Diffusion-weighted whole-body imaging with background body signal suppression/T2 image fusion for the diagnosis of colorectal polyp and cancer

Minoru Tomizawa 1,, Fuminobu Shinozaki 2, Yoshitaka Uchida 3, Katsuhiro Uchiyama 3, Kazunori Fugo 4, Takafumi Sunaoshi 2, Aika Ozaki 2, Eriko Sugiyama 2, Akira Baba 2, Daisuke Kano 2, Misaki Shite 2, Ryota Haga 2, Yoshiya Fukamizu 2, Satoshi Kagayama 2, Rumiko Hasegawa 5, Yoshinori Shirai 5, Yasufumi Motoyoshi 6, Takao Sugiyama 7, Shigenori Yamamoto 8, Takashi Kishimoto 4, Naoki Ishige 9
PMCID: PMC5348677  PMID: 28352344

Abstract

Diffusion-weighted whole-body imaging with background body signal suppression/T2 image fusion (DWIBS/T2) is useful for the diagnosis of cancer as it presents a clear contrast between cancerous and non-cancerous tissue. The present study investigated the limitations and advantages of DWIBS/T2 with regards to the diagnosis of colorectal polyp (CP) or cancer (CRC). The current study included patients diagnosed with CP or CRC following colonoscopy, who were subjected to DWIBS/T2 between July 2012 and March 2015. Patient records were analyzed retrospectively. Patients were subjected to DWIBS/T2 when they presented with abdominal cancers or inflammation. Colonoscopy was performed as part of screening, or if patients had suspected colon cancer or inflammatory bowel disease. A total of 8 male and 7 female patients were enrolled in the present study. All patients, with the exception of one who had been diagnosed with CRC following colonoscopy, had positive results and all patients diagnosed with CP following a colonoscopy, with the exception of one, had negative results on DWIBS/T2. Thus, CRC was detected by DWIBS/T2, while CP was not (P=0.0028). The diameter of CRC lesions was significantly larger than that of CP (P<0.0001) and that of lesions positive on DWIBS/T2 was significantly larger than that of negative lesions (P=0.0004). The depth of invasion tended to be greater for lesions positive on DWIBS/T2 compared with that of negative ones. This indicated that DWIBS/T2 may be suitable for the detection of CRC but not for detection of CP. The results of DWIBS/T2 may also be affected by lesion diameter and depth of invasion.

Keywords: colorectal cancer, colorectal polyp, depth of invasion

Introduction

Colorectal cancer (CRC) develops from colorectal polyps (CP) as a consequence of the adenoma-carcinoma sequence (1). CRC is frequently encountered in clinical settings and has a poor prognosis (2). To improve the prognosis of patients with CRC, a prompt and accurate diagnosis is required. Preliminary screening for CRC is performed by fecal occult blood testing and colonoscopy is performed for the final diagnosis (3). However, despite being preferable over all other available diagnostic modalities in terms of practicability and affordability, fecal occult blood testing is not a reliable technique (4). Therefore, colonoscopy is currently the gold standard for the diagnosis of CRC (5,6).

Diffusion-weighted whole-body imaging with background body signal suppression (DWIBS) images are obtained using multiple-signal averaging, pre-pulse fat suppression and heavy diffusion weighting during free breathing (7). DWIBS is based on DWI that utilizes the random movement of water at the molecular level, known as Brownian motion (8,9). This imaging technique provides a strong contrast between cancerous tissue and the surrounding non-cancerous tissue, which is useful in the detection and staging of cancer (10). One major limitation of DWIBS is that anatomical assessment may be difficult (11,12). To overcome this, fusion images of DWIBS and T2-weighed images (T2WI) are created by overlapping DWIBS images with T2WI (DWIBS/T2) images using an Extended MR Workspace workstation (10,13,14). DWIBS/T2 therefore clearly integrates functional and anatomical information.

The current study retrospectively analyzed DIWBS/T2 images from patients with CRC and CP to evaluate the advantages and limitations of this technique for the diagnosis of CRC and CP.

Materials and methods

Patients

Patient records from the Department of Gastroenterology at the National Hospital Organization Shimoshizu Hospital, including imaging results of patients from July 2012 until March 2015, were retrospectively analyzed. Each patient with suspected abdominal cancers or inflammation was subjected to DWIBS/T2 once for further examination, using magnetic resonance imaging (MRI), between July 2012 and March 2015. Colonoscopy was performed as part of screening or when patients had suspected colon cancer or inflammatory bowel disease. The enrolled patients were those who were subjected to colonoscopy and DWIBS/T2, and whose pathological diagnosis was available via biopsy, endoscopic mucosal resection, or surgical specimens. Patients were excluded from the present study when colonoscopy, DWIBS/T2, and pathological diagnosis were not available. A total of 15 patients were enrolled, comprising of eight male (mean age, 72.4±7.6 years) and seven female (mean age, 76.1±3.4 years) patients. Positron emission tomography/computed tomography (PET/CT) was performed for one patient with a colon polyp that was diagnosed as CRC via endoscopic mucosal resection. The present study was approved by the ethics committee of the National Hospital Organization Shimoshizu Hospital (Yotsukaido, Japan). The study was not considered a clinical trial since the procedures involved were performed as a part of routine clinical practice. Written, informed consent was obtained from all patients.

Diameter and depth of invasion

The diameter of the CP was measured using colonoscopy and the diameter of the CRC lesion was measured from surgical specimens using a ruler. Colonoscopy was performed using a CF-Q260A/I or PCF-Q260AI colonoscope (Olympus Corp., Tokyo, Japan). Therefore, the diameter of CRC lesions was only determined for patients who underwent surgery (n=5). Depth of invasion (mm) of the surgical specimen was determined by macroscopic examination by two pathologists.

MRI

All MRI studies were performed using a 1.5 Tesla scanner (Achieva, complementary software version 3.2.2; Philips Medical Systems, B.V., Eindhoven, The Netherlands). T1WI, T2WI and DWI were obtained with pulse sequences, as depicted in Table I. DWIBS/T2 images were constructed with an Extended MR WorkSpace (version 2.6.3.4; Philips Medical Systems). The DWI gradients were applied along the X, Y, and Z axes prior to and following a 180° inversion pre-pulse in order to obtain fat-saturated, isotropic images with DWI sensitivity. The following parameters were used for a single stack: b-value 0 and 800 sec/mm2; repetition time, echo time and inversion recovery, 6,960.79 and 150 msec, respectively; acquisition matrix, 176×115 and reconstruction matrix, 256; field of view (FOV) right/left: 530 mm, anterior/posterior: 349 mm and feet/head: 226 mm; slice thickness, 6 mm; size of reconstructed voxel, 2.07×2.08×6 mm3; four averages. One radiologist and one gastroenterologist analyzed the DWIBS/T2 images. To rule out T2 shine-through and distinguish malignant lesions from non-malignant causes of restricted diffusion, a positive apparent diffusion co-efficient (ADC) map was generated by ADC reduction (15).

Table I.

Pulse sequences used in the present study.

Parameter T1-weighted image T2-weighted image DWI (DWIBS)
Echo GRE Single-shot SE EPI SE
TR, msec Shortest 1,000 11,250
TE, msec First: 2.3 (out-phase, second: 4.6 (in-phase) 90 83
Flip angle, ° 75 90 90
NSA 1 1 4
Slice thickness, mm 8 8 5
Slice gap 1 1 0
Fat saturation None None SPAIR
Phase encoding direction Posterior-anterior Posterior-anterior Posterior-anterior

DWIBS, diffusion-weighted whole body imaging with background body signal suppression; GRE, gradient echo; SE, spin echo; EPI, echo planar imaging; SPAIR, spectral attenuated inversion recovery; NSA, number of signal averages; TE, echo time; TR, repetition time.

PET/CT imaging

PET/CT was performed at the Diagnostic PET Imaging Center, Department of Radiology, Sannoh Medical Center (Chiba, Japan) (16). Patients fasted overnight or for at least 6 h prior to injection of 4.0 MBq/kg fluorodeoxyglucose (18F-FDG). Images were obtained 60 min following the injection. Patients lay down in the supine position during the 60-min uptake period in order to minimize non-specific FDG uptake into the muscles. An integrated PET/CT system (Discovery ST; GE Healthcare Bio-Sciences, Pittsburg, PA, USA) was used for image acquisition. CT scanning was performed with 120 kV, 25 mA, a pitch of 1.75, a section thickness of 3.3 mm, a FOV of 50 cm and a matrix size of 512×512. Immediately following the unenhanced CT scan, a PET scan was performed with a section thickness of 3.3 mm, a matrix size of 128×128 and an acquisition time of 2.5 min. PET data sets were reconstructed iteratively using an ordered subsets expectation maximization algorithm and segmented attenuation correction (three iterations, 15 subsets) and the CT data. The standardized uptake value (SUV) was calculated using the following formula: SUV (g/ml) = regional radioactivity concentration (Bq/ml) / [injected dose (Bq) / body weight (g)] (17).

Colonoscopy

Colonoscopy was performed for patients with anemia, abdominal symptoms, a positive fecal occult blood test for screening or investigation of anemia, or for screening purposes. The devices used were CF-Q260A/I and PCF-Q260AL/I. Patients with CP were subjected to endoscopic mucosal resection as CP was considered to be a precancerous lesion. Submucosa was expanded via injection with physiological saline, and the mucosa was removed using a snare (SD-221U-25; Olympus Corp.) with a cushion of submucosa.

Statistical analysis

Values are expressed as the mean ± standard deviation unless otherwise specified. The χ2 test was applied to evaluate the correlation between the diagnosis of CP or CRC obtained by colonoscopy and the results of DWIBS/T2. One-way analysis of variance was applied to the diameters of CP or CRC and the diameter of positive or negative results on DWIBS/T2. P<0.05 was considered to indicate a statistically significant difference. JMP 10.0.2 (SAS Institute, Inc., Cary, NC, USA) was used for statistical analyses.

Results

Clinicopathological characteristics

Patient clinicopathological characteristics are listed in Table II. A total of 7 patients were diagnosed with CRC and 8 patients were diagnosed with CP. All patients diagnosed with CRC following colonoscopy, with the exception of one, had positive results on DWIBS/T2. CRC presented with a clear signal in the horizontal (Fig. 1A) and coronal (Fig. 1B) sections. Surgical treatment was not performed for patients seven and eight, since they had advanced CRC. They received conservative treatment and the lesion diameter and depth of invasion were therefore not investigated for these two patients. All patients diagnosed with CP following colonoscopy, with the exception of one had negative results following DWIBS/T2: A specimen obtained from patient one by endoscopic mucosal resection revealed a well-differentiated adenocarcinoma (Table II).

Table II.

Clinicopathological characteristics of patients.

Patient no. Gender Age, years Diagnosis Pathological diagnosis DWIBS/T2 findings Diameter, mm Depth of invasion Depth of invasion, mm
  1 M 67 CP Well (+) 15 M 3
  2 M 73 CRC Well (+) 60 SS 12
  3 F 78 CRC Well (+) 47 SS 12
  4 M 74 CRC Mod (−) 30 SS 3
  5 M 73 CRC Well (+) 63 SS 13
  6 F 73 CRC Well (+) 55 SS 10
  7 M 85 CRC Well (+) N.A. N.A. N.A.
  8 F 70 CRC Well (+) N.A. N.A. N.A.
  9 M 80 CP Adenoma (−) 4 N.A. N.A.
10 F 76 CP Adenoma (−) 12 N.A. N.A.
11 F 79 CP Adenoma (−) 5 N.A. N.A.
12 M 63 CP Adenoma (−) 10 N.A. N.A.
13 M 64 CP Adenoma (−) 5 N.A. N.A.
14 F 78 CP Adenoma (−) 5 N.A. N.A.
15 F 79 CP Adenoma (−) 7 N.A. N.A.

M, male; F, female; CP, colorectal polyp; CRC, colorectal cancer; DWIBS/T2, diffusion-weighted whole body imaging with background body signal suppression; Well, well-differentiated adenocarcinoma; Mod, moderately differentiated adenocarcinoma; (+), positive result; (−), negative result; M, mucous membrane; SS, subserosa; N.A., not analyzed.

Figure 1.

Figure 1.

DWIBS/T2 images of an ascending colon cancer. DWIBS/T2 was performed for a 73 year-old man with ascending colon cancer (patient two Table II). The (A) horizontal and (B) coronal section present positive results (indicated by arrows) in the ascending colon. DWIBS/T2, diffusion-weighted whole body imaging with background body signal suppression/T2 image fusion.

Positive DWIBS/T2 is associated with CRC

Results of the χ2 test revealed a significant association between the diagnosis of CP or CRC following colonoscopy and the results on DWIBS/T2 (P=0.0028; Table III). Seven patients diagnosed with CP had negative results, while six diagnosed with CRC had positive results following DWIBS/T2.

Table III.

Comparison between diagnosis of CP or CRC upon colonoscopy and results of DWIBS/T2.

DWIBS/T2 findings

Diagnosis on colonoscopy Negative (n) Positive (n) Total
CP   7   1   8
CRC   1   6   7
Total   8   7 15

A χ2 test was performed, indicating a significant association between the diagnosis of CP or CRC following colonoscopy and the results on DWIBS/T2 (P=0.0028). DWIBS/T2, diffusion-weighted whole body imaging with background body signal suppression/T2 image fusion; CP, colorectal polyp; CRC, colorectal cancer.

Greater lesion diameter is observed in CRC

To analyze the correlation between lesion diameter and detection of CP or CRC by DWIBS/T2, a scatter plot of lesion diameter in CRC and CP was generated (Fig. 2). The lesion diameter for CP (7.9±4.0 mm) was significantly smaller than that for CRC (51.0±13.2 mm; P<0.0001; Fig. 2). It was therefore implied that the diameter of the lesion may affect the detectability of CP or CRC when using DWIBS/T2.

Figure 2.

Figure 2.

Scatter plot of lesion diameter (mm) of CP or CRC. The lesion diameter was plotted against CP or CRC. The mean lesion diameter in the CP group (7.9±4.0 mm) was significantly smaller than that in the CRC group (51.0±13.2 mm, P<0.0001). CP, colorectal polyp; CRC, colorectal cancer. *P<0.0001.

Positive DWIBS/T2 is associated with lesion diameter

To determine the factors affecting the results of DWIBS/T2, scatter plots of lesion diameter (Fig. 3A) and depth of invasion (Fig. 3B) were generated for lesions which were positive and negative on DWIBS/T2. The mean diameter of lesions with negative results (9.8±8.6 mm) was significantly smaller than that of lesions with positive results on DWIBS/T2 (48.0±19.4 mm; P=0.0004). Furthermore, the depth of tumor invasion of the lesions negatively diagnosed on DWIBS/T2 was at the lower margin of the range of the positively diagnosed lesion however, the statistical significance could not be analyzed as there was only one patient with a negatively diagnosed lesion (patient fou Table II).

Figure 3.

Figure 3.

Scatter plot of diameter (mm) and depth of invasion (mm) of colorectal cancer. (A) Diameter (mm) and (B) depth of invasion (mm) were plotted against negative or positive results following diagnosis by DWIBS/T2. The mean diameter of lesions with negative results (9.8±8.6 mm) was significantly smaller than that of lesions with positive results on DWIBS/T2 (48.0±19.4 mm; P=0.0004). DWIBS/T2, diffusion-weighted whole body imaging with background body signal suppression/T2 image fusion. *P=0.0004.

Positive CP diagnosis with DWIBS/T2 and PET/CT

Patient one, who had been diagnosed with CP by colonoscopy (Fig. 4A and B) was positive on DWIBS/T2 (Fig. 4C). The PET/CT results for patient one were also indicative of CP (Fig. 4D). The patient's SUV of 18F-FDG was 2.5, indicating that uptake of 18F-FDG in their CP was 2.5 times greater than the expected uptake in the whole body if 18F-FDG was distributed equally. DWI (Fig. 4E) and PET findings (Fig. 4F) were similar.

Figure 4.

Figure 4.

DWIBS/T2 and PET/CT images of a patient with CP. A 67 year-old man presented abdominal pain (patient one; Table II) and was subjected to DWIBS/T2 to investigate the presence of cancer in the abdominal cavity. Colonoscopy was performed to examine the colon and rectum. (A) CP was located, (B) endoscopic mucosal resection was performed and tissue was microscopically examined (scale bar, 5 mm). The horizontal section on (C) DWIBS/T2 and (D) PET/CT revealed positive results, indicated by arrows, in the descending colon. Coronal section on (E) DWI and (F) PET exhibited positive results, indicated by arrows. CP, colorectal polyp; DWIBS/T2, diffusion-weighted whole body imaging with background body signal suppression/T2 image fusion; PET/CT, positron emission tomography/computed tomography.

Discussion

Conventional MRI has limitations with regard to tumor and node (N) staging, while DWI adds value, allowing for N staging (18). The ADC value is useful for distinguishing CRC from inflammatory bowel disease (19). Leufkens et al (20) performed DWI on patients who were scheduled for colonoscopy to evaluate the diagnostic accuracy of DWI for CP or CRC. It was demonstrated that for the detection of CRC or CP with a lesion diameter of >6 mm, the sensitivity and specificity of DWI were 80.0 and 72.7%, respectively (20). However, in the present study, DWIBS/T2 was able to distinguish between CRC and CP due to selectively detecting CRC. Furthermore, the mean lesion diameter in the CRC group was significantly larger than that in the CP group. Therefore, it was concluded that the lesion diameter may affect the results of DWIBS/T2. It has been indicated that the lesion diameter affects the results of ultrasonography for the diagnosis of CP and CRC (21).

It was also observed that the mean depth of invasion was greater in cases of CRC which were positively diagnosed on DWIBS/T2, compared with the depth of invasion in the one case that was negatively diagnosed; however, statistical significance could not be determined. It is therefore hypothesized that the pixel number may be a major factor affecting the diagnostic accuracy of DWIBS/T2.

For patient one, the results of DWIBS/T2, PET/CT, DWI and PET were similar. This indicated that DWIBS/T2 and PET/CT may be useful for detecting CP with well-differentiated adenocarcinoma, which is eligible for endoscopic mucosal resection. If the sensitivity of DWIBS/T2 for the diagnosis of large CP is proven to be superior to that of PET/CT, it may replace PET/CT as a diagnostic tool.

One limitation of the present study was the small sample siz therefore, further studies with larger samples are required to assess the correlation between the results of DWIBS/T2 and the depth of invasion.

In conclusion, lesion diameter and depth of invasion may affect the detectability of DWIBS/T2.

References

  • 1.Al-Sohaily S, Biankin A, Leong R, Kohonen-Corish M, Warusavitarne J. Molecular pathways in colorectal cancer. J Gastroenterol Hepatol. 2012;27:1423–1431. doi: 10.1111/j.1440-1746.2012.07200.x. [DOI] [PubMed] [Google Scholar]
  • 2.Brenner H, Kloor M, Pox CP. Colorectal cancer. Lancet. 2014;383:1490–1502. doi: 10.1016/S0140-6736(13)61649-9. [DOI] [PubMed] [Google Scholar]
  • 3.Stracci F, Zorzi M, Grazzini G. Colorectal cancer screening: Tests, strategies, and perspectives. Front Public Health. 2014;2:210. doi: 10.3389/fpubh.2014.00210. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Benton SC, Seaman HE, Halloran SP. Faecal occult blood testing for colorectal cancer screening: The past or the future. Curr Gastroenterol Rep. 2015;17:428. doi: 10.1007/s11894-015-0428-2. [DOI] [PubMed] [Google Scholar]
  • 5.Manser CN, Bachmann LM, Brunner J, Hunold F, Bauerfeind P, Marbet UA. Colonoscopy screening markedly reduces the occurrence of colon carcinomas and carcinoma-related death: A closed cohort study. Gastrointest Endosc. 2012;76:110–117. doi: 10.1016/j.gie.2012.02.040. [DOI] [PubMed] [Google Scholar]
  • 6.Lieberman DA, Holub JL, Morris CD, Logan J, Williams JL, Carney P. Low rate of large polyps (>9 mm) within 10 years after an adequate baseline colonoscopy with no polyps. Gastroenterology. 2014;147:343–350. doi: 10.1053/j.gastro.2014.04.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Takahara T, Imai Y, Yamashita T, Yasuda S, Nasu S, Van Cauteren M. Diffusion weighted whole body imaging with background body signal suppression (DWIBS): Technical improvement using free breathing, STIR and high resolution 3D display. Radiat Med. 2004;22:275–282. [PubMed] [Google Scholar]
  • 8.Sehy JV, Ackerman JJ, Neil JJ. Apparent diffusion of water, ions, and small molecules in the Xenopus oocyte is consistent with Brownian displacement. Magn Reson Med. 2002;48:42–51. doi: 10.1002/mrm.10181. [DOI] [PubMed] [Google Scholar]
  • 9.Koike N, Cho A, Nasu K, Seto K, Nagaya S, Ohshima Y, Ohkohchi N. Role of diffusion-weighted magnetic resonance imaging in the differential diagnosis of focal hepatic lesions. World J Gastroenterol. 2009;15:5805–5812. doi: 10.3748/wjg.15.5805. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Kwee TC, Takahara T, Ochiai R, Nievelstein RA, Luijten PR. Diffusion-weighted whole-body imaging with background body signal suppression (DWIBS): Features and potential applications in oncology. Eur Radiol. 2008;18:1937–1952. doi: 10.1007/s00330-008-0968-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Ohno Y, Koyama H, Onishi Y, Takenaka D, Nogami M, Yoshikawa T, Matsumoto S, Kotani Y, Sugimura K. Non-small cell lung cancer: Whole-body MR examination for M-stage assessment-utility for whole-body diffusion-weighted imaging compared with integrated FDG PET/CT. Radiology. 2008;248:643–654. doi: 10.1148/radiol.2482072039. [DOI] [PubMed] [Google Scholar]
  • 12.Fischer MA, Nanz D, Hany T, Reiner CS, Stolzmann P, Donati OF, Breitenstein S, Schneider P, Weishaupt D, von Schulthess GK, Scheffel H. Diagnostic accuracy of whole-body MRI/DWI image fusion for detection of malignant tumours: A comparison with PET/CT. Eur Radiol. 2011;21:246–255. doi: 10.1007/s00330-010-1929-x. [DOI] [PubMed] [Google Scholar]
  • 13.Sommer G, Wiese M, Winter L, Lenz C, Klarhöfer M, Forrer F, Lardinois D, Bremerich J. Preoperative staging of non-small-cell lung cancer: Comparison of whole-body diffusion-weighted magnetic resonance imaging and 18F-fluorodeoxyglucose-positron emission tomography/computed tomography. Eur Radiol. 2012;22:2859–2867. doi: 10.1007/s00330-012-2542-y. [DOI] [PubMed] [Google Scholar]
  • 14.Nechifor-Boilă IA, Bancu S, Buruian M, Charlot M, Decaussin-Petrucci M, Krauth JS, Nechifor-Boilă AC, Borda A. Diffusion weighted imaging with background body signal suppression/T2 image fusion in magnetic resonance mammography for breast cancer diagnosis. Chirurgia (Bucur) 2013;108:199–205. [PubMed] [Google Scholar]
  • 15.Wang Y, Miller FH, Chen ZE, Merrick L, Mortele KJ, Hoff FL, Hammond NA, Yaghmai V, Nikolaidis P. Diffusion-weighted MR imaging of solid and cystic lesions of the pancreas. Radiographics. 2011;31:E47–E64. doi: 10.1148/rg.313105174. [DOI] [PubMed] [Google Scholar]
  • 16.Tanaka S, Ikeda K, Uchiyama K, Iwamoto T, Sanayama Y, Okubo A, Nakagomi D, Takahashi K, Yokota M, Suto A, et al. [18F]FDG uptake in proximal muscles assessed by PET/CT reflects both global and local muscular inflammation and provides useful information in the management of patients with polymyositis/dermatomyositis. Rheumatology (Oxford) 2013;52:1271–1278. doi: 10.1093/rheumatology/ket112. [DOI] [PubMed] [Google Scholar]
  • 17.Grabinska K, Pelak M, Wydmanski J, Tukiendorf A, d'Amico A. Prognostic value and clinical correlations of 18-fluorodeoxyglucose metabolism quantifiers in gastric cancer. World J Gastroenterol. 2015;21:5901–5909. doi: 10.3748/wjg.v21.i19.5901. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Hotker AM, Garcia-Aguilar J, Gollub MJ. Multiparametric MRI of rectal cancer in the assessment of response to therapy: A systematic review. Dis Colon Rectum. 2014;57:790–799. doi: 10.1097/DCR.0000000000000127. [DOI] [PubMed] [Google Scholar]
  • 19.Avcu S, Bulut MD, Yavuz A, Bora A, Beyazal M. Value of DW-MRI ADC quantification of colonic wall lesions in differentiation of inflammatory bowel disease and colorectal carcinoma. Jpn J Radiol. 2014;32:6–13. doi: 10.1007/s11604-013-0260-2. [DOI] [PubMed] [Google Scholar]
  • 20.Leufkens AM, Kwee TC, van den Bosch MA, Mali WP, Takahara T, Siersema PD. Diffusion-weighted MRI for the detection of colorectal polyps: Feasibility study. Magn Reson Imaging. 2013;31:28–35. doi: 10.1016/j.mri.2012.06.029. [DOI] [PubMed] [Google Scholar]
  • 21.Tomizawa M, Shinozaki F, Hasegawa R, Fugo K, Shirai Y, Motoyoshi Y, Sugiyama T, Yamamoto S, Kishimoto T, Ishige N. Factors affecting the detection of colorectal cancer and colon polyps on screening abdominal ultrasonography. Hepatogastroenterology. 2015;62:295–298. [PubMed] [Google Scholar]