Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2017 Sep 1.
Published in final edited form as: Acad Radiol. 2016 Jun 14;23(9):1130–1136. doi: 10.1016/j.acra.2016.05.013

Instituting a Low Dose CT-Guided Lung Biopsy Protocol

Kimberly G Kallianos 1, Brett M Elicker 1, Travis S Henry 1, Karen G Ordovas 1, Janet Nguyen 1, David M Naeger 1
PMCID: PMC5344698  NIHMSID: NIHMS851358  PMID: 27317393

Abstract

Rationale and Objectives

We aimed to evaluate whether implementation of a low-dose CT-guided lung biopsy protocol, with the support of individual radiologists in the section, would lead to immediate and sustained decreases in radiation dose associated with CT-guided lung biopsies.

Materials and Methods

A low-dose CT-guided lung biopsy protocol was developed with modifications of kilovoltage peak (kVp), milliamperes (mA), and scan coverage. 413 CT guided lung biopsies were evaluated over a three-year period beginning in 2009, 175 performed with a standard protocol prior to development of a low dose protocol and 238 performed with a low dose protocol. The dose-length product (DLP) was recorded for each lung biopsy and retrospectively compared between the two protocols. Individual radiologist level DLPs were also compared before and after the protocol change.

Results

The mean biopsy dose decreased 64.4% with the low dose protocol (113.8 mGy-cm versus 319.7 mGy-cm; p <0.001). This decrease in radiation dose persisted throughout the entire 18 months evaluated following the protocol change. After the protocol change, each attending radiologist demonstrated a decrease in administered radiation dose. The diagnostic outcome rate and complication rate were unchanged over the interval.

Conclusion

Implementation of a low-dose CT-guided lung biopsy protocol resulted in an immediate reduction in patient radiation dose that was seen with all attending radiologists and persisted for at least 18 months. Such an intervention may be considered at other institutions wishing to reduce patient doses.

Keywords: Radiation dose, CT-guided lung biopsy, low-dose

Introduction

The last decade has brought increased awareness of the cancer-inducing risks of radiation sustained from medical imaging, and with it, a movement to decrease doses from modalities such as computed tomography (CT) [13]. CT-guided-biopsies can result in high doses by virtue of the repetitive imaging required with some techniques [4], however biopsies are prime candidates for dose reduction since image quality is less of a concern when used solely for needle guidance.

Previous work has shown the success of low-dose protocols in a variety of procedural settings, including CT-guided spine procedures [5, 6] and CT-guided lung biopsies [79]. However, the previous work with low-dose CT-guided lung biopsies was limited by small sample sizes, relatively short follow up periods to evaluate dose reductions, and limited analysis of the effect of a standardized low-dose protocol on multiple operators. This study examines the effect of a section-wide implementation of a low-dose CT-guided lung biopsy protocol in a high-volume referral center. This analysis includes an evaluation of the radiation doses over time and the radiation dose reductions achieved by individual attending radiologists.

Our hypotheses were that implementation of a low-dose CT-guided lung biopsy protocol with the support of individual radiologists in the section would lead to immediate and sustained decreases in radiation dose associated with CT-guided lung biopsies.

Materials and Methods

Low-dose CT-guided lung biopsy protocol

A low-dose CT-guided lung biopsy protocol was developed at our institution that included recommendations for multiple aspects of the biopsy including modifications of the kilovoltage peak (kVp), milliamperes (mA), and scan coverage. Development and implementation of the protocol was reached by consensus of all attending radiologists in the cardiothoracic section at our institution who performed biopsies during both the standard and low dose periods (n=6) [1012]. The protocol was divided into three phases of image acquisition for each biopsy: the planning phase, targeting phase, and post phase. The technical parameters of this consensus low dose protocol are provided in Table 1.

Table 1.

Low-Dose CT-guided lung biopsy protocol*

Phase Coverage Slice Thickness Voltage Tube Current
Planning Prior CTs should first be reviewed to select the optimal target. If the selected nodule is visible on the biopsy scout image then coverage approximately 0.5 cm above and below the lesion is acquired. If the lesion is not visible on the scout, a slightly larger range is acquired based on landmarks from prior CT. 2.5 mm 100 kVP for patient weight <140 lbs and arms up
120 kVp for patient weight >140 lbs and/or arms down
10 mA for scout
40 mA for patient weight <140 lbs and arms up
50 mA for patient weight >140 lbs and/or arms down
60 mA for patient weight >220 lbs and/or lesion at level of the liver
Targeting 9 images per iteration, then 7 images per iteration, then 5 images per iteration if possible 2.5 mm 100 kVP for patient weight <140lbs and arms up
120 kVp for patient weight >140 lbs and/or arms down
20 mA for patient weight <140 and arms up
25 mA for patient weight >140 and/or arms down
30 mA for patient weight >220 and/or lesion at level of the liver
Post Limited or whole chest 5 mm 100 kVP for patient weight <140lbs and arms up
120 kVp for patient weight >140 lbs and/or arms down
20 mA for patient weight <220 lbs
30 mA for patient weight >220 lbs
*

Typically using sharp reconstruction algorithm; kVP = peak kilovoltage, mA = milliamperes

All CT guided lung biopsies were performed on either a LightSpeed VCT or Discovery HD 750 (GE Healthcare, Milwaukee, Wisconsin). The same parameters were generally used on both scanners; adaptive statistical iterative reconstruction (ASIR) was not used. A standard coaxial technique was used for lung biopsies with a 19-gauge coaxial needle followed by 20–22 gauge FNA needles and 20-gauge core biopsy needles. The average number of samples per patient was 3.2, predominately fine needle aspirations (FNA) (average FNA samples 2.7 per patient versus average core biopsy samples 0.5 per patient). Feedback from an on-site pathologist was the main factor in determining the number and type of samples required. Biopsies were deemed diagnostic when the final pathologic analysis could establish a diagnosis of a specific benign entity or malignancy from the material obtained.

Retrospective data collection

Our institutional review board approved the creation of a Health Insurance Portability and Accountability Act (HIPAA) compliant database of CT-guided lung biopsies. All CT-guided lung biopsies performed during a 36-month period from September 2009 to August 2012 were included in the dataset. The dataset was divided into two groups: those undergoing biopsy in the 18 months prior to the protocol change (the standard-dose group), and those undergoing biopsy in the 18 months after the protocol change (the low dose group). Biopsies for which CT dose reports were not available were excluded. For all included biopsies, the following data were extracted:

  • Patient age

  • Attending radiologist

  • Lesion long axis size

  • Lesion depth from the pleura

  • Severity of emphysema (mild, moderate, severe)

  • Number of biopsy specimens obtained

  • Presence of biopsy complications: pneumothorax, chest tube placement, serious hemorrhage, and death

  • Radiation dose

  • Total coverage scanned, summed for all series (cm)

  • Total number of series performed

Radiation dose data was recorded in the form of dose-length product (DLP), which is the product of scan length in cm and the volume CT dose index (CTDIvol) for each individual scan event. Given each scan event can vary in the length covered and CTDIvol, DLP acts as a summary measurement. DLP is measured in miligray centimeter (mGy-cm) [13, 14].

For the purposes of analyzing radiation doses, the CT-guided lung biopsies were separated into three phases, planning, targeting, and post. The planning phase was defined as the initial CT acquisition(s) with a surface grid in place for lesion localization and selection of a needle path. The targeting phase was defined as the multiple acquisitions in which the needle was advanced into the lesion and biopsy samples were obtained. The post phase was defined as the CT acquisition after the needle was removed, which was used to evaluate for the presence of biopsy complications such as pneumothorax [15]. Radiation dose (DLP), total coverage (cm), and total number of series were recorded for each of the three phases. When multiple acquisitions were performed during a single phase, the scan coverage and DLP for each individual acquisition were summed.

Radiation doses (DLP) were also tabulated for each of the six attending radiologists who performed lung biopsies both before and after the protocol change. There was no systematic difference in the difficulty of biopsies performed by various attending radiologists because the attending assigned to biopsies each day was responsible for all scheduled procedures, regardless of difficulty, and the scheduling process did not take into account the attendings assigned to be on service on any given day. For the purposes of data presentation, radiologists were ordered based on the highest to lowest initial radiation doses.

The diagnosis of a post-procedure pneumothorax was made with a post phase CT, or via a standard post-biopsy chest radiograph which was obtained prior to discharge.

Statistical Analysis

Data analysis was performed using Stata version 14.0 (College Station, Texas). Summary statistics are reported as medians and interquartile ranges (IQR). Continuous parameters before and after the protocol change were compared using Mann–Whitney U tests due to non-normally distributed data. Chi-squared tests were used to test for equality between proportions. Comparison of continuous parameters over successive 6-month periods surrounding the protocol change was performed with the Kruskal-Wallis equality-of-populations rank test. Statistical significance was set at p < 0.05.

Results

A total of 446 biopsies were performed during the study period, and 33 (7.4%) were excluded because CT dose reports were not available, 27 from the standard dose group and 6 from the low dose group. 413 CT guided lung biopsies were included in our study, 175 performed with the standard protocol and 238 performed in the period after which the low dose protocol was instituted. Of note, overall imaging and procedure volume increased at our institution over the study period.

There was no significant difference in patient-specific parameters between the two groups including age, gender (Table 2A), as well as severity of emphysema (57.7% none, 17.7% mild, 17.1% moderate, 7,4% severe with the standard protocol versus 47.5% none, 18.5% mild, 25.2% moderate, and 8.8% severe with the low dose protocol, p=0.150). Lesion-specific parameters were also not significantly different between the standard and low dose biopsy protocols including lesion long axis size of 4.0 ± 2.5 cm versus 3.8 ± 2.7 cm (p=0.231) and lesion depth from the pleura of 1.8 ± 1.8 cm versus 2.1 ± 2.2 cm (p=0.185),

Table 2A.

Radiation dose before and after the CT-guided lung biopsy protocol change

Standard-Protocol (n= 175) Median [IQR] Low-Dose Protocol (n=238) Median [IQR] Percentage Change (%) P-value (Mann Whitney U)
Age (years) 67 [58–76] 66 [60–75] NA 0.803
Male (%) 114/ 175 (65.1%) 163/ 238 (68.5%) NA 0.475
Total DLP (mGy-cm) 319.7 [156.1–872.9] 113.8 [66.1–187.8] −64.4 <0.001
Total Coverage (cm) 492.5 [367.5–647.5] 467.5 [345–587.5] −5.1 0.149
Total Series # 9 [7–12] 12 [9–15] +33.3 <0.001
Planning
DLP (mGy-cm) 142.4 [51.7–317.9] 32.2 [15.5–72.9] −77.4 <0.001
Coverage (cm) 156.3 [112.5–225] 122.5 [90.5–175] −21.6 <0.001
Series # 1 [1–2] 2 [1–2] +100.0 0.001
Targeting
DLP (mGy-cm) 82.4 [34.0–282.2] 39.3 [23.5–60.6] −52.3 <0.001
Coverage (cm) 142.5 [75–225] 140 [90–221.5] −1.8 0.643
Series # 6 [4–9] 9 [6–12] +50.0 <0.001
Post
DLP (mGy-cm) 62.1 [23.4–208] 21.2 [12.4–42.7] −65.9 <0.001
Coverage (cm) 165.6 [82.5–260] 190 [100–260] +14.7 0.495
Series # 1 [1-1] 1 [1-1] 0.0 0.007*
*

Despite unchanged median and IQR, the distribution of biopsies with >1 # of series differed between the two groups

Total biopsy dose decreased an average of 64.4% under the low dose protocol compared to the standard protocol (DLP 113.8 mGy-cm versus 319.7 mGy-cm; p <0.001). Moreover, statistically significant dose reductions were observed in all three phases of the biopsy (77.4%, 52.3% and 65.8% for the planning, targeting and post-biopsy phases, respectively). Figure 1 depicts representative CT guided lung biopsy images using the standard versus low dose protocols.

Figure 1.

Figure 1

Representative biopsy images using the standard dose protocol (left, total DLP 568.12) and low dose protocol (right, total DLP 16.33).

The largest contributor to patient radiation dose using the standard-dose protocol was the planning phase with median DLP of 142.4 mGy-cm [IQR 51.7–917.9], while the largest contributor to patient radiation dose using the low dose protocol was the targeting phase, median DLP 39.3 mGy-cm [IQR 23.5–60.6].

The total number of series used to complete the biopsies significantly increased after the protocol change, median 12 [IQR 9–15] with the low-dose protocol compared to a median 9 [IQR 7–12] with the standard-dose protocol (p<0.001) (Table 2A). Total coverage in centimeters was not significantly different (p=0.149), however there was a significant decrease in coverage during the planning phase specifically (median 122.5 cm with the low-dose protocol versus a median 156.3 cm with the standard protocol) when analyzed alone, p<0.001.

To further delineate trends over time the information in Table 2A was subdivided into six 6-month periods (Table 2B, Figure 2). Radiation dose was relatively stable in the 18 months prior to the protocol change, and immediately decreased in the 6-month period afterward. This immediate decrease in radiation dose persisted throughout the subsequent two 6-month periods following the protocol change, thereby demonstrating 18 months total of reduced doses.

Table 2B.

Radiation dose over 6-month periods surrounding protocol change, Median [IQR].

12–18 months before (n=46) 6–12 months before (n=56) 6 months before (n=73) 6 months after (n=89) 6–12 months after (n=78) 12–18 months after (n=71) P-value (Kruskal-Wallis)
Total DLP (mGy-cm) 284.5 [165.7–859.2] 400.0 [193.3–1080.1] 309.2 [110.0–782.1] 102.3 [70.0–148.4] 120.8 [64.9–214.2] 117.8 [59.9–202.5] <0.001
Planning
DLP (mGy-cm) 118.6 [57.8–354.8] 178.4 [65.8–342.9] 144.8 [38.4–277.9] 31.7 [19.2–63.7] 34.2 [13.5–74.7] 31.6 [16.0–79.2] <0.001
Coverage (cm) 162.5 [132.5–240] 165.3 [110–223.4] 150 [100–202.5] 122.5 [90–170] 132.5 [92.5–192.5] 115 [90–162.5] <0.001
Series # 1 [1–2] 2 [1–2] 1 [1–2] 2 [1–2] 2 [1–2] 2 [1–2] 0.002
Targeting
DLP (mGy-cm) 93.2 [38.6–339.4] 67.0 [37.3–504.3] 83.6 [30.4–220.6] 34.7 [22.3–59.8] 40.5 [22.2–60.2] 42.3 [24.3–73.5] <0.001
Coverage (cm) 125 [75–245] 143.8 [72.5–232.5] 150 [90–210] 150 [90–225] 143.8 [100–221.5] 115 [70–200] 0.436
Series # 6 [4–10] 6 [4–9.5] 7 [5–9] 9 [6–10] 9 [6–12] 9 [6–13] <0.001
Post
DLP (mGy-cm) 67.3 [28.8–204.3] 104.5 [36.9–264.2] 37.3 [20.3–180.5] 20.8 [11.6–35.3] 29.0 [13.7–61.0] 19.7 [11.8–40.6] <0.001
Coverage (cm) 147.5 [50–245] 169 [95–270] 166.3 [90–277.5] 197.5 [98.1–260] 205 [115–272.8] 157.5 [80–262.5] 0.317
Series # 1 [1-1] 1 [1-1] 1 [1-1] 1 [1-1] 1 [1-1] 1 [1-1] 0.109

Figure 2.

Figure 2

Graphical representation of total DLP over 6-month periods surrounding the protocol change. Line represents median DLP with the upper and lower whisker lines representing the interquartile range. Kruskall Wallis test for differences p < 0.001.

Biopsy radiation doses using the standard protocol and low dose protocol were evaluated at the individual attending radiologist level. After the protocol change, each attending radiologist demonstrated a decrease in the radiation doses used (Figure 3). This decrease was statistically significant for three attendings (#1, 2, and 5). Before the protocol change there was large range of radiation doses used between attending radiologists; the difference from the lowest to highest median radiation doses was 387.9 mGy-cm. Afterward, radiation doses fell within a narrower range of 122.3 mGy-cm between the highest and lowest attending’s medians, a 68.5% decrease in the range of medians.

Figure 3.

Figure 3

Total DLP by attending radiologists before and after the protocol change. Bars represent median DLP, with the upper and lower whisker lines representing the interquartile range. Statistically significant difference at the p=0.05 level is denoted by an asterisk. The attending radiologists are presented in order from the highest to lowest median radiation doses used before the intervention.

After the protocol change, there was a non-statistically significant trend toward a decrease in pneumothorax rate: 64/238 (26.9%) using the low-dose protocol compared to 57/175 (32.6%) using the standard-dose protocol (p=0.210). There was also a trend toward decrease in the chest tube placement rate: 16/238 (6.7%) using the low-dose protocol compared to 21/175 (12.0%) using the standard-dose protocol compared (p=0.064). There was no significant difference in rates of serious hemorrhage between the two protocols, 1/175 versus 1/238, p = 0.827. There were no patient deaths with either protocol.

A diagnostic result (e.g. a pathologic diagnosis of cancer, infection, or a specific benign pathology) was achieved in 182/238 (76.5%) with the low dose protocol and 137/175 (78.3%) with the standard dose, p=0.664. There was a statically significant increase in number of biopsy samples obtained with the low dose protocol (3.5 ± 2.0 samples) compared to the standard dose protocol (2.8 ± 1.9), p<0.001.

Discussion

We instituted a low-dose CT guided biopsy protocol in our busy lung biopsy service that provided general guidelines for dose reduction techniques. The resultant “real world” effect of such a policy was a decrease in radiation dose overall and for each individual attending, with no change in biopsy success rate. This effect was sustained for the 18 months evaluated after the protocol change.

Our low-dose CT-guided lung biopsy protocol included recommendations for multiple aspects of the biopsy including modifications of the kilovoltage peak (kVp), milliamperes (mA) and scan coverage [1, 1012]. The recommendations to reduce CT coverage did not result in a statistically significant change in overall coverage, however there was a mild, statistically significant reduction in coverage during the planning phase when analyzed alone. Interestingly, there was a significant increase in the number of series performed during biopsies following the protocol change. The cause of the increase in series number was not directly investigated; however smaller coverage field may have been a contributing factor. Given that coverage and the series number did not dramatically decrease, the majority of the decrease in radiation dose achieved by our low-dose CT-guided lung biopsy protocol was the result of decreased exposures parameters (lower mA and/or kVP).

While radiation dose reduction was achieved in all phases of the biopsy, the greatest decrease in dose was seen in the planning phase. As a result, the targeting phase imparted the largest median radiation dose after the protocol change.

Prior studies of low dose lung biopsies have focused on technical success and complication rates following low-dose CT-guided biopsies. Similar to these studies, we did not observe a significant change in the percentage of biopsies achieving a diagnostic result following implementation of a low dose protocol. We did observe a lower overall diagnostic biopsy rate, both before and after the protocol change, compared to the range reported in prior studies of 92% to 95.8%, which may be related to our classifying nonspecific pathologies such as inflammation as “non-diagnostic” [16]. Additionally, our group is asked to attempt biopsies on small lesions (<1cm) in cases where the possible clinical benefit outweighs the increased risk of a non-diagnostic result. The complication rates reported in the literature ranging from 10% to 32% were similar to the complication rates observed in our low dose group [79], and complication rates did not increase at our institution over this interval (indeed, slightly lower complication rates were observed, likely unrelated to the change in biopsy protocol). Our protocol allowed for flexibility in the CT parameters depending upon patient specific factors and the preferences of the attending radiologists. Despite this flexibility, our median radiation dose of 113.8 mGy-cm falls within the range reported in the literature (10.98 mGy-cm to 133.3 mGy-cm) with our overall dose reduction of 64.4% also within the range of dose reductions reported in the literature of similar interventions (57.5 to 95%)[79].

To our knowledge there has been no previous investigation of low-dose CT-guided lung biopsies that included an assessment of long-term reductions in radiation dose. Our study demonstrated a dramatic effect that was seen within the first 6 months and that persisted for the full 18 months evaluated after the protocol change. This confirms that a low-dose CT-guided biopsy protocol can be successfully implemented with a sustained reduction in the doses used. In addition, our study is the first to perform analysis of CT-guided lung biopsy dose reduction at the individual attending radiologist level. In our section, all radiologists had a reduction in their radiation doses following the protocol change, 3 of which were statistically significant. Despite individual differences in biopsy technique, our data suggests interventions such as this can result in broad changes, standardizing the radiation dose of CT-guided lung biopsies and reducing variability between radiologists. Finally, while other studies have also evaluated overall radiation dose reduction, we additionally assessed which phases were most impacted by dose reduction techniques, specifically the planning phase.

This study has several limitations. First, our biopsies were performed by a single group of attending radiologists at one institution. Additionally, while we followed radiation dose for 18 months after the protocol change, further long-term follow-up could be performed to confirm continued success. This study employed a retrospective design, therefore prospective data gathering and randomization were not possible. We were not able to collect data on procedure duration as a part of this analysis, however we included number of biopsy samples collected as a proxy for procedure length. Of note, there was a statistically significant increase in number of samples obtained under the low dose protocol compared to the standard protocol, a large component of which is likely related to the increased utilization of molecular testing which requires additional samples. We suspect that this effect is not related to the radiation dose protocol change, however the lack of data on true procedure duration limits our ability to fully explore this finding. While specific care was taken to maintain the quality of the post biopsy scans in order to detect complications, the slightly lower pneumothorax detection rate on the post scans could be due to a lower sensitivity. Post procedure radiographs, which were used in both protocols, should have ensured similar sensitivities for clinically significant pneumothoraces.

Conclusions

In summary, implementation of a low-dose CT-guided lung biopsy protocol at our institution resulted in an immediate reduction in patient radiation dose that was seen with all attending radiologists and persisted for at least 18 months. The results of this study suggest that low-dose CT-guided lung biopsy protocols can be successfully implemented in real-world radiology practices. Awareness of the success of such an intervention is important for radiologists seeking to reduce patient radiation dose associated with their CT-guided lung biopsies.

Acknowledgments

Grants/funding: Dr. Kallianos is supported by the National Institutes of Health T32 Training Grant.

Abbreviations

CT

computed tomography

cm

centimeter

CTDIvol

volume CT dose index

DLP

dose length product

FNA

fine needle aspiration

HIPAA

Health Insurance Portability and Accountability Act

IQR

interquartile range

kVp

kilovoltage peak

mA

milliamperes

mGy

milliGray

Footnotes

Previous submissions/reports of similar type information: None.

Conflicts of interest: None

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Kalra MK, Sodickson AD, Mayo-Smith WW. CT Radiation: Key Concepts for Gentle and Wise Use. Radiographics. 2015;35:1706–1721. doi: 10.1148/rg.2015150118. [DOI] [PubMed] [Google Scholar]
  • 2.Smith-Bindman R, Lipson J, Marcus R, Kim KP, Mahesh M, Gould R, Berrington de Gonzalez A, Miglioretti DL. Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med. 2009;169:2078–2086. doi: 10.1001/archinternmed.2009.427. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Brenner DJ, Hall EJ. Computed tomography--an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–2284. doi: 10.1056/NEJMra072149. [DOI] [PubMed] [Google Scholar]
  • 4.Prosch H, Stadler A, Schilling M, Burklin S, Eisenhuber E, Schober E, Mostbeck G. CT fluoroscopy-guided vs. multislice CT biopsy mode-guided lung biopsies: accuracy, complications and radiation dose. Eur J Radiol. 2012;81:1029–1033. doi: 10.1016/j.ejrad.2011.01.064. [DOI] [PubMed] [Google Scholar]
  • 5.Shepherd TM, Hess CP, Chin CT, Gould R, Dillon WP. Reducing patient radiation dose during CT-guided procedures: demonstration in spinal injections for pain. AJNR Am J Neuroradiol. 2011;32:1776–1782. doi: 10.3174/ajnr.A2634. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Shpilberg KA, Delman BN, Tanenbaum LN, Esses SJ, Subramaniam R, Doshi AH. Radiation dose reduction in CT-guided spine biopsies does not reduce diagnostic yield. AJNR Am J Neuroradiol. 2014;35:2243–2247. doi: 10.3174/ajnr.A4053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Smith JC, Jin DH, Watkins GE, Miller TR, Karst JG, Oyoyo UE. Ultra-low-dose protocol for CT-guided lung biopsies. J Vasc Interv Radiol. 2011;22:431–436. doi: 10.1016/j.jvir.2011.01.431. [DOI] [PubMed] [Google Scholar]
  • 8.Adiga S, Athreya S. Safety, efficacy, and feasibility of an ultra-low dose radiation protocol for CT-guided percutaneous needle biopsy of pulmonary lesions: initial experience. Clin Radiol. 2014;69:709–714. doi: 10.1016/j.crad.2014.02.006. [DOI] [PubMed] [Google Scholar]
  • 9.Meng XX, Kuai XP, Dong WH, Jia NY, Liu SY, Xiao XS. Comparison of lung lesion biopsies between low-dose CT-guided and conventional CT-guided techniques. Acta Radiol. 2013;54:909–915. doi: 10.1177/0284185113485937. [DOI] [PubMed] [Google Scholar]
  • 10.Kalra MK, Maher MM, Toth TL, Hamberg LM, Blake MA, Shepard JA, Saini S. Strategies for CT radiation dose optimization. Radiology. 2004;230:619–628. doi: 10.1148/radiol.2303021726. [DOI] [PubMed] [Google Scholar]
  • 11.Litmanovich DE, Tack DM, Shahrzad M, Bankier AA. Dose Reduction in Cardiothoracic CT: Review of Currently Available Methods. RadioGraphics. 2014;34:1469–1489. doi: 10.1148/rg.346140084. [DOI] [PubMed] [Google Scholar]
  • 12.Kubo T, Ohno Y, Kauczor HU, Hatabu H. Radiation dose reduction in chest CT--review of available options. Eur J Radiol. 2014;83:1953–1961. doi: 10.1016/j.ejrad.2014.06.033. [DOI] [PubMed] [Google Scholar]
  • 13.McNitt-Gray MF. AAPM/RSNA Physics Tutorial for Residents: Topics in CT. RadioGraphics. 2002;22:1541–1553. doi: 10.1148/rg.226025128. [DOI] [PubMed] [Google Scholar]
  • 14.Huda W, Mettler FA. Volume CT Dose Index and Dose-Length Product Displayed during CT: What Good Are They? Radiology. 2011;258:236–242. doi: 10.1148/radiol.10100297. [DOI] [PubMed] [Google Scholar]
  • 15.Wu CC, Maher MM, Shepard JA. Complications of CT-guided percutaneous needle biopsy of the chest: prevention and management. AJR Am J Roentgenol. 2011;196:W678–682. doi: 10.2214/AJR.10.4659. [DOI] [PubMed] [Google Scholar]
  • 16.Haas BM, Elicker BM, Nguyen J, Ordovas KG, Jones KD, Henry TS, Naeger DM. Nondiagnostic Computed Tomography-guided Percutaneous Lung Biopsies Are More Likely When Infection Is Suspected. J Thorac Imaging. 2016;31:151–155. doi: 10.1097/RTI.0000000000000207. [DOI] [PubMed] [Google Scholar]

RESOURCES