Instrumented cardiac microphysiological devices via multi-material 3D printing

Johan U Lind 1,2,4, Travis A Busbee 1,2,4, Alexander D Valentine 1,2, Francesco S Pasqualini 1,2, Hongyan Yuan 1,2,3, Moran Yadid 1,2, Sung-Jin Park 1,2, Arda Kotikian 1,2, Alexander P Nesmith 1,2, Patrick H Campbell 1,2, Joost J Vlassak 2, Jennifer A Lewis 1,2,*, Kevin K Parker 1,2,*
PMCID: PMC5321777  NIHMSID: NIHMS819005  PMID: 27775708

Abstract

Biomedical research has relied on animal studies and conventional cell cultures for decades. Recently, microphysiological systems (MPS), also known as organs-on-chips, that recapitulate the structure and function of native tissues in vitro, have emerged as a promising alternative1. However, current MPS typically lack integrated sensors and their fabrication requires multi-step lithographic processes2. Here, we introduce a facile route for fabricating a new class of instrumented cardiac microphysiological devices via multi-material 3D printing. Specifically, we designed six functional inks, based on piezo-resistive, high conductance, and biocompatible soft materials that enable integration of soft strain gauge sensors within micro-architectures that guide the self-assembly of physio-mimetic laminar cardiac tissues. We validated that these embedded sensors provide non-invasive, electronic readout of tissue contractile stresses, inside cell incubator environments. We further applied these devices to study drug responses, as well as the contractile development of human stem cell derived laminar cardiac tissues over four weeks.


Current MPS models of muscle tissue rely on microscopy coupled with optical tracking analysis for assessing tissue contractile stress. For instance, muscular thin film (MTF) assays track changes in curvature of soft cantilever substrates induced by the contraction of a laminar tissue3, and micro-post assays measure the deflection of pillars supporting a micro-tissue4. Although these assays have proven valuable for short-term modeling of human disease and small-scale drug screening applications1,5,6, they are not well suited for higher throughput or longer term studies. Moreover, biomimetic microsystems are currently fabricated using soft material lithography-based techniques that require multiple steps, masks and dedicated tools4,79, which hinders rapid prototyping and customization. By contrast, multi-material 3D printing of viscoelastic inks enables a wide range of functional, structural and biological materials to be patterned and integrated in a single programmable manufacturing step1014.

Here, we introduce a fully 3D printed and instrumented microphysiological device that provides continuous electronic readout of the contractile stress of multiple laminar cardiac micro-tissues (Fig. 1a–b). Each device contains three key features: i) multilayer cantilevers, composed of a base layer, an embedded strain sensor, and a tissue-guiding layer, ii) electrical interconnects for readout, and iii) eight independent wells (Fig. 1c–i). The tissue-guiding layer promotes self-assembly of engineered physio-mimetic laminar tissues from neonatal rat ventricular myocytes (NRVMs) and human induced pluripotent stem cell derived cardiomyocytes (hiPS-CMs). Our microphysiological device facilitates tissue culture and non-invasive analyses of tissue contractile strength over several weeks, and facilitates drug studies inside a controlled incubator environment.

Figure 1. Device principle and microscale 3D-printing procedure.

Figure 1

(a) Principle sketch of the device. Contraction of an anisotropic engineered cardiac tissue (1) deflects a cantilever substrate (2), thereby stretching a soft strain gauge embedded in the cantilever. This generates a resistance change proportional to the contractile stress of the tissue (3). (b) The fully printed final device. Insert 1: Confocal microscopy image of immuno-stained laminar NRVM cardiac tissue on the cantilever surface. Blue: DAPI nuclei stain, White: α-actinin, scale bar 10 μm. Insert 2: Still images of a cantilever deflecting upon tissue contraction. Insert 3: Example resistance signal. (c–i) Automated printing of the device on a 2 inch x 3 inch glass slide substrate in 7 sequential steps. For each step, a corresponding still image from the printing procedure is displayed. For steps 1–5, a stylus profiling cross-sectional contour of the cantilever is additionally displayed. (c) In print step 1, a 0.5 μm Dextran thin film sacrificial layer is printed (d) In print step 2, a 3 μm TPU thin film cantilever base is printed (e) In print step 3, a 6.5 μm thick CB:TPU strain sensor loop is added to cantilever base. (f) In print step 4, a 1.5 μm TPU wire cover is added. (g) In print step 5, 20 μm tall, 60 μm wide PDMS micro-filaments are printed in slightly overlapping lines. The filaments constitute the top part of the cantilever and guide cardiomyocytes to form anisotropic laminar tissues. (h) In print step 6, electrical leads and contact are added using a high conductivity Ag:PA ink (i) In print step 7, covers to insulate exposed wires and wells to contain cells and media are printed using PDMS, PLA or ABS (See supplementary Fig S10).

To create this integrated device, six materials are patterned sequentially via direct ink writing multi-material 3D printing. To allow material integration at the microscale, the substrate topology and x-y-z positions of four individually addressable nozzles are determined by an automated calibration process (see supplementary movie S1). After this calibration, the device is printed in a single continuous procedure (Fig. 1c–i and supplementary movies S2–3).

The stress generated by laminar cardiac tissues is limited to 1–15 kPa5,8. Hence, to match this range, both the thickness and stiffness of each cantilever layer must be minimized. Towards this objective, we designed a series of highly dilute polymer-based inks. Their low solids content ensures patterning of thin individual layers (0.5 – 6.5 μm in thickness). Additionally, by tuning the evaporation rate of the carrier solvent solution, the ink viscosity and corresponding wetting and spreading behavior is controlled to achieve the desired lateral dimensions (supplementary Fig. S1–2). Using these inks, 0.5 μm dextran films are first printed (Fig. 1c). These serve as biocompatible, water-soluble sacrificial release layers that allow the final cantilevers to detach from the substrate and deflect freely. Next, using dilute thermoplastic polyurethane (TPU) inks, 3 μm thick cantilever bases, 6.5 μm thick strain gauge wires and 1.5 μm thick wire covers are printed in steps 2–4, respectively (Fig. 1d–f). The cantilever base and wire covers are printed using an unfilled TPU ink, while the strain gauge wires are printed using a TPU ink filled with 25 wt% carbon black nanoparticles (CB:TPU). The printed TPU-based features exhibit elastic mechanical properties with Young’s modulus of 1.6 MPa (supplementary Fig. S3), while CB:TPU features readily cure to form an elastic piezo-resistive material with Young’s modulus of 8.8 MPa and resistivity of 1.19 Ω·cm (supplementary Fig. S3–7). Several alternative fillers were investigated, including metals particles15 and carbon nanotubes16. However, we found that carbon black imparts the best combination of ink rheology, low stiffness and sensor hysteresis.

The remainder of the microphysiological device is printed using concentrated viscoelastic inks optimized for deposition of self-supporting structures (supplementary Fig. S8–9). Using a soft viscous PDMS ink (Young’s modulus of 1.28 MPa, see Supplementary Fig. S3), ~60 μm wide filaments are patterned on top of each cantilever (Fig. 1g). These comprise the majority of the cantilever thickness and serve as direct support below the physio-mimetic laminar cardiac tissue. Next, electrical leads and contact pads are printed using a high-conductivity, silver particle-filled, polyamide (Ag:PA) ink (Fig. 1h). Upon drying, the printed wires exhibit an electrical resistivity of 6.6x10−5 Ω·cm (supplementary Fig. S4), which ensures that the primary electrical resistance of the final device arises from the embedded CB:TPU strain gauges. Finally, the wire leads are covered with an insulating layer and eight individually addressable wells are printed using either a PDMS ink or biocompatible rigid thermoplastic polymers such as polylactid acid (PLA) or acrylonitrile butadiene styrene (ABS). The rigid polymers can be preferable for drug study applications, as they are less prone to bulk absorption of hydrophobic drugs than PDMS17 (Fig. 1i, supplementary Fig. S10). After printing, the devices are cured at 100°C and subsequently seeded with cardiomyocytes, which self-assemble into laminar tissues mimicking the structure of the native heart.

The musculature of the heart is composed of highly organized, structurally and electrically anisotropic layers18. To recapitulate this architecture in vitro, we printed a range of grooved microstructures by varying the spacing between curved ~20 μm x 60 μm (height x width) filaments and assessed their ability to guide the self-assembly of anisotropic laminar NRVM tissues (Fig. 2). We evaluated the degree of tissue structural anisotropy using immunofluorescent imaging of sarcomeric α-actinin and quantification of the sarcomere orientational order parameter (OOP) (Fig. 2a–f, supplementary Fig. S11)19. All substrates showed an OOP > 0.2 indicative of non-random sarcomere alignment. The highest OOP ~0.5 was observed for 60 μm filament spacing. Similarly, 60 μm filament spacing gave rise to the highest degree of electrophysiological anisotropy in the laminar NRVM tissues (Fig. 2g–i, supplementary Movies S4–7). For this spacing, action potential propagation was 2.7 times faster parallel to the grooves compared to the transverse direction. These data are in agreement with the longitudinal to transverse velocity ratio of 2.1 observed in the native ventricle18. Also, these anisotropic laminar tissues gave rise to unidirectional and concerted cantilever deflection and displayed the highest contractile stresses in the final devices (supplementary Fig. S12–14). Collectively, these results demonstrate that our printing technique can be employed to engineer laminar cardiac tissues with a range of ordered architectures. Notably, the OOP and electrophysiology data closely matched values obtained using semi-manual micro-contact printing and molding techniques7,20,21.

Figure 2. Micro-grooves guide cardiomyocyte self-assembly into anisotropic engineered tissues.

Figure 2

(a) Spacing of soft PDMS microfilaments (b) Sketch of micro-filaments guiding self-assembly of engineered cardiac tissue. (c) Stylus profilometer contours of substrates with filaments printed at 40, 60, 80 and 100 μm spacing. (d) Sarcomere OOP of laminar NRVM tissues developed on substrates with 40 (n=9), 60 (n=13), 80 (n=8) and 100 μm (n=10) filament spacing, error bars are S.E.M., *: P<0.05. (e–f) Representative confocal images from OOP dataset, Blue: DAPI nuclei stain, White: α-actinin (e) z-projection, scale bars 10 μm (f) x-z line scan. (g) Ratio between action potential (AP) propagation speed parallel and orthogonal to the grooves for laminar NRVM tissue developed on substrates with 40, 60, 80 and 100 μm filament spacing, (n≥3). Individual data points included (circles), error bars are S.E.M. (h) Representative activation time heat maps for AP data set, overlaid wide field microscope image of the samples as guide to the eye, scale bars 0.8 mm. Activation times normalized to maximum observed activation (tmax, Red), to account for tissue source variation. 2 Hz electrical point-stimulation is applied in top left corner of samples. Mean observed propagation speed parallel (Vpara) and orthogonal (Vortho) to grooves is displayed as vectors. (i) Normalized AP traces at four corners of activation map samples. Blue: at AP initiation corner, Red: At tmax-corner, Black: Corner parallel to grooves, Green: Corner orthogonal to grooves.

To accurately measure the stress exerted by the physio-mimetic laminar cardiac tissues, the embedded strain gauges within each cantilever must produce a reliable and detectable signal. We therefore evaluated the electro-mechanical properties of CB:TPU in a series of uniaxial strain tests (Fig. 3a–d, supplementary Fig. S5–7). To mimic the strains exerted by tissue contraction, 1–3 Hz triangular strains of 0.1% were applied to CB:TPU gauges. In this strain range, the gauges displayed a linear strain-resistance relationship with limited hysteresis and a corresponding gauge factor of 2.51 ± 0.09. In order to accurately convert resistance signals to stresses generated by the laminar cardiac tissues, we established a mechanical model of the cantilevers based on a multilayer version of Stoney’s equation (Fig. 3e–f and supplementary information). The mechanical model relies on the dimensions and Young’s moduli of each layer of the cantilever as well as the CB:TPU gauge factor. The model provides linear conversion factors between the relative changes in gauge resistance, changes in cantilever radius of curvature, and in turn, the longitudinal stresses generated by the laminar cardiac tissues. Applying these conversion factors, we obtained tissue twitch stress (σ), the differences between the systolic and diastolic stress, in the range of 7–15 kPa for laminar NRVM tissues, akin to prior reports3,7. To further verify the model, we compared the sensor readout with stress values concurrently obtained through optical tracking of cantilever deflection, and obtained highly similar stress values (Fig. 3g–h, supplementary Fig. S12–13, supplementary Movie S8–11). Yet, we observed one discrepancy between the optical tracking and electrical readout: at the initiation of tissue contraction, a small decrease in gauge resistance is seen, prior to the increase caused by cantilever deflection. We attribute this decrease to a brief axial compression of the gauge wire occurring during initial tissue contraction, while cantilever deflection is counteracted by inertial and viscous forces. Notably, this observation does not influence tissue twitch stress values. Together, these results demonstrate that the CB:TPU strain-gauges embedded in our fully printed platform provide reliable readouts of tissue contractile stress.

Figure 3. CB:TPU gauge factor, sensor readout and example drug-dose studies.

Figure 3

(a) Sketch of Instron test setup for determining CB:TPU gauge factor (GF). (b–d) Relative change in CB:TPU resistance upon triangular cyclic straining to 0.1% at 1 Hz. (c) Dark grey line: Observed relative resistance change. Red dotted line: Strain applied. (d) Relative resistance change vs. applied strain. Orange dotted line indicates linear fit to part of cycles with increasing strain yielding a gauge factor of 2.56. (e) Wide-field microscope images of cantilever bending upon tissue contraction. Minimum deflection corresponds to cardiac diastole (1) and peak deflection corresponding to systole (2). (f) Sketch of mechanical model applied to convert resistance change to stress generated by the tissue, see supplementary information. (g–h) Relative resistance changes (left axis) and corresponding calculated tissue stress (right axis) recorded from spontaneously beating cantilever. Blue dotted line in (h) indicates stress determined independently by optical tracking of the cantilever radius of curvature. (i) Representative traces of stress generated by a laminar NRVM tissue when tissue is exposed to verapamil and corresponding dose-response curve (n=4), error bars are S.E.M, stress normalized between maximal and minimal values, tissue paced at 1 Hz, apparent EC50 1.12x10−6 M. Individual data points included (circles) (j) Representative traces of stress generated by a laminar hiPS-CM tissue when exposed to isoproterenol and corresponding dose-response curve, (n=10), error bars are S.E.M, stress normalized between maximal and minimal values, tissue paced at 2 Hz, apparent EC50 2.74x10−9 M.

Data acquisition from cardiac microphysiological systems based on optical readout is laborious, requires dedicated microscopy setups and semi-automated image analysis software, and is limited by the image readout file sizes4,5. By contrast, our fully printed and instrumented cardiac devices provide a direct, non-invasive electronic readout of contractile stresses. Therefore, we could perform dose response studies of drugs that influence contraction strength or beat rate, directly inside a cell incubator, an ideal in vitro environment (Fig. 3i–j, supplementary Fig. S15–18). As examples, we carried out cumulative drug dose studies with the L-type calcium channel blocker verapamil and the β-adrenergic agonist isoproterenol. For laminar NRVM tissues we observed a negative inotropic response to verapamil, with an apparent EC50 of 1.15x10−6 M when paced at 1 Hz (Fig. 3i). We also observed a negative chronotropic effect for spontaneously beating tissues (supplementary Fig. S17), in good agreement with prior studies on isolated postnatal whole rat hearts22. Similarly, we observed a positive chronotropic response to isoproterenol for spontaneously beating laminar tissues, matching previous studies of engineered NRVM micro-tissues4 (supplementary Fig. S18). For hiPS-CM-based laminar tissues, we observed a positive inotropic response to isoproterenol, with an apparent EC50 of 2.69x10−9 M when paced at 2 Hz (Fig. 3j), resembling studies based on embryonic stem cell derived cardiomyocyte tissue23.

Beyond facilitating acute drug studies, our platform is well suited for extended studies to identify gradual changes in the contractile stress of engineered cardiac tissues, which can occur over the course of multiple weeks23,24. As a demonstration, we studied the in vitro contractile development of laminar tissues based on hiPS-CMs over the course of 28 days (Fig. 4a–g). During this period, the longitudinal contractile stress initially increased 4-fold from day 2 to 4, followed by a more gradual increase (Fig. 4a–b). Similarly, the spontaneous beating rate decreased (Fig. 4c), indicative of increased maturity25. The changes in twitch stress are mirrored by a structural development of the laminar hiPS-CM tissues (Fig. 4d–g, supplementary Fig. S19–20). From day 2 to day 14 the sarcomere OOP increased from 0.11 to 0.32 (Fig. 4e), while the sarcomere packing density (SPD, a measure of sarcomere periodic organization) increased from 0.07 to 0.2 (Fig. 4f), indicating that hiPS-CMs undergo sarcomerogenesis and myofibrillogenesis during culture19. Importantly, from day 14 to day 28, a significant increase in sarcomere length from 1.7 μm to 1.8 μm was observed (Fig. 4g), indicative of a more mature tissue26.

Figure 4. Long term hips-CM contractile development and thicker laminar NRVM tissue devices.

Figure 4

(a) Representative traces of contractile twitch stress generated by the laminar hiPS-CMs tissues at day 2, 14 and 28 (b–c) Contractile stress and spontaneous beat rate of hiPS-CMs tissues at day 2, 4, 8, 20 and 28 (n≥10) error bars are S.E.M. (d) Immuno-stained laminar hiPS-CMs tissues on device cantilevers at day 2 and day 28 after seeding, scale bars 10 μm. Blue: DAPI nuclei stain White: α-actinin stain, z-projection (e–g) SPD, OOP and sarcomere length of laminar hiPS-CM tissues at day 2, 14 and 28 (n=5) error bars are S.E.M., *: P<0.05. (h) 1: Modified device cantilever containing micro-pin and micro-well to support thicker laminar NRVM tissue, 2: Detail of cantilever with micro-pins 3: Still images of a cantilever deflecting upon tissue contraction. (i) 1: z-projection of immuno-stained thicker laminar tissue on the cantilever surface with micro-pin, scale bar 30 μm. 2: x-z line scan of thicker laminar tissue in grooves, scale bar 10 μm, Blue: DAPI nuclei stain, White: α-actinin, Red: actin. (j–k) Dose-response curves for thicker laminar NRVM tissues. Stress values for thicker tissues assumed linear proportional to relative resistance change: σ~ΔR/R0. Stress normalized between maximal and minimal values. (j) Dose-response curve for verapamil, (n=4), error bars are S.E.M. Individual data points included (circles), tissue paced at 1.5 Hz. Apparent EC50 7.90x10−7 M (k) Dose-response curve for isoproterenol, (n=3) error bars are S.E.M. Individual data points included (circles), tissue paced at 1.5 Hz. Apparent EC50 1.16x10−9 M.

While traditional muscular thin films mimic essential features of the cardiac architecture, cell-cell contacts are confined to a single cell layer. By contrast, a number of in vitro cardiac tissue models composed of thicker micro-tissues have been reported24,2729. To illustrate the versatility of our 3D printing methodology, we modified the MPS device to support thicker micro-tissues, approximately 4 cell layers thick (supplementary Fig. S21–24). This thickness matches that of the myocyte layers in a mammalian heart (i.e., ~ 4 cells in thickness per layer, each layer separated by connective tissue)30. To balance the increased contractile stress and mitigate tissue delamination, we printed thicker cantilevers that contained ~100 μm tall micro-pin arrays with tunable areal density (Fig. 4h–i, supplementary movie S13–15). Despite these modifications to both the cardiac tissue and cantilevers, the tissue remains orders of magnitude softer than the substrate. Hence, the basic assumptions of the mechanical model remain valid, and the sensor signal is directly proportional to mean tissue stress (see supplementary information). We carried out proof of principle isoproterenol and verapamil drug studies to illustrate the functional relevance of these thicker NRVM based tissues (Fig. 4j–k). We observed the expected positive and negative inotropic responses with apparent EC50 values comparable to earlier data from engineered 3D NRVM tissues and isolated postnatal whole rat hearts22,29.

Through multi-material 3D printing of a series of customized inks, we demonstrated the automated design and fabrication of instrumented cardiac microphysiological devices. The integrated sensors drastically simplify data acquisition and long term functional studies. Leveraging the ability to track the temporal development in tissue mechanics will enable new insights into tissue morphogenesis, pathogenesis, and drug-induced structural and functional remodeling. Our digital manufacturing approach is quite versatile, allowing for fabrication of a range of instrumented microphysiological devices. Notably, our approach facilitates rapid customization to match device geometries, mechanical and biochemical properties to a specific diseased state or a unique patient derived cell source. Our programmable microfabrication approach opens new avenues for in vitro tissue engineering, toxicology and drug screening research.

Methods

Ink formulations

The following ink formulations were used: Dextran ink: 10 g/ml Dextran (Sigma-Aldrich) was dissolved in 75:25 v:v, water: isopropanol, TPU ink: TPU 15 wt% Elastollan 35A (BASF) dissolved in 4:1 v:v tetrahydrofuran:dimethylformamide, CB:TPU ink: TPU 15 wt% Elastollan 35A, 5 wt% carbon black (Vulcan XC72R, Cabot) dissolved in 4:1 v:v tetrahydrofuran:dimethylformamide. Ag:Pa ink: 50 g Silver flakes (5–8 μm) mixed with 5.2 g 30 wt% versamid 973 solution (BASF) dissolved in pentanol, mixed with an additional 2.72 g of pentanol. Soft PDMS Ink: SE1700 (Dow-Corning) with 1:25 curing agent based ratio was applied for cantilever covers, micro-pins and micro-wells. Rigid PDMS ink: SE1700 (Dow-Corning), with 1:10 curing agent weight ratio, mixed 1:5 wt:wt with Sylgard 184 1:10 curing agent weight ratio (Dow-Corning) is used for wells and covers, or as gasket adhesive below PLA or ABS wells and covers.

Print procedure with integrated profiling

Printing was carried out using a three-axis motion-controlled stage (Aerotech) with 4 independent z-axes. Prior to printing, a custom automation system was applied to determine the substrate topology and relative x-y-z nozzle positions using an integrated laser profilometer, three CCD micrometers (Keyence), and custom machined fixtures. G-code generation and custom automation scripts were programmed using open-sourced python libraries (Mecode). Extrusion was carried out using syringes, dispense tips, and a digital pneumatic regulator (EFD). Dispense tip diameters: TPU ink: 100 μm, CB:TPU ink: 200 μm, Ag:PA ink: 250 μm, Soft PDMS ink: 30 μm, Rigid PDMS ink: 410 μm. For Dextran ink a refillable contact-pressure sensitive pen (0.7 mm tip, Montana) was utilized as deposition nozzle.

Device cell seeding and culture

Prior to cell seeding, devices were sterilized by UV-Ozone exposure for 8 mins. Subsequently, wells were incubated with a 50 μg/ml solution of Fibronectin (BD biosciences) in PBS for 1 hr. Fibronectin solution was aspirated and wells seeded with either primary NRVMs at a seeding density of a seeding density of 140 k/cm2 in 10% FBS in media 199 (Life Technologies) or hiPS-CMs (Cor4U, Axiogenesis, Germany) at 220 k/cm2 in designated Commercial Cor4U media (Axiogenesis, Germany). For thicker tissue version, NRVM are seeded at 1 M/cm2 adding cold BD Matrigel (BD biosciences) to final concentration 0.45–0.6 mg/ml, thus below gelling concentration. NRVMs were acquired from Sprague Dawley rats (n10 litters, per harvest) applying procedures approved by the Harvard University Animal Care and Use Committee, described in detail in supplementary information. hiPS-CMs were thawed, pre-plated and seeded following guidelines from supplier, as detailed in supplementary information. Cell media was changed at least every second day, applying 2% FBS in media 199 (Life Technologies) for NRVM culture, and designated Cor4U media for hiPS-CM culture.

Data acquisition

Data collection was performed using custom machined holder connecting device to a Keithley Multichannel DMM 3706a. Readouts were obtained either as 2-wire resistance recordings sampling at ≥60 Hz. A custom MATLAB (MathWorks, Natick, MA) code was applied for quantifying relative resistance changes upon tissue contraction applying peak detection and comparison with local baseline. For electrically paced samples a median filter (5) was applied. Linear conversion constants between relative resistance change, cantilever curvature and stress generated by tissue, were established using designated mechanical model.

Isoproterenol and verapamil Cumulative drug dose studies

Cumulative dosing of isoproterenol or verapamil (Sigma-Aldrich) cardiac drugs was performed on laminar NRVM and hiPS-CM tissues inside incubator. 700 μl serum free media was added to each well prior to drug dose experiments. A dilution series of the drugs in media (Life Technologies) was sequentially added in 7 μl doses. Tissue was incubated for 10 mins for each dose, prior to recording. For each dose at least 30 s were recorded per channel. Isoproterenol stocks were kept at 4°C prior to dosing. Pacing was applied using custom platinum wire electrodes. Each n denotes separate device wells with isolated tissue, sensor and media.

Tissue immuno-staining and structural analysis

All immunocytochemistry procedures were conducted at room temperature. Samples were first fixed with 4% PFA/PBS (v/v) solution for 15 mins and then permeabilized with 0.05% Triton-X/PBS (v/v) solution for 10 mins. Subsequently, samples were incubated for 1 hr with a monoclonal sarcomeric α-actinin (clone EA-53; Sigma-Aldrich) primary antibody, washed three times in PBS, and finally counterstained with: Alexa Fluor 488-conjugated anti-mouse secondary antibody, Alexa Fluor 633-conjugated Phalloidin and DAPI (Invitrogen). Samples were imaged using confocal microscopy, acquiring projected z-stack images of the wavy laminar tissues. The alignment and overall spatial organization of α-actinin positive structures in the immunostained digital images were evaluated with custom MATLAB (MathWorks, Natick, MA) code as previously described19.

Optical mapping experiments to determine tissue electrophysiology

AP propagation velocities for the engineered NRVM cardiac tissues were monitored using a modified tandem-lens macroscope (Scimedia, Costa Mesa, CA) equipped with a high speed camera (MiCAM Ultima, Scimedia, Costa Mesa, CA), a plan APO 1X objective, a collimator (Lumencor, Beaverton, OR) and a 200 mW Mercury lamp (X-Cite exacte, Lumen Dynamics, Canada). After 4 days culture in vitro, 10 mm x 10 mm engineered laminar NRVM tissues were incubated with 4 μM solution of a photovoltaic dye, RH237. (Invitrogen, Carlsbad, CA) for 5 mins at 37°C, and rinsed with Tyrode’s solution prior to recording. Recordings were acquired at a frame rate of 200 Hz. Electrical point simulation was applied to the corner of the tissue using two U-shaped platinum electrodes (Sigma-Aldrich, St. Louis, MO) located 0.5–1 mm above the tissue, applying 1–2 Hz, 5–10 V pulses of 10 ms duration using a pulse generator (MyoPacer Cell Stimulator, IonOptix, Milton, MA). Post-processing of data was conducted with custom MATLAB (MathWorks, Natick, MA) code as previously described21. A spatial filter with 3 × 3 pixels was applied to improve the signal-noise ratio similar. Activation time was calculated as the average time to maximum upstroke slope of pulses when continuously paced at 2 Hz during in a 5 s recording window.

Statistical analysis

For analysis of tissue SPD, OOP and sarcomeric length, One Way Analyses for Variance between the compositional groups were conducted using SigmaPlot (v12.0, Systat Software Inc.). All data sets passed Shapiro-Wilk normality tests and equal variance tests. For pairwise comparison Holm-Sidak method was applied. For all statistical analyses, p-values less than 0.05 were considered statistically significant. Sample sizes were chosen based on previous studies applying muscular thin film assays1921.

Code availability

Custom MATLAB scripts for signal detection and stress calculation can be found in supplementary information.

Supplementary Material

1

Acknowledgments

The authors thank Lori K. Sanders for her work on photography and time-lapse movies, Joss A. Goss for his assistance with fabrication of device holder and Jack Minardi for his development of Mecode, and his help with machine automation. This work was performed in part at the Center for Nanoscale Systems (CNS), a member of the National Nanotechnology Infrastructure Network (NNIN), which is supported by the National Science Foundation under NSF award no. ECS-0335765. CNS is part of Harvard University; This work was also supported by the National Center For Advancing Translational Sciences of the National Institutes of Health under Award Number UH3TR000522, the U. S. Army Research Laboratory and the U. S. Army Research Office under Contract No. W911NF-12-2-0036, and the Harvard University Materials Research Science and Engineering Center (MRSEC) award no. DMR-1420570. The views and conclusions contained in this document are those of the authors and should not be interpreted as representing the official policies, either expressed or implied, of the Army Research Office, Army Research Laboratory, the U.S. Government, or the the National Institutes of Health. The U.S. Government is authorized to reproduce and distribute reprints for Government purposes notwithstanding any copyright notation hereon.

Footnotes

Competing Financial Interests

The authors declare no competing financial interests

Author Contributions

J.U.L., T.A.B, J.A.L. and K.K.P designed the study. J.U.L. and T.A.B. designed device. T.A.B. coded 3D-print procedure and automation. J.U.L, T.A.B, A.K., and A.D.V. developed and characterized the printable materials. J.U.L, A.D.V. and T.A.B, optimized and printed devices. P.H.C. performed NRVM harvest and prepared culturing media. J.U.L., M.Y. and A.P.N. performed NRVM culture, drug-dose experiments, and data analysis. M.Y. and J.U.L. conducted hiPS-CM culture, experiments and data analysis. F.S.P and J.U.L. performed tissue staining, confocal imaging, and O.O.P. analysis. S-J.P., J.U.L. conducted optical mapping experiments and analysis. H.Y. and J.J.V. developed mechanical model of device. J.U.L, T.A.B., J.A.L, and K.K.P prepared illustrations and wrote manuscript. F.S.P and A.D.V contributed to writing manuscript.

References

  • 1.Bhatia SN, Ingber DE. Microfluidic organs-on-chips. Nature biotechnology. 2014;32:760–772. doi: 10.1038/nbt.2989. [DOI] [PubMed] [Google Scholar]
  • 2.Mammoto A, et al. Control of lung vascular permeability and endotoxin-induced pulmonary oedema by changes in extracellular matrix mechanics. Nature communications. 2013;4:1759. doi: 10.1038/ncomms2774. [DOI] [PubMed] [Google Scholar]
  • 3.Feinberg AW, et al. Muscular thin films for building actuators and powering devices. Science. 2007;317:1366–1370. doi: 10.1126/science.1146885. [DOI] [PubMed] [Google Scholar]
  • 4.Boudou T, et al. A microfabricated platform to measure and manipulate the mechanics of engineered cardiac microtissues. Tissue Engineering Part A. 2011;18:910–919. doi: 10.1089/ten.tea.2011.0341. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Wang G, et al. Modeling the mitochondrial cardiomyopathy of Barth syndrome with induced pluripotent stem cell and heart-on-chip technologies. Nature medicine. 2014;20:616–623. doi: 10.1038/nm.3545. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Hinson JT, et al. Titin mutations in iPS cells define sarcomere insufficiency as a cause of dilated cardiomyopathy. Science. 2015;349:982–986. doi: 10.1126/science.aaa5458. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Agarwal A, et al. Micropatterning Alginate Substrates for In Vitro Cardiovascular Muscle on a Chip. Advanced Functional Materials. 2013;23:3738–3746. doi: 10.1002/adfm.201203319. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Grosberg A, Alford PW, McCain ML, Parker KK. Ensembles of engineered cardiac tissues for physiological and pharmacological study: Heart on a chip. Lab on a Chip. 2011;11:4165–4173. doi: 10.1039/c1lc20557a. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Park SJ, et al. Phototactic guidance of a tissue-engineered soft-robotic ray. Science. 2016;353:158–162. doi: 10.1126/science.aaf4292. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Lewis JA. Direct ink writing of 3D functional materials. Advanced Functional Materials. 2006;16:2193–2204. [Google Scholar]
  • 11.Hardin JO, Ober TJ, Valentine AD, Lewis JA. Microfluidic Printheads for Multimaterial 3D Printing of Viscoelastic Inks. Advanced Materials. 2015;27:3279–3284. doi: 10.1002/adma.201500222. [DOI] [PubMed] [Google Scholar]
  • 12.Sun K, et al. 3D Printing of Interdigitated Li-Ion Microbattery Architectures. Advanced Materials. 2013;25:4539–4543. doi: 10.1002/adma.201301036. [DOI] [PubMed] [Google Scholar]
  • 13.Miller JS, et al. Rapid casting of patterned vascular networks for perfusable engineered three-dimensional tissues. Nature materials. 2012;11:768–774. doi: 10.1038/nmat3357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Kolesky DB, Homan KA, Skylar-Scott MA, Lewis JA. Three-dimensional bioprinting of thick vascularized tissues. Proceedings of the National Academy of Sciences. 2016;113:3179–3184. doi: 10.1073/pnas.1521342113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Matsuhisa N, et al. Printable elastic conductors with a high conductivity for electronic textile applications. Nature communications. 2015;6:7461. doi: 10.1038/ncomms8461. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Lipomi DJ, et al. Skin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes. Nature nanotechnology. 2011;6:788–792. doi: 10.1038/nnano.2011.184. [DOI] [PubMed] [Google Scholar]
  • 17.Berthier E, Young EW, Beebe D. Engineers are from PDMS-land, Biologists are from Polystyrenia. Lab on a Chip. 2012;12:1224–1237. doi: 10.1039/c2lc20982a. [DOI] [PubMed] [Google Scholar]
  • 18.Kléber AG, Rudy Y. Basic mechanisms of cardiac impulse propagation and associated arrhythmias. Physiological reviews. 2004;84:431–488. doi: 10.1152/physrev.00025.2003. [DOI] [PubMed] [Google Scholar]
  • 19.Pasqualini FS, Sheehy SP, Agarwal A, Aratyn-Schaus Y, Parker KK. Structural Phenotyping of Stem Cell-Derived Cardiomyocytes. Stem cell reports. 2015;4:340–347. doi: 10.1016/j.stemcr.2015.01.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Feinberg AW, et al. Controlling the contractile strength of engineered cardiac muscle by hierarchal tissue architecture. Biomaterials. 2012;33:5732–5741. doi: 10.1016/j.biomaterials.2012.04.043. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Sheehy SP, et al. Quality metrics for stem cell-derived cardiac myocytes. Stem cell reports. 2014;2:282–294. doi: 10.1016/j.stemcr.2014.01.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Oštádalová I, et al. Early postnatal development of contractile performance and responsiveness to Ca 2+, verapamil and ryanodine in the isolated rat heart. Journal of molecular and cellular cardiology. 1993;25:733–740. doi: 10.1006/jmcc.1993.1085. [DOI] [PubMed] [Google Scholar]
  • 23.Zhang D, et al. Tissue-engineered cardiac patch for advanced functional maturation of human ESC-derived cardiomyocytes. Biomaterials. 2013;34:5813–5820. doi: 10.1016/j.biomaterials.2013.04.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Zimmermann WH, et al. Three-dimensional engineered heart tissue from neonatal rat cardiac myocytes. Biotechnology and bioengineering. 2000;68:106–114. [PubMed] [Google Scholar]
  • 25.Reiser PJ, Portman MA, Ning XH, Moravec CS. Human cardiac myosin heavy chain isoforms in fetal and failing adult atria and ventricles. American Journal of Physiology-Heart and Circulatory Physiology. 2001;280:H1814–H1820. doi: 10.1152/ajpheart.2001.280.4.H1814. [DOI] [PubMed] [Google Scholar]
  • 26.Lundy SD, Zhu WZ, Regnier M, Laflamme MA. Structural and functional maturation of cardiomyocytes derived from human pluripotent stem cells. Stem cells and development. 2013;22:1991–2002. doi: 10.1089/scd.2012.0490. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Nunes SS, et al. Biowire: a platform for maturation of human pluripotent stem cell-derived cardiomyocytes. Nature methods. 2013;10:781–787. doi: 10.1038/nmeth.2524. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Xiao Y, et al. Microfabricated perfusable cardiac biowire: a platform that mimics native cardiac bundle. Lab on a Chip. 2014;14:869–882. doi: 10.1039/c3lc51123e. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Zimmermann WH, et al. Tissue engineering of a differentiated cardiac muscle construct. Circulation research. 2002;90:223–230. doi: 10.1161/hh0202.103644. [DOI] [PubMed] [Google Scholar]
  • 30.LeGrice IJ, et al. Laminar structure of the heart: ventricular myocyte arrangement and connective tissue architecture in the dog. American Journal of Physiology-Heart and Circulatory Physiology. 1995;269:H571–H582. doi: 10.1152/ajpheart.1995.269.2.H571. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1

RESOURCES