Abstract
Purpose
We propose a new gradient measurement technique based on dynamic single point imaging (SPI), which allows simple, rapid, and robust measurement of k-space trajectory.
Methods
To enable gradient measurement, we utilize the variable field of view (FOV) property of dynamic SPI which is dependent on gradient shape. First, 1D dynamic SPI data are acquired from a targeted gradient axis, and then relative FOV scaling factors between 1D images or k-spaces at varying encoding times are found. These relative scaling factors are the relative k-space position that can be used for image reconstruction. The gradient measurement technique can also be used to estimate the gradient impulse response function for reproducible gradient estimation as a linear-time invariant system.
Result
The proposed measurement technique was used to improve reconstructed image quality in 3D ultra-short echo, 2D spiral, and multi-echo bipolar gradient echo Cartesian imaging. In multi-echo bipolar gradient echo imaging, measurement of the k-space trajectory allowed the use of a ramp-sampled trajectory for improved acquisition speed (approximately 30%) and more accurate quantitative fat and water separation in a phantom.
Conclusion
The proposed dynamic SPI-based method allows fast k-space trajectory measurement with a simple implementation and no additional hardware for improved image quality.
Keywords: Single point imaging, gradient, calibration, GIRF, k-space trajectory, eddy current
INTRODUCTION
The gradient system is an essential component in modern clinical MR imaging. It performs temporal-spatial encoding of transverse magnetization through a spatially varying magnetic field. Gradient waveforms can be synthesized to perform a range of image encoding strategies including conventional Cartesian image encoding, as well as non-Cartesian acquisitions such as radial(1) and spiral(2) imaging. Unfortunately, there still exist many factors that inevitably cause distortions in the realized gradient magnetic field: eddy currents(3–6), imperfection induced by the power amplifier, and mechanical/thermal vibrations(7,8). Due to these undesired distortions, it is challenging in practice to realize the actual gradient field exactly as prescribed, which, if the prescribed gradient is assumed during reconstruction, can result in image artifacts (e.g., blurring, ringing, or phase error). This can be a critical issue in non-Cartesian acquisitions, and is further exacerbated in acquisition schemes with a long readout duration such as spiral or echo planar imaging (EPI)(9,10). In these cases, the k-space trajectory is prone to deviate from that prescribed due to accumulated error in the phase evolution resulting from the distorted gradient.
Methods to estimate the actual gradient shape and the resultant k-space trajectory have been previously presented in the MR literature, and can be classified as follows: Imaging based gradient measurement (IGM) and magnetic field monitoring (MFM). In IGM, a (typically) 1D imaging technique based on a specialized pulse sequence is exploited to measure the gradient shape. In MFM, several NMR-based field probes are placed inside the magnet bore and used to record field characteristics temporally and spatially (11–14). While this provides the most direct measurement of the gradient field, the use of additional external hardware adds complication and expense.
According to the methodology of data acquisition, IGM methods can be further classified into two categories: frequency encoding based gradient measurement (FGM) and phase encoding based gradient measurement (PGM). In FGM(15), off-centered selection of a thin slice is performed to the avoid signal dephasing effect caused by gradient, followed by measurement of the phase evolution over encoding time in the manner of frequency encoding. Although the efficacy of this measurement scheme has been verified in many critical studies(16,17), there still exist limitations such as the dependency on slice selection and T2* decay. Alternatively, in previously proposed PGM methods(18–21), the phase evolution is measured at a constant (and single) echo time after a RF pulse, which is advantageous in terms of reducing the impact of T2* decay. However, a series of RF pulses must be applied to measure the whole gradient, which requires extensive measurement times.
Once the gradient has been characterized, it can be used directly in image reconstruction to improve image quality. However, such a measurement is specific to the characterized gradient and pulse sequence parameters and is not typically generalizable to other acquisitions. To enable more accurate output waveforms, it is routine in current generation MR systems to perform pre-emphasis correction by inputting a filtered (or intentionally distorted) waveform into the gradient subsystem(22). Such methods rely on the assumption that gradient systems are generally characterizable as linear time invariant (LTI) systems(23). Unfortunately, these system-level corrections do not realize sufficiently accurate waveforms, which may necessitate the further use of the gradient measurement techniques described above. Thus, a more generalizable approach to gradient measurement is to utilize these techniques and the LTI concept to compute a gradient impulse response function (GIRF). This approach allows estimation of the distorted gradient shape instead of directly measuring the realized gradient. By acquiring a comprehensive measurement of the gradient (e.g., obtained from one of the methods described above), the GIRF can be determined as a unique finite impulse response filter, and then any gradient shape realized in the same gradient system can be analytically predicted by convolving an estimated GIRF with the prescribed gradient (24–26).
In this study, we have developed a new gradient measurement method utilizing 1D dynamic single point imaging (SPI)(27) performed across a range of phase encoding time delays, which does not require slice selection, additional equipment, or knowledge about the imaged subject. The field of view (FOV) in SPI changes over phase encoding time delay, exhibiting a variable FOV property under an applied gradient. In the proposed gradient measurement method, 1D SPI encoding is implemented in each gradient axis by linearly scaling the amplitude of a tested gradient with each TR (i.e., from −1x to +1x to implement phase encoding). Then, the FOV scaling factors between different phase encoding time delays are estimated by using k-space or image domain representations of the 1D SPI data. The FOV scaling factors represent relative encoding positions in k-space between two phase encoding time delays, and the 1st derivative of FOV scaling factors represents relative amplitude in the measured gradient. This measured k-space trajectory can then be utilized in reconstruction to improve image quality.
Three gradient-intensive sequences (ultrashort time echo (UTE) (1), spiral, and multi-echo bipolar gradient echo (GRE)) were tested to evaluate the efficacy of the new SPI-based gradient measurement scheme. In UTE, a center-out radial acquisition is used to minimize the achievable echo time and enable contrast for species with ultrashort T2*’s. In spiral imaging, the gradient waveforms are simultaneously designed to operate within the peak gradient slew rate and peak gradient amplitude limits. Finally, multi-echo bipolar GRE with ramp sampling for more robust reconstruction in quantitative fat-water imaging is demonstrated. GIRFs were also estimated using the proposed method, and applied to obtain an estimated k-space trajectory in UTE, spiral, and multi-echo bipolar GRE imaging.
METHODS
Theory
The proposed method is based on assumption that gradient distortion is a LTI function of the gradient input (23). In the proposed technique, 1D dynamic SPI is performed by linearly scaling with Np steps (−1x to 1x) the entire gradient waveform along a single axis to obtain Npx1 data. Note that dynamic SPI differs from conventional SPI approaches where multiple k-spaces are continuously acquired with a prescribed sampling rate while phase encoding gradients are on.(28–30) Figure 1-a shows an example of a trapezoidal readout gradient to be measured, and Figure 1-b shows the corresponding dynamic SPI encoding gradient used for the proposed calibration technique. The FOV at phase encoding time delay (tp, the elapsed time after RF excitation) in dynamic SPI is determined by the following equation(28,31,32):
(1) |
where Np is the number of phase encoding steps, γ is a gyromagnetic ratio, and G(τ) is an amplitude of maximum phase encoding gradient at time delay τ, for example, which is the trapezoidal gradient with solid line in Figure 1-b. Figure 1-c shows the RF pulse and data acquisition window. Note that in dynamic SPI, multiple images can be acquired at every ΔTs (sampling interval) which improves sampling efficiency over conventional SPI approaches. Figure 1-d shows how the SPI sampling trajectory changes, exhibiting a time-decreasing FOV. The minimum Np can be determined by the required FOV (typically larger than the diameter of imaged subject to avoid aliasing), fovR, and gradient amplitude, G(τ), as in the following equation:
(2) |
where T is the end of readout. Data is acquired using the same acquisition window as desired for imaging, and numerous k-spaces over encoding time can be obtained depending on the sampling rate.
FOV scaling search
For gradient measurement, a reference point at a certain encoding time is first selected among the SPI encoded data. Then, the relative FOV scaling factor directly reflects the relative k-space trajectory with respect to the k-space coordinate at reference encoding time, tr, as following equation shows:
(3) |
where t denotes a phase encoding time delay, and k(t) is a k-space position in the unit of cycle m−1 at encoding time, t. Note that any data point can be used for the reference encoding time, tr, however data acquired around the encoding time at which the SPI image has one-half of the required FOV (fovR) may be desired such that both k-space and image domain 1D profiles contain enough information (or resolution) for reliable estimation of FOV scaling factors. Two possible approaches to estimate the relative FOV scaling factor exist: a k-space domain or an image domain approach, which can be formulated as a minimization problem. k-Space based optimization is shown in the following equation:
(4) |
where K(t,k) denotes the magnitude of k-space encoded at phase encoding time t in 1D SPI, s is a real number that is local FOV scaling factor, and EK is an error function between two k-spaces. In practice since a finite number of data (Np) is acquired, data interpolation is used to synthesize the scaled k-space K(t,sk) from original k-space K(t,k). Image domain optimization is performed as follows:
(5) |
where I(t,x) denotes the magnitude of 1D image at encoding time t, and EI is an error function between two images. As in k-space domain, interpolation is applied to synthesize the scaled image I(t,x/s).
The 1D profile in the image domain has more resolution (more information that can be used in FOV scaling search) when the FOV is small (Figure 1-e), while the k-space profile shows a broader line-shape when the FOV is large (Figure 1-f). To exploit this property, the two estimates may be combined using a simple merging filter as shown in Figure 1-g.
Absolute k-space trajectory
As shown in equation 3 above, the estimated FOV scaling factor, FOVscale(t), represents the relative k-space position at encoding time t with respect to the k-space position at the reference encoding time tr. The absolute k-space trajectory, which allows reconstruction of the image at the prescribed FOV, can be calculated by simply scaling the relative k-space trajectory. Most trivially, this is done by scaling the unitless measured gradient waveform to match the prescribed gradient amplitude (e.g., in units of mTm−1). Note that potentially more accurate scaling could utilize the GIRF-distorted gradient waveform or consider only the plateaus of trapezoidal gradients. Alternatively, a phantom of known dimension could be scanned to obtain this calibration if the system gradient amplitude is inaccurate or unknown. If the targeted gradient system is well calibrated, the DC component at 0 Hz in Fourier transform of the measured and prescribed gradient can be used to obtain the scaling factor.
GIRF
We have utilized the SPI-based gradient measurement technique herein to estimate GIRF, using multiple triangular input gradients using methodology analogous to (24). Figure 2-a shows the pulse sequence diagram depicting the input gradients used for GIRF measurement. Note that a “pre-dephasing” gradient is prescribed before the input gradient to remove ambiguity in FOV scaling estimation due to the large FOV when SPI data is acquired near the center of k-space. We performed estimation of GIRF in the frequency domain (i.e., the transfer function) by dividing the measured output gradient by the prescribed input gradient after discrete Fourier transform to calculate a transfer function of the LTI system. The estimated transfer function is low-pass filtered to cut off unreliable, high frequency components using a reduced cosine filter.
Experimental setup
The imaging parameters for GIRF measurement, dynamic SPI based gradient measurement, and imaging experiments performed on three different scanners are shown in Table 1. For GIRF measurement of a 3T MR750 scanner (S1) (GE Healthcare, Waukesha, WI, USA) utilized 21 triangular gradients with amplitude between 7 mTm−1 and 33 mTm−1, and a slew rate = 200 mTm−1ms−1 were used. A pre-dephasing gradient with amplitude 21.7 mTm−1 was placed with 464 μs spacing before the largest input gradient. To measure the GIRF of a 1.5T Signa HDxt scanner (S2) (GE Healthcare, Waukesha, WI, USA), 15 triangular gradients with amplitudes equally spaced between 6.7 mTm−1 and 20 mTm−1 were prescribed. A pre-dephasing gradient with amplitude 11.4 mTm−1 was prescribed with 476 μs spacing before the largest input gradient. Slew rate = 118 mTm−1ms−1 was used for all triangular gradients. To measure the GIRF of a 3T Signa PET/MR scanner (S3) (GE Healthcare, Waukesha, WI, USA), 21 triangular gradients with amplitude between 10 mTm−1 and 30 mTm−1, and a slew rate = 118 mTm−1ms−1 were used. A pre-dephasing gradient with amplitude 21 mTm−1 was placed with 588 μs spacing before the largest input gradient. GIRFs for all scanners were measured with vender provided pre-emphasis correction turned on. A GE Healthcare 8-channel receive-only head coil was used, and a manufacturer-provided 15 cm spherical phantom (with no internal structure) was imaged. Total scan time to measure GIRF on each of the 3 scanners was 106 sec for S1, 75 sec for S2, and 106 sec for S3.
Table 1.
GE MR750 (S1) | GE Signa HDxt (S2) | GE Signa PET/MR (S3) | ||
---|---|---|---|---|
GIRF measurement | RF pulse | 24μs hardpulse | 24μs hardpulse | 24μs hardpulse |
Np | 401 | 401 | 401 | |
Flip angle (degree) | 6 | 6 | 6 | |
Slew rate (mTm−1ms−1) | 200 | 118 | 118 | |
# of input gradient | 21 | 15 | 21 | |
Pre-dephaser (mTm−1) | 21.7 | 11.4 | 21.0 | |
Input gradients (mTm−1) | 7~31 | 6.7~20 | 10~30 | |
Spacing (μs) | 464 | 476 | 588 | |
TR(ms) | 4.4 | 4.4 | 4.4 | |
Sampling rate (kHz) | 500 | 250 | 500 | |
# of data points | 988 | 378 | 982 | |
RF Coil | 8-ch receive only head coil | |||
Scan time (sec) | 106 | 75 | 106 | |
3D UTE | 2D Spiral | 3D Ramp sampling | ||
Imaging experiment | RF pulse | 24μs hardpulse | Sinc pulse | 24μs hardpulse |
Flip angle (degree) | 6 | 30 | 6 | |
TE (ms) | 0.09 | 2.42 | Cartesian: 1.06, 1.97, 2.87, 3.77 Ramp sampling: 0.85, 1.49, 2.13, 2.77 |
|
TR (ms) | 3.3 | 13 | Cartesian: 5.2 Ramp sampling: 3.7 |
|
Sampling rate (kHz) | 500 | 250 | 500 | |
# of data points | 415 | 512 | Cartesian: 1234 Ramp sampling: 1830 |
|
# of TR | 80000 | 48 | 101×31 | |
RF coil | 8-ch receive only head coil | Single CH T/R coil | 8-ch receive only head coil | |
Scan time | 4min 28 sec | 0.6 sec | Cartesian: 16.3 sec Ramp sampling: 11.6 sec |
|
SPI-based gradient measurement | Np | 401 | 401 | 401 |
TR (ms) | 3.3 | 13 | 3.7 | |
Reference time, tr (ms) | 0.29 | 2.94 for x-axis 3.18 for y-axis |
21 | |
Sampling rate (kHz) | 500 | 250 | 500 | |
# of data points | 415 | 512 | 1830 | |
RF coil | 8-ch receive only head coil | |||
Scan time | 4sec | quick: 42sec extensive: 385sec |
1.5sec |
3D radial UTE imaging was performed on system S1 using a GE Healthcare 8 channel receive-only head coil. k-Space was encoded using a center-out half radial trajectory, using the pulse sequence shown in Figure 2-b. After RF excitation using a 24μs hard pulse, 80,000 spokes were scanned at TE=90 μs (after RF coil deadtime) with an encoding duration of 588 μs. The maximum amplitude of readout gradient was 35 mTm−1, slew rate was 118 mTm−1ms−1, flip angle was 6 degrees, sampling bandwidth was 500 kHz, and TR was 3.3 ms. SPI-based gradient measurement was performed in the x, y, and z directions with Np = 401 using the same scan parameters. An image was reconstructed at FOV = 24×24×24 cm and 1×1×1 mm resolution using gridding. The brain of a human volunteer was scanned under approval from our institutions IRB, and a 15 cm spherical phantom was used for gradient measurement as in GIRF measurement. The scan time for UTE imaging was 4 min 28 sec, and the scan time to perform gradient measurement for all three gradient axes was 4 seconds in total.
2D spiral imaging was performed on GE Signa HDxt scanner (S2) using the pulse sequence in Figure 2-c. A single channel GE Healthcare transmit/receive head coil was used for imaging. A sinc pulse was used to achieve 30 degree flip angle, and a single axial slice at iso-center was obtained. 48 spiral arms with 512 readout points in an arm was encoded with a sampling bandwidth of 250 kHz. FOV was 12×12 cm, slice thickness was 8 mm, and spatial resolution was 1.04×1.04 mm. TR was 13 ms, and TE was 2.42 ms. SPI gradient measurement with Np = 401 was performed in two different ways for comparison: extensive and quick. All 48 different pairs of x and y gradients were measured with the extensive gradient measurement (385 seconds), while 2 pairs of x and y gradients were measured and reproduced to estimate trajectories for all 48 arms using a linear combination in the quick gradient measurement (42 seconds). In the quick measurement, the readout gradients in x and y-axis in the arm with rotational angle θ, gx(θ,t) and gy(θ,t), can be estimated by the following equation.
(6) |
Note that gx(0,t), , gy(0,t), and are measured gradient shapes, where can be interpreted as a x-gradient waveform realized in physical y-gradient, and is y-gradient waveform realized in physical x-gradient. A 15 cm spherical phantom was used for gradient measurement as for the GIRF measurement, and a manufacturer-provided resolution phantom was scanned and reconstructed with the nominal and measured (extensively and quickly) k-space trajectory.
In the experiment for multi-echo bipolar GRE imaging, non-selective 3D GRE imaging was performed on GE Signa PET/MR scanner (S3) using an GE Healthcare 8 channel receive-only head coil, with 2 mm spatial resolution and FOV = 6×26×20 cm. Phase encoding was performed in x and z direction with 61 and 201 phase encoding steps, respectively, and frequency encoding was performed in y direction. 8 vials containing 0%, 5%, 10%, 15%, 20%, 30%, 40%, and 50% fat with a T1 shortening agent were used for the experiment. Two different pulse sequences using the bipolar gradients shown in Figure 2-d were used to acquire 4 gradient echoes, where the readout gradients with a longer and shorter plateau were used for the conventional Cartesian sampling and ramp sampling, respectively. TE was 1061, 1965, 2869, and 3773 μs for Cartesian sampling and 847, 1487, 2127, and 2767 μs for ramp sampling. The maximum amplitude of the readout gradient for Cartesian or ramp sampling acquisition was respectively 22.6 mTm−1 or 33.0 mTm−1. Note that a gradient spoiler was applied in the readout direction by stretching the trapezoidal gradient at the end of the pulse sequence. The minimum TR was 5.2 ms for Cartesian sampling and 3.7 ms for ramp sampling (a 29% reduction), where scan time was 16.3 sec for Cartesian and 11.6 sec for ramp sampling. A 24 μs hard pulse with a 6 degree flip angle was used, and the sampling rate was 250 kHz and 500 kHz bandwidth for Cartesian and ramp sampling acquisitions respectively. SPI-based gradient measurement was performed for ramp sampling in the readout direction (y-axis) with Np = 401. The acquisition time for gradient measurement was 1.5 sec. A 15 cm spherical phantom was used for gradient measurement.
Data processing
Figure 3 shows a block diagram delineating the steps in the proposed method. The acquired data were processed in MATLAB (The Mathworks Inc, Natick, MA, USA). In the FOV scaling factor search stage, the reference encoding time, tr, is selected by referring to the nominal trajectory. Once the FOV scaling factors are estimated in k-space and image domain, they are combined using a merging filter. The merging filter was designed to have a linear slope in transition, where the width of transition was prescribed to cover five data points. Then, the absolute k-space trajectory is computed by scaling the relative k-space trajectory to physical gradient units. After estimating trajectories for the sampled k-space data, a convolution gridding reconstruction (33–36) was performed with oversampling ratio = 1.5 and a gridding kernel width = 5 pixels.
Note that any non-linear optimization method can be used to solve equation 4 and 5. In GIRF estimation, UTE imaging, and ramp sampling experiments, simple unconstrained optimization based on Nelder-Mead simplex (fminsearch in MATLAB) was used to estimate the FOV scaling factors. In the spiral imaging experiment, bounded nonlinear function optimization based on golden section search and parabolic interpolation(37) (fminbnd in MATLAB) was used for the FOV scaling search. The search range was set to ±7% of initial guess. The nominal k-space trajectory was used as an initial guess for the optimization in UTE imaging, and the GIRF-corrected trajectory was used as an initial guess for spiral imaging and ramp sampling. To scale the 1D image and k-space data in each iteration of optimization, cubic interpolation was applied. L2-norm and negative linear correlation were used as an error function for k-space (EK) and image (EI) respectively. Image domain SPI data was used for GIRF estimation, while k-space and image domain data was used for direct gradient measurement in UTE imaging, spiral imaging, and ramp sampling. To scale the unitless, relative gradient shape to an absolute gradient shape, estimated amplitudes in plateau of the first readout gradient was compared with the corresponding part in prescribed gradient shape in UTE imaging. In spiral imaging, entire gradient shape was compared with the GIRF-distorted gradient shape to scale the estimated gradient shape to obtain the correct FOV in image reconstruction. In multi-echo GRE imaging and all GIRF measurements in three systems (S1, S2, and S3), the DC component of Fourier transform measured and prescribed gradient waveform was used to obtain the absolute gradient shape.
In multi-echo bipolar GRE imaging, conventional Cartesian data (non-ramp sampled data) was directly reconstructed with no additional phase correction. Data acquired with ramp sampling was reconstructed using the GIRF-corrected trajectory or the measured trajectory. After reconstructing multi echo images, Iterative Decomposition of Water and Fat With Echo Asymmetry and Least-Squares Estimation (IDEAL) was applied to obtain fat and water separated images(38), and the fat fraction was calculated.
RESULTS
GIRF estimation
Figure 4-a shows one input gradient with amplitude of 33 mTm−1 following a pre-dephasing gradient, which was used for GIRF estimation in S1. Figure 4-b shows the normalized magnitude of the 1D SPI images over encoding time obtained in y-axis, and Figure 4-c shows the estimated FOV scaling factors. Figure 4-d shows the measured gradient shape. Figure 5-a,b,c shows the estimated transfer function in the Fourier transform domain and the corresponding GIRF in the time domain for three different systems, S1, S2, and S3, respectively. The magnitude and phase of the estimated transfer functions (discrete Fourier transform of GIRF) in the frequency domain and the corresponding GIRFs in the time domain are shown here, which shows suppressed noise up to 20 kHz. The GIRFs were obtained by low-pass filtering and performing the inverse discrete Fourier transform to the transfer function. The parameters for the low-pass filter was empirically determined to suppress the amplified noise, where full width half max was 44.4 kHz, and the transition band was 16 kHz. The off-centered peak in the GIRF implies a group delay of approximately 8 μs for S1 and S3 and 20 μs for S2, which matches with an empirically observed delay in the three systems in our group.
Ultra-short echo imaging
Figure 6-a shows the log magnitude of 1D k-space obtained using SPI encoding for gradient measurement in the x, y, and z-axis, and Figure 6-b shows the magnitude of the corresponding 1D SPI images that are normalized by the maximum amplitude in each encoding time. Under the mono-polar trapezoidal gradient, the encoded 1D k-space shows time-narrowing line shape over encoding, while the 1D image shows a time-broadening shape, where the both k-spaces and images exhibit time-decreasing FOV. Figure 6-c shows the GIRF measured trajectory, the SPI measured trajectory, and the nominal trajectory in physical x, y, z-axis, and a zoomed-in view, where group delay is shown between the nominal and the measured k-space trajectories. Figure 6-d show the UTE images reconstructed with the nominal, GIRF, and SPI measured trajectory for a sagittal and axial slice. The image reconstructed with the GIRF and SPI measured trajectory shows good quality with no visible imaging artifact such as ringing, while the image with the nominal trajectory exhibits ringing and mis-aligned image components.
Spiral imaging
Figure 7-a shows the nominal, GIRF measured, and SPI measured trajectory. Figure 7-b shows the image reconstructed using the nominal trajectory, and Figure 7-c shows images reconstructed with the delay-corrected trajectory where a group delay of 20 μs was used. Figure 7-d shows images reconstructed with trajectory estimated using GIRF. Figure 7-e and f show images reconstructed using the trajectory obtained by extensive and quick measurement, where all 48 spiral arms were measured independently or only 4 arms of measurement, respectively. Figure 7-g shows a difference image relative to the image using the extensively measured trajectory in Figure 7-e. The image reconstructed with the delay-uncorrected trajectory shows a severe blurring artifact, while the image with delay-corrected trajectory shows reduced blurring artifact, however there is remaining blurriness artifact as indicated by the red arrows. Images reconstructed using the GIRF-trajectory, extensively measured trajectory, or quickly measured trajectory show a much better result, while images with the measured trajectory show sharper edges as indicated by yellow arrows. The quick gradient measurement shows a comparable result to the full measurement as shown in the difference image in Figure 7-g, with 8.3% of the scan time compared to the extensive measurement time.
Multi-echo bipolar GRE imaging
Figure 8 shows fat and water separated images and the resultant fat fraction map obtained by three different imaging and reconstruction schemes: conventional Cartesian imaging, ramp sampling with nominal trajectory, ramp sampling with GIRF, and ramp sampling with SPI-based gradient measurement. The measured gradients show the raw data (unfiltered) of the estimated gradient waveform. The SNR measured from the separated water in all tubes, the mean and standard deviation of estimated fat fraction in each tube are shown in Table 2. In conventional Cartesian sampling, erroneous estimation of fat faction is shown due to the phase error between the echoes acquired with positive and negative readout gradient. In ramp sampling with GIRF-estimated or SPI-measured trajectory, the estimation of fat fraction is significantly improved, which is likely due to a combination of the improved accuracy in the k-space sampling position that center-aligns k-spaces across echoes as well as the reduced echo spacing, while use of the nominal trajectory yields an incorrect estimation of fat fraction with lower SNR than in ramp sampling with the corrected trajectory. The SNR measured in ramp sampling was comparable with the SNR in Cartesian sampling. The estimated fat fraction shows both GIRF and SPI-based gradient measurement method allow robust measurement of the readout gradient.
Table 2.
SNR | 0% fat (%) | 5% fat (%) | 10% fat (%) | 15% fat (%) | 20% fat (%) | 30% fat (%) | 40% fat (%) | 50% fat (%) | |
---|---|---|---|---|---|---|---|---|---|
Cartesian | 55.3 | 10.4 ±13.2 | 12.2 ±21.5 | 13.0 ±2.3 | 13.8 ±3.1 | 13.2 ±2.8 | 21.5 ±2.1 | 34.6 ±1.6 | 47.3 ±1.8 |
Ramp sampling with nominal trajectory | 45.9 | 37.9 ±1.9 | 42.3 ±1.7 | 46.7 ±1.9 | 50.4 ±1.8 | 48.1 ±2.0 | 43.5 ±2.1 | 42.6 ±1.7 | 48.2 ±1.7 |
Ramp sampling with GIRF | 56.1 | 5.9 ±2.1 | 6.2 ±2.0 | 9.9 ±2.4 | 14.8 ±2.6 | 19.7 ±2.6 | 29.0 ±1.6 | 39.6 ±1.5 | 50.6 ±1.7 |
Ramp sampling with SPI-based measurement | 56.4 | 5.4 ±2.1 | 6.2 ±2.2 | 10.3 ±2.3 | 15.3 ±3.0 | 19.8 ±2.5 | 29.1 ±1.7 | 39.8 ±1.5 | 50.3 ±1.8 |
DISCUSSION
The proposed SPI-based gradient measurement technique does not require any special hardware unlike other reported methods for PGM and MFM that need specialized equipment such as NMR field probes(11,12,20,39). While these methods have been shown to be viable techniques to measure gradient waveforms, the use of external hardware adds complication and may be cost prohibitive. In the proposed method, a 15 cm spherical phantom was used to perform gradient waveform measurement; however in theory, any object (including a human patient) can be used to perform gradient measurement. However, the imaged object does require definite boundaries (determined either by object size, coil sensitivity, or slice selectivity) to allow computation of the relative scale factors. The proposed method is robust to rigid motion of the object in two ways: the k-space scheme that only use magnitude that does not influenced by motion, and short measurement times that mitigate possible motion during the scan. The proposed SPI-based gradient measurement technique requires very minor modifications to the targeted pulse sequence, requiring only an additional loop during which the to be measured gradient amplitude is scaled for each gradient channel measured. Thus, this gradient waveform measurement can be added into existing sequences with little effort.
Compared with previously reported PGM(18–20) methods, where the number of RF pulses determines the resolution of estimated trajectory or gradient, the proposed technique allows higher resolution sampling of the gradient waveform (determined by the sampling bandwidth of the readout event). Therefore, the k-space position can be directly estimated from the measurement without any interpolation using the identical sampling rate for image acquisition. Furthermore, acquisition of the gradient measurement can be extremely rapid when the scan TR is short (e.g., ~1.5 sec per gradient axis). Furthermore, the use of the proposed methodology to perform in vivo measurement is feasible, where, for example, a database gradient measurement (e.g., obtained using a phantom) could be used to provide robust estimates for rapid gradient measurements. In vivo gradient measurement time for longer readouts (e.g., spiral and echo-planar imaging) could be further reduced by sub-sampling the number of phase encoding steps, particularly when a database calibration measurement is used. Note that in vivo calibration could be challenged when static field gradients due to magnetic field distortion is significant.
One additional feature of the SPI-based gradient measurement is that k-space trajectory can be independently recovered without knowing the history of previous k-space position. Theoretically, the FOV itself in a 1D SPI image is a direct and independent measure of k-space position as implied in equation 1. This allows flexibility to measure k-space trajectory in any type of pulse sequences even when the central region of k-space is not acquired as in hybrid encoding(40,41) and zero TE imaging(42–45). Furthermore, in the SPI-based method, noise or mis-estimation at earlier encoding times is not propagated to the later estimation since no integration or cumulative summation is required to obtain k-space position.
As demonstrated herein, the proposed gradient waveform method can be used to improve the quality of reconstructed images. This is particularly important for non-Cartesian imaging where small k-space trajectory errors lead to significant artifacts as seen in Figure 6. Another application is the use of the technique to measure trajectories in more conventional Cartesian-type acquisitions. In Figure 8, improvement is seen in chemical-shift encoded imaging (IDEAL) with a multi-echo bipolar readout by reconstructing to a measured trajectory which reduces artefactual phase due to an inaccurate k-space trajectory(46–48). Furthermore, ramp sampling reduced the total scan time by approximately 30%. The ability to perform a robust and rapid gradient measurement technique to enable ramp sampling, particularly for oblique slice orientations would be beneficial to many MRI sequences (e.g., fast gradient echo, echo-planar imaging, balanced SSFP, fast-spin echo) to provide moderate scan time reductions (20–30%) with minimal effect upon SNR.
The key idea of the proposed SPI-based gradient measurement technique is based on the assumption of linearity of the gradient distortion. This is typically a reasonable assumption when distortions are mostly caused by eddy currents that linearly scale with gradient amplitude. However, these conditions may be violated in certain instances such as gradient operation close to the limit of slew rate and amplitude, resulting in nonlinear distortion in the realized gradient shape, which is a limitation of the proposed method based upon this LTI assumption. Furthermore, the demonstration of the proposed method relies upon accurate gradient amplitude calibration (which is already a requirement to enable geometrically precise MR imaging). If mis-calibrated, global geometric distortions will be apparent relative to the physical x, y, and z gradient errors. However, this scaling can readily be compensated for by imaging a phantom with known geometry (for which the proposed method would be particularly well-suited).
Concomitant field effects are another nonlinear factor that cause deviations in the k-space trajectory. These time-variant, non-linear magnetic fields depend on gradient amplitudes and are inversely proportional to field strength. The concomitant field effects become more problematic in imaging with long readout duration and preparation gradients (e.g. velocity encoding). Despite the efficacy of the proposed gradient measurement method as shown in the experiments herein, it is difficult to directly measure the concomitant magnetic field or higher order fields since in the proposed method (and other FGM and PGM methods) the gradient measurement is performed independently in each axis, based on the linearity assumption of the gradient system. In practice, concomitant magnetic fields are modeled as second order approximations and several successful correction methods have been proposed and implemented (10,49–51). Thus, these additional correction terms could be readily applied to gradient measurement data obtained using the proposed method.
The proposed technique can be used as a calibration technique to estimate the GIRF without external hardware. As shown hereinbefore, the data collection for GIRF requires only 25~35 sec per gradient axis, making it very feasible for measurement during routine (e.g., daily or weekly) quality assurance imaging. In the present experiment using triangular gradient blips, the spectral resolution is inherently limited, and hence it may not be sensitive to long lasting eddy currents which is not visible in the estimated GIRF. This could potentially be overcome by applying frequency sweep methods (26,52), which we will explore in future work.
CONCLUSIONS
In this study, we implemented a robust and rapid gradient measurement method based on dynamic SPI, which allowed accurate measurement of k-space trajectory with high fidelity and no need of additional equipment to improve reconstructed image quality.
Acknowledgments
The research reported in this publication was supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under Award Number 1R21EB013770. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. We thank Diego Hernando, PhD for supplying the IDEAL code used herein.
References
- 1.Tyler DJ, Robson MD, Henkelman RM, Young IR, Bydder GM. Magnetic resonance imaging with ultrashort TE (UTE) PULSE sequences: technical considerations. J Magn Reson Imaging. 2007;25:279–89. doi: 10.1002/jmri.20851. [DOI] [PubMed] [Google Scholar]
- 2.Tan H, Meyer CH. Estimation of k-space trajectories in spiral MRI. Magn Reson Med. 2009;61:1396–1404. doi: 10.1002/mrm.21813. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Boesch CH. in Superconducting Magnets: Optimization of Corrections and Quantitative Characterization of Magnet / Gradient Systems. 1991;284:268–284. doi: 10.1002/mrm.1910200209. [DOI] [PubMed] [Google Scholar]
- 4.Jehenson P, Westphal M, Schuff N. Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field gradients in NMR systems. J Magn Reson. 1990;90:264–278. doi: 10.1016/0022-2364(90)90133-T. [DOI] [Google Scholar]
- 5.Ahn CB, Cho ZH. Analysis of the eddy-current induced artifacts and the temporal compensation in nuclear magnetic resonance imaging. IEEE Trans Med Imaging. 1991;10:47–52. doi: 10.1109/42.75610. [DOI] [PubMed] [Google Scholar]
- 6.Wysong RE, Madio DP, Lowe IJ. A novel eddy current compensation scheme for pulsed gradient systems. Magn Reson Med. 1994;31:572–575. doi: 10.1002/mrm.1910310517. [DOI] [PubMed] [Google Scholar]
- 7.Wu Y, Chronik Ba, Bowen C, Mechefske CK, Rutt BK. Gradient-induced acoustic and magnetic field fluctuations in a 4T whole- body MR imager. Magn Reson Med. 2000;44:532–6. doi: 10.1002/1522-2594(200010)44:4<532::aid-mrm6>3.0.co;2-q. [DOI] [PubMed] [Google Scholar]
- 8.Foerster BU, Tomasi D, Caparelli EC. Magnetic field shift due to mechanical vibration in functional magnetic resonance imaging. Magn Reson Med. 2005;54:1261–1267. doi: 10.1002/mrm.20695. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Du YP, Zhou XJ, Bernstein Ma. Correction of concomitant magnetic field-induced image artifacts in nonaxial echo-planar imaging. Magn Reson Med. 2002;48:509–515. doi: 10.1002/mrm.10249. [DOI] [PubMed] [Google Scholar]
- 10.Zhou XJ, Du YP, Bernstein Ma, Reynolds HG, Maier JK, Polzin Ja. Concomitant magnetic-field-induced artifacts in axial echo planar imaging. Magn Reson Med. 1998;39:596–605. doi: 10.1002/mrm.1910390413. [DOI] [PubMed] [Google Scholar]
- 11.Hammer BE. Magnetic field mapping with an array of nuclear magnetic resonance probes. Rev Sci Instrum. 1996;67:2378. doi: 10.1063/1.1147005. [DOI] [Google Scholar]
- 12.De Zanche N, Barmet C, Nordmeyer-Massner Ja, Pruessmann KP. NMR Probes for measuring magnetic fields and field dynamics in MR systems. Magn Reson Med. 2008;60:176–186. doi: 10.1002/mrm.21624. [DOI] [PubMed] [Google Scholar]
- 13.Barmet C, De Zanche N, Pruessmann KP. Spatiotemporal magnetic field monitoring for MR. Magn Reson Med. 2008;60:187–197. doi: 10.1002/mrm.21603. [DOI] [PubMed] [Google Scholar]
- 14.Barmet C, Wilm B, Pavan M, Pruessmann K. A third-order field camera with microsecond resolution for MR system diagnostics. Proc Intl Soc Mag Reson Med. 2009;4185:780. [Google Scholar]
- 15.Duyn JH, Yang Y, Frank Ja, van der Veen JW. Simple correction method for k-space trajectory deviations in MRI. J Magn Reson. 1998;132:150–153. doi: 10.1006/jmre.1998.1396. [DOI] [PubMed] [Google Scholar]
- 16.Papadakis NG, Martin KM, Pickard JD, Hall LD, Carpenter Ta, Huang CL. Gradient preemphasis calibration in diffusion-weighted echo-planar imaging. Magn Reson Med. 2000;44:616–624. doi: 10.1002/1522-2594(200010)44:4<616::AID-MRM16>3.0.CO;2-T. [pii] [DOI] [PubMed] [Google Scholar]
- 17.Johnson KM, Fain SB, Schiebler ML, Nagle S. Optimized 3D ultrashort echo time pulmonary MRI. Magn Reson Med. 2013;70:1241–1250. doi: 10.1002/mrm.24570. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Balcom BJ, Bogdan M, Armstrong RL. Single-Point Imaging of Gradient Rise, Stabilization, and Decay. J Magn Reson Ser A. 1996;118:122–125. doi: 10.1006/jmra.1996.0018. [DOI] [Google Scholar]
- 19.Goodyear DJ, Shea M, Beyea SD, Shah NJ, Balcom BJ. Single point measurements of magnetic field gradient waveform. J Magn Reson. 2003;163:1–7. doi: 10.1016/S1090-7807(03)00152-6. [DOI] [PubMed] [Google Scholar]
- 20.Han H, MacGregor RP, Balcom BJ. Pure phase encode magnetic field gradient monitor. J Magn Reson. 2009;201:212–217. doi: 10.1016/j.jmr.2009.09.011. [DOI] [PubMed] [Google Scholar]
- 21.Schneider JT, Haas M, Ruhm W, Hennig J, Ullmann P. Robust spatially selective excitation using radiofrequency pulses adapted to the effective spatially encoding magnetic fields. Magn Reson Med. 2011;65:409–421. doi: 10.1002/mrm.22635. [DOI] [PubMed] [Google Scholar]
- 22.Gach HM, Lowe IJ, Madio DP, Caprihan a, Altobelli Sa, Kuethe DO, Fukushima E. A programmable pre-emphasis system. Magn Reson Med. 1998;40:427–431. doi: 10.1002/mrm.1910400313. [DOI] [PubMed] [Google Scholar]
- 23.Brodsky EK, Samsonov Aa, Block WF. Characterizing and correcting gradient errors in non-Cartesian imaging: Are gradient errors Linear Time-Invariant (LTI)? Magn Reson Med. 2009;62:1466–1476. doi: 10.1002/mrm.22100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Vannesjo SJ, Haeberlin M, Kasper L, Pavan M, Wilm BJ, Barmet C, Pruessmann KP. Gradient system characterization by impulse response measurements with a dynamic field camera. Magn Reson Med. 2013;69:583–93. doi: 10.1002/mrm.24263. [DOI] [PubMed] [Google Scholar]
- 25.Vannesjo SJ, Haeberlin M, Kasper L, Pavan M, Wilm BJ, Barmet C, Pruessmann KP. Image reconstruction using the gradient impulse response for trajectory prediction. Magn Reson Med. 2013;69:583–93. doi: 10.1002/mrm.24263. [DOI] [PubMed] [Google Scholar]
- 26.Addy NO, Wu HH, Nishimura DG. Simple method for MR gradient system characterization and k-space trajectory estimation. Magn Reson Med. 2012;68:120–129. doi: 10.1002/mrm.23217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Emid S, Creyghton JHN. High resolution NMR imaging in solids. Phys B+C. 1985;128:81–83. doi: 10.1016/0378-4363(85)90087-7. [DOI] [Google Scholar]
- 28.Jang H, Subramanian S, Devasahayam N, Saito K, Matsumoto S, Krishna MC, McMillan AB. Single Acquisition Quantitative Single-Point Electron Paramagnetic Resonance Imaging. Magn Reson Med. 2013;70:1173–1181. doi: 10.1002/mrm.24886. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Kaffanke JB, Romanzetti S, Dierkes T, Leach MO, Balcom BJ, Jon Shah N. Multi-Frame SPRITE: A method for resolution enhancement of multiple-point SPRITE data. J Magn Reson. 2013;230C:111–116. doi: 10.1016/j.jmr.2013.01.008. [DOI] [PubMed] [Google Scholar]
- 30.Jang H, Matsumoto S, Devasahayam N, Subramanian S, Zhuo J, Krishna MC, McMillan AB. Accelerated 4D quantitative single point EPR imaging using model-based reconstruction. Magn Reson Med. 2015;73:1692–1701. doi: 10.1002/mrm.25282. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Matsumoto K, Subramanian S, Devasahayam N, Aravalluvan T, Murugesan R, Cook Ja, Mitchell JB, Krishna MC. Electron paramagnetic resonance imaging of tumor hypoxia: enhanced spatial and temporal resolution for in vivo pO2 determination. Magn Reson Med. 2006;55:1157–63. doi: 10.1002/mrm.20872. [DOI] [PubMed] [Google Scholar]
- 32.Subramanian S, Devasahayam N, Murugesan R, Yamada K, Cook J, Taube A, Mitchell JB, Lohman JaB, Krishna MC. Single-point (constant-time) imaging in radiofrequency Fourier transform electron paramagnetic resonance. Magn Reson Med. 2002;48:370–9. doi: 10.1002/mrm.10199. [DOI] [PubMed] [Google Scholar]
- 33.Pipe JG. Reconstructing MR images from undersampled data: data-weighting considerations. Magn Reson Med. 2000;43:867–75. doi: 10.1002/1522-2594(200006)43:6<867::aid-mrm13>3.0.co;2-2. [DOI] [PubMed] [Google Scholar]
- 34.Pipe JG, Menon P. Sampling density compensation in MRI: rationale and an iterative numerical solution. Magn Reson Med. 1999;41:179–86. doi: 10.1002/(sici)1522-2594(199901)41:1<179::aid-mrm25>3.0.co;2-v. [DOI] [PubMed] [Google Scholar]
- 35.Johnson KO, Pipe JG. Convolution kernel design and efficient algorithm for sampling density correction. Magn Reson Med. 2009;61:439–47. doi: 10.1002/mrm.21840. [DOI] [PubMed] [Google Scholar]
- 36.Beatty PJ, Nishimura DG, Pauly JM. Rapid gridding reconstruction with a minimal oversampling ratio. IEEE Trans Med Imaging. 2005;24:799–808. doi: 10.1109/TMI.2005.848376. [DOI] [PubMed] [Google Scholar]
- 37.Brent R. Algorithms for minimization without derivatives. Englewood Cliffs, New Jersey: Prentice-Hall; 1974. [Google Scholar]
- 38.Reeder SB, Pineda AR, Wen Z, Shimakawa A, Yu H, Brittain JH, Gold GE, Beaulieu CH, Pelc NJ. Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): application with fast spin-echo imaging. Magn Reson Med. 2005;54:636–44. doi: 10.1002/mrm.20624. [DOI] [PubMed] [Google Scholar]
- 39.Goora FG, Balcom BJ. Accuracy of the Magnetic Field Gradient Waveform Monitor ( MFGM ) Technique and Consequent Accuracy of Pre-Equalized Gradient Waveforms. :1–14. doi: 10.1002/cmr.b.000.21323. [DOI] [Google Scholar]
- 40.Grodzki DM, Jakob PM, Heismann B. Ultrashort echo time imaging using pointwise encoding time reduction with radial acquisition (PETRA) Magn Reson Med. 2012;67:510–8. doi: 10.1002/mrm.23017. [DOI] [PubMed] [Google Scholar]
- 41.Jang H, Wiens CN, McMillan AB. Ramped hybrid encoding for improved ultrashort echo time imaging. Magn Reson Med. 2015 doi: 10.1002/mrm.25977. 00:n/a–n/a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Weiger M, Pruessmann KP, Bracher A-K, Köhler S, Lehmann V, Wolfram U, Hennel F, Rasche V. High-resolution ZTE imaging of human teeth. NMR Biomed. 2012;25:1144–51. doi: 10.1002/nbm.2783. [DOI] [PubMed] [Google Scholar]
- 43.Weiger M, Brunner DO, Dietrich BE, Müller CF, Pruessmann KP. ZTE imaging in humans. Magn Reson Med. 2013;70:328–32. doi: 10.1002/mrm.24816. [DOI] [PubMed] [Google Scholar]
- 44.Wiesinger F, Sacolick LI, Menini A, Kaushik SS, Ahn S, Veit-haibach P, Delso G, Shanbhag DD. Zero TE MR Bone Imaging in the Head. Magn Reson Med. 2015 doi: 10.1002/mrm.25545. 00:n/a–n/a. [DOI] [PubMed] [Google Scholar]
- 45.Schieban K, Weiger M, Hennel F, Boss A, Pruessmann KP. ZTE imaging with enhanced flip angle using modulated excitation. Magn Reson Med. 2014;00:1–10. doi: 10.1002/mrm.25464. [DOI] [PubMed] [Google Scholar]
- 46.Yu H, Shimakawa A, McKenzie CA, Lu W, Reeder SB, Hinks RS, Brittain JH. Phase and amplitude correction for multi-echo water-fat separation with bipolar acquisitions. J Magn Reson Imaging. 2010;31:1264–1271. doi: 10.1002/jmri.22111. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Soliman AS, Wiens CN, Wade TP, McKenzie Ca. Fat quantification using an interleaved bipolar acquisition. Magn Reson Med. 2015 doi: 10.1002/mrm.25807. 00:n/a–n/a. [DOI] [PubMed] [Google Scholar]
- 48.Lu W, Yu H, Shimakawa A, Alley M, Reeder SB, Hargreaves Ba. Water-fat separation with bipolar multiecho sequences. Magn Reson Med. 2008;60:198–209. doi: 10.1002/mrm.21583. [DOI] [PubMed] [Google Scholar]
- 49.Bernstein Ma, Zhou XJ, Polzin Ja, King KF, Ganin A, Pelc NJ, Glover GH. Concomitant gradient terms in phase contrast MR: Analysis and correction. Magn Reson Med. 1998;39:300–308. doi: 10.1002/mrm.1910390218. [DOI] [PubMed] [Google Scholar]
- 50.King KF, Ganin A, Zhou XJ, Bernstein Ma. Concomitant gradient field effects in spiral scans. Magn Reson Med. 1999;41:103–112. doi: 10.1002/(SICI)1522-2594(199901)41:1<103::AID-MRM15>3.0.CO;2-M. [DOI] [PubMed] [Google Scholar]
- 51.Cheng JY, Santos JM, Pauly JM. Fast concomitant gradient field and field inhomogeneity correction for spiral cardiac imaging. Magn Reson Med. 2011;66:390–401. doi: 10.1002/mrm.22802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Vannesjo SJ, Dietrich BE, Pavan M, Brunner DO, Wilm BJ, Barmet C, Pruessmann KP. Field camera measurements of gradient and shim impulse responses using frequency sweeps. Magn Reson Med. 2014;72:570–583. doi: 10.1002/mrm.24934. [DOI] [PubMed] [Google Scholar]