Abstract
Background
Catheter-Associated Hospital-Acquired Infections (HAI's) are caused by biofilm-forming bacteria. Using a novel approach, we generated anti-infective barrier on catheters by charging them with Nitric Oxide (NO), a naturally-produced gas molecule. NO is slowly released from the catheter upon contact with physiological fluids, and prevents bacterial colonization and biofilm formation onto catheter surfaces.
Aims and methods
The aim of the study was to assess the anti-infective properties of NO-charged catheters exposed to low concentration (up to 103 CFU/ml) of microbial cells in-vitro. We assessed NO-charged tracheal tubes using Pseudomonas aeruginosa, dialysis and biliary catheters using Escherichia coli, and urinary catheters using E. coli, Candida albicans or Enterococcus faecalis. Safety and tolerability of NO-charged urinary catheters were evaluated in a phase 1 clinical study in 12 patients. Six patients were catheterized with NO-charged catheters (NO-group), followed by 6 patients catheterized with regular control catheters (CT-group). Comparison of safety parameters between the study groups was performed.
Results
NO-charged tracheal, dialysis biliary and urinary catheters prevented P. aeruginosa, E. coli and C. albicans attachment and colonization onto their surfaces and eradicated corresponding planktonic microbial cells in the surrounding media after 24–48 hours, while E. faecalis colonization onto urinary catheters was reduced by 1 log compared to controls. All patients catheterized with an NO-charged urinary catheter successfully completed the study without experiencing NO-related AE's or serious AE's (SAE's).
Conclusion
These data highlight the potential of NO-based technology as potential platform for preventing catheter-associated HAI's.
Introduction
Hospital-Acquired Infections (HAI's) affects millions of patients worldwide annually, resulting in increased morbidity and mortality, and inflicting substantial financial losses to health systems [1]. High frequency of HAI's such as Blood Stream Infection (BSI), Urinary Tract Infection (UTI) and Ventilator-Associated Pneumonia (VAP) are associated with the use of invasive medical devices such as central lines, urinary catheters and tracheal catheters, respectively [2, 3].
Bacteria can adhere to the surface of the catheter within hours from catheterization and colonize, creating a persistent environment called a biofilm [4–6]. These colonies consist of bacteria that have a profoundly different set of genes from those of planktonic bacteria [6], and resistant to standard antibiotics, disinfectants or germicides [7–8], augmenting the potential of these pathogens to cause infections in catheterized patients.
Nitric Oxide (NO) is a naturally-produced gas molecule which plays an important role in host-defense against a large variety of pathogens [9]. Several NO-releasing technologies were developed during the years. Amongst them NO-releasing nanoparticles and dendrimers were shown to have anti-microbial and anti-biofilm capabilities against multiple microbial species [10–12].
Using a novel approach, we generated an anti-infective barrier on catheters by charging them with NO. We have previously shown in-vitro that NO-charged urinary catheters that are immersed in urine slowly release low concentration of NO (μM levels) for at least 14 days, and prevent Escherichia coli colonization and biofilm formation onto their luminal and exterior surfaces [13]. NO-charged urinary catheters were found effective also in comparison with antibiotic-coated catheters (Nitrofurazone) and eradicated all bacteria in planktonic and biofilm states [14]. Moreover, NO-charged urinary catheters were superior compared to silver-coated catheters (that are clinically used) in preventing bacterial colonization and biofilm formation onto catheters surfaces [14].
Although NO is a natural gas and has a short half-life, prior studies demonstrated it may cause several unwanted side effects. NO gas at low concentrations of 5–40 ppm was found to be associated with decrease in blood pressure, inhibition of platelet aggregation, increased bleeding time and local side effects to its surrounding tissue such as uncontrolled erection (i.e. priapism) local skin irritation, and the formation of local skin edema and erythema [15–20].
The objectives of this study were to evaluate the anti-infective properties of several NO-charged catheter types in-vitro against low concentration (up to 103 CFU/ml) of common gram-negative, gram-positive bacteria and fungi, and to evaluate for the first time in men the safety and tolerability of NO-charged urinary catheters in patients undergoing radical prostatectomy that were catheterized for a period of 7–28 days.
Materials and methods
Charging of catheters with NO
Four different types of catheters were charged with NO in a previously described exposure chamber under proprietary conditions and using a proprietary technique [21]. These catheters were: Mallinckrodt™ hi-contour tracheal tube 4,5 Oral/Nasal 6,2 (COVIDIEN), HemoStream™ chronic dialysis catheter 15.5Fr (Angiotech), SKATER™ biliary drainage catheter 12Fr (Angiotech) and 18Fr silicone urinary catheters (Biometrix, Ltd, Israel). All catheter types were used in-vitro and the urinary catheter was used in-vivo. For each catheter type we used uncharged catheters or catheter sections as controls.
Measurement of NO release
To measure NO release following NO-charging each tracheal catheter was cut into 1-cm sections, and each dialysis, biliary and urinary catheter was cut into 2-cm sections. Each of the NO-charged and control catheter sections was placed in a tube and was immersed in sterile water. Tracheal catheter sections in 2.5ml, dialysis and biliary catheter sections in 4ml and urinary catheter sections in 2.25ml. Nitrite concentration in water was measured over 24 hours following immersion using Griess reagent [22]. Nitrite concentration (μM) was converted to NO concentration (ppm) as described elsewhere [13].
Bacterial preparation
Pseudomonas aeruginosa, E. coli, Enterococcus faecalis and Candida albicans were obtained from American Type Culture Collection (ATCC #14210, #25922, #29212 and #90028 respectively). P. aeruginosa were grown in Muller Hinton Broth (MHB) to a concentration of 1×108 CFU/ml and diluted with MHB to a working concentration of 102 CFU/ml. E. coli were grown in Luria Broth (LB) to a concentration of 2.5×108 CFU/ml and diluted with LB to working concentrations of 102 and 103 CFU/ml. E. faecalis and C. albicans were plated onto blood agar and sabouraud agar plates and incubated at 37°C for 24 and 48 hours, respectively. After incubation, E. faecalis and C. albicans colonies were taken and diluted with saline to a concentration of 1×108 or 1×105 (McFarland 0.5), respectively. Samples were further diluted with saline (2% LB) to a working concentration of 102 CFU/ml.
Microbial attachment and colonization onto catheters surfaces
2-cm sections of NO-charged and control catheters were immersed in suspensions contaminated with microbial cells; tracheal catheter sections were immersed in 1ml of MHB containing 102 CFU/ml of P. aeruginosa and incubated for 48-hours at 30°C. Dialysis and biliary catheter sections were immersed in 1.5ml of LB, containing 102, or 103 CFU/ml of E. coli and incubated for 24-hour at 37°C. Urinary catheter sections were immersed in 2.25ml of saline (2% LB) containing 102 CFU/ml of E.coli, C. albicans or E. faecalis and incubated for 24 hours at 37°C. Following incubation, each of the catheter sections was washed twice with sterile saline (0.9%). Then each of the catheter sections was qualitatively assessed for microbial attachment following rolling of the section onto Muller Hinton (MH) agar plates, LB agar plates, MacConkey agar plates or blood agar plates. Quantitative assessment for the number of colonizing microbial cells onto 1-cm of urinary catheter sections was performed after 24 hours of incubation with contaminated suspensions where sections were transferred into new tubes containing fresh saline solution and sonicated as described elsewhere [13].
Planktonic anti-microbial activity of NO-charged catheters
2-cm sections of NO-charged and control tracheal and urinary catheters were immersed in 1ml of MHB containing 102 CFU/ml of P. aeruginosa or 2.25ml of saline (2% LB) containing 102 CFU/ml of C. albicans or E. faecalis, respectively. Tracheal catheters were incubated for 48 hours at 30°C, and urinary catheters were incubated for 24 hours at 37°C. After incubation, samples were vortexed and aliquots were plated onto MH agar plates, MacConkey agar plates or blood agar plates and incubated overnight at 370 C. CFU's were counted and final bacterial load was calculated as CFU/ml.
Clinical study population
Patients undergoing radical prostatectomy surgery at the Urological Department, Rabin Medical Center (RMC) were selected by the principal investigator and screened for their eligibility. Inclusion criteria: 1) Male patients (>18yr). 2) Life expectancy of more than 12 months. 3) Patients scheduled for radical prostatectomy surgery. 4) The patient is willing and able to read, understand and sign the study specific informed consent form. Exclusion criteria: 1) any underlying disease involving the heart, lungs, skin, immunodeficiency or infection that could influence study results. 2) A urinary culture demonstrating UTI before radical prostatectomy surgery. 3) Known urethral stricture and a history of recurrent UTI's. Patients who met all inclusion criteria and none of the exclusion criteria were invited to participate in the study.
Ethics
This clinical study was conducted according to the laws, regulations and administrative provisions relating to the implementation of Good Clinical Practice (GCP) in the conduct of clinical trials on medicinal products for human use, as applicable by national legislation and according to the Israeli Ministry of Health (MOH) regulations. The clinical study was approved by the Investigational Review Board (IRB) of RMC and by the Israeli MOH. All patients eligible for participation in this study provided a written informed consent prior to enrollment.
Clinical study design
A prospective, phase I, open-label, case controlled single-center study performed at the Urology Department, Beilinson Hospital, RMC, Israel (registry name at ClinicalTrials.gov- NCT02277171) (Fig 1). Out of 15 patients enrolled (the number of enrolled patients was extended from 12 to 15 patients during the study, since 3 patients enrolled to the study were withdrawn prior to catheterization during radical prostatectomy surgery), 12 patients were catheterized with an 18Fr diameter silicone foley catheter for at least 1 day and were included in the study in the Intention To Treat (ITT) cohort. Six patients were catheterized with an NO-charged catheter (NO-group) during radical prostatectomy surgery. After all 6 patients from the NO-group completed the study, another group of 6 patients were catheterized during radical prostatectomy surgery with a regular non-charged control catheter (CT-group). Each patient was catheterized for 7–28 days, and was followed for 30–45 days after catheterization (catheterization time was extended during the study from 14 days to up to 28 days, and follow-up period was extended from 30 to up to 45 days after catheterization at the request of the urological department physicians). All catheterizations were performed by a physician. Patients from both groups received standard treatment in the hospitals urology department and in the outpatient clinic including anti-coagulation treatment prior to surgery and antibiotics during hospitalization. All deviations from protocol were approved by the IRB of RMC. Protocol was amended accordingly during the study and approved by IRB of RMC.
Safety assessments
Number and proportion of possible AE's or SAE's that may be associated with NO-charged urinary catheters including decreased in blood pressure, inhibition of platelet aggregation, increase in bleeding time, and local side effects such as unwanted erection (i.e. priapism) and local skin irritation were assessed in each patient from the NO-group throughout the study and were compared to the CT-group.
During the study vital signs (i.e. Blood pressure, Temperature) and coagulation (i.e. Prothrombin Time–PT, activated Partial Thromboplastin Time–aPTT, International Normalized Ratio–INR), Hematology (i.e. Platelets, White Blood Cells–WBC, hemoglobin and Neutrophils) and biochemistry (i.e. C-reactive) tests that may be in association with NO-releasing urinary catheter were taken prior to catheterization at screening for baseline evaluation, at hospitalization (daily), at catheter removal day and at follow-up visit. A comparison of parameters values between the groups and changes in parameters values from baseline levels during the study within each study group were evaluated and compared at the 1st day of catheterization (i.e. 24 hours after catheter insertion), at hospital discharge day, at catheter removal day and at follow-up visit. All lab personal were blinded to the study groups. All data for each individual patient was collected in paper Case Report Form (CRF). One copy of each CRF was kept by the investigator and the 2 copies were retrieved by the sponsor.
Statistical analysis
Fisher's exact test was used for analyzing the difference in proportions between the NO-group and CT-group. The Paired T-test for two means was applied for analyzing changes in parameters within a study group. The Non-parametric Mann-Whitney U-test was applied for analyzing differences in parameters between the NO-group and CT-group.
All data analysis was conducted on the ITT cohort. Last Observation Carried Forward (LOCF) approach was applied for all patients catheterized for at least 1 day (ITT cohort) to account for missing data at or prior to study termination.
Although statistical tests were conducted to compare between the groups and to analyze changes from baseline measurements at different time points during the study, the planned sample size was not expected to show statistical significance or statistical power, only demonstrate a safety profile. The data was analyzed using GraphPad Prism® version 5.0.
Results
Release of NO from charged catheters
NO has a short half-life in-vivo of a few seconds. Therefore, the level of more stable NO metabolite, nitrites, was used for indirect measurement of NO in fluids. As shown (Fig 2) NO release from tracheal, dialysis and biliary catheters was measured during 24 hours, with the largest amounts released during the first hours of immersion in water. After 24 hours the NO levels released per 1-cm of each of the NO-charged tracheal, dialysis and biliary catheters were 5.47 ppm, 6.69 ppm and 6.02 ppm, respectively. NO was not released from non-charged control sections (not shown).
NO-charged catheters prevent/reduce microbial attachment and colonization
To evaluate the ability of NO-charged catheters to prevent microbial attachment onto their surfaces, NO-charged and control catheter sections were exposed to suspensions contaminated with low concentration of different types of microbial cells; E. coli, C. albicans, E. faecalis and P. aeruginosa. Microbial attachment was evaluated after an incubation period of 24–48 hours. E. coli, C. albicans, E. faecalis and P. aeruginosa attached onto all corresponding control catheter sections. In contrast, E. coli, C. albicans and P. aeruginosa failed to attach onto NO-charged catheter sections, while E. faecalis attached onto NO-charged urinary catheter sections (Fig 3), albeit at lower affinity compared to non-charged catheters.
Quantitative evaluation of the number of colonizing microbial cells onto the surfaces of 1-cm sections of NO-charged and control urinary catheters was performed following an incubation period of 24 hours with E. coli, C. albicans and E. faecalis and sonication of the catheter sections in fresh saline solution to separate colonizing bacteria and fungi from their surfaces. As shown (Fig 4), E. coli and C. albicans colonization onto NO-charged urinary catheter sections was prevented; decreasing by approximately 5 logs and 2.5 logs compared to the control sections, respectively. E. faecalis colonization onto NO-charged urinary catheter sections decreased by 1 log compared to the control sections.
NO-charged catheters prevent planktonic microbial growth
Although NO has a small radius of action in solution we have showed that it possesses antimicrobial activity against E. coli in solution [13–14]. To evaluate whether NO-charged catheters possesses anti-microbial activity against other bacterial and fungi we exposed tracheal catheter sections to 102 CFU/ml of P. aeruginosa and urinary catheters to 102 CFU/ml of C. albicans and tested bacterial content in suspensions after 24–48 hours of incubation. P. aeruginosa suspensions incubated with control catheter sections had 3.09×108 CFU/ml and 5.6×108 CFU/ml after 24 hours and 48 hours of incubation, respectively, while control C. albicans suspensions had 5.17×102 CFU/ml after 24-hour incubation. In contrast, P. aeruginosa and C. albicans suspensions incubated with NO-charged catheters had 0 CFU/ml after 48 and 24 hours, respectively (Fig 5). Similarly, immersion of NO-charged dialysis and tracheal catheter sections in suspension containing 102 CFU/ml of E. coli resulted in complete eradication of planktonic bacteria after 24 hours, while suspensions incubated with control catheters had up to 108 CFU/ml of E. coli (data not shown).
Safety and tolerability of NO-charged urinary catheters–phase 1 clinical study
Proportion of patients who completed the study and NO-related adverse events
Patients were enrolled to this study from the 22nd of December, 2014 to the 30th of March, 2015. Study was completed at 12th of May, 2015 (end of follow-up period of the last patient). All patients from the NO-group successfully completed the study per protocol (Fig 1), without experiencing NO-related AE's (Table 1), while one patient from the CT-group failed to complete the study per protocol due to an SAE.
Table 1. AE's and SAE's recorded during the study.
CT-Group | NO-Group | ||||
---|---|---|---|---|---|
N | (%) | N | (%) | P Value | |
No. of Patient | 6 | 6 | |||
No. of Patients with AE's | 2 | (33%) | 2 | (33%) | 1.000 |
No. of Patients with SAE's | 1 | (16.66%) | 0 | (0%) | 1.000 |
AE's associated with NO catheter | N/A | 0 | (0%) | ||
SAE's associated with NO catheter | N/A | 0 | (0%) | ||
AE's (Severity) | |||||
Hematuria | 1 (Mild) | (16.66%) | 1 (Mild) | (16.66%) | 1.000 |
Bleeding | 1 (Severe) | (16.66%) | 0 | (0%) | 1.000 |
Hyperkalemia | 0 | (0%) | 1(Mild) | (16.66%) | 1.000 |
SAE's | |||||
Myocardial Infract | 1 | (16.66%) | 0 | (0%) | 1.000 |
Number and proportion of AE's and SAE's recorded during the study. Fisher's exact test was used for analyzing the difference in the number and the proportions of AE's and SAE's between both groups. p > 0.05, non-statistically significant difference. p < 0.05, statistically significant difference.
Physical examination
Since NO is associated with sustained penile erection and may cause local skin irritation and the formation of local skin edema and erythema to its surrounding tissue [15–20], a special emphasis was put on the genitourinary system in particular to the penis, scrotum or groin area during physical examinations performed at baseline, on a daily basis during hospitalization, at catheter removal day and at follow-up visit. No priapism or local irritation to the penis, scrotum or groin, were detected in both groups throughout the study. No other physical changes in the respiratory, cardiovascular or head, ears, eyes, nose and throat (HEENT) systems were observed in all patients from the NO-group, while the physical condition of one patient from the CT-group worsened during hospitalization due to myocardial infarct.
Vital signs and clinical laboratory evaluation
Vital signs measurements and biochemistry, hematology and coagulation tests were taken throughout the study and their values were compared between the CT-group and NO-group at the following time points: at baseline, at 1st day of catheterization, at hospital discharge day, at catheter removal day and at follow-up visit. In addition, the change of each parameter value from baseline at each time point was assessed within each particular group.
Baseline results of parameters that may be associated with NO and their corresponding normal ranges are presented (Table 2). As shown, no statistically significant differences were detected at baseline between the groups, and all tests results of patients from both groups were within the accepted normal ranges.
Table 2. Patient Characteristics and baseline measurements.
Characteristic | CT-Group | NO-Group | P Value | |
No. of patients | 6 | 6 | ||
Age (Yr) | 67 ± 3.39 | 68.33 ± 1.72 | 1.000 | |
NO-related Parameters | Mean±SEM | Mean±SEM | P Value | Normal range |
Blood Pressure (mmHg) | 101.9 ± 7.79 | 95.26 ± 1.39 | 0.393 | 80–120 |
Temperature (°C) | 36.75 ± 0.11 | 36.68 ± 0.060 | 0.685 | 36.5–37.2 |
PT (Sec) | 11.33 ± 0.30 | 10.63 ± 0.21 | 0.169 | 9.5–13.5 |
aPTT (Sec) | 32.65 ± 0.89 | 32.24 ± 1.62 | 0.792 | 24–40 |
INR | 0.92 ± 0.02 | 0.88 ± 0.01 | 0.146 | 0.8–1.2 |
Platelets (103/μl) | 297.50 ± 81.69 | 213.20 ± 12.21 | 0.748 | 130–400 |
Hemoglobin (g/dl) | 14.33 ± 0.53 | 14.83 ± 0.23 | 0.484 | 14–18 |
WBC (103/μl) | 6.51 ± 0.5 | 9.31 ± 1.28 | 0.093 | 4.8–10.8 |
Neutrophils absolute (103/μl) | 4.00 ± 0.5 | 4.08 ± 0.37 | 0.935 | 1.9–8.0 |
CRP (mg/dl) | 0.31 ± 0.10 | 0.25 ± 0.12 | 0.753 | 0–0.5 |
Baseline levels vital signs and coagulation, hematological, and biochemical parameters that may be associated with NO. All tests were taken prior to radical prostatectomy surgery. Differences in parameters values were compared between the Control group [CT-group] and the Nitric Oxide group [NO-group] using The Non-parametric Mann-Whitney U-test. Accepted normal ranges are presented Abbreviations: PT–Prothrombin Time; aPTT–Activated Partial Thromboplastin Time; INR–International Normalized Ratio; WBC–White Blood Cells; CRP–C—reactive protein. p > 0.05, non-statistically significant difference. p < 0.05, statistically significant difference.
Comparison results (CT vs NO group) of parameters that may be associated with NO throughout the study are presented at Table 3. Changes from baseline levels of these parameters within each group throughout the study are presented at Table 4.
Table 3. Comparison of parameters associated with NO-releasing catheter–CT versus NO Group.
CT-Group | NO-Group | ||||
---|---|---|---|---|---|
Parameters | N | Mean±SEM | N | Mean±SEM | P Value |
Blood Pressure (mmHg) | |||||
1st day of Catheterization | 6 | 96.47±6.206 | 6 | 90.68± 4.053 | 0.818 |
Hospital Discharge | 6 | 104.40±2.861 | 6 | 99.70±3.553 | 0.394 |
Catheter Removal | 6 | 106.8±5.455 | 6 | 102.40±5.495 | 0.521 |
Follow-up | 6 | 106±3.801 | 6 | 98.08± 3.595 | 0.261 |
Temperature (°C) | |||||
1st day of Catheterization | 6 | 36.86±0.086 | 6 | 36.81±0.164 | 0.470 |
Hospital Discharge | 6 | 36.87±0.096 | 6 | 36.99±0.095 | 0.375 |
Catheter Removal | 6 | 36.87±0.143 | 6 | 36.75±0.056 | 0.326 |
Follow-up | 6 | 36.88±0.130 | 6 | 36.77±0.080 | 0.470 |
PT (Sec) | |||||
1st day of Catheterization | 5 | 12.22±0.404 | 6 | 11.98±0.289 | 0.647 |
Hospital Discharge | 6 | 11.98±0.547 | 6 | 11.40± 0.330 | 0.589 |
Catheter Removal | 4 | 12.38±0.914 | 6 | 11.18±0.236 | 0.476 |
Follow-up | 5 | 11.62±0.723 | 6 | 11.08±0.199 | 1.000 |
Activated PTT (Sec) | |||||
1st day of Catheterization | 5 | 28.42±2.321 | 6 | 30.42±1.021 | 0.464 |
Hospital Discharge | 6 | 28.98±2.114 | 6 | 32.28±1.031 | 0.240 |
Catheter Removal | 4 | 30.70±0.930 | 6 | 34.15±1.411 | 0.114 |
Follow-up | 5 | 31.30± 1.123 | 6 | 33.63±1.476 | 0.329 |
INR | |||||
1st day of Catheterization | 5 | 0.99±0.032 | 6 | 0.99±0.020 | 0.855 |
Hospital Discharge | 6 | 0.97±0.044 | 6 | 0.94±0.024 | 1.000 |
Catheter Removal | 4 | 1.00±0.074 | 6 | 0.92±0.025 | 0.516 |
Follow-up | 5 | 0.94±0.059 | 6 | 0.89±0.016 | 0.926 |
Platelets (103/μl) | |||||
1st day of Catheterization | 6 | 273.00±105.00 | 6 | 176.30±9.982 | 0.699 |
Hospital Discharge | 6 | 273.50±90.330 | 6 | 193.50±13.40 | 0.818 |
Catheter Removal | 6 | 352.70±74.450 | 6 | 321.30± 20.480 | 0.699 |
Follow-up | 6 | 305.30±86.530 | 6 | 224.70±24.120 | 0.937 |
Hemoglobin (g/dl) | |||||
1st day of Catheterization | 6 | 11.95±0.532 | 6 | 13.23±0.161 | 0.092 |
Hospital Discharge | 6 | 12.72±0.711 | 6 | 13.37±0.437 | 0.521 |
Catheter Removal | 6 | 13.17±0.761 | 6 | 14.47±0.531 | 0.394 |
Follow-up | 6 | 13.28±0.789 | 6 | 14.43±0.438 | 0.240 |
WBC (103/μl) | |||||
1st day of Catheterization | 6 | 8.34±1.003 | 6 | 7.74±1.159 | 0.699 |
Hospital Discharge | 6 | 8.62±0.796 | 6 | 7.438±1.038 | 0.394 |
Catheter Removal | 6 | 8.09± 0.530 | 6 | 9.89±1.164 | 0.309 |
Follow-up | 6 | 6.75±0.755 | 6 | 9.41±1.099 | 0.093 |
Neutrophils (103/μl) | |||||
1st day of Catheterization | 6 | 6.50±0.800 | 6 | 6.9±1.248 | 0.937 |
Hospital Discharge | 6 | 6.45±0.712 | 6 | 5.567±0.382 | 0.394 |
Catheter Removal | 6 | 5.68±0.711 | 6 | 6.517±0.624 | 0.394 |
Follow-up | 6 | 4.63±0.847 | 6 | 5.017±0.240 | 0.297 |
CRP (mg/dl) | |||||
1st day of Catheterization | 5 | 2.75±0.951 | 6 | 3.57±0.551 | 0.430 |
Hospital Discharge | 6 | 7.86±3.797 | 6 | 6.96±1.003 | 0.485 |
Catheter Removal | 6 | 4.82±3.838 | 6 | 0.87±0.251 | 0.818 |
Follow-up | 6 | 4.65±3.865 | 6 | 0.42±0.154 | 0.240 |
Comparison of parameters values between the CT-group and the NO-group at 1st day of catheterization, at hospital discharge, at catheter removal day and at follow-up visit. Non-parametric Mann-Whitney U-test for independent samples was applied for analyzing differences in parameters between the NO-group and CT-group at each time point. p > 0.05, non-statistically significant difference. p < 0.05, statistically significant difference.
Table 4. Parameters associated with NO-releasing catheter–change from baseline within groups.
CT-Group | NO-Group | |||||||
---|---|---|---|---|---|---|---|---|
Parameters | N | Mean±SEM | % Change From Baseline | P value | N | Mean±SEM | % Change From Baseline | P value |
Blood Pressure (mmHg) | ||||||||
Baseline | 6 | 101.90±7.79 | 6 | 95.26±1.39 | ||||
1st day of Catheterization | 6 | 96.47±6.20 | -5.35 | 0.598 | 6 | 90.68±4.05 | -4.80 | 0.349 |
Hospital Discharge | 6 | 104.40±2.86 | 2.40 | 0.709 | 6 | 99.70±3.55 | 4.66 | 0.265 |
Catheter Removal | 6 | 106.8±5.45 | 4.74 | 0.364 | 6 | 102.40±5.49 | 7.44 | 0.317 |
Follow-Up | 6 | 106±3.801 | 4.04 | 0.514 | 6 | 98.08±3.595 | 2.97 | 0.564 |
Temperature (°C) | ||||||||
Baseline | 6 | 36.75±0.11 | 6 | 36.68±0.06 | ||||
1st day of Catheterization | 6 | 36.86±0.09 | 0.29 | 0.405 | 6 | 36.81±0.16 | 0.35 | 0.46 |
Hospital Discharge | 6 | 36.87±0.10 | 0.33 | 0.296 | 6 | 36.99±0.09 | 0.35 | 0.061 |
Catheter Removal | 6 | 36.87±0.14 | 0.32 | 0.598 | 6 | 36.75±0.06 | 0.19 | 0.394 |
Follow-Up | 6 | 36.88±0.130 | 0.36 | 0.299 | 6 | 36.77±0.080 | 0.23 | 0.434 |
PT (Sec) | ||||||||
Baseline | 6 | 11.33±0.307 | 6 | 10.63±0.219 | ||||
1st day of Catheterization | 5 | 12.22±0.404 | 9.18 | 0.019 | 6 | 11.98±0.289 | 12.70 | < 0.001 |
Hospital Discharge | 6 | 11.98±0.547 | 5.74 | 0.160 | 6 | 11.40±0.330 | 7.15 | 0.005 |
Catheter Removal | 4 | 12.38±0.914 | 6.80 | 0.292 | 6 | 11.18±0.236 | 5.17 | 0.005 |
Follow-Up | 5 | 11.62±0.723 | 3.88 | 0.402 | 6 | 11.08±0.199 | 4.23 | 0.048 |
aPTT (Sec) | ||||||||
Baseline | 6 | 32.65±0.896 | 5 | 32.24±1.627 | ||||
1st day of Catheterization | 5 | 28.42±2.321 | -15.19 | 0.083 | 6 | 30.42±1.021 | -6.33 | 0.035 |
Hospital Discharge | 6 | 28.98±2.114 | -11.21 | 0.150 | 6 | 32.28±1.031 | -0.06 | 0.988 |
Catheter Removal | 4 | 30.70±0.930 | -5.73 | 0.048 | 6 | 34.15±1.411 | 5.65 | 0.088 |
Follow-Up | 5 | 31.30±1.123 | -2.14 | 0.441 | 6 | 33.63±1.476 | 4.96 | 0.013 |
INR | ||||||||
Baseline | 6 | 0.92±0.022 | 6 | 0.88±0.011 | ||||
1st day of Catheterization | 5 | 0.99±0.032 | 7.61 | 0.055 | 6 | 0.99±0.020 | 12.50 | < 0.001 |
Hospital Discharge | 6 | 0.97±0.044 | 4.35 | 0.233 | 6 | 0.94±0.024 | 6.82 | 0.008 |
Catheter Removal | 4 | 1.00±0.074 | 6.52 | 0.309 | 6 | 0.92±0.025 | 4.55 | 0.101 |
Follow-Up | 5 | 0.94±0.059 | 2.17 | 0.603 | 6 | 0.89±0.016 | 2.27 | 0.233 |
Hemoglobin (g/dl) | ||||||||
Baseline | 6 | 14.33±0.539 | 6 | 14.83±0.229 | ||||
1st day of Catheterization | 6 | 11.95±0.532 | -16.61 | < 0.001 | 6 | 13.23±0.161 | -10.79 | 0.002 |
Hospital Discharge | 6 | 12.72±0.711 | -11.30 | 0.035 | 6 | 13.37±0.437 | -9.91 | 0.016 |
Catheter Removal | 6 | 13.17±0.766 | -8.14 | 0.038 | 6 | 14.47±0.531 | -2.49 | 0.517 |
Follow-Up | 6 | 13.28±0.789 | -7.33 | 0.075 | 6 | 14.43±0.438 | -2.70 | 0.347 |
WBC (103/μl) | ||||||||
Baseline | 6 | 6.51±0.497 | 6 | 9.307±1.283 | ||||
1st day of Catheterization | 6 | 8.34±1.003 | 28.11 | 0.042 | 6 | 7.74± 1.159 | 21.06 | 0.096 |
Hospital Discharge | 6 | 8.62±0.796 | 32.41 | 0.093 | 6 | 7.438±1.038 | 7.95 | 0.416 |
Catheter Removal | 6 | 8.09±0.530 | 24.27 | 0.125 | 6 | 9.89±1.164 | 21.06 | 0.091 |
Follow-Up | 6 | 6.75±0.755 | 3.69 | 0.843 | 6 | 9.41±1.099 | 5.37 | 0.548 |
Neutrophils (103/μl) | ||||||||
Baseline | 6 | 4.00±0.500 | 6 | 4.08± 0.368 | ||||
1st day of Catheterization | 6 | 6.50±0.800 | 62.50 | 0.015 | 6 | 6.90±1.248 | 69.12 | 0.041 |
Hospital Discharge | 6 | 6.45±0.712 | 61.25 | 0.074 | 6 | 5.57±0.382 | 35.05 | 0.053 |
Catheter Removal | 6 | 5.68±0.711 | 42.00 | 0.127 | 6 | 6.52±0.624 | 59.56 | 0.025 |
Follow-Up | 6 | 4.63±0.847 | 15.75 | 0.599 | 6 | 5.02±0.240 | 22.79 | 0.084 |
Platelets (103/μl) | ||||||||
Baseline | 6 | 297.5±81.7 | 6 | 213.2±12.21 | ||||
1st day of Catheterization | 6 | 273.0±105.0 | -8.24 | 0.416 | 6 | 176.3±9.982 | -17.27 | 0.005 |
Hospital Discharge | 6 | 273.5±90.3 | -8.07 | 0.179 | 6 | 193.50±13.4 | -9.23 | 0.067 |
Catheter Removal | 6 | 352.7±74.40 | 18.54 | 0.001 | 6 | 321.3±20.48 | 50.75 | 0.001 |
Follow-Up | 6 | 305.3±86.50 | 2.63 | 0.679 | 6 | 224.7±24.12 | 5.39 | 0.619 |
CRP (mg/dl) | ||||||||
Baseline | 5 | 0.31±0.100 | 5 | 0.25±0.120 | ||||
1st day of Catheterization | 5 | 2.75±0.951 | 867.74 | 0.084 | 6 | 3.57±0.551 | 1236.00 | 0.005 |
Hospital Discharge | 6 | 7.86±3.797 | 2841.94 | 0.114 | 6 | 6.96±1.003 | 2684.00 | 0.005 |
Catheter Removal | 6 | 4.82±3.838 | 1748.39 | 0.304 | 6 | 0.87±0.251 | 292.00 | 0.047 |
Follow-Up | 6 | 4.65±3.865 | 1693.55 | 0.321 | 6 | 0.42±0.154 | 88.00 | 0.205 |
Changes in parameters values from baseline levels during the study within both study groups were evaluated at the 1st day of catheterization, at hospital discharge, at catheter removal day and at follow-up visit. Paired T-test was applied for analyzing changes from baseline levels. p > 0.05, non-statistically significant change from baseline. p < 0.05, statistically significant change from baseline.
No statistically significant differences in all parameters associated with NO were detected between the study groups throughout the study (Table 3).
No statistically significant changes from baseline in blood pressure or temperature were observed within each study group throughout the study (Table 4). Both CT and NO groups exhibited an increase from baseline levels in the coagulation parameters PT and INR, and a decrease in aPTT during hospitalization. However, in the NO-group the change from baseline of these parameters was statistically significant. In both groups the levels of all coagulation parameters were between the normal ranges.
Hemoglobin levels significantly decreased from baseline in both groups during hospitalization, where in the CT-group a significant decrease was detected also at catheter removal day. The mean values of hemoglobin in the CT-group during hospitalization and follow-up visit were lower than the normal ranges by 16.7% and 7.3%, respectively.
WBC and neutrophils significantly increased from baseline by 28.1% and 62.5%, in the CT-group at the 1st day of catheterization, respectively. Similarly, an increase in WBC and neutrophils was also measured in the NO-group at the same corresponding time points. However, this increase was not statistically significant.
Platelets level decreased significantly from baseline levels by 17.3% in the NO-group at the 1st day of catheterization. In contrast, at catheter removal day a significant increase from baseline levels was detected in both groups.
CRP levels significantly increased in the NO-group by 1236%, 2684%, and 292% in comparison to baseline levels during the 1st day of catheterization, Hospital discharge, and catheter removal, respectively. CRP levels of the CT-group also increased during the same periods albeit without reaching statistical significance.
Excluding hemoglobin measurements, the mean values of all parameters during hospitalization, catheter removal and follow-up visits were within the normal ranges.
No significant differences in electrolytes, metabolites, renal function and liver function parameters were detected between the study groups (S1 Table). Changes from baseline levels were detected in several parameters in both groups, however all parameters at all time points were between the normal ranges (S2 Table).
Validation of NO release from NO-charged urinary catheters
In order to confirm that the catheters used in the clinical study release NO, for each NO-charged catheter delivered for catheterization, an internal control measurement was taken to assess the levels of NO released after 24-hour exposure to fluids. NO was released from all urinary catheters at a concentration of between 1.75–2.8 ppm per catheter after 24 hours.
Discussion
In this study we present pre-clinical data of the anti-infective properties of several types of NO-charged catheters in-vitro, and safety data from our phase 1 clinical study for the evaluation of the safety and tolerability of NO-charged urinary catheters in patients catheterized for 7–28 days.
In this study we chose to focus on the preventative capabilities of NO-charged catheters. Therefore the anti-infective properties of NO-charged catheters were tested against low concentration of bacteria and fungi which may attach onto the inner and outer surfaces of the catheters in small numbers during the first hours of catheterization. As shown, NO-charged catheters prevent gram-negative bacteria (E. coli and P. aeruginosa) and fungus (C. albicans) from attaching and colonizing onto their surfaces and exhibit some anti-microbial properties by eradicating small population of gram-negative bacteria and fungi in the surrounding media. This effect was more moderate in NO-charged urinary catheters tested against gram-positive bacteria (E. faecalis) where a 1 log reduction of in the number of colonizing bacteria was observed. In this case release of NO from the catheter had no effect on bacterial survival at the surrounding media.
Prevention/reduction of the colonization abilities of the microbial strains tested in this study are of clinical importance as biofilm-forming microbial cells are known to be play a significant role in the pathogenesis of catheter-associated HAI's and especially in UTI's and BSI's [23–25]. Most importantly, by reducing HAI's incidents cause by microbial pathogens it is reasonable to assume that NO-based charging technology may also reduce the amount of antibiotics given to patients, thus easing some of the concerns regarding excessive usage of antibiotics and the emergence of antibiotic-resistance microbial strains.
The anti-infective capabilities of NO-charged catheters demonstrated here were achieved against low concentrations of bacteria under a wide-range of NO concentrations released during the first few hours after immersion of the catheters in fluids. Although one may speculate that the anti-microbial protection exhibited here for tracheal, dialysis and biliary catheters may not last during the entire catheterization period since minimal levels of NO are released from these catheters after 24 hours, we believe that their ability to prevent early microbial colonization is an important step in preventing/reducing biofilm formation onto their surfaces and that this characteristic may have a dramatic impact on microbial susceptibility to various anti-microbial agents. Currently, we are working on increasing the duration of NO release from tracheal, dialysis and biliary catheters as already achieved for urinary catheters which release NO for a period of 14 days [13], so that they will better serve this important potential attribute.
The in-vitro data presented here were collected from three different experiments (Tracheal catheter study, chronic dialysis and biliary catheter, and urinary catheter study) that used several types of microbial cells and performed by different lab personal. The results presented in this study are in-line with data presented by Regev-Shoshani et al, which studied the NO-release profile from charged urinary catheters in urine and water, and their ability to prevent E. coli colonization and to elicit anti-microbial conditions in-vitro when exposed to low concentration of bacteria [13–14]. Taken together, NO-charged catheters characteristics demonstrated in this study and the consistency in the efficacy results supports the assumption that NO-charging technology might be effective in preventing/reducing catheter-associated HAI's and should be tested in clinical trials.
According to recent reports, UTI is a most common HAI; with Seventy-five percent of hospital-acquired UTI's are associated with catheterization [2]. Therefore, we chose to proceed with clinical development of NO-charged urinary catheters and test their safety and tolerability in catheterized patients.
NO-charged urinary catheters were found safe and well tolerated and met all primary endpoints; all patients catheterized with NO-charged urinary catheters successfully completed the study per protocol without experiencing AE's or SAE's associated with NO-charged catheters.
All physiological and coagulation (i.e. PT, aPTT and INR), hematological (i.e. Platelets, Hemoglobin, WBC, and Neutrophils) and biochemical (i.e. CRP) parameters that may be associated with NO-releasing catheters were compared between the groups throughout the study, and revealed no statistically significant differences.
Parameters values of CRP, PT, platelets, hemoglobin and WBC were changed in comparison to their baseline levels during the study. It seems that these changes are related to the radical prostatectomy surgery and to the medication given to the patients during the study, rather than to the inserted catheters. Indeed, post-operative elevation of CRP levels during hospitalization which is common after radical prostatectomy surgery [26], gradually decreased in both groups as time from catheterization proceeded. In addition, a gradual improvement of hemoglobin levels was also observed following an initial decrease observed during catheterization in both groups.
The decrease in blood coagulation parameters observed in both groups is primarily related to the anti-coagulative medication given to all patients prior to surgery to prevent venous thromboembolism, since all coagulation parameters improved in both groups as time from catheterization proceeded and when anti-coagulative medications were removed. Although, the decrease was markedly more evident in the NO-group, we assume that the effect of NO on blood coagulation is not clinically important since all parameters were between normal ranges at all time points, even though anti-coagulation medication was given. In addition, no bleeding episodes were documented in the NO-group throughout the study.
The safety data of NO-charged urinary catheters was expected since NO-charged urinary catheters release low concentration of NO. Although these concentrations are significantly lower (∼20-folds) than the NO levels approved by the Food and Drug administration (FDA) to treat infants suffering from persistent pulmonary hypertension of the newborn (PPHN) [27–28], they were found to elicit anti-microbial conditions in this and in previous studies [13–14].
Conclusions
The pre-clinical data and safety profile of NO-charged urinary catheters in humans confirm that these catheters can be considered safe for future efficacy studies and places NO-based charging technology as a potential platform for the prevention of catheter-associated HAI's.
Supporting information
Acknowledgments
We wish to thank Dan Sadeh, Kibbutz Beit-Alfa, Israel, for his technical assistance with charging the urinary catheters with Nitric Oxide and with the writing of the clinical trial protocol.
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
The study was fully funded by ENOX Ltd. Some of the authors hold positions at ENOX and took part in the study design, data collection, data analysis, and preparation of this manuscript.
References
- 1.World Health Organization (Who). Report on the Burden of Endemic HealthCare-Associated Infection Worldwide. WHO Libr Cat Data 2011:40. http://whqlibdoc.who.int/publications/2011/9789241501507_eng.pdf.
- 2.Center for Disease Control and Prevention. Healthcare-associated Infections (HAI's). http://www.cdc.gov/HAI/infectionTypes.html.
- 3.Koenig SM, Truwit JD. Ventilator-Associated Pneumonia: Diagnosis, Treatment, and Prevention. Clin Microbiol Rev 2006;19:637–57. doi: 10.1128/CMR.00051-05 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Costerton JW, Stewart PS, Greenberg EP. Bacterial biofilms: a common cause of persistent infections. Science 1999;284:1318–22. [DOI] [PubMed] [Google Scholar]
- 5.Dunne WM. Bacterial adhesion: seen any good biofilms lately? Clin Microbiol Rev 2002;15:155–66. doi: 10.1128/CMR.15.2.155-166.2002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Stoodley P, Sauer K, Davies DG, Costerton JW. Biofilms as complex differentiated communities. Annu Rev Microbiol 2002;56:187–209. doi: 10.1146/annurev.micro.56.012302.160705 [DOI] [PubMed] [Google Scholar]
- 7.Høiby N, Bjarnsholt T, Givskov M, Molin S, Ciofu O. Antibiotic resistance of bacterial biofilms. Int J Antimicrob Agents 2010;35:322–32. doi: 10.1016/j.ijantimicag.2009.12.011 [DOI] [PubMed] [Google Scholar]
- 8.Webb JS, Givskov M, Kjelleberg S. Bacterial biofilms: Prokaryotic adventures in multicellularity. Curr Opin Microbiol 2003;6:578–85. [DOI] [PubMed] [Google Scholar]
- 9.Schairer DO, Chouake JS, Nosanchuk JD, Friedman AJ. The potential of nitric oxide releasing therapies as antimicrobial agents. Virulence 2012;3:271–9. doi: 10.4161/viru.20328 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Hetrick EM, Shin JH, Paul HS, Schoenfisch MH. Anti-biofilm efficacy of nitric oxide-releasing silica nanoparticles. Biomaterials 2009;30:2782–9. doi: 10.1016/j.biomaterials.2009.01.052 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Worley BV, Slomberg DL, Schoenfisch MH. Nitric oxide-releasing quaternary ammonium-modified poly[amidoamine] dendrimers as dual action antibacterial agents. Bioconjug Chem 2014;25:918–27. doi: 10.1021/bc5000719 [DOI] [PubMed] [Google Scholar]
- 12.Worley BV, Schilly KM, Schoenfisch MH. Anti-Biofilm efficacy of dual-action nitric oxide-releasing alkyl chain modified poly[amidoamine] dendrimers. Mol Pharm 2015;12:1573–83. doi: 10.1021/acs.molpharmaceut.5b00006 [DOI] [PubMed] [Google Scholar]
- 13.Regev-Shoshani G, Ko M, Miller C, Av-Gay Y. Slow release of nitric oxide from charged catheters and its effect on biofilm formation by Escherichia coli. Antimicrob Agents Chemother 2010;54:273–9. doi: 10.1128/AAC.00511-09 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Regev-Shoshani G, Ko M, Crowe A, Av-Gay Y. Comparative efficacy of commercially available and emerging antimicrobial urinary catheters against bacteriuria caused by E. coli in vitro. Urology 2011;78:334–40. doi: 10.1016/j.urology.2011.02.063 [DOI] [PubMed] [Google Scholar]
- 15.Burnett AL. The role of nitric oxide in erectile dysfunction: implications for medical therapy. J Clin Hypertens 2006;8:53–62. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Paoloni JA, Appleyard RC, Nelson J, Murrell GA. Topical nitric oxide application in the treatment of chronic extensor tendinosis at the elbow: a randomized, double-blinded, placebo-controlled clinical trial. Am J Sports Med 2003;31:915–20. doi: 10.1177/03635465030310062901 [DOI] [PubMed] [Google Scholar]
- 17.Kermarrec N, Zunic P, Beloucif S, Benessiano J, Drouet L, Payen D. Impact of inhaled nitric oxide on platelet aggregation and fibrinolysis in rats with endotoxic lung injury. Role of cyclic guanosine 5’-monophosphate. Am J Respir Crit Care Med 1998;158:833–9. doi: 10.1164/ajrccm.158.3.9709097 [DOI] [PubMed] [Google Scholar]
- 18.Gries A, Herr A, Motsch J, et al. Randomized, placebo-controlled, blinded and cross-matched study on the antiplatelet effect of inhaled nitric oxide in healthy volunteers. Thromb Haemost 2000;83:309–15. [PubMed] [Google Scholar]
- 19.Gries a, Bode C, Peter K, et al. Inhaled nitric oxide inhibits human platelet aggregation, P-selectin expression, and fibrinogen binding in vitro and in vivo. Circulation 1998;97:1481–7. [DOI] [PubMed] [Google Scholar]
- 20.Hermann M, Flammer A, Lüscher TF. Nitric oxide in hypertension. J Clin Hypertens 2006;8:17–29. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Av-Gay Y, Miller C, Bohobot D, Regev-Shoshani G. Gaseous nitric oxide-sequestering products and processes of preparing same 2013; EP2571541 A2.
- 22.Green LC, Wagner DA, Glogowski J, Skipper PL, Wishnok JS, Tannenbaum SR. Analysis of nitrate, nitrite, and (15N) nitrate in biological fluids. Anal Biochem 1982;126:131–8. [DOI] [PubMed] [Google Scholar]
- 23.Nicolle LE. Catheter associated urinary tract infections. Antimicrob Resist Infect Control. 2014;3:23 doi: 10.1186/2047-2994-3-23 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Akbari F, Kjellerup BV. Elimination of bloodstream infections associated with candida albicans biofil in intravascular catheters. Pathogens. 2015;4:457–69. doi: 10.3390/pathogens4030457 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Olejnickova K, Hola V, Ruzicka F. Catheter- related infections caused by Pseudomonas aeruginosa: virulence factors involved and their relationships. Pathog Dis 2014;72:87–94. doi: 10.1111/2049-632X.12188 [DOI] [PubMed] [Google Scholar]
- 26.Wilbert DM, Schaerfe C, Strohmaier WL, Bichler KH. Evaluation of acute phase proteins for clinical monitoring of urinary tract infections and postoperatively. Investigative Urology 1991;4:81–85. [Google Scholar]
- 27.Davidson D, Barefield ES, Kattwinkel J, Dudell G, Damask M, Straube R, Rhines J, Chang CT. Inhaled nitric oxide for the early treatment of presistent pulmonary hypertension of the newborn: a randomized, double-masked, placebo-controlled, dose-response,multicenter study. The I-NO/PPHN Study Group. Pediatrics 1998;101:325–34. [DOI] [PubMed] [Google Scholar]
- 28.Wessel DL, Adatia I, Van Marter LJ, Thompson JE, Kane JW, Stark AR, Kourembanas S. Improved oxygenation in a randomized trial of inhaled nitric oxide for presistent pulmonary hypertension of the newborn. Pediatrics 1997;10:E7. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the paper and its Supporting Information files.