder J, Allen NB. Acute phase protein and cytokine levels in serum and saliva: A comparison of detectable levels and correlations in a depressed and healthy adolescent sample. Brain Behav Immun. 2013;34:164–175. doi: 10.1016/j.bbi.2013.08.010. [DOI] [PubMed] [Google Scholar]
  • 138.Baker DG, Nievergelt CM, O'Connor DT. Biomarkers of PTSD: Neuropeptides and immune signaling. Neuropharmacology. 2012;62:663–673. doi: 10.1016/j.neuropharm.2011.02.027. [DOI] [PubMed] [Google Scholar]
  • 139.Craig-Schapiro R, Fagan AM, Holtzman DM. Biomarkers of Alzheimer's disease. Neurobiol Dis. 2009;35:128–140. doi: 10.1016/j.nbd.2008.10.003. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES